Acessibilidade / Reportar erro

Development and validation of an HPLC method for the simultaneous determination of artesunate and mefloquine hydrochloride in fixed-dose combination tablets

The present study developed and validated an HPLC method for the simultaneous determination of artesunate (AS) and mefloquine hydrochloride (MQ) in fixed-dose combination tablets, according to ICH guidelines. The chromatographic separation was carried out on an XBridge C18 (250 x 4.6 mm i.d., 5 µm particle size, Waters) analytical column. The mobile phase included a 0.05 M monobasic potassium phosphate buffer (pH adjusted to 3.0 with phosphoric acid) and acetonitrile (50 + 50, v/v). The flow rate was 1.0 mL/min, and the run time was 13 minutes. A dual-wavelength approach was employed: AS detection was performed at 210 nm and MQ was detected at 283 nm, using a diode array detector. Stability of sample solutions was evaluated for 8 hours after preparation, during which time the solutions remained stable. Youden's test was employed to evaluate robustness. The method proved to be linear (r²>0.99), precise (RSD<2.0%), accurate, selective, and robust, proving to be appropriate for routine drug quality control analysis.

Artesunate; Mefloquine hydrochloride; High performance liquid chromatography; Fixed-dose combination tablets


Universidade de São Paulo, Faculdade de Ciências Farmacêuticas Av. Prof. Lineu Prestes, n. 580, 05508-000 S. Paulo/SP Brasil, Tel.: (55 11) 3091-3824 - São Paulo - SP - Brazil
E-mail: bjps@usp.br