Acessibilidade / Reportar erro

Development and validation of a new and economical stability indicating RP-HPLC method for cefixime trihydrate

ABSTRACT

The present work describes the development of a new high performance liquid chromatographic (HPLC) method for the determination of Cefixime trihydrate under different stress conditons as specified by ICH. For the analysis, a Phenomenex (250 x 4.6 mm, 5 µm particle size) ODS column and a SPD 20 A UV detector at 289 nm was used. The selected mobile phase was 10 mM disodium hydrogen phosphate (with 0.5% TEA, pH adjusted to 6.3 with OPA) and methanol in the ratio of 75:25 (v/v) in isocratic mode at a flow rate of 1 mL.min-1.The linear regression analysis data for the calibration plots showed good linear relationship with r2 = 0.9997 in the concentration range of 5-100 μg.mL-1. The stress degradation was performed using acid, alkali, water, hydrogen peroxide and uv light.

Uniterms:
High Performance Liquid Chromatography/quantitative analysis; Degradation; Stability-indicating; Cefixime trihydrate/determination.

Universidade de São Paulo, Faculdade de Ciências Farmacêuticas Av. Prof. Lineu Prestes, n. 580, 05508-000 S. Paulo/SP Brasil, Tel.: (55 11) 3091-3824 - São Paulo - SP - Brazil
E-mail: bjps@usp.br