Single-Layered Paper-Based Microfluidic Devices Made by Paint-Spraying Technique with Great Barrier Resistance for Colorimetric Assays

Abstract

The single-layered paint-spraying paper based microfluidic devices (sp-μPADs) have been for the first time integrated with the colorimetric assay. The polymethyl methacrylate (PMMA) board was applied as the mask instead of the iron mask, reducing the power requirement of the cutting machine. The paint was sprayed into the paper, which blocked the pores of the paper and constructed the hydrophobic barrier on the paper. The paint barrier was highly resistant to the various kind of chemicals and the minimum channel dimensions was 1.4 mm. By using the PMMA masks with different designs, the sp-μPADs with different designs have been fabricated and successfully applied to the single analyte assay of iron or bovine serum albumin (BSA) and the simultaneous assay of copper and nickel. Moreover, the feasibility of the multiplexed assay of eight analytes was also demonstrated. The analytical performances indicated that the fabricated sp-μPADs offered a promising platform in the colorimetric assays of either single or multiple analytes.

Keywords:
microfluidics; paper-based devices; paint spraying; colorimetry


Introduction

Microfluidic paper-based devices (μPADs) have become an alternative point-of-care (POC) analytical tool since the work reported by Whitesides and co-workers in 2007.11 Martinez, A. W.; Phillips, S. T.; Butte, M. J.; Whitesides, G. M.; Angew. Chem., Int. Ed. 2007, 46, 1318. The μPADs possess the advantages of simplicity, portability, miniaturization, in-field capability and low-cost with a comparison to the traditional measurement methods.22 Channon, R. B.; Nguyen, M. P.; Scorzelli, A. G.; Henry, E. M.; Volckens, J.; Dandy, D. S.; Henry, C. S.; Lab Chip 2018, 18, 793. In recent years, μPADs have been integrated with various kind of analytical techniques, such as mass spectrometry (MS),33 Jiang, Y.; Wang, P. C.; Locascio, L. E.; Lee, C. S.; Anal. Chem. 2001, 73, 2048.,44 Lee, J.; Soper, S. A.; Murray, K. K.; J. Mass Spectrom. 2009, 44, 579. electrochemistry,55 Nie, Z.; Nijhuis, C. A.; Gong, J.; Chen, X.; Kumachev, A.; Martinez, A. W.; Narovlyansky, M.; Whitesides, G. M.; Lab Chip 2010, 10, 477.,66 Sun, X.; Wang, H.; Jian, Y.; Lan, F.; Zhang, L.; Liu, H.; Ge, S.; Yu, J.; Biosens. Bioelectron. 2018, 105, 218. surface plasmon resonance (SPR),77 Hassani, A.; Skorobogatiy, M.; Opt. Express 2006, 14, 11616.,88 Malic, L.; Veres, T.; Tabrizian, M.; Biosens. Bioelectron. 2009, 24, 2218. colorimety,99 Wang, L.; Musile, G.; McCord, B. R.; Electrophoresis 2018, 39, 470.

10 Ellerbee, A. K.; Phillips, S. T.; Siegel, A. C.; Mirica, K. A.; Martinez, A. W.; Striehl, P.; Jain, N.; Prentiss, M.; Whitesides, G. M.; Anal. Chem. 2009, 81, 8447.
-1111 Li, B.; Fu, L.; Zhang, W.; Feng, W.; Chen, L.; Electrophoresis 2014, 35, 1152. fluorometry,1212 Fiedoruk-Pogrebniak, M.; Granica, M.; Koncki, R.; Talanta 2018, 178, 31.

13 Weng, X.; Neethirajan, S.; Microchim. Acta 2017, 184, 4545.
-1414 Qi, J.; Li, B.; Wang, X.; Fu, L.; Luo, L.; Chen, L.; Anal. Chem. 2018, 90, 11827. electrophoresis,1515 Coltro, W. K. T.; de Jesus, D. P.; da Silva, J. A. F.; do Lago, C. L.; Carrilho, E.; Electrophoresis 2010, 31, 2487. chromatography,1616 Lisowski, P.; Zarzycki, P. K.; Chromatographia 2013, 76, 1201. and chemiluminescence.1717 Caputo, D.; de Cesare, G.; Dolci, L. S.; Mirasoli, M.; Nascetti, A.; Roda, A.; Scipinotti, R.; IEEE Sens. J. 2013, 13, 2595.,1818 Kamruzzaman, M.; Alam, A. M.; Kim, K. M.; Lee, S. H.; Kim, Y. H.; Kabir, A. H.; Kim, G. M.; Dang, T. D.; Biomed. Microdevices 2013, 15, 195. Among these techniques, the colorimetry integrated with μPADs was most attractive, in which color or light intensity variations are monitored by a mobile phone camera or a color image scanner, and analyzed by an open source image processing program.1919 Koesdjojo, M. T.; Pengpumkiat, S.; Wu, Y.; Boonloed, A.; Huynh, D.; Remcho, T. P.; Remcho, V. T.; J. Chem. Educ. 2015, 92, 737. The colorimetric μPADs have been used for rapid qualitative and quantitative analysis in various fields such as biochemical analysis, food safety, environmental monitoring, and point-of-care clinical diagnostics.2020 Rossini, E. L.; Milani, M. I.; Carrilho, E.; Pezza, L.; Pezza, H. R.; Anal. Chim. Acta 2018, 997, 16.

21 Chabaud, K. R.; Thomas, J. L.; Torres, M. N.; Oliveira, S.; McCord, B. R.; Forensic Chem. 2018, 9, 35.

22 Taghizadeh-Behbahani, M.; Hemmateenejad, B.; Shamsipur, M.; Chem. Pap. 2018, 72, 1239.

23 Liu, C.; Gomez, F. A.; Miao, Y.; Cui, P.; Lee, W.; Talanta 2019, 194, 171.
-2424 da Silva, G. O.; de Araujo, W. R.; Paixao, T. R.; Talanta 2018, 176, 674.

One of the major challenges in the fabrication of μPADs is to pattern the hydrophilic paper with hydrophobic materials to define a reaction zone for the analyte measurement.2525 Cao, R.; Zhang, X.; Tan, W.; Shen, W.; Cellulose 2018, 25, 583. Up to now, many techniques have been developed for the fabrication of μPADs. The photolithography technique, while expensive, produces devices with high resolution.2626 Asano, H.; Shiraishi, Y.; Anal. Chim. Acta 2015, 883, 55. This method requires the usage of the clean room facilities including UV light sources, expensive photoresists, and oxygen plasma. Polydimethylsiloxane (PDMS) plotting,2727 Abdollahi-Aghdam, A.; Majidi, M. R.; Omidi, Y.; BioImpacts 2018, 8, 237. screen-printing,2828 Puneeth, S. B.; Goel, S.; IEEE Trans. Electron Devices 2019, 66, 3196. wax-printing,2929 Chiang, C. K.; Kurniawan, A.; Kao, C. Y.; Wang, M. J.; Talanta 2019, 194, 837. ink-jet printing3030 Zhang, H.; Smith, E.; Zhang, W.; Zhou, A.; Biomed. Microdevices 2019, 21, 48. and laser cutting3131 Casto, L. D.; Schuster, J. A.; Neice, C. D.; Baker, C. A.; Anal. Methods 2018, 10, 3616. need the specialized instruments and the accompanying consumables. Other groups3232 Cardoso, T. M.; Garcia, P. T.; Coltro, W. K.; Anal. Methods 2015, 7, 7311.

33 Pena-Pereira, F.; Lavilla, I.; Bendicho, C.; Talanta 2016, 147, 390.
-3434 Atabakhsh, S.; Ashtiani, S. J.; Microfluid. Nanofluid. 2019, 23, 69. have developed some instrumental-free techniques including stamping and drawing with permanent marker pen or eyeliner pencil. However, up-scaling such a production technique has been proven to be difficult. Therefore, it is highly desirable to develop a new low-cost approach, allowing for the rapid prototyping and the large scale manufacture of μPADs.

Spraying technology has become a new prototyping process for the μPADs manufacture emerging in recent years. Compared with the classic wax-printing technique, the spraying technology avoided the printing process, representing a major breakthrough in the μPADs manufacture.3535 Al-Qadiri, H. M.; Ovissipour, M.; Al-Alami, N.; Govindan, B. N.; Shiroodi, S. G.; Rasco, B.; J. Food Sci. 2016, 81, M1177. Coltro and co-workers3636 Cardoso, T. M. G.; de Souza, F. R.; Garcia, P. T.; Rabelo, D.; Henry, C. S.; Coltro, W. K. T.; Anal. Chim. Acta 2017, 974, 63. described a glue-spraying technique for fabricating the μPADs and the hydrophobic barriers were created by the available scholar glue. In order to produce high resistance barrier, the paper coated with glue was dried at 130 ºC for 60 min and then exposed under a halogen spotlight for 30 min. This tedious heating step significantly increased the time for the μPADs fabrication. Chailapakul and co-workers3737 Nurak, T.; Praphairaksit, N.; Chailapakul, O.; Talanta 2013, 114, 291. proposed a spraying method with paint for the μPADs fabrication. The paint is water resistant and good adhesive to various kinds of materials, and does not require the complex drying step.3838 Hu, R. G.; Zhang, S.; Bu, J. F.; Lin, C. J.; Song, G. L.; Prog. Org. Coat. 2012, 73, 129. The patterned iron mask was used to confine the barrier, proposing a higher requirement for cutting machines and undoubtedly increasing the fabrication cost. Additionally, the paint spraying μPADs developed by Chailapakul and co-workers3737 Nurak, T.; Praphairaksit, N.; Chailapakul, O.; Talanta 2013, 114, 291. was applied for the electrochemical detection of NiII. So far, no work has been reported on the colorimetric detection of one or more analytes with the paint spraying μPADs.

In this work, the single-layered paint-spraying μPADs (sp-μPADs) were integrated with the colorimetric method for the assays of one or more analytes. The patterned polymethyl methacrylate (PMMA) boards were used as the mask for the fabrication of sp-μPADs, which is cheaper than the iron mask. Spray paint was utilized to manufacture the two-dimensional channels in the filter paper. At first, the filter paper was masked with the patterned PMMA boards and then sprayed with the paint. After that, the painted filter paper was dried in an oven. Both the simple and complex fluidic patterns were successfully fabricated by the paint spraying prototyping process. The barrier resistance and the smallest channel size were investigated. Moreover, the validation of the sp-μPADs was demonstrated by detecting one or more analytes based on the color intensity analysis or the distance analysis.

Experimental

Chemicals and equipments

Whatman filter paper No. 1 was used for sp-μPADs fabrication, which was bought from Whatman International Ltd. (Maidstone, England). Spray paint with a volume of 400 mL per can (blue color or yellow color) was purchased from the local automotive products shop for constructing the channels. The design of the mask was drawn by the software of ZWCAD.3939 ZWCAD, v.18.0.55.0; Guangzhou Zhongwang Longteng Software Co., Ltd., China, 2018. The polymethyl methacrylate (PMMA) board (1 mm in thickness) was cut according to the design by a local laser cutting shop. The binder clips with a width of 19 mm were applied to fix the position of the PMMA boards on the filter paper. Bovine serum albumin (BSA), sodium lactate, tetrabromophenol blue (TBPB), chitosan with a viscosity of 400 mPa s, 3,3’,5,5’-tetramethylbenzidine (TMB), the universal indicator, glucose oxidase (GOx) from Aspergillus niger (> 180 U mg-1) and peroxidase from horseradish (HRP, > 200 units mg-1) were provided by Aladdin-Reagent Company (Shanghai, China). Lactate oxidase (LOx) from Aerococcus viridans was bought from Sigma-Aldrich (Shanghai, China). Fe(NO3)3•9H2O, 1,10-phenanthroline monohydrate (phen), ascorbic acid, glucose, potassium thiocyanate, sodium nitrite, citric acid, ethanol, acetic acid, sulfanilamide, N-(1-naphthyl) ethylenediamine dihydrochloride (NED), ethylenediaminetetraacetic acid disodium salt (EDTA), sodium diethyldithiocarbamate (DDTC), dimethylglyoxime, tris base, cupric nitrate, nickel nitrate, and other chemicals were purchased from Sinopharm Chemical Reagent Co., Ltd. (Shanghai, China). The deionized water was prepared by a Milli-Q system (Bedford, USA). The pictures of the sp-μPADs were obtained by a cellular phone (Huawei Nova, Huawei Technologies Co., Ltd, Shenzhen, China) in a photography box with a size of 40 × 40 × 40 cm. The photography box was bought from Shijiazhuang Ruying Film and Television Equipment Sales Co., Ltd. in Shijiazhuang, China and equipped with 84 LED light beads inside. When taking pictures, the cell phone was placed on the photo window at the top of the photography box, and the vertical distance from the prepared μPADs was 8 cm. The color intensity in the different color channels (i.e., red (R), green (G) and blue (B) channels) were measured by the ImageJ software.4040 ImageJ, v.1.50b; National Institutes of Health, USA, 2017. The purified drinking water produced by China Resources C’estbon Beverage Co., Ltd. in Shenzhen, China was employed as the complex sample for the recovery experiment.

sp-μPADs fabricated using a two-step paint spraying prototyping process

The schematic representation of the two-step paint spraying prototyping process for fabricating the device for the microzone assay was presented in Figure 1. One piece of filter paper was put in between two PMMA masks. The whole device was spray-painted for the first time. The filter paper was then taken out and heated at 60 ºC for 10 min. After that, the two patterned acrylic plates were rotated by 90 degrees and the dried filter paper was put in the middle of the two patterned acrylic plates. Then, the whole device was spray-painted for the second time. The filter paper was then taken out and dried again as mentioned above. A photograph of the actual developed μPAD was provided in Figure 1 and the channel’s pattern of the sp-μPADs can be easily changed by changing the mask with different designs.

Figure 1
The schematic representation of the two-step paint spraying prototyping process for fabricating the device.

The microzone assay of iron

Iron is a necessary nutrient to maintain the body growth and development.4141 Gammella, E.; Recalcati, S.; Rybinska, I.; Buratti, P.; Cairo, G.; Oxid. Med. Cell. Longevity 2015, 2015, 230182. Iron deficiency or overload could cause a bad influence on human health.4242 Kontoghiorghe, C. N.; Kontoghiorghes, G. J.; World J. Methodol. 2016, 6, 1. Thus, iron was detected to investigate the application value of the sp-μPADs. The reagent for iron assay was prepared by adding 100 mmol L-1 phen, 2% (m/m) ascorbic acid in sodium acetate buffer (20 mmol L-1, pH 4.6), using 95/5% (v/v) ethanol/water as the solvent. Then the reagent (2.5 μL) was pipetted into the microzone and dried at room temperature (RT) for 6 min. To generate a standard curve, the freshly prepared iron standards (2.5 μL) and the blank (ultrapure water) (2.5 μL) were added into the microzones of the μPAD. The pictures of the sp-μPADs were captured by a cell phone and the color intensities in different channels were analyzed by the ImageJ software.4040 ImageJ, v.1.50b; National Institutes of Health, USA, 2017.

The multiplexed assay of eight analytes

The eight analytes including BSA, glucose, nitrite, thiocyanate, lactate, pH, copper and nickel were used for the multiplexed assay in this work. For BSA assay, 0.5 μL of the reagents including citric acid (250 mmol L-1) and TBPB (3 mmol L-1) in 95% (v/v) ethanol were pipetted into the outer circular assay zone.4343 Garcia, P. T.; Cardoso, T. M. G.; Garcia, C. D.; Carrilho, E.; Coltro, W. K. T.; RSC Adv. 2014, 4, 37637. For the assays of glucose and lactate, 0.5 μL of chitosan (0.5% m/v) in acetic acid (2% v/v) was pipetted into the corresponding assay zones. The enzyme solutions in PBS (100 mmol L-1, pH 6.0) were composed of GOx (120 U mL-1) and HRP (30 U mL-1) for glucose assay and LOx (40 U mL-1) and HRP (300 U mL-1) for lactate assay, respectively. 0.5 μL of TMB solution (15 mmol L-1) was then pipetted into the assay zones.4444 Yu, P.; Deng, M.; Yang, Y.; Sensors 2019, 19, 4082.,4545 Dungchai, W.; Chailapakul, O.; Henry, C. S.; Anal. Chim. Acta 2010, 674, 227. For nitrite assay, the Griess reagent was composed of NED (10 mmol L-1), sulfanilamide (50 mmol L-1), and citric acid (330 mmol L-1). 0.5 μL of the Griess reagent was added into the assay zone.4444 Yu, P.; Deng, M.; Yang, Y.; Sensors 2019, 19, 4082. For thiocyanate assay, 0.5 μL of the solution containing 600 mmol L-1 FeIII was added into the outer circular assay zone.3333 Pena-Pereira, F.; Lavilla, I.; Bendicho, C.; Talanta 2016, 147, 390. For pH assay, 0.5 μL of the universal indicator were pipetted into the outer circular assay zone.4646 Koh, A.; Kang, D.; Xue, Y.; Lee, S.; Pielak, R. M.; Kim, J.; Hwang, T.; Min, S.; Banks, A.; Bastien, P.; Manco, M. C.; Wang, L.; Ammann, K. R.; Jang, K. I.; Won, P.; Han, S.; Ghaffari, R.; Paik, U.; Slepian, M. J.; Balooch, G.; Huang, Y.; Rogers, J. A.; Sci. Transl. Med. 2016, 8, 366ra165. For CuII assay, 0.5 μL of the solution containing 150 mmol L-1 EDTA and 5 mmol L-1 DDTC (pH 7.6) was pipetted into the outer circular assay zone.4747 Xie, G. L.; Yu, H.; Deng, M. H.; Zhao, X. L.; Yu, P.; Chem. Pap. 2019, 73, 1509. For nickel(II) assay, 0.5 μL of the sensing solution composed of dimethylglyoxime (50 mmol L-1) and tris base (50 mmol L-1) in 95/5% ethanol/H2O solvent was pipetted into the outer circular assay zone.4848 Cate, D. M.; Noblitt, S. D.; Volckens, J.; Henry, C. S.; Lab Chip 2015, 15, 2808. All these reagents were allowed to dry at RT for 10 min. Then, the model mixture containing the eight analytes (15 μL) was added to the circular zone in the center. After 10 min of reaction, the photo of the flower-shaped μPADs were taken using a smartphone.

Results and Discussion

Characterization of the fabricated sp-μPADs

The technical challenge in the sp-μPADs fabrication is how to control fluid flow in the channels.4949 Lim, W. Y.; Goh, B. T.; Khor, S. M.; J. Chromatogr. B 2017, 1060, 424. In order to achieve this goal, it is critical to construct the clear boundary to distinguish the hydrophilic area and the hydrophobic area. First, the hydrophobic property of the paint barrier was investigated. The water droplet was dropped on the surface of the sp-μPADs and the picture was shown in Figure 2A. The contact angle with water was 134º, indicating that the area deposited by paint exhibits hydrophobicity. The reason is that the paint blocked the pores in the filter paper, forming an hydrophobic interface on the surface of the paper. For the fabrication of device, it is critical to make sure that the paint fully penetrates into the filter paper. The fabricated μPAD was cut along the red dotted line and the cross section was observed by an optical microscope. It is clearly shown that the paint fully penetrates into the filter paper in the hydrophobic area and there is a well-defined boundary between the hydrophobic area and the hydrophilic area (Figure 2B). The manufacturing cost of a single device (including the cost of PMMA masks and laser cutting, filter paper, paint) is ca. US$ 0.17, manifesting the low cost of the paint spraying method. The channel size was calibrated and the results were presented in Figure 2C. The fabricated device for the channel size calibration possessed 12 radiating channels with the width ranging from 0.1 to 2 mm. It is noted that the smallest channel which was able to be filled in the entire length of the channel with red food dye, was revealed to be 1.4 mm with a designed width of 1.5 mm.

Figure 2
(A) The picture obtained after the water droplet was dropped on the surface of the sp-μPADs. (B) The cross section of the sp-μPADs captured under an optical microscope at 225× magnification. (C) The design of the device for the channel size calibration (a) and the fabricated device treated with the red food dye (b).

Investigation of the barrier resistance

The resistance when exposed to the various kinds of chemicals has been demonstrated. The paint with yellow color were used for the fabrication of the distance-based sp-μPADs to get a clear barrier view. The sp-μPADs used for this experiment included a circular zone for sample introduction with a radius of 4 mm and a channel with a width of 2 mm, whose size was marked in the Figure 3A. Different solutions (25 μL) with food dye was added in the circular zone and the optical micrographs were obtained 20 min later. The micrographs were presented in Figure 3B and indicated that the paint barrier exhibited the high chemical resistance towards the various kinds of chemicals such as acid and alkaline solutions, organic solvents, surfactants, buffer solution and the reducing reagent.

Figure 3
(A) The fabricated distance-based sp-μPADs. (B) The optical micrographs of the paint barrier after adding different solutions: (a) 150 mmol L-1 hydrochloric acid, (b) 150 mmol L-1 acetic acid, (c) 150 mmol L-1 sodium hydroxide, (d) 50% (v/v) methanol, (e) 50% (v/v) ethanol, (f) 5% (m/v) tetradecyltrimethylammonium bromide (TTAB), (g) 5% (m/v) triton X-100, (h) phosphate buffer solution (PBS) (100 mmol L-1 pH 7.4), (i) 150 mmol L-1 hydroxylamine hydrochloride.

The microzone assay performance for iron

The microzone array consisted of 9 zones arranged into 3 columns and 3 rows. To investigate the capability of the developed device, iron was used as the model analyte. The sp-μPADs for iron assay was feasible and the optimized parameter was obtained experimentally. The results shown in Figure S1 (Supplementary Information (SI) section) indicated that the B channel was selected as the analytical signal due to its high sensitivity. The results in Figure S2 (SI section) indicated that 100 mmol L-1 was the optimized concentration of phen for iron assay. Under the optimized conditions, a series of FeIII standards were pipetted into the microzones. The assay color intensity was analyzed using ImageJ software.4040 ImageJ, v.1.50b; National Institutes of Health, USA, 2017. It is noted from Figure 4a that the color in the assay zone gradually darken with the concentrations of FeIII increasing from 0 to 400 mg L-1, resulting in a notable decrease in the blue channel intensity. Figure 4b shows that a strong linear calibration plot (blue channel intensity = -0.7235CFe + 202, R2 = 0.9951) was obtained with a linear range from 5 to 150 mg L-1. The limit of detection (LOD) was calculated to be 2.2 mg L-1 by 3σ/S, σ is the standard deviation of the signal in nine parallel blank measurements and S is the slope of the calibration equation. For comparison, some representative chemosensors for iron detection are summarized in Table 1. It is noted that the performance of the sp-μPADs in this work were comparable with most of the literature listed in Table 1. The μPADs fabricated by Henry and co-workers5454 Mentele, M. M.; Cunningham, J.; Koehler, K.; Volckens, J.; Henry, C. S.; Anal. Chem. 2012, 84, 4474.,5555 Rattanarat, P.; Dungchai, W.; Cate, D.; Volckens, J.; Chailapakul, O.; Henry, C. S.; Anal. Chem. 2014, 86, 3555. presented the lower LOD for iron assay using the same reaction. The possible reason is that the poly(acrylic acid) and polyethylene glycol were added to immobilize the colored product and increase the hydrophilicity in the reservoir, resulting in the better assay performance.

Figure 4
(a) The results for iron assay. Inset: the pictures of the microzones after adding a series of FeIII standards. (b) Calibration curve for iron assay.

Table 1
The performance contrast of the different chemosensors for iron detection

To investigate the anti-interference ability of the fabricated sp-μPADs, the selectivity experiment was carried out using several common interferent ions as the interferents (CaII, CoII, KI, MgII, NaI, NiII, PbII, ZnII). Figure S3 (SI section) shows the images and the color values in the B channel of the assay zones with different ions. With a comparison to the blank, FeIII solution (100 mg L-1) produced an orange-red color in the detection zone and resulted in a significant decrease of the color intensity. The other ions (10000 mg L-1) produced no color with the reagent. The color intensities obtained in the interferents were almost the same as the intensity obtained in the blank, demonstrating the good selectivity of the sp-μPADs for FeIII detection. The repeatability was investigated by performing the determination of five concentration levels (5, 30, 50, 100, 150 mg L-1) with nine replicate measurements using the sp-μPADs. For the five concentration levels investigated, the relative standard deviations (RSD) were 4.6, 4.5, 3.6, 3.5 and 4.8%, respectively. The results demonstrated that the developed sp-μPADs performed a satisfactory repeatability.

To study the practicability of the sp-μPADs for iron assay, the recovery test was performed using the drinking water as the real world sample. Iron was not detected in the drinking water by our method. The possible reason may be that the concentration of iron in the sample was below the LOD, or there was no iron in the water sample. The water sample was spiked with the standard solutions of FeIII. The results of applying the assay to the real world sample are listed in Table S1 (SI section). The recoveries of FeIII ranged from 94.0 to 108.8%. In addition, all the RSD values of the recovery test were found to be below 4%, showing that the sp-μPADs were suitable for iron assay in the real world sample.

The simultaneous assay of copper and nickel

The applicability was also demonstrated by the simultaneous assay of copper and nickel. The assay reagents for the two ions assay were described in the “The multiplexed assay of eight analytes” sub-section. The fabricated device for this experiment was designed with two assay zones on both sides and one sample introduction zone in the center. All the circular zones in the device were designed with a diameter of 6 mm. The size of the channels between the circular zones was 5.4 mm (length) × 2 mm (width). 1.8 μL of the reagents were added in the assay zones on both sides and dried at RT for 10 min. 20 μL of the standard was then added to the circular zone in the center. The photos of the sp-μPADs were captured 10 min later and the color intensity was analyzed by the free image processing software. It is noted from Figure 5 that when the concentrations of the analytes increased, the colors in the assay zones became darker, indicating that more complexation between the analyte and reagents was produced. The linear range was 5-9 mmol L-1 for the copper assay and 1-6 mmol L-1 for the nickel assay, respectively. The detectable limit was 5 mmol L-1 for the copper assay and 1 mmol L-1 for the nickel assay, respectively. The results demonstrated that the sp-μPADs were successfully fabricated and could be extended to the simultaneous assay of two analytes.

Figure 5
(a) The photos of the sp-μPADs for the simultaneous assay of copper and nickel. (b) The calibration curve for copper assay. (c) The calibration curve for nickel assay.

The distance-based sp-μPADs for the detection of BSA

The distance-based sp-μPADs were also prepared by the paint spraying method, consisting of a circular reservoir with a radius of 4 mm for analyte introduction and a straight channel (2 × 25 mm) for the distance measurement. BSA was used as the model analyte to prove the applicability of the distance-based sp-μPADs. For the detection of BSA, the reagents including 250 mmol L-1 citric acid and 3 mmol L-1 TBPB were dissolved in 95% (v/v) ethanol. Then, 3 μL of the reagents were pipetted into the straight channel and dried at RT for 10 min. 25 μL of BSA standard was then added to the circular zone. After 5 min of reaction, the color distance was measured. The photos of the distance-based sp-μPADs for BSA assay are presented in the Figure 6a. The color of the reagents were yellow but produced a dark blue product that was easily visible after the reaction with BSA. Moreover, when the concentration of BSA increased, the color distance rose higher and higher, indicating a positive correlation between them. It is noted from Figure 6b that the linear range for BSA assay was 0.05-1 g L-1 and the visual limit of detection was 0.05 g L-1. The linearity for the detection of BSA was expressed by the squared correlation coefficient (R2), which equaled 0.9881. Although the first data point (0.05 g L-1, 1.7250 mm) reduces the correlation between the set of data, it was not removed. The reason is that the urinary albumin excretion (i.e., microalbuminuria) is an important marker for the early-stage renal failure when the levels of urine albumin are between 0.02 and 0.2 g L-1.5656 Martinez, A. W.; Phillips, S. T.; Carrilho, E.; Thomas, S. W.; Sindi, H.; Whitesides, G. M.; Anal. Chem. 2008, 80, 3699. Therefore, the distance assay of BSA sacrifices the linearity for the wide dynamic range, which partially overlaps the clinical range of urine albumin. The results demonstrated that the distance-based sp-μPADs were successfully fabricated by the paint spraying technique and exhibited a promising prospect for bioassay.

Figure 6
(a) The photos of the fabricated distance-based sp-μPADs for BSA assay. (b) The calibration curve for BSA assay.

The flower-shaped sp-μPADs fabricated by the paint-spraying method for the multiplexed assay

More complex design was employed for the device fabrication. The layout of the flower shaped sp-μPADs were manufactured with the spraying parameters listed as follows: the channel with a 2.0 mm width and a 4.5 mm length, the central analyte introduction reservoir with a diameter of 6.0 mm, the outer circular assay zone with a diameter of 4.0 mm. Figure 7 shows a photograph of the flower-shaped sp-μPADs for simultaneously detecting a model mixture containing the eight analytes including BSA (1 g L-1), glucose (2 mmol L-1), nitrite (1 mmol L-1), thiocyanate (10 mmol L-1), lactate (20 mmol L-1), pH (7), CuII (20 mmol L-1) and NiII (20 mmol L-1). These analytes are clinically relevant and commonly measured in biological systems such as urine or saliva, and their values are closely related to the human health. With a comparison to the blank, the eight analytes were successfully detected at the same time. Therefore, certain diseases could be quickly screened by the color changes obtained in the developed flower shaped sp-μPADs, providing a great analytical platform for the clinical testing.

Figure 7
The photograph of the flower-shaped sp-μPADs for the multiplexed assay of a model mixture with a volume of 15 μL: BSA (1 g L-1), glucose (2 mmol L-1), nitrite (1 mmol L-1), thiocyanate (10 mmol L-1), lactate (20 mmol L-1), pH (7), CuII (20 mmol L-1) and NiII (20 mmol L-1).

Conclusions

The simple, low-cost sp-μPADs have been fabricated by the paint spraying technique with simple or complex designs by changing the patterned PMMA mask. The applicability of the sp-μPADs was proved by implementing colorimetric detections of one or more analytes. The microzone assay performance for iron exhibited a LOD of 2.2 mg L-1 and a linear range of 5-150 mg L-1. The detectable limits for the simultaneous assay of copper and nickel were 5 mmol L-1 for copper assay and 1 mmol L-1 for nickel assay, respectively. The distance-based sp-μPADs were successfully fabricated and applied for the detection of BSA with a visual limit of detection of 0.05 g L-1. Furthermore, the possible application of the flower-shaped sp-μPADs were demonstrated by the multiplexed assay of eight analytes. Accordingly, the sp-μPADs offered a promising prospect on detecting single analyte or multiple analytes, showing the potential applications in environmental analysis and bioassays.

Supplementary Information

Supplementary information is available free of charge at http://jbcs.sbq.org.br as PDF file.

Acknowledgments

The authors are grateful to the financial support from the Student’s Platform for Innovation and Entrepreneurship Training Program (grant number S201910530018) and the Natural Science Foundation of Hunan Province (grant number 2018JJ3500). We are also grateful to Saichon Sumantukul and Matthew Kremer from Oregon State University for project review.

References

  • 1
    Martinez, A. W.; Phillips, S. T.; Butte, M. J.; Whitesides, G. M.; Angew. Chem., Int. Ed. 2007, 46, 1318.
  • 2
    Channon, R. B.; Nguyen, M. P.; Scorzelli, A. G.; Henry, E. M.; Volckens, J.; Dandy, D. S.; Henry, C. S.; Lab Chip 2018, 18, 793.
  • 3
    Jiang, Y.; Wang, P. C.; Locascio, L. E.; Lee, C. S.; Anal. Chem 2001, 73, 2048.
  • 4
    Lee, J.; Soper, S. A.; Murray, K. K.; J. Mass Spectrom. 2009, 44, 579.
  • 5
    Nie, Z.; Nijhuis, C. A.; Gong, J.; Chen, X.; Kumachev, A.; Martinez, A. W.; Narovlyansky, M.; Whitesides, G. M.; Lab Chip 2010, 10, 477.
  • 6
    Sun, X.; Wang, H.; Jian, Y.; Lan, F.; Zhang, L.; Liu, H.; Ge, S.; Yu, J.; Biosens. Bioelectron 2018, 105, 218.
  • 7
    Hassani, A.; Skorobogatiy, M.; Opt. Express 2006, 14, 11616.
  • 8
    Malic, L.; Veres, T.; Tabrizian, M.; Biosens. Bioelectron 2009, 24, 2218.
  • 9
    Wang, L.; Musile, G.; McCord, B. R.; Electrophoresis 2018, 39, 470.
  • 10
    Ellerbee, A. K.; Phillips, S. T.; Siegel, A. C.; Mirica, K. A.; Martinez, A. W.; Striehl, P.; Jain, N.; Prentiss, M.; Whitesides, G. M.; Anal. Chem 2009, 81, 8447.
  • 11
    Li, B.; Fu, L.; Zhang, W.; Feng, W.; Chen, L.; Electrophoresis 2014, 35, 1152.
  • 12
    Fiedoruk-Pogrebniak, M.; Granica, M.; Koncki, R.; Talanta 2018, 178, 31.
  • 13
    Weng, X.; Neethirajan, S.; Microchim. Acta 2017, 184, 4545.
  • 14
    Qi, J.; Li, B.; Wang, X.; Fu, L.; Luo, L.; Chen, L.; Anal. Chem 2018, 90, 11827.
  • 15
    Coltro, W. K. T.; de Jesus, D. P.; da Silva, J. A. F.; do Lago, C. L.; Carrilho, E.; Electrophoresis 2010, 31, 2487.
  • 16
    Lisowski, P.; Zarzycki, P. K.; Chromatographia 2013, 76, 1201.
  • 17
    Caputo, D.; de Cesare, G.; Dolci, L. S.; Mirasoli, M.; Nascetti, A.; Roda, A.; Scipinotti, R.; IEEE Sens. J. 2013, 13, 2595.
  • 18
    Kamruzzaman, M.; Alam, A. M.; Kim, K. M.; Lee, S. H.; Kim, Y. H.; Kabir, A. H.; Kim, G. M.; Dang, T. D.; Biomed. Microdevices 2013, 15, 195.
  • 19
    Koesdjojo, M. T.; Pengpumkiat, S.; Wu, Y.; Boonloed, A.; Huynh, D.; Remcho, T. P.; Remcho, V. T.; J. Chem. Educ. 2015, 92, 737.
  • 20
    Rossini, E. L.; Milani, M. I.; Carrilho, E.; Pezza, L.; Pezza, H. R.; Anal. Chim. Acta 2018, 997, 16.
  • 21
    Chabaud, K. R.; Thomas, J. L.; Torres, M. N.; Oliveira, S.; McCord, B. R.; Forensic Chem. 2018, 9, 35.
  • 22
    Taghizadeh-Behbahani, M.; Hemmateenejad, B.; Shamsipur, M.; Chem. Pap 2018, 72, 1239.
  • 23
    Liu, C.; Gomez, F. A.; Miao, Y.; Cui, P.; Lee, W.; Talanta 2019, 194, 171.
  • 24
    da Silva, G. O.; de Araujo, W. R.; Paixao, T. R.; Talanta 2018, 176, 674.
  • 25
    Cao, R.; Zhang, X.; Tan, W.; Shen, W.; Cellulose 2018, 25, 583.
  • 26
    Asano, H.; Shiraishi, Y.; Anal. Chim. Acta 2015, 883, 55.
  • 27
    Abdollahi-Aghdam, A.; Majidi, M. R.; Omidi, Y.; BioImpacts 2018, 8, 237.
  • 28
    Puneeth, S. B.; Goel, S.; IEEE Trans. Electron Devices 2019, 66, 3196.
  • 29
    Chiang, C. K.; Kurniawan, A.; Kao, C. Y.; Wang, M. J.; Talanta 2019, 194, 837.
  • 30
    Zhang, H.; Smith, E.; Zhang, W.; Zhou, A.; Biomed. Microdevices 2019, 21, 48.
  • 31
    Casto, L. D.; Schuster, J. A.; Neice, C. D.; Baker, C. A.; Anal. Methods 2018, 10, 3616.
  • 32
    Cardoso, T. M.; Garcia, P. T.; Coltro, W. K.; Anal. Methods 2015, 7, 7311.
  • 33
    Pena-Pereira, F.; Lavilla, I.; Bendicho, C.; Talanta 2016, 147, 390.
  • 34
    Atabakhsh, S.; Ashtiani, S. J.; Microfluid. Nanofluid. 2019, 23, 69.
  • 35
    Al-Qadiri, H. M.; Ovissipour, M.; Al-Alami, N.; Govindan, B. N.; Shiroodi, S. G.; Rasco, B.; J. Food Sci. 2016, 81, M1177.
  • 36
    Cardoso, T. M. G.; de Souza, F. R.; Garcia, P. T.; Rabelo, D.; Henry, C. S.; Coltro, W. K. T.; Anal. Chim. Acta 2017, 974, 63.
  • 37
    Nurak, T.; Praphairaksit, N.; Chailapakul, O.; Talanta 2013, 114, 291.
  • 38
    Hu, R. G.; Zhang, S.; Bu, J. F.; Lin, C. J.; Song, G. L.; Prog. Org. Coat. 2012, 73, 129.
  • 39
    ZWCAD, v.18.0.55.0; Guangzhou Zhongwang Longteng Software Co., Ltd., China, 2018.
  • 40
    ImageJ, v.1.50b; National Institutes of Health, USA, 2017.
  • 41
    Gammella, E.; Recalcati, S.; Rybinska, I.; Buratti, P.; Cairo, G.; Oxid. Med. Cell. Longevity 2015, 2015, 230182.
  • 42
    Kontoghiorghe, C. N.; Kontoghiorghes, G. J.; World J. Methodol. 2016, 6, 1.
  • 43
    Garcia, P. T.; Cardoso, T. M. G.; Garcia, C. D.; Carrilho, E.; Coltro, W. K. T.; RSC Adv. 2014, 4, 37637.
  • 44
    Yu, P.; Deng, M.; Yang, Y.; Sensors 2019, 19, 4082.
  • 45
    Dungchai, W.; Chailapakul, O.; Henry, C. S.; Anal. Chim. Acta 2010, 674, 227.
  • 46
    Koh, A.; Kang, D.; Xue, Y.; Lee, S.; Pielak, R. M.; Kim, J.; Hwang, T.; Min, S.; Banks, A.; Bastien, P.; Manco, M. C.; Wang, L.; Ammann, K. R.; Jang, K. I.; Won, P.; Han, S.; Ghaffari, R.; Paik, U.; Slepian, M. J.; Balooch, G.; Huang, Y.; Rogers, J. A.; Sci. Transl. Med. 2016, 8, 366ra165.
  • 47
    Xie, G. L.; Yu, H.; Deng, M. H.; Zhao, X. L.; Yu, P.; Chem. Pap 2019, 73, 1509.
  • 48
    Cate, D. M.; Noblitt, S. D.; Volckens, J.; Henry, C. S.; Lab Chip 2015, 15, 2808.
  • 49
    Lim, W. Y.; Goh, B. T.; Khor, S. M.; J. Chromatogr. B 2017, 1060, 424.
  • 50
    Singh, H.; Sindhu, J.; Khurana, J. M.; Sens. Actuators, B 2014, 192, 536.
  • 51
    Tripathy, S. K.; Woo, J. Y.; Han, C. S.; Sens. Actuators, B 2013, 181, 114.
  • 52
    Kaur, P.; Kaur, H.; Singh, K.; RSC Adv. 2013, 3, 64.
  • 53
    Apilux, A.; Dungchai, W.; Siangproh, W.; Praphairaksit, N.; Henry, C. S.; Chailapakul, O.; Anal. Chem 2010, 82, 1727.
  • 54
    Mentele, M. M.; Cunningham, J.; Koehler, K.; Volckens, J.; Henry, C. S.; Anal. Chem 2012, 84, 4474.
  • 55
    Rattanarat, P.; Dungchai, W.; Cate, D.; Volckens, J.; Chailapakul, O.; Henry, C. S.; Anal. Chem 2014, 86, 3555.
  • 56
    Martinez, A. W.; Phillips, S. T.; Carrilho, E.; Thomas, S. W.; Sindi, H.; Whitesides, G. M.; Anal. Chem 2008, 80, 3699.

Publication Dates

  • Publication in this collection
    01 Mar 2021
  • Date of issue
    Mar 2021

History

  • Received
    31 July 2020
  • Accepted
    30 Oct 2020
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br