Dissolved Organic Carbon Determination Using FIA and Photo-Fenton Reaction

Abstracts

The FIA-photo-Fenton system is based on the flow oxidation of the organic matter. A small amount of the sample containing H2O2 is injected into the acidic flow solution of Fe2+, which passes through a tubular PTFE reactor irradiated with UV light. The generated CO2 is quantified by a conductometric detector and is directly proportional to the dissolved organic carbon concentration in the sample. The optimization studies were performed using EDTA solutions. The average recovery of organic carbon was 83% with a relative standard deviation of 3.7% using a 1:5 molar ratio of Fe2+:H2O2, pH 2.0, 100 muL of sample injection and a liquid flow of 1 mL min-1. After optimization, the DOC concentration was quantified using 13 different organic compounds, where the average recovery was 90%. The rate of the analysis was in average 50 samples hour-1.

Photo-Fenton reaction; dissolved organic carbon; FIA


O sistema FIA-foto-Fenton é baseado na oxidação em fluxo da matéria orgânica. Uma pequena quantidade de amostra contendo H2O2 é introduzida no fluxo de uma solução ácida de Fe2+, que passa por um reator tubular de PTFE irradiado com luz UV. O CO2 gerado é quantificado condutometricamente e é diretamente proporcional à concentração de carbono orgânico dissolvido na amostra. Os estudos de otimização foram realizados empregando soluções de EDTA. A recuperação média de carbono orgânico foi de 83% com desvio padrão relativo de 3,7% empregando 100 miL de amostra, pH 2,0, razão molar entre Fe2+ e H2O2 de 1:5 e uma vazão de 1 mL.min-1. Após otimização as concentrações de DOC foram quantificadas em 13 diferentes compostos orgânicos, com uma recuperação média de 90%. A velocidade de análise foi em média 50 amostras/h.


Dissolved Organic Carbon Determination Using FIA and Photo-Fenton Reaction

Márcia M. Kondo1* * Author for correspondence , Maria A. S. V. Arcos2 and Marco T. Grassi3

1Departamento de Física e Química, Escola Federal de Engenharia de Itajubá, 37500-176, Itajubá – MG, Brazil; 2Departamento de Tecnologia Química, Universidade Federal do Maranhão, 65080-040, São Luis – MA, Brazil; 3Departamento de Química, Universidade Federal do Paraná - CP 19081, 81531-990, Curitiba – PR, Brazil

ABSTRACT

The FIA-photo-Fenton system is based on the flow oxidation of the organic matter. A small amount of the sample containing H2O2 is injected into the acidic flow solution of Fe2+, which passes through a tubular PTFE reactor irradiated with UV light. The generated CO2 is quantified by a conductometric detector and is directly proportional to the dissolved organic carbon concentration in the sample. The optimization studies were performed using EDTA solutions. The average recovery of organic carbon was 83% with a relative standard deviation of 3.7% using a 1:5 molar ratio of Fe2+:H2O2, pH 2.0, 100 mL of sample injection and a liquid flow of 1 mL min-1. After optimization, the DOC concentration was quantified using 13 different organic compounds, where the average recovery was 90%. The rate of the analysis was in average 50 samples hour-1.

Key words: Photo-Fenton reaction, dissolved organic carbon, FIA

INTRODUCTION

Total organic carbon (TOC) is an important parameter used as a tool to assess, for instance, the quality of a purified water. The determination and quantification of organic compounds in aqueous solution is not always an easy process which often, includes an specific analytical method. In this regard, the total organic carbon analysis can be used as an indirect method to quantify the concentration of a solution containing a single organic compound. When a more complex system is involved, TOC analysis shows the results of the sum of all organic compounds (Matthews et al., 1990).

These analyses are performed, usually, by two techniques, combustion at very high temperature, typically at 990° C using cobalt oxide as an oxidative catalyst, and photo-oxidation employing persulfate-ultraviolet irradiation. In both the methods, the organic carbon is converted to carbon dioxide (CO2) which can be analyzed by infrared analyzer or flame ionization detector (APHA-AAWWA-WPCF, 1995).

Recently, there has been numerous investigations (Huang et al., 1993; Venkatadri & Peters, 1993; Lunar et al., 2000) using alternative processes to oxidize organic compounds, called advanced oxidation processes (AOP). The active species generated are the highly reactive free radicals (Huang et al., 1993; Jardim et al., 1997), mainly hydroxyl radicals (), which are the initiator of a chain of reactions that leads to the total mineralization of the organic compounds to CO2, H2O and other inorganic.

The Fenton reaction is, among the AOPs, one of the most investigated. The reagent is a mixture of hydrogen peroxide and ferrous ion which produces according to the equation [1] (Huston & Pignatello, 1999; Pignatello, 1992; Engwall et al., 1999):

The literature shows that Fenton’s reagent is capable of efficiently degrade several organic compounds (Huang et al., 1993; Venkatadri & Peters, 1993; Lunar et al., 2000). In addition, Fenton reaction efficiency can be enhanced in the presence of UV irradiation (Pigantello, 1992). The iron product in equation [1] can be photoreduced regenerating Fe(II) and producing more , as shown in equation [2]:

The combination of Fenton reaction and UV irradiation is known as photo-assisted Fenton or photo-Fenton reaction. Bauer and Fallmann (1997) compared the wastewater treatment efficiency in a pilot scale employing AOP systems that can combine sunlight. The studied systems were TiO2/UV, Fe2+/H2O2/UV, Fe2+/O2/UV and Fe2+/O3/UV. The authors observed that among the studied system, the one using photo-Fenton reaction was the most effective and also pointed out that the estimated cost is cheaper than other AOP systems.

Several researchers have investigated the photo-Fenton reaction as an alternative and very promising wastewater treatment process. Doong and Chang (1998) investigated the degradation of organophosphorous pesticides, such as malathion, diazon and phorate, and observed that the degradation efficiency was more effective in a system using Fe2+/UV/H2O2 than in H2O2/UV system. The degradation of several pesticides was also investigated by Huston and Pignatello (1999). The authors showed that the pesticides were removed in less then 30 min and that the complete mineralization of this compounds was achieved in an average of 120 min. Compounds such as 4-chlorobiphenyl, 4,4’-dichlorobiphenyl Kuo & Lo, 1999), 2,4-chlorophenol, 2,4,6-trichlorophenol (Kuo et al., 1998), and colored compounds (Liao et al., 1999; Kang et al., 1999; Wu et al., 1999) were also employed as target ones for the organic degradation study using photo-Fenton reaction.

Matthews and co-workers (1990) described a new method for the determination of organic carbon using TiO2/UV, another powerful AOP system. The carbon dioxide produced was monitored by conductivity detection. Sample of 1 to 40 mL containing 0.1 to 30 mg mL-1 of organic carbon was recirculated through a spiral borosilicate glass tube coated with TiO2. A 20 W black light fluorescent lamp was used as UV source. The authors reported that 99% oxidation was achieved in less than 10 minutes of irradiation. They also mentioned that the main advantages of the apparatus are the low cost and ease operation.

In the present work we investigated the possibility of using Fenton and photo-Fenton reaction as an alternative method to quantify the dissolved organic carbon concentration. As mentioned before, one of the final products of the organic compounds degradation process, in both systems, is CO2. The constructed apparatus consisted of mineralizing the organics by photo-Fenton reaction using flow injection analysis (FIA) system. The CO2 generated due to the organic compounds mineralization was quantified by monitoring the increase of the conductivity. This arrangement is similar to the one used by Fadini (1995), where the author used a FIA-condutometric system and persulfate-ultraviolet irradiation to convert organic compounds to CO2. Under these conditions, Fadini obtained an average of 88% of dissolved organic carbon recovery and the equipment was able to process 14 samples per hour.

MATERIALS AND METHODS

Reagents

Inorganic carbon standard solution (1200 mg of C L-1), Na2CO3 (Vetec) was dried at 270° C for 1 hour and 10.6 g was dissolved with 1 L to purified water, obtained from Milli-Q Academic.

Hydrogen peroxide, 30% (Alfa Caesar) was standardized by permanganate titration with a concentration of 9.7 mol L-1.

Fe2+ solution (0.10 mol L-1) was prepared dissolving 27.80 g of FeSO4.7H2O (Reagen) in 1 L of an acidic solution containing 500 mmol L-1 of H2SO4 (Reagen).

Organic carbon standard solutions were prepared containing 1200 mg of C L-1. Organic compounds used in this work were EDTA (Reagen); methanol (Vetec); ethanol (Vetec); isopropanol (Nuclear); acetone (Vetec); phenol (Reagen); ascorbic acid (Reagen); hydroquinone (Vetec); pyridine (Vetec); glucose (Vetec); resorcinol (Vetec); caffeine (Nuclear), and 1,4-dioxane (Nuclear).

Equipment and Analysis

Samples containing a pre-determined amount of H2O2 were injected into a carrying flow of 10-3 mol L-1 Fe2+ acidic solution. This solution was allowed to pass through a 1.50 m length PTFE tube which was winded around the UV lamp (6 W, l = 254 nm). Using this tube the UV light can penetrate and irradiate the sample. During this course the organic compounds were oxidized to CO2 and H2O. The CO2 passed through a PTFE membrane to a purified water flow, in a diffusion cell, increasing the water conductivity, measured with a conductometer (CPH-100). The peak intensity was recorded (Instrumentos Científicos, C.G.). Figure 1 shows a schematic representation of the FIA-photo-Fenton system.


Organic carbon recovery study

Several organic compounds were investigated. For each one of them, solutions containing 6, 12, 24, 36 and 60 mg L-1 of organic carbon were prepared. To quantify the percent of organic carbon recovery, the results obtained for these samples were compared to a calibration curve, using the same carbon concentrations, prepared from a stock inorganic carbon (Na2CO3) standard solution.

RESULTS AND DISCUSSION

Fenton’s Reagent in FIA

A preliminary investigation was performed to determine the optimum configuration, and to know how the Fenton’s reagents (Fe2+ and H2O2 solutions) should be employed in FIA.

Injections of samples to an acidic Fe2+/H2O2 flow resulted in a severe instability, caused by the formation of, probably, oxygen gas bubbles. Oxygen can be formed due to the decomposition of hydrogen peroxide (Huang et al., 1993; Crittenden et al., 1999).

To avoid this problem, H2O2 was first introduced to the sample and then injected to the carrying flow of the Fe2+ solution. Using this condition, no bubbles were observed and the base line was kept stabilized during the analysis.

Fenton x Photo-Fenton

For this comparison study, EDTA was employed as the testing organic compound, and the FIA conditions used were 100 mL of sample volume, 1 mL min-1 of flow rate and 5 mmol L-1 H2O2.

As mentioned before, Fenton’s reagent is a strong oxidant that is capable to mineralize many organic compounds. The reaction time for the organic compounds mineralization depends on many parameters, such as the Fe2+/H2O2 concentration ratio, the chemical structure of testing compound, and pH, among others. The literature shows that this time could vary from 10 to over 120 min (Bandara et al., 1996; Blanco et al., 1999; Kitis et al., 1999). However, due to the short period of reaction allowed in this FIA settings (less than 2 minutes), the present oxidation efficiency was low. The performance was enhanced combining Fenton’s reagent and sample irradiation. Figure 2 shows the comparison of the oxidation results using Fenton and photo-Fenton reactions. Increase in the conductivity of the solution, indicated as peak height, was 9 times higher for the photo-Fenton system than for the Fenton system.


As already mentioned (Pignatello, 1992), the concentrations is higher in a photo-Fenton system compared to Fenton reaction, due to the photoreduction of Fe(OH)2+ to Fe2+, where more is generated. Moreover, under UV radiation, the H2O2 undergoes direct photolysis generating two hydroxyl radicals per each peroxide molecule (Topodurti et al., 1993; Legrini et al., 1993). The H2O2 has a maximum absorbance in the range of 210-240 nm, and even though one of the dominant emission wavelength of the mercury UV lamp is at 254 nm, it is possible to generate by direct photolysis of H2O2.

The results suggested that the Fenton reaction alone could not generate enough amounts of that would be capable to mineralize the organic compounds in the short time period of the FIA analysis. Under UV assisted Fenton reaction, more are produced and the oxidation of the organics to CO2 can be enhanced in the same period.

System Optimization

A. Fe2+/H2O2 Concentration Ratio. The literature shows that apparently there is not an unique ratio between H2O2 and Fe2+ concentration that could be used to oxidize organic compounds. Degradation studies using 2-chlorophenol and ionic surfactants were reported employing Fe2+:H2O2 molar ratio of 1:100 and 1:1, respectively (Huang et al., 1993; Kitis et al., 1999). Therefore, in this work, different Fe2+:H2O2 molar concentration ratios were investigated. The experiments were performed using 1.0 mL min-1 flow rate and 100 mL of sampling volume.

Figure 3 shows the average percentage recovery as organic carbon in EDTA solutions. The results showed a higher organic carbon recovery at a Fe2+:H2O2 molar ratio of 1:3; 1:4 and 1:5. At these ratios the average recovery was 70, 73 and 74%, respectively. Due to the slightly higher recovery obtained using 1:5 of Fe2+:H2O2 molar concentration ratio, this one was chosen to be used for further studies.


B. pH.

According to Huang et al. (1993), the hydrogen peroxide decomposition rate for Fenton reaction reached the maximum at a pH of 3.5. They attributed this phenomenon to the large catalytically active surface for H2O2 contact due to the progressive hydrolysis of the ferric iron. As a consequence, more free hydroxyl radicals were formed. For the photo-Fenton reaction, however, the optimum pH is at 2.8, where both Fe(III) species, present as Fe3+ and Fe(OH)2+, have approximately the same concentration. At lower pH values, the Fe(OH)2+ concentration decreases and at higher pH values, Fe(III) hydroxide precipitates (Pignatello, 1992). Nogueira and Guimarães (2000) also verified that the optimum pH range for photo-Fenton process was at range of 2.5-2.8, studying the degradation of dichloroacetic acid and 2,4-dichlorophenol.

Studies using FIA and condutometric detection (Jardim et al., 1990) to monitor the microorganism CO2 production in a liquid medium containing Escheririchia coli cultures showed better results using 10-2 mol L-1 of H2SO4. The pH of this solution was theoretically 1.7. Due to the fact that the present apparatus was a combination of photo-Fenton reaction in a FIA system, different pH values for the Fe2+ solution were studied. The optimum organic carbon recovery results were obtained using 5 mmol L-1 H2SO4 which resulted in pH value of 2.0. At higher pH values, the system presented an increase in instability.

C. Sampling Volume

The sample volume to be injected into FIA system was the other parameter optimized. Figure 4 shows the average organic carbon recovery obtained using four different sampling volumes. Results showed a higher dispersion in recovery tests when 50 or 200 mL of samples were injected. The optimum results were observed using 100 mL of injection volume, with an average organic carbon recovery of 81% and a 5.5% of variation coefficient among results.


D. Flow Rate

Several flow rates were employed. It was observed that the highest organic carbon recovery results were obtained using either 0.8 or 1.0 mL min-1. Statistically, using the T test at 95% confidence, there was no difference in the results among them. Lower values of flow rates resulted in system instability. At higher ones, the reagent diffusion into the sample cannot be totally accomplished and consequently, the oxidation of the organic could not be complete.

The following studies were performed using the flow rate of 0.8 mL min-1. The reason was that different organic compounds undergo oxidation at the different rates (Huston & Pignatello, 1999; Maletzky & Bauer, 1998; Krutzler & Bauer, 1999), therefore, the reaction time could be slightly higher at 0.8 mL min-1 than at 1.0 mL min-1. As a consequence, the oxidation percentage should also be higher.

Organic Carbon Recovery Study. Although the experimental parameters were optimized using EDTA as a test compound, for practical purpose, they were not modified for this study. Glucose, caffeine, pyridine, methanol, ethanol, isopropanol, resorcinol, phenol, acetone, hydroquinone, ascorbic acid and 1,4-dioxane were used, each one at the same concentration of organic carbon. Table 1 shows the average organic carbon recovery obtained for all the compounds studied. For most of the compounds investigated 100% recovery could be reached. Ascorbic acid recovery was in the range of 90%, EDTA and caffeine at 80% and isopropanol and acetone at 75%. It is known that the degradation rate is different for each compound (Jardim et al., 1997). Therefore, modifications of some of the system parameters, could probably enhance the percentage recovery of the organics that did not reach 100%.

Matthews and co-workers (1990) used TiO2/UV, another powerful AOP system, to oxidize organic compounds to CO2 and consequently determine the total organic carbon concentration. The authors investigated 11 compounds and reported that 99% oxidation was achieved in less than 10 minutes.

They reported that the method was suitable for samples containing 0.1 to 30 mg mL-1 of organic carbon and sample volumes of 1 to 40 mL.

Determination of DOC in real samples. Aqueous samples of four different sources were used to verify the efficiency of the FIA-photo-Fenton system in quantifying the DOC concentration. These results were compared to the ones obtained using a commercial TOC equipment (Shimadzu, TOC 5000). The comparison is presented in table 2. In general, it could be seen that there was not a significant difference among the values obtained using either methods.

Using a similar FIA apparatus, but employing persulfate/UV as the oxidant agent, Fadini (1995) compared the dissolved organic carbon results with the ones using the same commercial TOC equipment. The values obtained from Shimadzu TOC 5000 were on average 12% higher than the FIA system using persulfate/UV.

Although further studies have to be done in order to cover a wider range of dissolved organic carbon analysis, it is important to emphasize that the present system has some advantages. The rate of the analysis was on average 50 samples hour-1, using a sample volume of 100 mL. Moreover, using the present apparatus, there was no need for a special carrying gas neither for high temperatures to oxidize the organics, therefore, it could be used during field analysis.

ACKNOWLEDGEMENTS

This work was partially funded by CAPES and FAPEMA.

RESUMO

O sistema FIA-foto-Fenton é baseado na oxidação em fluxo da matéria orgânica. Uma pequena quantidade de amostra contendo H2O2 é introduzida no fluxo de uma solução ácida de Fe2+, que passa por um reator tubular de PTFE irradiado com luz UV. O CO2 gerado é quantificado condutometricamente e é diretamente proporcional à concentração de carbono orgânico dissolvido na amostra. Os estudos de otimização foram realizados empregando soluções de EDTA. A recuperação média de carbono orgânico foi de 83% com desvio padrão relativo de 3,7% empregando 100 mL de amostra, pH 2,0, razão molar entre Fe2+ e H2O2 de 1:5 e uma vazão de 1 mL.min-1. Após otimização as concentrações de DOC foram quantificadas em 13 diferentes compostos orgânicos, com uma recuperação média de 90%. A velocidade de análise foi em média 50 amostras/h.

Huang, C. P.; Dong, C. and Tang, Z. (1993), Advanced chemical oxidation: its present role and potential future in hazardous waste treatment. Waste Manag. 13, 361-377.

Received: September 21, 2000;

Revised: April 16, 2001;

Accepted: August 03, 2001.

  • APHA-AWWA-WPCF (1995), Standard Methods for the Examination of Water and Wastewater, American Public Health Association, Washington, DC, 19th ed. ch. 5.
  • Bandara, J.; Morrison, C.; Kiwi, J.; Pulgarin, C. and Peringer, P. (1996), Degradation/decoloration of concentrated solution of Orange II. Kinetics and quantum yield for sunlight induced reaction via Fenton type reagents. J. Photochem. Photobiol. A-Chem., 99, 57-66.
  • Bauer, R. and Fallmann, H. (1997), The photo-Fenton oxidation- a cheap and efficient wastewater treatment method. Res. Chem. Intermed 23, 341-354.
  • Blanco, J.; Malato, S.; Milow, B.; I. Maldonado, M.; Fallmann, H.; Krutzler, T. and Bauer, R. (1999), Techno-economical assessment of solar detoxification systems with compound parabolic collectors. J. Phys., 9, 259-264.
  • Crittenden, J. C.; Hu, S.; Hand, D. W. and Green, S. A. (1999), A kinetic model for H2O2/UV process in a completely mixed batch reactor. Wat. Res 33, 2315-2328.
  • Doong, R. A. and Chang, W. H. (1998), Photoassisted iron compound catalytic degradation of organophosphorous pesticides with hydrogen peroxide. Chemosphere 37, 2563-2572.
  • Engwall, M. A.; Pignatello, J. J. and Grasso, D. (1999), Degradation and detoxification of the wood preservatives creasote and pentachlorophenol in water by the photo-Fenton reaction. Wat. Res 33, 1151-1158.
  • Fadini, P. S. (1995), Measuring dissolved organic carbon in natural samples by flow injection analysis. Master Thesis, Universidade Estadual de Campinas, São Paulo, Brazil.
  • Huston, P. L. and Pignatello, J. J. (1999), Degradation of selected pesticide active ingredients and commerciall formulations in water by the photo-assisted Fenton reaction. Wat. Res. 33, 1238-1246.
  • Jardim, W. F.; Moraes, S. G. and Takiyama, M. M. K. (1997), Photocatalytic degradation of aromatic chlorinated compounds using TiO2: toxicity of intermediates. Wat. Res 31, 1728-1732.
  • Jardim, W. F.; Pasquini, C.; Guimarães, J. R. and Faria, L. C. (1990), Short-term toxicity test using Escherichia coli: monitoring CO2 production by flow injection analysis. Wat. Res. 24, 351-354.
  • Kang, S. F.; Liao, C. H. and Hung, H. P. (1999), Peroxidation treatment of dye manufacturing wastewater in the presence of ultraviolet light and ferrous ions. J. Hazard. Mater 65, 317-333.
  • Kitis, M.; Adams, C. D. and Daigger, G. T. (1999), The effects of Fentons reagent pretreatment on the biodegradability of nonionic surfactants. Wat. Res 33, 2561-2568.
  • Krutzler, T. and Bauer, R. (1999), Optimization of a photo-Fenton prototype reactor. Chemosphere, 38, 2517-2532.
  • Kuo, C. Y. and Lo, S. L. (1999), Oxidation of aqueous chorobiphenyls with photo-Fenton process. Chemosphere, 38, 2041-2051.
  • Kuo, C. Y.; Lo, S. L. and Chan, M. T. (1998), Oxidation of aqueous chlorophenols with photo-Fenton process. J. Environ. Sci. Health Part B-Pestic. Contam. Agric. Wastes, 33, 723-747.
  • Legrini, O.; Oliveros, E. and Braun, A. M. (1993), Photochemical processes for water treatment. Chem. Rev 93, 671-698.
  • Liao, C. H.; Kang, S. F. and Hung, H.P. (1999), Simultaneous removal of COD and color from dye manufacturing process wastewater using photo-Fenton oxidation process. J. Environ. Sci. Health Part A-toxic/hazard. Subst. Environ. Eng. 34, 989-1012.
  • Lunar, L.; Sicilia, D.; Rubio, S.; Pérez-Bendito, D. and Nickel, U. (2000), Degradation of photographic developers by Fentons reagent: condition optimization and kinetics for metol oxidation. Wat. Res. 34, 1791-1802.
  • Maletzky, P. and Bauer, R. (1998), The photo-Fenton method degradation of nitrogen containing organic compounds. Chemosphere, 37, 899-909.
  • Matthews, R. W.; Abdullah, M. and Low, G. K. C. (1990), Photocatalytic oxidation for total organic carbon analysis. Analytica Chem. Acta, 233, 171-179.
  • Nogueira, R. F. P. and Guimarães, J. R. (2000), Photodegradation of dichloroacetic acid and 2,4-dichlorophenol by ferrioxalate/H2O2 system. Wat. Res., 34, 895-901.
  • Pignatello, J. J. (1992), Dark and photoassisted Fe3+ catalyzed degradation of chlorophenox herbicides by hydrogen peroxide. Environ. Sci. Technol. 26, 944-951.
  • Topudurti, K. V.; Lewis, N. M. and Hirsh, S. R. (1993), The applicability of UV/oxidation technologies to treat contaminated groundwater. Environ. Progress, 12, 54-60.
  • Venkatadri, R. and Peters, R. W. (1993), Chemical oxidation technologies: ultraviolet light/hydrogen peroxide, Fentons reagent, and titanium dioxide-assisted photocatalysis. Hazard. Waste Hazard. Materials, 10, 107-149.
  • Wu, K. Q.; Xie, Y. D.; Zhao, Z. C. and Hidaka, H. (1999), Photo-Fenton degradation of a dye under visible light irradiation. J. Mol. Catal. A-Chem 144, 77-84.
  • *
    Author for correspondence
  • Publication Dates

    • Publication in this collection
      07 Oct 2002
    • Date of issue
      Mar 2002

    History

    • Received
      21 Sept 2000
    • Reviewed
      16 Apr 2001
    • Accepted
      03 Aug 2001
    Instituto de Tecnologia do Paraná - Tecpar Rua Prof. Algacyr Munhoz Mader, 3775 - CIC, 81350-010 Curitiba PR Brazil, Tel.: +55 41 3316-3052/3054, Fax: +55 41 3346-2872 - Curitiba - PR - Brazil
    E-mail: babt@tecpar.br