A multicommuted flow procedure for the determination of total and free cholesterol in eggs and human blood serum by chemiluminescence

Abstracts

A multicommuted flow-based procedure with detection by chemiluminescence for the determination of total and free cholesterol without changes in the flow manifold is proposed. Cholesterol esterase and cholesterol oxidase were both immobilized on glass beads via glutaraldehyde/(3-aminopropyl)-triethoxysilane and mini-columns containing the enzymes were used for online sample treatment. Cholesterol esters were cleaved to cholesterol and fatty acids at the packed reactor containing cholesterol esterase. The reactor containing cholesterol oxidase converted cholesterol to cholest-4-en-3-one also yielding hydrogen peroxide. Detection was based on the chemiluminescence produced by H2O2 in the hexacyanoferrate(III)-luminol system. Influence of both chemical and hydrodynamic variables on the chemiluminescence signals were investigated. The analytical curves were linear from 250 to 2500 mg L-1 and from 500 to 4000 mg L-1, for free and total cholesterol, respectively. Detection limits for both analytes were estimated as 60 mg L-1 at 99.7% confidence level. The sampling rate was 55 h-1 and reagent consumption was 350 µg of luminol and 2.6 mg of potassium hexacyanoferrate(III) per determination. The procedure developed was successfully applied for determination of cholesterol in eggs and in human blood serum with results in agreement with the reference spectrophotometric method at the 95% confidence level.

flow analysis; multicommutation; chemiluminescence; cholesterol; cholesterol esterase; cholesterol oxidase


Um sistema de análises em fluxo com multicomutação com detecção por quimiluminescência foi desenvolvido para a determinação de colesterol livre e total sem alterações no módulo de análises. Colesterol estearase e colesterol oxidase foram imobilizadas em esferas de vidro via glutaraldeído/3-aminopropiletoxisilano e mini-colunas contendo as enzimas foram usadas para tratamento das amostras em linha. Ésteres de colesterol foram clivados a colesterol e ácidos graxos na coluna empacotada com colesterol estearase. No reator contendo colesterol oxidase, o analito foi convertido a 4-colesten-3-ona e peróxido de hidrogênio. A detecção foi baseada na quimiluminescência produzida por H2O2 no sistema hexacianoferrato(III)-luminol. Os efeitos de variáveis químicas e hidrodinâmicas sobre o sinal de quimiluminescência foram avaliados. As curvas analíticas foram lineares entre 250 e 2500 mg L-1 e 500 e 4000 mg L-1, para colesterol livre e total, respectivamente. Limites de detecção para ambos os analitos foram estimados em 60 mg L-1 (99,7% de confiança). A frequência de amostragem foi de 55 h-1 e o consumo de reagentes foi de 350 µg de luminol e 2,6 mg de hexacianoferrato(III) de potássio por determinação. O procedimento desenvolvido foi aplicado à determinação de colesterol em ovos e soro sanguíneo humano gerando resultados concordantes com o procedimento espectrofotométrico de referência com nível de confiança de 95%.


ARTICLE

A multicommuted flow procedure for the determination of total and free cholesterol in eggs and human blood serum by chemiluminescence

Oldair D. LeiteI; Heberth J. VieiraII; Orlando Fatibello-FilhoIII; Fábio R. P. RochaIV,* * e-mail: frprocha@cena.usp.br

IInstituto de Ciências Ambientais e Desenvolvimento Sustentável, Universidade Federal da Bahia, 47805-100 Barreiras-BA, Brazil

IIUniversidade Federal do Pampa, 96412-420 Bagé-RS, Brazil

IIIDepartamento de Química, Universidade Federal de São Carlos, CP 676, 13560-970 São Carlos-SP, Brazil

IVCentro de Energia Nuclear na Agricultura, Universidade de São Paulo, CP 96, 13400-970 Piracicaba-SP Brazil

ABSTRACT

A multicommuted flow-based procedure with detection by chemiluminescence for the determination of total and free cholesterol without changes in the flow manifold is proposed. Cholesterol esterase and cholesterol oxidase were both immobilized on glass beads via glutaraldehyde/(3-aminopropyl)-triethoxysilane and mini-columns containing the enzymes were used for online sample treatment. Cholesterol esters were cleaved to cholesterol and fatty acids at the packed reactor containing cholesterol esterase. The reactor containing cholesterol oxidase converted cholesterol to cholest-4-en-3-one also yielding hydrogen peroxide. Detection was based on the chemiluminescence produced by H2O2 in the hexacyanoferrate(III)-luminol system. Influence of both chemical and hydrodynamic variables on the chemiluminescence signals were investigated. The analytical curves were linear from 250 to 2500 mg L-1 and from 500 to 4000 mg L-1, for free and total cholesterol, respectively. Detection limits for both analytes were estimated as 60 mg L-1 at 99.7% confidence level. The sampling rate was 55 h-1 and reagent consumption was 350 µg of luminol and 2.6 mg of potassium hexacyanoferrate(III) per determination. The procedure developed was successfully applied for determination of cholesterol in eggs and in human blood serum with results in agreement with the reference spectrophotometric method at the 95% confidence level.

Keywords: flow analysis, multicommutation, chemiluminescence, cholesterol, cholesterol esterase, cholesterol oxidase

RESUMO

Um sistema de análises em fluxo com multicomutação com detecção por quimiluminescência foi desenvolvido para a determinação de colesterol livre e total sem alterações no módulo de análises. Colesterol estearase e colesterol oxidase foram imobilizadas em esferas de vidro via glutaraldeído/3-aminopropiletoxisilano e mini-colunas contendo as enzimas foram usadas para tratamento das amostras em linha. Ésteres de colesterol foram clivados a colesterol e ácidos graxos na coluna empacotada com colesterol estearase. No reator contendo colesterol oxidase, o analito foi convertido a 4-colesten-3-ona e peróxido de hidrogênio. A detecção foi baseada na quimiluminescência produzida por H2O2 no sistema hexacianoferrato(III)-luminol. Os efeitos de variáveis químicas e hidrodinâmicas sobre o sinal de quimiluminescência foram avaliados. As curvas analíticas foram lineares entre 250 e 2500 mg L-1 e 500 e 4000 mg L-1, para colesterol livre e total, respectivamente. Limites de detecção para ambos os analitos foram estimados em 60 mg L-1 (99,7% de confiança). A frequência de amostragem foi de 55 h-1 e o consumo de reagentes foi de 350 µg de luminol e 2,6 mg de hexacianoferrato(III) de potássio por determinação. O procedimento desenvolvido foi aplicado à determinação de colesterol em ovos e soro sanguíneo humano gerando resultados concordantes com o procedimento espectrofotométrico de referência com nível de confiança de 95%.

Introduction

Cholesterol is a constituent of animal foods such as eggs and meat1,2 and excessive dietary uptake may hazardously increase its concentration in blood.3,4 Although small amounts are essential for the synthesis of hormones, high levels in blood increase the risks of heart diseases. Saturated fats rise cholesterol levels in blood while unsaturated fats contribute to its decrease.

The clinical analysis of cholesterol in serum samples is important in the diagnosis and prevention of a large number of clinical disorders such as hypertension and arteriosclerosis, with a clear correlation between total cholesterol in human blood serum and heart diseases.5 Several methods for cholesterol assay in biological samples and foods have been reported, including detection by spectrophotometry,4,6 amperometry,7,8 fluorimetry,9 potentiometry10 and chemiluminescence.11-13 In a previous work focusing on the flow-injection determination of total cholesterol in animal blood serum with detection by chemiluminescence,11 a solid-state detector of a commercial spectrophotometer was used for the analysis of offline diluted samples, thus making the analysis tedious and time consuming for a large lot of samples. Some works have focused the determination of both free and total cholesterol exploiting optical14,15 or amperometric16,17 detection.

Traditional analytical uses of chemiluminescence (CL) involve reactions that produce intense radiation emission that can be monitored with simple equipments. The light intensity is usually measured by a photomultiplier tube (PMT),13 but simpler and cheaper detectors like a photodiode can be used as well.18,19 The use of a photodiode detector and immobilized enzymes in flow systems present some relevant advantages such as simplicity, reliability, low-cost and lower reagent consumption.18,20

Multicommutation is a powerful alternative to increase the versatility of flow-based procedures, in which discrete commuting devices are employed for solution handling.21 This approach also presents the advantage of minimizing both reagent consumption and waste generation.21,22 Coupling a multicommuted flow procedure with chemiluminescence detection for the determination of cholesterol is thus an interesting alternative.

This paper reports on a CL system for the determination of total and free cholesterol in foods and human blood serum, based on the production of hydrogen peroxide by reaction of the analytes catalyzed by immobilized enzymes cholesterol esterase (ChE) and cholesterol oxidase (ChOx). The produced hydrogen peroxide oxidizes luminol in alkaline medium under catalysis of hexacyanoferrate(III) to produce the CL signals. The proposed procedure makes feasible the fast determination of both total and free cholesterol, minimizing costs, reagent consumption and waste generation.

Experimental

Reagents and solutions

All chemicals were of analytical-grade and solutions were prepared in deionized water obtained from a Millipore (Bedford, MA) Milli-Q system (model UV Plus Ultra-Low Organics Water). Cholesterol oxidase (ChOx) and cholesterol esterase (ChE) were purchased from Sigma (St. Louis, USA).

A 0.4 mol L-1 carbonate buffer solution was prepared by dissolving 10.6 g of Na2CO3 in water, the pH was adjusted to 11.0 with concentrated hydrochloric acid and the volume was made up to 250 mL with water. This solution was stored in an amber flask in the refrigerator at 4 ºC.

A 1.0×10-2 mol L-1 luminol solution (R1) was prepared by dissolving 443.0 mg of 5-amino-2,3-dihydro-1,4-phthalazinedione (Sigma) in 0.4 mol L-1 carbonate buffer solution at pH 11.0 and the volume was made up to 250 mL with the same buffer solution. This solution was stored in a refrigerator at 4 ºC by three days before use.23

A 0.1 mol L-1 potassium hexacyanoferrate(III) solution (R2) was daily prepared by dissolving 3.290 g of the salt and making the volume to 100 mL with water. A 0.01 mol L-1 stock cholesterol solution was prepared in 0.05 mol L-1 phosphate buffer solution at pH 7.5 also containing 10% (m/m) of Triton X-100 and 10% (m/m) isopropanol. This solution was heated at 65 ºC in a thermostated water bath for dissolution of cholesterol and was stable for at least 10 days when stored at 4 ºC in the dark. Cholesterol reference solutions were prepared by dilution of the stock with 0.05 mol L-1 phosphate buffer solution. Solutions 4.5% (v/v) (3-aminopropyl)triethoxysilane and 2.5% (m/v) glutaraldehyde were prepared in water by dilution of concentrated reagents.

A commercial kit used for the enzymatic determination of cholesterol was purchased from Labtest Diagnóstica (Lagoa Santa, MG, Brazil) and used in the reference procedure.

Apparatus

The multicommuted flow system consisted of a peristaltic pump (Ismatec, IPC-12) equipped with TygonTM pumping tubes to propel the solutions, six three-way solenoid valves (NResearch, 161T031), polyethylene tubing (0.8 mm i.d.) and acrylic joint points. System control and data acquisition were performed with a Pentium 800 MHz microcomputer equipped with a commercial electronic interface (National Instruments) and a lab-made electronic circuit similar to that previously described.24 The software was developed in Labview 7.0 (National Instruments) running under Windows XP.

Chemiluminescence measurements were carried out with a lab-made luminometer previously described.19 A spiral flow cell made of 100 cm coiled polyethylene tubing (ca. 500 µL inner volume) was placed in front of a photodiode (OSD 50-E) of 100 mm2 sensitive area. The polyethylene tubing is transparent to the electromagnetic radiation emitted by the luminol/H2O2 system (maximum within 420 and 440 nm).The flow cell and photodiode were protected from ambient light using an acrylic dark box.

The determination of cholesterol using the commercial kit was carried out with a Femto 482 spectrophotometer (São Paulo, Brazil).

Enzyme immobilization procedures

Cholesterol esterase (ChE) and cholesterol oxidase (ChOx) enzymes were immobilized on controlled pore silica. The beads were initially prepared by silanization of silica with (3-aminopropyl)-triethoxysilane and activation of the support with glutaraldehyde. Enzymes were then covalently bonded to glutaraldehyde.25

Initially, 3 mL of 4.5% (v/v) (3-aminopropyl)triethoxysilane aqueous solution was added to a mass of 0.20 g of treated silica. Then, the inorganic support was coated with (aminoalkyl)silane for 40 min in a vacuum desiccator at 25 oC. After removing the excess of the solution, the modified silica was allowed to react with 5 mL of 2.5% (m/v) glutaraldehyde aqueous solution for 20 min. After washing the modified silica with deionized water and 0.05 mol L-1 phosphate buffer solution at pH 7.0, a mass of 25 mg (120 U) of ChOx was coupled with residual aldehyde groups in phosphate buffer solution for 24 h at 4 ºC. This coupling reaction was monitored by determination of enzymatic activities of supernatant after reaction times of 4, 8, 12, 16 and 24 h. For ChE, a mass of 3 mg (25 U) was coupled with 0.20 g of pre-activated inorganic support as described for ChOx. Simultaneously, control experiments were carried out using 0.20 g of silica without activation at the same experimental conditions. So with the difference between the enzymatic activities obtained in each experiment it was possible to determine the amount of each enzyme immobilized (units of ChE or ChOx per gram of silica). For this purpose the protein content in the initial enzyme solution was determined before and after immobilization using the procedure described by Hartree.26 After that, the inorganic supports containing immobilized ChE and ChOx were washed with a 0.05 mol L-1 phosphate buffer solution in order to remove the excess of enzyme.

A mass of 200 mg of each immobilized enzyme was packed into two glass tubes (20 mm × 3 mm i.d.) and coupled to the multicommuted flow system. The columns with immobilized enzymes were stored in the same phosphate buffer solution at 4 ºC when not in use.

Sample preparation

Samples of eggs were acquired at the local market. Egg yolks were ground to pass through a 40-mesh sieve. Extraction was carried out by direct saponification as previously reported.27 An accurate mass of 500 mg of sample was heated under reflux with 5 mL of 2% (m/v) KOH in absolute ethanol in a water bath at 50 ºC. The heating time was 2 h instead of 1 h as previously employed.27 After cooling, 10 mL of water was added to the mixture and the unsaponificable fractions were extracted twice with 10 mL of hexane. The organic phase was washed with 20 mL of water and the volatile solvent was evaporated under reduced pressure. The resulting residue was diluted to 5.0 mL with a 0.05 mol L-1 phosphate buffer solution also containing 10% (m/m) isopropanol and 10% (m/m) Triton X-100.

The human blood serum samples (2.0 mL) were obtained from a local clinical analysis laboratory, being analyzed as received.

Flow diagram

The multicommuted flow system shown in Figure 1 was operated according to the valves switching course in Table 1 for determination of free or total cholesterol. The manifold comprised six three-way solenoid valves: one valve for each solution managed (V1, V2, V5 and V6), an auxiliary valve for sampling (V3) and a valve employed for stream redirecting (V4). In the position showed in Figure 1, all valves were switched off and the carrier stream was propelled through V1-V2-V4 towards the flow cell while the reagent solutions R1 and R2 were recycled back to their vessels. In order to define the sample aliquots, valves V1, V2 and V3 were simultaneously switched on, the carrier stream was sent back to its vessel and a sample aliquot was aspirated through V2 and V3 valves during 7.5 s (free cholesterol determination) or 3.0 s (total cholesterol determination). For free cholesterol determination, valve V4 was switched on during steps 2-4 and the sample zone was by-passed to the reactor containing ChE. After a suitable delay time (8.0 s), valves V5 and V6 were switched on during 8.0 s and the sample zone received the reagents (R1 and R2) at the confluence point z, being transported to the flow cell to measure the emitted radiation. For total cholesterol determination, the sample aliquot was transported towards the reactors containing ChE and ChOx by the carrier stream by switching all the valves off. Other steps were similar to those employed for free cholesterol determination. The temperature of the laboratory was maintained around 25 ºC with an air conditioner. All measurements were based on peak height and carried out in triplicate.


Results and Discussion

General aspects

A flow procedure based on multicommutation and chemiluminescence detection for the determination of free and total cholesterol is proposed. The enzymes were immobilized on silica beads, a suitable support in view of inertia to microbial attacks, stability and easy handling. The immobilization efficiencies were estimated by monitoring the protein concentration in the supernatant before and after the immobilization procedure. Concentrations of cholesterol oxidase (ChOx) and cholesterol esterase (ChE) in the solid support were 1.21 mg protein per g (5.8 U) and 7.1 mg protein per g (11.8 U), respectively. These values agreed with those previously reported in the literature.28,29

For total cholesterol determination, the hydrolysis of cholesterol ester by cholesterol esterase was the initial step (equation 1). The free cholesterol produced in this step was oxidized to cholest-4-en-3-one by cholesterol oxidase, with concomitant production of hydrogen peroxide (equation 2). The hydrogen peroxide produced was then quantified by oxidation of luminol in the presence of hexacyanoferrate(III) producing chemiluminescence (equation 3). Thus, the CL emission is proportional to the cholesterol concentration in the samples.

System optimization

Chemical and hydrodynamic variables were evaluated to establish the best conditions for determination of both analytes. The effect of sample volume under the CL intensity for determination of free cholesterol was studied in the range from 50 to 150 µL, by varying the sampling time (Table 1, step 1) with a constant sample flow-rate (1 mL min-1). In this study, the sample loop and the delay time were fixed in 30 cm and 8.0 s, respectively. The analytical signal increased linearly with the sample volume until 125 µL, in view of the lower sample dispersion, keeping almost constant for higher volumes. In this way, a sample volume of 125 µL was selected for further experiments, being the sample loop reduced to 25 cm. The aspiration was performed during 7.5 s, time enough to completely fill the sample loop. On the other hand, taking into account the concentration of total cholesterol expected in the samples, the sample volume was fixed in 50 µL for determination of this analyte. Time-based sampling was exploited in this step (3.0 s sampling time) to partially fill the sample loop. Sample dispersion in the sample loop (L) and the reactor coil (B) were adjusted to avoid previous sample dilution for determination of this analyte. If necessary, sample dilution could be carried out by the zone sampling approach, by directing part of the dispersed sample zone to waste through valve V3.30

The carrier flow rate affects sample residence time in the enzymatic columns (and thus the conversion rate) as well as the dilution by the reagents at the confluence point z. On the other hand, low flow rates of the carrier imply in longer delay times (Table 1, step 2) to assure reagent addition in the center of the sample zone. The effects of these variables were then simultaneously evaluated for determination of free and total cholesterol (Figure 2). For free cholesterol, equivalent CL signals were observed for carrier flow rates of 2.4 mL min-1 or 3.0 mL min-1 using a 10 s delay time. The effect of residence time in the enzymatic column was then less critical than the dilution at the confluence point, indicating that the conversion of cholesterol to cholest-4-en-3-one and hydrogen peroxide (equation 2) is fast. This was also observed for total cholesterol and in spite of the highest response for a carrier flow rate of 1.8 mL min-1 and 14 s delay time, a flow-rate of 2.4 mL min-1 and a delay time of 10 s were selected taking into account the determination of both analytes using the same flow manifold. This compromise condition did not significantly hinder the analytical signal, which was only ca. 7% lower than the obtained in the optimum condition.


The effects of the other variables that affect the analytical response were evaluated employing a 500 mg L-1 cholesterol solution (Figure 3). Luminol concentration was investigated in the range from 1.0×10-3 to 7.0×10-3 mol L-1 (Figure 3a). The CL emission increased with luminol concentration until 5.0×10-3, keeping constant for higher values in view of the self-absorption in excess of the reagent. Therefore, a 5.0×10-3 mol L-1 luminol solution was used in subsequent work. The analytical signal increased with the hexacyanoferrate(III) concentration up to 20×10-3 mol L-1 and then slightly decreased (Figure 3b). This effect was caused by the absorption of the CL radiation by the colored hexacyanoferrate(III) solution, which did not compensate the catalytic effect. Then, the 20×10-3 mol L-1 concentration was selected for the proposed flow procedure.


 




The influence of pH on the enzymatic reaction was investigated in the range from 6.5 to 8.0 employing 0.05 mol L-1 phosphate buffer solutions (Figure 3c). It was observed that maximum CL emission was achieved in the pH range from 7.0 to 7.5, in agreement with previous works. Thus, pH 7.0 was selected for the proposed flow procedure. As a 0.05 mol L-1 concentration was enough to keep the pH constant, solutions with higher buffer capacity were not evaluated.

In addition to the solution concentration and flow rates, the reagent amount in the sample zone also depends on the switching times of the corresponding valves. The effect was simultaneously evaluated for both reagents in the range 5.0-10 s, employing solutions flow-rates of 3.0 mL min-1 (Figure 3d). It was observed that the CL emission increased up to switching times of 8.0 s for both reagents, remaining constant to upper times. The volume of reagents employed under these conditions was 400 µL, which corresponds to 2.64 mg of potassium hexacyanoferrate(III) and 0.35 mg of luminol, respectively. The selected parameters are shown in Table 2.

Effect of concomitant species

Considering the application of the proposed flow procedure, the effect of several species, such as glucose and Na+ (1000 mg L-1), urea (200 mg L-1), uric acid, K+ and Ca2+ (100 mg L-1), ascorbic acid (0.1 mg L-1), Cu2+ and Zn2+ (0.01 mg L-1) and Fe2+ (0.005 mg L-1), was evaluated on the analytical signal for 2000 mg L-1 cholesterol. Ascorbic and uric acids caused an inhibition of 11 and 12%, respectively, at concentrations above the physiological.31 Despite Zn2+ ions caused an enhancement of CL emission (ca. 17%) when in the concentration of 0.01 mg L-1, the interference was not observed in the analysis of real samples, according to the results presented in the next section. This agreed with previous studies,32 in which the catalytic effect of zinc on the luminol oxidation was not observed in the presence of excess of another catalyst as employed in this work. Other species did not cause any interference on the determination of cholesterol.

Analytical characteristics and application

Transient signals for determination of free and total cholesterol, shown in Figure 4, demonstrate stable baselines and the repeatability of the results. The analytical response for free cholesterol determination was linear in the concentration range from 250 to 2500 mg L-1 and can be described by the equation: I = 43.0 + 0.325 × CChol (r = 0.999), in which I is the CL emission intensity in mV and CChol is the cholesterol concentration in mg L-1. For total cholesterol determination, the analytical response was linear in the range from 500 to 4000 mg L-1, being described by the equation I = 40.0 + 0.105 × CChol, r = 0.998. Coefficient of variation of 2.9% (n = 10) and a sampling throughput of 55 determinations h-1 were estimated. The detection limit was estimated as 60 mg L-1 at the 99.7% confidence level.


After a 6 h working period, no baseline drift was observed and only a slight variation of 5.6% of the response for 2500 mg L-1 cholesterol solution was observed. After 15 days of use of each enzymatic reactor, the sensitivity decreased to 80% of its initial value. During this period, at least 400 assays were performed with each reactor.

Accuracy of the proposed procedure was assessed by determining free cholesterol in eggs yolk and total cholesterol in human blood serum samples. The obtained results were compared with those provided by a conventional spectrophotometric procedure using a commercial enzymatic kit (Table 3). Results obtained employing both procedures were in agreement at the 95% confidence level, with an acceptable range of relative error (lower than 3%).

Analytical features of the proposed procedure were comparable with those achieved in previous works for determination of free and total cholesterol (Table 4).14-17 However, the multicommuted flow system allowed the determination of both analytes without system reconfiguration or previous sample dilution. The reagent consumption and waste generation were also minimized in view of the intermittent addition of the reagents and use of immobilized enzymes in a large number of assays.

Conclusions

The proposed flow procedure is robust and easily operated and has a great potential to clinical analysis. The procedure was implemented with inexpensive instrumentation by exploiting the multicommutation approach and a lab-made luminometer based on a photodiode. Low reagent consumptions and waste generation are other additional advantages. The method is simple, rapid and inexpensive and was successfully applied for the determination of free and total cholesterol in food and human blood serum samples.

Acknowledgments

The authors gratefully acknowledge fellowships and financial support from the Brazilian agencies FAPESP, CNPq and CAPES.

Submitted: January 14, 2010

Published online: May 27, 2010

FAPESP has sponsored the publication of this article.

  • 1. Bohac, C. E.; Rhee, K. S.; Cross, H. R.; Ono, K.; J. Food Sci. 1988, 53, 1642.
  • 2. Nogueira, G. C.; Bragagnolo, N.; Food Chem. 2002, 79, 267.
  • 3. Amundson, M. D.; Mingjie, Z.; J. Biochem. Biophys. Methods 1999, 38, 43.
  • 4. Krug, A.; Gobel, R.; Kellner, R.; Anal. Chim. Acta 1994, 287, 59.
  • 5. Cleeman, J. I.; Lenfant, C.; JAMA, J. Am. Med. Assoc. 1998, 280, 2099.
  • 6. Araújo, A. N.; Catita, J. A. M.; Lima, J. L. F. C.; Farmaco 1999, 54, 51.
  • 7. Nakaminami, T.; Kuwabata, S.; Yoneyama, H.; Anal. Chem 1999, 71, 1068.
  • 8. Trettnak, W.; Lionti, I.; Mascini, M.; Electroanalysis 1993, 5, 753.
  • 9. Amundson, D. M.; Zhou, M.; J. Biochem. Biophys. Methods 1999, 38, 43.
  • 10. Situmorang, M.; Alexander, P. W.; Hibbert, D. B.; Talanta 1999, 49, 639.
  • 11. Pires, C. K.; Reis, B. F.; Galhardo, C. X.; Martelli, P. B.; Anal. Lett. 2003, 36, 3011.
  • 12. Nabi, A.; Rashid, A.; Yaqoob, M.; Anal. Lett. 1996, 29, 2281.
  • 13. Huang, Y.; Zhang, C.; Zhang, Z.; Anal. Sci. 1999, 15, 867.
  • 14. Pineiro-Avila, G.; Salvador, A.; de la Guardia, M.; Analyst 1998, 123, 999.
  • 15. Krug, A.; Suleiman, A. A.; Guilbault, G. G.; Kellner, R.; Enzyme Microb. Technol. 1992, 14, 313.
  • 16. Yao, T.; Sato, M.; Kobayashi, Y.; Wasa, T.; Anal. Biochem. 1985, 149, 387.
  • 17. Adányi, N.; Váradi, M.; Eur. Food Res. Technol. 2003, 218, 99.
  • 18. Borges, E. P.; Fernandes, E. N.; Rocha, F. R. P.; Reis, B. F.; Quim. Nova 2002, 25, 1191.
  • 19. Leite, O. D.; Fatibello-Filho, O.; Vieira, H. J.; Rocha, F. R. P.; Cury, N. S. M.; Anal. Lett. 2007, 40, 3148.
  • 20. Rocha, F. R. P.; Ródenas-Torralba, E.; Reis, B. F.; Morales-Rúbio, A.; de la Guardia, M.; Talanta 2005, 67, 673.
  • 21. Rocha, F. R. P.; Reis, B. F.; Zagatto, E. A. G.; Lima, J. L. F. C.; Lapa, R. A.; Santos, J. L. M.; Anal. Chim. Acta 2002, 468, 119.
  • 22. Rocha, F. R. P.; Nóbrega, J. A.; Fatibello-Filho, O.; Green Chem. 2001, 3, 216.
  • 23. Kiba, N.; Ito, S.; Tachibana, M.; Tani, K.; Koizumi, H.; Anal. Sci. 2001, 17, 929.
  • 24. Ródenas-Torralba, E.; Rocha, F. R. P.; Reis, B. F.; Morales-Rubio, A.; de la Guardia, M.; J. Autom. Methods Manage. Chem. 2006, art. No. 20384.
  • 25. Weetall, H. H. In Methods in Enzymology; Mosbach, K., ed., vol. XLIV, Academic Press: New York, 1975, p. 135.
  • 26. Hartree, E. S.; Anal. Biochem. 1972, 48, 422.
  • 27. Jiang, Z.; Fenton, M.; Sim, J. S.; Poultry Sci. 1991, 70, 1015.
  • 28. Tabata, M; Endo, J.; Murachi, T; J. Appl. Biochem. 1981, 3, 84.
  • 29. Piñeiro-Avila, G.; Salvador, A.; de la Guardia, M.; Analyst 1998, 123, 999.
  • 30. Araújo, A. N.; Lima, J. L. F. C.; Reis, B. F.; Zagatto, E. A. G.; Anal. Chim. Acta 1995, 310, 447.
  • 31. Bishop, M. L.; Fody, E. P.; Schoeff, L.; Clinical Chemistry: Principles, Procedures and Correlations, Lippincott Willians & Wilkins: Philadelphia, 2005.
  • 32. Hu, Y.; Zhang, Z.; Luminescence 2008, 23, 338.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      10 Sept 2010
    • Date of issue
      2010

    History

    • Received
      14 Jan 2010
    • Accepted
      27 May 2010
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br