Intracardiac embolization of inferior vena cava filter associated with right atrium perforation and cardiac tamponade

Embolização de filtro de veia cava associado à perfuração de átrio direito e tamponamento cardíaco

Alexandre de Matos Soeiro, MD

Felipe Lourenço Fernandes, MD

Rafael Plens Teixeira, MD, MsC, PhD

Pedro Felipe Gomez Nicz, MD, PhD

Maria Carolina Feres de Almeida Soeiro

Carlos V. Serrano Jr., MD, PhD

Múcio Tavares de Oliveira Jr., MD, PhD

Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo (INCOR-HCFMUSP), São Paulo, SP, Brazil

ABSTRACT

Insertion of inferior vena cava filters has been well established in literature, reducing occurrence of pulmonary embolism after an episode of deep venous thrombosis in patients with contraindication to anticoagulation. There are a small number of complications related to procedure and embolization is rare. In this context, we described a case of intracardiac embolization associated with cardiac tamponade.

Key words: Vena Cava Filters; Vena Cava, inferior; Cardiac Tamponade

RESUMO

A colocação percutânea de filtro de veia cava inferior é bem estabelecida em literatura, reduzindo a chance de ocorrência de tromboembolismo pulmonar após um episódio de trombose venosa profunda em pacientes com contraindicações à anticoagulação. O índice de complicações relacionadas ao procedimento é pequeno, sendo embolização evento extremamente raro. Nesse contexto, descrevemos um caso de embolização intracardíaca de filtro de veia cava inferior associada ao tamponamento cardíaco.

Palavras-Chave: Veia Cava Inferior; Tamponamento Cardíaco; Filtros de Veia Cava

Abbreviations, acronyms & symbols
PE Pulmonary embolism
DVT Deep vein thrombosis
ICVF Inferior vena cava filter

INTRODUCTION

The treatment of choice for prevention of pulmonary embolism (PE) secondary to deep vein thrombosis (DVT) is anticoagulation; however, some patients may have contraindications to it. Without treatment, the chance of developing PE after an episode of DVT is approximately 40%[1]. Thus, for patients with contraindications to anticoagulation, there is the option of placing an Inferior Vena Cava Filter (ICVF) in an attempt to prevent PE[2]. The idea of implementing a filter to prevent clots reaching the pulmonary arteries was first described in 1958 by Hunter and DeWeese[3]. From this time until the present day, the models and placement techniques have evolved considerably, however the ICVF is not free of risks, and device-related complications can occur, e.g. in situ thrombosis (0-25%), occlusion of the ICVF (2-30%), recurrence of PE (0.5-6%), filter fracture (2-10%) and vena cava filter embolization (< 1%)[4,5].

Embolization of the filter to the heart or pulmonary artery is an extremely rare complication that can lead to symptoms ranging from hypotension, syncope, dyspnea or even cardiac arrest as the first manifestation[2,6]. The cause of embolization of the ICVF can vary greatly, and includes: placing the device in an unsuitable location, device failure and a vena cava with an enlarged diameter (>28 mm)[6].

In case of intracardiac embolization, therapeutic options involve surgical or endovascular removal or even maintenance of the vena cava filter in the same place[5], however the best treatment option is under discussion. Many authors suggested that the removal by open thoracotomy would be the best option[2].

Below we report a case of embolization of a vena cava filter in a patient treated at the Heart Institute in Sao Paulo.

CASE REPORT

A 47-years-old male who had a history of epilepsy was admitted to our service due to dyspnea and chest pain, that was worse during inspiration. Forty-five days prior to this admission he had been submitted to a subdural hematoma surgical drainage secondary to a same height fall. Following the procedure the patient developed DVT. Since he had clinical contraindication to anticoagulation,insertion of an ICVF was performed.

At entrance physical examination he was prostrated, pale and disoriented. Had a heart rate of 114 bpm, blood pressure of 108/80mmHg, weak pulse and capillary refill of less than 3 seconds. Cardiac examination showed hypophonetic and rhythmic heart sounds without murmurs. Pulmonary examination had diminished breath sounds at lung bases, a respiratory rate of 16 breaths per minute and an oxygen saturation of 94%. Extremities examination demonstrated an asymmetrical bilateral edema, that was larger in the left limb.

Electrocardiogram was normal. At initial laboratory studies he had hemoglobin of 9.7 g/dL, hematocrit of 31%, leukocytes 10.360/mm3, 351.000 platelets, CK-MB 1.12 ng/mL, troponin I 0.288 ng/ mL, urea 19mg/dL, creatinine 0.86 mg/dL,C-reactive protein (CRP) 272.22 mg/dL.

As the initial hypothesis was PE, the patient underwent CT angiography of the pulmonary arteries. Instead of showing signs of PE, CT showed an ICVF dislocation. It had moved to the right atrium. There was also right-sided pleural effusion and pericardial effusion (Figure 1).

Fig. 1 Tomography with the image of a vena cava filter located in transition of right atrium and right ventricle, before (A) and after (B) pericardial drainage 

Transthoracic echocardiogram showed signs of cardiac tamponade. In such way, emergent surgical pleural and pericardial drainage was performed.

After the procedure, CT and echocardiogram still showed moderate pericardial effusion, now with no signs of cardiac tamponade and a foreign body into the right atrium. In this moment the patient developed an early nosocomial infection thereby delaying the withdrawal of ICVF performed by open thoracotomy. During the procedure we observed that the ICVF had perforated the free wall of the right atrium (Figure 2).

Fig. 2 Inferior vena cava filter perforating the free wall of the right atrium 

Postoperatively, the patient developed septic shock, resolved after 33 days of antibiotics (Meropenem, Colistimethate, Linezolid, Fluconazole and amikacin at different times). He was discharged in good general condition, without anticoagulation due to the poor patient compliance and because the risk factor for DVT and PE, that was immobilization, had been resolved. After two months, he presented a new episode of DVT in the left medial gastrocnemius vein, at this time the anticoagulation therapy with warfarin was introduced.

DISCUSSION

Embolization of the IVCF to the heart or pulmonary artery is an extremely rare complication that can lead to symptoms ranging from hypotension, syncope, dyspnea or even cardiac arrest as the first manifestation[2,6]. The cause of embolization of the ICVF can vary greatly, and includes: placing the device in an unsuitable location, device failure and a vena cava with an enlarged diameter (>28 mm)[4-14].

In a series published by Ferris et al.[5], in which 324 patients had undergone to ICVF implantation, there were no cases of distal embolization. Similarly, another published case series with respect to 26 years of ICVF implants in 1731 patients at a center in the United States showed a complication rate of 0.3%, a rate of PE after the procedure of only 5.6% and no heart embolization.

The first description of cardiac tamponade associated with ICVF embolization was performed by Lahey et al.[13] in 1991. In this first published report, the authors associate the displacement of ICVF to its improper placement, with no evidence of fracture. As in our case, the patient had the IVCF removed by open surgery.

Differently to this first description and our case, the majority of reports in the literature with cardiac tamponade described the presence of fractured IVCF and intracardiac embolization of only parts of the filter[7-10]. Chandra et al.[8] reported there was no any well-defined correlation or factor associated with increased likelihood of intracardiac device embolization. Furthermore, the IVCF brands are variable and not necessarily associated with this complication[8,14].

Nicholson et al.[14] performed a single-center study in 2004, in which they published a series of 80 cases of patients who had IVCF placed. There were 5 cases of intracardiacembolization after filter fracture, and they were associated to two brands: BARD RECOVERY and BARD G2. Three of those patients had cardiac tamponade and one resulted in death. In all cases, the extraction was performed by open surgery[14].

Regarding the observed time between the IVCF placement and embolization, the literature is also inconclusive. Hussain et al.[11] described a case of embolization and cardiac tamponade after only four hours of filter implantation. However, the placement was performed in the subclavian vein, unlike our case. In another hand, most of the published series showed cases where embolization occurred months after the filter placement, usually associated with device fracture[7-10].

The therapeutic options in cases of cardiac embolization involve surgical removal, endovascular treatment or even conservative approach[3,7-14]. Many authors suggest that the withdrawal by open thoracotomy would be the best option[1,7,8,10-14]. Controversially, Vergara et al.[9] described a case of embolization associated with cardiac tamponade, in which the ICVF was removed percutaneously, suggesting that in selected patients, this is a plausible alternative to open surgery.

In our case, we opted for open surgical removal due to association with perforation and cardiac tamponade. However, the best treatment option is still uncertain due to the rarity of cases and it should be individualized[1].

We could not find in Brazilian literature descriptions of cases of IVCF embolization associated with cardiac tamponade. This makes this case unique nationwide, inferring the clinician to consider this adverse outcome in the indication of the procedure and as a complication of its placement[2,3].

CONCLUSION

Although embolization of the IVCF to the heart or pulmonary artery is an extremely rare complication (0.1-1.2%), clinicians should be aware of IVCF complications, in order to optimize the time-to-diagnosis and prevent further device-related complications[7,8].

Authors’ roles & responsibilities
AMS Analysis and interpretation of data; final approval of the manuscript; conception and design of the study; writing of the manuscript or revising it critically for its content
FLF Analysis and interpretation of data; final approval of the manuscript; writing of the manuscript or revising it critically for its content
RPT Final approval of the manuscript; writing of the manuscript or revising it critically for its content
PFGN Final approval of the manuscript; conception and design of the study; writing of the manuscript or revising it critically for its content
MCFAS Final approval of the manuscript; conception and design of the study; writing of the manuscript or revising it critically for its content
CVSJ Final approval of the manuscript; conception and design of the study; writing of the manuscript or revising it critically for its content
MTOJ Final approval of the manuscript; conception and design of the study; writing of the manuscript or revising it critically for its content

Work carried out at the Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo (INCOR-HCFMUSP), São Paulo, SP, Brazil.

No financial support.

REFERENCES

1. Sella DM, Oldenburg WA. Complications of inferior vena cava filters. Semin Vasc Surg. 2013;26(1):23-8. [ Links ]

2. Caplin DM, Nikolic B, Kalva SP, Ganguli S, Saad WE, Zuckerman DA; Society of Interventional Radiology Standards of Practice Committee. Quality improvement guidelines for performance of inferior vena cava filter placement for the prevention of pulmonary embolism. J Vasc Interv Radiol. 2011;22(11):1499-506. [ Links ]

3. Weinberg I, Kaufman J, Jaff MR. Inferior vena cava filters. JACC Cardiol Interv. 2013;6(6):539-47. [ Links ]

4. Owens CA, Bui JT, Knuttinen MG, Gaba RC, Carrillo TC, Hoefling N, et al. Intracardiac migration of inferior vena cava filters: review of published data. Chest. 2009;136(3):877-87. [ Links ]

5. Athanasoulis CA, Kaufman JA, Halpern EF, Waltman AC, Geller SC, Fan CM. Inferior vena cava filters: review of 26-year single-center clinical experience. Radiology. 2000;216(1):54-66. [ Links ]

6. Ferris EJ, McCowan TC, Carver DK, McFarland DR. Percutaneous inferior vena caval filters: follow-up of seven designs in 320 patients. Radiology. 1993;188(3):851-6. [ Links ]

7. Rogers NA, Nguyen L, Minniefield NE, Jessen ME, de Lemos JA. Fracture and embolization of an inferior vena cava filter strut leading to cardiac tamponade. Circulation. 2009;119(18):2535-6. [ Links ]

8. Chandra PA, Nwokolo C, Chuprun D, Chandra AB. Cardiac tamponade caused by fracture and migration of inferior vena cava filter. South Med J. 2008;101(11):1163-4. [ Links ]

9. Vergara GR, Wallace WF, Bennett KR. Spontaneous migration of an inferior vena cava filter resulting in cardiac tamponade and percutaneous filter retrieval. Catheter Cardiovasc Interv. 2007;69(2):300-2. [ Links ]

10. Saeed I, Garcia M, McNicholas K. Right ventricular migration of a recovery IVC filter's fractured wire with subsequent pericardial tamponade. Cardiovasc Intervent Radiol. 2006;29(4):685-6. [ Links ]

11. Hussain SM, McLafferty RB, Schmittling ZC, Zakaria AM, Ramsey DE, Larson JL, et al. Superior vena cava perforation and cardiac tamponade after filter placement in the superior vena cava--a case report. Vasc Endovascular Surg. 2005;39(4):367-70. [ Links ]

12. Bolton JW, Aldea GS. Perforation of the heart by an inferior vena cava filter. Mil Med. 1994;159(4):360-2. [ Links ]

13. Lahey SJ, Meyer LP, Karchmer AW, Cronin J, Czorniak M, Maggs PR, et al. Misplaced caval filter and subsequent pericardial tamponade. Ann Thorac Surg. 1991;51(2):299-300. [ Links ]

14. Nicholson W, Nicholson WJ, Tolerico P, Taylor B, Solomon S, Schryver T, et al. Prevalence of fracture and fragment embolization of Bard retrievable vena cava filters and clinical implications including cardiac perforation and tamponade. Arch Intern Med. 2010;170(20):1827-31. [ Links ]

Recebido: 30 de Dezembro de 2013; Aceito: 24 de Março de 2014

Correspondence address: Alexandre de Matos Soeiro, Instituto do Coração do Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo, Unidade Clínica de Emergência, Av. Dr. Eneas de Carvalho Aguiar, 44 - Cerqueira César, São Paulo, SP, Brazil - Zip code: 05403-900. E-mail: alexandre.soeiro@bol.com.br

Creative Commons License This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.