U-Pb dating by zircon dissolution method using chemical abrasion

Abstracts

Chemical abrasion was carried out on zircons grains of the Temora II standard for U-Pb dating prior to analyses using in situ Laser Ablation-MultiCollector Ion Coupled Plasma Mass Spectrometer (LA-ICPMS) followed by the Isotope Dissolution Thermal Ionization Mass Spectrometer (ID-TIMS) method. The proposed methodology was herein applied in order to reduce primarily the effects of secondary Pb loss, the presence of common lead and/or silicate impurities. Nine Temora II zircon grains were analyzed by the laser ablation method yielding an age of 418.3±4.3 Ma. Zircon grains of a same population were separated for chemical abrasion before dissolution and mass spectrometry analyses. Six fractions of them were separated for isotope dissolution using 235U-205Pb mixed spike after we have checked and assured the laboratory conditions of low blank values for total Pb of less than 2 pg/g. The obtained U-Pb zircon age by the ID-TIMS method was 415.7±1.8 Ma (error 0.43 %) based on four successful determinations. The results are consistent with the published ages for the Temora diorite (Temora I – 416.75±1.3 Ma; Temora II – 416.78±0.33 Ma) and established as 416±0.33 Ma. The technique is thus recommended for high precision U-Pb zircon analyses (error < 1 %), mainly for high resolution stratigraphic studies of Phanerozoic sequences.

chemical abrasion; ID-TIMS; LA-MC-ICPMS; U-Pb dating; zircon dissolution


Este trabalho apresenta a técnica de abrasão química em zircões do padrão Temora II aplicada em datação de U-Pb por Dissolução Isotópica e Espectrometria de Massa por Ionização Térmica, método DI-ESIT. O emprego deste método tem como princípio diminuir os efeitos da perda secundária de Pb radiogênico antes da dissolução. Em primeira instância, os zircões foram analisados pelo método in situ com uso de Microssonda Laser acoplada a um Espectrômetro de Massa Multicoletor com Plasma Acoplado Indutivamente (MC-ICP-MS), cujo objetivo foi identificar os grãos de zircão de fase de cristalização simples sem sobrecrescimento. Nove cristais de zircão analisados com microssonda a laser forneceram uma idade 418,3±4,3 Ma. Da mesma população foram separados cristais de zircão para datação utilizando a técnica de abrasão química e posterior análise pelo método DI-ESIT. Seis amostras de zircão foram separadas para diluição isotópica com adição de um traçador isotópico combinado U235-Pb205, em condições laboratoriais de brancos analíticos de Pb total menor que 2 pg/g. A idade U-Pb em zircão obtida foi 415,7±1,8 Ma (erro 0,43 %) com base em cinco amostras de zircão. Os resultados são concordantes com aquelas idades publicadas para o diorito Temora (Temora I – 416,75±1,3 Ma; Temora II – 416,78±0,33 Ma). A técnica descrita aqui é, portanto recomendada para ser aplicada em análises isotópicas de U-Pb com alta precisão (erros menor que 1%) para estudos de estratigrafia de alta resolução de sequências fanerozóicas.

abrasão química; ID-TIMS; LA-MC-ICPMS; datação U-Pb; dissolução de zircão


U-Pb dating by zircon dissolution method using chemical abrasion

Lucy TakeharaI; Farid Chemale JúniorII; Léo A. HartmannIII; Ivo A. DussinIV; Koji KawashitaIV

IServiço Geológico do Brasil – CPRM, SGAN 603, Conjunto "J" . Parte A , 1° andar, 70830-030, Brasília, DF, Brasil

IILaboratório de Geocronologia, Instituto de Geociências, Universidade de Brasília, Campus Darcy Ribeiro, s/n, 70904-970, Brasília, DF, Brasil

IIIInstituto de Geociências, Universidade Federal do Rio Grande do Sul, Av. Bento Gonçalves, 9500, Bairro Agronomia, 91501-970, Porto Alegre, RS, Brasil

IVCentro de Pesquisa Geocronológicas, Universidade de São Paulo, Rua do Lago, 562, 05508-080, São Paulo, SP, Brasil

Correspondence Correspondence to: Lucy Takehara E-mail: lucytakehara@gmail.com.br

ABSTRACT

Chemical abrasion was carried out on zircons grains of the Temora II standard for U-Pb dating prior to analyses using in situ Laser Ablation-MultiCollector Ion Coupled Plasma Mass Spectrometer (LA-ICPMS) followed by the Isotope Dissolution Thermal Ionization Mass Spectrometer (ID-TIMS) method. The proposed methodology was herein applied in order to reduce primarily the effects of secondary Pb loss, the presence of common lead and/or silicate impurities. Nine Temora II zircon grains were analyzed by the laser ablation method yielding an age of 418.3±4.3 Ma. Zircon grains of a same population were separated for chemical abrasion before dissolution and mass spectrometry analyses. Six fractions of them were separated for isotope dissolution using 235U-205Pb mixed spike after we have checked and assured the laboratory conditions of low blank values for total Pb of less than 2 pg/g. The obtained U-Pb zircon age by the ID-TIMS method was 415.7±1.8 Ma (error 0.43 %) based on four successful determinations. The results are consistent with the published ages for the Temora diorite (Temora I – 416.75±1.3 Ma; Temora II – 416.78±0.33 Ma) and established as 416±0.33 Ma. The technique is thus recommended for high precision U-Pb zircon analyses (error < 1 %), mainly for high resolution stratigraphic studies of Phanerozoic sequences.

Key words: chemical abrasion, ID-TIMS, LA-MC-ICPMS, U-Pb dating, zircon dissolution.

RESUMO

Este trabalho apresenta a técnica de abrasão química em zircões do padrão Temora II aplicada em datação de U-Pb por Dissolução Isotópica e Espectrometria de Massa por Ionização Térmica, método DI-ESIT. O emprego deste método tem como princípio diminuir os efeitos da perda secundária de Pb radiogênico antes da dissolução. Em primeira instância, os zircões foram analisados pelo método in situ com uso de Microssonda Laser acoplada a um Espectrômetro de Massa Multicoletor com Plasma Acoplado Indutivamente (MC-ICP-MS), cujo objetivo foi identificar os grãos de zircão de fase de cristalização simples sem sobrecrescimento. Nove cristais de zircão analisados com microssonda a laser forneceram uma idade 418,3±4,3 Ma. Da mesma população foram separados cristais de zircão para datação utilizando a técnica de abrasão química e posterior análise pelo método DI-ESIT. Seis amostras de zircão foram separadas para diluição isotópica com adição de um traçador isotópico combinado U235-Pb205, em condições laboratoriais de brancos analíticos de Pb total menor que 2 pg/g. A idade U-Pb em zircão obtida foi 415,7±1,8 Ma (erro 0,43 %) com base em cinco amostras de zircão. Os resultados são concordantes com aquelas idades publicadas para o diorito Temora (Temora I – 416,75±1,3 Ma; Temora II – 416,78±0,33 Ma). A técnica descrita aqui é, portanto recomendada para ser aplicada em análises isotópicas de U-Pb com alta precisão (erros menor que 1%) para estudos de estratigrafia de alta resolução de sequências fanerozóicas.

Palavras-chave: abrasão química, ID-TIMS, LA-MC-ICPMS, datação U-Pb, dissolução de zircão.

INTRODUCTION

The age determinations of geological events are a significant tool in basic and applied earth science studies. In the last two decades, the in situ, high resolution U-Pb dating by SHRIMP (Sensitive High Resolution Ion Microprobe) (Williams 1998, Rasmussen 2005) and LA-ICPMS (Laser Ablation Ion Coupled Plasma Mass Spectrometer) (Kosler et al. 2001, Jackson et al. 2004, Pickhardt et al. 2005) has brought new insights into the understanding of crystallization ages of minerals, especially zircon. The complexity of geological processes registered in the minerals can be dated with these techniques. These solid state, in situ determinations are useful to geoscientists because they can identify the timing of events such as rock generation, metamorphism and deformation in the crust and mantle.

Despite the great advance in U-Pb in situ dating with SHRIMP and LA-ICPMS methods, in some cases these techniques do not attain the required accuracy to date geologic events that occurred at the end of the Precambrian and in the beginning of the Phanerozoic. In these cases, more precise techniques are required, such as the isotopic dissolution (ID) methods in minerals using Thermal Ionization Mass Spectrometer (ID-TIMS). However, the total Pb blank (common Pb contamination) in the laboratory has to be less than 2 pg/g. Indeed, in good laboratory conditions the dating of one crystal or part of it by ID-TIMS may yield better precision than 0.1% of error, while the in situ dating with 25 µm beam gives 1% or more of error (Bowring and Schmitz 2003).

The chemical abrasion procedure in the zircon dissolution method is widely used because it may avoid the common Pb that is present in the external surface of the crystal. This procedure has improved the high resolution U-Pb dating.

Thus, the aim of this study is to present the U-Pb dating of zircons of the Temora diorite by ID-TIMS by applying the chemical abrasion technique to the analyzed zircon fractions before the isotope dissolution. The sample for this study was provided by the Australian Geological Survey Organization, and the diorite has been systematically used as standard for geochronological studies (Black et al. 2003). In order to separate the zircon grains with a single crystallization phase (zircon with overgrowth), U-Pb in situ dating by the LA-MC-ICP-MS method was carried out in the zircon concentrate.

ANALYTICAL PROCEDURES

After the crushing and milling of the diorite Temora II sample, zircon grains were separated by conventional procedures using heavy liquids and magnetic separator after concentration by hand panning. The most clear and inclusion-free zircon grains from the least magnetic fractions were handpicked for in situ LA-MC-ICP-MS analyses and exposed to chemical abrasion for subsequent U-Pb ID-TIMS analyses.

U-Pb in situ LA-MC-ICPMS ANALYSES

All zircon grains were mounted in epoxy in 2.5-cm-diameter circular grain mounts and polished to reveal the internal part of the grains. Images of zircons grains were obtained using binocular microscope Leica MZ 125 and backscattering images with the scanning electron microscope Jeol JSM 5800. Zircon grains were dated with a New Wave UP213 laser ablation microprobe coupled to a Neptune MC-ICP-MS at the Isotope Geology Laboratory of Universidade Federal do Rio Grande do Sul (UFRGS). Isotope data were acquired using a static mode with the spot size of 25 μm. The Laser-induced elemental fractional and instrumental mass discrimination were corrected by the GJ-1 reference zircon (Simon et al. 2004), following the measurement of two GJ-1 analyses to every ten sample zircon spots. The external error was calculated after the propagation error of the GJ-1 mean and the individual sample zircon (or spot). Laser operating conditions are summarized in Table I, and a detailed description of analytical method can be found in Chemale Jr. et al. (in press).

U-Pb ID-TIMS ANALYSES

Water and acid preparation and washing

In the Isotope Geology Laboratory of UFRGS clean laboratory facility (class 100), we obtained the best water and acid blanks for the U-Pb isotope dilution method, taking into account the suggestions of the researchers of Earth Time group (Bowring, personal communication).

The study started by improving the laboratory conditions in order to get low Pb blanks for the analyses. This involved the purification of water to prepare the acids and the cleaning of the polytetrafluoretilene (PTFE) vials, bomb and microcapsules to dissolve the zircon grains. In order to get pure acids, we produced purified water below 2 pg/g of total Pb, from initial Milli-Q® water with blank values lower than 10 pg/g.

To purify the Milli-Q® water, we used two techniques: (i) filtering process with anionic resin (10 ml each time) and (ii) bi-destillation of Milli-Q® water. The second technique was more effective and faster, producing a larger amount of ultrapure water with a blank of total Pb under 2 pg/g. After the preparation of water and acids with total Pb blank values (under 2 pg/g), we proceeded to the cleaning procedure of microcapsules and bombs. Each bomb and microcapsule was washed nine (9) times with the following acids sequence: HF conc + HNO3 7N at 200° C in the oven for one day (3 times); HCl 6N at 200° C in the oven for two days (3 times); and HF conc + HNO3 7N at 200° C in the oven for one day (3 times). In each acid washing exchange the bomb and microcapsules were washed three times with purified water with 2.5 Pb pg/g of total Pb.

Zircon pre-treatment (including chemical abrasion)

Zircon crystals from the same igneous population analyzed at Isotope Geology Laboratory of UFRGS by LA-MC-ICP-MS were separated by handpicking for ID-TIMS analyses. These crystals were clear and inclusion free. All zircon crystals were photographed with a binocular microscope. The isotope analyses were done after the following preparation steps.

The zircon grains were pre-treated following the procedures suggested by Mundil et al. (2004), which are: annealing (thermal treatment at 850°C for 36 h); making photograps of all crystals using the binocular microscope; chemical abrasion (HF conc in Teflon vials, as pressurized dissolution, at 200°C for 16 h); washing several times with tridistilled HNO3; cleaning with aqua regia in an ultrasonic equipment followed by tridistilled HNO3; making photographs of all crystals using the binocular microscope to estimated the sample weight; zircons were separated into groups of 6 to 8 grains, weighing approximately 10 μg each; and tgrouping the zircon grains that were then transferred to ultraclean microcapsules (PFTE Krogh type).

Zircon dissolution and U-Pb analyses

After the chemical abrasion, the separated zircon grains were dissolved in a microcapsule as follows: 10 drops of ultraclean HNO3 7N were added and heated in the hot plate for 30 minutes at 100° C; the zircon grains were washed three times with HNO3 7N, and a drop of acid plus zircon grains was left in the bottom of the microcapsule; 200 μl of 50 % HF was added with 4μl double spike 205Pb-233U-235U (UNB205PC) in the individual microcapsule. Three sets of five microcapsules each were placed on an elevated rack in a 125 ml digestion vessel, allowing the vapor transfer dissolution in a mixture with 50 % HF/HNO3 conc (30/1). The digestion vessel was placed in a metal jacket and left in an oven for six days at 220 °C. After the zircon dissolution, the sample + spike were transferred to Teflon® vials, which were dried with 20 μl of 3M HCl + 5μl 0.25 M H3PO4.The dried dissolved zircon samples were loaded in rhenium filament with silica gel + H3PO4.The U and Pb concentrations were determined in a VG-Sector 54 mass spectrometer with Daly detector and ion counting system of ORTEC, in a single mode. Pb was analyzed as metal and U as oxide. The isotope masses of 204Pb, 205Pb, 206Pb, 207Pbb, 208Pb, 235U and 238U were collected. Repeated analyses of NBS-981 and NBS-982 presented a Pb fractionation of < 0.1 % amu, similar to that of NBS 982 analyzed in the same turret and whose zircon samples were also analyzed. All data were reduced using the PBDAT program and Isoplot 3.1 (Ludwig 2003).

We dissolved seven zircon samples and two blanks. Two samples presented either very low 206Pb/204Pb isotope ratio (~200) or unreliable U-Pb data (sample MC 6, not shown in Table III), which we attribute to high Pb blank of the microcapsule or to contamination during the analyses.

RESULTS

U-Pb isotope analyses of nine zircon grains were determined by LA-MC-ICPMS and are shown in Table II.The data from eight zircon spots yielded a concordant age of 418.6±4.3 Ma (±1.02%) (Figure 1). The 206Pb/238U age is 418.1±7.1 Ma (±1.7 %, 95% conf.). This result is in good agreement with the age published for Temora II zircon obtained by ID-TIMS and SHRIMP methods (Black et al. 2004), although the calculated error is too high for geochronological studies of Phanerozoic rocks.


In the present study, the dating by the ID-TIMS method is more appropriate for zircon grains with only one crystallization phase. Thus, the ID-TIMS analyses of the same zircon population of Temora II standard were performed in two important steps before the zircon dissolution, which permitted to get high quality results (Table III). Ultrapure water with very low total Pb blank was used, and this demanded a long time to obtain (approximately 4-6 months). Indeed, it is not possible to obtain a high resolution U-Pb dating without a good laboratory blank (Bowring and Schmitz 2003).

Aother important step was the combination of zircon heating and chemical abrasion. This step showed that the results could be improved by leaching the common Pb present in the outer part of zircon grains. The heating procedure can seal small fractures, and, thus, homogenize the external surface, which was leached by chemical abrasion with HF vapor. The chemical abrasion procedure in this work is that described by Mundil et al. (2004).

The results are presented in Table III, which includes the analytical data of analyzed zircon grains and NBS 982 values obtained during the U-Pb isotope analyses.

Two blanks and seven zircon samples were analyzed. The obtained values for blanks (microcapsules without samples) were < 1pg/g of total Pb. Two zircon samples presented either very low 206Pb/204Pb isotope ratios (~200) or unreliable U-Pb data (sample MC 6), which we attribute to high Pb blank of the MC or to contamination during the analyses. From the five reliable analyses, two yielded concordant or almost concordant zircon fractions with concordia age at 416 Ma, and three discordant zircon fractions are used for age calculation (Figure 2). The age obtained is 415.7±1.8 Ma(±0.43%, 95 % conf.) (Figure 3), which is very close to the accepted TIMS age for the Temora I (416.75±1.3 Ma, Black et al. 2003) and Temora II (416.78±0.33 Ma, Black et al. 2004).



CONCLUSIONS

Zircon analyses for in situ dating by the LA-MC-ICPMS or SHRIMP method can be appropriate for choosing single zircon grains to be used for high resolution dating by the ID-TIMS method.

The analytical results obtained here by the LA-MC-ICPMS method have a great advantage because this is a quick analytical method (usually one day), although the error uncertainty is larger than 1 %. The obtained age and error can be used for regional geological studies or detrital zircon age determination, but it cannot solve some questions such as high resolution stratigraphic studies of Phanerozoic sequences.

On the other hand, the ID-TIMS method is a time consuming method as it takes at least seven days to be performed, and is very sensitive to laboratory conditions. The obtained age and error are more precise and accurate, with an error uncertainty < 0.5 %, so the method can be applied to high resolution, Phanerozoic stratigraphy. The results of the present study, with the ID-TIMS methods combined with a previous in situ zircon dating, i.e., LA-MC-ICPMS method, are therefore significant for U-Pb high precision analyses, because the total Pb blank of the laboratory is under 2 pg/g, and the dated zircon phase has a single crystallization phase. The chemical abrasion combined with step heating is also a recommended method since it considerably decreases the amount of common Pb in zircon grains before their dissolution.

ACKNOWLEDGMENTS

The authors would like to thank Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) for supporting the first author with a scholarship of Programa Nacional de Pós-Doutorado (PNPD) during this study. We thank Chris Foudoulis of the Geological Survey of Australia for providing Temora II sample and Márcio M. Pimentel for the U235-Pb205combined spike.

Manuscript received on February 18, 2010; accepted for publication on August 24, 2011

  • BLACK LP, KAMO SL, ALLEN CM, ALEINIKOFF JN, DAVIS DW, KORSCH RJ AND FOUDOULIS C. 2003. TEMORA 1: a new zircon standard for Phanerozoic UâPb geochronology. Chem Geol 200: 155-173.
  • BLACK LP ET AL. 2004. Improved 206Pb/238U microprobe geochronology by the monitoring of a trace-element-related matrix effect; SHRIMP, IDâTIMS, ELAâICPâMS and oxygen isotope documentation for a series of zircon standards. Chem Geol 205: 115-140.
  • BOWRING SA AND SCHMITZ MD. 2003. High precision zircon geochronology and the stratigraphic record. In: Hanchar JM and Hoskins PWO. Zircon: Experiments, Isotopes, and Trace Element Investigations. Rev Mineral Geochem 53: 305-326.
  • CHEMALE JR F, KAWASHITA K, DUSSIN IA, ÁVILA JN, JUSTINO D AND BERTOTTI AL. 2012. U-Pb zircon in situ dating with LA-MC-ICP-MS using a mixed detector configuration. An Acad Bras Cienc 84: 275-295.
  • JACKSON S, PEARSON NJ, GRIFFIN WL AND BELOUSOVA EA. 2004. The application of laser ablation-inductively coupled plasma-mass spectrometry to in situ UâPb zircon geochronology. Chem Geol 211: 47-69.
  • KOSLER J, TUBRETT M AND SYLVESTER P. 2001. Application of Laser Ablation ICP-MS to U-Th-Pb Dating of Monazite. Geostandards Newsletter 25(2): 375-386.
  • LUDWIG KR. 2003. User's manual for Isoplot 3.00 Berkeley Geochonology Centre, Berkeley.
  • MUNDIL R, LUDWIG KR, METCALFE I AND RENNE PR. 2004.
  • Age and timing of the Permian mass extinctions: U/Pb dating of closed-system zircons. Science 305(5691): 1760-1763.
  • PICKHARDT C, DIETZE H-J AND BECKER JHS. 2005. Laser ablation inductively coupled plasma mass spectrometry for direct isotope ratio measurements on solid samples. Int J Mass Spectrom 242: 273-280.
  • RASMUSSEN B. 2005. Radiometric dating of sedimentary rocks: application of diagenetic xenotime geochronology. Earth Sci Rev 68: 197-243.
  • SIMON E, JACKSON SE, PEARSONA NJ, GRIFFINA WL AND BELOUSOVA EA. 2004. The application of laser ablation-inductively coupled plasma-mass spectrometry to in situ UâPb zircon geochronology. Chem Geol 211: 47-69.
  • WILLIAMS IS. 1998. U-Th-Pb geochronology by ion micro-probe. In: McKibben MA, Shanks III WC and Ridley WI (Eds), Applications of Microanalytical Techniques to Understanding Mineralization Processes. Rev Econ Geol SEG, El Paso, Texas 7: 1-75.
  • Correspondence to:
    Lucy Takehara
    E-mail:
  • Publication Dates

    • Publication in this collection
      25 May 2012
    • Date of issue
      June 2012

    History

    • Received
      18 Feb 2010
    • Accepted
      24 Aug 2011
    Academia Brasileira de Ciências Rua Anfilófio de Carvalho, 29, 3º andar, 20030-060 Rio de Janeiro RJ Brasil, Tel: +55 21 3907-8100 - Rio de Janeiro - RJ - Brazil
    E-mail: aabc@abc.org.br