Acessibilidade / Reportar erro

Reliability of echographic examination for the study of optic nerve cupping

Confiabilidade do exame ecográfico como método de estudo da escavação do n ervo óptico

SUMMARY

Purpose:

To investigate the reliability and applicability of echography to detect optic nerve cupping smaller than or equal to 0.3 and greater than or equal to 0.9.

Patients and Method:

Fifty-three eyes from 30 patients, with transparent optic media, underwent contact B-scan echography to evaluate optic disc cupping. On biomicroscopy, twenty-six eyes presented small (Group 1) and 27 large (Group II) optic nerve cupping.

Results:

B-scan images of the optic disc were evaluated by two observers and ocular echography was considered reliable with a Kappa index of 0.77 (± 0.14). Sensitivity and specificity were 71.4% (53.4 - 81.1) and 96.2% (78.4 - 99.8), respectively, with a confidence interval of 95 % •The positive and negative predictive values found for the sample were 95.2 % (74.1 - 99.8) and 78.1 % (59.6 - 90.1), respectively.

Conclusion:

Echography is a reliable method to evaluate optic disc cupping when s; 0.3 and > 0.9. However, as a screening method it would be more important in the case of negative results.

Keywords:
Ophthalmic ultrasound; Optic nerve cupping; Glaucoma

RESUMO

Objetivo:

Este estudo tem o objetivo de avaliar a confiabilidade do exame ecográfico na avaliação de escavações do nervo óptico menores ou iguais a 0,3 e maiores ou iguais a 0,9.

Material e Métodos:

Foram analisados 53 olhos, de 30 pacientes, com meios ópticos transparentes, submetidos à ecografia ocular modo B para avaliação da escavação do disco óptico. À biomicroscopia, 26 possuíam escavação pequena (grupo l) e 27 escavação grande (grupo II).

Resultados:

Os exames impressos foram avaliados por dois observadores, e a ecografia ocular se mostrou confiável com índice Kappa de O, 77 ( ± O, 14). A sensibilidade e a especificidade, com intervalo de confiança de 95%, foram de 71,4% (53,4 - 81,1) e de 96,2% (78,4 - 99,8). O Valor preditivo positivo encontrado para a amostra foi de 95,2% (74,l - 99,8) e o negativo 78,lo/o (59,6 - 90,1).

Conclusão:

A ecografia é método confiável para avaliação da escavação do disco óptico permitindo a distinção entre o grupo 0,3 e o grupo > 0,9. No entanto, como método de "screening " seria mais importante nos resultados negativos.

Palavras-chave:
Ultra-som; Escavação do nervo óptico; Glaucoma

Full text available only in PDF format

  • Paper presented as a poster at the Annual Meeting of the Association for Research on Vision in Ophthalmology, Fort Lauderdale, FL, May 16, 1997.

REFERENCES

  • 1
    Cohen JS, Stone RD, Hetherington JJ, Bullock J. Glauco0matous cupping of the optic disc by ultrasonography. Am J Ophthalmol 1976;82:24-6.
  • 2
    Coleman DJ, Carroll FD. Evaluation of optic neuropathy with B-scan utlrasonography. Am J Ophthalmol 1972;74:915-20
  • 3
    Coleman DJ, Carroll FD. High resolution B-scan ultrasonography of ihe orbit. Arch Ophthalmol 1972;88:386-7.
  • 4
    Darnley-Fisch DA, Byrne SF, Hughes JR, Parrish II RK, Feuer WF. Contact B-scan echography in the assessment of optic nerve cupping. Am J Ophthalmol 1990;109:55-61.
  • 5
    Almeida NF, Rouguaumel MZA. A definição do caso na epidemiologia. In: Almeida MF, ed. Introdução à Epidemiologia Moderna. Salvador: Ápice Produtos de Conhecimento. Coedição ABRASCO 1980;27-47.
  • 6
    Sackett DL, Huynes RB, Guyatt GH, Tugwell P. The interpretation of diagnostic data. In: Sackett DL, ed. Clinical Epidemiology - A Basic Science for Clinical Medicine, 2nd ed. Boston/Toronto/London: Little, Brown and Company, 1991. p.69-152.
  • 7
    Fletcher RH, Fletcher SW, Wagner EH. Diagnostic test. In: Fletcher RH, Fletcher SW, eds. Clinical Epidemiology - the Essential. Baltimore, Williams & Wilkins, 1982. p.41-58.

Publication Dates

  • Publication in this collection
    Jun 1999
Conselho Brasileiro de Oftalmologia Rua Casa do Ator, 1117 - cj.21, 04546-004 São Paulo SP Brazil, Tel: 55 11 - 3266-4000, Fax: 55 11- 3171-0953 - São Paulo - SP - Brazil
E-mail: abo@cbo.com.br