Acessibilidade / Reportar erro

Comparative study on optical performance and visual outcomes between two diffractive multifocal lenses: AMO Tecnis ® ZMB00 and AcrySof ® IQ ReSTOR ® Multifocal IOL SN6AD1

Estudo comparativo de desempenho óptico e resultado visual entre duas lentes multifocais difrativas: Tecnis ZMB00 e AcrySof SN6AD1

ABSTRACT

Purpose:

To compare the optical performance and visual outcomes between two diffractive multifocal lenses: AMO Tecnis® ZMB00 and AcrySof® ReSTOR® SN6AD1.

Methods:

This prospective, non-randomized comparative study included the assessment of 74 eyes in 37 patients referred for cataract surgery and candidates for multifocal intraocular lens implants. Exclusion criteria included existence of any other eye disease, previous eye surgery, high axial myopia, preoperative corneal astigmatism of >1.00 cylindrical diopter (D), and intraoperative or postoperative complications. Ophthalmological evaluation included the measurement of uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), distance-corrected near visual acuity (DCNVA), and distance-corrected intermediate visual acuity (DCIVA), with analysis of contrast sensitivity (CS), wavefront, and visual defocus curve.

Results:

Postoperative UDVA was 0.09 and 0.08 logMAR in the SN6AD1 and ZMB00 groups, respectively (p=0.868); postoperative CDVA was 0.04 and 0.02 logMAR in the SN6AD1 and ZMB00 groups, respectively (p=0.68); DCIVA was 0.17 and 0.54 logMAR in the SN6AD1 and ZMB00 groups, respectively (p=0.000); and DCNVA was 0.04 and 0.09 logMAR in the SN6AD1 and ZMB00 groups, respectively (p=0.001). In both cases, there was an improvement in the spherical equivalent and UDVA (p<0.05). Under photopic conditions, the SN6AD1 group had better CS at low frequencies without glare (p=0.04); however, the ZMB00 group achieved better sensitivity at high frequencies with glare (p=0.003). The SN6AD1 and ZMB00 lenses exhibited similar behavior for intermediate vision, according to the defocus curve; however, the ZMB00 group showed a shorter reading distance than the SN6AD1 group. There were no significant differences regarding aberrometry between the two groups.

Conclusion:

Both lenses promoted better quality of vision for both long and short distances and exhibited a similar behavior for intermediate vision. The SN6AD1 and ZMB00 groups showed better results for CS under photopic conditions at low and high spatial frequencies, respectively.

Keywords:
Lenses, intraocular; Contrast sensitivity; Aberrometry; Visual acuity

RESUMO

Objetivo:

Comparar o desempenho óptico e resultado visual entre duas lentes multifocais difrativas: AMO Tecnis® ZMB00 e AcrySof® ReSTOR® SN6AD1.

Métodos:

O estudo prospectivo, comparativo não randomizado incluiu avaliação de 74 olhos em 37 pacientes com indicação de facectomia e candidatos a implante de lente multifocal. Critérios de exclusão foram: presença de outras doenças oculares; cirurgia ocular prévia; alta miopia axial; astigmatismo cornenano maior que 1,00 D cilíndrica; complicações intraoperatórias ou pós-operatórias. A avaliação oftalmológica contou com medida da acuidade visual para longe corrigida (CDVA) e não corrigida (UDVA), intermediária (DCIVA) e curta distância (DCNVA) com melhor correção óptica para longe; teste de sensibilidade ao contraste; análise de frente de onda e curva visual de Defocus.

Resultados:

A UDVA foi de 0,09 logMAR para o grupo SN6AD1 e 0,08 logMAR para o grupo ZMB00; com correção foi de 0,04 logMAR para SN6AD1 e 0,02 para o grupo ZMB00 (p>0,05). O grupo SN6AD1 obteve valor de 0,04 logMAR e o grupo ZMB00 de 0,09 logMAR para DCNVA (p<0,05) e DCIVA de 0,17 logMAR para SN6AD1 e 0,54 logMAR para ZMB00 (p=0,000). Houve melhora do equivalente esférico e da UDVA em ambos os grupos (p<0,05). Em condições fotópicas, o grupo SN6AD1 teve melhor sensibilidade ao contraste em baixas frequências sem ofuscamento (p<0,05), contudo grupo ZMB00 obteve melhor sensibilidade em altas frequências (p<0,05) com ofuscamento. As lentes SN6AD1 e ZMB00 obtiveram comportamento semelhantes para visão intermediária na curva de Defocus, porém, o grupo ZMB00 mostrou menor distância de leitura que o grupo SN6AD1. Não houve diferença com significância estatística relacionada à aberrometria na comparação dos dois grupos.

Conclusão:

As duas lentes promoveram melhor qualidade de visão para longe e perto e comportamento semelhante para visão intermediária. O grupo ZMB00 exibiu melhores resultados para sensibilidade ao contraste em condições fotópicas em alta frequência espacial e SN6AD1 em baixa frequência espacial.

Descritores:
Lentes intraoculares; Sensibilidades de contraste; Aberrometria; Acuidade visual

INTRODUCTION

Cataract surgery is the most common surgical procedure performed worldwide and has become increasingly reliable and reproducible since the addition of phacoemulsification. Only in the United States, two million procedures are annually performed due to an increase in early diagnoses and demands of patients who wish to maintain productivity and daily activities(11 Maxwell WA, Cionni RJ, Lehmann RP, Modi SS. Functional outcomes after bilateral implantation of apodized diffractive aspheric acrylic intraocular lenses with a +3.0 or +4.0 diopter addition power: Randomized multicenter clinical study. J Cataracr Refract Surg. 2009;35(12):2054-61.).

Multifocal intraocular lenses (IOLs) were introduced in the early 1980s and offer the benefit of promoting both near and distance vision simultaneously(22 Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.,44 Alfonso JF, Fernández-Vega L, Baamonde MB, Montés-Micó R. Prospective visual evaluation of apodized diffractive intraocular lenses. J Cataract Refract Surg. 2007;33(7): 1235-43.). However, after 30 years, adverse phenomena remain challenging, requiring careful patient selection since some patients could experience less clarity, decrease in contrast, and photopic phenomena when two images are overlapped(11 Maxwell WA, Cionni RJ, Lehmann RP, Modi SS. Functional outcomes after bilateral implantation of apodized diffractive aspheric acrylic intraocular lenses with a +3.0 or +4.0 diopter addition power: Randomized multicenter clinical study. J Cataracr Refract Surg. 2009;35(12):2054-61.,55 Woodward MA, Randleman JB, Stulting RD. Dissatisfaction after multifocal intraocular lens implantation. J Cataract Refract Surg. 2009;35(6):992-7.

6 Schmickler S, Bautista CP, Goes F, Shah S, Wolffsohn JS. Clinical evaluation of a multifocal aspheric diffractive intraocular lens. Br J Ophthalmol. 2013;97(12):1560-4.
-77 Rabsilber TM, Rudalevicius P, Jasinskas V, Holzer MP, Auffarth GU. Influence of +3.00 D and +4.00 D near addition on functional outcomes of a refractive multifocal intraocular lens model. J Cataract Refract Surg. 2013;39(3):350-7.). Moreover, the same optical quality cannot be expected as with a monofocal lens(33 de Vries NE, Nuijts RM. Multifocal intraocular lenses in cataract surgery: Literature review of benefits and side effects. J Cataract Refract Surg. 2013;39(2):268-78.,66 Schmickler S, Bautista CP, Goes F, Shah S, Wolffsohn JS. Clinical evaluation of a multifocal aspheric diffractive intraocular lens. Br J Ophthalmol. 2013;97(12):1560-4.).

Several types of materials and technologies have been employed in designing IOLs, with refractive, diffractive, and hybrid models being the most widely used by surgeons. Moreover, asphericity was introduced as a means to compensate positive corneal spherical aberration(22 Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.,33 de Vries NE, Nuijts RM. Multifocal intraocular lenses in cataract surgery: Literature review of benefits and side effects. J Cataract Refract Surg. 2013;39(2):268-78.). The addition promoted by the lens, whether it is +2.5 diopters (D), +3.0 D, or +4.0 D, influences the best reading distance and depth of focus. These different compositions culminate in distinctly different optical aberrometric behaviors(88 Oliveira CM, Ferreira A, Franco S. Wavefront analysis and Zernike polynomial decomposition for evaluation of corneal optical quality. J Cataract Refract Surg. 2012;38(2): 343-56.

9 Mello GR, Rocha KM, Santhiago MR, Smadja D, Krueger RR. Applications of wavefront technology. J Cataract Refract Surg. 2012;38(9):1671-83.
-1010 Hida WT, Nakano CT, Yamane IS, Motta AF, Tzeliks PF, Alencar LM, et al. Desempenho visual dos pacientes pseudofácicos com diferentes lentes intraoculares. Rev Bras Oftalmol. 2013;72(5):287-93.).

In this study, we compared visual parameters, optical performance, contrast sensitivity (CS), and wavefront analysis results between two different types of multifocal lenses currently available in the market: AcrySof® IQ ReSTOR® Multifocal IOL SN6AD1 (Alcon Laboratories, Inc., Irvine, CA, USA) and AMO Tecnis® ZMB00 (Abbott Medical Optics, Santa Ana, CA, USA).

The AcrySof® IQ ReSTOR® Multifocal IOL SN6AD1 is a single piece lens composed of foldable hydrophobic acrylic, aspheric, adding a negative aberration of -0.20 µm to the human eye. With a refractive-diffractive hybrid design and apodized diffractive anterior surface, it shows a better near-distance visual acuity for smaller pupils and better long-distance visual acuity for larger pupils(1111 Cillino G, Casuccio A, Pasti M, Bono V, Mencucci R, Cillino S. Working-age cataract patients: visual results, reading performance, and quality of life with three diffractive multifocal intraocular lenses. Ophthalmology. 2014;121(1):34-44.,1212 de Vries NE, Webers CA, Montés-Micó R, Ferrer-Blasco T, Nuijts RM. Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: comparative study. J Cataract Refract Surg. 2010;36(8): 1316-22.). It has a 3.6-mm diffractive optic zone, consisting of nine concentric rings that decrease from the center to the periphery (apodization), with a central ring of 0.856 mm and provides bifocality with a addition of 3 D, thus corresponding to 2.4 D in the lens plane(33 de Vries NE, Nuijts RM. Multifocal intraocular lenses in cataract surgery: Literature review of benefits and side effects. J Cataract Refract Surg. 2013;39(2):268-78.,1212 de Vries NE, Webers CA, Montés-Micó R, Ferrer-Blasco T, Nuijts RM. Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: comparative study. J Cataract Refract Surg. 2010;36(8): 1316-22.).

The AMO Tecnis® ZMB00 lens is an evolution of the ZM900 IOL, which is constructed of a single piece of foldable hydrophobic acrylic material, with a diffractive posterior surface and aspheric anterior surface crafted to add a negative spherical aberration of -0.27 µm. It has 32 concentric rings, with a central ring of 1 mm and promotes an addition of 4 D, thus corresponding to 3.2 D in the lens plane, regardless of pupil size(66 Schmickler S, Bautista CP, Goes F, Shah S, Wolffsohn JS. Clinical evaluation of a multifocal aspheric diffractive intraocular lens. Br J Ophthalmol. 2013;97(12):1560-4.,1111 Cillino G, Casuccio A, Pasti M, Bono V, Mencucci R, Cillino S. Working-age cataract patients: visual results, reading performance, and quality of life with three diffractive multifocal intraocular lenses. Ophthalmology. 2014;121(1):34-44.).

METHODS

This study was conducted in accordance with the ethical standards of clinical and surgical research and was approved by the ethics committee for Analysis of Research Projects (CAPPesq) of the Medical Staff of the Hospital Oftalmológico de Brasília and the Hospital das Forças Armadas, Brasília, Brazil. This prospective, non-randomized, comparative study included 74 eyes of 37 patients, selected from a list of cataract outpatients of the Hospital Oftalmológico de Brasília, who underwent cataract extraction with IOL implant between January 2011 and July 2013.

All surgeries were performed by the same surgeon (W.T.H.). The main incision was made in the steepest corneal meridian in three planes to encourage self-sealing. Continuous circular capsulorhexis and hydrodissection were performed and followed by viscoelastic soft-shell technique(1010 Hida WT, Nakano CT, Yamane IS, Motta AF, Tzeliks PF, Alencar LM, et al. Desempenho visual dos pacientes pseudofácicos com diferentes lentes intraoculares. Rev Bras Oftalmol. 2013;72(5):287-93.). Phacoemulsification was followed by irrigation and aspiration of the lenticular cortex and was concluded with implantation of one of the following lenses into the capsular bag: AcrySof® IQ ReSTOR® Multifocal IOL SN6AD1 or AMO Tecnis® ZMB00.

The inclusion criteria were (1) presence of a senile bilateral cataract, (2) interest in multifocal lens implant, (3) corneal astigmatism of ≤1.00 cylindrical diopter in both eyes, (4) pupil diameter of >3.0 mm under mesopic conditions, and (5) none of the following exclusionary factors: (a) existence of any other eye pathology or neuropathy that could decrease visual acuity, (b) CS or visual field sensitivity, (c) high axial myopia, (d) preoperative or postoperative complications, (e) concerns regarding an IOL implant inside the capsular bag, or (f) IOL decentration of >0.5 mm, as measured with a slit lamp.

We considered working and reading habits when choosing the lens for implantation. Patients with habits that demanded better intermediate vision were included in the IOL implantation group that aimed to achieve an addition of +3 D (SN6AD1). The remaining patients were included in the IOL group that aimed to achieve an addition of +4 D (ZMB00).

All patients underwent a complete ophthalmological evaluation on all visits, before and after surgery. Data from the 180-day visit of all patients were used for comparative analysis. Acuity of short-distance vision (33 cm) with and without refraction correction was measured using the reading table model of the Early Treatment Diabetic Retinopathy Study chart (ETDRS; Precision Vision, Ltd., Aurora, CO, USA). The same procedure was performed to test long-distance vision (4 m), although the chart was backlit to 100% contrast. All values were calculated according to the logMAR rating scale(1313 Hida WT, Motta AF, Kara-José Júnior N, Costa H, Tokunaga C, Cordeiro LN, et al. Estudo comparativo do desempenho visual e análise de frente de onda entre as lentes intra-oculares multifocais difrativas Tecnis® ZM900 e AcrySof® ResTor® SN60D3. Arq Bras Oftalmol. 2008;71(6):788-92.).

IOL dioptric power was calculated according to the optical interferometry technique using the IOLMaster® 500 Optical Biometer (Carl Zeiss, Jena, Germany), which sought emmetropia or the first positive value calculated by the following three formulas: Hoffer-Q for axial lengths of <22 mm, Holladay between 22 and 24.50 mm, and Sanders-Retzlaff-Kraff/T for axial lengths of >24.50 mm(1414 Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL. Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg. 2011;37(1):63-71.).

The postoperative scheme of medication (eye drops) was similar for all patients. The evaluation included the testing of monocular CS under photopic (with and without glare) and mesopic conditions using the Optec® 6500P Vision Screener (Stereo Optical Company, Inc., Chicago, IL USA) with spatial frequencies ranging between 1.5 and 18 cycles/degree and using the Functional Acuity Contrast Test chart provided by the manufacturer, which consists of columns that grow at different rates of 0.15 log units, which are then converted into base 10 logarithmic units for statistical analysis(22 Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.,1515 Jindra LF, Zemon V. Contrast sensitivity testing: a more complete assessment of vision. J Cataract Refract Surg. 1989;15(2):141-8.

16 Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.

17 Bautista CP, González DC, Gómez AC. Evolution of visual performance in 70 eyes implanted with the Tecnis(®) ZMB00 multifocal intraocular lens. Clin Ophthalmol (Auckland, N.Z.); 2012;6:403-7.

18 Hong YT, Kim SW, Kim EK, Kim T-i. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010;36(4):547-52.

19 Pesudovs K, Hazel CA, Doran RM, Elliott DB. The usefulness of Vistech and FACT contrast sensitivity charts for cataract and refractive surgery outcomes research. Br J Ophthalmol. 2004;88(1):11-6.
-2020 Alfonso JF, Fernández-Vega L, Baamonde MB, Montés-Micó R. Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens. J Cataract Refract Surg. 2007;33(3):430-8.).

The same assessment was analyzed across the wavefront function, and the monocular optical aberrations were gathered using an OPD-Scan III aberrometer (Nidek Co., Ltd., Gamagori, Japan). Unlike other studies, we sought to evaluate aberrometry under conditions of mesopic pupil in an attempt to detect the manifestation of wavefront function under physiological conditions without the use of pharmacological mydriasis(1313 Hida WT, Motta AF, Kara-José Júnior N, Costa H, Tokunaga C, Cordeiro LN, et al. Estudo comparativo do desempenho visual e análise de frente de onda entre as lentes intra-oculares multifocais difrativas Tecnis® ZM900 e AcrySof® ResTor® SN60D3. Arq Bras Oftalmol. 2008;71(6):788-92.). The following variables were collected: (1) total aberration, (2) high-order spherical aberration, (3) tilt, (4) coma, and (5) Strehl ratio(88 Oliveira CM, Ferreira A, Franco S. Wavefront analysis and Zernike polynomial decomposition for evaluation of corneal optical quality. J Cataract Refract Surg. 2012;38(2): 343-56.,99 Mello GR, Rocha KM, Santhiago MR, Smadja D, Krueger RR. Applications of wavefront technology. J Cataract Refract Surg. 2012;38(9):1671-83.). Finally, the visual monocular defocus curve was evaluated in long-distance visual acuity, corrected using the same ETDRS chart at a distance of 4 m, at intervals of 0.50 spherical diopters from -5.00 to +2.50 D(2121 Petermeier K, Messias A, Gekeler F, Szurman P. Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity. J Cataract Refract Surg. 2011;37(4):720-6.).

Significance was tested using the Tukey, Wilcoxon, Kruskal-Wallis, ANOVA, and chi-square tests by adjusting to a level of significance of 5% (p<0.05) and using SPSS for Windows statistical analysis software (IBM-SPSS, Inc., Chicago, IL, USA).

RESULTS

The study involved the participation of 37 subjects, 22 women (59.5%) and 15 men (40.5%), and a total of 74 eyes. There was homogeneity in the group distribution of lenses regarding to gender, age, axial length of the eye, and chosen diopter, as shown in table 1. The average preoperative spherical equivalence (SE) of all subjects was 1.58 D, and there was no relevant statistical difference between the groups (Table 1).

Table 1
Descriptive measures for age, implanted IOL refractive power (Diopters), axial length, preoperative spherical equivalent, preoperative and postoperative visual acuities for SN6AD1 and ZMB00 lenses

Preoperative uncorrected distance visual acuity (UDVA) was better in the SN6AD1 group than in the ZMB00 group (0.62 vs. 0.50 logMAR, p=0.058), although this difference was not statistically significant (Table 1). There were no statistically significant differences in postoperative data of SE, UDVA, or corrected long-distance visual acuity (CDVA) between the groups (Table 1).

There was a statistically significant difference in pre- and postoperative SE and in pre- and postoperative UDVA between groups (Table 1).

Considering short-distance vision, there was a significant difference in the average short-distance visual acuity, as analyzed with correction for long-distance (DCNVA) between the SN6AD1 and ZMB00 groups (0.04 vs. 0.09 logMAR, p=0.009), and in intermediate vision (0.17 vs. 0.54 logMAR, p=0.000) (Table 1 and Figure 1). Preoperative and postoperative spherical equivalent (SE) results are demonstrated in figure 2.

Figure 1
Preoperative long-distance visual acuity without (UDVA) correction, versus post-operative UDVA versus postoperative visual acuity for long-distance with (CDVA) correction for SN6AD1 and ZMB00 lenses.

Figure 2
Preoperative and postoperative spherical equivalent (SE), for SN6AD1 and ZMB00 lenses.

None of the patients required any type of short- or long-distance correction in their everyday life following surgery.

CS under photopic conditions without glare was better at a low frequency (1.5 cpd) for the SN6AD1 lens (p=0.039) and at a high frequency (18 cpd) for the ZMB00 lens (p=0.62). Under photopic conditions with glare, the ZMB00 lens performed better at high frequencies (18 cpd) with p = 0.003. Under mesopic conditions, the lenses exhibited similar behavior (p>0.05) (Figures 3, 4, and 5).

Figure 3
Comparison for the SN6AD1 and ZMB00 lenses with respect to the photopic situation without glare.

Figure 4
Comparison for the SN6AD1 lenses and the ZMB00, with respect to the photopic with glare situation.

Figure 5
Comparison for the SN6AD1 lenses and the ZMB00 lenses with respect to the mesopic with glare situation.

Regarding defocus curve, there was a difference in almost all of the distances assessed between lenses (converted into diopters, p<0.001). The ZMB00 lens showed peaks close to -3.0 D (33 cm) and 0.0 D, with an average visual acuity of 0.09 and 0.02 logMAR, respectively. The line corresponds to the SN6AD1 lens and showed peaks at -2.5 D (40 cm) and 0.0 D, with visual acuity of 0.06 and 0.02 logMAR, respectively. Regarding intermediate deflection vision, both occurred at -1.5 D (66 cm), with a visual acuity of 0.28 and 0.29 logMAR, respectively, but with no statistically significant difference from this point (p=0.68) to -2.5 D (40 cm) (p=0.12) (Figure 6).

Figure 6
Defocus curve for SN6AD1 lenses and ZMB00.

Wavefront analysis demonstrated that, in absolute values, the ZMB00 lens showed better results in the decomposition of the wavefront function; however, there was no statistically significant difference between groups (p>0.05) (Table 2).

Table 2
Comparison of wavefront function analysis between SN6AD1 and ZMB00

DISCUSSION

The analysis of groups distribution demonstrates homogeneity and enables comparisons between groups, indicating its suitability for comparison to other publications(77 Rabsilber TM, Rudalevicius P, Jasinskas V, Holzer MP, Auffarth GU. Influence of +3.00 D and +4.00 D near addition on functional outcomes of a refractive multifocal intraocular lens model. J Cataract Refract Surg. 2013;39(3):350-7.,1010 Hida WT, Nakano CT, Yamane IS, Motta AF, Tzeliks PF, Alencar LM, et al. Desempenho visual dos pacientes pseudofácicos com diferentes lentes intraoculares. Rev Bras Oftalmol. 2013;72(5):287-93.,2121 Petermeier K, Messias A, Gekeler F, Szurman P. Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity. J Cataract Refract Surg. 2011;37(4):720-6.

22 Ye PP, Li X, Yao K. Visual outcome and optical quality after bilateral implantation of aspheric diffractive multifocal, aspheric monofocal and spherical monofocal intraocular lenses: a prospective comparison. Int J Ophthalmol. 2013;6(3):300-6.

23 Yamauchi T, Tabuchi H, Takase K, Ohsugi H, Ohara Z, Kiuchi Y. Comparison of visual performance of multifocal intraocular lenses with same material monofocal intraocular lenses. PLoS One. 2013;8(6):e68236.
-2424 de Vries NE, Webers CA, Touwslager WR, Bauer NJ, de Brabander J, Berendschot TT, et al. Dissatisfaction after implantation of multifocal intraocular lenses. J Cataract Refract Surg. 2011;37(5):859-65.).

As demonstrated in other studies, the good quality of distance and near vision was attested through both types of lenses, regardless of which addition had been used. The improvement in UDVA, in addition to the excellent individual results, for both long and short distance confirmed the efficacy of the treatment and provided independence from glasses(11 Maxwell WA, Cionni RJ, Lehmann RP, Modi SS. Functional outcomes after bilateral implantation of apodized diffractive aspheric acrylic intraocular lenses with a +3.0 or +4.0 diopter addition power: Randomized multicenter clinical study. J Cataracr Refract Surg. 2009;35(12):2054-61.

2 Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.
-33 de Vries NE, Nuijts RM. Multifocal intraocular lenses in cataract surgery: Literature review of benefits and side effects. J Cataract Refract Surg. 2013;39(2):268-78.,66 Schmickler S, Bautista CP, Goes F, Shah S, Wolffsohn JS. Clinical evaluation of a multifocal aspheric diffractive intraocular lens. Br J Ophthalmol. 2013;97(12):1560-4.,1212 de Vries NE, Webers CA, Montés-Micó R, Ferrer-Blasco T, Nuijts RM. Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: comparative study. J Cataract Refract Surg. 2010;36(8): 1316-22.,1313 Hida WT, Motta AF, Kara-José Júnior N, Costa H, Tokunaga C, Cordeiro LN, et al. Estudo comparativo do desempenho visual e análise de frente de onda entre as lentes intra-oculares multifocais difrativas Tecnis® ZM900 e AcrySof® ResTor® SN60D3. Arq Bras Oftalmol. 2008;71(6):788-92.,1717 Bautista CP, González DC, Gómez AC. Evolution of visual performance in 70 eyes implanted with the Tecnis(®) ZMB00 multifocal intraocular lens. Clin Ophthalmol (Auckland, N.Z.); 2012;6:403-7.,2121 Petermeier K, Messias A, Gekeler F, Szurman P. Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity. J Cataract Refract Surg. 2011;37(4):720-6.

22 Ye PP, Li X, Yao K. Visual outcome and optical quality after bilateral implantation of aspheric diffractive multifocal, aspheric monofocal and spherical monofocal intraocular lenses: a prospective comparison. Int J Ophthalmol. 2013;6(3):300-6.
-2323 Yamauchi T, Tabuchi H, Takase K, Ohsugi H, Ohara Z, Kiuchi Y. Comparison of visual performance of multifocal intraocular lenses with same material monofocal intraocular lenses. PLoS One. 2013;8(6):e68236.,2525 Sood P, Woodward MA. Patient acceptability of the Tecnis multifocal intraocular lens. Clin Ophthalmol (Auckland, N.Z.). 2011;5:403-10.). Such inference is particularly evident when considering postoperative SE.

Analysis of DCNVA highlighted advantages of the SN6AD1 lens (p<0.05). However, when converting the Jaeger table visual scale to logMAR, values that transcend the understanding of the different visual levels for short distances may appear. Given that J1 corresponds to 0.00 logMAR and at the level immediately below, J2 corresponds to 0.1 logMAR, both groups presented average DCNVA level at J1 and better than J2(1111 Cillino G, Casuccio A, Pasti M, Bono V, Mencucci R, Cillino S. Working-age cataract patients: visual results, reading performance, and quality of life with three diffractive multifocal intraocular lenses. Ophthalmology. 2014;121(1):34-44.).

Considering CS under photopic conditions, the ZMB00 lens showed better results at high spatial frequencies (p<0.05). Low spatial frequencies are associated with the perception of motion (magnocellular pathway), whereas high frequencies are associated with the perception of detail (parvocellular pathway)(2626 Souza GS, Lacerda EM, Silveira VA, Araújo CS, Silveira LC. A visão através dos contrastes. Estudos Avançados. 2013;27(77):45-60.). Hida et al. compared the SN60D3 IOL to the ZM900 IOL and found similar results applicable to this study since SN6AD1 and ZMB00 lenses are the evolution of the SN60D3 and ZM900 IOLs(1313 Hida WT, Motta AF, Kara-José Júnior N, Costa H, Tokunaga C, Cordeiro LN, et al. Estudo comparativo do desempenho visual e análise de frente de onda entre as lentes intra-oculares multifocais difrativas Tecnis® ZM900 e AcrySof® ResTor® SN60D3. Arq Bras Oftalmol. 2008;71(6):788-92.). Cillino et al. compared the visual CSs of SN6AD1, SN6AD3, and ZMA00 lenses and found no statistical difference among the three groups. However, the study was performed using the Pelli-Robson method to evaluate CS, which has a greater chance of being influenced by the environment, unlike the automated method(1111 Cillino G, Casuccio A, Pasti M, Bono V, Mencucci R, Cillino S. Working-age cataract patients: visual results, reading performance, and quality of life with three diffractive multifocal intraocular lenses. Ophthalmology. 2014;121(1):34-44.). Afonso et al. found similar values for the SN6AD1 lens, as did de Vries et al. for average values at high frequencies (1.13 at 12 cpd and 0.65 at 18 cpd)(22 Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.,1212 de Vries NE, Webers CA, Montés-Micó R, Ferrer-Blasco T, Nuijts RM. Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: comparative study. J Cataract Refract Surg. 2010;36(8): 1316-22.).

The differences in Mesopic CS were not significant when comparing the two groups at any spatial frequency. This finding was consistent with those in other publications(22 Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.,1212 de Vries NE, Webers CA, Montés-Micó R, Ferrer-Blasco T, Nuijts RM. Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: comparative study. J Cataract Refract Surg. 2010;36(8): 1316-22.,1313 Hida WT, Motta AF, Kara-José Júnior N, Costa H, Tokunaga C, Cordeiro LN, et al. Estudo comparativo do desempenho visual e análise de frente de onda entre as lentes intra-oculares multifocais difrativas Tecnis® ZM900 e AcrySof® ResTor® SN60D3. Arq Bras Oftalmol. 2008;71(6):788-92.,2222 Ye PP, Li X, Yao K. Visual outcome and optical quality after bilateral implantation of aspheric diffractive multifocal, aspheric monofocal and spherical monofocal intraocular lenses: a prospective comparison. Int J Ophthalmol. 2013;6(3):300-6.).

It must be considered that under photopic and mesopic conditions, the two groups showed averages below their corresponding monofocal IOLs peers, as described in other studies. Considering CS, ZA9003 IOL has been considered better than the ZMA00 IOL(1010 Hida WT, Nakano CT, Yamane IS, Motta AF, Tzeliks PF, Alencar LM, et al. Desempenho visual dos pacientes pseudofácicos com diferentes lentes intraoculares. Rev Bras Oftalmol. 2013;72(5):287-93.,2222 Ye PP, Li X, Yao K. Visual outcome and optical quality after bilateral implantation of aspheric diffractive multifocal, aspheric monofocal and spherical monofocal intraocular lenses: a prospective comparison. Int J Ophthalmol. 2013;6(3):300-6.,2323 Yamauchi T, Tabuchi H, Takase K, Ohsugi H, Ohara Z, Kiuchi Y. Comparison of visual performance of multifocal intraocular lenses with same material monofocal intraocular lenses. PLoS One. 2013;8(6):e68236.). Nakano et al. and Vingolo et al. concluded the same by comparing SN60WF and SA60AT IOLs with SN60D3 and ZM900 IOLs(2727 Nakano CT, Hida WT, Kara-Jose Junior N, Motta AF, Fante D, Masson VF, et al. Comparison between OPD-scan results and contrast sensitivity of three intraocular lenses: spheric AcrySof SN60AT, aspheric AcrySof SN60WF and multifocal AcrySof Restor lens. Rev Bras Ophthalmol. 2009;68(4):216-22.,2828 Vingolo EM, Grenga P, Iacobelli L, Grenga R. Visual acuity and contrast sensitivity: AcrySof ReSTOR apodized diffractive versus AcrySof SA60AT monofocal intraocular lenses. J Cataract Refract Surg. 2007;33(7):1244-7.). This phenomenon could be explained through the analysis of the concept of multifocality promoted by these kind of lenses by splitting light energy into two foci and overlapping near and far images.

Afonso et al. correlated greater pupil diameters to better CS at all frequencies under photopic and mesopic conditions(2020 Alfonso JF, Fernández-Vega L, Baamonde MB, Montés-Micó R. Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens. J Cataract Refract Surg. 2007;33(3):430-8.). This idea seems plausible because a larger amount of light enters the optical system, which is proportional to the increase in pupil diameter. In our study, we did not perform analysis of any such correlation.

Considering visual defocus curve, it was clear that both groups reached two peaks of improved visual acuity corresponding to near and far distances. Considering near vision, significantly better vision was found close to 33 cm with the ZMB00 lens than with the SN6AD1 at this point, although the SN6AD1 lens demonstrated greater near vision at a 40-cm reading distance, but with no statistical significance when compared with the ZMB00 lens at this focus point. This is a different result from what was found at the near visual acuity test and may be related to pupillary function that improves the near visual acuity for refractive-diffractive hybrid design lenses.

When assessing intermediate vision of 66 cm, the two groups showed similar low performances when compared with the peak curve points of better vision. Schmickler et al. reached similar values when individually assessing the ZMB00 lens(66 Schmickler S, Bautista CP, Goes F, Shah S, Wolffsohn JS. Clinical evaluation of a multifocal aspheric diffractive intraocular lens. Br J Ophthalmol. 2013;97(12):1560-4.). Similarly, Afonso et al., de Vries et al., and Petermeier et al. depicted the same defocus curve pattern for the SN6AD1 lens(22 Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.,1212 de Vries NE, Webers CA, Montés-Micó R, Ferrer-Blasco T, Nuijts RM. Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: comparative study. J Cataract Refract Surg. 2010;36(8): 1316-22.,2121 Petermeier K, Messias A, Gekeler F, Szurman P. Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity. J Cataract Refract Surg. 2011;37(4):720-6.).

Nevertheless, analysis of the two IOL defocus patterns suggests that there should be a relative permanence of the curve near the better vision peaks, which indicates a reasonable depth of focus around these two top focal points. To be more specific, the reading distance might not be confined only to certain near distance, and it would have an amplitude of approximately 12 cm in diameter around the best near focal point. Although it seems little, it may make a difference for better adaptation of patients with more intermediate visual habits.

Regarding the results of aberrometry, both lenses were successful in decreasing corneal positive spherical aberration. However, self-criticism is necessary to evaluate the method employed. Because a multifocal lens simultaneously generates two distinct wavefronts, it may be difficult to achieve separate and accurate measurements with a particular device. Nevertheless, it is understood that the same difficulty in measurement was imposed upon both groups, making comparisons reasonable. Hida et al. compared the ZM900 and SN60D3 lenses and found similar aberrometric results, as did Nakano et al.(1010 Hida WT, Nakano CT, Yamane IS, Motta AF, Tzeliks PF, Alencar LM, et al. Desempenho visual dos pacientes pseudofácicos com diferentes lentes intraoculares. Rev Bras Oftalmol. 2013;72(5):287-93.,2727 Nakano CT, Hida WT, Kara-Jose Junior N, Motta AF, Fante D, Masson VF, et al. Comparison between OPD-scan results and contrast sensitivity of three intraocular lenses: spheric AcrySof SN60AT, aspheric AcrySof SN60WF and multifocal AcrySof Restor lens. Rev Bras Ophthalmol. 2009;68(4):216-22.). Moreno et al. evaluated a system for analyzing the wavefront of a double-passage Optical Quality Analysis System (Visiometrics, S.L., Cerdanyola, BCN, Spain) that could differentiate the two wavefronts and thereby assess the optical quality more accurately for better differentiation(2929 Moreno LJ, Piñero DP, Alió JL, Fimia A, Plaza AB. Double-pass system analysis of the visual outcomes and optical performance of an apodized diffractive multifocal intraocular lens. J Cataract Refract Surg. 2010;36(12):2048-55.).

It is important to emphasize the Strehl ratio variable. This value corresponds to the ratio between the peak of the measured point spread function (PSF) and a PSF peak in an aberration-free optical system. Values range from 0 to 1, where 1 represents an ideal optical system. The value usually decreases as the pupil increases, expanding the influence of optical aberrations that compromise quality. Larger or smaller values for similar pupil conditions correspond to better or worse visual quality(99 Mello GR, Rocha KM, Santhiago MR, Smadja D, Krueger RR. Applications of wavefront technology. J Cataract Refract Surg. 2012;38(9):1671-83.). Our results suggested equivalence between the two lenses. Artigas et al., compared the SN60D3 and ZM900 lenses and suggested better results for the SN60D3 lens, using an artificial eye as model(3030 Artigas JM, Menezo JL, Peris C, Felipe A, Díaz-Llopis M. Image quality with multifocal intraocular lenses and the effect of pupil size: comparison of refractive and hybrid refractive-diffractive designs. J Cataract Refract Surg. 2007;33(12):2111-7.).

In conclusion, the two lenses were effective in enabling better vision for both long and short distances, although neither exhibited excellent visual acuity on intermediate vision, despite this equivalence. Both groups had a reasonable range of focus around the two best focus points (near and far). The ZMB00 lens recorded better results in CS under photopic conditions at high frequencies, although both lenses demonstrated inferior results when compared to their monofocal peers described in past publications. The ZMB00 and SN6AD1 lenses are equivalent with regards to wavefront function, although more studies are needed to better assess aberrometry.

  • Funding: This study was supported by Hospital Oftalmológico de Brasília (HOB) and Centro de Estudos Oftalmológicos Renato Ambrósio (CEORA).
  • Approved by the following research ethics committee: Hospital das Forças Armadas, Brasília, DF, (#172.922).

REFERENCES

  • 1
    Maxwell WA, Cionni RJ, Lehmann RP, Modi SS. Functional outcomes after bilateral implantation of apodized diffractive aspheric acrylic intraocular lenses with a +3.0 or +4.0 diopter addition power: Randomized multicenter clinical study. J Cataracr Refract Surg. 2009;35(12):2054-61.
  • 2
    Alfonso JF, Fernández-Vega L, Blázquez JI, Montés-Micó R. Visual function comparison of 2 aspheric multifocal intraocular lenses. J Cataract Refract Surg. 2012;38(2):242-8.
  • 3
    de Vries NE, Nuijts RM. Multifocal intraocular lenses in cataract surgery: Literature review of benefits and side effects. J Cataract Refract Surg. 2013;39(2):268-78.
  • 4
    Alfonso JF, Fernández-Vega L, Baamonde MB, Montés-Micó R. Prospective visual evaluation of apodized diffractive intraocular lenses. J Cataract Refract Surg. 2007;33(7): 1235-43.
  • 5
    Woodward MA, Randleman JB, Stulting RD. Dissatisfaction after multifocal intraocular lens implantation. J Cataract Refract Surg. 2009;35(6):992-7.
  • 6
    Schmickler S, Bautista CP, Goes F, Shah S, Wolffsohn JS. Clinical evaluation of a multifocal aspheric diffractive intraocular lens. Br J Ophthalmol. 2013;97(12):1560-4.
  • 7
    Rabsilber TM, Rudalevicius P, Jasinskas V, Holzer MP, Auffarth GU. Influence of +3.00 D and +4.00 D near addition on functional outcomes of a refractive multifocal intraocular lens model. J Cataract Refract Surg. 2013;39(3):350-7.
  • 8
    Oliveira CM, Ferreira A, Franco S. Wavefront analysis and Zernike polynomial decomposition for evaluation of corneal optical quality. J Cataract Refract Surg. 2012;38(2): 343-56.
  • 9
    Mello GR, Rocha KM, Santhiago MR, Smadja D, Krueger RR. Applications of wavefront technology. J Cataract Refract Surg. 2012;38(9):1671-83.
  • 10
    Hida WT, Nakano CT, Yamane IS, Motta AF, Tzeliks PF, Alencar LM, et al. Desempenho visual dos pacientes pseudofácicos com diferentes lentes intraoculares. Rev Bras Oftalmol. 2013;72(5):287-93.
  • 11
    Cillino G, Casuccio A, Pasti M, Bono V, Mencucci R, Cillino S. Working-age cataract patients: visual results, reading performance, and quality of life with three diffractive multifocal intraocular lenses. Ophthalmology. 2014;121(1):34-44.
  • 12
    de Vries NE, Webers CA, Montés-Micó R, Ferrer-Blasco T, Nuijts RM. Visual outcomes after cataract surgery with implantation of a +3.00 D or +4.00 D aspheric diffractive multifocal intraocular lens: comparative study. J Cataract Refract Surg. 2010;36(8): 1316-22.
  • 13
    Hida WT, Motta AF, Kara-José Júnior N, Costa H, Tokunaga C, Cordeiro LN, et al. Estudo comparativo do desempenho visual e análise de frente de onda entre as lentes intra-oculares multifocais difrativas Tecnis® ZM900 e AcrySof® ResTor® SN60D3. Arq Bras Oftalmol. 2008;71(6):788-92.
  • 14
    Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL. Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg. 2011;37(1):63-71.
  • 15
    Jindra LF, Zemon V. Contrast sensitivity testing: a more complete assessment of vision. J Cataract Refract Surg. 1989;15(2):141-8.
  • 16
    Richman J, Spaeth GL, Wirostko B. Contrast sensitivity basics and a critique of currently available tests. J Cataract Refract Surg. 2013;39(7):1100-6.
  • 17
    Bautista CP, González DC, Gómez AC. Evolution of visual performance in 70 eyes implanted with the Tecnis(®) ZMB00 multifocal intraocular lens. Clin Ophthalmol (Auckland, N.Z.); 2012;6:403-7.
  • 18
    Hong YT, Kim SW, Kim EK, Kim T-i. Contrast sensitivity measurement with 2 contrast sensitivity tests in normal eyes and eyes with cataract. J Cataract Refract Surg. 2010;36(4):547-52.
  • 19
    Pesudovs K, Hazel CA, Doran RM, Elliott DB. The usefulness of Vistech and FACT contrast sensitivity charts for cataract and refractive surgery outcomes research. Br J Ophthalmol. 2004;88(1):11-6.
  • 20
    Alfonso JF, Fernández-Vega L, Baamonde MB, Montés-Micó R. Correlation of pupil size with visual acuity and contrast sensitivity after implantation of an apodized diffractive intraocular lens. J Cataract Refract Surg. 2007;33(3):430-8.
  • 21
    Petermeier K, Messias A, Gekeler F, Szurman P. Effect of +3.00 diopter and +4.00 diopter additions in multifocal intraocular lenses on defocus profiles, patient satisfaction, and contrast sensitivity. J Cataract Refract Surg. 2011;37(4):720-6.
  • 22
    Ye PP, Li X, Yao K. Visual outcome and optical quality after bilateral implantation of aspheric diffractive multifocal, aspheric monofocal and spherical monofocal intraocular lenses: a prospective comparison. Int J Ophthalmol. 2013;6(3):300-6.
  • 23
    Yamauchi T, Tabuchi H, Takase K, Ohsugi H, Ohara Z, Kiuchi Y. Comparison of visual performance of multifocal intraocular lenses with same material monofocal intraocular lenses. PLoS One. 2013;8(6):e68236.
  • 24
    de Vries NE, Webers CA, Touwslager WR, Bauer NJ, de Brabander J, Berendschot TT, et al. Dissatisfaction after implantation of multifocal intraocular lenses. J Cataract Refract Surg. 2011;37(5):859-65.
  • 25
    Sood P, Woodward MA. Patient acceptability of the Tecnis multifocal intraocular lens. Clin Ophthalmol (Auckland, N.Z.). 2011;5:403-10.
  • 26
    Souza GS, Lacerda EM, Silveira VA, Araújo CS, Silveira LC. A visão através dos contrastes. Estudos Avançados. 2013;27(77):45-60.
  • 27
    Nakano CT, Hida WT, Kara-Jose Junior N, Motta AF, Fante D, Masson VF, et al. Comparison between OPD-scan results and contrast sensitivity of three intraocular lenses: spheric AcrySof SN60AT, aspheric AcrySof SN60WF and multifocal AcrySof Restor lens. Rev Bras Ophthalmol. 2009;68(4):216-22.
  • 28
    Vingolo EM, Grenga P, Iacobelli L, Grenga R. Visual acuity and contrast sensitivity: AcrySof ReSTOR apodized diffractive versus AcrySof SA60AT monofocal intraocular lenses. J Cataract Refract Surg. 2007;33(7):1244-7.
  • 29
    Moreno LJ, Piñero DP, Alió JL, Fimia A, Plaza AB. Double-pass system analysis of the visual outcomes and optical performance of an apodized diffractive multifocal intraocular lens. J Cataract Refract Surg. 2010;36(12):2048-55.
  • 30
    Artigas JM, Menezo JL, Peris C, Felipe A, Díaz-Llopis M. Image quality with multifocal intraocular lenses and the effect of pupil size: comparison of refractive and hybrid refractive-diffractive designs. J Cataract Refract Surg. 2007;33(12):2111-7.

Publication Dates

  • Publication in this collection
    May-Jun 2016

History

  • Received
    23 Mar 2015
  • Accepted
    23 Jan 2016
Conselho Brasileiro de Oftalmologia Rua Casa do Ator, 1117 - cj.21, 04546-004 São Paulo SP Brazil, Tel: 55 11 - 3266-4000, Fax: 55 11- 3171-0953 - São Paulo - SP - Brazil
E-mail: abo@cbo.com.br