Measuring optic nerve sheath diameter using ultrasonography in patients with idiopathic intracranial hypertension

Medição do diâmetro da bainha do nervo óptico por ultrassonografia em pacientes com hipertensão intracraniana idiopática

Abstract

Background:

Idiopathic intracranial hypertension (IIH) is primarily a disorder of obese young women characterized by symptoms associated with raised intracranial pressure in the absence of a space-occupying lesion.

Objective:

To compare the mean optic nerve sheath diameter (ONSD) measured using ultrasonography (USG) in patients with idiopathic intracranial hypertension (IIH) and normal healthy individuals.

Methods:

A prospective study. Ninety-seven participants aged 18-80 years were divided into two groups as patients with IIH (n=47) and the control group (n=50). The ONSD was measured using ultrasound with a 10-MHz probe. ONSD was measured 3 mm behind the optic disc. Receiver operating characteristic (ROC) curve analysis was performed to determine patients with IIH using ONSD.

Results:

Body mass index was higher in the IIH group compared with the control group (p=0.001). The mean ONSD was statistically significantly thicker in the IIH group (6.4 mm) than in the control group (4.90 mm). The cut-off value of ONSD in patients with IIH was measured as 5.70 mm. There was a significant negative correlation between ONSD and age (r:-0.416 and p<0.001). There was a positive correlation between BMI and ONSD (r: 0.437 and p<0.001).

Conclusions:

Ultrasound can be a reliable, non-invasive and rapid tool to measure ONSD in monitoring patients with IIH. After the first diagnosis of IIH, based on neuroimaging and measuring intracranial pressure using invasive methods, ONSD can be used in treatment and follow-up.

Keywords:
Pseudotumor Cerebri; Ultrasonography; Optic Nerve

Resumo

Antecedentes:

A hipertensão intracraniana idiopática (HII) é primariamente um distúrbio de mulheres jovens obesas caracterizado por sintomas e sinais associados à pressão intracraniana elevada na ausência de uma lesão ocupante de espaço.

Objetivo:

Comparar o diâmetro médio da bainha do nervo óptico (ONSD) medido por ultrassonografia (USG) em pacientes com hipertensão intracraniana idiopática (HII) e indivíduos normais e saudáveis.

Métodos:

Estudo prospectivo. Noventa e sete participantes com idade entre 18-80 anos foram divididos em dois grupos: pacientes com HII (n=47) e o grupo controle (n=50). O ONSD foi medido por ultrassonografia com uma sonda de 10 MHz. O ONSD foi medido 3 mm atrás do disco óptico. A análise da curva ROC foi realizada para determinar pacientes com HII usando-se o ONSD.

Resultados:

O índice de massa corporal foi maior no grupo HII comparado ao grupo controle (p=0,001). O ONSD médio foi estatística e significativamente mais espesso no grupo HII (6,4 mm) do que no grupo controle (4,90 mm). O valor de corte do ONSD em pacientes com HII foi medido em 5,70 mm. Houve correlação negativa significativa entre ONSD e idade (r:-0,416 ep<0,001). Houve correlação positiva entre IMC e ONSD (r: 0,437 ep<0,001).

Conclusões:

A ultrassonografia pode ser uma ferramenta confiável, não invasiva e rápida para medir o ONSD no monitoramento de pacientes com HII. Após o primeiro diagnóstico de HII, com base em neuroimagem e na medida da pressão intracraniana por métodos invasivos, o ONSD pode ser utilizado no tratamento e acompanhamento.

Palavras-chave:
Pseudotumor Cerebral; Ultrassonografia; Nervo Óptico

INTRODUCTION

Idiopathic intracranial hypertension (IIH) is quite common in routine neuro-ophthalmology clinics. The diagnostic criteria for IIH were first introduced by Walter Dandy in 198511. Smith JL. Whence pseudotumor cerebri? J Clin Neuroophthalmol. 1985 Mar;5(1):55-6.. Clinically, it progresses with headaches and loss of vision-visual field in young-middle-aged women22. Rehman H, Khan MS, Nafees M, Rehman AU, Habib A. Optic nerve sheath diameter on sonography in idiopathic intracranial hypertension versus normal. J Coll Physicians Surg Pak. 2016 Sep;26(9):758-60..

In evaluating patients with IIH, lumbar puncture is generally preferred invasively, and magnetic resonance (MRI) and computed tomography (CT) noninvasively. Although invasive techniques are accurate and highly sensitive, they can cause adverse effects such as hemorrhage and infection that need to be managed33. Hagel S, Bruns T, Pletz MW, Engel C, Kalff R, Ewald C. External ventricular drain infections: risk factors and outcome. Interdiscip Perspect Infect Dis. 2014 Nov 17;2014:708531. https://doi.org/10.1155/2014/708531
https://doi.org/10.1155/2014/708531...
. However, CT and MRI are time-consuming, costly, and require patient transport. Therefore, evaluation of optic nerve sheath diameter (ONSD) using ultrasonography (USG), which provides low-cost and fast bedside examination, is a better option, especially in cases where patient transport is difficult, such as in the intensive care unit (ICU)44. Raboel PH, Bartek J Jr, Andresen M, Bellander BM, Romner B. Intracranial pressure monitoring: invasive versus non-invasive methods-a review. Crit Care Res Pract. 2012 Jun 8;2012:950393. https://doi.org/10.1155/2012/950393
https://doi.org/10.1155/2012/950393...
-77. Pershad J, Taylor A, Hall MK, Klimo P. Imaging strategies for suspected acute cranial shunt failure: a cost-effectiveness analysis. Pediatrics. 2017;140(2):e20164263. https://doi.org/10.1542/peds.2017-2852
https://doi.org/10.1542/peds.2017-2852...
. The technique is cheap and effective, and examinations take approximately 5 minutes at the bedside88. Soldatos T, Karakitsos D, Chatzimichail K, Papathanasiou M, Gouliamos A, Karabinis A. Optic nerve sonography in the diagnostic evaluation of adult brain injury. Crit Care. 2008 May 13;12(3):R67. https://doi.org/10.1186/cc6897
https://doi.org/10.1186/cc6897...
. ONSD has been measured as retrobulbar at a distance of 3 mm in most studies99. Bäuerle J, Lochner P, Kaps M, Nedelmann M. Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. J Neuroimaging. 2012 Jan;22(1):42-5. https://doi.org/10.1111/j.1552-6569.2010.00546.x
https://doi.org/10.1111/j.1552-6569.2010...
,1010. Ballantyne SA, O'Neill G, Hamilton R, Hollman AS. Observer variation in the sonographic measurement of optic nerve sheath diameter in normal adults. Eur J Ultrasound. 2002 Oct;15(3):145-9. https://doi.org/10.1016/s0929-8266(02)00036-8
https://doi.org/10.1016/s0929-8266(02)00...
. The optic nerve is surrounded by a dural sheath as part of the central nervous system. There is a small subarachnoid space of 0.1-0.2 mm between the dural sheath and white matter and communicating with the subarachnoid space surrounding the brain. When intracranial pressure (ICP) increases, the dural sheath expands and changes in the diameter of the sheath can be demonstrated using transocular USG55. Sahu S, Swain A. Optic nerve sheath diameter: a novel way to monitor the brain. J Neuroanaesth Crit Care. 2017 Feb 1;4(04):80109296. https://doi.org/10.4103/JNACC.JNACC_73_16
https://doi.org/10.4103/JNACC.JNACC_73_1...
.

This study was conducted to review the effectiveness of USG in the evaluation of patients with IIH and to compare the findings with other studies in the literature.

METHODS

This study was conducted as a case-control study in Ankara Provincial Health Directorate Beytepe Murat Erdi Eker State Hospital department of ophthalmology and neurology, Ankara, Turkey in the first quarter of 2021. The study protocol was approved by the ethics committee (registration number is E1/1541/2021). Fifty healthy volunteers and 47 patients with IIH were included in the study. The inclusion criteria for the study were as follows: age 18-80 years, diagnosed with IIH for a maximum period of 12 months, no additional disease, no drug history, no active or previous intraocular and orbital infections, no ophthalmologic disease other than refractive error, no history of eye-orbital-cranial surgery, less than -5.00 D (Diopter) and +3.00 D refractive error, no history of eye or head injury, no history of radiotherapy to the head and orbital region. Informed consent was obtained from all participants. The diagnosis of IIH was made by a neurologist according to the Dandy criteria. The patients underwent detailed neurologic examinations, ICP measurements, and lumbar puncture by a neurologist. Patients in the IIH group were also subjected to examinations by an ophthalmologist to exclude other causes of optic disc edema and diagnose papilledema. Participants were divided into two groups as patients with IIH and healthy volunteers. Measurements were taken in the supine position using a 10-MHz probe speed real-time ultrasound device (ultrasound scanner model E-Z Scan 5500+ by Sonomed Inc. NY) from the participants in both groups, with the probe placed in the superolateral of the globe with the upper eyelid closed. Only one eye of all participants was evaluated. The optic nerve head was visualized as a linear hypoechoic structure. ONSD was measured three times by the same investigator and the mean value was calculated. ONSD was measured 3 mm behind the optic nerve head (optic disk) as the transverse length of the optic nerve sheath (Figure 1).

Figure 1.
Transverse length of the optic nerve sheath, starting 3 mm behind the optic nerve head.

Statistical analysis

The data were recorded on a spreadsheet and analyses were performed after transfer to the SPSS version 25 software. Frequency and percentage values were used to describe categorical data, and Chi-square tests were used for comparison. The compatibility of continuous data to normal distribution was tested using the Shapiro-Wilk test. The continuous data did not conform to normal distribution and parametric assumptions were not met; thus these variables were described with median and 25-75% interquartile range (IQR) values. The Mann-Whitney U test was used to compare quantitative variables. Spearman’s correlation coefficient was used for the comparison of continuous variables. Receiver operating characteristic (ROC) curve analysis was performed to determine patients with IIH using ONSD. Any p-value of <0.05 was accepted as demonstrating significance.

Power Analysis was performed through the PASS 11 program. Using the ONSD values of the patient and control groups in the study by Rehman et al.22. Rehman H, Khan MS, Nafees M, Rehman AU, Habib A. Optic nerve sheath diameter on sonography in idiopathic intracranial hypertension versus normal. J Coll Physicians Surg Pak. 2016 Sep;26(9):758-60., it was concluded that each group should consist of at least 35 people with an 80% power level and 0.05 alpha errors.

RESULTS

The mean disease duration of patients with IIH was 3.85 ± 2.67 months. Ninety-seven eyes of 97 patients included in the study were examined. There was no significant difference between the two groups in terms of right and left eye distributions (p=0.917). In the IIH group, 36 (76.6%) people were female and 11 (23.4%) were male, and in the control group, 34 (68.0%) were female and 16 (32.0%) were male (Table 1).

Table 1.
Distribution of the two groups by sex and eye care included.

The mean age was 28 (range, 26-32) years in the IIH group, and 38 (range, 32-45) years in the control group; the IIH group was significantly younger (p=0.001). Body mass index (BMI) and blood pressure (diastolic) were higher in the IIH group compared with the control group (p=0.001 and p=0.039, respectively). The mean ONSD was statistically significantly thicker in the IIH group (6.4 mm) compared with the control group (4.90 mm) (p=0.001). There was no significant difference between the groups in terms of spherical equivalent (diopter), glycated hemoglobin (HbA1C) (mMol0/L), and blood pressure (systolic) values. The demographic data and examination findings of the two groups are given in Table 2. In the IHH group, 57.4% of the cases were receiving acetazolamide treatment.

Table 2.
Demographics and optic nerve sheath diameter values in controls and patients with Idiopathic intracranial hypertension.

Table 3 shows the correlation of the ONSD and demographic data. A high level of negative correlation was observed between ONSD and age (r: -0.416 and p<0.001). A high level of positive correlation was observed between BMI and ONSD (r: 0.437 and p<0.001). Moderate positive correlation was observed between HbA1C (mMol/L) and ONSD (r: 0.227 and p=0.025). There was no significant correlation between ONSD and disease duration, blood pressure, and spherical equivalent. The mean cerebrospinal fluid (CSF) pressure was 269.72 ± 66.29 mmH2O in the IHH group. A high level of positive correlation was determined between CSF pressure and ONSD (r: 0.740 and p<0.001).

Table 3.
Correlation of optic nerve sheath diameter and demographics.

According to the ROC analysis, the cut-off points of 5.70 mm, showed 100% sensitivity and 98% specificity (AUC: 0.999 [0.996-1.000]) (Figure 2).

Figure 2.
Receiver operating characteristic (ROC) analysis.

DISCUSSION

In our study, the efficiency of USG in the evaluation of patients with IIH and the measured mean ONSD value and cut-off value was compared with other studies in the literature. IIH is seen in young women. ONSD was highly correlated with age (r: -0.416, p<0.001) and BMI (r: 0.437, p<0.001). The cut-off value of ONSD in patients with IIH was determined as 5.70 mm.

ONSD measurement using USG is a simple, fast, non-invasive, and reliable method. Although lumbar puncture is the gold standard for measuring ICP, studies have reported that ONSD has a positive correlation with ICP1111. Kerscher SR, Schöni D, Neunhoeffer F, Wolff M, Haas-Lude K, Bevot A, et al. The relation of optic nerve sheath diameter (ONSD) and intracranial pressure (ICP) in pediatric neurosurgery practice - Part II: Influence of wakefulness, method of ICP measurement, intra-individual ONSD-ICP correlation and changes after therapy. Childs Nerv Syst. 2020 Jan;36(1):107-15. https://doi.org/10.1007/s00381-019-04336-4
https://doi.org/10.1007/s00381-019-04336...
-1313. Zhang X, Medow JE, Iskandar BJ, Wang F, Shokoueinejad M, Koueik J, et al. Invasive and noninvasive means of measuring intracranial pressure: a review. Physiol Meas. 2017 Jul 24;38(8):R143-82. https://doi.org/10.1088/1361-6579/aa7256
https://doi.org/10.1088/1361-6579/aa7256...
. ONSD values are measured at a fixed distance, and measurements taken as retrobulbar at a distance of 3 mm are considered reliable in the literature22. Rehman H, Khan MS, Nafees M, Rehman AU, Habib A. Optic nerve sheath diameter on sonography in idiopathic intracranial hypertension versus normal. J Coll Physicians Surg Pak. 2016 Sep;26(9):758-60.,99. Bäuerle J, Lochner P, Kaps M, Nedelmann M. Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. J Neuroimaging. 2012 Jan;22(1):42-5. https://doi.org/10.1111/j.1552-6569.2010.00546.x
https://doi.org/10.1111/j.1552-6569.2010...
,1414. Shirodkar CG, Munta K, Rao SM, Mahesh MU. Correlation of measurement of optic nerve sheath diameter using ultrasound with magnetic resonance imaging. Indian J Crit Care Med. 2015 Aug;19(8):466-70. https://doi.org/10.4103/0972-5229.162465
https://doi.org/10.4103/0972-5229.162465...
,1515. Kishk NA, Ebraheim AM, Ashour AS, Badr NM, Eshra MA. Optic nerve sonographic examination to predict raised intracranial pressure in idiopathic intracranial hypertension: the cut-off points. Neuroradiol J. 2018 Oct;31(5):490-5. https://doi.org/10.1177/1971400918789385
https://doi.org/10.1177/1971400918789385...
. In our study, the average ONSD values were calculated using B-mode USG from a retrobulbar 3 mm distance. However, as with all USG evaluations, it requires training. It has intra and interobserver variance, but these variations are small. In recent studies, the mean intraobserver variance was found as ± 0.1-0.2 mm, and the mean interobserver variance was ± 0.2-0.3 mm99. Bäuerle J, Lochner P, Kaps M, Nedelmann M. Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. J Neuroimaging. 2012 Jan;22(1):42-5. https://doi.org/10.1111/j.1552-6569.2010.00546.x
https://doi.org/10.1111/j.1552-6569.2010...
,1010. Ballantyne SA, O'Neill G, Hamilton R, Hollman AS. Observer variation in the sonographic measurement of optic nerve sheath diameter in normal adults. Eur J Ultrasound. 2002 Oct;15(3):145-9. https://doi.org/10.1016/s0929-8266(02)00036-8
https://doi.org/10.1016/s0929-8266(02)00...
. ONSD can also be measured using MRI, but the patient must be transported for MRI. Bedside evaluation of patients provides a great advantage, especially in ICPs and emergency departments. Studies comparing USG and MRI on this subject are also available in the literature. In a study comparing ONSD values measured using USG and MRI, it was concluded that ONSD values taken from a retrobulbar 3 mm distance showed a high level of correlation between the two methods1616. Price DA, Grzybowski A, Eikenberry J, Januleviciene I, Vercellin ACV, Mathew S, et al. Review of non-invasive intracranial pressure measurement techniques for ophthalmology applications. Br J Ophthalmol. 2020 Jul;104(7):887-92. https://doi.org/10.1136/bjophthalmol-2019-314704
https://doi.org/10.1136/bjophthalmol-201...
. Chen et al. showed ultrasonographic measurements of ONSD can dynamically and sensitively evaluate real-time ICP. In their study, ONSD measurements were performed approximately 5 min prior to and after a lumbar puncture. They found significant correlations between ONSD and ICP before lumbar punctures and between the median change in ONSD and the change in ICP1717. Chen L-M, Wang L-J, Hu Y, Jiang X-H, Wang Y-Z, Xing Y-Q. Ultrasonic measurement of optic nerve sheath diameter: a non-invasive surrogate approach for dynamic, real-time evaluation of intracranial pressure. Br J Ophthalmol. 2019 Apr;103(4):437-41. https://doi.org/10.1136/bjophthalmol-2018-312934
https://doi.org/10.1136/bjophthalmol-201...
. Padayachy et al. reported in child patients, the ONSD measurement with the best diagnostic accuracy for detecting an ICP ≥ 20 mmHg over the entire patient cohort was 5.5 mm, sensitivity 93.2 %, specificity 74 % and odds ratio of 39.31818. Padayachy LC, Padayachy V, Galal U, Gray R, Fieggen AG. The relationship between transorbital ultrasound measurement of the optic nerve sheath diameter (ONSD) and invasively measured ICP in children: Part I: repeatability, observer variability and general analysis. Childs Nerv Syst. 2016 Oct;32(10):1769-78. https://doi.org/10.1007/s00381-016-3067-5
https://doi.org/10.1007/s00381-016-3067-...
. Tekin Orkun et al. also indicated the mean cerebrospinal fluid opening pressure (37.75 ± 12.64 cm H2O) and the mean ONSD (5.94 ± 0.46 mm) were correlated in small sample size of pediatric IIH patients1919. Orgun LT, Atalay HT, Arhan E, Aydın K, Serdaroglu A. Optic nerve ultrasonography in monitoring treatment efficacy in pediatric idiopathic intracranial hypertension. Childs Nerv Syst. 2020 Jul;36(7):1425-33. https://doi.org/10.1007/s00381-019-04497-2
https://doi.org/10.1007/s00381-019-04497...
. In contrast to these studies, Lochner et al. reported no correlation was demonstrated between ONSD and cerebrospinal fluid opening pressure in adult IHH patients. They also found no differences in optic nerve diameter values between patients and control groups2020. Lochner P, Brio F, Zedde ML, Sanguigni S, Coppo L, Nardone R, et al. Feasibility and usefulness of ultrasonography in idiopathic intracranial hypertension or secondary intracranial hypertension. BMC Neurol. 2016 Jun 2;16:85. https://doi.org/10.1186/s12883-016-0594-3
https://doi.org/10.1186/s12883-016-0594-...
.

In our study, in accordance with the literature, ONSD values were significantly higher in the IIH group than in the control group22. Rehman H, Khan MS, Nafees M, Rehman AU, Habib A. Optic nerve sheath diameter on sonography in idiopathic intracranial hypertension versus normal. J Coll Physicians Surg Pak. 2016 Sep;26(9):758-60.,99. Bäuerle J, Lochner P, Kaps M, Nedelmann M. Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. J Neuroimaging. 2012 Jan;22(1):42-5. https://doi.org/10.1111/j.1552-6569.2010.00546.x
https://doi.org/10.1111/j.1552-6569.2010...
,1010. Ballantyne SA, O'Neill G, Hamilton R, Hollman AS. Observer variation in the sonographic measurement of optic nerve sheath diameter in normal adults. Eur J Ultrasound. 2002 Oct;15(3):145-9. https://doi.org/10.1016/s0929-8266(02)00036-8
https://doi.org/10.1016/s0929-8266(02)00...
. As frequently reported in the literature, IIH is mostly seen in young women with obesity11. Smith JL. Whence pseudotumor cerebri? J Clin Neuroophthalmol. 1985 Mar;5(1):55-6.,22. Rehman H, Khan MS, Nafees M, Rehman AU, Habib A. Optic nerve sheath diameter on sonography in idiopathic intracranial hypertension versus normal. J Coll Physicians Surg Pak. 2016 Sep;26(9):758-60.. Zheng et al. found a correlation between ONSD and BMI using high-resolution MRI. Their result indicates that the effects of BMI should be considered along with the ONSD during ICP monitoring. Meanwhile, the correlation index between ONSD and BMI was better than the ONSD in predicting IIH and could be used to obtain a more precise estimation of ICP2121. Zheng Y-M, Hao D-P, Tang G-Z, Zhou R-Z, Pang J, Dong C. High-resolution MRI assessment of optic nerve sheath diameter in adults: optic nerve sheath variation and a new diagnostic tool for intracranial hypertension. Acta Radiol. 2020 Oct 21:62(10):1397-403. https://doi.org/10.1177/0284185120966715
https://doi.org/10.1177/0284185120966715...
. As a different point of view, Lochner at al. claims to monitor the efficacy of diet and pharmacological treatment in IIH patients2222. Lochner P, Nardone R, Tezzon F, Coppo L, Brigo F. Optic nerve sonography to monitor treatment efficacy in idiopathic intracranial hypertension: a case report. J Neuroimaging. 2013 Oct;23(4):533-4. https://doi.org/10.1111/jon.12005
https://doi.org/10.1111/jon.12005...
. In our study, it was observed that the IIH group consisted mostly of young female patients with higher BMI values. In addition, a high level of correlation was observed between ONSD values and age (r: -0.416, p<0.001) and BMI (r: 0.437, p<0.001).

There is no single cut-off value for ONSD. This should be considered as a clear indication of abnormal or elevated ICP. In our study, the mean ONSD value was 6.40 mm in the IIH group and 4.90 mm in the control group. The best ONSD cut-off value indicating increased ICP was determined as 5.70 mm (100% sensitivity and 98% specificity) with an area under the curve (AUC) of 0.999. Shrestha et al. concluded that 95% of normal individuals in Nepal had an average ONSD value of 4.41 mm2323. Shrestha GS. Transorbital sonographic evaluation of normal optic nerve sheath diameter in healthy Nepalese adults. J Neuroanaesth Crit Care. 2016 Apr;3(2):115-8. https://doi.org/10.4103/2348-0548.182343
https://doi.org/10.4103/2348-0548.182343...
. Dubourg et al., in their meta-analysis, concluded that ONSD had a cut-off value of 5.10 mm2424. Dubourg J, Javouhey E, Geeraerts T, Messerer M, Kassai B. Ultrasonography of optic nerve sheath diameter for detection of raised intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2011 Jul;37(7):1059-68. https://doi.org/10.1007/s00134-011-2224-2
https://doi.org/10.1007/s00134-011-2224-...
. In a study conducted in China, Wang et al. reported that the mean ONSD value was 4.33 ± 0.38 mm in normal individuals and 6.61 ± 0.39 mm in patients with IIH2525. Wang L, Feng L, Yao Y, Wang Y, Chen Y, Feng J, et al. Optimal optic nerve sheath diameter threshold for the identification of elevated opening pressure on lumbar puncture in a Chinese population. PLoS One. 2015 Feb 9;10(2):e0117939. https://doi.org/10.1371/journal.pone.0117939
https://doi.org/10.1371/journal.pone.011...
. Kishk et al. measured the ONSD cut-off value as 6.05 mm (73.2% sensitivity and 91.4% specificity) with an AUC of 0.8501414. Shirodkar CG, Munta K, Rao SM, Mahesh MU. Correlation of measurement of optic nerve sheath diameter using ultrasound with magnetic resonance imaging. Indian J Crit Care Med. 2015 Aug;19(8):466-70. https://doi.org/10.4103/0972-5229.162465
https://doi.org/10.4103/0972-5229.162465...
. Fernando et al. showed in their study in patients with increased intracranial pressure for various reasons, the pooled AUC curve for ONSD sonography was 0.94 (0.91 to 0.96)2626. Fernando SM, Tran A, Cheng W, Rochwerg B, Taljaard M, Kyeremanteng K, et al. Diagnosis of elevated intracranial pressure in critically ill adults: systematic review and meta-analysis. BMJ. 2019 Jul 24;366:l4225. https://doi.org/10.1136/bmj.l4225
https://doi.org/10.1136/bmj.l4225...
. Li et al. reported in their study AUC analysis showed the ONSD of 5.6 mm was the best cutoff value with a sensitivity of 86% and a specificity of 71% for identifying high ICP2727. Li Z, Zhang XX, Yang HQ, Zhao LP, Jia JP, Sun F, et al. Correlation between ultrasonographic optic nerve sheath diameter and intracranial pressure. Zhonghua Yan Ke Za Zhi. 2018 Sep 11;54(9):683-7. https://doi.org/10.3760/cma.j.issn.0412-4081.2018.09.009
https://doi.org/10.3760/cma.j.issn.0412-...
. Del Saz-Saucedo et al. found that the best cut-off point for detecting raised ICP was 6.3 mms, with a sensitivity, specificity and positive likelihood ratio of 94.7%, 90.9% and 10.4, respectively. After a therapeutic lumbar puncture an 87% of cases had a partial reduction of ONSD values2828. del Saz-Saucedo P, Redondo-González O, Mateu-Mateu Á, Huertas-Arroyo R, García-Ruiz R, Botia-Paniagua E. Sonographic assessment of the optic nerve sheath diameter in the diagnosis of idiopathic intracranial hypertension. J Neurol Sci. 2016 Feb 15;361:122-7. https://doi.org/10.1016/j.jns.2015.12.032
https://doi.org/10.1016/j.jns.2015.12.03...
. According to the ROC analysis of our study, the cut-off point of 5.70 mm, showed 100% sensitivity and 98% specificity (AUC: 0.999 [0.996-1.000]). The sensitivity and specificity rates are the highest rates when compared to the other studies. Despite considerable debate on the normal and abnormal cut-off value of ONSD in different populations, this ultrasonographic measurement easy and noninvasive research method cannot be underestimated.

The limitation of our study is that when ONSD values are measured using USG, patients' ICPs are not measured, and they cannot be compared with ONSD values. Secondly, the distribution of age was different between the IHH and control groups. Some studies showed a correlation between age and ONSD2929. Cardim D, Czosnyka M, Chandrapatham K, Badenes R, Bertuccio A, Noto AD, et al. Effects of age and sex on optic nerve sheath diameter in healthy volunteers and patients with traumatic brain injury. Front Neurol. 2020 Aug 7;11:764. https://doi.org/10.3389/fneur.2020.00764
https://doi.org/10.3389/fneur.2020.00764...
, while others did not3030. Avci M, Kozacı N, Komut E, Komut S, Caliskan G, Tulubas G. The measurement of elderly volunteers’ optic nerve sheath diameters by ocular ultrasonography. Medicina (Kaunas). 2019 Jul 27;55(8):413. https://doi.org/10.3390/medicina55080413
https://doi.org/10.3390/medicina55080413...
. In our study, age was negatively correlated with ONSD. The heterogeneous distribution of age in the groups may also have affected the ONSD value. This was not taken into account when interpreting our results.

In conclusion, USG is a reliable, cheap, non-invasive and fast tool for measuring ONSD in the monitoring of patients with IIH. Although many studies have been conducted on the ONSD cut-off value as measured using USG, there is no consensus among authors yet. Our study made significant contributions to the literature in terms of determining ONSD cut-off values and demonstrating the diagnostic value of ONSD in patients with IIH. The values of the optic nerve sheath can replace cerebrospinal fluid pressure in the diagnosis or follow-up of these patients. Studies on this subject should be conducted on larger data sets and with a longer follow-up period. Considering the results of our study and comprehensive future studies, ONSD value may be considered as a preferable option in the diagnosis of IIH patients.

ACKNOWLEDGEMENTS

We are grateful to Beytepe Murat Erdi Eker State Hospital.

References

  • 1. Smith JL. Whence pseudotumor cerebri? J Clin Neuroophthalmol. 1985 Mar;5(1):55-6.
  • 2. Rehman H, Khan MS, Nafees M, Rehman AU, Habib A. Optic nerve sheath diameter on sonography in idiopathic intracranial hypertension versus normal. J Coll Physicians Surg Pak. 2016 Sep;26(9):758-60.
  • 3. Hagel S, Bruns T, Pletz MW, Engel C, Kalff R, Ewald C. External ventricular drain infections: risk factors and outcome. Interdiscip Perspect Infect Dis. 2014 Nov 17;2014:708531. https://doi.org/10.1155/2014/708531
    » https://doi.org/10.1155/2014/708531
  • 4. Raboel PH, Bartek J Jr, Andresen M, Bellander BM, Romner B. Intracranial pressure monitoring: invasive versus non-invasive methods-a review. Crit Care Res Pract. 2012 Jun 8;2012:950393. https://doi.org/10.1155/2012/950393
    » https://doi.org/10.1155/2012/950393
  • 5. Sahu S, Swain A. Optic nerve sheath diameter: a novel way to monitor the brain. J Neuroanaesth Crit Care. 2017 Feb 1;4(04):80109296. https://doi.org/10.4103/JNACC.JNACC_73_16
    » https://doi.org/10.4103/JNACC.JNACC_73_16
  • 6. Ohle R, McIsaac SM, Woo MY, Perry JJ. Sonography of the optic nerve sheath diameter for detection of raised intracranial pressure compared to computed tomography: a systematic review and meta-analysis. J Ultrasound Med. 2015 Jul;34(7):1285-94. https://doi.org/10.7863/ultra.34.7.1285
    » https://doi.org/10.7863/ultra.34.7.1285
  • 7. Pershad J, Taylor A, Hall MK, Klimo P. Imaging strategies for suspected acute cranial shunt failure: a cost-effectiveness analysis. Pediatrics. 2017;140(2):e20164263. https://doi.org/10.1542/peds.2017-2852
    » https://doi.org/10.1542/peds.2017-2852
  • 8. Soldatos T, Karakitsos D, Chatzimichail K, Papathanasiou M, Gouliamos A, Karabinis A. Optic nerve sonography in the diagnostic evaluation of adult brain injury. Crit Care. 2008 May 13;12(3):R67. https://doi.org/10.1186/cc6897
    » https://doi.org/10.1186/cc6897
  • 9. Bäuerle J, Lochner P, Kaps M, Nedelmann M. Intra- and interobsever reliability of sonographic assessment of the optic nerve sheath diameter in healthy adults. J Neuroimaging. 2012 Jan;22(1):42-5. https://doi.org/10.1111/j.1552-6569.2010.00546.x
    » https://doi.org/10.1111/j.1552-6569.2010.00546.x
  • 10. Ballantyne SA, O'Neill G, Hamilton R, Hollman AS. Observer variation in the sonographic measurement of optic nerve sheath diameter in normal adults. Eur J Ultrasound. 2002 Oct;15(3):145-9. https://doi.org/10.1016/s0929-8266(02)00036-8
    » https://doi.org/10.1016/s0929-8266(02)00036-8
  • 11. Kerscher SR, Schöni D, Neunhoeffer F, Wolff M, Haas-Lude K, Bevot A, et al. The relation of optic nerve sheath diameter (ONSD) and intracranial pressure (ICP) in pediatric neurosurgery practice - Part II: Influence of wakefulness, method of ICP measurement, intra-individual ONSD-ICP correlation and changes after therapy. Childs Nerv Syst. 2020 Jan;36(1):107-15. https://doi.org/10.1007/s00381-019-04336-4
    » https://doi.org/10.1007/s00381-019-04336-4
  • 12. Lochner P, Czosnyka M, Naldi A, Lyros E, Pelosi P, Mathur S, et al. Optic nerve sheath diameter: present and future perspectives for neurologists and critical care physicians. Neurol Sci. 2019 Dec;40(12):2447-57. https://doi.org/10.1007/s10072-019-04015-x
    » https://doi.org/10.1007/s10072-019-04015-x
  • 13. Zhang X, Medow JE, Iskandar BJ, Wang F, Shokoueinejad M, Koueik J, et al. Invasive and noninvasive means of measuring intracranial pressure: a review. Physiol Meas. 2017 Jul 24;38(8):R143-82. https://doi.org/10.1088/1361-6579/aa7256
    » https://doi.org/10.1088/1361-6579/aa7256
  • 14. Shirodkar CG, Munta K, Rao SM, Mahesh MU. Correlation of measurement of optic nerve sheath diameter using ultrasound with magnetic resonance imaging. Indian J Crit Care Med. 2015 Aug;19(8):466-70. https://doi.org/10.4103/0972-5229.162465
    » https://doi.org/10.4103/0972-5229.162465
  • 15. Kishk NA, Ebraheim AM, Ashour AS, Badr NM, Eshra MA. Optic nerve sonographic examination to predict raised intracranial pressure in idiopathic intracranial hypertension: the cut-off points. Neuroradiol J. 2018 Oct;31(5):490-5. https://doi.org/10.1177/1971400918789385
    » https://doi.org/10.1177/1971400918789385
  • 16. Price DA, Grzybowski A, Eikenberry J, Januleviciene I, Vercellin ACV, Mathew S, et al. Review of non-invasive intracranial pressure measurement techniques for ophthalmology applications. Br J Ophthalmol. 2020 Jul;104(7):887-92. https://doi.org/10.1136/bjophthalmol-2019-314704
    » https://doi.org/10.1136/bjophthalmol-2019-314704
  • 17. Chen L-M, Wang L-J, Hu Y, Jiang X-H, Wang Y-Z, Xing Y-Q. Ultrasonic measurement of optic nerve sheath diameter: a non-invasive surrogate approach for dynamic, real-time evaluation of intracranial pressure. Br J Ophthalmol. 2019 Apr;103(4):437-41. https://doi.org/10.1136/bjophthalmol-2018-312934
    » https://doi.org/10.1136/bjophthalmol-2018-312934
  • 18. Padayachy LC, Padayachy V, Galal U, Gray R, Fieggen AG. The relationship between transorbital ultrasound measurement of the optic nerve sheath diameter (ONSD) and invasively measured ICP in children: Part I: repeatability, observer variability and general analysis. Childs Nerv Syst. 2016 Oct;32(10):1769-78. https://doi.org/10.1007/s00381-016-3067-5
    » https://doi.org/10.1007/s00381-016-3067-5
  • 19. Orgun LT, Atalay HT, Arhan E, Aydın K, Serdaroglu A. Optic nerve ultrasonography in monitoring treatment efficacy in pediatric idiopathic intracranial hypertension. Childs Nerv Syst. 2020 Jul;36(7):1425-33. https://doi.org/10.1007/s00381-019-04497-2
    » https://doi.org/10.1007/s00381-019-04497-2
  • 20. Lochner P, Brio F, Zedde ML, Sanguigni S, Coppo L, Nardone R, et al. Feasibility and usefulness of ultrasonography in idiopathic intracranial hypertension or secondary intracranial hypertension. BMC Neurol. 2016 Jun 2;16:85. https://doi.org/10.1186/s12883-016-0594-3
    » https://doi.org/10.1186/s12883-016-0594-3
  • 21. Zheng Y-M, Hao D-P, Tang G-Z, Zhou R-Z, Pang J, Dong C. High-resolution MRI assessment of optic nerve sheath diameter in adults: optic nerve sheath variation and a new diagnostic tool for intracranial hypertension. Acta Radiol. 2020 Oct 21:62(10):1397-403. https://doi.org/10.1177/0284185120966715
    » https://doi.org/10.1177/0284185120966715
  • 22. Lochner P, Nardone R, Tezzon F, Coppo L, Brigo F. Optic nerve sonography to monitor treatment efficacy in idiopathic intracranial hypertension: a case report. J Neuroimaging. 2013 Oct;23(4):533-4. https://doi.org/10.1111/jon.12005
    » https://doi.org/10.1111/jon.12005
  • 23. Shrestha GS. Transorbital sonographic evaluation of normal optic nerve sheath diameter in healthy Nepalese adults. J Neuroanaesth Crit Care. 2016 Apr;3(2):115-8. https://doi.org/10.4103/2348-0548.182343
    » https://doi.org/10.4103/2348-0548.182343
  • 24. Dubourg J, Javouhey E, Geeraerts T, Messerer M, Kassai B. Ultrasonography of optic nerve sheath diameter for detection of raised intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2011 Jul;37(7):1059-68. https://doi.org/10.1007/s00134-011-2224-2
    » https://doi.org/10.1007/s00134-011-2224-2
  • 25. Wang L, Feng L, Yao Y, Wang Y, Chen Y, Feng J, et al. Optimal optic nerve sheath diameter threshold for the identification of elevated opening pressure on lumbar puncture in a Chinese population. PLoS One. 2015 Feb 9;10(2):e0117939. https://doi.org/10.1371/journal.pone.0117939
    » https://doi.org/10.1371/journal.pone.0117939
  • 26. Fernando SM, Tran A, Cheng W, Rochwerg B, Taljaard M, Kyeremanteng K, et al. Diagnosis of elevated intracranial pressure in critically ill adults: systematic review and meta-analysis. BMJ. 2019 Jul 24;366:l4225. https://doi.org/10.1136/bmj.l4225
    » https://doi.org/10.1136/bmj.l4225
  • 27. Li Z, Zhang XX, Yang HQ, Zhao LP, Jia JP, Sun F, et al. Correlation between ultrasonographic optic nerve sheath diameter and intracranial pressure. Zhonghua Yan Ke Za Zhi. 2018 Sep 11;54(9):683-7. https://doi.org/10.3760/cma.j.issn.0412-4081.2018.09.009
    » https://doi.org/10.3760/cma.j.issn.0412-4081.2018.09.009
  • 28. del Saz-Saucedo P, Redondo-González O, Mateu-Mateu Á, Huertas-Arroyo R, García-Ruiz R, Botia-Paniagua E. Sonographic assessment of the optic nerve sheath diameter in the diagnosis of idiopathic intracranial hypertension. J Neurol Sci. 2016 Feb 15;361:122-7. https://doi.org/10.1016/j.jns.2015.12.032
    » https://doi.org/10.1016/j.jns.2015.12.032
  • 29. Cardim D, Czosnyka M, Chandrapatham K, Badenes R, Bertuccio A, Noto AD, et al. Effects of age and sex on optic nerve sheath diameter in healthy volunteers and patients with traumatic brain injury. Front Neurol. 2020 Aug 7;11:764. https://doi.org/10.3389/fneur.2020.00764
    » https://doi.org/10.3389/fneur.2020.00764
  • 30. Avci M, Kozacı N, Komut E, Komut S, Caliskan G, Tulubas G. The measurement of elderly volunteers’ optic nerve sheath diameters by ocular ultrasonography. Medicina (Kaunas). 2019 Jul 27;55(8):413. https://doi.org/10.3390/medicina55080413
    » https://doi.org/10.3390/medicina55080413