Use of a minimally invasive uncalibrated cardiac output monitor in patients undergoing cesarean section under spinal anesthesia: report of four cases

Abstracts

BACKGROUND AND OBJECTIVES: Hemodynamic changes are observed during cesarean section under spinal anesthesia. Non-invasive blood pressure (BP) and heart rate (HR) measurements are performed to diagnose these changes, but they are delayed and inaccurate. Other monitors such as filling pressure and cardiac output (CO) catheters with external calibration are very invasive or inaccurate. The objective of the present study was to report the cardiac output measurements obtained with a minimally invasive uncalibrated monitor (LiDCO rapid) in patients undergoing cesarean section under spinal anesthesia. CASE REPORT: After approval by the Ethics Commission, four patients agreed to participate in this study. They underwent cesarean section under spinal anesthesia while at the same time being connected to the LiDCO rapid by a radial artery line. Cardiac output, HR, and BP were recorded at baseline, after spinal anesthesia, after fetal and placental extraction, and after the infusion of oxytocin and metaraminol. We observed a fall in BP with an increase of HR and CO after spinal anesthesia and oxytocin infusion; and an increase in BP with a fall in HR and CO after bolus of the vasopressor. CONCLUSIONS: Although this monitor had not been calibrated, it showed a tendency for consistent hemodynamic data in obstetric patients and it may be used as a therapeutic guide or experimental tool

Anesthesia, Spinal; Monitoring, Intraoperative; Hemodynamics; Cardiac Output; Hypotension; Cesarean Section


JUSTIFICATIVA E OBJETIVOS: Durante cesariana sob raquianestesia observam-se alterações hemodinâmicas. São realizadas medidas de pressão arterial (PA) não invasiva e de frequência cardíaca (FC) para diagnosticar essas alterações, mas com atraso e imprecisão. Outros monitores como cateteres de pressão de enchimento e débito cardíaco (DC) com calibração externa são muito invasivos ou imprecisos. O objetivo deste estudo foi relatar as medidas de débito cardíaco obtidas por um monitor minimamente invasivo não calibrado (LiDCO rapid) em pacientes submetidas à cesariana sob raquianestesia. RELATO DO CASO: Após aprovação da comissão de ética, quatro pacientes consentiram em participar do estudo. Elas foram submetidas à cesariana sob raquianestesia quando estavam conectadas ao LiDCO rapid por uma linha arterial radial. Os dados de DC, FC e BP foram registrados no momento basal, após instalação da raquianestesia, após extração fetal e placentária e após infusão de ocitocina e metaraminol. Foram observadas queda da PA, com aumento da FC e DC após a raquianestesia e infusão de ocitocina; e aumento da BP, com queda da FC e DC após bolus de vasopressor. CONCLUSÕES: Embora não calibrado, esse monitor produziu tendência de dados consistentes sobre a hemodinâmica de pacientes obstétricas e pode ser usado como guia terapêutico ou como ferramenta de pesquisa

ANESTESIA; ANESTESIA; CIRURGIA; COMPLICAÇÕES; CUIDADOS; TÉCNICAS DE MEDIÇÃO; TÉCNICAS DE MEDIÇÃO


JUSTIFICATIVA Y OBJETIVOS: Durante la cesárea bajo raquianestesia, se observaron alteraciones hemodinámicas. La medidas de presión arterial (PA), no invasiva y de frecuencia cardíaca (FC), para diagnosticar esas alteraciones se realizan, pero con atraso e imprecisión. Otros monitores como catéteres de presión de llenado y débito cardíaco (DC), con calibración externa son muy invasivos o imprecisos. El objetivo de este estudio, fue relatar las medidas de débito cardíaco obtenidas por un monitor mínimamente invasivo no calibrado (LiDCO rapid), en pacientes sometidas a la cesárea bajo raquianestesia. RELATO DEL CASO: Después de la aprobación por parte de la comisión de ética, cuatro pacientes estuvieron de acuerdo en participar en el estudio. Se sometieron entonces a la cesárea bajo raquianestesia cuando estaban conectadas al LiDCO rapid por una línea arterial radial. Los datos de DC, FC y BP se registraron al momento basal, después de la instalación de la raquianestesia, después de la extracción fetal y placentaria, y después de la infusión de ocitocina y metaraminol. Observamos una caída de la PA, con un aumento de la FC y DC posterior a la raquianestesia e infusión de ocitocina; y un aumento de la BP, con una caída de la FC y DC después de un bolo de vasopresor. CONCLUSIONES: Aunque no esté calibrado, ese monitor nos ofreció datos consistentes sobre la hemodinámica de las pacientes obstétricas y puede ser usado como guía terapéutico o como una herramienta de investigación

ANESTESIA; ANESTESIA; CIRUGÍA; COMPLICACIONES; CUIDADOS; TÉCNICAS DE MEDICIÓN; TÉCNICAS DE MEDICIÓN


CLINICAL INFORMATION

IMedical Degree from FMUSP; Supervisor of Anestesia Obstétrica HC-FMUSP

IIProfessor of FMUSP; Professor of the Disciplina de Anestesiologia of FMUSP

IIIAnesthesiologist; Centro Obstétrico - HCFMUSP

IVMaster's and Medical Degree from Faculdade de Medicina da USP; Technical Director of Serviço de Pré-Natal da Clínica Obstétrica do HC FMUSP; responsible for the sector of Cardiopatia e Gravidez. Member of the sectors of Hipertensão e Trombose na Gravidez

VMedical Degree from FMUSP; Professor of the Disciplina de Obstetrícia of FMUSP

VIProfessor of FMUSP; Disciplina de Obstetrícia of FMUSP

Correspondence to

SUMMARY

BACKGROUND AND OBJECTIVES: Hemodynamic changes are observed during cesarean section under spinal anesthesia. Non-invasive blood pressure (BP) and heart rate (HR) measurements are performed to diagnose these changes, but they are delayed and inaccurate. Other monitors such as filling pressure and cardiac output (CO) catheters with external calibration are very invasive or inaccurate. The objective of the present study was to report the cardiac output measurements obtained with a minimally invasive uncalibrated monitor (LiDCO rapid) in patients undergoing cesarean section under spinal anesthesia.

CASE REPORT: After approval by the Ethics Commission, four patients agreed to participate in this study. They underwent cesarean section under spinal anesthesia while at the same time being connected to the LiDCO rapid by a radial artery line. Cardiac output, HR, and BP were recorded at baseline, after spinal anesthesia, after fetal and placental extraction, and after the infusion of oxytocin and metaraminol. We observed a fall in BP with an increase of HR and CO after spinal anesthesia and oxytocin infusion; and an increase in BP with a fall in HR and CO after bolus of the vasopressor.

CONCLUSIONS: Although this monitor had not been calibrated, it showed a tendency for consistent hemodynamic data in obstetric patients and it may be used as a therapeutic guide or experimental tool.

Keywords: Anesthesia, Spinal; Monitoring, Intraoperative; Hemodynamics; Cardiac Output; Hypotension; Cesarean Section.

INTRODUCTION

Spinal anesthesia has been consolidating as the technique of choice for cesarean section worldwide 1 due to the quality of the blockade and safety of the technique, using small anesthetic mass and not requiring ventilatory assistance and tracheal intubation.

However, it produces a non-selective blockade: the synapses in sensorial, motor, and autonomic fibers are temporarily interrupted. The autonomous blockade is regional, but intense enough to lead to severe hemodynamic changes, such as hypotension. The sympathetic blockade is not the only cause of hypotension; aortocaval compression by the fetus 2 and infusion of oxytocin 3 also contribute.

The diagnosis of hemodynamic compromise during cesarean section under spinal anesthesia is usually non-invasive, such as heart rate and non-invasive blood pressure monitors, used as substitute markers of maternal cardiac output 4. However, the delay and the lack of accuracy of these parameters in reflecting the fall in uterine-placental blood flow are significant 5. Cardiac filling pressure catheters, besides being very invasive, also may be unreliable. Cardiac output (CO) is considered one of the most accurate parameters to detect these changes 6 and has better correlation with uterine-placental blood flow 7. With the advent of minimally invasive equipment to measure CO, the opportunity to investigate this parameter in the obstetric population has emerged. However, most of these equipments require external calibration resulting in a more cumbersome operation and greater invasiveness. Therefore, the objective of the present study was to observe the behavior of CO measured by a minimally invasive uncalibrated monitor (LiDCO rapid) in patients undergoing cesarean section under spinal anesthesia.

METHODS

After approval by the Ethics Committee of the institution, patients referred to the pre-delivery unit of the Obstetric Center were questioned whether they wanted to participate in the study. Detailed information on the protocol and procedures emphasizing the voluntariness of participation without any repercussions on treatment in case they did not wish to participate was provided. Four patients agreed to participate and they signed an informed consent. Exclusion criteria, besides refusal to give consent, were not established.

The patient was transferred to the operating room where venous cannulation was performed with an 18G catheter. This was followed by puncture of the radial artery with a 22G venous catheter for invasive blood pressure monitoring, which was connected to a minimally invasive uncalibrated cardiac output (CO) monitor (LiDCO rapid - Pulse CO system, LiDCO Ltda, Cambridge, GB). This equipment uses the algorithm of pulse power analysis with correction for anthropometric characteristics for individual arterial complacency, and external calibration was not performed. Thus, it converts changes in blood pressure over time on an estimate of the systolic volume and CO (nominal value), which can be used in the analysis of tendencies 8.

The first measures of CO, BP, and heart rate (HR) were recorded with the patient in partial left lateral decubitus over a 15 degree wedge under the right gluteal region, considered the baseline measurements. The patient was, then, anesthetized with the standardized spinal anesthesia for cesarean section of the institution consisting of 15 mg of 0.5% hyperbaric bupivacaine associated with 80 µg of 0.02% morphine through a 27G Whitacre needle. The patient was once more placed on a 15 degree wedge on the right gluteal region resulting in uterine dislocation to the left 9 until after the surgical fields were placed, at which time the wedge was removed. Crystalloids were not administered before the spinal anesthesia 10, but immediately after it, resulting in co-hydration with 10 mL.kg-1 of Ringer's lactate 11. Blood pressure was constantly monitored and any fall was immediately corrected to guarantee fetal well-being 12 with a bolus of 200 g of metaraminol IV 13, a predominantly alpha-agonist vasopressor 14,15, and the measurements were recorded at this moment as "araminol". When incision of the skin was performed, hemodynamic parameters were recorded as "anesthesia", and we expected to be recording the peak of the sympathetic blockade. As soon as the fetus was extracted, followed by delivery of the afterbirth, new measurements were recorded as "birth". According to the protocol of the institution, 3 IU IV 16 of oxytocin over at least 60 seconds were infused, followed by recording of the measurements as "oxytocin". The infusion was repeated according to the evaluation of the uterine tonus by the obstetrician, up to a limit of 9 IU 17. No patient required more oxytocin or other classes of uterotonics, such as ergot derivatives or prostaglandins. The "final" measurement was recorded when closing of the aponeurosis. The arterial catheter was removed and pressure was applied to the puncture site for five minutes, and the patient was followed-up for eight hours for detection of any vascular complications. Data was stored and tabulated on the memory card of the LiDCO rapid monitor and treated by its specific software, the LiDCO view.

CASE 1

This is a 39-year old, 86 kg, 165 cm, gesta 6 (2 cesarean sections and 3 abortions) with a diagnosis of a 34-week twin pregnancy, and acute fetal distress, which were the indications for cesarean section. Her physical status was classified as ASA II, since she had gestational diabetes on insulin, and hypertension controlled with alpha-methyldopa (1.5 g.day-1). The case developed according to that described in "methods", and the measurements are shown in Table I.

CASE 2

This is a 27-year old female, 89 kg, 162 cm, gesta 1, with diagnosis of a single fetus pregnancy on pelvic presentation, in labor, with indication of cesarean section. She was classified as physical status (ASA) II, since she had gestational diabetes controlled with diet. The case developed according to that described in "methods", and her measurements are shown in Table II.

CASE 3

This is a 31-year old female, 106 kg, 179 cm, gesta 2, with a history of cesarean section, classified as ASA II, with a diagnosis of gestational diabetes controlled with diet, controlled systemic lupus erythematosus, fetal macrosomia, and iterativeness, and cesarean section was indicated. Measurements obtained after the onset of the surgery are shown in Table III.

CASE 4

This is a 32-year old female, 63 kg, 149 cm, physical status ASA I, gesta 2, with a history of cesarean section, with a diagnosis of twin pregnancy in which one fetus was on pelvic presentation, with indication for cesarean section. Measurements are shown in Table IV.

DISCUSSION

In general, the behavior of hemodynamic variables observed during interventions in patients undergoing cesarean sections under spinal anesthesia was:

1. After installation of the blockade, a fall in MAP with occasional elevation of HR and CO is observed;

2. When using the vasopressor metaraminol, a predominantly alpha-agonist drug, an increase in MAP with fall in HR and CO is observed;

3. When the fetus is extracted, the behavior is not uniform, with occasional elevation of CO and fall in MAP;

4. After infusion of oxytocin, marked changes are observed, such as fall in MAP besides an increase in HR and CO.

In case 1, after the blockade, measurements are close to baseline possibly because the sympathetic blockade did not achieve its peak or the effectiveness of conservative measures like dislocation of the uterus to the left and co-hydration. Unlike cases 2 and 4, who presented initially a compensatory increase of CO secondary to an increase in HR without compromising the BP, but which later manifested with a significant fall in BP. Since the beginning, case 3 showed significant hemodynamic changes with fall in MAP and increase in HR and CO. In response to the fall in MAP, boluses of the alphaagonist vasopressor metaraminol were used, resulting in a typical response 18, with an increase in MAP and fall in HR and CO, most likely due to an increase in afterload, except in case 2 in which only after fetal extraction a recovery in MAP was observed. Regarding delivery, in all cases, only after fetal extraction a similar response was observed with an increase in CO, which can be explained by self-transfusion resulting from uterine involution and contraction, as well as reduction in aortocaval compression, which increased preload. Oxytocin has vasodilator actions, with fall in MAP and increase in HR and CO, which were observed in all cases except in case 2. At the end, parameters returned to baseline levels most likely due to all interventions with less intense actions. The behavior of MAP demonstrated that its variations, most likely due to changes in systemic vascular resistance (SVR), are secondary to vasodilator (spinal anesthesia and oxytocin) and vasoconstrictor (metaraminol) actions of interventions performed during cesarean sections that affect arterial tonus and not due to bleeding. Regarding the absence of SVR measurements as a limiting factor of this study, we decide not to use it, since it is an invasive measurement because it requires a right atrium catheter to measure pressure, which is required to its calculation.

Therefore, physiologic situations are observed, such as aortocaval decompression after birth and self-transfusion of uterine blood, that impose a great overload on the cardiovascular system with an increase of venous return, which is accommodated through the cardiac valves and pulmonary blood vessels. However, this may be a problem in patients with low cardiovascular reserve or with cardiopathies, such as mitral and aortic stenosis, besides pulmonary hypertension, resulting in pulmonary edema or cardiac failure. The same can be seen in situations that are induced like spinal anesthesia and oxytocin infusion in which normal patients respond with an increase in CO and fall in SVR and venous return, but in cardiopathies, these responses should be controlled or even avoided.

This cardiac output monitor produced data compatible with those of prior publications 3,6,8,18,19, and in this scenery, external calibration with data obtained by thermodilution or lithium dilution was dispensable, especially when analysis is made based on tendencies and nominal values and not absolute values. Measurements produced can orient regarding a possible cause of hemodynamic changes and indicate the most appropriate treatment for the case, besides signaling reckless conducts, such as rapid infusion of oxytocin. They can also be useful in patients with limited cardiovascular reserve, since they cannot tolerate significant changes in SVR and they cannot compensate for changes in CO.

To conclude, it is a fact that HR, BP, and the communication with the awaken patient assessing her well-being are important parameters and the most commonly used in obstetric anesthesia. However, the minimally invasive uncalibrated CO monitors can have an important role in hemodynamic management of gravidas, especially high-risk gravidas with associated comorbidities 4 or as a research tool.

REFERENCES

  • 1. Shibli KU, Russell IF - A survey of anaesthetic techniques used for caesarean section in the UK in 1997. Int J Obstet Anesth, 2000;9:160-7.
  • 2. Hartley H, Seed PT, Ashworth H et al. - Effect of lateral versus supine wedged position on development of spinal blockade and hypotension. Int J Obstet Anesth, 2001;10:182-8.
  • 3. Archer TL, Knape K, Liles D et al. - The hemodynamics of oxytocin and other vasoactive agents during neuraxial anesthesia for cesarean delivery: findings in six cases. Int J Obstet Anesth, 2008;17:247-54.
  • 4. Dyer RA, James MF - Maternal hemodynamic monitoring in obstetric anesthesia. Anesthesiology, 2008;109:765-7.
  • 5. Langesaeter E - Is it more informative to focus on cardiac output than blood pressure during spinal anesthesia for cesarean delivery in women with severe preeclampsia? Anesthesiology, 2008;108:771-2.
  • 6. Langesaeter E, Rosseland LA, Stubhaug A - Continuous invasive blood pressure and cardiac output monitoring during cesarean delivery: a randomized, double-blind comparison of low-dose versus high-dose spinal anesthesia with intravenous phenylephrine or placebo infusion. Anesthesiology, 2008;109:856-63.
  • 7. Robson SC, Boys RJ, Rodeck C et al. - Maternal and fetal haemodynamic effects of spinal and extradural anaesthesia for elective caesarean section. Br J Anaesth, 1992;68:54-59.
  • 8. Archer TL, Heitmeyer JD - Perioperative hemodynamics obtained by pulse contour analysis facilitated the management of a patient with chronic hypertension, renal insufficiency, and superimposed preeclampsia during cesarean delivery. J Clin Anesth, 2010;22:274-9.
  • 9. Sharwood-Smith G, Drummond GB - Hypotension in obstetric spinal anaesthesia: a lesson from pre-eclampsia. Br J Anaesth, 2009;102:291-4.
  • 10. Rout CC, Akoojee SS, Rocke DA et al. - Rapid administration of crystalloid preload does not decrease the incidence of hypotension after spinal anaesthesia for elective caesarean section. Br J Anaesth, 1992;68:394-7.
  • 11. Ngan Kee WD, Khaw KS, Ng FF - Prevention of hypotension during spinal anesthesia for cesarean delivery: an effective technique using combination phenylephrine infusion and crystalloid cohydration. Anesthesiology, 2005;103:744-50.
  • 12. Macarthur A, Riley ET - Obstetric anesthesia controversies: vasopressor choice for postspinal hypotension during cesarean delivery. Int Anesthesiol Clin, 2007;45:115-32.
  • 13. Ngan Kee WD, Khaw KS, Ng FF et al. - Prophylactic phenylephrine infusion for preventing hypotension during spinal anesthesia for cesarean delivery. Anesth Analg, 2004;98:815-21.
  • 14. Lee SW, Khaw KS, Ngan Kee WD et al. - Management of hypotension in obstetric spinal anaesthesia. Br J Anaesth, 2009;103:457-8.
  • 15. Vallejo MC, Ramanathan S - Should alpha-agonists be used as first line management of spinal hypotension? Int J Obstet Anesth, 2003;12:243-5.
  • 16. Carvalho JC, Balki M, Kingdom J et al. - Oxytocin requirements at elective cesarean delivery: a dose-finding study. Obstet Gynecol, 2004;104:1005-10.
  • 17. Tsen LC, Balki M - Oxytocin protocols during cesarean delivery: time to acknowledge the risk/benefit ratio? Int J Obstet Anesth, 2010;19:243-5.
  • 18. Dyer RA, Piercy JL, Reed AR et al. - Hemodynamic changes associated with spinal anesthesia for cesarean delivery in severe preeclampsia. Anesthesiology, 2008;108:802-11.
  • 19. Dyer RA, Reed AR, van Dyk D et al. - Hemodynamic effects of ephedrine, phenylephrine, and the coadministration of phenylephrine with oxytocin during spinal anesthesia for elective cesarean delivery. Anesthesiology, 2009;111:753-65.
  • Use of a minimally invasive uncalibrated cardiac output monitor in patients undergoing cesarean section under spinal anesthesia: report of four cases

    Fernando Bliacheriene, TSAI; Maria José Carvalho Carmona, TSAII; Cristina de Freitas Madeira BarrettiIII; Cristiane Maria Federicci HaddadIII; Elaine Soubhi MouchalwatIII; Maria Rita de Figueiredo Lemos BortolottoIV; Rossana Pulcineli Vieira FranciscoV; Marcelo ZugaibVI
  • Publication Dates

    • Publication in this collection
      28 Sept 2011
    • Date of issue
      Oct 2011

    History

    • Received
      27 Oct 2010
    • Accepted
      31 Jan 2011
    Sociedade Brasileira de Anestesiologia R. Professor Alfredo Gomes, 36, 22251-080 Botafogo RJ Brasil, Tel: +55 21 2537-8100, Fax: +55 21 2537-8188 - Campinas - SP - Brazil
    E-mail: bjan@sbahq.org