Factors that interfere with the response of nurses in the monitoring of clinical alarms

ABSTRACT

The aim of the present study was to reflect on nursing practice regarding the monitoring of clinical alarms.

Objective:

The objective of the present study was to identify and synthesize the best empirical evidence found on factors that influence the response of nurses regarding clinical alarms.

Method:

An integrative literature review was conducted with searches undertaken in ten electronic databases restricted to the period from 2005 to 2016.

Results:

Eight articles were included by cross-checking the descriptors selected.

Conclusion:

In the analysis of the studies, the following possible factors that might interfere with the response of nurses in the monitoring of clinical alarms were found: high number of false alarms, inaudibility of alarms due to the competition of sounds, difficulty in distinguishing the urgency of alarms, and increase in noise caused by the raise in the number of alarms.

Descriptors:
Clinical Alarms; Noise; Nursing; Professional Competence; Safety

RESUMO

O presente estudo tem como propósito refletir sobre a prática de enfermagem no que concerne à temática da monitorização dos alarmes clínicos. Objetivo: Pretendemos identificar e sintetizar, a melhor, a evidência empírica produzida sobre os fatores que influenciam a resposta dos enfermeiros perante os alarmes clínicos.

Método:

Perspetivou-se um estudo de revisão integrativa da literatura. Selecionámos um conjunto de dez bases de dados eletrónicas, delimitou-se a pesquisa ao período temporal de 2005 a 2016.

Resultados:

Por meio de uma estratégia de cruzamento dos descritores selecionados, foram incluídos oito artigos.

Conclusão:

Na análise dos estudos, reconhecemos como possíveis fatores que interferem na resposta dos enfermeiros na monitorização dos alarmes clínicos: o grande número de falsos alarmes, a inaudibilidade dos alarmes por causa da competição de sons, a dificuldade de distinguir a urgência dos alarmes e o aumento do ruído provocado pelo aumento do número de alarmes.

Descritores:
Alarmes Clínicos; Ruído; Enfermagem; Competência Profissional; Segurança

RESUMEN

El presente estudio tiene como propósito reflexionar sobre la práctica de enfermería en lo que concierne a la temática del monitoreo de las alarmas clínicas.

Objetivo:

Pretendemos identificar y sintetizar, la mejor, la evidencia empírica producida sobre los factores que influyen la respuesta de los enfermeros frente a las alarmas clínicas.

Método:

Se tuvo una perspectiva de un estudio de revisión integradora de la literatura. Seleccionamos un conjunto de diez bases de datos electrónicos, se delimitó la pesquisa al periodo temporal de 2005 a 2016.

Resultados:

Por medio de una estrategia de cruzamiento de los descriptores seleccionados, fueron incluidos ocho artículos.

Conclusión:

En el análisis de los estudios, reconocemos como posibles factores que interfieren en la respuesta de los enfermeros en el monitoreo de las alarmas clínicos: el gran número de falsas alarmas, lo inaudible de las alarmas por causa de la competición de sonidos, la dificultad de distinguir la urgencia de las alarmas y el aumento de ruido provocado por el aumento del número de alarmas.

Descriptores:
Alarmas Clínicas; Ruido; Enfermería; Competencia Profesional; Seguridad

INTRODUCTION

In recent years, there has been a significant increase in specialization of care and consequent acquisition of more technologically sophisticated equipment in hospitals. All this equipment is provided with optical and audible alarms that, in addition to the background noise of devices, create a potentially uncomfortable environment for patients and caregivers(11 Cropp AJ, Woods LA, Raney D, Bredle DL. Name that tone: the proliferation of alarms in the intensive care unit. Chest. 1994 Apr;105(4):1217-20.).

In most hospitals, noise levels exceed the levels recommended by the World Health Organization (WHO) of 40 decibels (dB) during the day and 30 dB during the night(22 Berglund B. Lindvall T, Schwela DH. Guidelines for Community Noise. World Health Organization [Internet]. 1999[cited 2014 Dec 18]. Available from: www.who.int/docstore/peh/noise/guidelines2.html
www.who.int/docstore/peh/noise/guideline...
). Cvach(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.) reinforces that noise levels in hospitals have increased significantly since 1960.

Nowadays, an endless number of devices that sound incessantly demands our attention. Cardiac monitors, ventilators, heating systems, feeding pumps, syringes and infusion pumps, among others, are appanage of daily practice.

Considering that nurses are professionals who are permanently close to patients, they are the population most exposed to noise coming from alarms, and both (nurses and patients) might be exposed to about 700 alarms of cardiac monitors per day(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.), which might represent occupational risks, compromise the safety of care, and affect patients’ recovery.

The value of technology is allied with human competence, since clinical alarms are considered an essential and life-saving key tool. The Joint Commission on Accreditation of Healthcare Organizations defines clinical alarm as “any alarm that is intended to protect the individual receiving care or alert the staff that the individual is at increased risk and needs immediate assistance”(44 Phillips J, Barnsteiner J. Clinical alarms: improving efficiency and effectiveness. Crit Care Nurs Q. 200;28(4):317-23.).

The purpose of equipment alarms is to alert professionals for potential problems and serious or dangerous situations. However, they might also compromise the quality of work of nurses and patient safety due to the abundance of false positives. False positive alarms are understood as those that do not assume clinical relevance at the time or are caused by technical problems or artifacts(55 Imhoff M, Kuhls S. Alarm algorithms in critical care monitoring. Anesth Analg. 2006;102(5):1525-37.). These may be generated not only by the equipment’s hypersensitivity, but also by the inadequacy of the parameters’ limits to the clinical condition of each patient. Therefore, unnecessary noise might lead to nurses’ desensitization, thus making them ignore, silence or even disconnect alarms(66 Graham KC, Cvach M. Monitor alarm fatigue: standardizing use of physiological monitors and decreasing nuisance alarms. Am J Crit Care. 2010 Jan;19(1):28-34.).

The intensity and frequency of alarms in hospitals are of unquestionable importance. Many studies show that the increase in the number of alarms created a unsafe, noisy and annoying environment, especially in intensive care units and operating rooms(77 Gurses AP, Martinez EA, Bauer L, Kim G, Lubomski LH, Marsteller JA, Pennathur PR, Goeschel C, Pronovost PJ, Thompson D. Using human factors engineering to improve patient safety in the cardiovascular operating room. Work [Internet]. 2012[cited 2014 Dec 18];41(Suppl1):1801-4. Available from: http://content.iospress.com/articles/work/wor0388
http://content.iospress.com/articles/wor...
).

Considering that humans’ cognitive abilities have limits, each time nurses respond to an alarm, their attention is redirected, and care related to other patients need to be reprogrammed. Several researchers described this phenomenon as the “cry wolf” effect, in which the human behavior in the response to alarms is adjusted according to the rate of false alarms perceived(88 Bliss JP, Dunn, MC. Behavioural implications of alarm mistrust as a function of task workload. Ergonomics [Internet]. 2000[cited 2014 Dec 18];43(9):1283-300. Available from: http://www.tandfonline.com/doi/abs/10.1080/001401300421743
http://www.tandfonline.com/doi/abs/10.10...
-99 Korniewicz DM, Clark T, David Y. A national online survey on the effectiveness of clinical alarms. Am J Crit Care [Internet]. 2008[cited 2014 Dec 18];17(1):36-41. Available from: http://ajcc.aacnjournals.org/content/17/1/36.long
http://ajcc.aacnjournals.org/content/17/...
). When there is a low rate of false alarms, human response is more appropriate, whereas with high rates of false alarms, natural human response is to respond with less frequency. This is an adaptive human mechanism triggered to manage the overload of cognitive resources and consequently of attention.

Therefore, the purpose of the present study was to reflect on nursing practice regarding the monitoring of clinical alarms based on the following guiding question: Which factors interfere with the response of nurses in the monitoring of clinical alarms?

OBJECTIVE

Identifying factors that interfere with the performance of nurses, as well as behavior changes that might be adopted in order to optimize the provision of care were considered important in this study. The choice for the topic is explained by the concern for the safety of critically ill patients, who depend on the continuous monitoring and surveillance of clinical alarms, for which nurses are responsible.

In addition to answering the research question, the objective of the present study was to identify and synthesize the best empirical evidence found on factors that influence the response of nurses regarding clinical alarms.

METHOD

Ethical aspects

No conflicts of interest were found. It is worth mentioning that the authors who provided scientific support throughout the development of the study were identified, as well as the citation of their references, in order to value and support their intellectual property.

Study design

Considering the topic and objectives of the present study, an integrative literature review was carried out.

Study protocol

Two descriptors were identified for the development of the present study: “Clinical alarms” and “Nurs*”, validated by means of the Medical Subject Headings - MeSH. Subsequently, a set of ten electronic databases was selected: CINAHL Complete, MEDLINE Complete, Database of Abstracts of Reviews of Effects, Cochrane Central Register of Controlled Trials, Cochrane Database of Systematic Reviews, Cochrane Methodology Register, Library, Information Science & Technology Abstracts, MedicLatina, Health Technology Assessments, and NHS Economic Evaluation Database. The search was carried out in Portuguese and English languages and restricted to articles from 2005 to 2016. Data collection occurred between July and August 2016.

Sample and inclusion criteria

The PI[C]OD method was used for definition of selection criteria and composition of the sample: participants (and clinical condition), intervention, outcomes, and design (type of study), as presented in Chart 1

Chart 1
Inclusion criteria of the studies to be selected
(1010 Ramalho A. Manual para redacção de estudos e projectos de revisão sistemática com e sem metanálise. Estrutura funções e utilização na investigação em enfermagem. Coimbra: Formasau; 2005.).

The initial sample was made up of 156 studies. Later, they were evaluated and selected considering criteria established by means of their title and abstract. Full reading was carried out if justified by the abovementioned criteria.

In this study, descriptors were submitted to cross-checking among them, and the advanced search form available at the aforementioned databases was used as strategy, thus obtaining 156 articles (excluding repeated ones). Of the total, 71 were excluded by reading their titles, 63 by reading their abstracts, and 14 by reading the full articles. Therefore, eight articles were included. After that, analysis, evaluation, and synthesis of the empirical evidence were carried out, considering the aforementioned selection criteria as a guiding line. The information obtained was organized in order to highlight the most relevant aspects of the phenomenon in study.

RESULTS

The presentation of the scientific articles selected was outlined with the aim of organizing evidences produced (Chart 2). The results were categorized by title, year, country of origin, design/number of participants, interventions, and findings/main conclusions. The articles were organized based on the year of publication in order to evidence the most recent knowledge.

Chart 2
Synthesis of the evidence found

DISCUSSION

The results from the phenomenon in study were discussed towards a critical reflection guided by its objectives. In order to enhance this analysis, not only studies from the sample selected were discussed, but also others from the same thematic area, which were considered relevant, even though they did not meet the selection criteria.

Among the eight articles selected, which were all extracted from scientific journals, three used a quantitative methodological approach, three had a qualitative approach, one was an integrative review, and one was a systematic literature review. All articles selected were international. During the research, national studies referring to this theme were not found, which might demonstrate the little interest existing in the management of alarms by healthcare professionals. Of the total search results, most articles were not associated with the objectives of this study, which is the reason why they were disregarded.

The reduced size of the sample might suggest that practices of nurses regarding clinical alarms were not a topic much studied in the last ten years.

In hospitals, the number of special care units, such as intensive care units and operating rooms is increasing. These are environments where patients are connected to ventilatory devices and surrounded by a high number of medical equipment that produces different types of alarms, according to the condition of patients and the equipment.

Studies such as the one by Korniewaz et al.(99 Korniewicz DM, Clark T, David Y. A national online survey on the effectiveness of clinical alarms. Am J Crit Care [Internet]. 2008[cited 2014 Dec 18];17(1):36-41. Available from: http://ajcc.aacnjournals.org/content/17/1/36.long
http://ajcc.aacnjournals.org/content/17/...
) showed that adverse events (incidents that result in harm for patients) associated with alarms still occur, in spite of the abundance of alarms in devices. In the comments of the 406 nurses who participated in the study of Honan et al., one adverse event (caused by the lack of attention to alarms, and which resulted in a patient’s death) and six near-miss events (incident that did not cause injury, illness or harm - but had potential to do it) were found(1212 Honan L, Funk M, Maynard M, Fahs D, Clark T, David Y. Nurses’ perspectives on clinical alarms. Am J Crit Care [Internet]. 2015[cited 2014 Dec 18];24(5):387-95. Available from: http://ajcc.aacnjournals.org/content/24/5/387.long
http://ajcc.aacnjournals.org/content/24/...
).

The complexity in the management of clinical alarms might be one of the explanations for the alarming frequency of adverse events, which shows the importance in identifying factors that interfere with the response of professionals to alarms.

In the analysis of the studies selected, the following possible factors that might interfere with the response of nurses in the monitoring of clinical alarms were found: high number of false alarms; inaudibility of alarms due to the competition of sounds; difficulty in distinguishing the urgency of alarms; and increase in noise caused by the raise in the number of alarms. Therefore, a brief analysis of the aforementioned factors was carried out.

Increase in noise caused by the raise in the number of alarms

The review by Konkani et al.(1515 Konkani A, Oakley B, Bauld TJ. Reducing hospital noise: a review of medical device alarm management. Biomed Instrum Technol [Internet]. 2012[cited 2014 Dec 18];46(6):478-87. Available from: http://www.aami-bit.org/doi/pdf/10.2345/0899-8205-46.6.478
http://www.aami-bit.org/doi/pdf/10.2345/...
) mentions a study by Sengpiel in 2011, which showed that the sound of most equipment alarms ranges in the interval of 60 to 70 dB and some exceed 80 dB - the equivalent to the sound heard from 25 meters of a busy freeway.

Intensive care patients and professionals report that the noise in these units is high, and studies confirm this. There are several noise sources; however, conversation among healthcare professionals (reaching values between 59 and 90 dB) and equipment alarms are often mentioned as the most disturbing for patients. The study by Lawson et al.(1616 Lawson N, Thompson K, Saunders G, Saiz J, Richardson J, Brown D, et al. Sound intensity and noise evaluation in a critical care unit. Am J Crit Care [Internet]. 2010[cited 2014 Dec 18];19(6):e88-98. Available from: http://ajcc.aacnjournals.org/content/19/6/e88
http://ajcc.aacnjournals.org/content/19/...
) showed that noise peaks of alarms inside patients’ rooms are high and they increase as the level of alarm volume increases. It is worth mentioning that, although there was an increase in the noise in rooms, a corresponding increase in noise in adjacent rooms was not found. In the measurements, the average noise level inside patients’ rooms was mostly lower than 45 dB. Noise peaks were always higher than 85 dB. Closing the door of adjacent rooms did not reduce noise peaks. Peaks and average noise levels did not differ systematically during 24 hours of measurement.

Honan et al. mention in their study that patients and their families described continuous alarms as “disturbing”, “frightful”, and inducing unnecessary panic. In the same study, nurses reported that “noise pollution”, resulting partly from the high number of alarms, might be associated with comorbidities, such as “increase in patients’ anxiety”, “lack of sleep”, “ICU psychoses” and “delirium”(1212 Honan L, Funk M, Maynard M, Fahs D, Clark T, David Y. Nurses’ perspectives on clinical alarms. Am J Crit Care [Internet]. 2015[cited 2014 Dec 18];24(5):387-95. Available from: http://ajcc.aacnjournals.org/content/24/5/387.long
http://ajcc.aacnjournals.org/content/24/...
).

The high number of false alarms

Alarm systems are highly sensitive, but with little specificity, resulting in high rates of false alarms (about 99.4%). The proliferation of monitoring equipment and the frequency of sensory overload caused by the noise of alarms represent a risk for the development of desensitization of professionals to alarms and consequently for their silencing or even deactivation, as ways to alleviate the problem(1414 Gazarian PK. Nurses’ response to frequency and types of electrocardiography alarms in a non-critical care setting: a descriptive study. Int J Nurs Stud [Internet]. 2014[cited 2014 Dec 18];51(2):190-7. Available from: http://www.journalofnursingstudies.com/article/S0020-7489(13)00157-0/abstract
http://www.journalofnursingstudies.com/a...
).

Parallel to the high sensitivity, if limit levels of the monitored parameter are defined with a very short interval, true but clinically insignificant alarms might occur. These alarms are known as “nuisance” alarms. The study by Sowan et al. showed that most nurses interviewed strongly agreed that “nuisance” alarms are frequent, interrupt the care provided to patients, and reduce the reliability of alarms(1111 Sowan AK, Gomez TM, Tarriela AF, Reed CC, Paper BM. Changes in default alarm settings and standard in-service are insufficient to improve alarm fatigue in an intensive care unit: a pilot project. JMIR Hum Factors [Internet]. 2016[cited 2014 Dec 18];3(1):e1. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797663/pdf/humanfactors_v3i1e1.pdf
https://www.ncbi.nlm.nih.gov/pmc/article...
)). When alarms are considered that way, professionals might disconnect and silence them, or ignore the warning that is intended to make a safer environment. Instead of creating a safer environment, a high number of “nuisance” alarms assumes the opposite effect, resulting in desensitization(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.,1111 Sowan AK, Gomez TM, Tarriela AF, Reed CC, Paper BM. Changes in default alarm settings and standard in-service are insufficient to improve alarm fatigue in an intensive care unit: a pilot project. JMIR Hum Factors [Internet]. 2016[cited 2014 Dec 18];3(1):e1. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797663/pdf/humanfactors_v3i1e1.pdf
https://www.ncbi.nlm.nih.gov/pmc/article...
).

This problem assumes special attention regarding low-priority alarms. For example, alarms of ECG electrodes wrongly applied or that unstick, or alarms of oximeter sensors that accidentally dislocate, typically have tones that are less audible than high-priority alarms. Nurses might not notice and consequently not respond to these alarms, and the non-monitoring of a particular parameter may lead to an eventual non-detection of patients’ critical conditions. Sometimes, professionals might act improperly to avoid frequent “nuisance” alarms, such as by reducing the volume of alarms, extending the limits of alarms beyond the acceptable interval, or even deactivating them(99 Korniewicz DM, Clark T, David Y. A national online survey on the effectiveness of clinical alarms. Am J Crit Care [Internet]. 2008[cited 2014 Dec 18];17(1):36-41. Available from: http://ajcc.aacnjournals.org/content/17/1/36.long
http://ajcc.aacnjournals.org/content/17/...
).

In most situations, alarms distract and interfere with the ability of professionals in effectively performing other critical activities. Alarms also contribute to the desensitization of nurses to devices; in such a way that alarms for “true” events have less probability of calling the attention of the staff. In the study by Korniewicz et al., 77% of the participants agreed, or strongly agreed that “nuisance” alarms interrupt the sequence of care practice. A high number of false alarms is often observed and they contribute to desensitization, lack of reliability, and lack of response from professional caregivers (Cvach(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.), citing Lawless).

Studies support the idea that the increase in the number of alarms reduces “faith” of healthcare professionals in alarms, at the same time it increases the noise level in the hospital environment(1515 Konkani A, Oakley B, Bauld TJ. Reducing hospital noise: a review of medical device alarm management. Biomed Instrum Technol [Internet]. 2012[cited 2014 Dec 18];46(6):478-87. Available from: http://www.aami-bit.org/doi/pdf/10.2345/0899-8205-46.6.478
http://www.aami-bit.org/doi/pdf/10.2345/...
).

The study by Gazarian(1414 Gazarian PK. Nurses’ response to frequency and types of electrocardiography alarms in a non-critical care setting: a descriptive study. Int J Nurs Stud [Internet]. 2014[cited 2014 Dec 18];51(2):190-7. Available from: http://www.journalofnursingstudies.com/article/S0020-7489(13)00157-0/abstract
http://www.journalofnursingstudies.com/a...
) showed that practices of nurses regarding the management of alarms is an area with significant weaknesses. The practice in verifying parameters and volume of alarms at the beginning of shifts ensures that parameters are pre-set and audible. Not following this procedure at the beginning of shifts might have consequences. In the abovementioned study, in 7 of the 18 observations, this procedure was neglected. These observations also showed the difficulty and complexity in managing alarm systems, as well as limitations of nurses in managing several alarms and tasks that compete with these regarding priority.

During observations of monitors’ data and alarms triggered, the author found that these alarms had a minimum impact on the care plan of patients. In 53 of the 161 alarms triggered, the alarm was artifact. This not only results in data without meaning, but also represents a noisy sound and possible interruption of nurses’ regular work, probably withdrawing them from a more significant care procedure.

Inaudibility of alarms and competition of sounds

An alarm is considered audible when it can be heard by healthcare professionals in their environment, which includes background noise such as ventilators, feeding pumps, and people talking. On the other hand, alarms from different equipment compete among each other when triggered at the same time. This range of situations makes the hearing of alarms difficult. Another aspect that must be considered is the identification of alarms regarding their type (for example, if it is about low blood pressure or high heart rate) and the equipment’s source that generates them. However, identifying the urgency of alarms is also necessary. Nurses have to distinguish and correctly identify which equipment is generating the alarm (source), with regard to each specific alarm and its urgency. All these aspects must be considered by equipment manufacturers for their standardization, thus making them more effective and universal among manufacturers(1515 Konkani A, Oakley B, Bauld TJ. Reducing hospital noise: a review of medical device alarm management. Biomed Instrum Technol [Internet]. 2012[cited 2014 Dec 18];46(6):478-87. Available from: http://www.aami-bit.org/doi/pdf/10.2345/0899-8205-46.6.478
http://www.aami-bit.org/doi/pdf/10.2345/...
).

The IEC 60601-1-8 standard that manufacturers must follow, regarding the sound and function of alarms, suggests simple and melodic alarm sounds that enable the distinction of eight different alarm sources (meeting the ability of humans in distinguishing from 5 to 7 categories of sound) and their classification as high-, medium- or low-priority(1717 Block FE Jr. Point-Counterpoint: IEC 62D Medical Monitor Standards Will Make Alarms Worse. Biomed Instrum Technol. 2011 Spring;Suppl:60, 62-3.). However, some studies suggest the review of this standard, indicating that melodic alarm sounds are difficult to identify and distinguish when tasks overlap. These studies concluded that the capture of alarms with melodic sounds is weak and that nurses react faster and more accurately to medium-priority alarms, even though high-priority alarms sound more urgent(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.).

In a study about the audibility of alarms of infusion pumps, the author concluded that these alarms are audible enough and might compete with the environmental noise when the doors of rooms are opened. However, their audibility becomes significantly reduced when the doors of rooms are closed or when there is higher background noise. Therefore, it is important to have signaling devices of adjunct alarms to ensure their audibility(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.).

In the study by Funk et al.(1313 Funk M, Clark JT, Bauld TJ, Ott JC, Coss P. Attitudes and practices related to clinical alarms. Am J Crit Care [Internet]. 2014[cited 2014 Dec 18];23(3):e9-e18. Available from: http://ajcc.aacnjournals.org/content/23/3/e9.long
http://ajcc.aacnjournals.org/content/23/...
), statements such as “difficulty in identifying the alarm source” and “difficulty in hearing alarms when triggered” increased regarding the degree of importance in the questionnaire undertaken in 2011, compared to that undertaken in 2005-2006. The increase in the importance of these two dimensions might be associated with the increase in the number of patients monitored, lack of standardized alarms, and increasing use of equipment with alarms. The authors also indicated the high background noise in hospitals as a factor that contributes to inaudibility.

There is a false argument that the increase in volume levels of alarm sounds facilitates their hearing in adjacent rooms, mostly assisted by the design of some units (where nursing rooms are adjacent to patients’ units). However, Lawson et al.(1616 Lawson N, Thompson K, Saunders G, Saiz J, Richardson J, Brown D, et al. Sound intensity and noise evaluation in a critical care unit. Am J Crit Care [Internet]. 2010[cited 2014 Dec 18];19(6):e88-98. Available from: http://ajcc.aacnjournals.org/content/19/6/e88
http://ajcc.aacnjournals.org/content/19/...
) concluded that levels of these alarms, when measured in adjacent rooms, do not register a higher level of hearing of alarms, thus only causing noise for patients and inaudibility for nurses.

Difficulty in distinguishing the urgency of alarms

The urgency of alarms perceived contributes to the response of nurses to alarms. However, nurses use additional strategies to determine their responses, including the severity rate of patients, signal duration, rare occurrence of alarms, and their workload(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.).

Cited by Cvach(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.), Bitan found that professionals’ probability of response is directly proportional to the alarm response, which is based on the perception rate of alarms as true. If an alarm system is perceived as 90% reliable, the response rate will be approximately 90%; if the alarm system is perceived as 10% reliable, the response rate will be approximately 10%.

Nurses respond to alarms for different reasons and not only because they trigger. Nurses prioritize their interventions through the evaluation of the urgency of alarms regarding the severity rate of patients and have a greater tendency to react to alarms of longer duration and those considered rare. As workload or the complexity of interventions and care to be provided increase, the response to alarms and performance of interventions decline. Therefore, the duration of signals assumes an important influence; however, workload, severity of patients, and complexity of interventions might lead to other reaction/response strategies(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.).

Both the design of devices’ alarm systems and the level of knowledge of professionals contribute to the inappropriate setting of alarms, which justify effective and continuous training reinforced by the study by Sowan et al., in which 50% of nurses reported the need for training on cardiac monitors(1111 Sowan AK, Gomez TM, Tarriela AF, Reed CC, Paper BM. Changes in default alarm settings and standard in-service are insufficient to improve alarm fatigue in an intensive care unit: a pilot project. JMIR Hum Factors [Internet]. 2016[cited 2014 Dec 18];3(1):e1. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797663/pdf/humanfactors_v3i1e1.pdf
https://www.ncbi.nlm.nih.gov/pmc/article...
). Although many alarm systems seem apparently easy, their functioning is not well understood by the staff. A specific case is the existence of several ways to deceive alarm systems of physiological monitoring. An action can silence an existing alarm; another can deactivate all alarms simultaneously; or another can also deactivate alarms indefinitely. The misunderstanding of these differences might lead to inappropriate actions for given circumstances, a situation that might trigger adverse events and affect patient safety(99 Korniewicz DM, Clark T, David Y. A national online survey on the effectiveness of clinical alarms. Am J Crit Care [Internet]. 2008[cited 2014 Dec 18];17(1):36-41. Available from: http://ajcc.aacnjournals.org/content/17/1/36.long
http://ajcc.aacnjournals.org/content/17/...
).

Studies such as those by Mackenzie and Galbrun(1818 MacKenzie D, Galbrun L. Noise Levels and Noise Sources in Acute Care Hospital Wards. Build Serv Eng Res Technol. 2007; 28(2), 117-31.), Funk et al.(1313 Funk M, Clark JT, Bauld TJ, Ott JC, Coss P. Attitudes and practices related to clinical alarms. Am J Crit Care [Internet]. 2014[cited 2014 Dec 18];23(3):e9-e18. Available from: http://ajcc.aacnjournals.org/content/23/3/e9.long
http://ajcc.aacnjournals.org/content/23/...
), and Honan et al.(1212 Honan L, Funk M, Maynard M, Fahs D, Clark T, David Y. Nurses’ perspectives on clinical alarms. Am J Crit Care [Internet]. 2015[cited 2014 Dec 18];24(5):387-95. Available from: http://ajcc.aacnjournals.org/content/24/5/387.long
http://ajcc.aacnjournals.org/content/24/...
) show that hospitals became noisy, thus hearing and identifying the sound of an alarm and providing an appropriate response in due time can sometimes become a difficult problem to solve.

Measures to reduce false alarms

Nurses can implement strategies to reduce the number of unnecessary alarms such as: to improve the preparation of patients’ skin before the application of electrodes; substitute electrodes daily, as well as equipment batteries; proceed with the interruption of alarms for a short period of time before providing some specific care; customize limits of alarms to patients to prevent not operable alarms; and consider the establishment of standard limits specific for the unit. They should also make informed decisions regarding the time when monitoring begins and ends, in order to prevent unnecessary monitoring associated with false alarms(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.,1212 Honan L, Funk M, Maynard M, Fahs D, Clark T, David Y. Nurses’ perspectives on clinical alarms. Am J Crit Care [Internet]. 2015[cited 2014 Dec 18];24(5):387-95. Available from: http://ajcc.aacnjournals.org/content/24/5/387.long
http://ajcc.aacnjournals.org/content/24/...
-1313 Funk M, Clark JT, Bauld TJ, Ott JC, Coss P. Attitudes and practices related to clinical alarms. Am J Crit Care [Internet]. 2014[cited 2014 Dec 18];23(3):e9-e18. Available from: http://ajcc.aacnjournals.org/content/23/3/e9.long
http://ajcc.aacnjournals.org/content/23/...
).

The setting of alarms to the real needs of patients ensures their validity and provides an early warning for potential critical conditions. Registering alarm parameters in the clinical process of patients proved to be an effective intervention to improve compliance regarding alarms settings(33 Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.,1414 Gazarian PK. Nurses’ response to frequency and types of electrocardiography alarms in a non-critical care setting: a descriptive study. Int J Nurs Stud [Internet]. 2014[cited 2014 Dec 18];51(2):190-7. Available from: http://www.journalofnursingstudies.com/article/S0020-7489(13)00157-0/abstract
http://www.journalofnursingstudies.com/a...
).

Hospitals must have appropriate policies to ensure training of healthcare professionals in the correct use of devices and alarm systems. The design of intensive care units must be thought and structured in a way that alarms can always be heard by the staff(99 Korniewicz DM, Clark T, David Y. A national online survey on the effectiveness of clinical alarms. Am J Crit Care [Internet]. 2008[cited 2014 Dec 18];17(1):36-41. Available from: http://ajcc.aacnjournals.org/content/17/1/36.long
http://ajcc.aacnjournals.org/content/17/...
,1111 Sowan AK, Gomez TM, Tarriela AF, Reed CC, Paper BM. Changes in default alarm settings and standard in-service are insufficient to improve alarm fatigue in an intensive care unit: a pilot project. JMIR Hum Factors [Internet]. 2016[cited 2014 Dec 18];3(1):e1. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797663/pdf/humanfactors_v3i1e1.pdf
https://www.ncbi.nlm.nih.gov/pmc/article...
).

Reduction of false alarms must result in a significant decrease in the frequency of alarms, with proportional increase in relevant clinical alarms. With less alarms, the response time of nurses to those clinically relevant will be shorter. The future impact will probably be a more effective use of the time by nurses, a more peaceful environment, and less sentinel events associated with monitoring alarms(1313 Funk M, Clark JT, Bauld TJ, Ott JC, Coss P. Attitudes and practices related to clinical alarms. Am J Crit Care [Internet]. 2014[cited 2014 Dec 18];23(3):e9-e18. Available from: http://ajcc.aacnjournals.org/content/23/3/e9.long
http://ajcc.aacnjournals.org/content/23/...
).

Limitations of the study

The present study presented as limitation the lack of studies associated with the theme explored.

Similarly, the choice of descriptors, databases, and languages for the development of the study might have affected the results obtained.

CONCLUSION

Monitoring of clinical alarms has the purpose of previously warning changes in the clinical condition of patients. However, excessive false alarms might lead to desensitization of professionals, as well as to the interruption of work dynamics. Nurses, through care practice, are one of the professional groups most exposed to the problem of clinical alarms, given their continuous performance with patients.

Based on the research undertaken, critical factors that interfere with the efficiency in the performance of alarms include artifacts of sensors, the reduced consideration for the human factor in equipment design, complexity of alarm systems, inappropriate design of facilities, environmental noise, and little training/knowledge of the staff regarding alarms and equipment with alarms.

Several studies on alarms in the hospital environment agree regarding the existence of excessive low-priority and false alarms that pollute the sound environment, interfere with communication, reduce concentration, increase the probability of errors, and reduce the response rate to alarms.

The development of observational studies and behavioral analysis of nurses regarding clinical alarms are suggested for further studies, with the purpose of optimizing patient safety, thus contributing to the continuous improvement in quality of care.

REFERÊNCIAS

  • 1
    Cropp AJ, Woods LA, Raney D, Bredle DL. Name that tone: the proliferation of alarms in the intensive care unit. Chest. 1994 Apr;105(4):1217-20.
  • 2
    Berglund B. Lindvall T, Schwela DH. Guidelines for Community Noise. World Health Organization [Internet]. 1999[cited 2014 Dec 18]. Available from: www.who.int/docstore/peh/noise/guidelines2.html
    » www.who.int/docstore/peh/noise/guidelines2.html
  • 3
    Cvach M. Monitor alarm fatigue: an integrative review. Biomed Instrum Technol. 2012 Jul-Aug;46(4):268-77.
  • 4
    Phillips J, Barnsteiner J. Clinical alarms: improving efficiency and effectiveness. Crit Care Nurs Q. 200;28(4):317-23.
  • 5
    Imhoff M, Kuhls S. Alarm algorithms in critical care monitoring. Anesth Analg. 2006;102(5):1525-37.
  • 6
    Graham KC, Cvach M. Monitor alarm fatigue: standardizing use of physiological monitors and decreasing nuisance alarms. Am J Crit Care. 2010 Jan;19(1):28-34.
  • 7
    Gurses AP, Martinez EA, Bauer L, Kim G, Lubomski LH, Marsteller JA, Pennathur PR, Goeschel C, Pronovost PJ, Thompson D. Using human factors engineering to improve patient safety in the cardiovascular operating room. Work [Internet]. 2012[cited 2014 Dec 18];41(Suppl1):1801-4. Available from: http://content.iospress.com/articles/work/wor0388
    » http://content.iospress.com/articles/work/wor0388
  • 8
    Bliss JP, Dunn, MC. Behavioural implications of alarm mistrust as a function of task workload. Ergonomics [Internet]. 2000[cited 2014 Dec 18];43(9):1283-300. Available from: http://www.tandfonline.com/doi/abs/10.1080/001401300421743
    » http://www.tandfonline.com/doi/abs/10.1080/001401300421743
  • 9
    Korniewicz DM, Clark T, David Y. A national online survey on the effectiveness of clinical alarms. Am J Crit Care [Internet]. 2008[cited 2014 Dec 18];17(1):36-41. Available from: http://ajcc.aacnjournals.org/content/17/1/36.long
    » http://ajcc.aacnjournals.org/content/17/1/36.long
  • 10
    Ramalho A. Manual para redacção de estudos e projectos de revisão sistemática com e sem metanálise. Estrutura funções e utilização na investigação em enfermagem. Coimbra: Formasau; 2005.
  • 11
    Sowan AK, Gomez TM, Tarriela AF, Reed CC, Paper BM. Changes in default alarm settings and standard in-service are insufficient to improve alarm fatigue in an intensive care unit: a pilot project. JMIR Hum Factors [Internet]. 2016[cited 2014 Dec 18];3(1):e1. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797663/pdf/humanfactors_v3i1e1.pdf
    » https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4797663/pdf/humanfactors_v3i1e1.pdf
  • 12
    Honan L, Funk M, Maynard M, Fahs D, Clark T, David Y. Nurses’ perspectives on clinical alarms. Am J Crit Care [Internet]. 2015[cited 2014 Dec 18];24(5):387-95. Available from: http://ajcc.aacnjournals.org/content/24/5/387.long
    » http://ajcc.aacnjournals.org/content/24/5/387.long
  • 13
    Funk M, Clark JT, Bauld TJ, Ott JC, Coss P. Attitudes and practices related to clinical alarms. Am J Crit Care [Internet]. 2014[cited 2014 Dec 18];23(3):e9-e18. Available from: http://ajcc.aacnjournals.org/content/23/3/e9.long
    » http://ajcc.aacnjournals.org/content/23/3/e9.long
  • 14
    Gazarian PK. Nurses’ response to frequency and types of electrocardiography alarms in a non-critical care setting: a descriptive study. Int J Nurs Stud [Internet]. 2014[cited 2014 Dec 18];51(2):190-7. Available from: http://www.journalofnursingstudies.com/article/S0020-7489(13)00157-0/abstract
    » http://www.journalofnursingstudies.com/article/S0020-7489(13)00157-0/abstract
  • 15
    Konkani A, Oakley B, Bauld TJ. Reducing hospital noise: a review of medical device alarm management. Biomed Instrum Technol [Internet]. 2012[cited 2014 Dec 18];46(6):478-87. Available from: http://www.aami-bit.org/doi/pdf/10.2345/0899-8205-46.6.478
    » http://www.aami-bit.org/doi/pdf/10.2345/0899-8205-46.6.478
  • 16
    Lawson N, Thompson K, Saunders G, Saiz J, Richardson J, Brown D, et al. Sound intensity and noise evaluation in a critical care unit. Am J Crit Care [Internet]. 2010[cited 2014 Dec 18];19(6):e88-98. Available from: http://ajcc.aacnjournals.org/content/19/6/e88
    » http://ajcc.aacnjournals.org/content/19/6/e88
  • 17
    Block FE Jr. Point-Counterpoint: IEC 62D Medical Monitor Standards Will Make Alarms Worse. Biomed Instrum Technol. 2011 Spring;Suppl:60, 62-3.
  • 18
    MacKenzie D, Galbrun L. Noise Levels and Noise Sources in Acute Care Hospital Wards. Build Serv Eng Res Technol. 2007; 28(2), 117-31.

Publication Dates

  • Publication in this collection
    Jan-Feb 2017

History

  • Received
    12 Jan 2016
  • Accepted
    14 Aug 2016
Associação Brasileira de Enfermagem SGA Norte Quadra 603 Conj. "B" - Av. L2 Norte 70830-102 Brasília, DF, Brasil, Tel.: (55 61) 3226-0653, Fax: (55 61) 3225-4473 - Brasília - DF - Brazil
E-mail: reben@abennacional.org.br