Evaluation of the autonomic function in patients with hypertrophic cardiomyopathy with and without syncope

Abstracts

BACKGROUND: Several mechanisms may be involved in the trigger of syncope in patients with hypertrophic cardiomyopathy (HCM), including hemodynamic collapses that might be related to an autonomic imbalance. OBJECTIVE: To evaluate and compare the autonomic function of patients presenting HCM with unexplained syncope (US) to those without syncope. METHODS: Thirty-seven patients were included, 16 with US and 21 without syncope. Their autonomic function was assessed by spontaneous and phenylephrine induced baroreflex sensitivity (BRS), by heart rate variability (HRV) in time domain during 24-hour Holter and in frequency domain (spectral analysis), both in supine position and at 70º head-up tilt (HUT). RESULTS: The spontaneous BRS was similar in both groups (16,46 ± 12,99 vs. 18,31 ± 9,88 ms/mmHg, p = 0,464), as was phenylephrine-induced BRS (18,33 ± 9,31 vs. 15,83 ± 15,48 ms/mmHg, p = 0,521). No differences were observed in SDNN (137,69 ± 36,62 vs . 145,95 ± 38,07 ms, p=0,389). The group presenting syncope had a significantly lower RMSSD (24,88±10,03 vs. 35,58 ± 16,43 ms, p = 0,042) and a tendency to lower pNN50 (4,51 ± 3,78 vs . 8,83 ± 7,98%, p =0,085) and lower values of the high frequency component of HRV spectral analysis at rest (637,59±1.295,53 vs. 782,65±1.264,14ms2, p=0,075). No significant difference was observed in response to HUT (p = 0,053). HUT sensitivity, specificity and accuracy in identifying the etiology of US in HCM patients were 6%, 66% and 40%, respectively. CONCLUSIONS: A lower parasympathetic tone was observed in HCM patients with US, but the clinical relevance of this finding remains unclear. HUT is not a valuable tool for evaluating the origin of syncope in these patients, mainly because of its poor specificity.

Cardiomyopathy, hypertrophic; syncope; autonomic nervous system


FUNDAMENTO: Diversos mecanismos podem estar envolvidos no desencadeamento da síncope em pacientes com cardiomiopatia hipertrófica (CMH), incluindo colapsos hemodinâmicos que podem estar relacionados a um desequilíbrio autonômico. OBJETIVO: Avaliar e comparar a função autonômica de pacientes que apresentam CMH com síncope inexplicada (SI) com os que não apresentam síncope. MÉTODOS: Trinta e sete pacientes foram incluídos, sendo 16 com SI e 21 sem síncope. Sua função autonômica foi avaliada por sensibilidade barorreflexa (SB) espontânea e induzida por fenilefrina, pela variabilidade da frequência cardíaca (VFC) no domínio do tempo durante o Holter de 24 horas e no domínio da frequência (análise espectral), ambos em decúbito dorsal e no teste de inclinação (TI) a 70º. RESULTADOS: A SB espontânea mostrou-se semelhante em ambos os grupos (16,46 ± 12,99 vs. 18,31 ± 9,88 ms/mmHg, p = 0,464), assim como a SB induzida por fenilefrina (18,33 ± 9,31 vs. 15,83 ± 15,48 ms/mmHg, p = 0,521). Não foram observadas diferenças no SDNN (137,69 ± 36,62 vs . 145,95 ± 38,07 ms, p = 0,389). O grupo com síncope apresentou um RMSSD significativamente menor (24,88 ± 10,03 vs. 35,58 ± 16,43 ms, p = 0,042) e tendência a menor pNN50 (4,51 ± 3,78 vs . 8,83 ± 7,98%, p =0,085) e a menores valores do componente de alta frequência da análise espectral da VFC em repouso (637,59 ± 1.295,53 vs. 782,65 ± 1.264,14 ms2 , p = 0,075). Nenhuma diferença significativa foi observada em resposta ao TI (p = 0,053). A sensibilidade, especificidade e acurácia do TI na identificação da etiologia da SI em pacientes com CMH foram 6%, 66% e 40%, respectivamente. CONCLUSÃO: Observou-se tônus parassimpático mais baixo em pacientes com CMH e SI, mas a relevância clínica deste achado ainda não está clara. O TI não é uma ferramenta vantajosa para avaliar a origem da síncope em tais doentes, principalmente por causa da sua baixa especificidade.

Cardiomiopatia Hipertrófica; Síncope; Sistema Nervoso Autônomo


ORIGINAL ARTICLE

IInstituto do Coração - Hospital das Clínicas da FMUSP, São Paulo, SP - Brasil

IICentral Brasileira de Holter, São Paulo, SP - Brasil

Mailing Address

ABSTRACT

BACKGROUND: Several mechanisms may be involved in the trigger of syncope in patients with hypertrophic cardiomyopathy (HCM), including hemodynamic collapses that might be related to an autonomic imbalance.

OBJECTIVE: To evaluate and compare the autonomic function of patients presenting HCM with unexplained syncope (US) to those without syncope.

METHODS: Thirty-seven patients were included, 16 with US and 21 without syncope. Their autonomic function was assessed by spontaneous and phenylephrine induced baroreflex sensitivity (BRS), by heart rate variability (HRV) in time domain during 24-hour Holter and in frequency domain (spectral analysis), both in supine position and at 70º head-up tilt (HUT).

RESULTS: The spontaneous BRS was similar in both groups (16,46 ± 12,99 vs. 18,31 ± 9,88 ms/mmHg, p = 0,464), as was phenylephrine-induced BRS (18,33 ± 9,31 vs. 15,83 ± 15,48 ms/mmHg, p = 0,521). No differences were observed in SDNN (137,69 ± 36,62 vs . 145,95 ± 38,07 ms, p=0,389). The group presenting syncope had a significantly lower RMSSD (24,88±10,03 vs. 35,58 ± 16,43 ms, p = 0,042) and a tendency to lower pNN50 (4,51 ± 3,78 vs . 8,83 ± 7,98%, p =0,085) and lower values of the high frequency component of HRV spectral analysis at rest (637,59±1.295,53 vs. 782,65±1.264,14ms2, p=0,075). No significant difference was observed in response to HUT (p = 0,053). HUT sensitivity, specificity and accuracy in identifying the etiology of US in HCM patients were 6%, 66% and 40%, respectively.

CONCLUSIONS: A lower parasympathetic tone was observed in HCM patients with US, but the clinical relevance of this finding remains unclear. HUT is not a valuable tool for evaluating the origin of syncope in these patients, mainly because of its poor specificity.

Keywords: Cardiomyopathy, hypertrophic; syncope; autonomic nervous system.

Introduction

Unexplained syncope is considered a major risk factor for sudden death in patients with hypertrophic cardiomyopathy (HCM), especially in young people1. Nevertheless, its low sensitivity and specificity is explained by the several possible mechanisms involved in the pathogenesis of this symptom, which may or may not be associated with the underlying cardiac disease2-7.

It has recently been demonstrated that patients with HCM may have an abnormal vascular control, presenting with hypotension during exercise8,9 or even during daily activities10, as a result of an exaggerated drop in systemic vascular resistance, because of an impaired forearm vasoconstriction or a paradoxical vasodilatation. This abnormality may determine syncope or near syncope10 and is probably related to a deficiency in the afferent limb of the cardiopulmonary reflex arc11, suggesting that an autonomic nervous system (ANS) dysfunction may play a role in the development of these symptoms.

Despite these recent reports, few studies have compared HCM patients with and without history of unexplained syncope. Considering the possible involvement of the autonomic nervous system in the pathogenesis of this disorder, the aim of this study was to evaluate and compare the autonomic nervous system function, represented by heart rate variability, baroreflex sensitivity and response to orthostatic stress, in HCM patients with and without unexplained syncope.

Methods

Patients were prospectively selected between April 2007 and August 2009. The diagnosis of HCM was based on the finding of a LV hypertrophy equal to or greater than 15 mm in the absence of either cardiac or systemic disease that could explain this abnormality. Exclusion criteria were an age older than 60 years (n = 4), a cardiac rhythm other than a sinus rhythm (n = 4), syncope episodes associated with atrial arrhythmias (n = 3), previous miectomy (n = 1), an outflow tract gradient greater than 70 mmHg (n = 2), contraindications for interrupting antiarrhythmic drugs (n = 1), systolic dysfunction (n = 1) and blood pressure exceeding 160/90 mmHg at the first evaluation. The patients were divided in two groups based on their history of previous syncope episodes. Therefore, the final study group involved 37 patients, 16 with and 21 without previous syncope episodes (Tables 1 and 2). In the group presenting syncope, it was not possible to identify arrhythmic causes during the 24-hour Holter monitoring, and no significant outflow tract obstruction (defined as gradient greater than 50 mmHg) had been observed by echocardiogram at rest, except in one patient who presented an outflow tract gradient of 68 mmHg. Intraventricular gradient greater than 30 mmHg was found in 5 (33,3%) patients in syncope group and in 3 (14,3%) patients without syncope (p = 0,175).

Among the patients submitted to exercise test (93,8% of syncope group patients and 90,5% of without syncope group), we found that abnormal blood pressure response to exercise was similar between the groups (26,67 vs. 21,05%, p = 0,618).

Study protocol

The evaluation was performed after the hospital's Ethical Committee approval. Written informed consent was obtained from all patients. Patients arrived between 8 and 9 AM, after a fasting period of at least 6 hours. All cardioactive medications were discontinued for at least five half-lives before the study.

Initial evaluation

Patients were asked to answer a personal and family history questionnaire. Data from a rest electrocardiogram, echocardiogram, brain natriuretic peptide dosage and blood pressure (BP) response to exercise were then collected. Blood pressure response was considered abnormal if there was a drop or if the increase in the systolic BP was lower than 20 mmHg.

Blood pressure and electrocardiogram monitoring

The electrocardiogram and beat-to-beat BP measurements were obtained and analyzed by the Task Force System (CNSystems Medizintechnik GmbH, Graz, Austria). After monitoring and with a peripheral venous access in an antecubital vein, the patient remained resting in the supine position for at least 15 minutes.

Spontaneous baroreflex sensitivity (BRS)

After the rest period, 15 consecutive minutes of continuous systolic BP and heart rate were registered. The spontaneous BRS was analyzed by the sequence method, which is based on the identification of the spontaneous occurrence of three or more beats that are characterized by a progressive increase in the systolic BP associated with prolongation of the RR interval, or by a progressive decrease in the systolic BP associated with a reduction of the RR interval. This identification was automatic and the data were considered eligible when modifications of the systolic BP and the RR interval were equal to or greater than 1 mmHg e 5ms, respectively.

Phenylephrine induced baroreflex sensitivity

Patients were maintained on continuous BP and electrocardiogram monitoring. The vasoconstrictor agent phenylephrine was administered intravenously (2 to 4 mcg/kg), at least three bolus in 10-minute intervals, to produce an increase in the systolic BP of 15 to 40 mmHg. The consecutive values of the systolic BP and the corresponding RR intervals, considered one beat later, were plotted, and then a linear regression graphic was constructed. The inclination of this line represented the quantitative measurement of the BRS. We considered only those ramps in which the correlation coefficient (r2) was greater than 0.60. The average of these measurements was calculated for each patient.

Head-up tilting (HUT)

Patients were exposed to 70 degree tilting for 40 minutes on an electrically operated tilt table with a footboard support and two security belts, one at the level of the thorax and the other at the level of the knees. In case of a positive response to HUT, the table was immediately returned to a supine position. The European Society of Cardiology criteria were used to define the responses to HUT12.

Heart rate variability (HRV)

The HRV was acquired in time and frequency domains. The time domain variables of HRV considered were the standard deviation of normal RR intervals (SDNN), the root mean square of successive differences (RMSSD) and the percentage of adjacent normal RR intervals that differ by at least 50 ms (pNN50) during 24 hours of electrocardiogram recording. In the frequency domain, the high (0.15-0.4Hz), low (0.04-0.15Hz) and very low (0-0.04Hz) frequency bands and the total power density of the spectrum were considered, in absolute values and in normalized units, at rest (for 15 minutes) and in the first 5 minutes of 70 degree tilting.

Data analysis

Categorical variables are expressed as absolute (n) and relative frequencies (%). The association of these variables and the previous history of syncope were evaluated by the chi-squared test, adjusted by Fisher statistics. Quantitative variables are expressed as the mean value ± standard deviation and median; the comparison between the groups was performed using the Mann-Whitney test, once not all variables studied were normally distributed. For the variables studied at two moments in time, the Mann-Whitney test was used to verify the differences between the groups at the moments studied. Then, the Wilcoxon test was applied to verify the differences between the two times of observation. A p value < 0.05 was considered significant.

Results

Spontaneous baroreflex sensitivity

No significant difference was found in the spontaneous BRS when the patients with and without syncope were compared (16.46 ± 12.99 vs. 18.31 ± 9.88 ms/mmHg; p = 0.464). When only the values of spontaneous BRS characterized by progressive systolic BP increases and RR interval prolongations were considered, no statistically significant difference was found between the groups (18.48 ± 17.42 vs. 18.95 ± 11.20 ms/mmHg; p = 0.373). A similar response was noted when the values characterized by the decrease of the systolic BP and RR interval were considered (15.17 ± 9.32 vs. 17.45 ± 9.77 ms/mmHg; p = 0.524). Data from one patient (without syncope) were omitted, because that patient did not fulfill the method criteria for identification of the spontaneous sequences.

Phenylephrine induced baroreflex sensitivity

When comparing patients with syncope and without syncope there was no significant difference in the phenylephrine-induced BRS (18.33 ± 9.31 vs. 15.48 ± 15.48 ms/mmHg; p = 0.521). Data from three patients (two without syncope and one with syncope) were excluded because the r2 was lower than 0.60.

Head-up tilting (HUT)

HUT was positive in eight patients (40%), who presented hypotension and referred near syncope, one of them (6%) with previous history of syncope and seven without previous syncope (33.33%), p = 0.053. This finding attributed to HUT the following characteristics for detection of syncope etiology in this population: a sensitivity of 6%, a specificity of 66%, a positive predictive value of 12.5%, a negative predictive value of 48% and an accuracy of 40%. Among the positive responses, two were vasodepressive and six were mixed.

Heart rate variability

Analyses in time domain

The mean duration of the electrocardiographic monitoring was 23 h 55 min. During analysis, 2.8% of the data were lost due to interferences or artifacts. We did not find statistically significant differences in the mean heart rate between the groups (77.75 ± 10.33 vs. 80.48 ± 12.36 bpm; p = 0.701). For the analyses of HRV, data from two patients were excluded, both without syncope, because of a significant percentage of artifacts during the recording. When the values of SDNN were considered, no significant difference was found between the groups (137.69 ± 36.62 vs. 145.95 ± 38.07 ms; p = 0.389). The group with syncope had significantly lower values of RMSSD than the group without syncope (24.88 ± 10.03 vs. 35.58 ± 16.43 ms; p = 0,042) and we observed a tendency toward higher values of pNN50 in patients without syncope (4.51 ± 3.78 vs. 8.83 ± 7.98 %; p = 0.085).

Analyses in frequency domain

As shown in Table 3, we did not find any statistically significant differences between the groups when considering the frequency domain variables of the HRV studied. Note that the syncope group presented a tendency to lower values of the high frequency component in the supine position (p = 0.075). When the influences of postural changes on the referred parameters were studied, orthostatic posture caused a similar significant increase in the LF/HF ratio in both groups (p = 0.010 for the group with syncope; p = 0.021 for the group without syncope).

Discussion

To identify the origin of syncope in HCM patients is still a challenge. It may be caused by several mechanisms, like arrhythmias or ventricular outflow tract obstruction, but in the majority of cases it may result of a combination of factors influencing the circulatory balance. It is well known that the ANS plays a major role in cardiovascular homeostasis, once the vagal and sympathetic modulation and balance represents the main mechanism of an adequate brain perfusion maintenance.

The prognostic value of an adequate ANS function is already well established in some cardiac diseases13,14. Many years ago, Gilligan et al15 observed that patients with HCM had lower parasympathetic activity (measured by HR variation during deep breathing and by Valsalva maneuver) than normal individuals, compatible with an afferent parasympathetic limb alteration. Other studies have demonstrated that HCM patients may present an altered hemodynamic behavior during exercise characterized by a drop in systemic vascular resistance and hypotension8,9, and suggested that this could be a consequence of the activation of ventricular mechanoreceptors leading to a sympathetic withdrawal to resistance vessels. Prasad et al10 have demonstrated that this phenomenon of hypotension occurs also during routine activities and even at rest, when they studied a group of patients presenting frequent syncope or near syncope and noted that patients with these symptoms and hypotensive episodes during BP monitoring had a lower spontaneous BRS.

Spontaneous BRS measurements

In this study, we did not find a significant difference in the spontaneous BRS between HCM patients with and without syncope. These conflicting results may be explained by the inclusion criteria of the two studies. Patients with syncope of our study had less episodes of syncope and we did not include patients with near syncope. On the other hand, patients from Prasad's study had more episodes and had at least one in the previous month. Another possible explanation for the difference in our findings is that, in our study, the measurements of spontaneous BRS were obtained during rest and in a supine position. In contrast, the other study's data were acquired on different occasions during 24 hours of monitoring. For this reason, one may suggest that the BRS might have more variations in its adaptation capacity over the course of the day.

Phenylephrine induced baroreflex sensitivity

The reflex response of HR to phenylephrine - a vasoselective pressor agent - has been extensively used to evaluate BRS since its initial description in 196916, although it has been poorly studied in HCM patients. In our study, no differences were observed in the phenylephrine induced BRS between patients with and without syncope. In the unique report published previously that used this method, Thomson et al11 compared HCM patients to normal people and observed similar results between the groups.

Head-up tilting (HUT)

We found no statistical difference in positive responses to HUT in HCM patients with and without unexplained syncope, but with a trend toward more positive tests in the non-syncope group. Other authors have published comparable findings, suggesting that HUT is not a good tool to evaluate the etiology of syncope in HCM patients10,17,18. Only Gilligan et al19 found a greater frequency of positive responses in patients with a previous history of syncope. It is important to emphasize that these authors used isoproterenol infusion in cases with an initial negative response to passive tilting and it is supposed that HCM patients are more susceptible to mechanoreceptor activation and Bezold-Jarisch reflex deflagration when submitted to excessive adrenergic stimulation.

HUT had poor sensitivity, specificity and accuracy in identifying the etiology of syncope in this sample. We must remember that patients with unexplained syncope may have many reasons for a hemodynamic collapse, including arrhythmias. Finally, it has already been demonstrated that neurally mediated syncope corresponds to a small percentage of cases among individuals suffering from structural heart disease. This differs completely from what is found among patients with normal hearts, in whom neurally mediated syncope represents the most common cause of hemodynamic collapses.

Heart rate variability

HRV has been extensively studied in cardiovascular diseases, and its prognostic value is well established in most of these disorders, especially in ischemic heart disease13,20. In this study, we observed that SDNN, an estimative of total HRV, was similar between the groups. However, RMSSD values were significantly lower in patients with a history of syncope (p = 0.042); pNN50 values tended to be lower in the same group (p = 0.085). When considering the parameters analyzed in the frequency domain, we observed a tendency to lower values of the high frequency component during a supine position in the syncope group. RMSSD, pNN50 and the high frequency component of spectral analyses are measures that represent especially the parasympathetic activity.

A decrease in vagal activity has been previously related in patients with HCM when compared to normal people21-23. This decrease in parasympathetic function was correlated to more severe clinical presentations. Döven et al22 observed lower values of SDNN, RMSSD and pNN50 in patients in the functional class III e IV and in those with a greater degree of outflow tract obstruction. A lower vagal activity was noted in HCM patients, especially in those with a history of syncope, thoracic pain during effort, dyspnea and mitral regurgitation21.

Counhihan et al24 observed that specific parameters of vagal activity, including pNN50 and the high frequency component of spectral analyses, were lower in patients presenting atrial arrhythmias on ambulatory ECG recordings. The same authors also found lower measures of HRV, global and specific to vagal activity (pNN50 and high frequency component), in patients with non-sustained ventricular tachycardia on 24h Holter monitoring. The decrease in the high frequency component of the spectral analyses was the only component of the HRV related to greater chances of cardiac events (death or hospitalization because of heart failure), after 28 months of follow-up, in the study published by Kawasaki et al25.

On the other hand, Bittencourt et al26 compared HCM patients with normal volunteers and did not find significant differences in HRV between the groups, when considering short electrocardiographic recordings.

We did not find, after a literature review, any study that has specifically compared the HRV in HCM patients with and without previous syncope episodes. Nevertheless, Bonaduce et al21 have already identified a lower vagal activity in symptomatic HCM patients, including the ones presenting syncope.

It is still speculative how a decrease in the parasympathetic activity could be associated with syncope in these patients. It is possible that it acts modulating the onset of atrial and ventricular arrhythmias. Maybe it could be considered a marker of a more severe presentation of the disease, as it has been demonstrated in other cardiomyopathies.

Study limitations

It is important to consider the small number of patients enrolled in the study. Despite being in a tertiary center, with a specialized HCM outpatients unit, it is a rare disease and unexplained syncope was not a very frequent symptom presented by our population of patients. Additionally, we cannot rule out an arrhythmic cause of syncope in this population. Continuous, prospective prolonged electrocardiographic monitoring is strongly recommended to confirm or exclude arrhythmic events during the follow-up. Concerning this issue, we must note that previous episodes of atrial fibrillation were more frequent in patients presenting syncope in our sample.

Conclusions

Our study suggests that HCM patients presenting unexplained syncope have lower vagal activity than those without syncope.

We also concluded that HUT is not a good tool in diagnosing syncope in these patients. This finding supports the idea that this method should not be recommended for routine evaluation of unexplained syncope in this specific population, mainly because of the low specificity of the test.

Conception and design of the research: Macatrão-Costa MF, Arteaga-Fernandez E, Hachul D; Acquisition of data: Macatrão-Costa MF, Arteaga-Fernandez E, Brito FS, Hachul D; Analysis and interpretation of the data: Macatrão-Costa MF, Brito FS, Hachul D; Statistical analysis and Writing of the manuscript: Macatrão-Costa MF; Critical revision of the manuscript for intellectual content: Arteaga-Fernandez E, Darrieux F, Melo SL, Scanavacca M, Sosa E, Hachul D.

Potential Conflict of Interest

No potential conflict of interest relevant to this article was reported.

Sources of Funding

There were no external funding sources for this study.

Study Association

This article is part of the thesis of Doctoral submitted by Milena Frota Macatrão-Costa, from Faculdade de Medicina da Universidade de São Paulo.

References

  • 1. Gersh BJ, Maron BJ, Bonow RO, Dearani JA, Fifer MA, Link MS, et al; American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines; American Association for Thoracic Surgery; American Society of Echocardiography; American Society of Nuclear Cardiology; Heart Failure Society of America; Heart Rhythm Society; Society for Cardiovascular Angiography and Interventions; Society of Thoracic Surgeons. 2011 ACCF/AHA guideline for the diagnosis and treatment of hypertrophic cardiomyopathy: executive summary: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation. 2011;124(24):2761-96.
  • 2. Desai DM, Bhat GS, Daxini BV, Sharma S. Complete heart block as a cause of syncope in hypertrophic cardiomyopathy. J Assoc Physicians India. 1991;39(12):965-6.
  • 3. Wang DW, Deng YB. Hypertrophic cardiomyopathy complicated by severe bradycardias: a pedigree report. Clin Cardiol. 2002;25(2):76-80.
  • 4. Brembilla-Perrot B, Terrier de La Chaise A, Beurrier D. Paroxysmal atrial fibrillation: main cause of syncope in hypertrophic cardiomyopathy. Arch Mal Coeur Vaiss. 1993;86(11):1573-8.
  • 5. Dilsizian V, Bonow RO, Epstein SE, Fananapazir L. Myocardial ischemia detected by thalium scintigraphy is frequently related to cardiac arrest and syncope in young patients with hypertrophic cardiomyopathy. J Am Coll Cardiol. 1993;22(3):796-804.
  • 6. Etoh T, Matsuda J, Hanada Y, Takenaga M, Ishikawa T, Koiwaya Y, et al. Recurrent syncope induced by left ventricular outflow tract obstruction: demonstration in a patient with hypertrophic obstructive cardiomyopathy. J Cardiol. 1997;30(3):143-7.
  • 7. Buja G, Melacini P, Folino A, Miorelli M, Nava A. Spontaneous and induced vasodepressor/vasovagal syncope in hypertrophic cardiomyopathy. Clin Cardiol. 1992;15(5):387-9.
  • 8. Frenneaux MP, Counihan PJ, Caforio AL, Chikamori T, McKenna WJ. Abnormal blood pressure response during exercise in hypertrophic cardiomyopathy. Circulation. 1990;82(6):1995-2002.
  • 9. Counihan PJ, Frenneaux MP, Webb DJ, McKenna WJ. Abnormal vascular responses to supine exercise in hypertrophic cardiomyopathy. Circulation. 1991;84(2):686-96.
  • 10. Prasad K, Williams L, Campbell R, Elliot PM, McKenna WJ, Frenneaux M. Episodic syncope in hypertrophic cardiomyopathy: evidence for inappropriate vasodilation. Heart. 2008;94(10):1312-7.
  • 11. Thomson HL, Morris-Thurgood J, Atherton J, Frenneaux M. Reduced cardiopulmonary baroreflex sensitivity in patients with hypertrophic cardiomyopathy. J Am Coll Cardiol. 1998;31(6):1377-82.
  • 12. Brignole M, Menozzi C, Del Rosso A, Costa S, Gaggioli G, Bottoni N, et al. New classification of haemodynamics of vasovagal syncope: beyond the VASIS classification: analysis of the pre-syncopal phase of the tilt test without and with nitroglycerin challenge. Vasovagal Syncope International Study. Europace. 2000;2(1):66-76.
  • 13. La Rovere MT, Bigger JT Jr, Marcus FI, Mortara A, Schwartz PJ; Baroreflex sensitivity and heart-rate variability in prediction of total cardiac mortality after myocardial infarction. Atrami (Autonomic Tone and Reflexes After Myocardial Infarction) Investigators. Lancet. 1998;351(9101):478-84.
  • 14. Mortara A, La Rovere MT, Pinna GD, Prpa A, Maestri R, Febo O, et al. Arterial baroreflex modulation of heart rate in chronic heart failure: clinical and hemodynamic correlates and prognostic implications. Circulation. 1997;96(10):3450-8.
  • 15. Gilligan DM, Chan WL, Sbarouni E, Nihoyannopoulos P, Oakley CM. Autonomic function in hypertrophic cardiomyopathy. Br Heart J. 1993;69(6):525-9.
  • 16. Smyth HS, Sleight P, Pickering GW. Reflex regulation of arterial pressure during sleep in man: a quantitative method of assessing baroreflex sensitivity. Circ Res. 1969;24(1):109-21.
  • 17. Manganelli F, Betocchi S, Ciampi Q, Storto G, Losi MA, Violante A, et al. Comparison of hemodynamic adaptation to orthostatic stress in patients with hypertrophic cardiomyopathy with and without syncope and in vasovagal syncope. Am J Cardiol. 2002;89(12):1405-10.
  • 18. Sneddon JF, Slade A, Seo H, Camm J, McKenna WJ. Assessment of the diagnostic value of head-up tilt testing in the evaluation of syncope in hypertrophic cardiomyopathy. Am J Cardiol. 1994;73(8):601-4.
  • 19. Gilligan DM, Nihoyannopoulos P, Chan WL, Oakley CM. Investigation of a hemodynamic basis for syncope in hypertrophic cardiomyopathy: use of a head up tilt test. Circulation. 1992;85(6):2140-8.
  • 20. Kleiger RE, Miller JP, Bigger JT Jr, Moss AJ. Decreased heart rate variability and its association with increased mortality after acute myocardial infarction. Am J Cardiol. 1987;59(4):256-62.
  • 21. Bonaduce D, Petretta M, Betocchi S, Ianniciello A, Marciano F, Apicella C, et al. Heart rate variability in patients with hypertrophic cardiomyopathy: association with clinical and echocardiographic features. Am Heart J. 1997;134(2 Pt 1):165-72.
  • 22. Döven O, Sayin T, Güldal M, Karaoguz R, Oral D. Heart rate variability in hypertrophic obstructive cardiomyopathy: association with functional classification and left ventricular outflow gradients. Int J Cardiol. 2001;77(2-3):281-6.
  • 23. Mörner S, Wiklund U, Rask P, Olofsson BO, Kazzam E, Waldeström A. Parasympathetic dysfunction in hypertrophic cardiomyopathy assessed by heart rate variability: comparison between short-term and 24-h measurements. Clin Physiol Funct Imaging. 2005;25(2):90-9.
  • 24. Counihan PJ, Fei L, Bashir Y, Farrel TG, Haywood GA, McKenna WJ. Assessment of heart rate variability in hypertrophic cardiomyopathy: association with clinical and prognostic features. Circulation. 1993;88(4 Pt 1):1682-90.
  • 25. Kawasaki T, Azuma A, Sakatani T, Hadase M, Kamitani T, Kawasaki S, et al. Prognostic value of heart variability in patients with hypertrophic cardiomyopathy. J Electrocardiol. 2003;36(4):333-8.
  • 26. Bittencourt MI, Benchimol Barbosa PR, Drumond Neto C, Bedirian R, Barbosa EC, Brasil F, et al. Avaliação da função autonômica na cardiomiopatia hipertrófica. Arq Bras Cardiol. 2005;85(6):388-96.
  • Evaluation of the autonomic function in patients with hypertrophic cardiomyopathy with and without syncope

    Milena Frota Macatrão-CostaI; Edmundo Arteaga-FernandezI; Fábio Sandoli de BritoII; Francisco DarrieuxI; Sissy Lara de MeloI; Mauricio ScanavaccaI; Eduardo SosaI; Denise HachulI
  • Publication Dates

    • Publication in this collection
      11 Mar 2013
    • Date of issue
      Feb 2013

    History

    • Received
      03 May 2012
    • Accepted
      04 Sept 2012
    • Reviewed
      20 Aug 2012
    Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
    E-mail: revista@cardiol.br