Subcutaneous Tissue Thickness is an Independent Predictor of Image Noise in Cardiac CT

Abstracts

Background:

Few data on the definition of simple robust parameters to predict image noise in cardiac computed tomography (CT) exist.

Objectives:

To evaluate the value of a simple measure of subcutaneous tissue as a predictor of image noise in cardiac CT.

Methods:

86 patients underwent prospective ECG-gated coronary computed tomographic angiography (CTA) and coronary calcium scoring (CAC) with 120 kV and 150 mA. The image quality was objectively measured by the image noise in the aorta in the cardiac CTA, and low noise was defined as noise < 30HU. The chest anteroposterior diameter and lateral width, the image noise in the aorta and the skin-sternum (SS) thickness were measured as predictors of cardiac CTA noise. The association of the predictors and image noise was performed by using Pearson correlation.

Results:

The mean radiation dose was 3.5 ± 1.5 mSv. The mean image noise in CT was 36.3 ± 8.5 HU, and the mean image noise in non-contrast scan was 17.7 ± 4.4 HU. All predictors were independently associated with cardiac CTA noise. The best predictors were SS thickness, with a correlation of 0.70 (p < 0.001), and noise in the non-contrast images, with a correlation of 0.73 (p < 0.001). When evaluating the ability to predict low image noise, the areas under the ROC curve for the non-contrast noise and for the SS thickness were 0.837 and 0.864, respectively.

Conclusion:

Both SS thickness and CAC noise are simple accurate predictors of cardiac CTA image noise. Those parameters can be incorporated in standard CT protocols to adequately adjust radiation exposure.

Sternum / radiation effects; Radiation injuries; Computed tomography; Artifacts


Fundamento:

Há poucos dados sobre a definição de parâmetros simples e robustos para predizer artefato de imagem em tomografia computadorizada (TC) cardíaca.

Objetivos:

Avaliar o valor da simples medida da espessura do tecido subcutâneo (espessura pele-esterno) como preditor de artefato de imagem em TC cardíaca.

Métodos:

O estudo avaliou 86 pacientes submetidos a angiotomografia computadorizada cardíaca (ATCC) com sincronização prospectiva com ECG e avaliação de escore de cálcio coronário com 120 kV e 150 mA. A qualidade da imagem foi medida objetivamente pelo artefato de imagem na aorta em ATCC, sendo 'artefato baixo' definido como aquele < 30 UH. Os diâmetros torácicos anteroposterior e laterolateral, o artefato de imagem na aorta e a espessura pele-esterno foram medidos como preditores de artefato em ATCC. A associação de preditores e artefato de imagem foi avaliada usando-se correlação de Pearson.

Resultados:

A dose média de radiação foi 3,5 ± 1,5 mSv. O artefato de imagem médio na ATCC foi de 36,3 ± 8,5 UH, sendo o artefato de imagem médiona fase sem contraste do exame de 17,7 ± 4,4 UH. Todos os preditores foram independentemente associados com artefato em ATCC. Os melhores preditores foram espessura pele-esterno, com correlação de 0,70 (p < 0,001), e artefato de imagem na fases em contraste,com correlação de 0,73 (p < 0,001). Ao avaliar a habilidade de predizer artefato de imagem baixo, as áreas sob a curva ROC para o artefato de imagem na fases em contraste e para a espessura pele-esterno foram 0,837e 0,864, respectivamente.

Conclusão:

Tanto espessura pele-esterno quanto artefato de escore de cálcio são preditores simples e precisos de artefato de imagem em ATCC. Tais parâmetros podem ser incorporados aos protocolos de TC padrão para ajustar adequadamente a exposição à radiação.

Esterno / efeitos de radiação; Lesões por radiações; Tomografia computadorizada; Artefatos


Introduction

Cardiac computed tomographic angiography (CTA) is a useful and accurate tool to evaluate the coronary arteries and cardiac structure11. Taylor AJ, Cerqueira M, Hodgson JM, Mark D, Min J, O'Gara P, et al. ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 Appropriate Use Criteria for Cardiac Computed Tomography A Report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the Society of Cardiovascular Computed Tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance. J Am Coll Cardiol. 2010;56(22):1864-94.. Although cardiac CTA is safe, the use of radiation poses a small risk22. Brenner DJ, Hall EJ. Computed tomography-an increasing source of radiation exposure. N Engl J Med. 2007;357(22):2277-84., which has raised concerns, because of the increasing number of medical tests using radiation33. Gerber TC, Carr JJ, Arai AE, Dixon RL, Ferrari VA, Gomes AS, et al. Ionizing radiation in cardiac imaging: a science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention. Circulation. 2009;119(7):1056-65.. In particular, younger patients and women are at a particularly higher risk of long-term complications following radiation exposure44. Einstein AJ, Henzlova MJ, Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA. 2007;298(3):317-23..

The reduction in radiation exposure has been the aim of many recent advances in cardiac CTA, including the use of tube current modulation55. Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, et al. Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009;301(5):500-7., prospectively ECG-triggered axial scan66. Hausleiter J, Meyer TS, Martuscelli E, Spagnolo P, Yamamoto H, Carrascosa P, et al. Image quality and radiation exposure with prospectively ECG-triggered axial scanning for coronary CT angiography: the multicenter, multivendor, randomized PROTECTION-III study. JACC Cardiovasc Imaging. 2012;5(5):484-93. and newer acquisition modes77. Qin J, Liu LY, Fang Y, Zhu JM, Wu Z, Zhu KS, et al. 320-detector CT coronary angiography with prospective and retrospective electrocardiogram gating in a single heartbeat: comparison of image quality and radiation dose. Br J Radiol. 2012;85(1015):945-51., as well as the adequate adjustment of the kV and mA used during image acquisition55. Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, et al. Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009;301(5):500-7.,88. Bischoff B, Hein F, Meyer T, Hadamitzky M, Martinoff S, Schomig A, et al. Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study. JACC Cardiovasc Imaging. 2009;2(8):940-6.,99. Imai K, Ikeda M, Kawaura C, Aoyama T, Enchi Y, Yamauchi M. Dose reduction and image quality in CT angiography for cerebral aneurysm with various tube potentials and current settings. Br J Radiol. 2012;85(1017):e673-81.. Additionally, studies have shown that the combination of several techniques1010. Starck G, Lönn L, Cederblad Å, Forssell-Aronsson E, Sjöström L, Alpsten M. A method to obtain the same levels of CT image noise for patients of various sizes, to minimize radiation dose. Br J Radiol. 2002;75(890):140-50. and exposure estimation based on complex calculations are also helpful in optimizing radiation exposure1111. Smedby Ö, Fredrikson M, De Geer J, Borgen L, Sandborg M. Quantifying the potential for dose reduction with visual grading regression. Br J Radiol. 2013;86(1021):20110784.. More recently, the use of iterative reconstruction techniques have also been proposed as a potential dose sparing technique1212. Bittencourt M, Schmidt B, Seltmann M, Muschiol G, Ropers D, Daniel W, et al. Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience. Int J Cardiovasc Imaging. 2011;27(7):1081-7.

13. Desai GS, Thabet A, Elias AY, Sahani DV. Comparative assessment of three image reconstruction techniques for image quality and radiation dose in patients undergoing abdominopelvic multidetector CT examinations. Br J Radiol. 2013;86(1021):20120161.

14. Tatsugami F, Matsuki M, Nakai G, Inada Y, Kanazawa S, Takeda Y, et al. The effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography. Br J Radiol. 2012;85(1016):e378-82.
-1515. Qi LP, Li Y, Tang L, Li YL, Li XT, Cui Y, et al. Evaluation of dose reduction and image quality in chest CT using adaptive statistical iterative reconstruction with the same group of patients. Br J Radiol. 2012;85(1018):e906-11..

Although effective, the inadequate or excessive use of those techniques may affect image quality and result in limited or inadequate scans. The main adverse effect of lowering the radiation dose is the image noise increase caused by a reduction in the number of photons that reach the detectors33. Gerber TC, Carr JJ, Arai AE, Dixon RL, Ferrari VA, Gomes AS, et al. Ionizing radiation in cardiac imaging: a science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention. Circulation. 2009;119(7):1056-65..

The main cause of increased noise in cardiac CTA is the interposed extracardiac structures. Therefore, many protocols rely on measures associated with the patient's body constitution, e. g. body mass index (BMI)1616. Yoshimura N, Sabir A, Kubo T, Lin PJ, Clouse ME, Hatabu H. Correlation between image noise and body weight in coronary CTA with 16-row MDCT. Acad Radiol. 2006;13(3):324-8., patient's chest circumference1717. Ghoshhajra BB, Engel LC, Major GP, Verdini D, Sidhu M, Karolyi M, et al. Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr. 2011;5(4):240-6. and diameter1818. Voros S, Rivera JJ, Berman DS, Blankstein R, Budoff MJ, Cury RC, et al; Society for Atherosclerosis Imaging and Prevention Tomographic Imaging and Prevention Councils; Society of Cardiovascular Computed Tomography. Guideline for minimizing radiation exposure during acquisition of coronary artery calcium scans with the use of multidetector computed tomography: a report by the Society for Atherosclerosis Imaging and Prevention Tomographic Imaging and Prevention Councils in collaboration with the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr. 2011;5(2):75-83. to adjust the radiation parameters. Guidelines currently recommend the use of chest anteroposterior (AP) diameter or chest lateral width to estimate the radiation dose1818. Voros S, Rivera JJ, Berman DS, Blankstein R, Budoff MJ, Cury RC, et al; Society for Atherosclerosis Imaging and Prevention Tomographic Imaging and Prevention Councils; Society of Cardiovascular Computed Tomography. Guideline for minimizing radiation exposure during acquisition of coronary artery calcium scans with the use of multidetector computed tomography: a report by the Society for Atherosclerosis Imaging and Prevention Tomographic Imaging and Prevention Councils in collaboration with the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr. 2011;5(2):75-83., although this strategy has not yet been validated or compared to other techniques.

Thus, in this study we aimed at objectively evaluating the association of BMI, chest AP diameter, and chest lateral width with the amount of noise in cardiac CTA images. Additionally, we hypothesized that the subcutaneous tissue measurement at the level of the sternum and the noise in non-contrast images would better correlate with image noise.

Methods

We enrolled 86 consecutive patients who underwent prospective ECG-gated cardiac CTA and calcium scoring with the same tube potential and tube current following the same injection protocol. All coronary cardiac CTAs were performed to evaluate known or suspected coronary artery disease. The study was approved by the local institutional ethics committee, and all participants signed the written informed consent. Data on patient characteristics and clinical information were collected prospectively as part of an institutional database.

All patients with a heart rate above 60 bpm received oral beta-blockers prior to image acquisition. Cardiac CTA scans were performed in a 64 row multi-slice scanner (Brilliance 64, Philips Healthcare, Best, Netherlands). The scout images were acquired with 120 kV and 30 mA. After the scout images, all patients also underwent a prospectively ECG-gated calcium scoring with a tube potential of 120 kV and a tube current of 55 mA. The contrast-enhanced cardiac CTA was performed with a collimation of 64 x 0.625 mm, gantry rotation time of 400 milliseconds with tube current of 150 mA and tube potential of 120 kV. The slice thickness was selected as 0.8 mm, increment of 0.4 mm, using 100 ml of iodine contrast (Ultravist 370, Bayer, Germany) injected with a dual head injector (Medrad Inc., U.S.A.) at a rate of 6 ml/seconds followed by 60 ml of saline at the same rate, using a 18 gauge in the antecubital vein. Automated bolus tracking was used by placing a circular region of interest in the descending aorta and acquisition was triggered when the average attenuation value in the region of interest reached 150 Hounsfield Unit (HU). The 75% R-R interval image was used for image reconstruction and coronary analysis. Images were reconstructed using standard filtered back projection and a standard kernel.

Factors related to patient characteristics and injection protocols that affect the aorta contrast were considered. We enrolled patients with similar heart rate, varying from 50-60 bpm, with no history of heart failure and we used the same injection protocol for the whole sample, with a total of 100 ml iodine contrast at 6 ml/s, followed by 60 ml of saline solution at 6 ml/s, using an 18 gauge in the antecubital vein. The image noise in computed tomography (CT) was defined as the standard deviation measured with a region of interest of 1 cm2 in the ascending aorta (Figure 1A). The chest AP diameter was measured on the chest digital axial cross-section image with a line passing through the middle of the heart (Figure 1B), the skin-sternum (SS) thickness was measured from the skin to the sternum in the middle intermammary level (Figure 1C), and the chest lateral width was measured on the topogram from skin to skin at the level of the left hemidiaphragm (Figure 1D). The image noise calculation in the aorta was also measured as standard deviation units, with a region of interest of 1 cm2 in the prospective calcium score using a fixed tube potential of 120 kV and a tube current of 55 mA (Figure 1E). The BMI was calculated as the weight in kilograms divided by the squared height in meters.

Figure 1
A) Aorta noise measurement with an 1-cm2 region of interest in the ascending aorta. B) Chest anteroposterior diameter measurement. C) Skin-sternum thickness measurement. D) Chest lateral width measurement. E) Calcium score noise measurement.

Statistical Analysis

All continuous variables have been inspected for normality and presented as mean ± standard deviations as no significant departures from normality were detected. Categorical variables were presented in absolute and relative (%) frequencies. The relationship between image noise in the cardiac CT and BMI, chest AP diameter, SS thickness, chest lateral width and aorta noise in calcium score imaging was evaluated using Pearson correlation and linear regression. For comparing Pearson correlations, the Fisher's z approximation was used. Multivariate linear regression models were built to identify the best combination of predictors to estimate cardiac CTA noise. To define the best prediction model, the lowest adjusted R2 value was used. Additionally, the image noise was dichotomized as 'low noise', if SD < 30 HU, and 'high noise', if SD > 30, as in previous reports1919. Ghafourian K, Younes D, Simprini LA, Weigold WG, Weissman G, Taylor AJ. Scout view X-ray attenuation versus weight-based selection of reduced peak tube voltage in cardiac CT angiography. JACC Cardiovasc Imaging. 2012;5(6):589-95.,2020. Gao J, Li J, Earls J, Li T, Wang Q, Dai R. Individualized tube current selection for 64-row cardiac CTA based on analysis of the scout view. Eur J Radiol. 2011;79(2):266-71.. Receiver-operating characteristic (ROC) curve analysis to predict the 'low noise' images was performed for each of the predictors. A p-value of less than 0.05 was considered significant. The Bonferroni adjustment was applied to adjust for multiple comparisons. The Stata program, version 12, (StataCorp) was used for all statistical analyses.

Results

Table 1 shows the characteristics of the patient population. The mean total radiation dose, including CAC and cardiac CTA was 3.5 ± 1.5 mSv; the mean image noise in the CT was 36.3 ± 8.5 HU, and the mean image noise in the non-contrast scan was 17.7 ± 4.4 HU.

Table 1
Baseline characteristics of the population

All individual parameters were significantly associated with the noise in the cardiac CTA. Pearson correlation values are shown in Table 2, and the correlation matrix is shown in Figure 2. In a pairwise comparison, the CAC noise, BMI and SS thickness have a more significant correlation with cardiac CTA noise than any of the chest diameters (p < 0.001 for all comparisons). The correlation of CAC noise with outcome showed a trend towards being more significant than BMI (p = 0.05), whereas the CAC noise and SS thickness (p = 0.16), as well as the SS thickness and BMI correlations (p = 0.13) were not significantly different. In the multivariate analysis, the best prediction model included the association of CAC noise and SS thickness, with an adjusted R2 of 0.61 (p < 0.0001).

Table 2
Pearson correlation of the predictors vs image noise in cardiac computed tomographic angiography
Figure 2
Scatter plots of each of the predictors vs. cardiac CTA noise, including the regression lines. A) AP diameter; B) Lateral width; C) BMI; D) CAC noise; and E) Skin-sternum thickness.

Similarly, the best independent predictor of 'low noise' was SS thickness with an area under the ROC curve (AUC) of 0.86 (95% confidence interval (CI): 0.79 - 0.94) (Figure 3), followed by CAC noise (AUC of 0.84; 95% CI: 0.75 - 0.93), BMI (AUC of 0.80; 95% CI: 0.69 - 0.91), chest AP diameter (AUC of 0.62, 95% CI: 0.49 - 0.75), and chest lateral width (AUC of 0.59; 95% CI: 0.47 - 0.73) (Figure 2).

Figure 3
Receiver operating characteristic curves for each of the predictors as an estimator of ‘low image noise’ in cardiac CTA. The area under the ROC curve value is presented in the legend for each parameter. APD: anteroposterior diameter; LW: lateral width; SS: skin-sternum thickness; BMI: body mass index; CAC: calcium score.

Discussion

The present study assessed two new parameters (CAC noise estimation and SS thickness) to better predict the image noise in contrast-enhanced cardiac CTA versus the currently recommended parameters of chest diameter, although the results of the new measures as compared to BMI did not reach statistical significance. Our study demonstrated that both parameters are better predictors of image noise than the currently recommended chest diameter techniques. Additionally, both parameters improved the image noise prediction when associated with the currently best predictor (BMI). Finally, they also had the best performance to discriminate cardiac CTA scans with a 'low noise', defined as a SD <30 HU.

Although the best fitting model included both parameters, the increase in R2 by combining both parameters was small, and we believe a more parsimonious model including either one of the two variables or BMI in a univariate model would be better suited to routine radiation exposure adjustments.

The identification of a single parameter as a better predictor of image noise has been the focus of recent publications1717. Ghoshhajra BB, Engel LC, Major GP, Verdini D, Sidhu M, Karolyi M, et al. Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr. 2011;5(4):240-6.,2020. Gao J, Li J, Earls J, Li T, Wang Q, Dai R. Individualized tube current selection for 64-row cardiac CTA based on analysis of the scout view. Eur J Radiol. 2011;79(2):266-71.,2121. Schuhbaeck A, Schaefer M, Marwan M, Gauss S, Muschiol G, Lell M, et al. Patient-specific predictors of image noise in coronary CT angiography. J Cardiovasc Comput Tomogr. 2013;7(1):39-45.. Those studies have focused on measuring the subcutaneous tissue as a predictor of image noise. Gao et al2020. Gao J, Li J, Earls J, Li T, Wang Q, Dai R. Individualized tube current selection for 64-row cardiac CTA based on analysis of the scout view. Eur J Radiol. 2011;79(2):266-71. have used a measurement of the X-ray attenuation of the thorax in the scout images of patients undergoing cardiac CTA. Their data suggest that this measurement is a better predictor of cardiac CTA image noise than BMI or weight. In the study by Schuhbaeck et al2121. Schuhbaeck A, Schaefer M, Marwan M, Gauss S, Muschiol G, Lell M, et al. Patient-specific predictors of image noise in coronary CT angiography. J Cardiovasc Comput Tomogr. 2013;7(1):39-45. a more advanced measurement of the entire chest wall at the level of the aortic root was used, and the authors concluded that such measure is significantly better than other predictors, including BMI2121. Schuhbaeck A, Schaefer M, Marwan M, Gauss S, Muschiol G, Lell M, et al. Patient-specific predictors of image noise in coronary CT angiography. J Cardiovasc Comput Tomogr. 2013;7(1):39-45.. Finally, Ghoshhajra et al1717. Ghoshhajra BB, Engel LC, Major GP, Verdini D, Sidhu M, Karolyi M, et al. Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr. 2011;5(4):240-6. have measured the chest area for the same purposes1717. Ghoshhajra BB, Engel LC, Major GP, Verdini D, Sidhu M, Karolyi M, et al. Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr. 2011;5(4):240-6.. The main limitation of those studies was the inclusion of different acquisition protocols with different radiation exposures. Since those parameters were defined based on patients' characteristics, both studies might have overestimated the association of the measurements with the final cardiac CTA noise.

Our study restricted the analysis to similar patients submitted to the same acquisition protocol (with the same kV, mA, contrast rate and acquisition mode). In our sample, patients with heart failure were excluded and the image acquisition was performed with similar heart rates (range, 50-60 bpm), adjusting the factors that could affect aorta contrast enhancement. The main reason for this choice is that kV, mA and acquisition mode are usually chosen based on the patient's weight, BMI and heart rate. With this restriction we avoid the spurious association between predictors and outcome, due to the indication bias of selecting the most appropriate parameters prior to the image acquisition based on patients' characteristics.

Previous studies have demonstrated that image-based parameters are more precise than other patients' characteristics, such as BMI. Our study also showed that SS thicknesses and the direct measurement of noise in non-contrast images corroborate those prior data. These simple measurements take less than one minute and can be performed as part of a usual cardiac CTA acquisition flow without additional radiation or cost.

As demonstrated by Ghoshhajra et al1717. Ghoshhajra BB, Engel LC, Major GP, Verdini D, Sidhu M, Karolyi M, et al. Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr. 2011;5(4):240-6., those measurements lead to a better use of radiation. Not only is radiation exposure reduced in most cases, but precise measurements can lead to the use of higher doses in appropriate cases, which should be the goal of cardiac imaging2222. Huda W, Vance A. Patient radiation doses from adult and pediatric CT. AJR Am J Roentgenol. 2007;188(2):540-6..

Among the various radiation-reducing techniques, the recently introduced use of automated attenuation-based tube potential selection was able to reduce radiation exposure by about 25%2323. Winklehner A, Goetti R, Baumueller S, Karlo C, Schmidt B, Raupach R, et al. Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness. Invest Radiol. 2011;46(12):767-73.. This automated method is used in image attenuation as part of an automated algorithm to define radiation exposure. The SS thickness can be seen as a simplified version of the concept of chest attenuation as a predictor of image noise. Nevertheless, additional data is needed to define the appropriate adjustments in mA and kV of the current measurements.

Our study must, however, be read in the context of its design. First, to allow precise correlation of patients' measurements and noise estimation, only a fixed protocol was used. Therefore, the definition of the adequate mA and kV adjustments of these measures cannot be evaluated in the current data. Second, although simple, these measurements add an extra step to image acquisition. Third, this can only be used in patients who undergo a non-contrast scan prior to the cardiac CT. In cases where the non-contrast scan is not performed, the non-contrast noise cannot be calculated, but the SS thickness can still be measured in the contrast timing images.

Conclusion

Both SS thickness and CAC noise are simple accurate predictors of cardiac CTA image noise. Those parameters can be incorporated in standard CT protocols to adequately adjust radiation exposure.

Author contributions

Conception and design of the research: Staniak HL, Sharovsky R, Bensenor I, Lotufo PA, Bittencourt MS; Acquisition of data: Staniak HL, Sharovsky R, Bittencourt MS; Analysis and interpretation of the data: Staniak HL, Sharovsky R, Pereira AC, Bittencourt MS; Statistical analysis: Staniak HL, Sharovsky R, Pereira AC, Bensenor I, Lotufo PA, Bittencourt MS; Writing of the manuscript: Staniak HL, Bittencourt MS; Critical revision of the manuscript for intellectual content: Staniak HL, Sharovsky R, Pereira AC, Castro CC, Bensenor I, Lotufo PA, Bittencourt MS.

Referências

  • 1
    Taylor AJ, Cerqueira M, Hodgson JM, Mark D, Min J, O'Gara P, et al. ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 Appropriate Use Criteria for Cardiac Computed Tomography A Report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the Society of Cardiovascular Computed Tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance. J Am Coll Cardiol. 2010;56(22):1864-94.
  • 2
    Brenner DJ, Hall EJ. Computed tomography-an increasing source of radiation exposure. N Engl J Med. 2007;357(22):2277-84.
  • 3
    Gerber TC, Carr JJ, Arai AE, Dixon RL, Ferrari VA, Gomes AS, et al. Ionizing radiation in cardiac imaging: a science advisory from the American Heart Association Committee on Cardiac Imaging of the Council on Clinical Cardiology and Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention. Circulation. 2009;119(7):1056-65.
  • 4
    Einstein AJ, Henzlova MJ, Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA. 2007;298(3):317-23.
  • 5
    Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC, et al. Estimated radiation dose associated with cardiac CT angiography. JAMA. 2009;301(5):500-7.
  • 6
    Hausleiter J, Meyer TS, Martuscelli E, Spagnolo P, Yamamoto H, Carrascosa P, et al. Image quality and radiation exposure with prospectively ECG-triggered axial scanning for coronary CT angiography: the multicenter, multivendor, randomized PROTECTION-III study. JACC Cardiovasc Imaging. 2012;5(5):484-93.
  • 7
    Qin J, Liu LY, Fang Y, Zhu JM, Wu Z, Zhu KS, et al. 320-detector CT coronary angiography with prospective and retrospective electrocardiogram gating in a single heartbeat: comparison of image quality and radiation dose. Br J Radiol. 2012;85(1015):945-51.
  • 8
    Bischoff B, Hein F, Meyer T, Hadamitzky M, Martinoff S, Schomig A, et al. Impact of a reduced tube voltage on CT angiography and radiation dose: results of the PROTECTION I study. JACC Cardiovasc Imaging. 2009;2(8):940-6.
  • 9
    Imai K, Ikeda M, Kawaura C, Aoyama T, Enchi Y, Yamauchi M. Dose reduction and image quality in CT angiography for cerebral aneurysm with various tube potentials and current settings. Br J Radiol. 2012;85(1017):e673-81.
  • 10
    Starck G, Lönn L, Cederblad Å, Forssell-Aronsson E, Sjöström L, Alpsten M. A method to obtain the same levels of CT image noise for patients of various sizes, to minimize radiation dose. Br J Radiol. 2002;75(890):140-50.
  • 11
    Smedby Ö, Fredrikson M, De Geer J, Borgen L, Sandborg M. Quantifying the potential for dose reduction with visual grading regression. Br J Radiol. 2013;86(1021):20110784.
  • 12
    Bittencourt M, Schmidt B, Seltmann M, Muschiol G, Ropers D, Daniel W, et al. Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience. Int J Cardiovasc Imaging. 2011;27(7):1081-7.
  • 13
    Desai GS, Thabet A, Elias AY, Sahani DV. Comparative assessment of three image reconstruction techniques for image quality and radiation dose in patients undergoing abdominopelvic multidetector CT examinations. Br J Radiol. 2013;86(1021):20120161.
  • 14
    Tatsugami F, Matsuki M, Nakai G, Inada Y, Kanazawa S, Takeda Y, et al. The effect of adaptive iterative dose reduction on image quality in 320-detector row CT coronary angiography. Br J Radiol. 2012;85(1016):e378-82.
  • 15
    Qi LP, Li Y, Tang L, Li YL, Li XT, Cui Y, et al. Evaluation of dose reduction and image quality in chest CT using adaptive statistical iterative reconstruction with the same group of patients. Br J Radiol. 2012;85(1018):e906-11.
  • 16
    Yoshimura N, Sabir A, Kubo T, Lin PJ, Clouse ME, Hatabu H. Correlation between image noise and body weight in coronary CTA with 16-row MDCT. Acad Radiol. 2006;13(3):324-8.
  • 17
    Ghoshhajra BB, Engel LC, Major GP, Verdini D, Sidhu M, Karolyi M, et al. Direct chest area measurement: a potential anthropometric replacement for BMI to inform cardiac CT dose parameters? J Cardiovasc Comput Tomogr. 2011;5(4):240-6.
  • 18
    Voros S, Rivera JJ, Berman DS, Blankstein R, Budoff MJ, Cury RC, et al; Society for Atherosclerosis Imaging and Prevention Tomographic Imaging and Prevention Councils; Society of Cardiovascular Computed Tomography. Guideline for minimizing radiation exposure during acquisition of coronary artery calcium scans with the use of multidetector computed tomography: a report by the Society for Atherosclerosis Imaging and Prevention Tomographic Imaging and Prevention Councils in collaboration with the Society of Cardiovascular Computed Tomography. J Cardiovasc Comput Tomogr. 2011;5(2):75-83.
  • 19
    Ghafourian K, Younes D, Simprini LA, Weigold WG, Weissman G, Taylor AJ. Scout view X-ray attenuation versus weight-based selection of reduced peak tube voltage in cardiac CT angiography. JACC Cardiovasc Imaging. 2012;5(6):589-95.
  • 20
    Gao J, Li J, Earls J, Li T, Wang Q, Dai R. Individualized tube current selection for 64-row cardiac CTA based on analysis of the scout view. Eur J Radiol. 2011;79(2):266-71.
  • 21
    Schuhbaeck A, Schaefer M, Marwan M, Gauss S, Muschiol G, Lell M, et al. Patient-specific predictors of image noise in coronary CT angiography. J Cardiovasc Comput Tomogr. 2013;7(1):39-45.
  • 22
    Huda W, Vance A. Patient radiation doses from adult and pediatric CT. AJR Am J Roentgenol. 2007;188(2):540-6.
  • 23
    Winklehner A, Goetti R, Baumueller S, Karlo C, Schmidt B, Raupach R, et al. Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness. Invest Radiol. 2011;46(12):767-73.
  • Study Association This study is not associated with any post-graduation program.
  • Sources of Funding There were no external funding sources for this study.

Publication Dates

  • Publication in this collection
    01 Nov 2013
  • Date of issue
    Jan 2014

History

  • Received
    27 Mar 2013
  • Reviewed
    27 June 2013
  • Accepted
    07 Aug 2013
Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br