A Comparison of Radiation Dose Between Standard and 3D Angiography in Congenital Heart Disease

Abstracts

Background:

The use of three-dimensional rotational angiography (3D-RA) to assess patients with congenital heart diseases appears to be a promising technique despite the scarce literature available.

Objectives:

The objective of this study was to describe our initial experience with 3D-RA and to compare its radiation dose to that of standard two-dimensional angiography (2D-SA).

Methods:

Between September 2011 and April 2012, 18 patients underwent simultaneous 3D-RA and 2D-SA during diagnostic cardiac catheterization. Radiation dose was assessed using the dose-area-product (DAP).

Results:

The median patient age and weight were 12.5 years and 47.5 Kg, respectively. The median DAP of each 3D-RA acquisition was 1093µGy.m2 and 190µGy.m2 for each 2D-SA acquisition (p<0.01). In patients weighing more than 45Kg (n=7), this difference was attenuated but still significant (1525 µGy.m2 vs.413µGy.m2, p=0.01). No difference was found between one 3D-RA and three 2D-SA (1525µGy.m2 vs.1238 µGy.m2, p = 0.575) in this population. This difference was significantly higher in patients weighing less than 45Kg (n=9) (713µGy.m2 vs.81µGy.m2, P = 0.008), even when comparing one 3D-RA with three 2D-SA (242µGy.m2, respectively, p<0.008). 3D-RA was extremely useful for the assessment of conduits of univentricular hearts, tortuous branches of the pulmonary artery, and aorta relative to 2D-SA acquisitions.

Conclusions:

The radiation dose of 3D-RA used in our institution was higher than those previously reported in the literature and this difference was more evident in children. This type of assessment is of paramount importance when starting to perform 3D-RA.

Heart Defects, Congenital; Coronary Angiography/ radiation effects; Angiography,Digital Subtraction/methods; Catheterization


Fundamento:

O uso da angiografia rotacional tridimensional (AR-3D) na avaliação de pacientes portadores de cardiopatia congênita parece ser promissor, apesar de haver pouca literatura disponível.

Objetivos:

O objetivo deste estudo foi descrever nossa experiência inicial com AR-3D, comparando sua dose de radiação com a da angiografia padrão bidimensional (AP-2D).

Métodos:

Entre setembro de 2011 e abril de 2012, 18 pacientes foram submetidos simultaneamente a AR-3D e AP-2D durante cateterização cardíaca diagnóstica. A dose de radiação foi avaliada através do produto dose-área (DAP).

Resultados:

A mediana de idade e de peso dos pacientes foi de, respectivamente, 12,5 anos e 47,5 kg. O DAP mediano de cada aquisição em AR-3D foi de 1093µGy.m2, e de 190µGy.m2 para cada aquisição em AP-2D (p < 0,01). Nos pacientes que pesavam mais de 45 kg (n=7), essa diferença foi menor, mas ainda significativa (1525µGy.m2 versus 413 µGy.m2, p = 0,01). Não houve diferença entre uma AR-3D e três AP-2D (1525µGy.m2 versus1238µGy.m2, p = 0,575) nesta população. Essa diferença foi significativamente maior em pacientes que pesavam menos de 45 kg (n = 9) (713 µGy.m2 versus 81 µGy.m2, p = 0,008), mesmo quando se comparou uma AR-3D com três AP 2D (242µGy.m2, respectivamente, p < 0,008). A AR-3D mostrou-se extremamente útil na avaliação de condutos de corações univentriculares, ramos pulmonares tortuosos e aorta em relação às aquisições em AP-2D.

Conclusões:

A dose de radiação da AR-3D em nossa instituição foi maior que a anteriormente relatada na literatura, sendo essa diferença mais evidente em crianças. Esse tipo de avaliação é de fundamental importância quando se começa a realizar AR-3D.

Cardiopatias Congênitas; Angiografia Coronária/efeitos de radiação; Angiografia Digital/métodos; Cateterismo


Introduction

Computed tomography angiography, or three-dimensional rotational angiography (3D-RA), was introduced in the neurovascular field to improve the assessment of intracranial aneurysms and to guide interventional neurovascular procedures. 3D-RA quickly became the gold standard for an accurate diagnosis of intracranial aneurysms1Missler U, Hundt C, Wiesmann M, Mayer T, Bruckmann H. Three-dimensional reconstructed rotational digital subtraction angiography in planning treatment of intracranial aneurysms. Eur Radiol. 2000;10(4):564-8.. This technique is also used in electrophysiological and peripheral vascular procedures2Seymour HR, Matson MB, Belli AM, Morgan R, Kyriou J, Patel U. Rotational digital subtraction angiography of the renal arteries: technique and evaluation in the study of native and transplant renal arteries. Br J Radiol. 2001;74(878):134-41. , 3Orlov MV, Hoffmeister P, Chaudhry GM, Almasry I, Gijsbers GH, Swack T, et al. Three-dimensional rotational angiography of the left atrium and esophagus--A virtual computed tomography scan in the electrophysiology lab? Heart Rhythm. 2007;4(1):37-43. and has been tested for coronary evaluation4Agostoni P, Biondi-Zoccai G, Van Langenhove G, Cornelis K, Vermeersch P, Convens C, et al. Comparison of assessment of native coronary arteries by standard versus three-dimensional coronary angiography. Am J Cardiol. 2008;102(3):272-9..

Despite all the recent progress with new diagnostic modalities for congenital heart diseases, cardiac catheterization continues to be of great value in this field because of its diagnostic accuracy and therapeutic advantages. The use of 3D-RA to assess patients with congenital heart diseases appears to be a promising technique despite the scarce literature available5Glatz AC, Zhu X, Gillespie MJ, Hanna BD, Rome JJ. Use of angiographic CT imaging in the cardiac catheterization laboratory for congenital heart disease. JACC Cardiovasc Imaging. 2010;3(11):1149-57.

Panzer J, Taeymans Y, De Wolf D. Three-dimensional rotational angiography of a patient with pulmonary atresia intact septum and coronary fistulas. Pediatr Cardiol. 2008;29(3):686-7.
- 7Rigatelli G, Zamboni A, Cardaioli P. Three-dimensional rotational digital angiography in a complicated case of patent ductusarteriosustranscatheter closure. Catheter Cardiovasc Interv. 2007;70(6):900-3.. Notwithstanding the potential benefits of 3D-RA images in comparison to those of 2D-SA, the radiation dose must be evaluated carefully to ensure that the long-term deleterious effects of radiation to both patients and the medical team are not increased.

The objectives of the present study were to report our initial experience with 3D-RA in patients with congenital heart diseases; to compare its radiation dose to that of standard two-dimensional procedures in the same patient; and to report our initial impressions of the defects best evaluated with this tool.

Methods

Sample

This is a retrospective study that comprised our initial experience with 3D-RA in 18 patients referred for diagnostic cardiac catheterization between September 2011 and April 2012. The objective of the procedure was basically to evaluate the aorta, pulmonary arteries and cavopulmonary connections, apart from one patient with patent ductus arteriosus.

Procedure and Radiation Exposure

All children underwent general anesthesia with orotracheal intubation. Adult patients routinely underwent conscious sedation and local anesthesia for diagnostic procedures, which did not pose any problems for three-dimensional acquisition. All patients were under pacemaker overdrive suppression during 3D-RA to allow adequate filling of the structures studied, for a preselected time frame. All patients or their parents gave informed consent prior to the procedure.

Initially, patients routinely underwent standard two-dimensional imaging. Subsequent 3D-RA was performed to compare and better understand the benefits of this technique in the same patient. As we gained experience, 3D-RA images and reconstruction became easier and 2D-SA was performed only if necessary after evaluating the reconstructed three-dimensional images. This approach was adopted in six patients undergoing only one complementary two-dimensional angiography, and one patient underwent only 3D-RA, without the need for additional images. The cumulative dose (CD) was measured in all procedures. The total dose-area-product (DAP) in each angiography and number of angiographies, as well as the quality and impact of 2D-SA and 3D-RA images, were also compared.

All procedures were performed using the same equipment. The 3D-RA images were acquired with 200-degree rotation using a C-arm-mounted flat-panel monoplane angiographic system (Artis zee, Siemens Medical Solutions, Forchheim, Germany). The images were reconstructed using a commercially available software (syngo DynaCT, Siemens Medical Solutions). The radiation settings were changed from those recommended by the manufacturer in order to reduce X-ray exposure. Three-dimensional rotational image acquisition was performed with a dose of 0.17 µGy/f, a frame rate of 30 frames/s, and tube voltage of 70 Kv. Post-processing reconstruction was performed on a separate workstation in the interventional suite (Leonardo Dyna CT, Siemens Medical Solutions).

In adults, 2D-SA imaging was obtained with a dose of 0.17 µGy/f at 10 to 15 frames/s, and tube voltage of 81 Kv. In children weighing less than 40 Kg, the dose was 0.14 µGy per image, frame rate of 30 frames/s, and tube voltage of 73 Kv. The anti-scatter grid was removed before the procedure in children weighing less than 20 Kg.

To evaluate radiation exposure, the software automatically calculates DAP separately for each 2D-SA or 3D-RA and for fluoroscopic images. The amount of contrast media used was assessed in this study. Institutional review board approval was obtained for this retrospective analysis.

Statistical Analysis

Descriptive statistics, including median and interquartile range, were calculated for continuous variables not normally distributed. The results were compared using the Wilcoxon test (95% confidence level). All statistical calculations were performed using SPSS Statistics 19 (SPSS Inc, Chicago, USA).

Results

Sample Characteristics

The median age of the patients included in the study was 12.5 years (ranging from 1 to 44 years) and the median weight was 47.5 Kg (ranging from 9 to 100 Kg). The diagnoses included 7 patients with tetralogy of Fallot; 5 patients with aortic coarctation; 2 patients previously undergoing Fontan anastomosis; 1 patient with truncus arteriosus; 1 patient with patent ductus arteriosus; 1 patient with pulmonary artery stenosis; and 1 patient with dilatation of the ascending aorta (Table 1).

Table 1
Demographics of Patients Included In The Study

Radiation Dose

The median CD in both 2D-SA and 3D-RA was 171 mGy (interquartile range 40.6 to 1071 mGy). The median total DAP for the 3D-RA protocol was 1093 µGy.m2 (interquartile range 701 to 1767 µGy.m2), and the median DAP for the two-dimensional study in the same patients was 360 µGy.m2 (interquartile range 200 to 1049 µGy.m2). The DAP of two-dimensional angiography alone was 190 µGy.m2 (interquartile range 76 to 413 µGy.m2, p = 0.01).

The difference seemed to be similar in patients weighing more than 45 kilograms. In this population, the median DAP for each 3D-RA acquisition was 1525 µGy.m2 (interquartile range 1074 to 2031 µGy.m2) compared to 413 µGy.m2 in each 2D-SA acquisition in the same patient (p = 0.01). If we consider a mean of three two-dimensional angiographies for an adequate procedure, no difference was found in radiation exposure between two-dimensional study and 3D-RA (1525 µGy.m2 vs. 1238 µGy.m2, p = 0.575)

However, if we only evaluate patients weighing less than 45 Kg, this difference is even greater. The median DAP from a single 3D-RA acquisition in this population was 713 µGy.m2 (interquartile range 422 to 1090 µGy.m2) and the median DAP for each two-dimensional acquisition in the same patient was 81 µGy.m2 (p = 0.008). If we consider a mean of three angiographies for an adequate study, the median DAP is 242 µGy.m2 (interquartile range 183 to 423 µGy.m2). This difference is still favorable to 2D-SA acquisitions in this population (p = 0.008). Some of these results are summarized in Table 2.

Table 2
Comparison between DAP of one 3D-RA and one and three 2D-SA

Imaging Quality and Additional Information

Of 11 patients undergoing both a complete two-dimensional standard evaluation and subsequently a 3D-RA, we verified that 3D-RA provided additional information in 3 (27%). Apart from 1 patient with patent ductus arteriosus, all other images were acceptable and at least comparable to those of 2D-SA.

Discussion

Children and adults with congenital heart disease are increasingly more frequently undergoing successive diagnostic or therapeutic radiation-based procedures. The harmful cumulative long-term effects of X-ray exposure are well known and extensively described in the literature. DAP has been previously correlated with peak skin dose and can be used as a reliable and online parameter to guide radiation exposure during cardiac catheterization8Bacher K, Bogaert E, Lapere R, De Wolf D, Thierens H. Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation. 2005;111(1):83-9.

El Sayed MH, Roushdy AM, El Farghaly H, El Sherbini A. Radiation exposure in children during the current era of pediatric cardiac intervention. Pediatr Cardiol. 2012;33(1):27-35.
- 1010 Grech V, Grech M, Debono J, Xuereb RG, Fenech A. Greater radiation savings at higher body mass indexes with dual axis rotational coronary angiography. Catheter Cardiovasc Interv. 2013;81(1):170-1..

New advances in technology intend to improve diagnostic accuracy without increasing the risks for patients1111 Chambers CE, Fetterly KA, Holzer R, Lin PJ, Blankenship JC, Balter S, et al. Radiation safety program for the cardiac catheterization laboratory. Catheter Cardiovasc Interv. 2011;77(4):546-56. , 1212 Smith BG, Tibby SM, Qureshi SA, Rosenthal E, Krasemann T. Quantification of temporal, procedural, and hardware-related factors influencing radiation exposure during pediatric cardiac catheterization. Catheter Cardiovasc Interv. 2012;80(6):931-6..

The use of 3D-RA was initially described for the assessment of intracranial aneurysms and was also reported in electrophysiological and peripheral vascular procedures, as well as for the assessment of coronary arteries1Missler U, Hundt C, Wiesmann M, Mayer T, Bruckmann H. Three-dimensional reconstructed rotational digital subtraction angiography in planning treatment of intracranial aneurysms. Eur Radiol. 2000;10(4):564-8. , 4Agostoni P, Biondi-Zoccai G, Van Langenhove G, Cornelis K, Vermeersch P, Convens C, et al. Comparison of assessment of native coronary arteries by standard versus three-dimensional coronary angiography. Am J Cardiol. 2008;102(3):272-9. , 1313 Gailloud P, Oishi S, Carpenter J, Murphy KJ. Three-dimensional digital angiography: new tool for simultaneous three-dimensional rendering of vascular and osseous information during rotational angiography. AJNR Am J Neuroradiol. 2004;25(4):571-3.

14 Bridcut RR, Murphy E, Workman A, Flynn P, Winder RJ. Patient dose from 3D rotational neurovascular studies. Br J Radiol. 2007;80(953):362-6.

15 Akhtar M, Vakharia KT, Mishell J, Gera A, Ports TA, Yeghiazarians Y, et al. Randomized study of the safety and clinical utility of rotational vs. standard coronary angiography using a flat-panel detector. Catheter Cardiovasc Interv. 2005;66(1):43-9.

16 Raman SV, Morford R, Neff M, Attar TT, Kukielka G, Magorien RD, et al. Rotational X-ray coronary angiography. Catheter Cardiovasc Interv. 2004;63(2):201-7.

17 Grech M, Debono J, Xuereb RG, Fenech A, Grech V. A comparison between dual axis rotational coronary angiography and conventional coronary angiography. Catheter Cardiovasc Interv. 2012;80(4):576-80.
- 1818 Maddux JT, Wink O, Messenger JC, Groves BM, Liao R, Strzelczyk J, et al. Randomized study of the safety and clinical utility of rotational angiography versus standard angiography in the diagnosis of coronary artery disease. Catheter Cardiovasc Interv. 2004;62(2):167-74..

The rationale for incorporating this technology in the field of congenital heart disease was to obtain additional information not available from 2D-SA. Despite the scarce information in the literature5Glatz AC, Zhu X, Gillespie MJ, Hanna BD, Rome JJ. Use of angiographic CT imaging in the cardiac catheterization laboratory for congenital heart disease. JACC Cardiovasc Imaging. 2010;3(11):1149-57. , 1919 Glöckler M, Koch A, Greim V, Shabaiek A, Rüffer A, Cesnjevar R, et al. The value of flat-detector computed tomography during catheterisation of congenital heart disease. Eur Radiol. 2011;21(12):2511-20. , 2020 Kapins CEB, Coutinho RB, Barbosa FB, Silva CMC, Lima VC, Carvalho AC. Use of 3D rotational angiography (3D-RA) in congenital heart disease: experience with 53 patients. Rev Bras Cardiol Invasiva. 2010;18(2):199-203., the use of 3D-RA in this context reduces the number of necessary acquisitions and brings the possibility of increasing accuracy in the assessment of patients with defects with a complex anatomy; however, it is not yet well established whether 3D-RA adds information on these defects.

In 2010, Kapins et al2020 Kapins CEB, Coutinho RB, Barbosa FB, Silva CMC, Lima VC, Carvalho AC. Use of 3D rotational angiography (3D-RA) in congenital heart disease: experience with 53 patients. Rev Bras Cardiol Invasiva. 2010;18(2):199-203. first described a series of 53 cases of diagnostic catheterization using 2D-SA and 3D-RA in patients with congenital heart disease and a mean weight of 28.5Kg2020 Kapins CEB, Coutinho RB, Barbosa FB, Silva CMC, Lima VC, Carvalho AC. Use of 3D rotational angiography (3D-RA) in congenital heart disease: experience with 53 patients. Rev Bras Cardiol Invasiva. 2010;18(2):199-203.. At least half of the patients benefited from this new technology in the planning of their further surgical or percutaneous procedures. No differences were found in DAP between two-dimensional angiographies and 3D-RA (356 x 374 µGy.m2) 2020 Kapins CEB, Coutinho RB, Barbosa FB, Silva CMC, Lima VC, Carvalho AC. Use of 3D rotational angiography (3D-RA) in congenital heart disease: experience with 53 patients. Rev Bras Cardiol Invasiva. 2010;18(2):199-203..

Glatz et al. described the use of 3D-RA in 41 cases of congenital heart disease5Glatz AC, Zhu X, Gillespie MJ, Hanna BD, Rome JJ. Use of angiographic CT imaging in the cardiac catheterization laboratory for congenital heart disease. JACC Cardiovasc Imaging. 2010;3(11):1149-57.. Despite providing high-quality diagnostic imaging in over 70% of cases, this technique provided additional information in less than 30% of the cases studied. After using a phantom test to measure the DAP for different settings, the authors found a radiation dose of approximately 122 µGy.m2 for a 10 year-old child and 471 µGy.m2 for an adult male.

In 2011, Glockler et al1919 Glöckler M, Koch A, Greim V, Shabaiek A, Rüffer A, Cesnjevar R, et al. The value of flat-detector computed tomography during catheterisation of congenital heart disease. Eur Radiol. 2011;21(12):2511-20. evaluated and reported their experience in 62 patients with a median weight of 14.4 Kg undergoing rotational angiography for the assessment of pulmonary arteries, aorta, and cavopulmonary connections. In this study, the images were considered superior to those of conventional angiography in 90.3% of cases and very useful for guiding interventional procedures using overlaying three-dimensional images. The median DAP from each run was 111 µGy.m2.

More recently, Berman et al2121 Berman DP, Khan DM, Gutierrez Y, Zahn EM. The use of three-dimensional rotational angiography to assess the pulmonary circulation following cavo-pulmonary connection in patients with single ventricle. Catheter Cardiovasc Interv. 2012;80(6):922-30. described the usefulness of 3D-RA in patients previously undergoing cavopulmonary connections with a clear understanding of the mechanism of pulmonary stenosis2121 Berman DP, Khan DM, Gutierrez Y, Zahn EM. The use of three-dimensional rotational angiography to assess the pulmonary circulation following cavo-pulmonary connection in patients with single ventricle. Catheter Cardiovasc Interv. 2012;80(6):922-30.. The median DAP was 306 µGy.m2 in this population with a median weight of 15.7Kg. The authors reported unpublished data with even smaller doses in three-dimensional procedures after working with the company and physicists.

The lower dose of radiation exposure compared to that used in our study can be explained by the lower median weight of patients in all these studies reported. This linear dose variation among patients of different weights has already been described in the literature8Bacher K, Bogaert E, Lapere R, De Wolf D, Thierens H. Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation. 2005;111(1):83-9.

El Sayed MH, Roushdy AM, El Farghaly H, El Sherbini A. Radiation exposure in children during the current era of pediatric cardiac intervention. Pediatr Cardiol. 2012;33(1):27-35.
- 1010 Grech V, Grech M, Debono J, Xuereb RG, Fenech A. Greater radiation savings at higher body mass indexes with dual axis rotational coronary angiography. Catheter Cardiovasc Interv. 2013;81(1):170-1..

In the present study, the median DAP of one run of 3D-RA was higher than that of a two-dimensional view alone. However, if we consider that multiple two-dimensional projections are sometimes necessary, this difference is considerably reduced and the benefits from 3D-RA can surpass the slightly increased radiation dose.

In the subgroup of patients with less than 45Kg, the difference detected between both groups was much higher. Even comparing to 3 two-dimensional angiographies, the difference in radiation exposure was favorable to the standard technique and may not justify the benefits obtained from this technique in this particular population.

However, if we consider only patients weighing more than 45 Kg, the difference between each 3D-RA and 2D-SA alone was not prohibitive. Considering the need for multiple two-dimensional views, this difference was similar, or even favorable to 3D-RA and can support the recommendation of the use of 3D-RA for the assessment of patients with aortic coarctation or Fontan anastomosis, and for the evaluation of the right ventricle and pulmonary arteries.

The difference observed in the pediatric population and the higher radiation exposure, when compared to that reported in the literature, led our local physicist to work hard together with the manufacturer's technicians to reduce radiation exposure from 3D-RA performed in our institution. This effort eventually resulted in an important reduction in the radiation dose and reinforced the importance of this assessment whenever any center intends to start performing tridimensional angiographies. As already described by Berman et al2121 Berman DP, Khan DM, Gutierrez Y, Zahn EM. The use of three-dimensional rotational angiography to assess the pulmonary circulation following cavo-pulmonary connection in patients with single ventricle. Catheter Cardiovasc Interv. 2012;80(6):922-30., this confirms that cooperation between local staff and companies is a process that must be continuously evaluated and the goal is to achieve the lowest possible radiation dose without losing image quality. If not adequately surveyed, the dose can be prohibitive and may not justify the benefits obtained from the technique in certain patient populations. We strongly recommend this evaluation to all institutions that perform 3D-RA. Further prospective studies quantifying the radiation dose both on the patient's skin level and on the operator are necessary to answer some questions regarding radiation exposure in 3D-RA, primarily in children with congenital heart disease, and to determine whether rotational angiography can safely replace 2D-SA acquisitions. Akhtar described the presence of a learning curve for rotational angiography with benefits obtained in the latter half of rotational procedures1515 Akhtar M, Vakharia KT, Mishell J, Gera A, Ports TA, Yeghiazarians Y, et al. Randomized study of the safety and clinical utility of rotational vs. standard coronary angiography using a flat-panel detector. Catheter Cardiovasc Interv. 2005;66(1):43-9.. In the field of congenital heart diseases, the learning curve is essential for elucidating which cases will benefit from rotational 3D-DSA studies without requiring additional two-dimensional acquisitions.

Three-dimensional rotational angiography is very useful in patients diagnosed with aortic coarctation. A single acquisition can demonstrate the size of the ascending aorta, features from the aortic arch hardly viewed in two-dimensional studies, and the presence of aneurysm in patients previously undergoing percutaneous or surgical procedures (Figure 1). In one patient, additional information on aortic arch tortuosity was available because of the spatial resolution provided by three-dimensional reconstruction. The possibility of evaluating the ascending aorta with the same contrast injection is useful in patients at an increased risk of progressive dilatation. An additional 2D-SA injection in these patients does not seem to be useful to provide more information from three-dimensional reconstruction.

Figure 1
Bare stent previously implanted in a patient with aortic coarctation with an adequate late result

In the assessment of the pulmonary arteries, the possibility of reconstruction can allow visualization of the spatial relationship not feasible in two-dimensional procedures and is useful for guiding or planning percutaneous and surgical interventions. Considering eight patients previously operated on for pulmonary circulation (7 patients with tetralogy of Fallot and 1 patient with truncus arteriosus) and one patient with native pulmonary artery stenosis, three-dimensional assessment was elucidative with clear and useful images of the pulmonary branches, subvalvar region, and previously implanted stents (Figure 2). The images were considered superior to 2D-SA acquisitions in patients with an aneurysmatic right ventricular outflow tract and those with tortuous branches of the pulmonary artery, providing optimal assessment of these lesions. Finally, stent fractures and neointimal hyperplasia were better evaluated in two-dimensional angiographies, with little additional information to be obtained from 3D-RA.

Figure 2
Late post-operative period of truncus arteriosus with a tubular stenosis of the left pulmonary artery.

In patients with Fontan anastomosis, the slow flow from the right atrial or conduit anastomosis sometimes makes visualization of the pulmonary arteries difficult and requires a lot of different two-dimensional projections. This issue is not a concern in 3D-RA and only one acquisition can delineate the anastomosis and pulmonary branches clearly. In the two patients previously undergoing Fontan anastomosis, 3D-RA provided better images than four or five two-dimensional angiographies in different projections. Careful placement of the angiographic catheter ensured good visualization of the conduit, pulmonary arteries, and Glenn anastomosis.

The only case in which 3D-RA did not provide good visualization was that of a patient with ductus arteriosus. The defect could not be well reconstructed and we suspect it was a technical problem related to the learning curve period.

The fact that intravascular stenosis like neointimal hyperplasia with or without stent fractures could be better visualized in two-dimensional views could be related to failure of contrast filling of this part of the vessel during the acquisition time. Glatz has already described the "dropout" phenomenon in areas of tight stenosis or stented regions and warned about the risks of misdiagnosis that not uncommonly occur with computed tomography or cardiac magnetic resonance and that could also be present during 3D-RA5Glatz AC, Zhu X, Gillespie MJ, Hanna BD, Rome JJ. Use of angiographic CT imaging in the cardiac catheterization laboratory for congenital heart disease. JACC Cardiovasc Imaging. 2010;3(11):1149-57..

Limitations

The retrospective single-center design is a limitation of our study. Moreover, the patients included in this study comprised the initial experience with 3D-RA in our institution, and only 11 patients underwent both 3D-RA and a complete 2D-SA study. However, we considered it was necessary due to the scarce literature available in the beginning of the experience. From our point of view, the lack of information on the benefits of and radiation exposure from three-dimensional angiography in the assessment of congenital heart diseases justified the use of both approaches in the same patient. The amount of contrast used was not assessed, which is another limitation of the study and can be an additional benefit of this approach.

Conclusion

The use of 3D-RA seems to be of great value in the field of congenital heart diseases. The radiation dose of 3D-RA performed in our institution was higher than that of 2D-SA and this difference was more evident in children. The radiation dose was even higher than that previously reported in the literature for 3D-RA. The information obtained from this initial experience and the comparison with the literature available were of paramount importance to work on radiation dose reduction.

  • Study Association
    This study is not associated with any thesis or dissertation work.
  • Author contributions
    Conception and design of the research: Manica JLL, Medeiros RF; Acquisition of data: Manica JLL, Rossi Filho RI, Fischer LS, Medeiros RF, Broetto G, Borges MS; Analysis and interpretation of the data: Manica JLL, Fischer LS, Medeiros RF, Broetto G; Statistical analysis: Manica JLL; Writing of the manuscript: Manica JLL, Rossi Filho RI; Critical revision of the manuscript for intellectual content: Manica JLL, Rossi Filho RI, Borges MS.
  • Sources of Funding
    There were no external funding sources for this study.

References

  • 1
    Missler U, Hundt C, Wiesmann M, Mayer T, Bruckmann H. Three-dimensional reconstructed rotational digital subtraction angiography in planning treatment of intracranial aneurysms. Eur Radiol. 2000;10(4):564-8.
  • 2
    Seymour HR, Matson MB, Belli AM, Morgan R, Kyriou J, Patel U. Rotational digital subtraction angiography of the renal arteries: technique and evaluation in the study of native and transplant renal arteries. Br J Radiol. 2001;74(878):134-41.
  • 3
    Orlov MV, Hoffmeister P, Chaudhry GM, Almasry I, Gijsbers GH, Swack T, et al. Three-dimensional rotational angiography of the left atrium and esophagus--A virtual computed tomography scan in the electrophysiology lab? Heart Rhythm. 2007;4(1):37-43.
  • 4
    Agostoni P, Biondi-Zoccai G, Van Langenhove G, Cornelis K, Vermeersch P, Convens C, et al. Comparison of assessment of native coronary arteries by standard versus three-dimensional coronary angiography. Am J Cardiol. 2008;102(3):272-9.
  • 5
    Glatz AC, Zhu X, Gillespie MJ, Hanna BD, Rome JJ. Use of angiographic CT imaging in the cardiac catheterization laboratory for congenital heart disease. JACC Cardiovasc Imaging. 2010;3(11):1149-57.
  • 6
    Panzer J, Taeymans Y, De Wolf D. Three-dimensional rotational angiography of a patient with pulmonary atresia intact septum and coronary fistulas. Pediatr Cardiol. 2008;29(3):686-7.
  • 7
    Rigatelli G, Zamboni A, Cardaioli P. Three-dimensional rotational digital angiography in a complicated case of patent ductusarteriosustranscatheter closure. Catheter Cardiovasc Interv. 2007;70(6):900-3.
  • 8
    Bacher K, Bogaert E, Lapere R, De Wolf D, Thierens H. Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation. 2005;111(1):83-9.
  • 9
    El Sayed MH, Roushdy AM, El Farghaly H, El Sherbini A. Radiation exposure in children during the current era of pediatric cardiac intervention. Pediatr Cardiol. 2012;33(1):27-35.
  • 10
    Grech V, Grech M, Debono J, Xuereb RG, Fenech A. Greater radiation savings at higher body mass indexes with dual axis rotational coronary angiography. Catheter Cardiovasc Interv. 2013;81(1):170-1.
  • 11
    Chambers CE, Fetterly KA, Holzer R, Lin PJ, Blankenship JC, Balter S, et al. Radiation safety program for the cardiac catheterization laboratory. Catheter Cardiovasc Interv. 2011;77(4):546-56.
  • 12
    Smith BG, Tibby SM, Qureshi SA, Rosenthal E, Krasemann T. Quantification of temporal, procedural, and hardware-related factors influencing radiation exposure during pediatric cardiac catheterization. Catheter Cardiovasc Interv. 2012;80(6):931-6.
  • 13
    Gailloud P, Oishi S, Carpenter J, Murphy KJ. Three-dimensional digital angiography: new tool for simultaneous three-dimensional rendering of vascular and osseous information during rotational angiography. AJNR Am J Neuroradiol. 2004;25(4):571-3.
  • 14
    Bridcut RR, Murphy E, Workman A, Flynn P, Winder RJ. Patient dose from 3D rotational neurovascular studies. Br J Radiol. 2007;80(953):362-6.
  • 15
    Akhtar M, Vakharia KT, Mishell J, Gera A, Ports TA, Yeghiazarians Y, et al. Randomized study of the safety and clinical utility of rotational vs. standard coronary angiography using a flat-panel detector. Catheter Cardiovasc Interv. 2005;66(1):43-9.
  • 16
    Raman SV, Morford R, Neff M, Attar TT, Kukielka G, Magorien RD, et al. Rotational X-ray coronary angiography. Catheter Cardiovasc Interv. 2004;63(2):201-7.
  • 17
    Grech M, Debono J, Xuereb RG, Fenech A, Grech V. A comparison between dual axis rotational coronary angiography and conventional coronary angiography. Catheter Cardiovasc Interv. 2012;80(4):576-80.
  • 18
    Maddux JT, Wink O, Messenger JC, Groves BM, Liao R, Strzelczyk J, et al. Randomized study of the safety and clinical utility of rotational angiography versus standard angiography in the diagnosis of coronary artery disease. Catheter Cardiovasc Interv. 2004;62(2):167-74.
  • 19
    Glöckler M, Koch A, Greim V, Shabaiek A, Rüffer A, Cesnjevar R, et al. The value of flat-detector computed tomography during catheterisation of congenital heart disease. Eur Radiol. 2011;21(12):2511-20.
  • 20
    Kapins CEB, Coutinho RB, Barbosa FB, Silva CMC, Lima VC, Carvalho AC. Use of 3D rotational angiography (3D-RA) in congenital heart disease: experience with 53 patients. Rev Bras Cardiol Invasiva. 2010;18(2):199-203.
  • 21
    Berman DP, Khan DM, Gutierrez Y, Zahn EM. The use of three-dimensional rotational angiography to assess the pulmonary circulation following cavo-pulmonary connection in patients with single ventricle. Catheter Cardiovasc Interv. 2012;80(6):922-30.