Comparison between MDCT and Grayscale IVUS in a Quantitative Analysis of Coronary Lumen in Segments with or without Atherosclerotic Plaques

Abstracts

Background:

The diagnostic accuracy of 64-slice MDCT in comparison with IVUS has been poorly described and is mainly restricted to reports analyzing segments with documented atherosclerotic plaques.

Objectives:

We compared 64-slice multidetector computed tomography (MDCT) with gray scale intravascular ultrasound (IVUS) for the evaluation of coronary lumen dimensions in the context of a comprehensive analysis, including segments with absent or mild disease.

Methods:

The 64-slice MDCT was performed within 72 h before the IVUS imaging, which was obtained for at least one coronary, regardless of the presence of luminal stenosis at angiography. A total of 21 patients were included, with 70 imaged vessels (total length 114.6 ± 38.3 mm per patient). A coronary plaque was diagnosed in segments with plaque burden > 40%.

Results:

At patient, vessel, and segment levels, average lumen area, minimal lumen area, and minimal lumen diameter were highly correlated between IVUS and 64-slice MDCT (p < 0.01). However, 64-slice MDCT tended to underestimate the lumen size with a relatively wide dispersion of the differences. The comparison between 64-slice MDCT and IVUS lumen measurements was not substantially affected by the presence or absence of an underlying plaque. In addition, 64-slice MDCT showed good global accuracy for the detection of IVUS parameters associated with flow-limiting lesions.

Conclusions:

In a comprehensive, multi-territory, and whole-artery analysis, the assessment of coronary lumen by 64-slice MDCT compared with coronary IVUS showed a good overall diagnostic ability, regardless of the presence or absence of underlying atherosclerotic plaques.

Coronary Artery Disease, Plaque, Atherosclerotic / diagnosis; Multidetector Computed Tomography / utilization; Ultrasonography / utilization


Fundamento:

A comparação do rigor diagnóstico da TCMD e da USIV foi pouco descrita, estando restrita principalmente a estudos que avaliaram segmentos com placas ateroscleróticas já documentadas.

Objetivos:

Este estudo objetiva avaliar o desempenho diagnóstico da tomografia computadorizada com 64 colunas de detectores (tomografia computadorizada multidetector- TCMD) e da ultrassonografia intravascular (USIV) em escala de cinza na verificação das dimensões da luz coronária, em um contexto mais amplo, incluindo também segmentos coronários sadios e com patologia leve.

Métodos:

A TCMD foi realizada em todos os pacientes antes da realização da USIV, com um intervalo < 72 horas entre os dois exames. Imagens de USIV foram obtidas de pelo menos uma coronária, independente da presença de estenose luminal durante a angiografia. Um total de 21 pacientes foram incluídos, com imagens de 70 vasos (comprimento total 114,6 ± 38.3 mm por paciente). Placas coronárias foram diagnosticada em segmentos com carga de placas > 40%.

Resultados:

Uma alta correlação entre as medidas de TCMD e USIV para área luminar média, área luminar mínima e diâmetro luminar mínimo foi encontrada no nível de paciente, vaso e segmento (p-valor < 0,01 para todas as correlações). A TCMD, no entanto, tendeu a subestimar o tamanho luminar com uma dispersão de diferenças relativamente ampla. A comparação entre as medidas da luz por TCMD e USIV não foi substancialmente afetada pela presença ou ausência de placa subjacente. Além disso, a TCMD mostrou boa precisão geral na detecção de parâmetros associados a lesões limitantes de fluxo.

Conclusão:

Em uma análise compreensiva e multi-focal da luz coronária, demonstramos bom desempenho diagnóstico da TCMD, quando comparada a USIV, independente da presença de placas ateroscleróticas adjacentes.

Doença da Artéria Coronariana; Placa Aterosclerótica / diagnóstico; Tomografia Computadorizada Multidetectores / utilização; Ultrassonografia / utilização


Condensed Abstract

A total of 21 patients (70 vessels; total length 114.6 ± 38.3 mm per patient) underwent 64-slice MDCT and IVUS imaging. At patient, vessel, and segment levels, average lumen area, minimal lumen area, and minimal lumen diameter were highly correlated between IVUS and 64-slice MDCT. The comparison between 64-slice MDCT and IVUS lumen measurements was not substantially affected by the presence or absence of an underlying plaque (defined as IVUS plaque burden > 40%). In addition, 64-slice MDCT showed good global accuracy for the detection of IVUS parameters associated with flow‑limiting lesions.

List of Abbreviations

EEL External elastic lamina

HU Hounsfield unit

IVUS Intravascular ultrasound

MDCT Multidetector computed

ROC Receiver operating characteristic

CTA Computed tomography angiography

Introduction

Multidetector computed tomography (MDCT) is a reliable noninvasive method for the detection of coronary luminal stenosis. However, most validation studies have compared MDCT with catheter-based angiography, a method that is known to have limitations1Meijer AB, O YL, Geleijns J, Kroft LJ. Meta-analysis of 40- and 64-MDCT angiography for assessing coronary artery stenosis. AJR Am J Roentgenol. 2008;191(6):1667-75.,2Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I, et al. Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med. 2008;359(22):2324-36.. Intravascular ultrasound (IVUS) is currently considered the gold standard for the in vivo assessment of coronary luminal dimensions. Compared with quantitative angiography, morphologic data from IVUS imaging have shown a better correlation with coronary flow measurements and noninvasive detection of ischemia3Briguori C, Anzuini A, Airoldi F, Gimelli G, Nishida T, Adamian M, et al. Intravascular ultrasound criteria for the assessment of the functional significance of intermediate coronary artery stenoses and comparison with fractional flow reserve. Am J Cardiol. 2001;87(2):136-41.. Moreover, IVUS is a valuable prognostic parameter for patients deferred from coronary angioplasty4Abizaid AS, Mintz GS, Mehran R, Abizaid A, Lansky AJ, Pichard AD, et al. Long-term follow-up after percutaneous transluminal coronary angioplasty was not performed based on intravascular ultrasound findings: importance of lumen dimensions. Circulation. 1999;100(3):256-61..

A number of studies have previously compared MDCT with intravascular ultrasound, indicating a good conformity between these two methods5Springer I, Dewey M. Comparison of multislice computed tomography with intravascular ultrasound for detection and characterization of coronary artery plaques: a systematic review. Eur J Radiol. 2009;71(2):275-82.,6Joshi SB, Okabe T, Roswell RO, Weissman G, Lopez CF, Lindsay J, et al. Accuracy of computed tomographic angiography for stenosis quantification using quantitative coronary angiography or intravascular ultrasound as the gold standard. Am J Cardiol. 2009;104(8):1047-51.. Recent meta-analysis data showed that, compared with IVUS, coronary computed tomography angiography (CTA) appears to be highly accurate for the estimation of luminal area, percentage of area stenosis, plaque volume, and plaque area and for the detection of plaque7Fischer C, Hulten E, Belur P, Smith R, Voros S, Villines TC. Coronary CT angiography versus intravascular ultrasound for estimation of coronary stenosis and atherosclerotic plaque burden: a meta-analysis. J Cardiovasc Comput Tomogr. 2013;7(4):256-66.. However, most studies evaluated specific regions of interest within the coronary tree, restricting the analysis to segments with atherosclerotic plaques by IVUS5Springer I, Dewey M. Comparison of multislice computed tomography with intravascular ultrasound for detection and characterization of coronary artery plaques: a systematic review. Eur J Radiol. 2009;71(2):275-82.,6Joshi SB, Okabe T, Roswell RO, Weissman G, Lopez CF, Lindsay J, et al. Accuracy of computed tomographic angiography for stenosis quantification using quantitative coronary angiography or intravascular ultrasound as the gold standard. Am J Cardiol. 2009;104(8):1047-51.,7Fischer C, Hulten E, Belur P, Smith R, Voros S, Villines TC. Coronary CT angiography versus intravascular ultrasound for estimation of coronary stenosis and atherosclerotic plaque burden: a meta-analysis. J Cardiovasc Comput Tomogr. 2013;7(4):256-66.. It is important to note that, in clinical practice, MDCT images are commonly analyzed following an all-segment scheme without the aid of a coronary map to focus the assessment on specific regions of interest.

It is important to evaluate the diagnostic performance of MDCT in a comprehensive manner, including both coronary segments with atherosclerosis as well as portions without significant disease. Therefore, the present study compared 64-slice MDCT with IVUS for the quantitative assessment of coronary luminal dimensions following an inclusive, whole-artery analytic strategy.

Methods

Overall Protocol Design and Study Population

The study included patients with known coronary artery disease who were scheduled for elective coronary stent implantation. Intravascular ultrasound examination was performed at the time of the percutaneous procedure, before any therapeutic intervention. At least one major epicardial vessel without previous angioplasty or surgical treatment was examined with IVUS, but operators were strongly encouraged to acquire IVUS images from all three coronary territories. MDCT was performed 72 h prior to the IVUS procedure as part of the study protocol. The study protocol was approved by the local ethics committee, and written informed consent was obtained from each patient.

Ivus Imaging and Analysis

The IVUS examination was carried out after the administration of intracoronary nitrate, utilizing a 20 MHz electronic multi-array 2.9F catheter (Eagle Eye®, Volcano Corporation Inc., San Diego, CA, USA) connected to a dedicated console (InVision Gold®, Volcano Corporation Inc., San Diego, CA, USA). Acquisition was performed during motorized pullback at a constant speed of 0.5 mm/s (R-100® pullback device, Volcano Corporation Inc., San Diego, Ca, USA). Off-line IVUS analyses were performed by operators blinded to other patients’ characteristics, using a dedicated software (pcVH 2.2 ®; Volcano Corporation Inc., San Diego, CA, USA) for the measurement of vessels. The luminal and external elastic lamina (EEL) boundaries were traced semi‑automatically to obtain area measurements. Total plaque plus media area was calculated as EEL area minus lumen area. Plaque burden was calculated as the plaque plus media area divided by the EEL area, multiplied by 100. A coronary plaque was diagnosed to be present in all segments with plaque burden > 40%. A coronary segment was considered to be free of atherosclerosis or having only mild atherosclerotic plaque when the plaque burden was ≤ 40%.

Multidetector Computer Tomography

Patients with a heart rate greater than 65 bpm received up to 15 mg of intravenous metoprolol before the acquisition of MDCT images, unless contraindicated. Sublingual nitrate was given to all patients prior to the acquisition of MDCT images8Decramer I, Vanhoenacker PK, Sarno G, Van Hoe L, Bladt O, Wijns W, et al. Effects of sublingual nitroglycerin on coronary lumen diameter and number of visualized septal branches on 64-MDCT angiography. AJR Am J Roentgenol. 2008;190(1):219-25.. All scans were performed utilizing a 64-slice MDCT scanner (Aquillion 64TM, Toshiba Medical Systems, Japan). Acquisition protocols were applied as previously described9Miller JM, Dewey M, Vavere AL, Rochitte CE, Niinuma H, Arbab-Zadeh A, et al. Coronary CT angiography using 64 detector rows: methods and design of the multi-centre trial CORE-64. Eur Radiol. 2009;19(4):816-28.. Patients first underwent calcium scanning performed with prospective electrocardiogram gating and a detector collimation of 64 x 3.0 mm. The trigger delay for prospective gating was adjusted to the heart rate in order to obtain the images during the rest period of the coronary arteries. Only patients with a total calcium score < 600 were included. After calcium scanning, the CT coronary angiography protocol was performed using a 64 x 0.5 mm detector collimation and a minimum gantry rotation time of 400 ms, and the scanner settings including pitch were adjusted according to patient’s sex and weight9Miller JM, Dewey M, Vavere AL, Rochitte CE, Niinuma H, Arbab-Zadeh A, et al. Coronary CT angiography using 64 detector rows: methods and design of the multi-centre trial CORE-64. Eur Radiol. 2009;19(4):816-28.. After the intravenous injection of contrast medium (Iopamidol, 370 mg iodine/ml, Bracco) and after reaching a threshold of 180 Hounsfield units (HU), at the descending aorta, the helical scan for CT coronary angiography was performed.

Quantitative coronary analysis was performed with a dedicated workstation (Vitrea2, v3.9 loaded with the Coronary Vessel Probe and SURE Plaque software; all by Vital Images Inc, Plymouth, MN, USA). All analyses were performed by operators blinded to clinical and IVUS characteristics, following a predefined analytic plan. First, the average lumen attenuation was measured in the ascending aorta (at the level of the coronary ostia) and used as the parameter to set the threshold for the detection of the coronary lumen borders. Then, the lumen and the vessel borders were semiautomatically reconstructed for the measurement of their areas (Figure 1); thus, permitting the calculation of the plaque plus media area (vessel area minus lumen area) and the percent of plaque burden (plaque plus media area divided by the vessel area, multiplied by 100).

Figure 1
Schematic representation of quantitative coronary analysis by 64-slice MDCT (above) and IVUS (below). The upper images show the longitudinal reconstruction for both methods. The lower images show illustrative vessel cross-sections with the outer vessel and lumen (inner) boundaries tracings.

IVUS and MDCT Imaging Coregistration and Segment Analysis

Intravascular ultrasound and 64-slice MDCT images were matched by using reproducible vessel landmarks (e.g., aorto-ostial junction, side branches, etc.) as anatomical axial references. Visible landmarks through the target vessel length were annotated for both IVUS and 64-slice MDCT, and all vessel portions were carefully scrutinized to match the exact counterpart (Figure 1). The left main coronary was computed as a separate vessel for analysis. Repeated measurements by the same operators were performed for 27 segments, randomly selected, and correlation coefficients were used to determine intra-observer and inter-observer variability.

Statistical Analysis

The present study is an exploratory analysis without a prior formal sample size calculation. Nevertheless, the chosen sample size of 20 patients would be sufficient to demonstrate a significant correlation of any given continuous parameter with a Pearson coefficient of 0.65at an alfa of 0.05 and a power of 0.9. Statistical analyses were performed with software SPSS 21 (IBM Corp. Released 2012, IBM SPSS Statistics for Windows, Version 21.0. Armonk, NY, USA). Categorical variables are presented as percentages and continuous variables as means and standard deviations. Continuous variables were analyzed for correlation using the Pearson’s correlation test. The method proposed by Bland and Altman1010 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10.,1111 Bland JM, Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995;346(8982):1085-7. was utilized to evaluate the differences between the measurements of 64-slice MDCT and IVUS for mean lumen area, minimal lumen area, and minimal lumen diameter. In Bland and Altman’s method, the average of the two measurements (64-slice MDCT and IVUS) is plotted against the difference between them (64-slice MDCT minus IVUS) with 95% limits of agreement calculated to evaluate the measurement concordance. For all comparison tests, p < 0.05 was considered significant.

Results

The study population consisted of 21 patients with 70 imaged vessels (3.3 vessels per patient) and a final total length of 114.6 ± 38.3 mm per patient (ranging from 39.5 mm to 181.0 mm) (Table 1). All patients had coronary artery disease with percutaneous coronary intervention indication. Men were 71% of the study population, and mean age was 56 years. There was a high prevalence of diabetes mellitus (48%) and hypertension (71%) (Table 1).

Table 1
Patients characteristics (n = 21)

All CT scans were considered of diagnostic quality and all vessel segments > 2 mm were considered for analysis (there were no unevaluable segments). A total of 72 plaques were identified by IVUS (plaque burden > 40%) with 95 remaining coronary segments without plaque or with mild atherosclerotic disease (plaque burden ≤ 40%).

At patient and vessel levels, the measurements of average lumen area and minimal lumen area were highly correlated between IVUS and 64-slice MDCT (correlation coefficient: patient level 0.75 and 0.78, vessel level 0.91 and 0.93, p < 0.01 for all comparisons) (Table 2). At the patient level there was a moderate correlation between minimal lumen diameter measured by IVUS and by 64-slice MDCT (correlation coefficient: 0.64, p = 0.002) (Table 2). At the segment level, the correlation between 64-slice MDCT and IVUS was similar in segments with or without a plaque burden > 40% (Table 2). However, the average absolute difference in measurements tended to be larger among coronary segments without atherosclerotic plaque (Table 2 and Figure 3). Overall, 64-slice MDCT underestimated the lumen size with a relatively wide dispersion of the differences (Table 2 and Figure 2). The correlation coefficients for repeated IVUS and 64-slice MDCT lumen area tracings were 0.90 and 0.91, respectively Table 3 shows the diagnostic performance of 64-slice MDCT for detecting luminal areas < 4.0 mm2 and lumen diameters < 1.8 mm by IVUS, cutoff values previously shown to be markers of flow-restrictive coronary lesions3Briguori C, Anzuini A, Airoldi F, Gimelli G, Nishida T, Adamian M, et al. Intravascular ultrasound criteria for the assessment of the functional significance of intermediate coronary artery stenoses and comparison with fractional flow reserve. Am J Cardiol. 2001;87(2):136-41.. Interestingly, the accuracy to detect these IVUS luminal cutoffs was not notably affected by the presence or absence of atherosclerotic plaques (Table 3).

Table 2
Quantitative coronary lumen parameters by IVUS and by 64-slice MDCT at patient, vessel, and segment levels
Table 3
64-slice MDCT sensitivity, specificity, negative predictive value, positive predictive value, and global predictive accuracy * * Global predictive accuracy calculated as the area under the receiver operating characteristic curve (C-index) for the detection of luminal area < 4.0 mm2 and luminal diameter < 1.8 mm by IVUS IVUS parameters previously shown to be associated with flow-limiting lesions.(3)
Figure 2
Bland-Altman graphs for 64-slice MDCT and IVUS measurements at patient and vessel levels. Mean lumen area (A), minimal lumen area (B), and minimal lumen diameter (C) in per-patient (left) and per-vessel analyses (right). The solid lines represent the mean difference in measurements, and the traced lines represent the upper and lower 95% confidence intervals for the difference in measurements.
Figure 3
Bland-Altman graphs for 64-slice MDCT and IVUS measurements at segment level. Mean lumen area (A), minimal lumen area (B), and minimal lumen diameter (C) for segments without significant plaques (IVUS plaque burden ≤ 40%; left) and for segments with atherosclerotic plaque (IVUS plaque burden > 40%; right). The solid lines represent the mean difference in measurements, and the traced lines represent the upper and lower 95% confidence intervals for the diference in measurements.

Discussion

Currently, the primary clinical role of 64-slice MDCT is the evaluation of coronary luminal obstructions. This study reports that the assessment of coronary lumen by 64-slice MDCT shows excellent correlation with IVUS, with a good overall diagnostic ability to quantify coronary luminal dimensions regardless of the presence or absence of underlying atherosclerotic plaques.

Al though signi f icant ly cor related wi th IVUS, the measurements by 64-slice MDCT slightly underestimate the lumen area and the minimal lumen diameter with a relatively wide margin of error. In a previous study, Cademartiri et al. showed that lumen attenuation measured by CTA significantly affected the measured plaque attenuation1212 Cademartiri F, Mollet NR, Runza G, Bruining N, Hamers R, Somers P, et al. Influence of intracoronary attenuation on coronary plaque measurements using multislice computed tomography: observations in an ex vivo model of coronary computed tomography angiography. Eur Radiol. 2005;15(7):1426-31.. The higher the lumen attenuation, the higher the plaque attenuation is. In contrast, calcium attenuation and surrounding fat attenuation were not significantly affected. On the basis of this finding, a semiautomatic tool for tracing coronary lumen and plaque borders based on aortic lumen attenuation could underestimate the luminal area of segments with plaque and reduce the specificity of CTA to identify non-significantly diseased segments by IVUS. Nevertheless, 64-slice MDCT showed a good global predictive accuracy for the detection of an IVUS luminal area < 4.0 mm2 and an IVUS minimal lumen diameter < 1.8 mm, parameters previously shown to be markers of flow-limiting coronary lesions3Briguori C, Anzuini A, Airoldi F, Gimelli G, Nishida T, Adamian M, et al. Intravascular ultrasound criteria for the assessment of the functional significance of intermediate coronary artery stenoses and comparison with fractional flow reserve. Am J Cardiol. 2001;87(2):136-41.. In particular, 64-slice MDCT showed excellent sensitivity and negative predictive value for the identification of patients or vessels with the above characteristics at the expense of poor specificity and positive predictive value. These findings reinforce the primary clinical role of 64-slice MDCT for screening and ruling out, rather than ruling in, coronary luminal obstruction.

Our findings were obtained in the context of a blinded whole-artery 64-slice MDCT assessment, with no previous indication for focusing on a specific plaque-containing segment. It is important to highlight that this analytic approach emulates the way 64-slice MDCT is interpreted in clinical practice. The fact that the diagnostic performance of 64-slice MDCT for lumen assessment was not substantially different for segments with or without plaques when using IVUS is reassuring and strengthens the value of 64-slice MDCT as a clinical tool. It must be emphasized that the assessment of segments with minimal coronary disease is of clinical relevance, because of the potential impact on future disease progression and acute events and because of the importance of analyzing segments beyond the stenosis site when planning the therapeutic strategy.

Limitations

It must be emphasized that the present study is mainly a methodological analysis with limited immediate clinical application. The main limitation of our study is the relatively small sample size included in the analysis. The patient population analyzed had a relatively high risk profile, which may limit our ability to extrapolate the results of this study to other clinical scenarios. Exclusion of patients with coronary calcium score > 600 could have overestimated the correlation between IVUS and CTA for measurements of coronary lumen area. In a previous study, the presence of calcification was found to be independently correlated to coronary CTA inaccuracy in evaluating coronary lumen area1313 Kruk M, Noll D, Achenbach S, Mintz GS, Pregowsky J, Kaczmarska E, et al. Impact of coronary artery calcium characteristics on accuracy of CT angiography. JACC Cardiovasc Imaging. 2014;7(1):49-58.. Each patient contributed with multiple measurements; thus, affecting data independence. Nonetheless, such limitation is a drawback potentially found in every study aiming to evaluate the imaging of coronary atherosclerotic disease.

Conclusion

This study demonstrates that 64-slice MDCT has a good correlation with IVUS for the evaluation of the coronary artery lumen in segments with significant plaques as well as in segments without plaques. Our data suggest that 64-slice MDCT can be used by the clinician to comprehensively define coronary artery anatomy in patients with coronary artery disease, thereby improving clinical risk stratification and facilitating effective clinical care.

Funding Sources: Pedro A. Lemos has received research grants from The State of São Paulo Research Foundation (FAPESP), Sao Paulo, Brazil, and from the National Council for Scientific and Technological Development (CNPq), Brazil. João L. A. Falcão has received a Doctorate Scholarship from FAPESP.

  • Sources of Funding
    This study was funded by Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP).
  • Study Association
    This article is part of the doctoral thesis of João Luiz de Alencar Araripe Falcão by Universidade de São Paulo.

References

  • 1
    Meijer AB, O YL, Geleijns J, Kroft LJ. Meta-analysis of 40- and 64-MDCT angiography for assessing coronary artery stenosis. AJR Am J Roentgenol. 2008;191(6):1667-75.
  • 2
    Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I, et al. Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med. 2008;359(22):2324-36.
  • 3
    Briguori C, Anzuini A, Airoldi F, Gimelli G, Nishida T, Adamian M, et al. Intravascular ultrasound criteria for the assessment of the functional significance of intermediate coronary artery stenoses and comparison with fractional flow reserve. Am J Cardiol. 2001;87(2):136-41.
  • 4
    Abizaid AS, Mintz GS, Mehran R, Abizaid A, Lansky AJ, Pichard AD, et al. Long-term follow-up after percutaneous transluminal coronary angioplasty was not performed based on intravascular ultrasound findings: importance of lumen dimensions. Circulation. 1999;100(3):256-61.
  • 5
    Springer I, Dewey M. Comparison of multislice computed tomography with intravascular ultrasound for detection and characterization of coronary artery plaques: a systematic review. Eur J Radiol. 2009;71(2):275-82.
  • 6
    Joshi SB, Okabe T, Roswell RO, Weissman G, Lopez CF, Lindsay J, et al. Accuracy of computed tomographic angiography for stenosis quantification using quantitative coronary angiography or intravascular ultrasound as the gold standard. Am J Cardiol. 2009;104(8):1047-51.
  • 7
    Fischer C, Hulten E, Belur P, Smith R, Voros S, Villines TC. Coronary CT angiography versus intravascular ultrasound for estimation of coronary stenosis and atherosclerotic plaque burden: a meta-analysis. J Cardiovasc Comput Tomogr. 2013;7(4):256-66.
  • 8
    Decramer I, Vanhoenacker PK, Sarno G, Van Hoe L, Bladt O, Wijns W, et al. Effects of sublingual nitroglycerin on coronary lumen diameter and number of visualized septal branches on 64-MDCT angiography. AJR Am J Roentgenol. 2008;190(1):219-25.
  • 9
    Miller JM, Dewey M, Vavere AL, Rochitte CE, Niinuma H, Arbab-Zadeh A, et al. Coronary CT angiography using 64 detector rows: methods and design of the multi-centre trial CORE-64. Eur Radiol. 2009;19(4):816-28.
  • 10
    Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10.
  • 11
    Bland JM, Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995;346(8982):1085-7.
  • 12
    Cademartiri F, Mollet NR, Runza G, Bruining N, Hamers R, Somers P, et al. Influence of intracoronary attenuation on coronary plaque measurements using multislice computed tomography: observations in an ex vivo model of coronary computed tomography angiography. Eur Radiol. 2005;15(7):1426-31.
  • 13
    Kruk M, Noll D, Achenbach S, Mintz GS, Pregowsky J, Kaczmarska E, et al. Impact of coronary artery calcium characteristics on accuracy of CT angiography. JACC Cardiovasc Imaging. 2014;7(1):49-58.

Publication Dates

  • Publication in this collection
    27 Jan 2015
  • Date of issue
    Apr 2015

History

  • Received
    03 June 2014
  • Reviewed
    13 June 2014
  • Accepted
    13 Oct 2014
Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br