Acessibilidade / Reportar erro

Long-Term Follow-Up of Patients after Percutaneous Coronary Intervention with Everolimus-Eluting Bioresorbable Vascular Scaffold

Abstract

Background:

Bioresorbable vascular scaffolds (BVS) were developed to improve the long-term results of percutaneous coronary intervention, restoring vasomotion.

Objectives:

To report very late follow-up of everolimus-eluting Absorb BVS (Abbott Vascular, Santa Clara, USA) in our center.

Methods:

Observational retrospective study, in a single Brazilian center, from August 2011 to October 2013, including 49 patients submitted to Absorb BVS implantation. Safety and efficacy outcomes were analyzed in the in-hospital and very late follow-up phases (> 2 years).

Results:

All 49 patients underwent a minimum follow-up of 2.5 years and a maximum of 4.6 years. Mean age was 56.8 ± 7.6 years, 71.4% of the patients were men, and 26.5% were diabetic. Regarding clinical presentation, the majority (94%) had stable angina or silent ischemia. Device success was achieved in 100% of cases with 96% overall procedure success rate. Major adverse cardiovascular events rate was 4% at 30 days, 8.2% at 1 year, and 12.2% at 2 years, and there were no more events until 4.6 years. There were 2 cases of thrombosis (1 subacute and 1 late).

Conclusions:

In this preliminary analysis, Absorb BVS showed to be a safe and effective device in the very late follow-up. Establishing the efficacy and safety profiles of these devices in more complex scenarios is necessary.

Keywords:
Percutaneous Coronary Intervention; Absorbable Implants / utilization; Everolimus; Coronary Artery Disease; Clinical Evolution

Resumo

Fundamento:

Os suportes vasculares bioabsorvíveis (SVB) foram desenvolvidos com o intuito de melhorar os resultados da intervenção coronária percutânea a longo prazo, restabelecendo-se a vasomotricidade.

Objetivos:

Reportar o seguimento muito tardio do implante do SVB eluidor de everolimus Absorb® (Abbot Vascular, Santa Clara, EUA) em nosso centro.

Métodos:

Estudo observacional, retrospectivo, em um único centro brasileiro, que incluiu 49 pacientes submetidos ao implante do SVB Absorb® entre agosto/2011 e outubro/2013. Foram analisados os desfechos de segurança e eficácia na fase hospitalar e bastante tardia (> 2 anos).

Resultados:

Todos os 49 pacientes completaram um seguimento mínimo de 2,5 anos, sendo o máximo de 4,6 anos. A média de idade foi 56,8 ± 7,6 anos, sendo 71,4% da população estudada do sexo masculino e 26,5% composta por diabéticos. Considerando a apresentação clínica, a grande maioria (94%) tinha angina estável ou isquemia silenciosa. Obteve-se sucesso do dispositivo em 100% dos casos e do procedimento, em 96%. A taxa de eventos cardiovasculares maiores foi de 4% aos 30 dias, de 8,2% em 1 ano, e de 12,2% em 2 anos, sem mais eventos até 4,6 anos. Houve 2 casos de trombose (1 subaguda e 1 tardia) até o último seguimento.

Conclusões:

Nesta análise preliminar, o uso do SVB Absorb® mostrou-se seguro e eficaz no seguimento bastante tardio. Faz-se necessário estabelecer o perfil de eficácia e segurança destes dispositivos em cenários de maior complexidade.

Palavras-chave:
Intervenção Coronária Percutânea; Implantes Absorvíveis / utilização; Doença Arterial Coronariana; Everolimo; Evolução Clínica

Introduction

In the era of drug-eluting stents (DES), percutaneous coronary intervention (PCI) significantly improved clinical outcomes, with a reduction in excessive neointimal proliferation by adding antiproliferative agents. The permanent presence of intracoronary metal devices and long-lasting polymers, however, can delay natural vascular healing, resulting in constant inflammatory response and unfavorable clinical outcomes.11 Daemen J, Wenaweser P, Tsuchida K, Abrecht L, Vaina S, Morger C, et al. Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two-institutional cohort study. Lancet. 2007;369(9562):667-78.

2 Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, et al. Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation. 2007;115(18):2435-41.
-33 Kotani J, Awata M, Nanto S, Uematsu M, Oshima F, Minamiguchi H, et al. Incomplete neointimal coverage of sirolimus-eluting stents: angioscopic findings. J Am Coll Cardiol. 2006;47(10):2108-11.

Bioresorbable vascular scaffolds (BVS), thus, appeared as an alternative to those permanent prostheses: they can maintain the mechanical properties of metallic DES in the first months, and then be completely reabsorbed, eliminating possible adverse effects of their presence in the coronary arteries.

Recently developed, the Absorb BVS (Abbot Vascular, Santa Clara, USA) is aimed at meeting the above-mentioned criteria, maintaining the efficacy profile of last-generation metallic DES. The Absorb BVS was assessed in humans for the first time in the ABSORB clinical trial (cohorts A and B), with promising results.44 Ormiston JA, Serruys PW, Regar E, Dudek D, Thuesen L, Webster MW, et al. A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial. Lancet. 2008;371(9616):899-907.

5 Serruys PW, Onuma Y, Ormiston JA, de Bruyne B, Regar E, Dudek D, et al. Evaluation of the second generation of a bioresorbable everolimus drug-eluting vascular scaffold for treatment of de novo coronary artery stenosis: six-month clinical and imaging outcomes. Circulation. 2010;122(22):2301-12.
-66 Onuma Y, Serruys PW, Gomez J, de Bruyne B, Dudek D, Thuesen L, et al. Comparison of in vivo acute stent recoil between the bioresorbable everolimus-eluting coronary scaffolds (revision 1.0 and 1.1) and the metallic everolimus-eluting stent. Catheter Cardiovasc Interv. 2011;78(1):3-12.

Based on those results, the ABSORB EXTEND study, a multicenter single-arm study, has been conducted in 56 centers of several countries, aimed initially at including around 800 patients and at assessing the safety and performance of the Absorb BVS in a larger and more diversified population, as compared to that of initial studies, with more complex lesions.77 Abizaid A, Ribamar Costa J Jr, Bartorelli AL, Whitbourn R, van Geuns RJ, Chevalier B, et al; ABSORB EXTEND investigators. The ABSORB EXTEND study: preliminary report of the twelve-month clinical outcomes in the first 512 patients enrolled. EuroIntervention. 2015;10(12):1396-401.

The present analysis reports the very late follow-up (>2 years) of the first patients submitted to Absorb BVS implantation in Brazil, as part of the ABSORB EXTEND multicenter registry.

Methods

Study design and target population

The present study included the patients treated with Absorb BVS between August/2011 and October/2013, in a tertiary cardiological center in Brazil, who were included in the international multicenter single-arm study, ABSORB EXTEND study, as part of the first 512 patients recruited in 56 centers of Europe, Australia, New Zealand, Japan, Hong Kong, Malaysia, Singapore, Latin America and Canada.

It is worth noting that the participation in the EXTEND registry marks the beginning of the Brazilian experience with that new technology. This study was financed by Abbot Vascular, Santa Clara, USA. The Ethics Committee on Research of our institution approved the study protocol, and all patients provided written informed consent.

Inclusion/exclusion criteria

Patients with the following characteristics were included in the study: age ≥ 18 years; evidence of myocardial ischemia, such as stable or unstable angina; silent ischemia; and functional test or transient alterations on 12-lead electrocardiography compatible with ischemia.

The patients had up to two de novo lesions that could be percutaneously treated, each located in separate native epicardial vessels. The lesions should be in a native coronary vessel, whose target-vessel diameter was ≥ 2.0 mm and ≤ 3.3 mm, and whose target-lesion extension was ≤ 28 mm, both assessed by use of on-line quantitative coronary angiography (QCA) or intracoronary ultrasound (ICUS). The target lesions should be in an artery or branch of significant caliber and stenosis should be visually estimated ≥ 50% and < 100%, with TIMI (Thrombolysis in Myocardial Infarction) flow ≥ 1. Previous PCI in a non-target vessel was allowed, if performed at least 30 days after the index procedure or planned for 6 months after the index procedure; PCI in target-vessel lesions were allowed if performed at least 6 months before the index procedure or planned to 6 months after the index procedure.

Patients with the following characteristics were excluded from the study: previous acute myocardial infarction (AMI) up to 3 days before the index procedure; arrhythmias with hemodynamic instability; left ventricular ejection fraction < 30%; chronic renal failure; left main coronary artery lesions; lesions in arterial or venous grafts; in-stent restenosis; bifurcation lesions; total occlusion (TIMI flow 0); and significant calcification or excessive tortuosity.

Device

We used the Absorb BVS, the same device used in cohort B of the ABSORB study.88 Oberhauser JP, Hossainy S, Rapoza RJ. Design principles and performance of bioresorbable polymeric vascular scaffolds. EuroIntervention. 2009;5 Suppl F:F15-22.,99 Okamura T, Garg S, Gutierrez-Chico JL, Shin ES, Onuma Y, Garcia-Garcia HM, et al. In vivo evaluation of stent strut distribution patterns in the bioabsorbable everolimus-eluting device: an OCT ad hoc analysis of the revision 1.0 and revision 1.1 stent design in the ABSORB clinical trial. EuroIntervention. 2010;5(8):932-8. The Absorb platform is composed by the polymer poly-L-lactic acid (PLLA), the antiproliferative drug everolimus (Novartis Pharmaceuticals Corporation, Basel, Switzerland), and a matrix of poly-D, L-lactic acid (PDLLA), at a 1:1 ratio, forming an amorphous matrix covered with 100µ everolimus/cm2. Both PLLA and PDLLA are metabolized and resorbed in the body. PDLLA is expected to be completely resorbed by the arteries in 9 months, while PLLA, in approximately 36 months. During resorption, the chains with PLLA and PDLLA are hydrolyzed, the last product of that reaction being lactic acid, biologically metabolized via Krebs cycle.55 Serruys PW, Onuma Y, Ormiston JA, de Bruyne B, Regar E, Dudek D, et al. Evaluation of the second generation of a bioresorbable everolimus drug-eluting vascular scaffold for treatment of de novo coronary artery stenosis: six-month clinical and imaging outcomes. Circulation. 2010;122(22):2301-12.

At the time the patients were included in this study, Absorb devices were available only in two diameters (2.5 and 3.0 mm) and two lengths (18 and 28 mm).

Procedure

All procedures were performed electively, in accordance with current guidelines. The lesions were treated with the usual intervention techniques, which required pre-dilatation with a shorter balloon, with a diameter 0.5 mm smaller than that of the device used. The Absorb's deployment pressure should never exceed the manufacturer's maximum nominal reference value.

Post-dilatation was subjected to need and operator's assessment. It was performed with non-compliant balloons, within the expansion limits of the BVS (post-dilatation balloons should not exceed 0.5 mm the nominal diameter of the implanted BVS).

Preprocedural dual antiplatelet therapy comprised an attack dose of acetylsalicylic acid (300 mg) and clopidogrel (300 mg), at least 24 hours before the procedure, or 600 mg if < 24 hours. After the intervention, acetylsalicylic acid was prescribed indefinitely and clopidogrel (75 mg/day) was maintained for at least 6 months.

Quantitative coronary angiography and intracoronary ultrasound

The recommended limits of the target-vessel's diameter were established by use of on-line QCA on distal and proximal maximal luminal diameter (Dmax), the Dmax being assessed in the distal and proximal portions of the target segment to be coated with the BVS, or by use of ICUS. Overlapping of the BVS was allowed for lesions > 22 mm and ≤ 28 mm, with a recommended limit of 1-4 mm.

Follow-up

Clinical follow-up, via outpatient clinic consultation or telephone, was mandatory at day 30 (± 7 days), 6 months (± 14 days) and 1, 2 and 3 years (± 28 days), following the ABSORB EXTEND study protocol. After that, routine return visits were recommended. Minimum follow-up was 2.5 years. All adverse events and symptoms, such as angina, details of subsequent PCIs, as well as medication use and changes, were collected in the period. The patients did not undergo a new protocol coronary angiography, being only reassessed in case of clinical indication due to symptoms or evidence of ischemia.

Study outcomes

All outcomes were adjudicated by an independent clinical events committee abiding by the protocol definitions based on the Academic Research Consortium (ARC).1010 Cutlip DE, Windecker S, Mehran R, Boam A, Cohen DJ, van Es GA, et al; Academic Research Consortium. Clinical end points in coronary stent trials: a case for standardized definitions. Circulation. 2007;115(17):2344-51.

Clinical success comprised device's success (based on the target lesion) and procedural success (assessed in each patient). In addition, it included scaffold thrombosis (ST), cardiovascular death, AMI (either related or not to the target vessel) and revascularization rate (target-lesion or target-vessel revascularization, or total revascularization). In addition, combined outcome rates, considering ischemia-driven (ID) major adverse cardiovascular events (MACE) (ID-MACE), ID target-vessel failure (ID-TVF), ID target-vessel revascularization and ID target-lesion revascularization (ID-TLR), were assessed.

The device's success was defined as successful device's deployment in the target lesion and successful withdrawal of the BVS delivery system, with residual stenosis < 50% assessed via QCA (or visual estimate, when QCA was unavailable).

The procedure's success was defined as device's success with no ID-MACE during hospitalization for up to 7 days after the procedure. If there were two lesions, both should meet the success criteria.

Cardiac death was defined as any death of cardiac cause, such as AMI, low output syndrome, and lethal arrhythmia. Unattended death and death of unknown cause were classified as cardiac death. This included the deaths related to the procedure.

The classification of AMI and the diagnostic criteria were defined based on the pre-established protocol:1111 Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD, et al. Third universal definition of myocardial infarction. Eur Heart J. 2012;33(20):2551-67. Q-wave AMI, characterized by the development of a new pathological Q wave; Non-Q-wave AMI, defined as elevation of creatine phosphokinase (CK) levels ≥ 2 times the upper limit of normality with concomitant increase in CK-MB in the absence of new pathological Q waves.

The revascularization events were defined as follows:

  • - ID-MACE: composed of cardiac death, Q-wave/non-Q-wave AMI, target-lesion revascularization via PCI or coronary artery bypass graft (CABG);

  • - ID-TVF: composed of cardiac death, AMI with and without Q wave, target-vessel revascularization via PCI or CABG;

  • - ID-TLR: defined as any new PCI in the target lesion, either percutaneous or CABG in the target vessel with positive functional ischemia, ischemic symptoms or angiography evidencing lumen diameter at stenosis ≥ 50% by use of QCA, or revascularization of a target lesion with diameter ≥ 70% by use of QCA without ischemic symptoms or functional test.

Scaffold thrombosis was categorized as acute (< 1 day), subacute (1-30 days), late (> 30 days and < 1 year) and very late (>1 year), and defined based on the ARC guidelines as follows:1010 Cutlip DE, Windecker S, Mehran R, Boam A, Cohen DJ, van Es GA, et al; Academic Research Consortium. Clinical end points in coronary stent trials: a case for standardized definitions. Circulation. 2007;115(17):2344-51. definite (acute coronary syndrome and pathological or angiographic confirmation of the BVS thrombosis) or likely (death of unknown cause ≤ 30 days or AMI related to the target vessel without angiographic confirmation).

Statistical analysis

Continuous variables with normal distribution were expressed as mean and standard deviation. Categorical variables were expressed as absolute numbers and percentages. The SPSS program (Statistical Package for the Social Science, Chicago, USA), version 19, was used for data tabulation.

Results

The present study represents the analysis of 49 patients (53 lesions/57 BVS) included in the ABSORB EXTEND study and submitted to PCI with Absorb BVS implantation, at a Brazilian tertiary cardiology center. Clinical 1-year follow-up was obtained in 100% of the cases, while 2-year follow-up, in 97.9% of the cases. Mean follow-up was 3.59 ± 0.72 years (2.5-4.6 years).

Table 1 shows the demographic and clinical characteristics of the population studied. The patients' mean age was 56.8 ± 7.6 years, most of them were men (71.4%), and 26.5% of the population studied had diabetes. In addition, only 6.1% of the patients had more than one target lesion, and 6.1% of the patients presented with clinical findings of acute coronary syndrome (55.1%, stable angina; 38.8%, silent ischemia). Neither ST-segment elevation AMI nor recent AMI occurred.

Table 1
Demographic and clinical characteristics

Table 2 illustrates the angiographic characteristics of the lesions treated and the procedure. Most lesions treated were in the anterior descending coronary artery (46.9%), followed by the right coronary (32.6%) and circumflex (26.5%) arteries. The mean grade of stenosis was 76.0 ± 8.5%. By use of on-line QCA or ICUS, the lesions had a mean diameter of 2.92 ± 0.28 mm (range, 2.2-3.5 mm) and a mean extension of 15.98 ± 5.55 mm (range, 7-28 mm).

Table 2
Angiographic and procedural characteristics

The device's clinical success was 100%, while the procedure's clinical success was 96% (47/49) in the 49 patients submitted to PCI with Absorb implantation. Two patients (4%) had periprocedural AMI while hospitalized.

Table 3 shows the clinical outcome data at 30 days and 1 year, and the very late follow-up of the patients. At 30 days, the MACE rate was 4% because of the periprocedural AMI rate. Cardiac mortality, target-vessel revascularization and non-target-vessel revascularization was 0%.

Table 3
Clinical outcomes in early, middle-term and long-term follow-up

At 1 year, the MACE rate was 8.2%, because of cardiac death and need for revascularization of the target vessel (but not of the target lesion) via PCI in one patient, the global AMI rate being maintained as 0%. At 2 years, the MACE rate was 12.2% because of a non-Q AMI event related to the target vessel and one in-stent restenosis event requiring target-lesion revascularization. From 2 years of follow-up till now, there were neither cardiovascular nor cerebrovascular events, and the accumulated MACE rate remained as 12.2% among the patients followed up till almost 5 years.

Regarding device's thrombosis and based on the ARC criteria, the findings were as follows: one case of definite subacute thrombosis 13 days after implantation, need for urgent surgical vascular procedure and a new angiographic study with an unsuccessful recanalization attempt; and one case of likely late thrombosis 34 days after PCI (sudden death episode). After one year, there was no additional case of thrombosis.

Discussion

In this initial experience, at a single center, the Absorb BVS performed well in the long run, with a very low target-vessel failure rate.

In the past 3 years, more than 60,000 patients were treated with Absorb BVS worldwide, despite the lack of a robust randomized study comparing it with contemporary drug-eluting stents.1212 Abizaid A, Costa JR Jr, Gibson CM. Bioresorbable vascular scaffolds in daily clinical practice: is the essential really invisible to the eyes? J Am Coll Cardiol. 2015;65(8):802-4.

The assessment of Absorb BVS has begun with the ABSORB cohort studies A and B and clinical trial.1313 Dudek D, Onuma Y, Ormiston JA, Thuesen L, Miquel-Hebert K, Serruys PW. Four-year clinical follow-up of the ABSORB everolimus-eluting bioresorbable vascular scaffold in patients with de novo coronary artery disease: the ABSORB trial. EuroIntervention. 2012;7(9):1060-1.,1414 Onuma Y, Dudek D, Thuesen L, Webster M, Nieman K, Garcia-Garcia HM, et al. Five-year clinical and functional multislice computed tomography angiographic results after coronary implantation of the fully resorbable polymeric everolimus-eluting scaffold in patients with de novo coronary artery disease: the ABSORB cohort A trial. JACC Cardiovasc Interv. 2013;6(10):999-1009. After changes in the device's design and structure, the device's current version began to be used in cohort B, involving 101 patients, and showed a 1-year late lumen loss of 0.27 mm, the 2-year follow-up evidencing a MACE rate of 6.8% and no device's thrombosis.55 Serruys PW, Onuma Y, Ormiston JA, de Bruyne B, Regar E, Dudek D, et al. Evaluation of the second generation of a bioresorbable everolimus drug-eluting vascular scaffold for treatment of de novo coronary artery stenosis: six-month clinical and imaging outcomes. Circulation. 2010;122(22):2301-12.,1515 Ormiston JA, Serruys PW, Onuma Y, van Geuns RJ, de Bruyne B, Dudek D, et al. First serial assessment at 6 months and 2 years of the second generation of absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study. Circ Cardiovasc Interv. 2012;5(5):620-32. At 5 years, the Absorb's structures were no longer discernible on optical tomography or ICUS, the MACE rate being 11%, with no evidence of thrombosis.1616 Serruys PW, Ormiston J, van Geuns RJ, de Bruyne B, Dudek D, Christiansen E, et al. A polylactide bioresorbable scaffold eluting everolimus for treatment of coronary stenosis: 5-year follow-up. J Am Coll Cardiol 2016;67(7):766-76.

The initial analysis of the first 512 patients recruited in the ABSORB EXTEND registry, in a 1-year follow-up, confirms the efficacy of Absorb BVS, with very low incidence of ID-MACE (4.8%), ID-TVF (4.4%) and device's thrombosis (0.8%).77 Abizaid A, Ribamar Costa J Jr, Bartorelli AL, Whitbourn R, van Geuns RJ, Chevalier B, et al; ABSORB EXTEND investigators. The ABSORB EXTEND study: preliminary report of the twelve-month clinical outcomes in the first 512 patients enrolled. EuroIntervention. 2015;10(12):1396-401. At 3 years, with 250 patients, the MACE rate was 9.3%, the ID-TVF, 10.1%, and thrombosis, 1.2%.1717 Smits PC. ABSORB EXTEND: BVS Safe Out to 3 Years. In: Transcatheter Cardiovascular Therapeutics, Washington, DC, September 13-17; 2014. [Internet]. [Accessed in 2016 Dec 15]. Available from: http://www.practiceupdate.com/conference/145/2
http://www.practiceupdate.com/conference...

In our study, the MACE rate in a very late follow-up was equivalent, with no event after 2 years, corroborating the theory that the major benefit of the BVS occurs in the long run, with both low rate of events and the likelihood of new revascularization and BVS assessment by use of non-invasive imaging techniques.

Regarding the comparison with the results of drug-eluting metal stents, no long-term follow-up study has been published. In a recent meta-analysis encompassing the last four randomized studies comparing Absorb BVS with the everolimus-eluting metal stent Xience® (Abbot Vascular, Santa Clara, USA), ABSORB II,1818 Serruys PW, Chevalier B, Dudek D, Cequier A, Carrie D, Iniguez A, et al. A bioresorbable everolimus-eluting scaffold versus a metallic everolimus-eluting stent for ischaemic heart disease caused by de-novo native coronary artery lesions (ABSORB II): an interim 1-year analysis of clinical and procedural secondary outcomes from a randomised controlled trial. Lancet. 2015;385(9962):43-54. ABSORB III,1919 Ellis SG, Kereiakes DJ, Metzger DC, Caputo RP, Rizik DG, Teirstein PS, et al; ABSORB III Investigators. Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med. 2015;373(20):1905-15. ABSORB Japan2020 Kimura T, Kozuma K, Tanabe K, Nakamura S, Yamane M, Muramatsu T, et al. A randomized trial evaluating everolimus-eluting Absorb bioresorbable scaffolds vs. everolimus-eluting metallic stents in patients with coronary artery disease: ABSORB Japan. Eur Heart J. 2015;36(47):3332-42. and ABSORB China,2121 Gao R, Yang Y, Han Y, Huo Y, Chen J, Yu B, et al. Bioresorbable vascular scaffolds versus metallic stents in patients with coronary artery disease: ABSORB China trial. J Am Coll Cardiol. 2015;66(21):2298-309. the relative combined outcomes rates at the end of the first year did not differ between the Absorb and Xience groups (11.9% vs. 10.6%, respectively, p=0.38). Target-vessel AMI was significantly higher in the Absorb group as compared to the Xience group (5.1% vs. 3.3%, respectively, p=0.04), due partially to the higher rate of periprocedural AMI and partially to the higher rate of ST (definite or likely) in the Absorb group (1.3% vs. 0.6%, respectively, p=0.08). The results were similar after multivariate analysis adjusted to baseline characteristics, and were consistent even in the analysis of most subgroups.2222 Stone GW, Gao R, Kimura T, Kereiakes DJ, Ellis SG, Onuma Y, et al. 1-year outcomes with the Absorb bioresorbable scaffold in patients with coronary artery disease: a patient-level, pooled meta-analysis. Lancet. 2016;387(10025):1277-89.

The EVERBIO-II Trial (Comparison of Everolimus- and Biolimus-Eluting Stents With Everolimus-Eluting Bioresorbable Vascular Scaffold Stents II), a single-center study, involved 240 patients randomized at the 1:1:1 proportion for everolimus-eluting stent, biolimus-eluting stent or Absorb BVS. In a 2-year follow-up, the MACE rate related to the device was 13% in the everolimus- and biolimus-eluting stent groups vs. 21% in the Absorb group (p=0.12), and the related MACE rate was 32% vs. 35%, respectively (p=0.67), with only one ST event in the Absorb group and none in the DES groups (p=0.33). Thus, once again DES were considered non-inferior to BVS.2323 Puricel S, Arroyo D, Corpataux N, Baeriswyl G, Lehmann S, Kallinikou Z, et al. Comparison of everolimus- and biolimus-eluting coronary stents with everolimus-eluting bioresorbable vascular scaffolds. J Am Coll Cardiol. 2015;65(8):791-801.,2424 Puricel S, Arroyo D, Corpatausc N, Baeriswyl G, Lehmann S, Kallinikou Z, et al. Comparison of everolimus- and biolimus-eluting coronary stents with everolimus-eluting bioresorbable vascular scaffolds: 2-year outcomes of the EVERBIO II Trial. J Am Coll Cardiol. 2015;65(8):79-81.

Regarding other BVSs, the DESolve® NX (Elixir Medical Corporation, Sunnyvale, USA) was the only BVS with late follow-up and recently published results. At 2 years, that new device showed the following rates: MACE, 7.4%; isolate cardiac death, 2.5%; AMI, 0.8%; target-lesion revascularization, 4.1%; and target-lesion failure, 7.4%. In addition, the thrombosis rate was minimal (0.8%).2525 Abizaid A, Costa RA, Schofer J, Ormiston J, Maeng M, Witzenbichler B, et al. Serial multimodality imaging and 2-year clinical outcomes of the Novel DESolve Novolimus-Eluting Bioresorbable Coronary Scaffold System for the Treatment of Single De Novo Coronary Lesions. JACC Cardiovasc Interv. 2016;9(6):565-74.

Tamburino et al., using a complex statistical analysis, have assessed the database of the GHOST-EU Registry (Gauging coronary Healing with biOresorbable Scaffolding plaTforms in EUrope), with 1,189 patients treated with Absorb BVS in Europe and 5,034 patients treated with everolimus-eluting metal stent (Xience) of the XIENCE V Registry in the USA. After propensity score matching, 905 pairs of patients were identified with similar characteristics. Of the total of 1,810 patients, there was no difference between the Absorb and Xience groups concerning the risk of MACE within 1 year (5.8% vs. 7.6%, respectively, p=0.12). Cardiac death was less likely to occur in the Absorb group (0.7% vs. 1.9%, p=0.03) and there was a tendency towards reduction in AMI in the Absorb group as compared to the Xience group (2.4% vs. 4.0%, p=0.07). In addition, there was no difference in target-vessel revascularization (4.6% vs. 3.5%, p=0.22) and definite or likely thrombosis (1.8% vs. 1.1%) between the Absorb and Xience groups, respectively.2626 Tamburino C, Capranzano P, Gori T, Latib A, Lesiak M, Nef H, et al. 1-year outcomes of everolimus-eluting bioresorbable scaffolds versus everolimus-eluting stents: a propensity-matched comparison of the GHOST-EU and XIENCE V USA Registries. JACC Cardiovasc Interv. 2016;9(5):440-9. In most studies, the ST cases occurred in the immediate post-procedural period (<30 days), and cases after the sixth month were rare, as observed in the cohort reported.

Limitations

This was a retrospective and observational study, having, thus, obvious limitations. The sample was small, with low clinical and anatomical complexity, following the ABSORB EXTEND study protocol.

Conclusions

In this case series, Absorb BVS implantation was associated with a low incidence of adverse events, mainly in the very long-term follow-up (> 2 years). However, larger studies with a higher number of patients and more complex scenarios are necessary to confirm these preliminary observations.

  • Sources of Funding
    This study was funded by Abbot Vascular.
  • Study Association
    This study is not associated with any thesis or dissertation work.

References

  • 1
    Daemen J, Wenaweser P, Tsuchida K, Abrecht L, Vaina S, Morger C, et al. Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two-institutional cohort study. Lancet. 2007;369(9562):667-78.
  • 2
    Finn AV, Joner M, Nakazawa G, Kolodgie F, Newell J, John MC, et al. Pathological correlates of late drug-eluting stent thrombosis: strut coverage as a marker of endothelialization. Circulation. 2007;115(18):2435-41.
  • 3
    Kotani J, Awata M, Nanto S, Uematsu M, Oshima F, Minamiguchi H, et al. Incomplete neointimal coverage of sirolimus-eluting stents: angioscopic findings. J Am Coll Cardiol. 2006;47(10):2108-11.
  • 4
    Ormiston JA, Serruys PW, Regar E, Dudek D, Thuesen L, Webster MW, et al. A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial. Lancet. 2008;371(9616):899-907.
  • 5
    Serruys PW, Onuma Y, Ormiston JA, de Bruyne B, Regar E, Dudek D, et al. Evaluation of the second generation of a bioresorbable everolimus drug-eluting vascular scaffold for treatment of de novo coronary artery stenosis: six-month clinical and imaging outcomes. Circulation. 2010;122(22):2301-12.
  • 6
    Onuma Y, Serruys PW, Gomez J, de Bruyne B, Dudek D, Thuesen L, et al. Comparison of in vivo acute stent recoil between the bioresorbable everolimus-eluting coronary scaffolds (revision 1.0 and 1.1) and the metallic everolimus-eluting stent. Catheter Cardiovasc Interv. 2011;78(1):3-12.
  • 7
    Abizaid A, Ribamar Costa J Jr, Bartorelli AL, Whitbourn R, van Geuns RJ, Chevalier B, et al; ABSORB EXTEND investigators. The ABSORB EXTEND study: preliminary report of the twelve-month clinical outcomes in the first 512 patients enrolled. EuroIntervention. 2015;10(12):1396-401.
  • 8
    Oberhauser JP, Hossainy S, Rapoza RJ. Design principles and performance of bioresorbable polymeric vascular scaffolds. EuroIntervention. 2009;5 Suppl F:F15-22.
  • 9
    Okamura T, Garg S, Gutierrez-Chico JL, Shin ES, Onuma Y, Garcia-Garcia HM, et al. In vivo evaluation of stent strut distribution patterns in the bioabsorbable everolimus-eluting device: an OCT ad hoc analysis of the revision 1.0 and revision 1.1 stent design in the ABSORB clinical trial. EuroIntervention. 2010;5(8):932-8.
  • 10
    Cutlip DE, Windecker S, Mehran R, Boam A, Cohen DJ, van Es GA, et al; Academic Research Consortium. Clinical end points in coronary stent trials: a case for standardized definitions. Circulation. 2007;115(17):2344-51.
  • 11
    Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD, et al. Third universal definition of myocardial infarction. Eur Heart J. 2012;33(20):2551-67.
  • 12
    Abizaid A, Costa JR Jr, Gibson CM. Bioresorbable vascular scaffolds in daily clinical practice: is the essential really invisible to the eyes? J Am Coll Cardiol. 2015;65(8):802-4.
  • 13
    Dudek D, Onuma Y, Ormiston JA, Thuesen L, Miquel-Hebert K, Serruys PW. Four-year clinical follow-up of the ABSORB everolimus-eluting bioresorbable vascular scaffold in patients with de novo coronary artery disease: the ABSORB trial. EuroIntervention. 2012;7(9):1060-1.
  • 14
    Onuma Y, Dudek D, Thuesen L, Webster M, Nieman K, Garcia-Garcia HM, et al. Five-year clinical and functional multislice computed tomography angiographic results after coronary implantation of the fully resorbable polymeric everolimus-eluting scaffold in patients with de novo coronary artery disease: the ABSORB cohort A trial. JACC Cardiovasc Interv. 2013;6(10):999-1009.
  • 15
    Ormiston JA, Serruys PW, Onuma Y, van Geuns RJ, de Bruyne B, Dudek D, et al. First serial assessment at 6 months and 2 years of the second generation of absorb everolimus-eluting bioresorbable vascular scaffold: a multi-imaging modality study. Circ Cardiovasc Interv. 2012;5(5):620-32.
  • 16
    Serruys PW, Ormiston J, van Geuns RJ, de Bruyne B, Dudek D, Christiansen E, et al. A polylactide bioresorbable scaffold eluting everolimus for treatment of coronary stenosis: 5-year follow-up. J Am Coll Cardiol 2016;67(7):766-76.
  • 17
    Smits PC. ABSORB EXTEND: BVS Safe Out to 3 Years. In: Transcatheter Cardiovascular Therapeutics, Washington, DC, September 13-17; 2014. [Internet]. [Accessed in 2016 Dec 15]. Available from: http://www.practiceupdate.com/conference/145/2
    » http://www.practiceupdate.com/conference/145/2
  • 18
    Serruys PW, Chevalier B, Dudek D, Cequier A, Carrie D, Iniguez A, et al. A bioresorbable everolimus-eluting scaffold versus a metallic everolimus-eluting stent for ischaemic heart disease caused by de-novo native coronary artery lesions (ABSORB II): an interim 1-year analysis of clinical and procedural secondary outcomes from a randomised controlled trial. Lancet. 2015;385(9962):43-54.
  • 19
    Ellis SG, Kereiakes DJ, Metzger DC, Caputo RP, Rizik DG, Teirstein PS, et al; ABSORB III Investigators. Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med. 2015;373(20):1905-15.
  • 20
    Kimura T, Kozuma K, Tanabe K, Nakamura S, Yamane M, Muramatsu T, et al. A randomized trial evaluating everolimus-eluting Absorb bioresorbable scaffolds vs. everolimus-eluting metallic stents in patients with coronary artery disease: ABSORB Japan. Eur Heart J. 2015;36(47):3332-42.
  • 21
    Gao R, Yang Y, Han Y, Huo Y, Chen J, Yu B, et al. Bioresorbable vascular scaffolds versus metallic stents in patients with coronary artery disease: ABSORB China trial. J Am Coll Cardiol. 2015;66(21):2298-309.
  • 22
    Stone GW, Gao R, Kimura T, Kereiakes DJ, Ellis SG, Onuma Y, et al. 1-year outcomes with the Absorb bioresorbable scaffold in patients with coronary artery disease: a patient-level, pooled meta-analysis. Lancet. 2016;387(10025):1277-89.
  • 23
    Puricel S, Arroyo D, Corpataux N, Baeriswyl G, Lehmann S, Kallinikou Z, et al. Comparison of everolimus- and biolimus-eluting coronary stents with everolimus-eluting bioresorbable vascular scaffolds. J Am Coll Cardiol. 2015;65(8):791-801.
  • 24
    Puricel S, Arroyo D, Corpatausc N, Baeriswyl G, Lehmann S, Kallinikou Z, et al. Comparison of everolimus- and biolimus-eluting coronary stents with everolimus-eluting bioresorbable vascular scaffolds: 2-year outcomes of the EVERBIO II Trial. J Am Coll Cardiol. 2015;65(8):79-81.
  • 25
    Abizaid A, Costa RA, Schofer J, Ormiston J, Maeng M, Witzenbichler B, et al. Serial multimodality imaging and 2-year clinical outcomes of the Novel DESolve Novolimus-Eluting Bioresorbable Coronary Scaffold System for the Treatment of Single De Novo Coronary Lesions. JACC Cardiovasc Interv. 2016;9(6):565-74.
  • 26
    Tamburino C, Capranzano P, Gori T, Latib A, Lesiak M, Nef H, et al. 1-year outcomes of everolimus-eluting bioresorbable scaffolds versus everolimus-eluting stents: a propensity-matched comparison of the GHOST-EU and XIENCE V USA Registries. JACC Cardiovasc Interv. 2016;9(5):440-9.

Publication Dates

  • Publication in this collection
    09 Jan 2017
  • Date of issue
    Feb 2017

History

  • Received
    13 May 2016
  • Reviewed
    28 Sept 2016
  • Accepted
    24 Oct 2016
Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br