Reprocessing of Medical Products in Electrophysiology

Keywords
Recycling; Electrophysiologic Techniques / cardiac; Catheter Ablation; Cost Savings

Electrophysiological procedures use high-cost multipolar electrode catheters which can be reprocessed. The reuse thereof has been performed by electrophysiology services in Europe, United States, Latin America and also in our midst. In fact, prior studies have proved that there is an actual cost decrease11 Tessarolo F, Disertoric M, Caola I, Guarrera GM, Favaretti C, Nollo G. Health technology assessment on reprocessing single-use catheters for cardiac electrophysiology: results of a three-years study. Conf Proc IEEE Eng Med Biol Soc. 2007;2007:1758-61.,22 Medical Data Institute. International Institute for Health Economics [internet]. [Accessed in 2016 Aug 16]. Available from: http://www.md.institute.com/cms/ressorts/medizintechnologie/aufbereitung-medizinprodukte/Pressemitteilung-Gutachten-bestaetigt-Wiederaufbereitung-von-Einwegprodukten-senkt-Kosten-EN.pdf
http://www.md.institute.com/cms/ressorts...
and have also attested to the safety and efficacy of such practice,33 Avitall B, Khan M, Krum D, Jazayeri M, Hare J. Repeated use of ablation catheters: a prospective study. J Am Coll Cardiol. 1993;22(5):1367-72.

4 Dunnigan A, Roberts C, McNamara M, Benson DW Jr, Benditt DG. Success of reuse of cardiac electrode catheters. Am J Cardiol. 1987;60(10):807-10.

5 Mickelsen S, Mickelsen C, MacIndoe C, Jaramillo J, Bass S, West G, et al. Trends and patterns in electrophysiologic and ablation catheter reuse in the United States. Am J Cardiol. 2001;87(3):351-3.

6 O'Donoghue S, Platia EV. Reuse of pacing catheters: a survey of safety and efficacy. Pacing Clin Electrophysiol. 1988;11(9):1279-80.

7 Ross DL. Re-use of electrode catheters labeled as single use for clinical cardiac electrophysiological studies. Aust N Z J Med. 1997;26(5):632-5.

8 Blomstrom-Lundqvist, C. The safety of reusing ablation catheters with temperature control and the need for a validation protocol and guidelines for reprocessing. Pacing Clin Electrophysiol. 1998;21(12):2563-70.

9 Bank, Alan J, Berry James M, Wilson RF, Lester BR. Acceptance criteria for reprocessed AcuNav catheters: comparison between functionality testing and clinical image assessment. Ultrasound Med Biol. 2009;35(3):507-14.

10 Lester BR, Boser NP, Miller K, Schnapf M, Jacques BA, Sullivan BF, et al. Reprocessing and sterilization of single-use electrophysiological catheters: removal of organic carbon and protein surface residues. J AOAC Int. 2009;92(4):1165-73.

11 Druce, JD, Russell JS, Birch CJ, Yates LA, Harper RW, Smolich JJ. A decontamination and sterilization protocol employed during reuse of cardiac electrophysiology catheters inactivates human immunodeficiency virus. Infect Control Hosp Epidemiol. 2003;24(3):184-90.
-1212 Lindsay BD, Kutalek SP, Cannom DS, Hammill SC, Naccarelli GV; NASPE Task Force on Reprocessing of Electrophysiological Catheters. North American Society of Pacing and Electrophysiology. Reprocessing of electrophysiology catheters: clinical studies, regulations, and recommendations. A report of the NASPE Task Force on Reprocessing of Electrophysiological Catheters. North American Society of Pacing and Electrophysiology. Pacing Clin Electrophysiol. 2001;24(8 Pt1):1297-305 observing rates of complication and therapeutic results similar to the ones obtained with first-use electrophysiology devices. The growing concern with sustainability and no waste, associated with the efficacy and safety already demonstrated, increasingly stimulate the practice of reprocessing single-use medical devices throughout the world.

The American Society of Cardiac Arrhythmias issued a favorable opinion to the reprocessing of electrophysiological devices to the FDA - Food And Drug Adminstration1313 U.S. Department of Health and Human Services Food and Drug Division of Dental, Infection Control, and General Hospital Devices. Reprocessing and reuse of single use devices: review prioritization scheme and enforcement priorities for single use devices reprocessed by third parties and hospitals" Docket nº 0053 [internet]. [Accessed in 2016 Sep 13]. Available from: http://www.fda.gov/ohrms/dockets/dailys/00/apr00/041300/c000021.pdf
http://www.fda.gov/ohrms/dockets/dailys/...
as did the GAO - Government Accountability Office, a federal oversight entity of the United States.1414 U.S. Government Accountability Office Single-use medical devices: little available evidence of harm from reuse, but oversight warranted [Internet]. [Accessed in 2016 Aug 13]. Available from: http://www.gao.gov/new.items/he00123.pdf
http://www.gao.gov/new.items/he00123.pdf...

In Brazil, the reprocessing of such products was regulated by the National Health Surveillance Agency (ANVISA), through a Resolution by the Collegiate Board (RDC) 1561515. Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). Resolução - RDC n° 156, de 11 de agosto de 2006. Reprocessamento de produtos para a saúde [Internet]. 2015 set. 17 [Acesso em 2016 jun 01]. Disponível em: http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-rdc-n-156-de-11-de-agosto-de-2006
http://www20.anvisa.gov.br/segurancadopa...
and Special Resolution (RE) 2605,1616 Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). RE n° 2605, de 11 de agosto de 2006. Reprocessamento de produtos para a saúde [Internet]. 2015 set. 17 [Acesso em 2016 jun 01]. Disponível em: http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-re-n-2605-de-11-de-agosto-de-2006.
http://www20.anvisa.gov.br/segurancadopa...
both published in 2006. The RDC 156 establishes that the authorization of reprocessing single-use medical devices, should be at the time of registration in Brazil.1515. Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). Resolução - RDC n° 156, de 11 de agosto de 2006. Reprocessamento de produtos para a saúde [Internet]. 2015 set. 17 [Acesso em 2016 jun 01]. Disponível em: http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-rdc-n-156-de-11-de-agosto-de-2006
http://www20.anvisa.gov.br/segurancadopa...
Despite the fact that most of the manufacturers labeled their products as single-use, ANVISA demands the submission of documents that substantiate the reasons for not reprocessing. Once the manufacturer's arguments are proved and accepted, the words "Reprocessing Forbidden" must be included in the label of that certain product. Also, RE 26051616 Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). RE n° 2605, de 11 de agosto de 2006. Reprocessamento de produtos para a saúde [Internet]. 2015 set. 17 [Acesso em 2016 jun 01]. Disponível em: http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-re-n-2605-de-11-de-agosto-de-2006.
http://www20.anvisa.gov.br/segurancadopa...
lists 66 materials classified as materials whose reprocessing is invariably forbidden. We stress that said list does not contain any product used in the electrophysiological procedures routine.

In 2013, ANVISA issued Technical Note No. 001/20131717 Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). Nota Técnica n° 001/2013/GEMAT/GGTPS/ANVISA. Informações de rotulagem de produtos para saúde [Internet]. 2013 out. 03 [Acesso em 2016 jun 01]. Disponível em: http://profissional.sbhci.org.br/reprocessamento-de-materiais/
http://profissional.sbhci.org.br/reproce...
reiterating the validity of the reprocessing rules published in 2006, in reply to the users' recurrent doubts and demands for clarifications, as per the following excerpt from the resolution: "demands and questions regarding the correct interpretation to be given to the contents of the labels of product for a single use, available in the market, has become increasingly frequent". Currently, in spite of this notice, doubts still persist with regard to the understanding of the rules in force. Due to that, we have made a detailed analysis of the labels of materials routinely used here in electrophysiology procedures, with the purpose of assessing possible incongruences that justify misunderstandings and interpretation errors.

For such analysis, we analyzed the contents of the labels of materials used in electrophysiological procedures, written in Portuguese, available in ANVISA's database http://www.anvisa.gov.br/scriptsweb/correlato/correlato_rotulagem.htm. We included labels from 7 manufacturers that registered products intended for electrophysiology with ANVISA. Once the website had been accessed, we typed the name of the manufacturers in field "Supplier's Name", obtaining a complete list of medical products each manufacturer. Afterwards, we chose only the labels of the products used in electrophysiological procedures. The labels were then printed out, numbered and grouped according to their similarity with regard to physical characteristics and technical applicability, classified as: 1) fixed-curve diagnostic catheter; 2) deflectable-curve diagnostic catheter; 3) circular catheter or high-density mapping catheter; 4) non-irrigated ablation catheter; 5) irrigated ablation catheter; 6) introducers and sheaths; 7) transseptal needle; and 8) intracardiac echocardiography catheter. Labels and/or records with more than one kind of product were attached more than once, that is, one for each of the products to which they corresponded, according to the applicability and characteristic thereof.

The products were then classified into 5 groups:

  1. G1 - reprocessing permitted;

  2. G2 - reprocessing forbidden;

  3. G3 - irregular condition, for not complying with the labeling recommendations set forth in RDC 156;

  4. G4 - conflicting information; and

  5. G5 - no label in ANVISA's database.

This classification was based on the RDC 156, which recommends that the labels should contain only the words: "Reprocessing Forbidden" or "The manufacturer recommends single use"1515. Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). Resolução - RDC n° 156, de 11 de agosto de 2006. Reprocessamento de produtos para a saúde [Internet]. 2015 set. 17 [Acesso em 2016 jun 01]. Disponível em: http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-rdc-n-156-de-11-de-agosto-de-2006
http://www20.anvisa.gov.br/segurancadopa...
. Thus, the products whose labels did not contain the expression "Reprocessing Forbidden" were defined as G1, and they may or may not contain the words "The manufacturer recommends single use". In G2, the products whose labels carried the expression "Reprocessing Forbidden" were included, in spite of the presence of any other word or information. In G3, labels with expressions "Single Use", "Product for Single Use", "Do Not Re-Sterilize", "Discard after using" and "Destroy after using" were included, even if accompanied by expression "The manufacturer recommends single use", given that, pursuant to Technical Note No. 001/2013,1717 Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). Nota Técnica n° 001/2013/GEMAT/GGTPS/ANVISA. Informações de rotulagem de produtos para saúde [Internet]. 2013 out. 03 [Acesso em 2016 jun 01]. Disponível em: http://profissional.sbhci.org.br/reprocessamento-de-materiais/
http://profissional.sbhci.org.br/reproce...
these sentences are considered to not be in conformity with the rules of the regulatory agency. The products that had 2 or more labels, with recommendations differing from one another and/or irregular, were classified as G4. And lastly, the products not in ANVISA's database were classified as G5.

The products included in G1 and G2 were considered to have their labels in conformity with ANVISA's rules, while those classified as G3, G4 and G5 were considered to not be in conformity.

For each group of products with the same applicability and characteristic, and whose labels were in conformity with ANVISA (G1 and G2), it was also assessed whether they were uniform with regard to reprocessing prohibition or not.

Lastly, physical labels were compared by sampling with the labels in ANVISA's database, to assess whether the information contained in both sources matched.

The labeling research was made from July 25, to August 25, 2016 and identified 121 products used in electrophysiological procedures, registered with ANVISA's database, totaling 116 labels (Table 1). Forty-five labels (37.2%) were classified as reprocessing permitted (G1); 41 (33.9%) as reprocessing forbidden (G2); 28 (23.1%) were irregular (G3); 3 (2.5%) had two or more labels with conflicting information (G4) and lastly, four products (3.3%) did not have labels in ANVISA's database (G5). We were then able to note that 86 (71.1%) labels were in conformity, whereas 34 (28.9%) were not in conformity with RDC 156.

Table 1
Labels of medical products grouped according to similarity of characteristics and applicability

The analysis of sub-groups of products with similar characteristics and the same applicability, included in G1 and G2 (labels in conformity with RDC 156), showed that only the intracardiac echocardiography catheter was uniform with regard to the reprocessing recommendations. In this specific case, all six existing types had in their labels the words "The manufacturer recommends single use", which characterizes, therefore, a reprocessing permission. Other products did not have parity in the contents of their labels (Table 1).

Three products were classified as G4, of which one was a fixed-curve diagnostic catheter, one was a deflectable-curve diagnostic catheter and another a non-irrigated ablation catheter with bidirectional curve, all of which were from different manufacturers. The three products had more than one label catalogued in ANVISA's database, under the same registration number and with different recommendations. For the fixed-curve diagnostic catheter (ANVISA registration 10192030102), three labels were found, with the following information: "Reprocessing Forbidden", "The manufacturer recommends single use" and "Product for Single Use", which, pursuant to the RDC 156, mean, respectively, reprocessing forbidden, reprocessing allowed and irregular information. The deflectable-curve diagnostic catheter (ANVISA registration 10341350368) and the ablation catheter (ANIVISA registration 10332340206), for their turn, had two labels, with the following words: "Reprocessing Forbidden" and "The manufacturer recommends single use", which are contradictory instructions.

Lastly, nine physical labels of products used in electrophysiology (Table 2) were analyzed. In six of them, no reprocessing information was found. Upon assessing these six labels in ANVISA's database, we were able to ascertain that one of them contained the expression "The manufacturer recommends single use"; three of them contained the words "Reprocessing Forbidden", and in the other one, the information was not in conformity with RDC 156. In addition, one product (transseptal introducer sheath, registered with ANVISA under No. 10332340208) did not have a label in ANVISA's database.

Table 2
Comparative analysis of physical labels vs. labels in ANVISA’s database

As we were able to note in this analysis, in spite of the fact that the reprocessing of materials used in electrophysiological procedures is allowed and regulated by ANVISA, there are important incongruences in the labels, in a number of products that is not trifling, which may generate mistaken interpretations by the users, and consequently the improper reprocessing of said materials.

The contents of 34 labels (28.9%) from ANVISA's database, which are not in conformity with RDC 156, require urgent adaptation.

We consider it to be extremely important for this information, defined upon the registration of the product, to be clear and irrefutable, thus making sure a quick and correct indentification of medical products with regard to the use thereof. In this regard, the labels should contain a single expression, clearly defining the situation of each medical product: "reprocessing forbidden" or "reprocessing allowed". We are also of the opinion that the criteria used to classify the product must be standardized and ensure the equity of information contained in the physical labels and in ANVISA's database.

Also, it is our opinion that the technical information submitted to ANVISA by the manufacturers, in justification of the prohibition of reprocessing a certain product, upon the registration thereof, must be accessible to the users for them to be aware of it.

For these reasons, the Brazilian Society of Cardiac Arrhythmias (SOBRAC) met with the suppliers of electrophysiological products available in the domestic market and suggested an immediate review of the information contained in the labels, so as to adapt them to ANVISA's standards and make the information unequivocal.

In summary, after the conduction of this research, it was possible to reach the following conclusions and suggestions:

  1. The reprocessing and reuse of medical products in electrophysiology is permitted in Brazil and regulated by ANVISA, through RDC 156;

  2. A thorough analysis of the labels found inconsistencies that could entail misinterpretations and improper decisions by the users with regard to the compliance with RDC 156, even if unintentional.

  3. In the current scenario, while these incongruences are not rectified, it is necessary for the healthcare services that reprocess such products to make a stringent and systematic assessment of both product labels: the physical one and the one in ANVISA's database, with the purpose of identifying the ones that are not in conformity with the RDC 156, as well as those that contain differing instructions, thus avoiding mistakes in the compliance with ANVISA's orders.

  • Sources of Funding
    There were no external funding sources for this study.
  • Study Association
    This study is not associated with any thesis or dissertation work.

References

  • 1
    Tessarolo F, Disertoric M, Caola I, Guarrera GM, Favaretti C, Nollo G. Health technology assessment on reprocessing single-use catheters for cardiac electrophysiology: results of a three-years study. Conf Proc IEEE Eng Med Biol Soc. 2007;2007:1758-61.
  • 2
    Medical Data Institute. International Institute for Health Economics [internet]. [Accessed in 2016 Aug 16]. Available from: http://www.md.institute.com/cms/ressorts/medizintechnologie/aufbereitung-medizinprodukte/Pressemitteilung-Gutachten-bestaetigt-Wiederaufbereitung-von-Einwegprodukten-senkt-Kosten-EN.pdf
    » http://www.md.institute.com/cms/ressorts/medizintechnologie/aufbereitung-medizinprodukte/Pressemitteilung-Gutachten-bestaetigt-Wiederaufbereitung-von-Einwegprodukten-senkt-Kosten-EN.pdf
  • 3
    Avitall B, Khan M, Krum D, Jazayeri M, Hare J. Repeated use of ablation catheters: a prospective study. J Am Coll Cardiol. 1993;22(5):1367-72.
  • 4
    Dunnigan A, Roberts C, McNamara M, Benson DW Jr, Benditt DG. Success of reuse of cardiac electrode catheters. Am J Cardiol. 1987;60(10):807-10.
  • 5
    Mickelsen S, Mickelsen C, MacIndoe C, Jaramillo J, Bass S, West G, et al. Trends and patterns in electrophysiologic and ablation catheter reuse in the United States. Am J Cardiol. 2001;87(3):351-3.
  • 6
    O'Donoghue S, Platia EV. Reuse of pacing catheters: a survey of safety and efficacy. Pacing Clin Electrophysiol. 1988;11(9):1279-80.
  • 7
    Ross DL. Re-use of electrode catheters labeled as single use for clinical cardiac electrophysiological studies. Aust N Z J Med. 1997;26(5):632-5.
  • 8
    Blomstrom-Lundqvist, C. The safety of reusing ablation catheters with temperature control and the need for a validation protocol and guidelines for reprocessing. Pacing Clin Electrophysiol. 1998;21(12):2563-70.
  • 9
    Bank, Alan J, Berry James M, Wilson RF, Lester BR. Acceptance criteria for reprocessed AcuNav catheters: comparison between functionality testing and clinical image assessment. Ultrasound Med Biol. 2009;35(3):507-14.
  • 10
    Lester BR, Boser NP, Miller K, Schnapf M, Jacques BA, Sullivan BF, et al. Reprocessing and sterilization of single-use electrophysiological catheters: removal of organic carbon and protein surface residues. J AOAC Int. 2009;92(4):1165-73.
  • 11
    Druce, JD, Russell JS, Birch CJ, Yates LA, Harper RW, Smolich JJ. A decontamination and sterilization protocol employed during reuse of cardiac electrophysiology catheters inactivates human immunodeficiency virus. Infect Control Hosp Epidemiol. 2003;24(3):184-90.
  • 12
    Lindsay BD, Kutalek SP, Cannom DS, Hammill SC, Naccarelli GV; NASPE Task Force on Reprocessing of Electrophysiological Catheters. North American Society of Pacing and Electrophysiology. Reprocessing of electrophysiology catheters: clinical studies, regulations, and recommendations. A report of the NASPE Task Force on Reprocessing of Electrophysiological Catheters. North American Society of Pacing and Electrophysiology. Pacing Clin Electrophysiol. 2001;24(8 Pt1):1297-305
  • 13
    U.S. Department of Health and Human Services Food and Drug Division of Dental, Infection Control, and General Hospital Devices. Reprocessing and reuse of single use devices: review prioritization scheme and enforcement priorities for single use devices reprocessed by third parties and hospitals" Docket nº 0053 [internet]. [Accessed in 2016 Sep 13]. Available from: http://www.fda.gov/ohrms/dockets/dailys/00/apr00/041300/c000021.pdf
    » http://www.fda.gov/ohrms/dockets/dailys/00/apr00/041300/c000021.pdf
  • 14
    U.S. Government Accountability Office Single-use medical devices: little available evidence of harm from reuse, but oversight warranted [Internet]. [Accessed in 2016 Aug 13]. Available from: http://www.gao.gov/new.items/he00123.pdf
    » http://www.gao.gov/new.items/he00123.pdf
  • 15.
    Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). Resolução - RDC n° 156, de 11 de agosto de 2006. Reprocessamento de produtos para a saúde [Internet]. 2015 set. 17 [Acesso em 2016 jun 01]. Disponível em: http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-rdc-n-156-de-11-de-agosto-de-2006
    » http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-rdc-n-156-de-11-de-agosto-de-2006
  • 16
    Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). RE n° 2605, de 11 de agosto de 2006. Reprocessamento de produtos para a saúde [Internet]. 2015 set. 17 [Acesso em 2016 jun 01]. Disponível em: http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-re-n-2605-de-11-de-agosto-de-2006.
    » http://www20.anvisa.gov.br/segurancadopaciente/index.php/legislacao/item/resolucao-re-n-2605-de-11-de-agosto-de-2006.
  • 17
    Ministério da Saúde. Agência Nacional de Vigilância Sanitária. (ANVISA). Nota Técnica n° 001/2013/GEMAT/GGTPS/ANVISA. Informações de rotulagem de produtos para saúde [Internet]. 2013 out. 03 [Acesso em 2016 jun 01]. Disponível em: http://profissional.sbhci.org.br/reprocessamento-de-materiais/
    » http://profissional.sbhci.org.br/reprocessamento-de-materiais/

Publication Dates

  • Publication in this collection
    Feb 2017

History

  • Received
    26 Sept 2016
  • Reviewed
    11 Oct 2016
  • Accepted
    27 Oct 2016
Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br