Serodiagnosis of Trypanosoma cruzi Infection Using the New Particle Gel Immunoassay - ID-PaGIA Chagas

Abstract

The ID-Chagas test is a particle gel immunoassay (PaGIA). Red coloured particles are sensitised with three different synthetic peptides representing antigen sequences of Trypanosoma cruzi: Ag2, TcD and TcE. When these particles are mixed with serum containing specific antibodies, they agglutinate. The reaction mixture is centrifuged through a gel filtration matrix allowing free agglutinated particles to remain trapped on the top or distributed within the gel. The result can be read visually. In order to investigate the ability of the ID-PaGIA to discriminate negative and positive sera, 111 negative and 119 positive, collected in four different Brazilian institutions, were tested by each of the participants. All sera were previously classified as positive or negative according to results obtained with three conventional tests (indirect immunofluorescence, indirect hemaglutination, and enzime linked immunosorbent assay). Sensitivity rates of ID-PaGIA varied from 95.7% to 97.4% with mean sensitivity of 96.8% and specificity rates varied from 93.8 to 98.8% with mean specificity of 94.6%. The overall Kappa test was 0.94. The assay presents as advantages the simplicity of operation and the reaction time of 20 min. In this study, ID-PaGIA showed to be highly sensitive and specific.

Chagas' disease; immunodiagnosis; serological tests; particle gel immunoassay


Serodiagnosis of Trypanosoma cruzi Infection Using the New Particle Gel Immunoassay - ID-PaGIA Chagas

Vol. 94(1): 77-82

Ana Rabello/+, Alejandro O Luquetti*, Eliana F Moreira**, Maria de Fátima Gadelha***, José Alisson dos Santos**** Laércio de Melo****, Peter Schwind*****

Centro de Pesquisas René Rachou-Fiocruz, Caixa Postal 1743, 30190-002 Belo Horizonte, MG, Brasil *Universidade de Goiás, Goiânia, GO, Brasil **Hemocentro de Pernambuco, Recife, PE, Brasil ***Fundação Ezequiel Dias, Belo Horizonte, MG, Brasil ****DiaMed Brasil, Belo Horizonte, MG, Brasil *****DiaMed

AG-Cressier sur Morat, Switzerland

The ID-Chagas test is a particle gel immunoassay (PaGIA). Red coloured particles are sensitised with three different synthetic peptides representing antigen sequences of Trypanosoma cruzi: Ag2, TcD and TcE. When these particles are mixed with serum containing specific antibodies, they agglutinate. The reaction mixture is centrifuged through a gel filtration matrix allowing free agglutinated particles to remain trapped on the top or distributed within the gel. The result can be read visually. In order to investigate the ability of the ID-PaGIA to discriminate negative and positive sera, 111 negative and 119 positive, collected in four different Brazilian institutions, were tested by each of the participants. All sera were previously classified as positive or negative according to results obtained with three conventional tests (indirect immunofluorescence, indirect hemaglutination, and enzime linked immunosorbent assay). Sensitivity rates of ID-PaGIA varied from 95.7% to 97.4% with mean sensitivity of 96.8% and specificity rates varied from 93.8 to 98.8% with mean specificity of 94.6%. The overall Kappa test was 0.94. The assay presents as advantages the simplicity of operation and the reaction time of 20 min. In this study, ID-PaGIA showed to be highly sensitive and specific.

Key words: Chagas' disease - immunodiagnosis - serological tests-particle gel immunoassay

The present epidemiological data concerning the burden of Chagas' disease claim for new strategies for disease management and control. According to the World Health Organization (WHO 1997), 16-18 million people are currently infected, and 25% of the total population in Central and South America are at risk (TDR, Progress Report 95-96). The most important route of transmission of this endemic disease caused by the protozoan parasite Trypanosoma cruzi is by contact of a human being with an infected triatomine bug. Transfusional transmission is considered the second main route of infection and in some endemic regions, where vectorial transmission is under control or in countries where triatomine vectors are not settled, blood transfusion becomes the most important mechanism of disease transmission (Moncayo 1992, Dias & Coura 1997). Screening of infected blood, in blood banks, aiming the interruption of the transfusional transmission of this disease represents an important measure among the control priorities (Schmunis 1991). Interruption of transmission in Southern Cone, Argentina, Brazil, Chile and Uruguay, due to initiatives of coverage of blood banks by screening, is expected in 1998. The target in Colombia, Ecuador and Peru was to implement blood screening by 1997. Emphasis on blood bank is being ensured in Costa Rica, El Salvador, Guatemala, Mexico, Nicaragua, Panama and Honduras. Screening is now mandatory in Bolivia, where prevalence of T. cruzi infected blood in blood banks has been as high as 63% (Schmunis 1985).

In chronic stage of the infection, detection of the parasite in blood by direct parasitological examination, hemoculture or xenodiagnosis is highly specific, but sensitivity varies from 34% to 85% depending on technical conditions (Cedillos et al. 1970, 1982, Cerisola et al. 1971, Luz et al. 1994). Thus, detection of serum antibodies to T. cruzi antigens has been the main method for the diagnosis of Chagas' disease.

Among several serological tests available the most commonly used are the indirect hema-glutination assay (IHA), the indirect immunofluorescent test (IIFT), the direct agglutination (DA) and the enzyme linked immunosorbent assay (ELISA). Most of these serological tests are based on reagents prepared from whole parasites or complex mixtures of antigens; technical procedures may differ from one laboratory to the other, so variable results are obtained.

Different defined T. cruzi antigens, purified proteins from parasite lysates and recombinant antigens have been produced since the late 80s. Eleven of them showed sensitivity varying from 82 to 100% and specificity varying from 83 to 100% when tested in a double blind multicentric study (Moncayo & Luquetti 1990).

The development of diagnostic methods that combine reliable antigens, as synthetic peptides, with a simple and reproducible assay is highly desirable for endemic disease control programs. In this work, a novel immunodiagnostic assay for Chagas' disease immunodiagnosis, the ID-Chagas antibody test is described and its performance is compared to the conventional serology.

Sample size - In order to investigate the ability of the ID-PaGIA Chagas test to discriminate negative and positive sera, a sample size of 100 negative and 100 positive serum samples was required using as parameters a minimum Kappa agreement of 0.80 and an expected tests concordance of 90% (Landis & Koch 1977).

Two hundred and thirty samples, 119 positive and 111 negative, collected in four different Brazilian institutions were interchanged and all tested by each of the participants of the respective institution: (i) Centro de Pesquisas René Rachou-Fiocruz (positive sera from Bambuí, endemic area of the State of Minas Gerais, negative sera from Ibituruna, a non-endemic area for Chagas' disease); (ii) Universidade Federal de Goiás (frozen samples from the Laboratório de Pesquisa da Doença de Chagas, Faculdade de Medicina, that is the reference for local blood banks and patients care units); (iii) Hemocentro de Pernambuco (samples recently collected from patients living in Recife and from blood donors) and (iv) Fundação Ezequiel Dias, Funed, in Belo Horizonte (sera from blood donors and health basic care in Minas Gerais). Negative samples were all from normal individuals.

Conventional serological tests - Three serological tests (IIFT, IHA and ELISA) were performed. The IIFT was performed according to Camargo and Hoshino-Shimizu (1974). T. cruzi epimastigotes were used as antigen and fluorescein isothiocyanate conjugated sheep anti-human immunoglobulin G, (Fluoline G, Biolab Diagnóstica SA, Rio de Janeiro) was used as the second antibody. The Chagas HAI-Immunoserum kit (Immunoserum-Tecnologia Imunológica Indústria e Comércio, São Paulo, Brazil) was used for IHA and the ELISA-Hemobio Chagas (Embrabio, Empresa Brasileira de Biotecnologia, São Paulo, Brazil) or Chagnostika kit (Organon Teknika, São Paulo, Brazil) were used for the ELISA test.

All sera included were split into two groups, according to the serological results obtained: those belonging to non infected individuals and those from infected people. Conventional serology that presented discordant results was confirmed by the Laboratory of Chagas' disease from the University of Goiás for all sera that presented discrepant results.

ID-Chagas antibody test - The ID-Chagas antibody test is a particle gel immunoassay (PaGIA). The ID Micro Typing System (gel test) which is a well known system for immuno-hematology laboratorial procedures, monitors hemagglutination reactions by the focusing and spatial separation of erythrocytes within a gel matrix. In the PaGIA system (Fig. 1) erythrocytes are replaced by coloured polymer particles coated with the target antigens. When antigen coated particles are mixed with serum containing specific antibodies, particles do agglutinate. The reaction mixture is centrifuged through a gel filtration matrix allowing free non-agglutinated particles to deposit on the bottom of the microtube while agglutinated particles remain trapped on top or distributed within the gel. The result can be read visually.

Particle preparation - The standard PaGIA beads consist of red coloured particles which serve as a universal carrier (Hobbs et al., manuscript in preparation). Beads were coated with an optimisedmixture of three different synthetic peptides representing antigenic sequences from T. cruzi trypomastigote antigens: Ag2, TcD and TcE previously described (Peralta et al. 1994, Reed 1996). The synthetic peptide-coated solid phase beads were stored refrigerated as a ready-to-use 0.15% solids (w/v) suspension. The same batch of sensitized beads was used in all institutions.

ID-Cards - The ID-Micro Typing System uses specially designed ID-Cards. Each card has six reaction chambers containing a buffered Sephacryl gel matrix.

Test procedure - After appropriate identification of serum the aluminium foil that covers the ID-Cards is removed and the ID-Cards are placed vertically in the cards holders. The polymer particles reagent tube is mixed in a vortex thoroughly for 5 sec and 25 ml of plasma or serum are then added into the particle suspension in the microtube. Reactants are mixed by aspirating and expelling one volume of particles/serum mixture within the reaction chamber. The ID-Cards are incubated for 10 min at room temperature. After incubation, ID-Cards are centrifuged for 10 min in a special ID-Centrifuge and reaction are read and recorded.

Interpretation of the results - Positive results, which indicate the presence of detectable levels of anti-T. cruzi antibodies are typically illustrated by the presence of a tight red line on the top of the gel surface or by the presence of agglutinated beads distributed within the gel matrix. Negative results were illustrated by the presence of a tight pellet of particles at the bottom of the microtube following centrifugation with no agglutinated particles present within or in the top of the gel (Fig. 2).

Results of ID-PaGIA Chagas were considered positive or negative by independent visual interpretation agreement of three technicians not aware of conventional serology classification.

Statistical analysis - Sensitivity and specificity of the ID-Chagas antibody test were calculated based on conventional serology criteria: all three positive or all three negative conventional serology tests. Kappa agreement was calculated to evaluate agreement between conventional serology and ID-Chagas test (Landis & Koch 1977).

Sensitivity rates varied from 95.7% to 97.4% with mean sensitivity of 96.8% and specificity rates varied from 93.8 to 98.8% with mean specificity of 94.6%. The overall Kappa test was 0.94. Comparisons of results performed by each of the institutions can be seen in Table I.

The grouped results from each institution are shown in Table II. Inter-laboratory reproducibility was high since of 476 tests done with the 119 positive sera, six sera gave false negative results by all institutions and of 444 tests done with the 111 negative sera, 11 sera gave false positive results by all institutions.

After appropriate training of the technicians visual interpretation was easy and fast. Only seven sera out of the 230 were differently interpreted by one of the three technicians which gives a percentual agreement of 96.9%. From these seven sera, five were considered positive by two and doubtful by one technician. The other two sera were considered positive by two technicians and negative by the third one.

In this study, the new PaGIA - ID-PaGIA Chagas - has been compared to three conventional immunoassays for the Chagas' disease serodiagnosis. Although high sensitivity and specificity have been found for each of the institutions involved, differences among them could be observed. A higher frequency of false positive and false negative results occurred in Goiás. Several variables that may have influenced the test performance cannot be ruled out, for instance, false-negative results may be related to parasite strains and/or individual host immune response. Since detailed epidemiological and clinical analyses were beyond the scope of this work other studies are required to investigate the regional differences.

Criteria of positivity was based on three serological positive and negative sera in order to have the best attainable serological definition. Further studies will address the comparison of the ID-PaGIA Chagas with each of the three conventional serological tests.

Different circumstances make the standardization of Chagas' disease serodiagnosis difficult. Ideally, the more efficient tests may result from the standardization of pure immunodominant fractions of T. cruzi antigens to replace the complex extracts of the parasite. Sensitivity with defined antigens may be improved by the association of different antigenic sites selected according to complementary reactivity and non-specific reactions may be avoided by the use of appropriate antigens. With the use of synthetic peptides batch to batch variations may be largely reduced and the development of simple format methods that lessens test steps and handling may offer increased safety.

The specificity of existing diagnostic tests is known to be adversely affected by cross-reactivity with sera from patients infected with other Trypanosoma spp., Leishmania spp., syphilis, Amoeba spp., Giardia spp. (Schottelius 1982). The use of purified and better-defined antigens improves sensitivity and specificity of serological tests. The diagnostic efficacy of T. cruzi recombinant antigens has been evaluated by different laboratories. In a study coordinated by the WHO specificity and sensitivity of 10 out of 17 recombinant antigens tested equal or higher than 0.80 with sensitivity ranging from 95 to 100% and specificity ranging from 86 to 100% (Schmunis 1991). Although several recombinant antigens displayed high sensitivity and specificity some of them were unable to detect anti-T. cruzi antibodies in four out of 148 chronic chagasic patients (Pan et al. 1992).

The cross reactivity of TcD and PEP2 peptides has been studied by Peralta et al. (1994). The specificity was 99% in an ELISA test using the association of these two antigens tested for samples from patients with cutaneous leishmaniasis, visceral leishmaniasis, leprosy, tuberculosis and for samples from healthy individuals.

In the ID-Chagas antibody test, the use of well defined antigens in the form of highly purified synthetic peptides serves to minimise such cross-reactivity. Cross-reactivity with visceral leishmaniasis was observed only in three of 25 sera obtained from patients with visceral leishmaniasis (Schwind, pers. commun.).

The ID-PaGIA format presents a fast and simple alternative to ELISA because only two sequential pipetting steps are needed as it is devoid of multiple reagent addition and wash steps intrinsic to the ELISA technique. The exposure to potentially infectious material inherent to all previously described techniques is significantly reduced as the technique requires only the single addition of an undiluted sample along with the non-infectious particle reagent. Thereafter, all reagents remain within the test card which may be resealed to completely enclose the samples. The cost of the test is comparable to an ELISA.

In this study, direct comparisons between the ID-PaGIA and conventional serology have shown that the ID-Chagas antibody test is highly sensitive and specific. The assay presents as advantages the simplicity of operation and the reaction time of 20 min.

To Áureo Almeida and Cláudia Correa Carli from the Centro de Pesquisas René Rachou, Suelene B Nascimento Tavares and Rosangela Amaral de Oliveira from Universidade Federal de Goiás, Flávia Cunha from Hemocentro de Pernambuco, Ildiko de Miranda Pinto, Sonia Conceição Machado Diniz, and Sylvia Vignoli Cordeiro from DiaMed-Brasil and Myriam Brodard from DiaMed-Switzerland for technical assistance. To Dr Mark Roberts and Rod Hobbs for development of the polymer particles.

  • Camargo ME, Hoshino-Shimizu S 1974. Metodologia sorológica na infecção pelo

    Trypanosoma cruzi.

    Rev Goiana de Med 20: 47-65

    .

  • Cedillos RA, Dimas D, Hernandes AY 1970. Blood concentration method in the diagnosis of Chagas' disease.

    Rev Lat Amer Microbiol 12: 200-203.

  • Cedillos RA, Torrealba JW, Tonn RJ, Mosca W, Ortegon A 1982. El xenodiagnóstico artificial en la Enfermedad de Chagas.

    Bol Of Sanit Panam 93: 240-248.

  • Cerisola JA, Rohwedder R, Del Prado CE 1971. Rendimiento del xenodiagnóstico en la infection chagásica cronica humana utilizando ninfas de diferentes espécies de triatomineos.

    Bol Chil Parasitol 26: 57-58.

  • Dias JCP, Coura JR 1997. Epidemiologia, p. 33-66. In JCP Dias & JR Coura (eds)

    , Clínica e Terapêutica da Doença de Chagas. Uma Abordagem Prática para o Clínico Geral, Fiocruz, Rio de Janeiro.

  • Landis JR, Koch GG 1977. The measurement of observer agreement for categorical data.

    Biometrics 33: 159-174.

  • Luz ZM, Coutinho MG, Cançado JR, Krettli AU 1994. Hemoculture: sensitive technique in the detection of

    Trypanosoma cruzi in chagasic patients in the chronic phase of Chagas disease.

    Rev Soc Bras Med Trop 27: 143-148.

  • Moncayo A 1992. Chagas' disease: epidemiology and prospects for interruption of transmission in the Americas.

    World Health Stat Qu 45: 276-279.

  • Moncayo A, Luquetti AO 1990. Multicentre double blind study for evaluation of

    Trypanosoma cruzi defined antigens as diagnostic reagents.

    Mem Inst Oswaldo Cruz 85: 489-495.

  • Pan AA, Rosenberg GB, Hurley MK, Schock GJ, Chu VP, Aiyappa A 1992. Clinical evaluation of an EIA for the sensitive and specific detection of serum antibody to

    Trypanosoma cruzi (Chagas' disease).

    J Infect Dis

    165: 585-588.

  • Peralta JM, Teixeira MGM, Shreffler WG, Pereira JB, Burns Jr JM, Sleath PR, Reed S 1994. Serodiagnosis of Chagas' disease by enzime-linked immuno-sorbent assay using two synthetic peptides as antigens.

    J Clin Micribiol 32: 971-974.

  • Reed S 1996. Coumpounds and methods for the detection of

    T. cruzi infection. Patent application WO 96/29605.

  • Schmunis GA 1985. Chagas' disease and blood transfusion.

    Prog Clin Biol Res

    182: 127-145.

  • Schmunis GA 1991.

    Trypanosoma cruzi, the etiologic agent of Chagas' disease: status in the blood supply in endemic and non endemic countries.

    Transfusion 31: 547-557.

  • Schnaidman BB, Yoshida N, Gorin PA, Travassos LR 1986. Cross-reactive polysaccharides from

    Trypanosoma cruzi and fungi (especially

    Dactylium dendroides).

    J Protozool 33:186-191.

  • Schottelius J 1982. The identification by lectins of two strain groups of

    Trypanosoma cruzi.

    Z Parasitenkd

    68: 147-154.

  • WHO 1997. Tropical Disease Research: Progress Report 95-96, Geneva, 134 pp.

+Corresponding author. Fax: +55-31-295.3115. E-mail: ana@netra.cpqrr.fiocruz.br

Received 12 May 1998

Accepted 3 November 1998

Figure 1

Figure 2

  • Camargo ME, Hoshino-Shimizu S 1974. Metodologia sorológica na infecçăo pelo Trypanosoma cruzi Rev Goiana de Med 20: 47-65.
  • Cedillos RA, Dimas D, Hernandes AY 1970. Blood concentration method in the diagnosis of Chagas' disease. Rev Lat Amer Microbiol 12: 200-203.
  • Cedillos RA, Torrealba JW, Tonn RJ, Mosca W, Ortegon A 1982. El xenodiagnóstico artificial en la Enfermedad de Chagas. Bol Of Sanit Panam 93: 240-248.
  • Cerisola JA, Rohwedder R, Del Prado CE 1971. Rendimiento del xenodiagnóstico en la infection chagásica cronica humana utilizando ninfas de diferentes espécies de triatomineos. Bol Chil Parasitol 26: 57-58.
  • Dias JCP, Coura JR 1997. Epidemiologia, p. 33-66. In JCP Dias & JR Coura (eds), Clínica e Terapęutica da Doença de Chagas. Uma Abordagem Prática para o Clínico Geral, Fiocruz, Rio de Janeiro.
  • Landis JR, Koch GG 1977. The measurement of observer agreement for categorical data. Biometrics 33: 159-174.
  • Luz ZM, Coutinho MG, Cançado JR, Krettli AU 1994. Hemoculture: sensitive technique in the detection of Trypanosoma cruzi in chagasic patients in the chronic phase of Chagas disease. Rev Soc Bras Med Trop 27: 143-148.
  • Moncayo A 1992. Chagas' disease: epidemiology and prospects for interruption of transmission in the Americas. World Health Stat Qu 45: 276-279.
  • Moncayo A, Luquetti AO 1990. Multicentre double blind study for evaluation of Trypanosoma cruzi defined antigens as diagnostic reagents. Mem Inst Oswaldo Cruz 85: 489-495.
  • Pan AA, Rosenberg GB, Hurley MK, Schock GJ, Chu VP, Aiyappa A 1992. Clinical evaluation of an EIA for the sensitive and specific detection of serum antibody to Trypanosoma cruzi (Chagas' disease). J Infect Dis 165: 585-588.
  • Peralta JM, Teixeira MGM, Shreffler WG, Pereira JB, Burns Jr JM, Sleath PR, Reed S 1994. Serodiagnosis of Chagas' disease by enzime-linked immuno-sorbent assay using two synthetic peptides as antigens. J Clin Micribiol 32: 971-974.
  • Reed S 1996. Coumpounds and methods for the detection of T. cruzi infection. Patent application WO 96/29605.
  • Schmunis GA 1985. Chagas' disease and blood transfusion. Prog Clin Biol Res 182: 127-145.
  • Schmunis GA 1991. Trypanosoma cruzi, the etiologic agent of Chagas' disease: status in the blood supply in endemic and non endemic countries. Transfusion 31: 547-557.
  • Schnaidman BB, Yoshida N, Gorin PA, Travassos LR 1986. Cross-reactive polysaccharides from Trypanosoma cruzi and fungi (especially Dactylium dendroides). J Protozool 33:186-191.
  • Schottelius J 1982. The identification by lectins of two strain groups of Trypanosoma cruzi Z Parasitenkd 68: 147-154.
  • WHO 1997. Tropical Disease Research: Progress Report 95-96, Geneva, 134 pp.

Publication Dates

  • Publication in this collection
    04 Feb 1999
  • Date of issue
    Jan 1999

History

  • Received
    12 May 1998
  • Accepted
    03 Nov 1998
Instituto Oswaldo Cruz, Ministério da Saúde Av. Brasil, 4365 - Pavilhão Mourisco, Manguinhos, 21040-900 Rio de Janeiro RJ Brazil, Tel.: (55 21) 2562-1222, Fax: (55 21) 2562 1220 - Rio de Janeiro - RJ - Brazil
E-mail: memorias@fiocruz.br