Effect of additional filtration on radiation doses and image quality in videofluoroscopic studies

Abstracts

OBJECTIVE: The purpose of thys study was to investigate the effect of the addition of aluminum (1 mm) and copper (0.4 mm) filters on effective radiation doses and image quality in videofluoroscopy. MATERIALS AND METHODS: An ionization chamber coupled with an electrometer was added to x-ray tube to measure the kerma area product with 65 kV and 0.7 mA technique, without and with additional filtration. Low contrast, gray scale and spatial resolution were measured utilizing Leeds test objects. Fifteen volunteers underwent pharynx study, ten without and five with aluminum and cooper filters associated, and had the kerma area product/minute compared. RESULTS: The specified filters addition, either separated or associated, allowed an expressive decrease in kerma area product besides an actual improvement in the videofluoroscopic images quality determined by a better gray tones differentiation and increased brightness contrast ratio in the gray curve. CONCLUSION: Additional aluminum and copper filters interposition, especially when associated, results in improved image quality with expressive reduction in the required radiation doses.

Videofluoroscopy; Kerma area product; X-ray; Radiological image; Radiation dose


OBJETIVO: Investigar o efeito da adição de filtros de alumínio (1 mm) e cobre (0,4 mm) na redução das doses efetivas de radiação e na qualidade das imagens em exames videofluoroscópicos. MATERIAIS E MÉTODOS: Ao tubo de raios X adicionou-se câmara de ionização conectada a um eletrômetro para medir o produto kerma-área, com técnica de 65 kVp e 0,7 mA, sem e com adição dos filtros. Foi medida resolução espacial, a de baixo contraste e tons de cinza, utilizando os objetos de teste de Leeds. Quinze voluntários tiveram o produto kerma-área/minuto do estudo faríngeo comparados, dez com filtração e base e cinco com adição dos filtros associados. RESULTADOS: A adição dos filtros separados ou associados produziu expressiva redução do produto kerma-área, com ganho na qualidade das imagens videofluoroscópicas determinado pela maior separação dos tons de cinza e aumento da relação brilho/contraste da curva de cinza. CONCLUSÃO: A interposição adicional de filtros de alumínio e cobre, em especial quando associados, melhora a qualidade das imagens, com expressiva redução das doses de radiação necessárias à sua geração.

Videofluoroscopia; Produto kerma-área; Raios X; Imagem radiológica; Dose de radiação


ORIGINAL ARTICLES


Effect of additional filtration on radiation doses and image quality in videofluoroscopic studies*

Milton Melciades Barbosa CostaI; João Luiz Leocadio da NovaII; Lucía Viviana CanevaroIII

IPhD, Titular Professor in charge of Laboratório de Motilidade Digestiva e Imagem do Instituto de Ciências Biomédicas do Centro de Ciências da Saúde da Universidade Federal do Rio de Janeiro (ICB/CCS-UFRJ), Rio de Janeiro, RJ, Brazil

IIPhD in Sciences of Communication, Associate Professor at Universidade Federal Fluminense (UFF), Niterói, RJ, Brazil

IIIPhD, Physicist at Instituto de Radioproteção e Dosimetria/Comissão Nacional de Energia Nuclear (IRD/CNEN), Rio de Janeiro, RJ, Brazil

Mailing address

ABSTRACT

OBJECTIVE: The purpose of thys study was to investigate the effect of the addition of aluminum (1 mm) and copper (0.4 mm) filters on effective radiation doses and image quality in videofluoroscopy.

MATERIALS AND METHODS: An ionization chamber coupled with an electrometer was added to x-ray tube to measure the kerma area product with 65 kV and 0.7 mA technique, without and with additional filtration. Low contrast, gray scale and spatial resolution were measured utilizing Leeds test objects. Fifteen volunteers underwent pharynx study, ten without and five with aluminum and cooper filters associated, and had the kerma area product/minute compared.

RESULTS: The specified filters addition, either separated or associated, allowed an expressive decrease in kerma area product besides an actual improvement in the videofluoroscopic images quality determined by a better gray tones differentiation and increased brightness contrast ratio in the gray curve.

CONCLUSION: Additional aluminum and copper filters interposition, especially when associated, results in improved image quality with expressive reduction in the required radiation doses.

Keywords: Videofluoroscopy; Kerma area product; X-ray; Radiological image; Radiation dose.

INTRODUCTION

TV fluoroscopy has replaced dark room fluoroscopy(1,2). The use of image intensifiers and the video camera has brought a sensitive improvement in images quality and a significant reduction of radiation exposure both for technicians and patients(2-5). The transition from the visualization of the images on a fluoroscopic screen in a totally dark room to the visualization on a television monitor, besides the doses reduction, represented the birth of video-fluoroscopy.

Videofluoroscopy allows the recording of fluoroscopic images on magnetic media. Its value resides in the feasibility of real-time analysis and reanalysis of the recorded dynamics, without the need for additional exposures. In the study of swallowing and related disorders, the method is considered and accepted as the best one amongst the available ones for the diagnosis and follow-up of dysphagias(6-9). Radiation doses employed in the performance of such examinations are considered to be low or totally acceptable, considering the amount of information that the method provides. However, it is important to know the doses administered to the patient, besides implementing measures for optimization of the practice(2,4,10-12).

If appropriately prescribed, radiological examinations produce invaluable information for disorders diagnosis and follow-up. However, the benefits from the use of x-rays should not lead to an attitude of underestimating their risks. The International Commission on Radiological Protection (ICRP Publication 103)(13) understands that when the benefits of the radiological examination are clinically justifiable, the associated radiation exposure risks are acceptable. The general foundations of radiological protection define as one of its main objectives, the determination of an acceptable level of radiation exposure, without unduly limiting the benefits of the use of x-radiation. However, the magnitude of such doses shall be kept as low as reasonably feasible. Dose limits are not applicable in cases of medical patient's exposure; for such individuals, diagnostic reference levels are utilized. The dose reference level shall be the lowest sufficient exposure to obtain the required information for the indication and allowed by the appropriately proposed radiological ex-amination(13-16).

Due to the fact that no radiologic equipment has been specifically developed for this method, videofluoroscopy relies on apparatuses such as seriography and angiography units, and C-arms, all of them in principle developed for other types of examinations. Aluminum filters coupled with the x-ray tubes of such equipment filter the low-intensity radiation and attenuate higher intensity radiations, concentrating the intensity levels of the primary beam on the desired radiation range.

In specific cases, an increase in filtration may be desirable in order to eliminate low energy photons, which is not done in equipment designed for multiple types of exams. The recommendation is that the total filtration in radiological diagnostic equipment be of at least 2.5 mm of aluminum equivalent(17,18).

The relevance of additional filtration of the main x-ray beam was highlighted by Watanabe(19) as one of the primary means of reducing the radiation exposure dose for radiodiagnosis, and the author further observes that, with additional filtration produced by 0.08 mm of copper, one obtains radiographic images with a 14% smaller radiation dose, with the same quality of films that received a 2.00 mm of aluminum filtered radiation dose.

Nicholson et al.(20) have obtained a dose reduction between 75% and 80% by removing the antiscattering grid and using additional filtration of 0.7 mm steel.

Morrell et al.(21) have identified, in barium enema, a mean dose reduction in the order of 57% (from 17.7 Gy.cm2 to 7.6 Gy.cm2) by using additional filtration with 0.3 mm of copper. They also report that they did not observe any significant image degradation using the Leeds TO10 test object, with the radiological images being analyzed by a specialized radiologist.

Haiter(22), utilizing filters made from aluminum and zinc alloys in dental x-ray apparatuses, have obtained a decrease in air kerma rate of up to 18.53%, without changes in the films contrast. Gonçalves(23), on dental radiographic films, have observed a reduction of air kerma rate of up to 47.33% in relation to the aluminum filter, with the same level of images contrast. Similarly, still with copper and aluminum alloy, Bóscolo et al.(24) have observed a significant reduction of the air kerma rate without a comparable loss of images contrast.

The objective of the present study was to investigate, on videofluoroscopic sequences, the effect of adding aluminum and copper filters at the window of the x-ray tube, and evaluating the radiation dose reduction as well as the images quality using Leeds test objects at videofluoroscopic studies of the cervical region in volunteers.

MATERIALS AND METHODS

The baseline equipment utilized was a Philips BV 22 C-arm (Philips; Holland) with a 100 kV, 20 mA maximum intensity bulb, Philips LR24424 intensifier (Philips; Holland), total nominal filtration of 2.5 mm of aluminum equivalent and 2.4 mm focal point, two-pulse rectification, fixed anode and 10º angle. The TV system of the equipment is of black & white type, based on the NTSC standard, comprising a black & white 20" Philips monitor (Philips; Holland) and a black & white CCD Sony Mythos B/W (Sony; USA) camera (0.1 lux; f = 3.6mm; 400-line resolution) coupled with the image intensifier.

The following devices were coupled with the baseline equipment: a VHS Panasonic NV-MV 40 video recorder (Panasonic; Brazil); a Philips DVD recorder DVDR 3455H (Philips; USA); a color 13" Panasonic CT-1383VY monitor (Panasonic; Mexico).

The images generated by the radiological equipment are captured on a video output (BNC) of the black & white monitor and directed for simultaneous analogical recording (VHS) and digital recording (DVD), with control images being displayed on the color monitor screen.

To the x-ray tube cone a Diamentor M2 PTW (PTW; Freiburg, Germany) a connected ionization chamber was added to measure the kerma-area product (KAP), a quantity defined as the product of the radiation quantity emitted by the x-ray tube by the irradiated area(25,26). Additional aluminum and copper filters were alternatively placed before the ionization chamber, with the objective of evaluating the additional filtration effect. In this equipment, the size of the radiation field was maintained fixed by the collimation system (Figure 1).


KAP and Leeds test objects

The KAP measurement was performed with the technique of 65 kV (2.4 mmAl half-value layer) and 0.7 mA (nominal), as well as the spatial resolution evaluation, low-contrast and gray scale evaluation, using the TOR(TVF), TO10 and GS2 Leeds test objects(27).

The free space between the tube cone and the external surface of the image intensifier is 80 cm. In order to measure KAP, a 1 mm thick copper plate was placed at 60 cm from the external surface of the cone, which had been modified for the insertion of the ionization chamber, however keeping its original dimensions. This plate simulates the attenuation that would be produced by an adult weighting 70 kg and also serves as a protection for the image intensifier.

The KAP and image quality measurements were performed as follows: a) radiological equipment standard filtration with no additional filtration; b) equipment standard filtration + additional filtration with a 1 mm thick aluminum plate; c) equipment standard filtration + additional 0.4 mm thick copper plate filtration; d) equipment standard filtration + 1 mm thick aluminum plate + 0.4 mm thick copper plate. With the objective of comparing the KAP values, the irradiation time utilized in the four situations was (127 ± 1) seconds, a long and continuous time sufficient for obtaining the KAP values in a way to express similarities or differences in a reliable manner.

In order to evaluate spatial and low contrast resolution, the Leeds TOR(TVF) and TO10 test objects were attached, one at a time, to the entry surface of the image intensifier. The spatial resolution was evaluated by identification of the group of line pairs up to the point where the lines of the test object (Figure 2) were distinct(28). The low contrast resolution was analyzed by the identification of the different circles that could be visualized in relation to the density of the plate containing them (background) (Figure 3). Because of the evaluation of the image quality in terms of spatial resolution and low contrast were performed with test objects with semi-quantitative characteristics, the evaluations were simultaneously carried out by three observers.



The gray tones were analyzed on images obtained from rectangular gray scales contained in the center of the GS2 Leeds test object (Figure 4). The object was placed in the entry of the image intensifier and was irradiated with the 65 kV and 0.7 mA technique, generating an image with different gray tones for each density level on value scale visualized in steps. These values were obtained by the analysis of the video signal by using the luminance histogram analysis function of software Adobe Photoshop® 5.5 (Adobe; USA). Separate measurements were performed for each of the gray tones in 460-pixel samples of the video signal, corresponding to each density of the rectangular bar. On each histogram the software calculates and provides the standard deviation for the measurement, expressing the dispersion around the mean level. The analysis for the entire bar was repeated for the four filtering situations. Data were recorded on a Microsoft Office Excel® 2003 (Microsoft Corp.; USA) worksheet, allowing the generation of graphs in a logarithmic scale. The luminance analysis of the video signal for determined areas was also utilized for the ten-step sets in the different filtration conditions, in 17,995 pixels samples related to the video signal of the capture of the whole rectangular bar.


With the objective of estimating the effect of the different filtrations, the software Report 78 SRS-780 (Institute of Physics and Engineering Medicine; UK)(29) was utilized for analyzing x-ray spectra with a 65 keV peak.

The volunteers

The pharynx of 15 male volunteers with ages between 27 and 56 years were evaluated. The individuals, all of them professional wind musicians, were submitted to videofluoroscopic study for morpho-functional analysis of the pharyngeal dynamics during effort.

Such project was approved by the Committee for Ethic in Research of the institution in compliance with the standards proposed by the World Medical Association (WMA Helsinki Declaration, Finland, 1995, supplemented by the 52nd WMA General Assembly, Edinburgh, United Kingdom, 2000, with amendments in Washington, 2002 and Tokyo, 2004). The musicians' pharynges were evaluated on left lateral and anteroposterior views, with the same equipment utilized for KAP measurements and studies with the Leeds test objects. Ten of the volunteers had already had their pharynx previously evaluated at the baseline conditions of the equipment, without additional filtration, and with the exposure time and KAP recorded by the PTW-Diamentor M2 electrometer (PTW; Freiburg, Germany) coupled with the C-arm. The remaining five volunteers were submitted to the same protocol, except for the association with aluminum (1 mm) and copper (0.4 mm) filtration. The examination times were similar in most cases. However, for the benefit of standardization, the KAP per minute of examination time (KAP rate) was considered for each patient.

RESULTS

KAP and Leeds test objects

Table 1 demonstrates the technique (65 kV, 0.7 mA) and the x-radiation exposure time (in seconds) with the four proposed filtration conditions, observing the KAP (cGy.cm2) and spatial resolution (TOR (TVF) Leeds object) variations. Table 2 demonstrates the technique, exposure time and filtration conditions for observation of the variations in low-contrast ratio (TO10 Leeds object).

The graphic on Figure 5 was obtained with the Report 78 SRS-780(29) software demonstrating the different x-ray spectra with 65 keV peak for the different filtrations. One observes the elimination of low-energy x-ray photons, as well as the beam hardening (dislocation of the spectrum peak to higher energies, with higher beam penetration).


The luminance value results and respective standard deviations, obtained by the histograms of separate measurements of each one of the gray tones, with 460 pixels of the video signal, with the Leeds GS2 test object, are expressed on Table 3, which also includes the calculation of the standard deviation percentages. The logarithmic curve obtained from this table allows the visualization of the gray-scale curve enhancement, in particular when the associated filtration of aluminum and copper is utilized. The greater contribution of the aluminum filter in the darker gray tones is highlighted (Figure 6).


The gray tone histogram of the rectangular bar contained in the Leeds GS2 test object, allows the visualization, for each one of the different filtrations, the enhancement of the separation between the gray levels of each step, when one compares the histogram without filtration with the additional filtration histogram, respectively at the left and at the right on Figure 7.


The volunteers

The KAP measurements, total and per minute, obtained in the evaluation of the volunteers submitted to pharyngeal videofluoroscopy without additional filtration, are shown on Table 4, and those with additional filtration (copper + aluminum), on Table 5.

DISCUSSION

Radiodiagnostic studies represent the greatest cause of exposure to artificial sources of ionizing radiation(30,31). However, the daily contact with the radiation sources causes a lenient attitude towards the inherent risks(32). Additionally, it is important to mention the already mentioned possibility of high doses generated and delivered by defective equipment(12,13). In spite of the conscience for the need of using the lowest possible doses, examples of violation of this concept are repetitive(34,35).

Videofluoroscopy is the most appropriate radiological method for the study of dynamic biological processes, allowing noninvasive morpho-functional recording of the phenomena with low radiation doses as compared with fluoroscopy used for extended periods, for later radiographic recording, or with cineradiography which requires considerably higher doses(2,12,36). A further reduction of required and delivered doses, maintaining the diagnostic quality without loss in the images quality, adds value to the method and contributes to the definition of reference levels for the videofluoroscopic swallowing study (4,10- 12). Such levels, still to be definitely defined, tend to be established at lower values due to the significant dose reduction achieved with the use of additional filtration such as those observed in the present study, and which have also been reported by other studies(19-24).

With same test object, the same technical parameters (60 kV, 0.7 mA), and similar exposure times and varying filtration systems, the authors observed that as the aluminum, copper or association of these two filters were added, significant KAP reduction was achieved. With additional copper (0.4 mm) and aluminum (1 mm) filtration the observed reduction in KAP was over 60% as compared with that obtained without additional filtration, with no significant loss in image quality on the semi-qualitative analyses.

The obtained energy distribution curves allowed the observation of a curve's peak dislocation to the right, which means that as the filtration increases, the quantity of low energy photons in the spectrum decreases, reducing the total dose delivered to the patient with no loss in the image quality. This dislocation of the energy peak of the photons to higher values (beam hardening) caused the remaining beam spectrum to have a higher penetration capability.

The interposition of attenuating plates could interfere in the low contrast and spatial resolution of the images. However, the authors observed that the measured values for these parameters did not change with the interposition of the additional filtration. However, low contrast and spatial resolution, most frequently explored in literature are semi-quantitative tests that present a certain degree of subjectivity. Frequently, health professionals associate the image quality with their capacity of seeing and identifying abnormalities on the images.

Thus, an objective quantification of the image quality may be better defined by the analysis of the luminance values and respective standard deviations, obtained by the gray tone histograms of the video signal obtained from the Leeds GS2 test objects. Luminance and gray levels are parameters utilized to identify the best quality of images(37-39).

The gray scale has been widely utilized in the study of video and television systems for more than 60 years, and has been adopted by several centers in the evaluation of radiological image systems(40). This standard, defined by the Electronic Industries Association (standard RS-189) and later adopted by the Society of Motion Picture of Television Engineering, allows for the evaluation of the behavior of the response curve between black and white in video capture and display devices. The relevance of this evaluation must be highlighted, as the cathode ray tubes utilized in video monitors, do not have a linear response to the black-to-white excursion, which requires a previous compensation in the capture system. This compensation system is defined as the gamma parameter or factor, and is measured by the curvature of the gray scale in relation to a hypothetical straight line representing the black-to-white excursion. As the gamma factor increases, the black & white curve differentiates further from a straight line.

It is important to observe that the variation of the gamma factor changes the response of intermediate gray tones, while adjustment in brightness and contrast acts primarily in the limits of white and black on the same scale. Although these are different parameters, there is a strong interaction among gamma factor, brightness and contrast. That is the reason why the high and low contrast response must be evaluated in association with the intermediate gray tone response.

The Electronic Industries Association scale for image quality control in professional television, as well as the Leeds test object used in the present study present the same ten-level gray scale. In this standard there is a 10% variation between each step or level, in a way to cover the entire transition between black and white. Each step has a maximum value deviation or differential gain, ranging between 2% and 4%, that was evaluated by the standard deviation of the gray tones of each step.

For practical ends, the current intensity (mA) can be related to the black level, and the difference in potential (kV) to the white level. For that reason, when there is greater or lesser opacity of a radiological image under analysis, adjustments in kV or mA allow for improvements in the image sharpness.

Table 2 and the respective graphic (Figure 6) demonstrate the gray scale modified by all the filters utilized. Mid-gray tones are more distinctive. Tone separation was greater in the mid values of the scale than those observed in the curve extremities (values for black and white). Both copper and aluminum filters decreased the standard deviation for each gray level, increasing the separation of each level on the scale. The association of two filters contributed even further for the levels separation, as evidenced by the standard deviation percentage for each one of the gray levels.

The analysis of the gray tone set histograms of the whole bar in the different simulated situations allows a clearer visualization of tones separation. Tones that in some cases were covered by the measurement dispersion of a given step became sharper within the distribution. The improvement in the definition between the different steps enhanced the frequency peaks and valleys between each one of the measurements presented on the histogram on Figure 7.

All the added filtrations produced changes in the gray scale curve, increasing the value of the gamma factor and amplifying the spectrum of the represented tones. The mean gray level ranged from 125.24 ± 49.10 to 136.96 ± 52.29. Aluminum was the greatest contributor for the curve expansion, increasing the difference between the highest and lowest level represented in the scale by ten gray tones. Such curve expansion, with an increase in represented levels, in the case of association between aluminum and copper plates increases to 161.23 the obtained difference between the highest level of white and the lowest level of black represented in the Leeds test object scale.

The closest tones to white are more sharply enhanced when the filtration is made by the interposition of the copper filter alone or in association with the aluminum filter. This filtration enhances areas that are more permeable to x-radiation, where the tone separation effectively increases the demonstration of areas whose transparency was very close. This increase in contrast favors a better evaluation of the videofluoroscopic image.

The studies performed with all volunteers had a same objective and followed the same protocol with anteroposterior and left lateral views. In order to independently evaluate the results regarding the different irradiation times, the KAP rates (cGy.cm2/ min) were calculated as shown on Tables 3 and 4, allowing the normalization of the radiation exposure time, independently from the total exposure time at the end of each volunteer examination.

The first ten volunteers were evaluated without any additional filtration besides the inherent one, and the exposure of this group was compared with that of the 5 volunteers group for which aluminum and copper filters in association were added to the inherent filtration.

The mean radiation exposure time observed in the first group (4.2 minutes) is very close to twice the exposure time (2.3 minutes) required for the performance of the same evaluation protocol with the second group. How can this difference in required time for the same protocol be explained?

These videofluoroscopic examinations were aimed at understanding and recording the structural response of the pharynx to the increase in the pressure generated by the transit of two or three deglutitions of known volumes of 100% barium sulfide solution, and also the increase in pharyngeal pressure increase caused by forced expiration against resistance in healthy wind musicians.

The establishment of the study protocol and morpho-functional analysis of the images during their recording ended up generating exposures above recommended and/or required for the protocol accomplishment, since the events could be recorded and analyzed in a considerably shorter time without any loss of information.

Certainly, this reduction of the time required for the study performance had as one of its components, a greater experience in the execution of the protocol, however, undoubtedly, the knowledge of the fact that the radiation doses were being recorded brought a greater diligence in the protocol accomplishment, with a reduction of the total radiation exposure time and dose.

From the calculation of KAP rates, one observes that the second group exposure is 2.5 times lower than that observed in the first group. This KAP reduction can now be attributed to the additional filtration produced by the association of the aluminum and copper plates. The KAP reduction of 2.5 times is practically the same observed with the interposition of the 1 mm copper plate, which was 2.7 times smaller when the copper and aluminum filters are associated, as compared with the inherent filtration.

In what concerns the quality of images obtained in the volunteer studies with the association of copper and aluminum plates, the improvement in the image quality with a better gray level resolution was perceived. Such subjective perception is confirmed by the results of the semi-quantitative test results obtained with the use of the TOR (TVF) and TO10 Leeds test objects and by the gray scale luminance study of the GS2 Leeds object.

Besides demonstrating gains in images quality, such findings enhance the perception of the fact that the consciousness of being measuring the radiation and the use of the plates is an important factor in the determination of the reference levels to be established for videofluoroscopic and certainly for the other fluoroscopic studies.

CONCLUSIONS

The interposition of additional copper and aluminum filtration, particularly when associated, improves the images quality, with considerable reduction in the radiation doses required for their generation. These facts allow the conclusion that the addition of the filters to the radiological system is an important factor to be considered in the performance of videofluoroscopic studies.

Acknowledgements

To Mr. Tadeu José Dias, technician in charge of SJTR Equipamentos Médico Hospitalar Ltda., for the support in the adaptation and implementation of the necessary changes in the equipment for video-fluoroscopic study of deglutition, and to Leslie Silva Pereira, of Instituto de Radiodosimetria do Rio de Janeiro, for the support in the utilization of the software Report 78.

REFERENCES

  • 1. Hofmann FW, Marhoff P. High resolution fluoroscopic equipment. In: Quality control in diagnostic radiology. Conference Reports Series-26. London: The Hospital Physicists' Association; 1976. p. 27-9.
  • 2. Costa MMB, Nova JLL, Carlos MT, et al. Videofluoroscopia: um novo método. Radiol Bras. 1992;25:11-8.
  • 3. Airth GR. Image requirements. In: Quality control in diagnostic radiology. Conference Reports Series-26. London: The Hospital Physicists' Association; 1976. p. 26-7.
  • 4. Costa MMB, Canevaro LV, Azevedo ACP. Análise dosimétrica do método videofluoroscópico aplicado ao estudo da dinâmica da deglutição. Radiol Bras. 2000;33:353-7.
  • 5. American Association of Physicists in Medicine. AAPM Report No. 74. Quality control in diagnostic radiology. Report of Task Group #12. Diagnostic X-ray Imaging Committee. Madison: Medical Physics Publishing; 2002.
  • 6. Beck TJ, Gayler BW. Image quality and radiation levels in videofluoroscopy for swallowing studies: a review. Dysphagia. 1990;5:118-28.
  • 7. Noll SF, Bender CE, Nelson MC. Rehabilitation of patients with swallowing disorders. In: Braddom RL, editor. Physical medicine & rehabilitation. 2nd ed. Philadelphia: Saunders; 2000. p.535-60.
  • 8. Costa MMB. Uso de bolo contrastado sólido, líquido e pastoso no estudo videofluoroscópico da dinâmica da deglutição. Radiol Bras. 1996;29:35-9.
  • 9. Macedo-Filho ED. Avaliação endoscópica da deglutição no diagnóstico da disfagia orofaríngea. In: Costa MMB, Lemme EPO, Koch HA, organizadores. Temas em deglutição e disfagia: abordagem multidisciplinar. Rio de Janeiro: Supraset; 1998. p. 77-82.
  • 10. Costa MMB, Canevaro LV, Azevedo ACP. Dosimetric assessment of swallowing examinations with videofluoroscopy. In: International Conference on the Radiological Protection of Patients in Diagnostic and Interventional Radiology, Nuclear Medicine and Radiation Therapy; 2001 Mar; Malaga, Spain.
  • 11. Costa MMB, Canevaro LV, Azevedo ACP, et al. Valores típicos do "produto dose área" (DAP) obtidos durante o estudo videofluoroscópico da deglutição. Radiol Bras. 2003;36:17-20.
  • 12. Costa MMB. Dose referência: quantificação das doses de raios-X necessárias ao estudo videofluoroscópico da deglutição. SIICSalud, Expertos del Iberoamérica; 2004. [acessado em 3 de junho de 2009]. Disponível em: www.siicsalud.com/des/des042/05414002.htm
  • 13. International Commission on Radiological Protection. ICRP Publication 103: The 2007 Recommendations of the International Commission on Radiological Protection. Oxford: Elsevier; 2007.
  • 14. International Commission on Radiological Protection. ICRP Publication 73: Radiological protection and safety in medicine. Oxford: Elsevier; 1997.
  • 15. European Commission. Guidance on diagnostic reference levels (DRLs) for medical exposures. Radiation Protection 109 Report. General Directorate Environment, Nuclear Safety and Civil Protection. Luxembourg: Commission of the European Community; 1999.
  • 16. Faulkner K, Bosmans H, O'Brien R, et al. Optimisation of dose and performance in interventional and digital imaging. Radiat Prot Dosimetry. 2005;117:1-2.
  • 17. Brasil. Ministério da Saúde. Secretaria de Vigilância Sanitária. Diretrizes de proteção radiológica em radiodiagnóstico médico e odontológico. Portaria nş 453, de 1ş de junho de 1998. Diário Oficial da União, Brasília, 2 de junho de 1998.
  • 18. Economides S, Hourdakis CJ, Kalivas N, et al. Image quality evaluation and patient dose assessment of medical fluoroscopic X-ray systems: a national study. Radiat Prot Dosimetry. 2008;129:419-25.
  • 19. Watanabe PCA. Filtração adicional da radiação X em odontologia. Estudo comparativo entre filtros de alumínio e cobre [tese de doutorado]. São Paulo: Universidade de São Paulo; 1999.
  • 20. Nicholson RA, Thornton A, Akpan M. Radiation dose reduction in paediatric fluoroscopy using added filtration. Br J Radiol. 1995;68:296-300.
  • 21. Morrell RE, Rogers AT, Jobling JC, et al. Barium enema: use of increased copper filtration to optimize dose and image quality. Br J Radiol. 2004; 77:116-22.
  • 22. Haiter CFS. Estudo da liga de alumínio-zinco como filtração alternativa para a radiação X [dissertação de mestrado]. Piracicaba: Universidade Estadual de Campinas; 2000.
  • 23. Gonçalves A. Estudo da liga de alumínio-cobre, como filtração alternativa para radiação X [tese de doutorado]. Piracicaba: Universidade Estadual de Campinas; 2000.
  • 24. Bóscolo FN, Gonçalves A, Rollo JMDA, et al. Effects of aluminum-copper alloy filtration on photon spectra, air kerma rate and image contrast. Braz Dent J. 2005;15:214-9.
  • 25. International Atomic Energy Agency. Dosimetry in diagnostic radiology: an international code of practice. Technical Reports Series No. 457. Vienna: International Atomic Energy Agency; 2007.
  • 26. Martin CJ. Radiation dosimetry for diagnostic medical exposures. Radiat Prot Dosimetry. 2008; 128:389-412.
  • 27. Cowen AR, Clarke OF, Coleman NJ, et al. Leeds X-ray test objects instruction manual. Leeds: University of Leeds; 1992.
  • 28. Tapiovaara MJ. Review of relationships between physical measurements and user evaluation of image quality. Radiat Prot Dosimetry. 2008;129:244-8.
  • 29. Cranley K, Gilmore BJ, Fogarty GWA, et al. Catalogue of diagnostic X-ray spectra and other data. IPEM Report 78. SRS-780. York: Institute of Physics and Engineering in Medicine; 1997.
  • 30. Pan American Health Organization. Protección del paciente en radiodiagnóstico: informe del Comité de la CIPR, adoptado por la Comisión en mayo de 1982. Cuaderno Técnico 3. Washington: Pan American Health Organization; 1987.
  • 31. Oliveira ML. Doses de entrada na pele de pacientes em radiologia pediátrica. Expertos de Iberoamérica, abril 2005. [acessado em 22 de junho de 2009]. Disponível em: http://www.siicsalud.com/dato/dat043/05425015.htm
  • 32. Vinhas L. Os raios não conhecem fronteiras. Instituto de Radiodosimetria da Comissão Nacional de Energia Nuclear, em vídeo educativo. NUTES - Núcleo de Tecnologia Educacional para Saúde. Rio de Janeiro: Universidade Federal do Rio de Janeiro; 1990.
  • 33. Magalhães R. Os raios não conhecem fronteiras. Secretaria de Saúde do Estado do Rio de Janeiro, em vídeo educativo. NUTES - Núcleo de Tecnologia Educacional para Saúde. Rio de Janeiro: Universidade Federal do Rio de Janeiro; 1990.
  • 34. Freitas MB, Yoshimura EM. Dose measurements in chest diagnostic X rays: adult and paediatric patients. Radiat Prot Dosimetry. 2004;111:73-6.
  • 35. Travassos LV, Boechat MCB, Santos EM, et al. Avaliação das doses de radiação em uretrocistografia miccional de crianças. Radiol Bras. 2009; 42:21-5.
  • 36. Koch HA. Métodos radiológicos de investigação das doenças disfágicas. In: Costa MMB, Lemme EMO, Koch HA, organizadores. Deglutição e disfagia: abordagem multidisciplinar. Rio de Janeiro: Supraset; 1998. p. 93-8.
  • 37. Nawfel RD, Chan KH, Wagenaar DJ, et al. Evaluation of video gray-scale display. Med Phys. 1992;19:561-7.
  • 38. Caronna G, Santoro G, Testa P. Preference study of peak luminance vs. resolution for color television tubes. SMPTE J. 1998;107:538-43.
  • 39. Nova JLL. Qualidade da imagem. In: Costa MMB, Lemme EMO, Koch HA, organizadores. Deglutição e disfagia: abordagem multidisciplinar. Rio de Janeiro: Supraset; 1998. p. 29-34.
  • 40. Holmes B, Liggins R. A television x-ray image amplifier. SMPTE J. 1965;74:102-4.
  • Endereço para correspondência:
    Dr. Milton M.B. Costa
    Laboratório de Motilidade Digestiva e Imagem, Departamento de Anatomia, ICB/CCS-UFRJ. Cidade Universitária
    Ilha do Fundão. Rio de Janeiro, RJ, Brasil, 21941-590
    E-mail:
  • *
    Trabalho realizado no Laboratório de Motilidade Digestiva e Imagem do Instituto de Ciências Biomédicas do Centro de Ciências da Saúde da Universidade Federal do Rio de Janeiro (ICB/CCS-UFRJ), Rio de Janeiro, RJ, Brasil. Suporte financeiro: Fundação Educacional Charles Darwin.