Quantitation of genistein and genistin in soy dry extracts by UV-Visible spectrophotometric method

Resumo

This paper describes the development and validation of an UV-Visible spectrophotometric method for quantitation of genistein and genistin in soy dry extracts, after reaction with aluminum chloride. The method showed to be linear (r²= 0.9999), precise (R.S.D. < 2%), accurate (recovery of 101.56%) and robust. Seven samples of soy dry extracts were analyzed by the spectrophotometric validated method and by RP-HPLC. Genistein concentrations determined by spectrophotometry (0.63% - 16.05%) were slightly higher than values obtained by HPLC analysis (0.40% - 12.79%); however, the results of both methods showed a strong correlation.

isoflavones; spectrophotometry; RP-HPLC


isoflavones; spectrophotometry; RP-HPLC

ARTIGO

Quantitation of genistein and genistin in soy dry extracts by UV-Visible spectrophotometric method

Isabela da Costa César* * e-mail: isaccesar@bol.com.br ; Fernão Castro Braga; Cristina Duarte Vianna-Soares; Elzíria de Aguiar Nunan; Gérson Antônio Pianetti; Ligia Maria Moreira-Campos

Departamento de Produtos Farmacêuticos, Faculdade de Farmácia, Universidade Federal de Minas Gerais, Av. Pres. Antônio Carlos, 6627, 31270-901 Belo Horizonte - MG, Brasil

ABSTRACT

This paper describes the development and validation of an UV-Visible spectrophotometric method for quantitation of genistein and genistin in soy dry extracts, after reaction with aluminum chloride. The method showed to be linear (r2= 0.9999), precise (R.S.D. < 2%), accurate (recovery of 101.56%) and robust. Seven samples of soy dry extracts were analyzed by the spectrophotometric validated method and by RP-HPLC. Genistein concentrations determined by spectrophotometry (0.63% - 16.05%) were slightly higher than values obtained by HPLC analysis (0.40% - 12.79%); however, the results of both methods showed a strong correlation.

Keywords: isoflavones; spectrophotometry; RP-HPLC.

INTRODUCTION

Phytoestrogens are plant natural products regarded to possess estrogenic effects. Soy isoflavones are the most frequently used phytoestrogens and have received large attention over the last years due to their potential ability to reduce the symptoms associated to menopause.1 Additionally, they are considered effective to prevent chronic pathologies such as osteoporosis, cardiovascular diseases and hormone-related cancers.2 Hence, soy dry extracts have been widely employed as pharmaceutical raw materials for manufacturing isoflavone supplements, used for the treatment of menopausal symptoms and the above mentioned chronic diseases.

Daidzein and genistein are the major isoflavones found in soy extracts, along with the minor product glycitein (Figure1). They also occur naturally in soy as their glycosides, named daidzin, genistin and glycitin, respectively.3 When orally administered, isoflavones glycosides undergo enzymatic hydrolysis in the small intestine, releasing the aglycones, the moieties which are absorbed and present biologic effects.4In vitro studies for these aglycones have demonstrated that genistein exhibits the highest affinity for estrogenic receptors,5 being approximately ten times more active in vivo than daidzein.6 Therefore, genistein can be accounted as the main isoflavone responsible for the biological activity of soy extracts.


Many methods reported in literature for the analysis of soy isoflavones are based on high performance liquid chromatography (HPLC)7-10 and gas chromatography-mass spectrometry (GC-MS).11-14 They have been developed to quantify several isoflavones simultaneously, however, involve laborious processes of sample preparation, separation, detection and data analyses. We have recently developed and validated a RP-HPLC method for quantifying isoflavone aglycones released after acid hydrolysis and demonstrated its application for quantifying isoflavone aglycones in the analysis of soy dry extract samples used as pharmaceutical raw material.15

Some methods reported in literature16 and pharmacopoeias17,18 to the assay of flavonoids are based on the UV absorbance determination of the resulting complex with aluminum chloride (AlCl3). Flavonoids containing 5-hydroxy-4-keto, 3-hydroxy-4-keto or o-dihydroxyl systems are able to chelate with AlCl3 and the reaction is disclosed by a bathochromic shift of the bands in the UV-Visible spectrum.19,20 Among the major isoflavones found in soy extracts, only genistein and its glycoside, genistin, present the structural feature required to react with AlCl3. As far as we know, no spectrophotometric method for the quantitative assay of genistein or other isoflavones has been reported to date.

The increasing consumption of soy extract as pharmaceutical products and supplements demands the development of alternative methods for the routine analysis of soy isoflavones. Within this context, the aim of this study was to develop and validate a simple and rapid UV-Visible method to quantify genistein and its glycoside genistin in soy dry extracts, after reaction with AlCl3. Samples of soy dry extract were analyzed by the developed method and the results obtained were statistically compared with those previously determined by RP-HPLC.15

EXPERIMENTAL

Reagents and materials

Genistein reference compound was purchased from ChromaDex (Santa Ana, CA, USA). Seven samples of soy dry extracts standardized to contain 40% (w/w) of total isoflavones, from six different suppliers, were kindly donated by compounding pharmacies from Belo Horizonte, MG, Brazil. Two batches of supplier 2 were assayed. Methanol and acetic acid were purchased from Merck (Darmstadt, Germany) and aluminum chloride (analytical reagent grade) was obtained from Sigma (St. Louis, MO, USA).

Instrumental and analytical conditions

UV-Visible spectrophotometric analyses were carried out on a Shimadzu UV160A spectrophotometer. After reaction with AlCl3, the UV-Visible spectrum of genistein and soy dry extract solutions were recorded in the range of 200 to 500 nm. In the spectral analysis, the wavelength 382 nm was defined for the quantitation of genistein and genistin in soy dry extract samples. The time interval for recording the UV spectrum after complex formation was also investigated to determinate the optimal time for reaction and to evaluate the complex stability. Hence, spectrophotometric determination of soy dry extract solutions was carried out immediately after the AlCl3 addition and at regular intervals of 10 min, until 240 min. The interval time of 10 min was defined for measuring absorbance of solutions after AlCl3 addition.

The HPLC analyses were carried out on an HP1100 system Agilent (Santa Clara, CA, USA), composed of quaternary pump, auto sampler, photodiode array detector (DAD) and HP ChemStation software. The chromatographic conditions employed were previously described by César et al..15 A C18 endcapped Lichrospher column (250 x 4.6 mm I.D.; 5 µm particle size) was employed (Merck, Darmstadt, Germany), at 30 °C. The mobile phase was composed of 0.1% acetic acid and methanol (52:48), at a flow rate of 1.0 mL/min. The injection volume was 20 µL and UV-detection was performed at 254 nm.

Preparation of solutions for spectrophotometric analysis

Genistein standard solution: approximately 15 mg of genistein reference compound were accurately weighed in a 100 mL volumetric flask. Methanol (80 mL) was added, the solution was sonicated for 5 min and volume was adjusted to 100 mL with methanol. An aliquot (5 mL) was transferred to a volumetric flask (25 mL) and 1 mL of AlCl3 2% (w/v) in methanol was added. The flask volume was completed with methanol, to obtain a solution at 30 µg/mL.

Soy dry extract sample solution: approximately 50 mg of soy dry extract sample, previously dried at 105 °C for 4 h, were accurately weighed in a 50 mL volumetric flask. Methanol (40 mL) was added and the solution was sonicated for 5 min. The volume was adjusted to 50 mL with methanol and filtered. An aliquot of the solution (5 mL) was transferred to a volumetric flask (25 mL) and 1 mL of AlCl3 2% (w/v) in methanol was added. The flask volume was completed with methanol.

Blank solutions for genistein standard and soy dry extract sample solutions were similarly prepared, without the addition of AlCl3.

Validation of the spectrophotometric method

The optimized method was validated according to ICH guidelines for the validation of analytical methods.21

Linearity

Three replicates of genistein standard solutions at 150 µg/mL were diluted in methanol to six different concentrations (1.5; 6; 18; 30; 42 and 54 µg/mL). An aliquot (1 mL) of AlCl3 2% (w/v) in methanol was added to each diluted solution. A calibration curve for concentration versus absorbance was plotted and the obtained data were subjected to regression analysis using the Least Square Method.

Precision

The intra-day precision was evaluated by analyzing six replicates of soy dry extract solutions (n=6), at 100% of the test concentration (30 µg/mL). Similarly, the inter-day precision was evaluated in two consecutive days (n=12). The concentration of genistein in soy dry extracts was determined after reaction with AlCl3, and the relative standard deviation (R.S.D.) was calculated.

Accuracy

Standard solutions of genistein, at three different concentration levels, were added to soy dry extracts samples before reaction with AlCl3. At each level, solutions were prepared in triplicate and the recovery percentage was determined.

Specificity

It was evaluated regarding the presence of other major isoflavones (daidzein and glycitein) in soy dry extracts. Solutions of those isoflavones were prepared at 30 µg/mL, following addition of 1 mL of AlCl3 2% (w/v) in methanol. The UV-Visible spectra were recorded in the range of 200 to 500 nm, to evaluate the presence of absorption band at 382 nm, indicative of complex formation.

Robustness

Six sample solutions were prepared and analyzed using the established conditions and by variation of the following analytical parameters: added volume of AlCl3 solution (0.5-1.5 mL) and two methanol suppliers. Genistein contents and R.S.D. were determined for each condition. The obtained data were submitted to statistical analysis (Student-t test) at 0.05 significance level.

Detection and quantitation limits

Diluted genistein standard solutions were analyzed, at decreasing concentrations, in the range of 1.00 to 0.10 µg/mL. To assess the limit of detection, absorbance values and UV-Visible spectra were evaluated, whereas the quantitation limit was settled as the lower concentration which provided responses with adequate linearity and precision (R.S.D.< 2.0%). The detection and quantitation limits were also calculated by the standard deviation and the slope of the calibration curve and the values were compared with those obtained by the response of the diluted solutions.

Analysis of soy dry extracts samples

Seven commercial samples of soy dry extract were analyzed using the developed spectrophotometric method and the RP-HPLC method described by César et al..15 Samples solutions were prepared in quintuplicate for the spectrophotometric analyses, as described in Soy dry extract sample solution. Solution absorbances were recorded at 382 nm, 10 min after AlCl3 addition, employing the correspondent blank solutions as reference. For the chromatographic analyses, 50 mg of soy dry extract were accurately weighed in a 100 mL volumetric flask, in quintuplicate, following the addition of 3.0 M HCl in ethanol (80 mL). The solution was sonicated for 5 min and submitted to steam bath for 40 min. The flask volume was completed, and 5 mL were transferred to a 25 mL volumetric flask. The volume was completed with mobile phase and 20 µL were injected onto the chromatograph.

RESULTS AND DISCUSSION

The UV-Visible spectrum recorded for a solution of genistein after reaction with AlCl3 showed an intense absorption band with maximum wavelength at 382 nm, which was not found in the UV-Visible spectrum of genistein alone (Figure 2). As expected, this absorption band indicates complex formation between genistein and AlCl3. A similar absorption band was present in the UV-Visible spectrum of a soy dry extract solution, after reaction with AlCl3, attesting the presence of genistein and its glycoside in the analyzed sample.


The time interval for spectrophotometric detection of the complex was evaluated. The UV absorbance of genistein/AlCl3 product reaches its maximum after 5 min of reaction. Its stability was observed to last for 240 min, since no significant variation in absorption was detected during this period. Based on these findings, a period of 10 min after AlCl3 addition was defined as the optimal detection time for quantitation of genistein and genistin in soy dry extract samples.

Validation of the spectrophotometric method

Linearity

A linear correlation (r2 = 0.9999) was found between the absorbance values and the concentration of genistein solutions, after reaction with AlCl3, in the assayed range. The regression analysis data are shown in Table 1. The significance of the intercept obtained in the calibration curve was tested and this parameter was not statistically significant (p>0.05). Hence, it can be considered that the curve passes through the origin. The low R.S.D. value (0.76%) indicated the high precision of the calibration curve.

Precision

In the intra-day precision analyses (n=6), the mean content of genistein and genistin, expressed as genistein, was 15.83% (R.S.D.=0.65%). For the inter-day precision (n=12), the obtained mean was 15.95% (R.S.D.=1.16%). R.S.D. values lower than 2.0% attested the precision of the method.

Accuracy

It was investigated by means of a standard addition experiment, at three concentration levels in triplicate (n=9). The mean recovery of 101.56% (R.S.D. = 0.35%) assured the method accuracy.

Specificity

No absorption band was found at 382 nm in the UV-Visible spectra of daidzein and glycitein, recorded after addition of AlCl3. The absence of chelatogenic systems in those compounds disables the reaction with AlCl3. Therefore, the method showed adequate specificity for genistein and its glycoside, regarding the other major isoflavones (daidzein and glycitein) in soy dry extracts samples.

Robustness

Statistical analysis showed no significant difference between results obtained employing analytical conditions established for the method and for experiments in which variations were introduced. Thus, the method showed to be robust for variable volumes of AlCl3 added, in the range of 0.5 to 1.5 mL, as well as for two different suppliers of methanol used.

Detection and quantitation limits

The detection limit calculated using parameters of the calibration curve was 0.25 µg/mL. In the spectrophotometric analysis of diluted solutions, it was possible to identify the absorption band at 382 nm in the concentration 0.15 µg/mL, with a corresponding absorbance value of 0.003A. The quantitation limit determined based on the parameters of the calibration curve was 0.76 µg/mL. However, a genistein standard solution at 0.40 µg/mL provided absorbance values around 0.006 A, with adequate precision (R.S.D.<2.0%) and linearity and therefore, this value can be considered the quantitation limit of the method. Hence, detection and quantitation limits calculated by the parameters of the calibration curve were close, but slightly higher than the values obtained by the spectrophotometric analysis of diluted solutions.

Analysis of soy dry extract samples

Commercial samples of soy dry extracts employed as pharmaceutical raw material, standardized to contain 40% (w/w) of total isoflavones, were analyzed by both spectrophotometric and chromatographic methods. The results obtained by the spectrophotometric method correspond to both genistein and genistin contents, expressed as genistein. In the RP-HPLC analysis, the results are expressed as genistein content after hydrolysis, corresponding to the amounts of genistein and its hydrolyzed glycoside, genistin (Table 2).

Data on Table 2 indicate higher contents of genistein and genistin determined by spectrophotometry in comparison to the concentrations assayed by the RP-HPLC method. Analysis of the obtained results by Student's t-test indicated statistically significant difference between genistein contents determined by spectrophotometry and RP-HPLC, for all assayed samples (p<0.05). However, a clear proportionality between the contents obtained by both methods can be observed in the analyzed samples. The different sample preparation procedures, including solvents and temperature, might contribute to this observed content variation between the UV-Vis and RP-HPLC methods. Besides, the content specifications are not similar; the result of the spectrophotometric method is expressed as genistein content, corresponding to the total content of genistein and genistin, while the results of the chromatographic method are expressed as genistein content after hydrolysis. Hence, these differences regarding the two method specifications may interfere in a direct comparison of the results and demonstrate the necessity of considering distinct specifications for each method.

The developed UV-Visible method showed to be selective regarding the other isoflavones present in higher amounts in soy dry extracts (daidzein and glycitein). However, such extracts may contain other flavonoids and phenylpropanoids in lower concentrations and we cannot discharge the potential interference of those compounds in the detection by the spectrophotometric method, making it less specific when compared to RP-HPLC. On the other hand, the spectrophotometric method is more economical, in terms of cost and time, than the HPLC method.

The analyzed extracts are standardized to contain 40% (w/w) of total isoflavones; however, the contents of individual isoflavones are not controlled and may fluctuate due to the influence of several factors, including soy variety, period of crop collection and geographical location.4 Hence, the analyzed samples showed high variations in genistein contents assayed in both methods. Considering that differences in biological effects of individual isoflavones are well documented, and genistein is supposed to be the most active compound,22 such variations might interfere in the efficacy of soy extracts based treatments, whenever reduced amounts of genistein are found.

CONCLUSIONS

Reaction with aluminum chloride followed by UV-Visible detection was demonstrated to be a simple and suitable technique for quantitation of genistein and genistin in soy dry extract samples. The developed method complied with all validation parameters and showed to be feasible and of low cost. Hence, it can be applied by compounding pharmacies and small laboratories in routine quality control analyses or may be used as a preliminary assay to evaluate the genistein content in soy extracts.

ACKNOWLEDGEMENTS

The authors thank FAPEMIG, for financial support.

Recebido em 4/5/07; aceito em 1/7/08; publicado na web em 5/11/08

  • 1. Howes, L. G.; Howes, J. B.; Knight, B. C.; Maturitas 2006, 55, 203.
  • 2. Anderson, J. J. B.; Anthony, M. S.; Cline, J. M.; Washburn, S. A.; Garner, S. C.; Public Health Nutr. 1999, 2, 489.
  • 3. Duncan, A. M.; Phipps, W. R.; Kurzer, M. S.; Best Pract. Res. Clin. Obstet. Gynecol 2003, 17, 253.
  • 4. Dixon, R. A.; Annual Rev. Plant Biol. 2004, 55, 225.
  • 5. Wober, J.; Weisswange, I.; Vollmer, G.; J. Steroid Biochem. Mol. Biol. 2002, 83, 227.
  • 6. Branham, W. S.; Dial, S. L.; Moland, C. L.; J. Nutr. 2002, 132, 658.
  • 7. Delmonte, P.; Perry, J.; Rader, J. I.; J. Chromatogr., A 2006, 1107, 59.
  • 8. Chen, L. J.; Zhao, X.; Plummer, S.; Tang, J.; Games, D. E.; J. Chromatogr., A 2005, 1082, 60.
  • 9. Klejdus, B.; Mikelová, R.; Petrlová, J.; Potesil, D.; Adam, V.; Stiborová, M.; Hodek, P.; Vacek, J.; Kizek, R.; Kubán, V.; J. Agric. Food Chem. 2005, 53, 5848.
  • 10. Penalvo, J. L.; Nurmi, T.; Adlercreutz, H.; Food Chem. 2004, 87, 297.
  • 11. Thompson, L. U.; Boucher, B. A.; Liu, Z.; Cotterchio, M.; Kreiger, N.; Nutr. Cancer 2006, 54, 184.
  • 12. Kuo, H. W.; Ding, W. H.; J. Chromatogr., A 2004, 1027, 67.
  • 13. Grace, P. B.; Taylor, J. I.; Botting, N. P.; Fryatt, T.; Oldfield, M. F.; Bingham, S. A.; Anal. Biochem. 2003, 315, 114.
  • 14. Wang, C. C.; Prasain, J. K.; Barnes, S.; J. Chromatogr., B: Anal. Technol. Biomed. Life Sci. 2002, 777, 3.
  • 15. César, I. C.; Braga, F. C.; Vianna-Soares, C. D.; Nunan, E. A.; Pianetti, G. A.; Condessa, F. A.; Barbosa, T. A. F.; Moreira-Campos, L. M.; J. Chromatogr., B: Anal. Technol. Biomed. Life Sci. 2006, 836, 74.
  • 16. Kosalec, I.; Bakmaz, M.; Pepeljnjak, S.; Vladimir-Knezevic, S.; Acta Pharm. 2004, 54, 65.
  • 17
    British Pharmacopoeia, 5th ed., Her Majesty's Stationary Office: London, 2005.
  • 18
    Farmacopéia Brasileira, 4ª ed., Atheneu: São Paulo, 1988.
  • 19. Mabry, T. J.; Markham, K. R.; Thomas, M. B.; The systematic identification of flavonoids, Springer-Verlag: New York, 1970.
  • 20. Harbone, J. B.; Mabry, T. J.; Mabry, H.; The Flavonoids, Chapman and Hall: London, 1975.
  • 21
    Validation of Analytical Procedures: Methodology - ICH Harmonized Tripartite Guideline; International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human Use, 1996.
  • 22. Kuiper, G. G. J. M.; Lemmen, J. G.; Carlsson, B.; Corton, J. C.; Safe, S. H.; Van Der Saag, P. T.; Van Der Burg, B.; Gustafsson, J.; Endocrinology 1998, 139, 4252.
  • *
    e-mail:
  • Datas de Publicação

    • Publicação nesta coleção
      16 Jan 2009
    • Data do Fascículo
      2008

    Histórico

    • Aceito
      01 Jul 2008