CROMATOGRAFIA POR INTERAÇÕES HIDROFILICAS (HILIC): ESTADO DA ARTE E APLICAÇÕES

HYDROPHILIC INTERACTION CHROMATOGRAPHY (HILIC): STATE OF THE ART AND APPLICATIONS

Resumo

Hydrophilic interaction liquid chromatography (HILIC) has been gaining increased attention for its effective separation of highly polar compounds, which include carbohydrates, amino acids, pharmaceutical compounds, proteins, glycoproteins, nucleosides, etc. Polar compounds are usually poorly retained on reverse-phase liquid chromatography (RP-HPLC) columns or have poor solubility in the apolar mobile phase of normal-phase high performance liquid chromatography (NP-HPLC). Since HILIC uses organic solvents such as ACN or MeOH ( > 70%), also used in RP-HPLC and polar stationary phases similar to NP-HPLC (bare silica, diol, amino, amide, saccharide, zwitterionic stationary phases, etc.), it represents a hybrid of the two separation modes. The high organic content in the MP leads to good compatibility with mass spectrometry (MS), increasing the detectivity. This review describes the fundamentals of HILIC and highlights some interesting applications.

Keywords:
hydrophilic interaction chromatography; separation of polar compounds; HILIC separations; high performance liquid chromatography; polar stationary phases


Keywords:
hydrophilic interaction chromatography; separation of polar compounds; HILIC separations; high performance liquid chromatography; polar stationary phases

INTRODUÇÃO

A cromatografia líquida no modo fase reversa (RP-HPLC, do inglês reversed phase high performance liquid chromatography), onde se emprega uma fase estacionária (SP, do inglês stationary phase) mais apolar que a fase móvel (MP, do inglês mobile phase) é a técnica analítica de separação mais utilizada.11 Maldaner, L.; Collins, C. H.; Jardim, I. C. S. F.; Quim. Nova2010 , 33, 1559. Porém, uma das limitações do uso da fase reversa é a baixa retenção de moléculas polares ou iônicas. O uso da cromatografia líquida no modo de fase normal (NP-HPLC, do inglêsnormal phase high performance liquid chromatography), na qual se emprega SP mais polares que as fases móveis, pode solucionar essa limitação. No entanto, a solubilidade de moléculas polares em MP apolares não aquosas é limitada, restringindo a aplicabilidade desta. E em alguns casos, a cromatografia de íons não pode ser aplicada devido a ausência de cargas dos solutos. Para minimizar esses problemas, na década de 90, Alpert22 Alpert, A. J.; J. Chromatogr.1990 , 499, 177. introduziu a cromatografia por interações hidrofílicas (HILIC, do ingles hydrophilic interaction chromatography). A HILIC é uma modalidade de separação na qual é empregada uma SP mais polar do que em RP-HPLC e MP com pouca concentração de água ou tampão, sendo normalmente > 70% em solvente orgânico, tipicamente acetonitrila.33 Gama, M. R.; Silva, R. G. C.; Collins, C. H. C.; Bottoli, C. B. G.;Trends Anal. Chem.2012 , 37, 48.

Dessa forma, HILIC apresenta modo de separação ortogonal ao modo em RP-HPLC, pois os mecanismos de separação do modo HILIC e do modo RP são independentes entre si. Com isso há a possibilidade de se explorar diferentes perfis de seletividade, o que permite a sua aplicação em cromatografia líquida bidimensional abrangente (LC × LC), com a vantagem de apresentar menor incompatibilidade do eluente da primeira dimensão (1D) com o eluente da segunda dimensão (2D), quando comparado a um sistema ortogonal com NP-HPLC, podendo ser utilizada no modo de duas dimensões para separação de amostras complexas na qual a amostra é submetida a dois mecanismos de separação diferentes e complementares com a finalidade de fornecer um maior poder de separação e distribuição dos analitos nas duas dimensões, aumentando o poder de resolução.44 Mogollón, G. S. N.; Lima, P. F.; Gama, M. R.; Furlan, M. F.; Braga, S. C. G. N.; Prata, P. S.; Jardim, I. C. S. F.; Augusto, F.; Quim. Nova2014 , 37, 1680.

Além disso, seu uso com detectores de espectrometria de massas, devido à alta proporção de solvente orgânico na MP,55 Lin, H.; Ou, J.; Zhang, Z.; Dong, J.; Wu, M.; Zou, H.; Anal. Chem. 2012 , 84, 2721.

6 Buszewski, B.; Noga, S.; Anal. Bioanal. Chem.2012 , 402, 231.

7 Cubbon, S.; Antonio, C.; Wilson, J.; Thomas-Oates, J.; Mass Spectrom. Rev.2010 , 29, 671.

8 Spagou, K.; Tsoukali, H.; Raikos, N.; Gika, H.; Wilson, I. D.; Theodoridis, G.; J. Sep. Sci.2010 , 33, 716.
-99 Xu, R. N.; Rieser M. J.; El-Shourbagy, T. A.;Bioanalysis2009 , 1, 239. vem possibilitando um incremento anual de trabalhos utilizando essa modalidade de cromatografia (Figura 1), principalmente devido à necessidade crescente de separação de compostos altamente polares (com valores negativos delog P: coeficiente de partição octanol-água) em análises clínicas e de compostos farmacêuticos, nas áreas da metabolômica, proteômica, glicômica, etc. (Figura 2).

Figura 1
Número de publicações científicas obtidas através da busca do termo “hydrophilic interaction chromatography” no portal ISI/Web of Science de 1990 até 2014
Figura 2
Exemplo de seleção do tipo de modalidade cromatográfica em função da polaridade dos analitos medida pelo log P

As SP utilizadas em HILIC1010 Zhang, F.; Shen, G.; Ji, S.; Yang, B.; J. Liq. Chromatogr. Relat. Technol.2015 , 38, 349. são fases polares tradicionalmente conhecidas como a sílica pura,1111 Naidong, W.; Shou, W.; Chen, L.Y.; Jiang, X. L.; J. Chromatogr. B2001 , 754, 387.,1212 Dejaegher, B.; Heyden, Y. V.; J. Sep. Sci.2010 , 33, 698. sílica modificada1313 Jandera, P.; Anal. Chim. Acta2011 , 692, 1. com grupos triazóis, cianos, aminos,1414 Oyler, A. R.; Armstrong, B. L.; Cha, J. Y.; Zhou, M. X.; Yang, Q.; Robison, R. I.; Dunphy, R.; Burinsky, D. J.; J. Chromatogr. A1996 , 724, 378.,1515 Tolstikov, V.V.; Fiehn, O.; Anal. Biochem.2002 , 301, 298. amidas, ácidos sulfônicos, carboxilatos, diols,1616 Jandera, P.; Hajek, T.; J. Sep. Sci.2009 , 32, 3603. poliidroxietilaspartamida,22 Alpert, A. J.; J. Chromatogr.1990 , 499, 177. e SP à base de ciclodextrinas,1616 Jandera, P.; Hajek, T.; J. Sep. Sci.2009 , 32, 3603.,1717 Armstrong, D. W.; Jin, H. L.; J. Chromatogr.1989 , 462, 219. ou polímeros,33 Gama, M. R.; Silva, R. G. C.; Collins, C. H. C.; Bottoli, C. B. G.;Trends Anal. Chem.2012 , 37, 48. além de novos materiais como compostos zwiteriônicos, os quais apresentam cargas positivas e negativas em sua estrutura.1818 Kawachi, Y.; Ikegami, T.; Takubo, H.; Ikegami, Y.; Miyamoto, M.; Tanaka, N.; J. Chromatogr. A2011 , 1218, 5903.

19 Wang, C.; Jiang, C.; Armstrong, D. W.; J. Sep. Sci.2008 , 31, 1980.
-2020 Qiao, L.; Dou, A.; Shi, X.; Li, H.; Shan, Y.; Lu, X.; Xu, G.;J. Chromatogr. A2013 , 1286, 137. Mais recentemente SPs à base de líquidos iônicos2121 Pino, V.; Afonso, M. A.; Anal. Chim. Acta2012 , 714, 20.,2222 Qiao, L.; Li, H.; Shan, Y.; Wang, S.; Shi, X.; Lu, X.; Xu, G.;J. Chromatogr. A2014 , 1330, 40. vêm sendo utilizadas, além de SP com diversas funcionalidades preparadas por reações do tipoclick chemistry.2323 Guo, Z. M.; Lei, A. W.; Zhang, Y. P.; Xu, Q.; Xue, X. Y.; Zhang, F. F.; Liang, X. M.; Chem. Commun.2007 , 2491.,2424 Fu, Q.; Guo, Z.; Liang, T.; Zhang, X.; Xu, Q.; Liang, X.;Anal. Methods2010 , 2, 217.

A utilização do modo de separação HILIC, também denominada na época de seu surgimento de cromatografia líquida de "fase reversa-reversa" ou cromatografia "aquosa de fase normal", propicia algumas vantagens quando comparada a outros modos de separação como a fase normal ou fase reversa:2525 Ikegami, T.; Tomomatsu, K.; Takubo, K.; Horie, K.; Tanaka, N.;J. Chromatogr A2008 , 474, 1184.

26 Pesek, J. J., Matyska, M. T.; In Hydrophilic Interaction Liquid Chromatography (HILIC) and Advanced Applications; Wang, P. G.; He, W., eds.; CRC Press: Boca Raton, 2011, ch. 2.

27 Weng, N.; J. Chromatogr. B2003 , 796, 209.
-2828 Weng, N.; Shou, W.; Chen, Y.; Jiang, X.; J. Chromatogr. B2001 , 754, 387.

  1. Alta concentração de solvente orgânico na MP, o que diminui a pressão do sistema visto que a viscosidade da MP também dimimui. A alta concentração de solvente também aumenta a detectabilidade quando se utiliza o detector de espectrometria de massas, uma vez que o processo de ionização dos analitos é favorecido;

  2. Maiores retenções e eficiências de separação de analitos polares com baixa retenção no modo em fase reversa, como peptídeos e ácidos nucleicos, sem a necessidade de utilização de aditivos tipo par-iônico;

  3. Maior preservação da SP pelo uso de menores quantidades de água na MP, evitando o colapso de sua superfície, principalmente em fases estacionárias com caráter mais hidrofóbico;

  4. Compatibilidade com técnicas de extração que utilizam altas concentrações de solvente orgânico, como a extração em fase sólida (SPE, do inglêssolid phase extraction) ou extração líquido-líquido, o que facilita e simplifica a etapa de preparação de amostra.

HILIC também pode ser conhecida por "cromatografia de fase normal aquosa" uma vez que essa modalidade de separação envolve uma fase estacionária polar similar a algumas das SP utilizado em cromatografia com fase normal, e a porção aquosa na fase móvel atua como o solvente forte, de forma oposta ao que ocorre na cromatografia líquida em fase reversa.2929 Hemstrom, P.; Irgum, K.; J. Sep. Sci.2006 , 29, 1784. Por outro lado, HILIC pode ser relacionada ao modo de separação em fase normal de compostos altamente polares ou iônicos, mas sob condições típicas da fase reversa. Dessa forma, muitos autores descrevem essa modalidade como um eixo de conexão entre a fase normal (NP) e a fase reversa (RP) (Figura 3).

Figura 3
Relação entre HILIC e as demais modalidades principais de cromatografia líquida. Adaptado da ref. 6

BREVE HISTÓRICO

Em 1951 Gregor et al.3030 Gregor, H. P.; Collins, F. C.; Pope, M.; J. Colloid Sci. 1951 , 6, 304.descreveram as interações envolvendo a camada de água presente em uma SP a base de resina trocadora de íons. Os trabalhos de Samuelson e Sjöström3131 Samuelson, O.; Sjöström, E.; Sven. Kem. Tidskr.1952 , 64, 305. e Rückert e Samuelson3232 Rückert, H.; Samuelson, O.; Sven. Kem. Tidskr.1954 , 66, 337. também relacionaram interações entre os analitos e a camada de água presente na SP trocadora de íons. Um pouco depois, Linden e Lawhead,3333 Linden, J. C.; Lawhead, C. L.; J. Chromatogr.1975 , 105, 125. em 1975, utilizaram colunas polares com SP trocadora de ânions na separação de sacarídeos com uma MP com elevada concentração de solvente orgânico e Palmer,3434 Palmer, J. K.; Anal. Lett. 1975 ,8, 215. no mesmo ano, utilizou uma coluna à base de carboidrato para separar açucares com uma MP com alta concentração de acetonitrila (ACN).

Sequencialmente, em 1982 e 1985, Verhaar e Kuster3535 Verhaar, L. A. T.; Kuster, B. F. N.; J. Chromatogr.1982 , 234, 57. e Nikolov e Reilly,3636 Nikolov, Z. L.; Reilly, P. J.; J. Chromatogr.1985 , 325, 287.respectivamente, separaram misturas de açucares utilizando colunas polares com grupos amino baseadas em suporte de sílica, utilizando uma mistura de água/acetonitrila como MP.

No entanto, a utilização do termo e do acrônimo HILIC é atribuída a Alpert22 Alpert, A. J.; J. Chromatogr.1990 , 499, 177. no seu trabalho clássico de 1990 sobre essa modalidade de separação em cromatografia líquida, no qual foram separados peptídeos, ácidos nucleicos e outros compostos polares. Este autor propôs o termo como uma variação da cromatografia em fase normal quando utilizou MP com alta concentração de solvente orgânico e mecanismo de retenção baseado na interação hidrofílica entre os analitos polares e a camada de água superficialmente imobilizada na SP também polar.

MECANISMOS DE SEPARAÇÃO

Em HILIC a MP utilizada é similar à utilizada no modo em RP, descrita como uma mistura de água/solvente orgânico, tipicamente ACN, porém, diferentemente da RP, com alta concentração de solvente orgânico e pouca concentração de água (ou tampão aquoso) (5-40%).33 Gama, M. R.; Silva, R. G. C.; Collins, C. H. C.; Bottoli, C. B. G.;Trends Anal. Chem.2012 , 37, 48.,66 Buszewski, B.; Noga, S.; Anal. Bioanal. Chem.2012 , 402, 231.,2626 Pesek, J. J., Matyska, M. T.; In Hydrophilic Interaction Liquid Chromatography (HILIC) and Advanced Applications; Wang, P. G.; He, W., eds.; CRC Press: Boca Raton, 2011, ch. 2.,2929 Hemstrom, P.; Irgum, K.; J. Sep. Sci.2006 , 29, 1784. Assim como o modo em NP, HILIC utiliza SP mais polares que facilmente apresentam uma camada de água superficialmente imobilizada ou adsorvida e que retém solutos igualmente polares por meio de interações hidrofílicas por meio de um equilíbrio de partição. Da mesma forma, acredita-se que podem ocorrer também interações do tipo ligação de hidrogênio entre os grupos funcionais polares dos analitos e da SP2929 Hemstrom, P.; Irgum, K.; J. Sep. Sci.2006 , 29, 1784.,3737 Yoshida, T.; J. Biochem. Biophys.Methods2004 , 60, 265.

38 Yoshida, T.; J. Chromatogr. A1998 , 811, 61.

39 Bell, D. S.; LCGC North Am.2015 , 33, 90.
-4040 Liang, T.; Fu, Q.; Shen, A.; Wang, J.; Xin, H.; Guo, K. Y.; Liang, X.; J. Chromatogr. A2015 , 1388, 110. e interações eletrostáticas entre a SP e grupos funcionais ionizados dos analitos (Figura 4). Adicionalmente, podem também ocorrer interações do tipo forças de van der Waals entre a porção hidrofóbica da SP e grupos não polares dos analitos,33 Gama, M. R.; Silva, R. G. C.; Collins, C. H. C.; Bottoli, C. B. G.;Trends Anal. Chem.2012 , 37, 48. e adsorção dos analitos na superfície do adsorvente.

Figura 4
Tipos de interações presentes na cromatografia por interações hidrofílicas (HILIC): partição hidrofílica dos solutos da amostra entre a fase móvel (orgânica) e a camada aquosa sobre a superfície da fase estacionária; interação entre os átomos polares presentes nos solutos e a superfície polar da fase estacionária (interações por ligações de hidrogênio), interações eletrostáticas e hidrofóbicas. Na ilustração o soluto está representado pela molécula da citosina; a SP é a sílica pura em (a) e, em (b) do tipo zwitteriônica a base de sílica funcionalizada com grupos sulfobetaína. A MP utilizada foi ACN/H2O

Como resultado desse mecanismo triplo de interação, na HILIC há a inversão na ordem de retenção dos compostos, em que os compostos mais polares são os mais retidos em comparação com a RP, na qual os compostos polares são os menos retidos (Figura 5).

Figura 5
Separação de uma mistura de saponinas derivadas de triterpenos em HILIC e em RP. Detecção UV a 210 nm. Fase móvel: H2O (A) e ACN (B). Gradiente HILIC: 0-10 min, A:B (15:85, v/v)→A:B (20:80, v/v), 10-20 min, A:B (20:80, v/v)→A:B (35:65, v/v). Gradiente RP: 0-20 min, A:B (80:20, v/v)→A:B (70:30, v/v), 20-35 min, A:B (70:30, v/v)→A:B (50/50, v/v). Coluna cromatográfica Click TE-RA (a base de mercapto-sílica modificada com rebaudioside A (150 mm × 4.6 mm i.d., 5μm), labmade). Analitos: (1) ginsenoside Mc, (2) gypenoside XIII, (3) gin-senoside Rg3, (4) notoginsenoside Fe, (5) notoginsenoside Fd, (6) ginsenoside Rd ,(7) ginsenoside Rc, (8) notoginsenoside Fa. Adaptado da Ref. 40

FASES ESTACIONÁRIAS

Atualmente, o número de SP disponíveis para HILIC é significativo e vêm crescendo a cada dia. Há no mercado de colunas cromatográficas SP especialmente preparadas para essa modalidade cromatográfica, com diversos grupos funcionais que propiciam a retenção e seletividade adequadas para analitos polares de acordo com o tipo de aplicação a ser desenvolvida. A priori, uma SP tipo HILIC deve conter grupos polares para interação de analitos com polaridade semelhante.4141 Garcia-Gomez, D.; Rodriguez-Gonzalo, E.; Carabias-Martinez, R.;Trends Anal. Chem.2013 , 47, 111. Ainda hoje, as SP mais utilizadas são as baseadas em sílica, incluindo sílica híbrida, partículas superficialmente porosas sub-2 µm e sílica monolítica4242 McCalley, D. V.; J. Chromatogr. A2010 , 1217, 3408.

43 Chen, Y.; Wang, K.; Liu, Y.; Yang, H.; Yao, S.; Chen, B.; Nie, L.; Xu, G. M.; Electrophoresis2013 , 34, 1877.
-4444 Li, R.; Yuan, Q.; Huang, Y.; Chin. J. Chromatogr.2014 , 32, 675. ou suportes de sílica modificados quimicamente com grupamentos neutros, tais como ciano, diol, amida, alquilamida, etc., ou carregados (amina, grupos sulfônicos) ou, até, com grupos zwitteriônicos (ZIC), que tem demonstrado bom desempenho para retenção de compostos polares iônicos ou não-iônicos.66 Buszewski, B.; Noga, S.; Anal. Bioanal. Chem.2012 , 402, 231.,4545 Zhang, F.; Shen, G.; Ji, S.; Yang, B.; J. Liq. Chromatogr. Relat. Technol.2015 , 38, 349.

Fases estacionárias produzidas por " click chemistry"

As tendências dos últimos anos na síntese e preparação de SP para HILIC incluem as reações por click chemistry,4646 Chu, C. H.; Liu, R. H.; Chem. Soc. Rev.2011 , 40, 2177.,4747 Kolb, H. C.; Finn, M. G.; Sharpless, B.; Angew. Chem. Int. Ed.2001 , 40, 2004. o que representa preparar SP por meio de reações entre grupos funcionais que reagem de forma rápida, seletiva (daí o termo do inglês click) e que estaria relacionado ao "encaixe" entre as moléculas de reagentes em reações que incluem duas etapas, com altos rendimentos, utilizando condições reacionais brandas. A click chemistry também é utilizada para síntese e preparo de compostos ou materiais nas áreas de biociências,4848 Grammel, M.; Hang, H. C.; Nat. Chem. Biol.2013 , 40, 2004.,4949 Xie, R.; Hong, S.; Chen, X.; Curr. Opin. Chem. Biol.2013 , 17, 747.farmacêutica5050 Kolb, H. C.; Sharpless, B.; Drug Discovery Today2003 , 8, 1128.,5151 Thirumurugan, P.; Matosiuk, D.; Jozwiak, K.; Chem. Rev.2013 , 113, 4905. e de materiais.5252 Xi, W.; Scott, T. F.; Kloxin, C. J.; Bowman, C. N.; Adv. Funct. Mater.2014 , 24, 2572.

Como exemplo de SP preparada por reações do tipo click chemistrypode-se destacar o trabalho de Guo et al.,5353 Guo, Z. M.; Liu, Y. F.; Xu, J. Y.; Xu, Q.; Xue, X. Y.; Zhang, F. F.; Ke, Y. X.; Liang, X. M.; Lei, A. W.; J. Chromatogr. A2008 , 1191, 78. que modificaram a superfície da sílica com grupos polares de polietilenoglicol (PEG), seguidos pela modificação do mesmo suporte com glicopeptídeos,5454 Zhao, Y. Y.; Yu, L.; Guo, Z. M.; Li, X. L.; Liang, X. M.;Anal Bioanal. Chem. 2011 ,399, 3359.β-cyclodextrinas5555 Guo, Z.; Jin, Y.; Liang, T.; Liu, Y.; Xu, Q.; Liang, X.; Le, A.;J. Chromatogr. A2011 , 1216, 257. ou fosfolipídios.5656 Zhao, Y. Y.; Li, X. L.; Yan, J. Y.; Guo, Z. M.; Liang, X. M.;Anal. Methods2012 , 4, 1244.

Outro exemplo interessante de preparação de SP por meio de reações do tipoclick chemistry foi reportado por Peng et al.,5757 Peng, X. T.; Lui, T.; Ji, S. X.; Feng, Y .Q.; J. Sep. Sci.2013 , 36, 2571. os quais sintetizaram uma nova SP a base de sílica modificada com grupos tiol e carboxila (Figura 6a), que apresentou perfil de separação satisfatório para a caracterização cromatográfica de 13 nucleosídeos e suas bases, além de quatro vitaminas hidrossolúveis.

Figura 6
Reações do tipo click chemistry envolvidas na preparação da SP HILIC baseada em sílica modificada com grupos tiol e carboxila e o seu desempenho na separação de 13 nucleosídeos em (a) e o desempenho de uma SP preparada por click chemistry modificada com grupos amino na separação de quatro compostos orgânicos com características básicas em (b). Dimensões das colunas cromatográficas: 150 mm × 4.6 mm em (a) e 100 mm × 2.1 mm em (b). Analitos: (1) timina, (2) uracila, (3) timidina, (4)desoxiadenosina, (5) uridina, (6) adenina, (7) adenosina, (8) citosina, (9) deoxicitidina, (10) guanina, (11) desoxiguanosina, (12) citidina e (13) guanosina em (a). Em (b), (1) difenilamina, (2)cafeína, (3) teofilina, (4) diprofilina. Adaptadas das refs. 57 e 58

Liu et al.5858 Liu, Y.; Du, Q.; Yang, B.; Zhang, F.; Chang, C.; Liang, X.;Analyst2012 , 137, 1624. prepararam uma SP a base de grupos amino imobilizados sobre sílica via reações do tipoclick chemistry com bom desempenho para separação de compostos polares, incluindo nucleosídeos, compostos orgânicos com características ácidas e básicas. A SP ainda separou ânions de natureza inorgânica no modo cromatografia de íons (Figura 6b).

Fases estacionárias monolíticas convencionais e capilares

Colunas monolíticas são um meio contínuo de separação, formadas por um corpo único, poroso, em formato de haste, utilizados em cromatografia líquida (cromatografia líquida em escala convencional, eletrocromatografia capilar e cromatografia líquida capilar) como alternativa as colunas com SP particuladas. Essas colunas possuem uma estrutura bi-porosa com macroporos (com diâmetros superiores a 50 nm) e mesoporos (2 a 50 nm) que permite a utilização de vazões mais altas, devido à alta permeabilidade do material, o que proporciona análises mais rápidas, sem o comprometimento da eficiência de separação e sem aumento significativo da pressão no sistema cromatográfico. Além disso, apresentam fácil preparação e modificação química de sua superfície e não necessitam do uso defrits para isolamento da SP.5959 Legido-Quigley, C.; Marlin, L. D; Melin, V.; Manz, A.; Smith, N. W.;Electrophoresis2003 , 24, 917.

60 Moravcová, D.; Jandera, P.; Urban, J.; Planeta, J.; J. Sep. Sci.2004 , 27, 789.

61 Faria, A. M.; Bottolli, C. B. G.; Jardim, I. C. S. F.; Collins, C. H.; Quim. Nova2006 , 29, 300.

62 Svec, F.; Geiser, L.; LCGCNorth Am.2006 , 24, 22.

63 Carr, C. D.; LCGC Column Technology Supplement2006 , 4, 44.

64 Ali, I.; Aboul-Enein, H. Y.; Gupta, V. K.;Nanochromatography and Nanocapillary Electrophoresis: Pharmaceutical and Environmental Analyses, John Wiley & Sons: Hoboken, 2009, ch. 2.

65 Wu, R.; Zou, H.; M. T. In Monolithic Chromatography and Its Modern Applications; Wang, P. G.; He, W., eds.; ILM Publications: Hertfordshire, 2010, ch. 9.

66 Silva, R. G. C.; Collins, C. H.; Bottolli, C. B. G.; Quim. Nova2012 , 34, 849.

67 Silva, C. G. A.; Collins, C. H.; Bottolli, C. B. G.;Microchem. J.2014 , 116, 249.
-6868 Liang, Y.; Zhang, L.; Zhang, Y.; Anal. Bioanal. Chem.2013 , 405, 2095.

Colunas monolíticas polares a base de sílica, introduzidas por Tanaka et al.,6969 Tanaka, N.; Kobayashi, H.; Nakanishi, K.; Minakuchi, H.; Ishizuka, N.; Anal. Chem.2001 , 73, 420A. vem sendo utilizadas no modo HILIC devido a suas propriedades físico-químicas como estabilidade em solventes orgânicos utilizados em cromatografia, reduzido diâmetro de poros e elevada área superficial. Esses monolitos são preparados pelo processo sol-gel (PSG), que envolve reações de hidrólise de um alcóxido (geralmente tetraetoxisilano, TEOS, ou tetrametoxisilano, TMOS), chamado de precursor, seguida por reações de condensação e policondensação. O tamanho dos poros do material é controlado por polímeros orgânicos, chamados de agentes porogênicos, sendo que o mais utilizado é o polietilenoglicol (PEG), que é dissolvido em água contendo um catalisador apropriado, que pode ser um ácido ou base (normalmente ácido acético ou uréia). A reação se dá primeiramente pela formação de um sol, suspensão coloidal de espécies sólidas em um líquido. Em seguida, há a formação do gel a partir da policondensação do sol, levando a uma estrutura úmida. Após, a estrutura porosa bimodal é obtida por meio de tratamento térmico, no qual sucessivas evaporações e secagens são realizadas para decompor os constituintes orgânicos e estabilizar a superfície do gel hidrofílico de sílica.6161 Faria, A. M.; Bottolli, C. B. G.; Jardim, I. C. S. F.; Collins, C. H.; Quim. Nova2006 , 29, 300.,7070 Tanaka, N.; Kobayashi, H.; Ishizuka, N.; Minakuchi, H.; Nakanishi, K.; Hosoya, K.; Ikegami, T.; J. Chromatogr. A2002 , 965, 35.

71 Brinker, C. J.; Scherer, G. W.; The Physics and Chemistry of Sol-Gel Processing, Academic Press: San Diego, 1990, ch. 1.
-7272 Miyabe, K.; Guiochon, G.; J. Sep. Sci.2004 , 27, 853.Alternativamente, pode-se substituir o alcóxido de silício por alcóxidos metálicos de zircônio ou titânio, resultando em materiais com maior estabilidade química do que os de sílica, que resistem apenas na faixa 2 < pH < 8, inclusive podendo ser utilizados em separações no modo HILIC utilizando a cromatografia líquida em escala convencional ou capilar.6767 Silva, C. G. A.; Collins, C. H.; Bottolli, C. B. G.;Microchem. J.2014 , 116, 249.,7373 Randon, J.; Huguet, S.; Piram, A.; Puy, G.; Demesmay, C.; Rocca, J. L.; J. Chromatogr. A2006 , 1109, 19.

74 Randon, J.; Huguet, S.; Demesmay, C.; Berthold, A.; J. Chromatogr. A2010 , 1217, 1496.

75 Randon, J.; Guerrin, J. F.; Rocca, J. L.; J. Chromatogr. A2008 , 1214, 186.

76 Konishi, J.; Fujita, K.; Nakanishi, K.; Hirao, K.; Morisato, K.; Miyazaki, S.; J. Chromatogr. A2009 , 1216, 7375.

77 Zhou, T.; Lucy, C. A.; J. Chromatogr. A2008 , 1187, 87.
-7878 Zhou, T.; Lucy, C. A.; J. Chromatogr. A2010 , 1217, 82.

Um exemplo interessante foi descrito por Zhou e Lucy,7878 Zhou, T.; Lucy, C. A.; J. Chromatogr. A2010 , 1217, 82. que separaram, utilizando o modo HILIC, uma mistura de ácidos carboxílicos altamente polares, incluindo ácidos com mais de uma carboxila na estrutura, utilizando uma coluna à base de titânia (TiO2) e MP ACN:tampão acetato em pH 6 ou 11, como pode ser observado nas Figuras 7a e 7b.

Figura 7
Separações de ácidos monocarboxílicos polares em (a) e separação de ácidos multi-carboxílicos em (b). Condições cromatográficas em (a): vazão: 1 mL min-1; MP: ACN: tampão acetato de sódio 10 mmol L-1 (95:5 v/v), pH 6; detecção 254 nm. Condições cromatográficas em (b): vazão: 1 mL min-1; MP: ACN: tampão acetato de sódio 3 mmol L-1 (50:50 v/v), pH 11; detecção 254 nm. Em (a) e (b) foi utilizada coluna à base de titânia (TiO2) (150 mm × 4.6 mm, 5 μm). Adaptada da ref. 78

Fases estacionárias à base de líquidos iônicos

Líquidos iônicos (IL, do inglês ionic liquids) são compostos geralmente formados por um cátion orgânico e um ânion inorgânico ou orgânico. Estes compostos também são chamados de sais fundidos à temperatura ambiente, que apresentam baixíssima pressão de vapor e propriedades de solvatação elevada.7979 Consorti, C. S.; Souza, R. F.; Dupont, J.; Suarez, P. A. Z.;Quim. Nova2001 , 24, 830.

80 Anderson, J. L.; Armstrong D. W.; Anal Chem.2003 , 1575, 4851.
-8181 Bertoti; A. R.; Ferreira, J. C. N.; Quim. Nova2009 , 32, 1934.

A baixa volatilidade dos IL não é uma propriedade crítica em cromatografia líquida e a sua natureza não molecular resulta em seletividades únicas tanto como aditivos de MP, para suprimir interações indesejáveis com os grupos silanóis ácidos ou como seletores químicos de SP, devido à possibilidade de imobilização sobre materiais considerados suportes cromatográficos, combinando, assim, suas características e propriedades com as desses materiais aplicados tanto no modo em fase normal, como em fase reversa.8282 Liu, J.; Jonsson, J. A.; Jiang, G.; Trends Anal. Chem.2005 , 24, 20.

83 Pandey, S.; Anal. Chim. Acta2006 , 556, 38.

84 Sun, P.; Armstrong, D. W.; Anal. Chim. Acta2010 , 661, 1.

85 Berthod, A.; Ruiz-Angel, M. J.; Broch-Carda, S.; J. Chromatogr. A2008 , 1184, 6.

86 van Meter, D. S.; Stuart, O. D.; Carle, A. B.; Stalcup, A. M.;J. Chromatogr. A2008 , 1191, 67.

87 Pino, V.; Afonso, A. M.; Anal. Chim. Acta2012 , 714, 20.

88 Xin, B.; Hao, J.; Chem. Soc. Rev.2014 , 43, 7171.

89 Zhang, M.; Liang, X.; Jiang, S.; Qiu, H.; Trends Anal. Chem.2014 , 53, 60.
-9090 Sun, M.; Feng, J.; Luo, C.; Liu, X.; Jiang, S.; Anal. Bioanal. Chem.2014 , 406, 2651.

Em HILIC, a modificação de suportes cromatográficos com IL já foi realizada com grupos imidazol,9191 Qui, H.; Mallik, A. K.; Takafuji, M.; Jiang, S.; Ihara, H.;Analyst2012 , 137, 2553. glucaminium9292 Qiao, L.; Wang, S.; Li, H.; Shan, Y.; Dou, A.; Shi, X.; Xu, G.;J. Chromatogr. A2014 , 1360, 240. e tricatiônicos.9393 Qiao, L.; Shi, X.; Lu, X.; Xu, G.; J. Chromatogr. A2015 , 1396, 62. Quiao et al.9393 Qiao, L.; Shi, X.; Lu, X.; Xu, G.; J. Chromatogr. A2015 , 1396, 62. prepararam e caracterizaram físico-química e cromatograficamente uma SP baseada no IL tri-imidazol quimicamente ligado na superfície de sílica porosa. A eficiência de separação e seletividade foi investigada por meio da separação de solutos nucleosídeos, bases orgânicas e flavonoides, e foi demonstrado que colunas cromatográficas baseadas em IL apresentam excelente potencial para serem utilizados em HILIC. Da mesma forma, foi demonstrada a versatilidade das SP a base de IL, que foram utilizadas em fase reversa com seletividade superior a uma SP do tipo alquila com grupos C8.

Seleção de colunas em HILIC

Com o grande número de SP e colunas cromatográficas disponíveis comercialmente, somada àquelas produzidas em laboratório, tipo labmade, surge a necessidade de se prever o comportamento de uma determinada SP frente à separação de analitos com características específicas, através do estudo de suas interações moleculares com a SP e MP.

Em HILIC, para se otimizar a retenção dos analitos polares, sugere-se a observação dos valores de log P e compará-lo com o grau de polaridade da SP. Por outro lado, alguns diagramas, como apresentados na Figura 8, desenvolvido pelos fabricantes de colunas, podem ser uma ferramenta interessante para seleção de SP para uma aplicação específica.

Figura 8
Diagrama para seleção de uma SP tipo HILIC de acordo com as propriedades químicas dos analitos a serem separados. Adaptada das refs. 94 e 95

Assim, de acordo com a Figura 8, SP com elevada hidrofilicidade, com grupos amida e zwiteriônicas (localizadas mais a direita no diagrama), deveriam ser escolhidas para separar compostos mais polares e hidrofílicos. Para a separação de analitos básicos, as fases recomendadas seriam as mais acima do eixo y, a saber, SP com alta capacidade de interação com cátions, à base de sílica. No entanto, para retenção de compostos ácidos, SP com elevada capacidade de interação com ânions são as mais apropriadas, como as que apresentam grupos amina. E por fim, SP que não apresentam propriedades de troca iônica, como as modificadas com grupos ureia, são as mais recomendadas para a separação de misturas com analitos neutros, ácidos, básicos.9494 https://www.thermoscientific.com/content/tfs/en/about-us/general-landing-page/hilic-overview.html, acessado em Novembro de 2015.
https://www.thermoscientific.com/content...
,9595 http://www.bioszeparacio.hu/_user/browser/File/Thermo/Applik%C3%A1ci%C3%B3/PN20957_HPLC_2014_HILIC_Methods.pdf, acessado em Novembro de 2015.
http://www.bioszeparacio.hu/_user/browse...

Por outro lado, assim como em cromatografia líquida no modo fase reversa, ou em cromatografia com fluido supercrítico, pode-se utilizar metodologias e modelos matemáticos para caracterizar SP por meio do estudo das interações moleculares que ocorrem durante o processo de separação cromatográfica.9696 Vitha, M.; Carr, P. W.; J. Chromatogr. A2006 , 1126, 143.

97 West, C.; Lesellier, E.; J. Chromatogr. A2008 , 1191, 21.

98 West, C.; Fougère, L.; J.; Lesellier, E.; J. Chromatogr. A2008 , 1189, 227.

99 West, C.; Ogden, J.; Lesellier, E.; J. Chromatogr. A2009 , 1216, 5600.

100 West, C.; Lesellier, E.; J. Chemom.2012 , 26, 52.
-101101 Galea, C.; Mangelings, D.; Heyden, Y. V.; Anal. Chim. Acta2015 , 886, 1.

Schuster e Lindner102102 Schuster, G.; Lindner, W.; J. Chromatogr. A2013 , 1273, 73.,103103 Schuster, G.; Lindner, W.; J. Chromatogr. A2013 , 1301, 98. foram os pioneiros na utilização do modelo dos parâmetros de solvatação, LSER (do inglês linear solvation energy relationship,) para classificar e avaliar o comportamento de 23 colunas cromatográficas tipo HILIC em duas faixas de pH, 3 e 5. Nesse trabalho, o coeficiente de ajuste do modelo (r2) apresentou valores entre 0,487 e 0,854, devido à utilização de descritores de espécies iônicas na sua forma neutra. Para sanar essa limitação, Wang et al.104104 Wang, J.; Guo, Z.; Shen, A.; Yu, L.; Xiao, Y.; Xue, X.; Zhang, X.; Liang, X.; J. Chromatogr. A2015 , 1398, 29. propuseram uma adaptação ao modelo de Schuster e Lindner, baseado na retenção de 41 solutos representando 5 propriedades relacionadas aos mecanismos de retenção em HILIC: h, habilidade de realizar partição hidrofílica; a, acidez; b, basicidade; c, capacidade de realizar troca catiônica e d, capacidade de realizar troca aniônica, o que proporcionou um modelo matemático mais adequado, com coeficiente de ajuste do modelo (r2) ≥ 0,990 e com baixo desvio padrão (SD), ≤ 0,041. Além disso, com base nos coeficientes de regressão do modelo relacionados a estas interações, a seletividade de SP HILIC foi plotada num diagrama do "tipo aranha" (do inglês, spider diagram).105105 Lesellier, E.; J. Chromatogr. A2015 , 1389, 49. Esse diagrama serve para guiar usuários e pesquisadores na seleção de colunas mais adequadas para a separação de um analito com propriedade específica e baseado nas propriedades estudadas em HILIC (Figura 9).

Figura 9
Diagrama tipo aranha das 23 colunas cromatográficas caracterizadas por Wang et al.,104 utilizando a HILIC para os cinco tipos de interações avaliadas pelo modelo dos parâmetros de solvatação: h, habilidade de realizar partição hidrofílica; a: acidez; b: basicidade; c: capacidade de realizar troca catiônica; d: capacidade de realizar troca aniônica. Adaptada da ref. 104

De acordo com a Figura 9, SP neutras e zwiteriônicas (ZIC) estão na mesma região do diagrama e, por isso, apresentam comportamento de retenção semelhante, ao passo que FE trocadoras de cátions estariam mais distantes no diagrama por apresentarem interações diferenciadas das colunas neutras e tipo ZIC. Por outro lado, a proximidade das SP do tipo Click TE-GSH (4.6 mm) e Click TE-GSH (3.0 mm) indicam que as dimensões da coluna cromatográfica (no caso diâmetro de coluna) têm papel insignificante no perfil de seletividade cromatográfica em HILIC. Assim, poderia-se selecionar uma coluna específica de acordo com o tipo de interação desejada para uma dada separação, em função das propriedades químicas dos analitos.

FASES MÓVEIS

Solventes em HILIC

A fase móvel clássica para separações em HILIC consiste em uma solução contendo um solvente aprótico como a ACN e água (com proporção de ACN > 70%, geralmente entre 70-95%). A ordem de força cromatográfica dos solventes em HILIC é dada por:66 Buszewski, B.; Noga, S.; Anal. Bioanal. Chem.2012 , 402, 231.,1313 Jandera, P.; Anal. Chim. Acta2011 , 692, 1.,2929 Hemstrom, P.; Irgum, K.; J. Sep. Sci.2006 , 29, 1784.,106106 http://www.nacalai.co.jp/global/cosmosil/pdf/cosmosil_hilic_application_notebook.pdf, acessado em Novembro de 2015.
http://www.nacalai.co.jp/global/cosmosil...
,107107 Lanças, F. M.; Sci. Chromatogr.2010 , 2, 49.

Acetona < isopropanol ca. propanol < acetonitrila < etanol < metanol < água

Metanol é o solvente orgânico clássico para ser utilizado em RP-HPLC, no entanto, em HILIC sua força cromatográfica é maior e, por isso, a retenção dos analitos é bastante diminuída (Figura 10a). Nesse sentido, acetonitrila, por ser um solvente mais fraco, foi selecionada como "solvente clássico" de HILIC, uma vez que, em contraste à utilização de metanol ou outros alcoóis, sua estrutura não favorece interações por ligações de hidrogênio e não proporciona competição com a água pelos sítios ativos da fase estacionária, o que diminui a estabilidade da camada de água utilizada para separação dos compostos em HILIC.

Figura 10
Perfil de retenção de compostos polares utilizando como MP MeOH:10 mmol L-1 CH3COONH4 (90:10 v/v) em (a) e utilizando ACN:10 mmol L-1CH3COONH4 (90:10 v/v) em (b). Coluna cromatográfica: Cosmosil HILIC (sílica modificada com grupos triazóis). Analitos: eritrol, leucina, tris(hidroximetil)aminometano, e ácido guanidoacético. Adaptada da referência 106

Além disso, independente do solvente orgânico, um gradiente típico de HILIC se inicia com elevada proporção de solvente orgânico e, por isso, compostos mais retidos são eluídos com o aumento na proporção de água na MP, uma vez que a retenção aumenta quando o conteúdo de água na MP diminui.

Uso de fases móveis tamponadas

Quando se analisa compostos ionizáveis, faz-se necessário o uso de tampões, na concentração comumente utilizada na fase reversa (5-100 mmol L-1), como acetato, fosfato, formiato, etc., mantendo-se a elevada proporção de solvente orgânico. O pH do tampão geralmente fica entre 2-7,5 unidades e o uso de tampões com pH mais elevado aumenta a retenção (Figura 10b).

Como se pode observar na Figura 10b, SP com suportes de sílica podem ser utilizadas na faixa de 2 < pH < 8, visto que em pH acima de 8 ocorre a dissolução da sílica e o colapso da SP e em pH menores que 2 ocorre a hidrólise da ligação siloxano (Si-O-Si-C), com a quebra da ligação entre a fase ligada e o suporte cromatográfico, resultando em uma perda contínua de fase, com subsequente perda de retenção do analito. Por isso, não se recomenda em HILIC uso de SP com suportes de sílica ou com sílica pura quando se utiliza tampões acima ou abaixo dessa faixa de pH.108108 Kirkland, J. J.; van Straten, M. A.; J. Chromatogr. A1995 , 691, 3.

109 Kirkland, J. J.; Henderson, J. W.; DeStefano, J. J.; van Straten, M. A.; Claessens, H. A.; J. Chromatogr. A1997 , 762, 97.
-110110 Ma, L.; Carr, P. W.; Anal. Chem. 2007 ,79, 4681.Para separação de solutos ácidos, deve-se evitar o uso de tampões a base de formiato de amônio por apresentarem baixa retenção neste tipo de MP. Nesse caso, recomenda-se o uso de tampões à base de ácido trifluoroacético, pois diminuem a repulsão entre os solutos carregados negativamente e os silanóis residuais da SP base de sílica.66 Buszewski, B.; Noga, S.; Anal. Bioanal. Chem.2012 , 402, 231.

Uso de aditivos

O uso de ácido etilenodiamino tetra-acético (EDTA, do inglês,ethylenediamine tetraacetic acid) na concentração de 5 mmol L-1 ou de tampão citrato pH 7 (10 mmol L-1) tem sido reportada106106 http://www.nacalai.co.jp/global/cosmosil/pdf/cosmosil_hilic_application_notebook.pdf, acessado em Novembro de 2015.
http://www.nacalai.co.jp/global/cosmosil...
como uma alternativa interessante para suprimir interações indesejáveis entre os silanóis ácidos presentes no suporte e os analitos básicos, o que leva ao alargamento dos picos cromatográficos e perda da simetria.

Para analitos ácidos, quando se utiliza o detector espectrômetro de massas e ionização por eletrospray (ESI-MS), ácido ftálico na concentração de 0,08 mmol L-1 aumenta a detectabilidade devido ao desfavorecimento da formação de adutos de sódio com os analitos. Também foi reportado o aumento da simetria dos picos cromatográficos.111111 Qi, W. S.; Guan, Q.; Sun, T. Q.; Cao, Y. J.; Zhang, L.; Guo, Y.;Anal. Chim. Acta2015 , 870, 75.

HILIC ACOPLADA A ESPECTROMETRIA DE MASSAS

A alta compatibilidade da HILIC com a espectrometria de massas, devido à elevada porcentagem de solvente orgânico na MP, tem elevado o número de aplicações da cromatografia líquida acoplada à espectrometria de massas, o que pode ser evidenciado por suas aplicações em diversas áreas da ciência como em química ambiental,112112 Taujenis, L.; Olsauskaite, V.; Padarauskas, A.; Acta Chromatogr.2015 , 27, 373.

113 Fan, H.; Qiu, J. B.; Fan, L.; Li, A. F.; Environ. Sci. Pollut. Res.2015 , 22, 5943.
-114114 Rodin, I.; Baygildiev, T.; Stavrianidi, A.; Braun, A.; Rybalchenko, I.; Shpigun, O.; Chromatographia2015 , 78, 585. de alimentos,115115 Inoue, K.; Tanada, C.; Sakamoto, T.; Tsutsui, H.; Akiba, T.; Min, J. Z.; Todoroki, K.; Yamano, Y.; Toyo'oka, T.; Food Chem.2015 , 184, 318.

116 Ghfar, A. A.; Wabaidur, S. M.; Ahmed, A. Y. B. H.; Alothman, Z. A.; Khan, M. R.; Al-Shaalan, N. H.; Food Chem. 2015 ,176, 487.

117 Roullier-Gall, C.; Witting, M.; Tziotis, D.; Ruf, A.; Gougeon, R. D.; Schmitt-Kopplin, P.; Tetrahedron2015 , 71, 2983.
-118118 Le Maux, S.; Nongonierma, A.; FitzGerald, B.; Richard, J.;Food Chem. 2015 , 173, 847. na área farmacêutica,119119 Vidmar, J.; Martincic, A.; Milacic, R.; Scancar, J.;Talanta2015 , 138, 1.

120 Martens-Lobenhoffer, J.; Bode-Boger, S. M.; J. Chromatogr. B2015 , 990, 164.

121 Kasagic-Vujanovic, I.; Jancic-Stojanovic, B.; Rakic, T.; Ivanovic, D.; J. Liq. Chromatogr. Relat. Technol.2015 , 38, 919.
-122122 Kahsay, G.; Song, H. Y.; Van Schepdael, A.; Cabooter, D.; Adams, E.;J. Pharm. Biomed. Anal.2014 , 87, 142. na área clínica,111111 Qi, W. S.; Guan, Q.; Sun, T. Q.; Cao, Y. J.; Zhang, L.; Guo, Y.;Anal. Chim. Acta2015 , 870, 75.,123123 Meyer-Monath, M.; Chatellier, C.; Cabooter, D.; Rouget, F.; Morel, I.; Lestremau, F.; Talanta2015 , 138, 231.

124 Holcapek, M.; Cervena, B.; Cifkova, E.; Lisa, M.; Chagovets, V.; Vostalova, J.; Bancirova, M.; Galuszka, J.; Hill, M.; J. Chromatogr. B2015 , 990, 52.
-125125 Wang, S. H.; Wu, H. Y.; Huang, X. L.; Geng, P. W.; Wen, C. C.; Ma, J. S.; Zhou, Y. F.; Wang, X. Q.; J. Chromatogr. B2015 , 990, 118. e, mais recentemente, na área ômica (proteômica, glicômica, lipidômica, pepdômica, etc.)126126 Magiera, S.; Baranowski, J.; J. Pharm. Biomed. Anal.2015 , 109, 171.

127 Konstantina S.; Tsoukali, H.; Nikolaos R.; Gika, H.; Wilson, I. D.; Georgios, T.; J. Sep. Sci.2010 , 33, 716.

128 Cao, M. S.; Fraser, K.; Huege, J.; Featonby, T.; Rasmussen, S.; Jones, C.; Metabolomics2015 , 11, 696.

129 Meissen, J. K.; Hirahatake, K. M.; Adams, S. H.; Fiehn, O.;Metabolomics2015 , 11, 707.

130 Tufi, S.; Lamoree, M.; de Boer, J.; Leonards, P.; J. Chromatogr. B2015 , 1395, 79.

131 Vorkas, P. A.; Isaac, G.; Anwar, M. A.; Davies, A. H.; Want, E. J.; Nicholson, J. K.; Holmes, E.; Anal. Chem.2015 , 87, 4184.
-132132 Teleki, A.; Sanchez-Kopper, A.; Takors, R. Anal. Biochem.2015 , 475, 4.

Assim como na modalidade em fase reversa, o método mais utilizado para ionizar os analitos é a ionização por eletrospray, devido ao estado físico líquido da MP.

Em metabolômica, HILIC acoplada à MS tem se apresentado como uma ferramenta importante e complementar a cromatografia líquida em fase reversa, para compostos polares que não são retidos ou têm baixa retenção. O acoplamento com o espectrômetro de massas permite a elucidação estrutural dos metabólitos uma vez que o objetivo principal é fornecer o mais abrangente conjunto de perfis de metabólitos o quanto possível para os diferentes tipos de matrizes (fluidos biológicos, tecidos, células, etc.).88 Spagou, K.; Tsoukali, H.; Raikos, N.; Gika, H.; Wilson, I. D.; Theodoridis, G.; J. Sep. Sci.2010 , 33, 716.

A MS e a espectroscopia de ressonância magnética nuclear (NMR, do inglêsnuclear magnetic resonance) são atualmente as principais tecnologias analíticas para a geração de dados para estudos na área de metabolômica, sendo que a MS ainda fornece uma importante vantagem: a possibilidade do fácil e direto acoplamento com a cromatografia. Nesse sentido, HILIC-MS proporciona não só a retenção de compostos polares, mas a possibilidade do estudo de novas interações relativas à química complementar das SP. Da mesma forma, a alta porcentagem de solvente orgânico na MP permite a detecção de diversos analitos em ambos os modos de ionização, positiva ou negativa, maximizando o número de compostos a serem estudados.

Os estudos de Godejohann,133133 Godejohann M.; J. Chromatogr. A2006 , 1156, 87. por exemplo, ilustram a complementaridade da HILIC em relação a RP-HPLC na separação e identificação de metabólitos polares endógenos em urina de ratos tipo "Zucker", criados em laboratório para serem utilizados como modelos genéticos para o estudos de obesidade, utilizando o detector MS (Figura 11). Nesse trabalho, os compostos creatinina, creatina, uracil, trigoneline e ácido hipúrico foram separados no modo HILIC utilizando uma coluna polar de sílica e o perfil de separação comparado com aquele que utilizou uma coluna tipo C18 em modo fase reversa, ambos utilizando como detector espectrômetro de massas com analisador por tempo de vôo (do inglês, time-of-flight). A utilização dos dois modos de separação permitiu a obtenção de resultados complementares em relação à retenção dos compostos analisados e os analitos, que eluiram praticamente junto com o volume morto em fase reversa, porém, foram efetivamente separados em HILIC, o que permitiu se determinar o perfil metabolômico de ratos magros e ratos obesos em amostras de urina.

Figura 11
Perfil de separação dos metabólitos de ratos obesos em (a) e magros em (b) separados por HILIC e identificados por MS. A extração do íon m/z 154,517 revelou melhor detectividade desse analito em HILIC quando comparado com o modo em RP. Adaptada da ref. 133

APLICAÇÕES RECENTES

Algumas aplicações recentes da HILIC estão sumarizadas na Tabela 1, na qual se pode observar que as possibilidades de análise com HILIC são amplas e variadas, desde a separação de metabólitos ou biomarcadores, fármacos, drogas de abuso ou seus metabólitos em matrizes complexas como alimentos, plasma, amostras ambientais, ou cabelos, separados em colunas cromatográficas de sílica pura ou modificada com grupos polares, ciclodextrinas, líquidos iônicos e zwiteriônicos, e utilizando detectores mais simples, como ultravioleta (UV), por arranjos de diodos (DAD, do inglês, diode array detector), de espalhamento de luz (ELSD, do inglês, evaporative light scattering detector) ou ainda detectores mais sofisticados como o MS.

Tabela 1
Aplicações recentes da separação de compostos por HILIC

CONSIDERAÇÕES FINAIS

HILIC representa uma nova e recente modalidade de separação da cromatografia líquida e tem como principal vantagem à ortogonalidade em relação à cromatografia líquida em fase reversa. HILIC tem sido utilizada na separação de compostos polares em diversas áreas da ciência como a área farmacêutica, de alimentos, da química ambiental, em toxicologia, metabolômica, entre outras.

Uma das áreas de pesquisa dentro de separações que mais tem contribuído para o aumento das publicações no meio acadêmico e a aplicação da HILIC na indústria se deve à pesquisa e o desenvolvimento de novas fases estacionárias com características químicas e de retenção baseadas em interações polares, como, por exemplo, ligações de hidrogênio. Nesse sentido, há muito que se pesquisar para se produzir fases estacionárias cada vez mais seletivas e com características ainda mais apropriadas. O estudo das interações moleculares que ocorrem entre analitos, MP e SP, baseados principalmente na utilização de modelos matemáticos - como o modelo dos parâmetros de solvatação que descreve essas interações a fim de elucidar com mais detalhes os mecanismos que regem a retenção no modo HILIC, bem como permite a previsão do perfil de separação baseado nas características físico-químicas dos analitos a fim de se selecionar uma SP mais adequada para a separação de um grupo específico ou classes de analitos, também merece destaque.

AGRADECIMENTOS

À FAPESP pela bolsa de pós-doutorado (processo 2012/23518-2) e auxílio financeiro (processo 2013/25933-0). Esta é uma contribuição do Instituto Nacional de Química Bioanalítica (INCTBio).

REFERÊNCIAS

  • 1
    Maldaner, L.; Collins, C. H.; Jardim, I. C. S. F.; Quim. Nova2010 , 33, 1559.
  • 2
    Alpert, A. J.; J. Chromatogr.1990 , 499, 177.
  • 3
    Gama, M. R.; Silva, R. G. C.; Collins, C. H. C.; Bottoli, C. B. G.;Trends Anal. Chem.2012 , 37, 48.
  • 4
    Mogollón, G. S. N.; Lima, P. F.; Gama, M. R.; Furlan, M. F.; Braga, S. C. G. N.; Prata, P. S.; Jardim, I. C. S. F.; Augusto, F.; Quim. Nova2014 , 37, 1680.
  • 5
    Lin, H.; Ou, J.; Zhang, Z.; Dong, J.; Wu, M.; Zou, H.; Anal. Chem 2012 , 84, 2721.
  • 6
    Buszewski, B.; Noga, S.; Anal. Bioanal. Chem.2012 , 402, 231.
  • 7
    Cubbon, S.; Antonio, C.; Wilson, J.; Thomas-Oates, J.; Mass Spectrom. Rev.2010 , 29, 671.
  • 8
    Spagou, K.; Tsoukali, H.; Raikos, N.; Gika, H.; Wilson, I. D.; Theodoridis, G.; J. Sep. Sci.2010 , 33, 716.
  • 9
    Xu, R. N.; Rieser M. J.; El-Shourbagy, T. A.;Bioanalysis2009 , 1, 239.
  • 10
    Zhang, F.; Shen, G.; Ji, S.; Yang, B.; J. Liq. Chromatogr. Relat. Technol.2015 , 38, 349.
  • 11
    Naidong, W.; Shou, W.; Chen, L.Y.; Jiang, X. L.; J. Chromatogr. B2001 , 754, 387.
  • 12
    Dejaegher, B.; Heyden, Y. V.; J. Sep. Sci2010 , 33, 698.
  • 13
    Jandera, P.; Anal. Chim. Acta2011 , 692, 1.
  • 14
    Oyler, A. R.; Armstrong, B. L.; Cha, J. Y.; Zhou, M. X.; Yang, Q.; Robison, R. I.; Dunphy, R.; Burinsky, D. J.; J. Chromatogr. A1996 , 724, 378.
  • 15
    Tolstikov, V.V.; Fiehn, O.; Anal. Biochem2002 , 301, 298.
  • 16
    Jandera, P.; Hajek, T.; J. Sep. Sci.2009 , 32, 3603.
  • 17
    Armstrong, D. W.; Jin, H. L.; J. Chromatogr1989 , 462, 219.
  • 18
    Kawachi, Y.; Ikegami, T.; Takubo, H.; Ikegami, Y.; Miyamoto, M.; Tanaka, N.; J. Chromatogr. A2011 , 1218, 5903.
  • 19
    Wang, C.; Jiang, C.; Armstrong, D. W.; J. Sep. Sci.2008 , 31, 1980.
  • 20
    Qiao, L.; Dou, A.; Shi, X.; Li, H.; Shan, Y.; Lu, X.; Xu, G.;J. Chromatogr. A2013 , 1286, 137.
  • 21
    Pino, V.; Afonso, M. A.; Anal. Chim. Acta2012 , 714, 20.
  • 22
    Qiao, L.; Li, H.; Shan, Y.; Wang, S.; Shi, X.; Lu, X.; Xu, G.;J. Chromatogr. A2014 , 1330, 40.
  • 23
    Guo, Z. M.; Lei, A. W.; Zhang, Y. P.; Xu, Q.; Xue, X. Y.; Zhang, F. F.; Liang, X. M.; Chem. Commun.2007 , 2491.
  • 24
    Fu, Q.; Guo, Z.; Liang, T.; Zhang, X.; Xu, Q.; Liang, X.;Anal. Methods2010 , 2, 217.
  • 25
    Ikegami, T.; Tomomatsu, K.; Takubo, K.; Horie, K.; Tanaka, N.;J. Chromatogr A2008 , 474, 1184.
  • 26
    Pesek, J. J., Matyska, M. T.; In Hydrophilic Interaction Liquid Chromatography (HILIC) and Advanced Applications; Wang, P. G.; He, W., eds.; CRC Press: Boca Raton, 2011, ch. 2.
  • 27
    Weng, N.; J. Chromatogr. B2003 , 796, 209.
  • 28
    Weng, N.; Shou, W.; Chen, Y.; Jiang, X.; J. Chromatogr. B2001 , 754, 387.
  • 29
    Hemstrom, P.; Irgum, K.; J. Sep. Sci2006 , 29, 1784.
  • 30
    Gregor, H. P.; Collins, F. C.; Pope, M.; J. Colloid Sci 1951 , 6, 304.
  • 31
    Samuelson, O.; Sjöström, E.; Sven. Kem. Tidskr1952 , 64, 305.
  • 32
    Rückert, H.; Samuelson, O.; Sven. Kem. Tidskr1954 , 66, 337.
  • 33
    Linden, J. C.; Lawhead, C. L.; J. Chromatogr1975 , 105, 125.
  • 34
    Palmer, J. K.; Anal. Lett 1975 ,8, 215.
  • 35
    Verhaar, L. A. T.; Kuster, B. F. N.; J. Chromatogr1982 , 234, 57.
  • 36
    Nikolov, Z. L.; Reilly, P. J.; J. Chromatogr1985 , 325, 287.
  • 37
    Yoshida, T.; J. Biochem. BiophysMethods2004 , 60, 265.
  • 38
    Yoshida, T.; J. Chromatogr. A1998 , 811, 61.
  • 39
    Bell, D. S.; LCGC North Am.2015 , 33, 90.
  • 40
    Liang, T.; Fu, Q.; Shen, A.; Wang, J.; Xin, H.; Guo, K. Y.; Liang, X.; J. Chromatogr. A2015 , 1388, 110.
  • 41
    Garcia-Gomez, D.; Rodriguez-Gonzalo, E.; Carabias-Martinez, R.;Trends Anal. Chem.2013 , 47, 111.
  • 42
    McCalley, D. V.; J. Chromatogr. A2010 , 1217, 3408.
  • 43
    Chen, Y.; Wang, K.; Liu, Y.; Yang, H.; Yao, S.; Chen, B.; Nie, L.; Xu, G. M.; Electrophoresis2013 , 34, 1877.
  • 44
    Li, R.; Yuan, Q.; Huang, Y.; Chin. J. Chromatogr.2014 , 32, 675.
  • 45
    Zhang, F.; Shen, G.; Ji, S.; Yang, B.; J. Liq. Chromatogr. Relat. Technol.2015 , 38, 349.
  • 46
    Chu, C. H.; Liu, R. H.; Chem. Soc. Rev.2011 , 40, 2177.
  • 47
    Kolb, H. C.; Finn, M. G.; Sharpless, B.; Angew. Chem. Int. Ed.2001 , 40, 2004.
  • 48
    Grammel, M.; Hang, H. C.; Nat. Chem. Biol.2013 , 40, 2004.
  • 49
    Xie, R.; Hong, S.; Chen, X.; Curr. Opin. Chem. Biol.2013 , 17, 747.
  • 50
    Kolb, H. C.; Sharpless, B.; Drug Discovery Today2003 , 8, 1128.
  • 51
    Thirumurugan, P.; Matosiuk, D.; Jozwiak, K.; Chem. Rev.2013 , 113, 4905.
  • 52
    Xi, W.; Scott, T. F.; Kloxin, C. J.; Bowman, C. N.; Adv. Funct. Mater.2014 , 24, 2572.
  • 53
    Guo, Z. M.; Liu, Y. F.; Xu, J. Y.; Xu, Q.; Xue, X. Y.; Zhang, F. F.; Ke, Y. X.; Liang, X. M.; Lei, A. W.; J. Chromatogr. A2008 , 1191, 78.
  • 54
    Zhao, Y. Y.; Yu, L.; Guo, Z. M.; Li, X. L.; Liang, X. M.;Anal Bioanal. Chem 2011 ,399, 3359.
  • 55
    Guo, Z.; Jin, Y.; Liang, T.; Liu, Y.; Xu, Q.; Liang, X.; Le, A.;J. Chromatogr. A2011 , 1216, 257.
  • 56
    Zhao, Y. Y.; Li, X. L.; Yan, J. Y.; Guo, Z. M.; Liang, X. M.;Anal. Methods2012 , 4, 1244.
  • 57
    Peng, X. T.; Lui, T.; Ji, S. X.; Feng, Y .Q.; J. Sep. Sci.2013 , 36, 2571.
  • 58
    Liu, Y.; Du, Q.; Yang, B.; Zhang, F.; Chang, C.; Liang, X.;Analyst2012 , 137, 1624.
  • 59
    Legido-Quigley, C.; Marlin, L. D; Melin, V.; Manz, A.; Smith, N. W.;Electrophoresis2003 , 24, 917.
  • 60
    Moravcová, D.; Jandera, P.; Urban, J.; Planeta, J.; J. Sep. Sci.2004 , 27, 789.
  • 61
    Faria, A. M.; Bottolli, C. B. G.; Jardim, I. C. S. F.; Collins, C. H.; Quim. Nova2006 , 29, 300.
  • 62
    Svec, F.; Geiser, L.; LCGCNorth Am.2006 , 24, 22.
  • 63
    Carr, C. D.; LCGC Column Technology Supplement2006 , 4, 44.
  • 64
    Ali, I.; Aboul-Enein, H. Y.; Gupta, V. K.;Nanochromatography and Nanocapillary Electrophoresis: Pharmaceutical and Environmental Analyses, John Wiley & Sons: Hoboken, 2009, ch. 2.
  • 65
    Wu, R.; Zou, H.; M. T. In Monolithic Chromatography and Its Modern Applications; Wang, P. G.; He, W., eds.; ILM Publications: Hertfordshire, 2010, ch. 9.
  • 66
    Silva, R. G. C.; Collins, C. H.; Bottolli, C. B. G.; Quim. Nova2012 , 34, 849.
  • 67
    Silva, C. G. A.; Collins, C. H.; Bottolli, C. B. G.;Microchem. J.2014 , 116, 249.
  • 68
    Liang, Y.; Zhang, L.; Zhang, Y.; Anal. Bioanal. Chem.2013 , 405, 2095.
  • 69
    Tanaka, N.; Kobayashi, H.; Nakanishi, K.; Minakuchi, H.; Ishizuka, N.; Anal. Chem.2001 , 73, 420A.
  • 70
    Tanaka, N.; Kobayashi, H.; Ishizuka, N.; Minakuchi, H.; Nakanishi, K.; Hosoya, K.; Ikegami, T.; J. Chromatogr. A2002 , 965, 35.
  • 71
    Brinker, C. J.; Scherer, G. W.; The Physics and Chemistry of Sol-Gel Processing, Academic Press: San Diego, 1990, ch. 1.
  • 72
    Miyabe, K.; Guiochon, G.; J. Sep. Sci.2004 , 27, 853.
  • 73
    Randon, J.; Huguet, S.; Piram, A.; Puy, G.; Demesmay, C.; Rocca, J. L.; J. Chromatogr. A2006 , 1109, 19.
  • 74
    Randon, J.; Huguet, S.; Demesmay, C.; Berthold, A.; J. Chromatogr. A2010 , 1217, 1496.
  • 75
    Randon, J.; Guerrin, J. F.; Rocca, J. L.; J. Chromatogr. A2008 , 1214, 186.
  • 76
    Konishi, J.; Fujita, K.; Nakanishi, K.; Hirao, K.; Morisato, K.; Miyazaki, S.; J. Chromatogr. A2009 , 1216, 7375.
  • 77
    Zhou, T.; Lucy, C. A.; J. Chromatogr. A2008 , 1187, 87.
  • 78
    Zhou, T.; Lucy, C. A.; J. Chromatogr. A2010 , 1217, 82.
  • 79
    Consorti, C. S.; Souza, R. F.; Dupont, J.; Suarez, P. A. Z.;Quim. Nova2001 , 24, 830.
  • 80
    Anderson, J. L.; Armstrong D. W.; Anal Chem.2003 , 1575, 4851.
  • 81
    Bertoti; A. R.; Ferreira, J. C. N.; Quim. Nova2009 , 32, 1934.
  • 82
    Liu, J.; Jonsson, J. A.; Jiang, G.; Trends Anal. Chem.2005 , 24, 20.
  • 83
    Pandey, S.; Anal. Chim. Acta2006 , 556, 38.
  • 84
    Sun, P.; Armstrong, D. W.; Anal. Chim. Acta2010 , 661, 1.
  • 85
    Berthod, A.; Ruiz-Angel, M. J.; Broch-Carda, S.; J. Chromatogr. A2008 , 1184, 6.
  • 86
    van Meter, D. S.; Stuart, O. D.; Carle, A. B.; Stalcup, A. M.;J. Chromatogr. A2008 , 1191, 67.
  • 87
    Pino, V.; Afonso, A. M.; Anal. Chim. Acta2012 , 714, 20.
  • 88
    Xin, B.; Hao, J.; Chem. Soc. Rev.2014 , 43, 7171.
  • 89
    Zhang, M.; Liang, X.; Jiang, S.; Qiu, H.; Trends Anal. Chem.2014 , 53, 60.
  • 90
    Sun, M.; Feng, J.; Luo, C.; Liu, X.; Jiang, S.; Anal. Bioanal. Chem.2014 , 406, 2651.
  • 91
    Qui, H.; Mallik, A. K.; Takafuji, M.; Jiang, S.; Ihara, H.;Analyst2012 , 137, 2553.
  • 92
    Qiao, L.; Wang, S.; Li, H.; Shan, Y.; Dou, A.; Shi, X.; Xu, G.;J. Chromatogr. A2014 , 1360, 240.
  • 93
    Qiao, L.; Shi, X.; Lu, X.; Xu, G.; J. Chromatogr. A2015 , 1396, 62.
  • 94
    https://www.thermoscientific.com/content/tfs/en/about-us/general-landing-page/hilic-overview.html, acessado em Novembro de 2015.
    » https://www.thermoscientific.com/content/tfs/en/about-us/general-landing-page/hilic-overview.html
  • 95
    http://www.bioszeparacio.hu/_user/browser/File/Thermo/Applik%C3%A1ci%C3%B3/PN20957_HPLC_2014_HILIC_Methods.pdf, acessado em Novembro de 2015.
    » http://www.bioszeparacio.hu/_user/browser/File/Thermo/Applik%C3%A1ci%C3%B3/PN20957_HPLC_2014_HILIC_Methods.pdf
  • 96
    Vitha, M.; Carr, P. W.; J. Chromatogr. A2006 , 1126, 143.
  • 97
    West, C.; Lesellier, E.; J. Chromatogr. A2008 , 1191, 21.
  • 98
    West, C.; Fougère, L.; J.; Lesellier, E.; J. Chromatogr. A2008 , 1189, 227.
  • 99
    West, C.; Ogden, J.; Lesellier, E.; J. Chromatogr. A2009 , 1216, 5600.
  • 100
    West, C.; Lesellier, E.; J. Chemom.2012 , 26, 52.
  • 101
    Galea, C.; Mangelings, D.; Heyden, Y. V.; Anal. Chim. Acta2015 , 886, 1.
  • 102
    Schuster, G.; Lindner, W.; J. Chromatogr. A2013 , 1273, 73.
  • 103
    Schuster, G.; Lindner, W.; J. Chromatogr. A2013 , 1301, 98.
  • 104
    Wang, J.; Guo, Z.; Shen, A.; Yu, L.; Xiao, Y.; Xue, X.; Zhang, X.; Liang, X.; J. Chromatogr. A2015 , 1398, 29.
  • 105
    Lesellier, E.; J. Chromatogr. A2015 , 1389, 49.
  • 106
    http://www.nacalai.co.jp/global/cosmosil/pdf/cosmosil_hilic_application_notebook.pdf, acessado em Novembro de 2015.
    » http://www.nacalai.co.jp/global/cosmosil/pdf/cosmosil_hilic_application_notebook.pdf
  • 107
    Lanças, F. M.; Sci. Chromatogr.2010 , 2, 49.
  • 108
    Kirkland, J. J.; van Straten, M. A.; J. Chromatogr. A1995 , 691, 3.
  • 109
    Kirkland, J. J.; Henderson, J. W.; DeStefano, J. J.; van Straten, M. A.; Claessens, H. A.; J. Chromatogr. A1997 , 762, 97.
  • 110
    Ma, L.; Carr, P. W.; Anal. Chem 2007 ,79, 4681.
  • 111
    Qi, W. S.; Guan, Q.; Sun, T. Q.; Cao, Y. J.; Zhang, L.; Guo, Y.;Anal. Chim. Acta2015 , 870, 75.
  • 112
    Taujenis, L.; Olsauskaite, V.; Padarauskas, A.; Acta Chromatogr.2015 , 27, 373.
  • 113
    Fan, H.; Qiu, J. B.; Fan, L.; Li, A. F.; Environ. Sci. Pollut. Res.2015 , 22, 5943.
  • 114
    Rodin, I.; Baygildiev, T.; Stavrianidi, A.; Braun, A.; Rybalchenko, I.; Shpigun, O.; Chromatographia2015 , 78, 585.
  • 115
    Inoue, K.; Tanada, C.; Sakamoto, T.; Tsutsui, H.; Akiba, T.; Min, J. Z.; Todoroki, K.; Yamano, Y.; Toyo'oka, T.; Food Chem2015 , 184, 318.
  • 116
    Ghfar, A. A.; Wabaidur, S. M.; Ahmed, A. Y. B. H.; Alothman, Z. A.; Khan, M. R.; Al-Shaalan, N. H.; Food Chem 2015 ,176, 487.
  • 117
    Roullier-Gall, C.; Witting, M.; Tziotis, D.; Ruf, A.; Gougeon, R. D.; Schmitt-Kopplin, P.; Tetrahedron2015 , 71, 2983.
  • 118
    Le Maux, S.; Nongonierma, A.; FitzGerald, B.; Richard, J.;Food Chem 2015 , 173, 847.
  • 119
    Vidmar, J.; Martincic, A.; Milacic, R.; Scancar, J.;Talanta2015 , 138, 1.
  • 120
    Martens-Lobenhoffer, J.; Bode-Boger, S. M.; J. Chromatogr. B2015 , 990, 164.
  • 121
    Kasagic-Vujanovic, I.; Jancic-Stojanovic, B.; Rakic, T.; Ivanovic, D.; J. Liq. Chromatogr. Relat. Technol.2015 , 38, 919.
  • 122
    Kahsay, G.; Song, H. Y.; Van Schepdael, A.; Cabooter, D.; Adams, E.;J. Pharm. Biomed. Anal.2014 , 87, 142.
  • 123
    Meyer-Monath, M.; Chatellier, C.; Cabooter, D.; Rouget, F.; Morel, I.; Lestremau, F.; Talanta2015 , 138, 231.
  • 124
    Holcapek, M.; Cervena, B.; Cifkova, E.; Lisa, M.; Chagovets, V.; Vostalova, J.; Bancirova, M.; Galuszka, J.; Hill, M.; J. Chromatogr. B2015 , 990, 52.
  • 125
    Wang, S. H.; Wu, H. Y.; Huang, X. L.; Geng, P. W.; Wen, C. C.; Ma, J. S.; Zhou, Y. F.; Wang, X. Q.; J. Chromatogr. B2015 , 990, 118.
  • 126
    Magiera, S.; Baranowski, J.; J. Pharm. Biomed. Anal.2015 , 109, 171.
  • 127
    Konstantina S.; Tsoukali, H.; Nikolaos R.; Gika, H.; Wilson, I. D.; Georgios, T.; J. Sep. Sci.2010 , 33, 716.
  • 128
    Cao, M. S.; Fraser, K.; Huege, J.; Featonby, T.; Rasmussen, S.; Jones, C.; Metabolomics2015 , 11, 696.
  • 129
    Meissen, J. K.; Hirahatake, K. M.; Adams, S. H.; Fiehn, O.;Metabolomics2015 , 11, 707.
  • 130
    Tufi, S.; Lamoree, M.; de Boer, J.; Leonards, P.; J. Chromatogr. B2015 , 1395, 79.
  • 131
    Vorkas, P. A.; Isaac, G.; Anwar, M. A.; Davies, A. H.; Want, E. J.; Nicholson, J. K.; Holmes, E.; Anal. Chem.2015 , 87, 4184.
  • 132
    Teleki, A.; Sanchez-Kopper, A.; Takors, R. Anal. Biochem.2015 , 475, 4.
  • 133
    Godejohann M.; J. Chromatogr. A2006 , 1156, 87.
  • 134
    Sfakianaki, A. G.; Stalikas, C. D.; Food Chem2015 , 184, 188.
  • 135
    Zuo, R., Zhou, S.; Zuo, Y.; Deg, Y.; Food Chem2015 , 182, 242.
  • 136
    Trivedi, D. K.; Iles, R. K.; Biomed. Chromatogr.2015 , 29, 240.
  • 137
    Rubio, N. C.; Hastedt, M.; Gonzalez, J.; Pragst, F.; Int. J. Legal Med.2015 , 129, 69.
  • 138
    Yuan, Y.; Liao, Z. J.; Wang, F. L.; Zhang, H. L.; Guo, Y.; Li, R. P.; J. Liq. Chromatogr. Relat. Technol.2013 , 36, 958.

Datas de Publicação

  • Publicação nesta coleção
    Fev 2016

Histórico

  • Recebido
    09 Jun 2015
  • Aceito
    22 Set 2015
Sociedade Brasileira de Química Secretaria Executiva, Av. Prof. Lineu Prestes, 748 - bloco 3 - Superior, 05508-000 São Paulo SP - Brazil, C.P. 26.037 - 05599-970, Tel.: +55 11 3032.2299, Fax: +55 11 3814.3602 - São Paulo - SP - Brazil
E-mail: quimicanova@sbq.org.br