ANÁLISE SIMULTÂNEA DOS INDICADORES BIOLÓGICOS DE EXPOSIÇÃO AOS SOLVENTES ETILBENZENO, ESTIRENO, TOLUENO E XILENO NA URINA POR CLAE-UV

SIMULTANEOUS ANALYSIS OF BIOLOGICAL INDICATORS OF EXPOSURE TO ETHYLBENZENE, STYRENE, TOLUENE AND XYLENE SOLVENTS IN URINE BY HPLC-UV

Resumo

A method for simultaneous analysis of phenylglyoxylic (PGA), mandelic (MA), hippuric (HA), orto-, meta- and para-methylhippuric (o-, m- and p-MHA) acids in urine, by high performance liquid chromatography with UV detection was developed and validated. The substances are biotransformation products used as biological indicators to assess the exposure of workers, respectively, to ethylbenzene, styrene, toluene and xylene. The urinary metabolites were separated using a C18 column and a mixture of 5 mM phosphate buffer pH 2.6:acetonitrile (92:8, v/v) as mobile phase. Urine samples were extracted with dichloromethane:ethyl acetate (70:30, v/v). The method used phenacetin as internal standard and was linear in the interval of 100-500 µg mL-1 for PGA and MA and 150-700 µg mL-1 for HA and MHA. The detection limits in µg mL-1, were 19.8 for PGA, 1.7 for MA, 4.1 for HA, 3.9 for o-MHA, 3.3 for m-MHA and 7.3 for p-MHA. Intra- and inter-assay precisions (as relative standard deviation) were all less than 15% and accuracy (as relative standard error) did not exceed 12.3%. The recovery was higher than 65% for all metabolites. The developed method will be applied to biological evaluation of workers exposure to these solvents.

Keywords:
volatile organic compounds; urine; biological monitoring; HPLC-UV


Keywords:
volatile organic compounds; urine; biological monitoring; HPLC-UV

INTRODUÇÃO

Os compostos orgânicos etilbenzeno, estireno, tolueno e xileno são amplamente utilizados como solventes em indústrias de plástico, resina, borracha e farmacêutica.11 Crandall, M. S.; Am. Ind. Hyg. Assoc. J. 1981, 42, 499.

2 Seedorff, L.; Olsen, E.; Ann. Occup. Hyg. 1990, 34, 371.

3 Truchon, G.; Brodeur, J.; Gerin, M.; J. Anal. Toxicol. 1990, 14, 227.
-44 Szúcs, S.; Tóth, L.; Legoza, J.; Sárváry, A.; Ádány, R.; Arch. Toxicol. 2002, 76, 560. Também são constituintes de combustíveis, tintas, vernizes e colas.44 Szúcs, S.; Tóth, L.; Legoza, J.; Sárváry, A.; Ádány, R.; Arch. Toxicol. 2002, 76, 560.

5 2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011.
-66 Onchoi, C.; Kongtip, P.; Yoosook, W.; Chantanajul, S.; Southeast Asian J. Trop. Med. Public Health 2008, 39, 1164. Trabalhadores expostos a níveis elevados desses solventes podem apresentar efeitos adversos, como irritação (da pele e do trato respiratório), alterações do sistema nervoso central, alterações funcionais nos nervos ótico e óptico.55 2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011. Portanto, a monitorização da exposição, tanto ambiental quanto biológica, são atividades complementares e fundamentais na prevenção de tais efeitos. A monitorização ambiental avalia o risco à saúde do indivíduo exposto através da concentração dos solventes na zona respiratória, comparando-a aos Limites de Exposição Ocupacional. Por sua vez, a monitorização biológica pode avaliar a dose interna do composto para verificar se a exposição do trabalhador está em níveis aceitáveis, por comparação com o respectivo Limite para o Indicador Biológico utilizado. Para os solventes, os marcadores biológicos (ou indicadores biológicos) podem ser a própria substância ou os seus metabólitos presentes em matrizes biológicas.77 Teass, A. W.; Biagini, R. E.; DeBord, D. G.; Hull, R D. Em Manual of Analytical Methods; Eller, P. M., ed.; NIOSH: Cincinnati, 1998, cap. F.

A principal via de absorção do etilbenzeno, estireno, tolueno e xileno é a via respiratória e a eliminação ocorre, principalmente, através dos metabólitos urinários.55 2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011. Os principais metabólitos do etilbenzeno e do estireno são o ácido mandélico (AM) e o ácido fenilglioxílico (AFG). No entanto, esses metabólitos não são indicadores específicos, pois também são formados quando há exposição a outras substâncias, por exemplo, estirenoglicol, óxido de estireno, fenilglicol, acetofenona, fenilglicina.55 2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011. O principal metabólito urinário do orto-, meta- e para-xileno são os ácidos orto-, meta- e para-metil hipúricos (o-, m- e p-AMH), respectivamente, os quais são completamente excretados dentro de 1 ou 2 dias depois da exposição, fato que limita a utilização desses metabólitos como indicadores biológicos apenas a exposições recentes.88 http://www.atsdr.cdc.gov/ToxProfiles/tp.asp?id=296&tid=53, acessada em Julho 2016.
http://www.atsdr.cdc.gov/ToxProfiles/tp....
Para o tolueno, o principal metabólito é o ácido hipúrico (AH). Os valores de referência (VR) e os índices biológicos máximos permitidos (IBMP) desses indicadores biológicos estabelecidos pela Norma Regulamentadora nº 7 (NR-7) e os BEIs® - Biological Exposure Indices propostos pela ACGIH (American Conference of Governmental Industrial Hygienists) - são apresentados na Tabela 1.55 2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011.,99 http://www.mtps.gov.br/images/Documentos/SST/NR/NR7.pdf, acessada em Julho 2016.
http://www.mtps.gov.br/images/Documentos...

Tabela 1
Indicadores biológicos de exposição ao etilbenzeno, estireno, tolueno e xileno na urina

Este estudo teve por objetivo desenvolver e validar um método simples para análise simultânea de AFG, AM, AH, o-AMH, m-AMH e p-AMH em urina por cromatografia líquida de alta eficiência com detecção por ultravioleta (CLAE-UV).

Esta Nota Técnica foi apresentada no 8º Congresso de Extensão Universitária da UNESP e seu resumo expandido encontra-se disponível online.1010 Esta Nota Técnica foi apresentada no 8º Congresso de Extensão Universitária da UNESP e seu resumo expandido encontra-se disponível no endereço http://200.145.6.205/index.php/congressoextensao/8congressoextensao/paper/viewFile/322./182
http://200.145.6.205/index.php/congresso...

PARTE EXPERIMENTAL

Reagentes

Os padrões do ácido fenilglioxílico (AFG, 97%), ácido mandélico (AM, 99%), ácido hipúrico (AH, 98%), ácido metil hipúrico (orto -AMH, meta-AMH e para-AMH, 98%) e fenacetina (padrão interno) foram obtidos da Sigma Aldrich (St Louis, MO, EUA). Os solventes metanol (J.T. Baker, Ecatepec, Estado do México, México), acetato de etila (Sigma Aldrich, St. Louis, MO, EUA), diclorometano (J. T. Baker, Center Valley, PA, EUA) e acetonitrila (J.T. Baker, Center Valley, PA, EUA) utilizados foram grau CLAE. O ácido o-fosfórico 85% (Synth, Diadema, Brasil), fosfato de sódio monobásico (Synth, Diadema, Brasil) e ácido clorídrico 37% (Qhemis, Brasil) são reagentes de grau analítico. A água utilizada em todas as soluções foi obtida através de um sistema de purificação Milli-Q Plus (Millipore, Belford, MA, EUA).

As soluções-padrão estoque foram preparadas separadamente dissolvendo os padrões em metanol para obter a concentração de 14 mg mL-1 de AFG e AM e 20 mg mL-1 de AH, o-AMH, m-AMH e p-AMH. O padrão interno fenacetina foi preparado na concentração de 2 mg mL-1 em metanol. Os padrões de calibração foram preparados diluindo as soluções-padrão estoque em metanol para obter concentrações variando de 400 a 2000 µg mL-1 para AFG e AM (400, 800, 1200, 1400 e 2000 µg mL-1) e de 600 a 2800 µg mL-1 para AH, o-AMH, m-AMH e p-AMH (600, 1000, 1400, 1960, 2800 µg mL-1).

Instrumentação

A análise dos metabólitos foi realizada por CLAE-UV utilizando um cromatógrafo líquido de alta eficiência Perkin Elmer, composto de um sistema binário de bombas Flexar LC Pump, degaseificador e detector Flexar UV/VIS. A coluna utilizada foi Brownlee C18 SSP (100 × 3,0 mm; 2,7 µm) da Perkin Elmer. O sistema de cromatografia foi controlado pelo programa Chromera UV/Vis (Perkin Elmer). A mistura de tampão fosfato 5 mM (pH 2,6):acetonitrila (92:8 v/v) foi utilizada como fase móvel, em fluxo de 0,6 mL min-1. A detecção por ultravioleta foi realizada a 254 nm nos primeiros 3,0 min e a 225 nm até o final da análise para todos os metabólitos.

Preparo das amostras

Alíquotas de 200 µL de urina e 20 µL de solução de fenacetina 2 mg mL-1 (padrão interno) foram adicionadas em um microtubo de 2 mL. A amostra foi acidificada com 20 µL de ácido clorídrico 1 M e submetida à extração líquido-líquido com 1 mL da mistura de diclorometano:acetato de etila (70:30, v/v). As amostras foram agitadas por 30 min e posteriormente centrifugadas a 10.600 × g por 10 min, a 4 ºC. Seiscentos microlitros da fase orgânica foram evaporados a secura em banho seco a 70 ºC. O resíduo foi reconstituído com 600 µL de fase móvel, filtrado em membrana de 0,45 µm (Analítica, São Paulo, Brasil) e uma alíquota de 20 µL do filtrado foi injetada no sistema CLAE-UV.

Curvas de calibração e linearidade

Três curvas de calibração para análise de AFG, AM, AH, o-AMH, m-AMH e p-AMH foram preparadas em alíquotas de 200 µL de urina branco coletada de indivíduos não expostos aos solventes precursores dos metabólitos em estudo enriquecidas com 50 µL de cada padrão de calibração. As curvas de calibração incluíram a análise de amostra branco e amostras com as seguintes concentrações finais de 100, 200, 300, 350 e 500 µg mL-1 para AFG e AM; 150, 250, 350, 490 e 700 µg mL-1 para AH, o-AMH, m-AMH e p-AMH em urina. As amostras da curva de calibração foram processadas e analisadas conforme descrito previamente.

Controles de qualidade foram preparados enriquecendo amostras de urina branco com três diferentes níveis de concentração (100, 200 e 420 µg mL-1 para AFG e AM; 140, 250 e 588 µg mL-1 para AH, o-AMH, m-AMH e p-AMH).

Limites de detecção (LD) e quantificação (LQ)

Os limites de detecção e de quantificação foram obtidos como recomenda o National Institute for Occupational Safety and Health (NIOSH).1111 Kennedy, E. R.; Fischbach, T. J.; Song, R.; Eller, P. M.; Shulman, S. A.; Guidelines for Air Sampling and Analytical Method Development and Evaluation, NIOSH: Cincinnati, 1995, apêndice 3. Em resumo, foram preparados padrões de calibração de baixa concentração em metanol variando de 100 a 2000 µg mL-1 para AFG, 6 a 120 µg mL-1 para AM, 4 a 80 µg mL-1 para AH e 30 a 600 µg mL-1 para o-AMH, m-AMH e p-AMH. Três curvas de calibração foram preparadas em alíquotas de 200 µL de urina branco coletada de indivíduos não expostos enriquecidas com 50 µL de cada padrão de calibração de baixa concentração. A triplicata da curva de calibração de baixa concentração foi utilizada para obter a equação de regressão linear (Equação 1) e o erro padrão da regressão (Sy) (Equação 2). O erro padrão da regressão e o coeficiente angular da equação de regressão linear foram usados para o cálculo do limite de detecção (Equação 3). O limite de quantificação foi calculado como 3,33 vezes o limite de detecção (Equação 4).

(1) Y = a X + b

(2) Sy = Σ ( Y ̂ i Y ) 2 ( N 2 )

(3) L D = 3 × S y c o e f i c i e n t e a n g u l a r

(4) LQ = 3 , 33 × LD

onde: Ŷi é a resposta predita de cada concentração do analito (X); Yi é a resposta obtida (Y); N é o número de padrões de calibração de baixa concentração (N=5)

Recuperação

A recuperação foi determinada enriquecendo alíquotas de 200 µL de urina branco com concentrações finais na urina de 200 e 350 µg mL-1 de AFG e AM, 250 e 490 µg mL-1 de AH, o-AMH, m-AMH e p-AMH, e 20 µL de solução de fenacetina (2 mg mL-1). As amostras (n=5) foram extraídas como descrito anteriormente. A recuperação foi calculada comparando a área dos picos das amostras extraídas conforme o protocolo analítico com a área obtida a partir das soluções-padrão de mesma concentração não submetidas ao processo de extração, sendo esta última considerada 100%.

Precisão e exatidão

A precisão e exatidão intra e interensaios foram investigadas utilizando os controles de qualidade de baixa (CQB), média (CQM) e alta (CQA) concentração analisados em quintuplicata em 3 dias diferentes. A precisão foi expressa como desvio padrão relativo (coeficiente de variação, CV) (Equação 5) e a exatidão como erro padrão relativo (EPR) (Equação 6).

(5) C V = d e s v i o p a d r ã o × 100 c o n c e n t r a ç ã o m é d i a e x p e r i m e n t a l

(6) E P R = ( c o n c e n t r a ç ã o m é d i a e x p e r i m e n t a l - v a l o r n o m i n a l ) × 100 v a l o r n o m i n a l

RESULTADOS E DISCUSSÃO

No presente estudo, foi desenvolvido e validado um método utilizando CLAE-UV capaz de analisar simultaneamente os indicadores biológicos de exposição aos solventes etilbenzeno, estireno, tolueno e xileno. Vários métodos para analisar AFG, AM, AH, o-AMH, m-AMH e p-AMH na urina foram descritos na literatura, porém poucos são capazes de analisar todos os metabólitos simultaneamente.44 Szúcs, S.; Tóth, L.; Legoza, J.; Sárváry, A.; Ádány, R.; Arch. Toxicol. 2002, 76, 560.,66 Onchoi, C.; Kongtip, P.; Yoosook, W.; Chantanajul, S.; Southeast Asian J. Trop. Med. Public Health 2008, 39, 1164.,1212 Ogata, M.; Taguchi, T.; Int. Arch. Occup. Environ. Health 1987, 59, 263.

13 Ohashi, Y.; Mamiya, T.; Mitani, K.; Wang, B.; Takigawa, T.; Kira, S.; Kataoka, H.; Anal. Chim. Acta 2006, 566, 167.
-1414 Antunes, M. V.; Patuzzi, A. L. M.; Linden, R.; Quim. Nova 2008, 31, 1865. Os métodos que utilizam cromatografia gasosa acoplada à espectrometria de massas têm sido amplamente utilizados. Entretanto, tais métodos exigem um processo de derivatização durante o preparo da amostra.44 Szúcs, S.; Tóth, L.; Legoza, J.; Sárváry, A.; Ádány, R.; Arch. Toxicol. 2002, 76, 560.,1313 Ohashi, Y.; Mamiya, T.; Mitani, K.; Wang, B.; Takigawa, T.; Kira, S.; Kataoka, H.; Anal. Chim. Acta 2006, 566, 167. Entre os métodos que utilizam a cromatografia líquida para a análise simultânea dos mesmos analitos, há aquele em que a separação dos analitos foi observada apenas após 50 minutos,1212 Ogata, M.; Taguchi, T.; Int. Arch. Occup. Environ. Health 1987, 59, 263. aquele que utiliza grande volume de amostra (2 mL) e solventes orgânicos ou não utilizam padrão interno.66 Onchoi, C.; Kongtip, P.; Yoosook, W.; Chantanajul, S.; Southeast Asian J. Trop. Med. Public Health 2008, 39, 1164.,1212 Ogata, M.; Taguchi, T.; Int. Arch. Occup. Environ. Health 1987, 59, 263.,1414 Antunes, M. V.; Patuzzi, A. L. M.; Linden, R.; Quim. Nova 2008, 31, 1865.

A Figura 1 apresenta os cromatogramas obtidos com o presente método. Os metabólitos foram completamente separados em menos de 30 minutos, com exceção do m-AMH e p- AMH, os quais coeluem (Figura 1A). No entanto, a separação completa desses isômeros não se faz necessária para a presente aplicação, uma vez que é recomendado o uso da soma dos isômeros do AMH como indicador biológico para a exposição a xilenos.55 2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011.,99 http://www.mtps.gov.br/images/Documentos/SST/NR/NR7.pdf, acessada em Julho 2016.
http://www.mtps.gov.br/images/Documentos...
O método mostrou-se seletivo, uma vez que não foram observados picos interferentes no tempo de retenção dos metabólitos, como pode ser observado ao comparar o cromatograma de uma urina branco (Figura 1C) com o cromatograma da solução-padrão (Figura 1B). Para que um método seja considerado seletivo, as respostas de picos interferentes próximo ao tempo de retenção do analito devem ser inferiores a 20% da resposta do analito nas amostras de LIQ.

Figura 1
Cromatogramas dos seis metabólitos obtidos com o presente método (a) Urina não exposta enriquecida com solução-padrão (300 µg mL-1 de AFG e AM e 350 µg mL-1 de AH, o-AMH, m-AMH e p-AMH), (b) Solução-padrão (300 µg mL-1 de AFG e AM e 350 µg mL-1 de AH, o-AMH, m-AMH e p-AMH) e (c) Urina branco [1. AFG, 2. AM, 3. AH, 4. o-AMH, 5. m-AMH, 6. p-AMH, 7. Fenacetina (P.I.)]

A análise de uma urina branco proveniente de indivíduo não exposto ao tolueno apresenta um pico de AH (Figura 1C), que foi encontrado na urina humana provavelmente devido ao consumo de alimentos que contêm ácido benzoico ou benzoato como conservador.55 2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011.,1515 Ogawa, M.; Suzuki, Y.; Endo, Y.; Kawamoto, T.; Kayama, F.; Ind. Health. 2011, 49, 195.-1616 Antunes, M. V.; Niederauer, C. G.; Linden, R.; Clin. Biochem. 2013, 46, 1276. Esse conservador está presente em refrigerantes, sucos de frutas, margarina, molho de tomate, enlatados.1717 Tfouni, S. A. V.; Toledo, M. C. F.; Food Addit. Contam. 2002, 19, 647.

18 Gonzalez, K. C.; Sagebin, F. R.; Oliveira, P. G.; Glock, L.; Thiesen, F. V.; Ciência & Saúde Coletiva 2010, 15, 1637.
-1919 Alves, A. C.; Rangel, C. V.; Pratte-Santos, R.; Revista Sapientia 2012, 11, 34. Um estudo identificou os refrigerantes como a principal fonte de benzoato no Brasil, representando mais de 80% da ingestão diária do conservador.1717 Tfouni, S. A. V.; Toledo, M. C. F.; Food Addit. Contam. 2002, 19, 647. O ácido benzoico ou benzoato é metabolizado no fígado e nos rins formando o AH.2020 Badenhorst, C. P. S.; Erasmus, E.; Van Der Sluis, R.; Nortje, C.; Van Dijk, A. A.; Drug Metab. Rev. 2014, 46, 343. A alimentação e a variação interindividual no metabolismo do ácido benzoico influenciam diretamente na quantidade basal de AH presente na urina. A ACGIH, atualmente, não recomenda o uso do AH como indicador biológico para a exposição a tolueno, pois a elevada quantidade basal de AH pode dificultar a interpretação no limite de exposição atualmente recomendado para o tolueno, de 75 mg m-3. No entanto, no Brasil a legislação ainda prevê o uso do AH como indicador biológico de exposição ao tolueno.99 http://www.mtps.gov.br/images/Documentos/SST/NR/NR7.pdf, acessada em Julho 2016.
http://www.mtps.gov.br/images/Documentos...
Vale ressaltar que a legislação brasileira referente aos indicadores biológicos de exposição ao tolueno não é atualizada desde 1994.99 http://www.mtps.gov.br/images/Documentos/SST/NR/NR7.pdf, acessada em Julho 2016.
http://www.mtps.gov.br/images/Documentos...

Trabalhos anteriores sugerem realizar a análise de AFG a 254 nm, pois o metabólito apresenta uma resposta melhor do que a 225 nm.1212 Ogata, M.; Taguchi, T.; Int. Arch. Occup. Environ. Health 1987, 59, 263.,2121 Ogata, M.; Sugihara, R.; Int. Arch. Occup. Environ. Health 1978, 42, 11. Nos cromatogramas apresentados na Figura 2 é possível observar a diferença na resposta do AFG quando analisado a 225 nm (Figura 2A) e a 254 nm (Figura 2B). Portanto, o presente método propõe a análise do AFG a 254 nm e os demais metabólitos a 225 nm.

Figura 2
Cromatogramas utilizando diferentes comprimentos de onda (a) Análise utilizando apenas λ = 225 nm e (b) Análise utilizando λ = 254 nm nos primeiros 3 min e λ = 225 nm a partir dos 3 min até o final da corrida [1. AFG, 2. AM, 3. AH, 4. o-AMH, 5. m-AMH, 6. p-AMH, 7. Fenacetina (P.I.)]

A curva de calibração foi gerada através do gráfico da razão entre área do analito/área do padrão interno vs. a concentração do analito. As curvas de calibração para o AFG, AM, AH, o-AMH, m-AMH e p-AMH em urina enriquecida foram lineares nos intervalos de concentração estudados (Tabela 2).

Tabela 2
Parâmetros da curva de calibração, limite de detecção (LD) e limite de quantificação (LQ) dos metabólitos

O ensaio de recuperação está relacionado a eficiência da extração e foi superior a 65% para todos os metabólitos e o padrão interno e não variou em função da concentração (Tabela 3). Embora a porcentagem de recuperação próxima a 100% tanto para o analito como para o padrão interno seja desejável, porcentagens inferiores são admitidas desde que a recuperação seja reprodutível entre diferentes concentrações do analito. Os limites de detecção e quantificação foram calculados a partir da equação de regressão linear e do erro padrão da regressão. Os limites de detecção obtidos para AFG, AM, AH, o-AMH, m-AMH e p-AMH foram de 19,83 µg mL-1, 1,73 µg mL-1, 4,12 µg mL-1, 3,93 µg mL-1, 3,29 µg mL-1 e 7,29 µg mL-1, respectivamente (Tabela 2) e são compatíveis com a aplicação do método para avaliação biológica da exposição ocupacional aos solventes estireno, etilbenzeno, xilenos e tolueno.

Tabela 3
Recuperação do AFG, AM, AH, o-AMH, m-AMH, p-AMH e fenacetina adicionados em urina não exposta

Alíquotas de urina coletada de indivíduo não exposto foram enriquecidas com AFG, AM, AH, o-AMH, m-AMH e p-AMH e analisadas em quintuplicata para determinar a precisão e exatidão intra e interensaio. Todos os metabólitos apresentaram coeficiente de variação menor que 15%. A exatidão, expressa como erro padrão relativo, apresentou variação menor que 12,3% para os três níveis de concentração testados (Tabela 4).

Tabela 4
Precisão e exatidão intra e interensaio do método de análise de AFG, AM, AH e AMH em urina por CLAE-UV (n=5)

CONCLUSÃO

Foi desenvolvido e validado um método utilizando CLAE-UV capaz de analisar simultaneamente os indicadores biológicos AM, AFG, AH, o-AMH, m-AMH e p-AMH em urina. A sensibilidade, precisão e exatidão do método mostraram-se compatíveis com a sua aplicação na monitorização biológica da exposição ocupacional aos solventes etilbenzeno, estireno, tolueno e xileno conforme recomendado pela NR-7 e ACGIH.

AGRADECIMENTOS

Os autores agradecem a Pró-Reitoria de Extensão Universitária da UNESP (PROEX) e a Coordenadoria de Saúde e Segurança do Trabalhador e Sustentabilidade Ambiental (COSTSA).

REFERÊNCIAS

  • 1
    Crandall, M. S.; Am. Ind. Hyg. Assoc. J. 1981, 42, 499.
  • 2
    Seedorff, L.; Olsen, E.; Ann. Occup. Hyg. 1990, 34, 371.
  • 3
    Truchon, G.; Brodeur, J.; Gerin, M.; J. Anal. Toxicol. 1990, 14, 227.
  • 4
    Szúcs, S.; Tóth, L.; Legoza, J.; Sárváry, A.; Ádány, R.; Arch. Toxicol. 2002, 76, 560.
  • 5
    2011 TLVs and BEIs with 7th Edition Documentation 2011 [Cd-room], ACGIH: Cincinatti, 2011.
  • 6
    Onchoi, C.; Kongtip, P.; Yoosook, W.; Chantanajul, S.; Southeast Asian J. Trop. Med. Public Health 2008, 39, 1164.
  • 7
    Teass, A. W.; Biagini, R. E.; DeBord, D. G.; Hull, R D. Em Manual of Analytical Methods; Eller, P. M., ed.; NIOSH: Cincinnati, 1998, cap. F.
  • 8
    http://www.atsdr.cdc.gov/ToxProfiles/tp.asp?id=296&tid=53, acessada em Julho 2016.
    » http://www.atsdr.cdc.gov/ToxProfiles/tp.asp?id=296&tid=53
  • 9
    http://www.mtps.gov.br/images/Documentos/SST/NR/NR7.pdf, acessada em Julho 2016.
    » http://www.mtps.gov.br/images/Documentos/SST/NR/NR7.pdf
  • 10
    Esta Nota Técnica foi apresentada no 8º Congresso de Extensão Universitária da UNESP e seu resumo expandido encontra-se disponível no endereço http://200.145.6.205/index.php/congressoextensao/8congressoextensao/paper/viewFile/322./182
    » http://200.145.6.205/index.php/congressoextensao/8congressoextensao/paper/viewFile/322./182
  • 11
    Kennedy, E. R.; Fischbach, T. J.; Song, R.; Eller, P. M.; Shulman, S. A.; Guidelines for Air Sampling and Analytical Method Development and Evaluation, NIOSH: Cincinnati, 1995, apêndice 3.
  • 12
    Ogata, M.; Taguchi, T.; Int. Arch. Occup. Environ. Health 1987, 59, 263.
  • 13
    Ohashi, Y.; Mamiya, T.; Mitani, K.; Wang, B.; Takigawa, T.; Kira, S.; Kataoka, H.; Anal. Chim. Acta 2006, 566, 167.
  • 14
    Antunes, M. V.; Patuzzi, A. L. M.; Linden, R.; Quim. Nova 2008, 31, 1865.
  • 15
    Ogawa, M.; Suzuki, Y.; Endo, Y.; Kawamoto, T.; Kayama, F.; Ind. Health. 2011, 49, 195.
  • 16
    Antunes, M. V.; Niederauer, C. G.; Linden, R.; Clin. Biochem. 2013, 46, 1276.
  • 17
    Tfouni, S. A. V.; Toledo, M. C. F.; Food Addit. Contam. 2002, 19, 647.
  • 18
    Gonzalez, K. C.; Sagebin, F. R.; Oliveira, P. G.; Glock, L.; Thiesen, F. V.; Ciência & Saúde Coletiva 2010, 15, 1637.
  • 19
    Alves, A. C.; Rangel, C. V.; Pratte-Santos, R.; Revista Sapientia 2012, 11, 34.
  • 20
    Badenhorst, C. P. S.; Erasmus, E.; Van Der Sluis, R.; Nortje, C.; Van Dijk, A. A.; Drug Metab. Rev. 2014, 46, 343.