Contact force-guided catheter ablation for the treatment of atrial fibrillation: a meta-analysis of randomized, controlled trials

Abstract

Contact force (CF) sensing technology allows real-time monitoring during catheter ablation for atrial fibrillation (AF). However, the effect of CF sensing technology on procedural parameters and clinical outcomes still needs clarification. Because of the inconsistent results thus far in this area, we performed a meta-analysis to determine whether CF sensing technology can improve procedural parameters and clinical outcomes for the treatment of AF. Studies examining the benefits of CF sensing technology were identified in English-language articles by searching the MEDLINE, Web of Science, and Cochrane Library databases (inception to May 2015). Ten randomized, controlled trials involving 1834 patients (1263 males, 571 females) were included in the meta-analysis (681 in the CF group, 1153 in the control group). Overall, the ablation time was significantly decreased by 7.34 min (95%CI=-12.21 to -2.46; P=0.003, Z test) in the CF group compared with the control group. CF sensing technology was associated with significantly improved freedom from AF after 12 months (OR=1.55, 95%CI=1.20 to 1.99; P=0.0007) and complications were significantly lower in the CF group than in the control group (OR=0.50, 95%CI=0.29 to 0.87; P=0.01). However, fluoroscopy time analysis showed no significantly decreased trend associated with CF-guided catheter ablation (weighted mean difference: -2.59; 95%CI=-9.06 to 3.88; P=0.43). The present meta-analysis shows improvement in ablation time and freedom from AF after 12 months in AF patients treated with CF-guided catheter ablation. However, CF-guided catheter ablation does not decrease fluoroscopy time.

Atrial fibrillation; Contact force sensing; Meta-analysis; Randomized, controlled trials


Introduction

Radiofrequency catheter ablation (RFCA) is a potentially curative method for treatment of atrial fibrillation (AF). The 2012 guidelines recommend catheter ablation in cases of symptomatic paroxysmal AF or for those intolerant to antiarrhythmic medication (11. Calkins H, Kuck KH, Cappato R, Brugada J, Camm AJ, Chen SA, et al. 2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace 2012; 14: 528-606, doi: 10.1093/europace/eus027.
https://doi.org/10.1093/europace/eus027...
). Pulmonary vein isolation remains the cornerstone of the strategy for all catheter ablation procedures (22. Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation - developed with the special contribution of the European Heart Rhythm Association. Europace 2012; 14: 1385-1413.). Probably the most important limitation of RFCA for AF is the rate of recurrence owing to electrical reconnection of the pulmonary veins (33. Tilz RR, Chun KR, Schmidt B, Fuernkranz A, Wissner E, Koester I, et al. Catheter ablation of long-standing persistent atrial fibrillation: a lesson from circumferential pulmonary vein isolation. J Cardiovasc Electrophysiol 2010; 21: 1085-1093, doi: 10.1111/j.1540-8167.2010.01799.x.
https://doi.org/10.1111/j.1540-8167.2010...
,44. Ouyang F, Antz M, Ernst S, Hachiya H, Mavrakis H, Deger FT, et al. Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation 2005; 111: 127-135, doi: 10.1161/01.CIR.0000151289.73085.36.
https://doi.org/10.1161/01.CIR.000015128...
). This recurrence is associated with poor quality of life; and, if these reconnections were eliminated, long-term success rates would improve (55. Matsuo S, Yamane T, Date T, Inada K, Kanzaki Y, Tokuda M, et al. Reduction of AF recurrence after pulmonary vein isolation by eliminating ATP-induced transient venous re-conduction. J Cardiovasc Electrophysiol 2007; 18: 704-708, doi: 10.1111/j.1540-8167.2007.00842.x.
https://doi.org/10.1111/j.1540-8167.2007...
,66. Yamane T, Matsuo S, Date T, Lellouche N, Hioki M, Narui R, et al. Repeated provocation of time- and ATP-induced early pulmonary vein reconnections after pulmonary vein isolation: eliminating paroxysmal atrial fibrillation in a single procedure. Circ Arrhythm Electrophysiol 2011; 4: 601-608, doi: 10.1161/CIRCEP.110.960138.
https://doi.org/10.1161/CIRCEP.110.96013...
).

Contact force (CF), which occurs between the catheter tip and target tissue, is a crucial determinant of the lesion characteristics of radiofrequency. Several studies (77. Kuck KH, Reddy VY, Schmidt B, Natale A, Neuzil P, Saoudi N, et al. A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm 2012; 9: 18-23, doi: 10.1016/j.hrthm.2011.08.021.
https://doi.org/10.1016/j.hrthm.2011.08....
,88. Kumar S, Chan M, Lee J, Wong MC, Yudi M, Morton JB, et al. Catheter-tissue contact force determines atrial electrogram characteristics before and lesion efficacy after antral pulmonary vein isolation in humans. J Cardiovasc Electrophysiol 2014; 25: 122-129, doi: 10.1111/jce.12293.
https://doi.org/10.1111/jce.12293...
) have reported that the use of a CF-guided catheter may significantly improve procedural parameters and clinical outcomes. However, CF sensing technology does not improve clinical outcomes in patients with paroxysmal AF (99. Stabile G, Solimene F, Calo L, Anselmino M, Castro A, Pratola C, et al. Catheter-tissue contact force values do not impact mid-term clinical outcome following pulmonary vein isolation in patients with paroxysmal atrial fibrillation. J Interv Card Electrophysiol 2015; 42: 21-26, doi: 10.1007/s10840-014-9947-2.
https://doi.org/10.1007/s10840-014-9947-...
). Therefore, we performed a meta-analysis to determine whether the use of CF sensing technology can improve procedural parameters and clinical outcomes in the treatment of AF.

Material and Methods

Study search strategy

Studies that investigated the benefits of CF sensing technology were identified in English-language articles by searching the MEDLINE, Web of Science, and Cochrane Library databases (inception to May 2015). We used the search terms: "atrial fibrillation", "contact force", "pulmonary vein isolation", and "catheter ablation".

Inclusion criteria

We identified eligible articles on the basis of the following inclusion criteria: 1) study design (randomized, controlled trials), 2) target population (AF patients who underwent RFCA), 3) intervention (studies involving CF-guided catheter ablation), and 4) outcomes (procedural parameters and clinical outcomes were examined in the target population). When multiple studies shared the same subject population, we included only the most recent study.

Data extraction

Two reviewers independently extracted data from all eligible studies fulfilling the inclusion criteria. Disagreement was resolved by discussion between the two reviewers. We extracted the following data from the included studies: the first author, publication date, region of origin, number of patients, ablation device, study design, and baseline characteristics. For data that were not provided in the main text, the required information was obtained in part from the supplementary online appendix.

Statistical methods

The Cochrane Collaboration meta-analysis review methodology was used for this study. Continuous variables with a normal distribution are reported as means±SD. The effects of continuous variables were evaluated as weighted mean differences (WMDs). Odds ratios (ORs) were used as summary statistics for discontinuous variables. The presence of heterogeneity across studies was evaluated. P≤0.10 was considered to be significant for statistical heterogeneity (1010. Higgins JP, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med 2002; 21: 1539-1558, doi: 10.1002/sim.1186.
https://doi.org/10.1002/sim.1186...
). All statistical analyses were performed with RevMan version 4.2.2, which is available from the Cochrane Collaboration website (http://www.cochrane.org/cochrane/hbook/htm).

Results

Identification of studies

A total of 364 potentially eligible citations were identified using our search strategy. After the initial screening, 33 relevant articles were selected for further review. Among these, 23 articles were excluded according to the inclusion criteria. Ten randomized, controlled trials involving 1834 patients were eventually identified and included in the meta-analysis (681 in the CF group, 1153 in the control group) (1111. Akca F, Janse P, Theuns DA, Szili-Torok T. A prospective study on safety of catheter ablation procedures: contact force guided ablation could reduce the risk of cardiac perforation. Int J Cardiol 2015; 179: 441-448, doi: 10.1016/j.ijcard.2014.11.105.
https://doi.org/10.1016/j.ijcard.2014.11...
1212. Andrade JG, Monir G, Pollak SJ, Khairy P, Dubuc M, Roy D, et al. Pulmonary vein isolation using "contact force" ablation: the effect on dormant conduction and long-term freedom from recurrent atrial fibrillation - a prospective study. Heart Rhythm 2014; 11: 1919-1924, doi: 10.1016/j.hrthm.2014.07.033.
https://doi.org/10.1016/j.hrthm.2014.07....
1313. Jarman JW, Panikker S, DAS M, Wynn GJ, Ullah W, Kontogeorgis A, et al. Relationship between contact force sensing technology and medium-term outcome of atrial fibrillation ablation: a multicenter study of 600 patients. J Cardiovasc Electrophysiol 2015; 26: 378-384, doi: 10.1111/jce.12606.
https://doi.org/10.1111/jce.12606...
1414. Jourda F, Providencia R, Marijon E, Bouzeman A, Hireche H, Khoueiry Z, et al. Contact-force guided radiofrequency vs. second-generation balloon cryotherapy for pulmonary vein isolation in patients with paroxysmal atrial fibrillation - a prospective evaluation. Europace 2015; 17: 225-231, doi: 10.1093/europace/euu215.
https://doi.org/10.1093/europace/euu215...
1515. Marijon E, Fazaa S, Narayanan K, Guy-Moyat B, Bouzeman A, Providencia R, et al. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results. J Cardiovasc Electrophysiol 2014; 25: 130-137, doi: 10.1111/jce.12303.
https://doi.org/10.1111/jce.12303...
1616. Martinek M, Lemes C, Sigmund E, Derndorfer M, Aichinger J, Winter S, et al. Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing Clin Electrophysiol 2012; 35: 1312-1318, doi: 10.1111/j.1540-8159.2012.03503.x.
https://doi.org/10.1111/j.1540-8159.2012...
1717. Sciarra L, Golia P, Natalizia A, de Ruvo E, Dottori S, Scara A, et al. Which is the best catheter to perform atrial fibrillation ablation? A comparison between standard ThermoCool, SmartTouch, and Surround Flow catheters. J Interv Card Electrophysiol 2014; 39: 193-200, doi: 10.1007/s10840-014-9874-2.
https://doi.org/10.1007/s10840-014-9874-...
1818. Sigmund E, Puererfellner H, Derndorfer M, Kollias G, Winter S, Aichinger J, et al. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement - a case-matched comparison in 198 patients. Pacing Clin Electrophysiol 2015; 38: 201-208, doi: 10.1111/pace.12549.
https://doi.org/10.1111/pace.12549...
1919. Wakili R, Clauss S, Schmidt V, Ulbrich M, Hahnefeld A, Schussler F, et al. Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation. Clin Res Cardiol 2014; 103: 97-106, doi: 10.1007/s00392-013-0625-7.
https://doi.org/10.1007/s00392-013-0625-...
2020. Wutzler A, Huemer M, Parwani AS, Blaschke F, Haverkamp W, Boldt LH. Contact force mapping during catheter ablation for atrial fibrillation: procedural data and one-year follow-up. Arch Med Sci 2014; 10: 266-272, doi: 10.5114/aoms.2014.42578.
https://doi.org/10.5114/aoms.2014.42578...
).

Characteristics of the studies

Table 1 shows the characteristics of the 10 clinical trials that were published between 2012 and 2015. The duration of follow-up ranged from the time of discharge to 12 months. Baseline characteristics of the two study groups were well balanced with respect to baseline features. In the CF group, the SmartTouch (Biosense Webster, USA) or Tacticath (Endosense SA, Switzerland) ablation catheter was used, while a standard catheter or cryoballoon catheter was used in the control group.

Procedural parameters

Table 2 shows the main results of the pooled WMDs and ORs in the meta-analysis. The procedural parameters mainly included ablation time, overall duration of the procedure, and fluoroscopy time. Data on ablation time were available in nine randomized, controlled trials. CF sensing technology was associated with a significantly decreased ablation time (WMD=-7.34; 95% confidence interval [CI]=-12.21 to -2.46; P=0.003; Figure 1). An analysis of overall duration of the procedure indicated that CF-guided catheter ablation advantageously reduced the overall duration (WMD=-19.43; 95%CI=-29.61 to -9.26; P=0.0002; Table 2). However, the WMD estimate indicated that CF sensing technology did not significantly decrease the fluoroscopy time compared with in the control group (WMD=-2.59; 95%CI=-9.06 to 3.88; P=0.43; Figure 2).

Figure 1
Ablation time in the CF group compared to the control group with catheter ablation of atrial fibrillation (WMD=-7.34; 95%CI=-12.21 to -2.46; P=0.003; Z test). CF: contact force; WMD: weighted mean difference. For reference details, see Table 1.
Figure 2
Forest plot of fluoroscopy time meta-analysis of contact force (CF) sensing technology in catheter ablation of atrial fibrillation (weighted mean difference [WMD]: -2.59; 95%CI=-9.06 to 3.88; P=0.43; Z test). For reference details, see Table 1.

Clinical outcomes

Efficacy analysis of freedom from AF after 12 months showed that CF-guided catheter ablation was associated with a significant improvement in freedom from AF (OR=1.55; 95%CI=1.20 to 1.99; P=0.0007; Figure 3). Additionally, CF sensing technology significantly decreased complications (OR=0.50; 95%CI=0.29 to 0.87; P=0.01; Figure 4).

Figure 3
Cumulative OR estimate of freedom from atrial fibrillation after 12 months in the contact force (CF) group compared to the control group (OR=1.55; 95%CI=1.20 to 1.99; P=0.0007; Z test). For reference details, see Table 1.
Figure 4
Comparison of complications in the contact force (CF) group compared to the control group (OR=0.50; 95%CI=0.29 to 0.87; P=0.01; Z test). For reference details, see Table 1.

Sensitivity analysis

A single study involved in the meta-analysis was deleted each time to reflect the influence of the individual data set on the pooled WMDs or ORs. We found that the corresponding WMDs or ORs were not materially altered, which indicated that our results were statistically robust.

Publication bias

Begg's funnel plot and Egger's test were performed to determine the publication bias of the literature. Funnel plot analysis did not show any evidence of obvious asymmetry and possible publication bias should not have substantially influenced the results of this meta-analysis.

Discussion

CF-guided catheter ablation is a new technology for ablating AF. Therefore, the effect of this technology on procedural parameters and clinical outcomes needs to be determined. To circumvent the reversibility of pulmonary vein isolation, ensuring that first-time RF lesions are optimal and result in permanent myocardial tissue damage is important (2121. Kowalski M, Grimes MM, Perez FJ, Kenigsberg DN, Koneru J, Kasirajan V, et al. Histopathologic characterization of chronic radiofrequency ablation lesions for pulmonary vein isolation. J Am Coll Cardiol 2012; 59: 930-938, doi: 10.1016/j.jacc.2011.09.076.
https://doi.org/10.1016/j.jacc.2011.09.0...
). Optimal contact between the catheter tip and tissue during RF ablation appears to be important for providing the best radiofrequency ablation lesions (2222. Shah D, Lambert H, Langenkamp A, Vanenkov Y, Leo G, Gentil-Baron P, et al. Catheter tip force required for mechanical perforation of porcine cardiac chambers. Europace 2011; 13: 277-283, doi: 10.1093/europace/euq403.
https://doi.org/10.1093/europace/euq403...
2323. Neuzil P, Reddy VY, Kautzner J, Petru J, Wichterle D, Shah D, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol 2013; 6: 327-333, doi: 10.1161/CIRCEP.113.000374.
https://doi.org/10.1161/CIRCEP.113.00037...
2424. Reddy VY, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C, et al. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm 2012; 9: 1789-1795, doi: 10.1016/j.hrthm.2012.07.016.
https://doi.org/10.1016/j.hrthm.2012.07....
). Currently, several parameters are used to gauge adequate contact, such as pure tactile feedback, impedance, and intracardiac electrograms. Nevertheless, these parameters are unable to precisely ensure CF because of differences in the catheter shaft, deflection, torque, and types of sheaths that are used in the procedure. Additionally, electrogram reduction is a particularly poor predictor of transmurality (2525. Okumura Y, Johnson SB, Bunch TJ, Henz BD, O'Brien CJ, Packer DL. A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. J Cardiovasc Electrophysiol 2008; 19: 632-640, doi: 10.1111/j.1540-8167.2008.01135.x.
https://doi.org/10.1111/j.1540-8167.2008...
2626. Kumar S, Haqqani HM, Chan M, Lee J, Yudi M, Wong MC, et al. Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans. Heart Rhythm 2013; 10: 962-969, doi: 10.1016/j.hrthm.2013.03.022.
https://doi.org/10.1016/j.hrthm.2013.03....
2727. Yokoyama K, Nakagawa H, Shah DC, Lambert H, Leo G, Aeby N, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol 2008; 1: 354-362, doi: 10.1161/CIRCEP.108.803650.
https://doi.org/10.1161/CIRCEP.108.80365...
).

Appropriate CF is important for catheter ablation because insufficient CF results in ineffective delivery of radiofrequency, whereas excessive CF can cause collateral tissue injury. Continuous CF monitoring has many advantages for catheter ablation. First, this monitoring ensures that the catheter is correctly placed, including application and direction. Second, continuous CF monitoring can monitor the catheter's movements in real time. Therefore, taking the abovementioned factors into account, using CF sensing technology during RFCA for AF is a useful development.

Our meta-analysis demonstrated that CF sensing technology could not only reduce ablation time and the overall duration of procedure, but also improve freedom from AF after 12 months. Although CF-sensing did not reduce fluoroscopy time, we believe that this will eventually be achieved as operators gain more experience in the future. With regard to operators, real-time CF measurements will allow them to favorably modulate the CF and thus the lesion size. Therefore, a point with a CF that is too low or too high because of difficulty in catheter positioning could potentially be compensated for by changing the radiofrequency power or the duration (2828. Thiagalingam A, D'Avila A, Foley L, Guerrero JL, Lambert H, Leo G, et al. Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. J Cardiovasc Electrophysiol 2010; 21: 806-811, doi: 10.1111/j.1540-8167.2009.01693.x.
https://doi.org/10.1111/j.1540-8167.2009...
).

Even though we performed a pooled analysis, which included 10 randomized, controlled trials, some caution is needed in the interpretation of our results. Overall, the meta-analysis was not performed using individual patient data. Therefore, prespecified data were partly extracted from the studies for analysis. Several studies with a small sample size were included, which were likely to involve selected cohorts of patients and operators. Additionally, potential heterogeneity among the clinical trials due to varying inclusion criteria, the definition of variables, and different ablation devices, cannot be excluded.

In summary, the present meta-analysis showed improvement in ablation time and freedom from AF after 12 months in AF patients treated with CF-guided catheter ablation. However, CF-guided catheter ablation does not decrease fluoroscopy time. Because of the relatively small sample size of our meta-analysis, larger-scale, prospectively designed, randomized, double-blinded trials should be carried out to clarify the potential benefits of CF-guided catheter ablation for the treatment of AF.

Acknowledgments

This study was supported in part by grants from the National Natural Science Foundation in China (#81100157 and #81470496).

References

  • 1
    Calkins H, Kuck KH, Cappato R, Brugada J, Camm AJ, Chen SA, et al. 2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace 2012; 14: 528-606, doi: 10.1093/europace/eus027.
    » https://doi.org/10.1093/europace/eus027
  • 2
    Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation - developed with the special contribution of the European Heart Rhythm Association. Europace 2012; 14: 1385-1413.
  • 3
    Tilz RR, Chun KR, Schmidt B, Fuernkranz A, Wissner E, Koester I, et al. Catheter ablation of long-standing persistent atrial fibrillation: a lesson from circumferential pulmonary vein isolation. J Cardiovasc Electrophysiol 2010; 21: 1085-1093, doi: 10.1111/j.1540-8167.2010.01799.x.
    » https://doi.org/10.1111/j.1540-8167.2010.01799.x
  • 4
    Ouyang F, Antz M, Ernst S, Hachiya H, Mavrakis H, Deger FT, et al. Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation 2005; 111: 127-135, doi: 10.1161/01.CIR.0000151289.73085.36.
    » https://doi.org/10.1161/01.CIR.0000151289.73085.36
  • 5
    Matsuo S, Yamane T, Date T, Inada K, Kanzaki Y, Tokuda M, et al. Reduction of AF recurrence after pulmonary vein isolation by eliminating ATP-induced transient venous re-conduction. J Cardiovasc Electrophysiol 2007; 18: 704-708, doi: 10.1111/j.1540-8167.2007.00842.x.
    » https://doi.org/10.1111/j.1540-8167.2007.00842.x
  • 6
    Yamane T, Matsuo S, Date T, Lellouche N, Hioki M, Narui R, et al. Repeated provocation of time- and ATP-induced early pulmonary vein reconnections after pulmonary vein isolation: eliminating paroxysmal atrial fibrillation in a single procedure. Circ Arrhythm Electrophysiol 2011; 4: 601-608, doi: 10.1161/CIRCEP.110.960138.
    » https://doi.org/10.1161/CIRCEP.110.960138
  • 7
    Kuck KH, Reddy VY, Schmidt B, Natale A, Neuzil P, Saoudi N, et al. A novel radiofrequency ablation catheter using contact force sensing: Toccata study. Heart Rhythm 2012; 9: 18-23, doi: 10.1016/j.hrthm.2011.08.021.
    » https://doi.org/10.1016/j.hrthm.2011.08.021
  • 8
    Kumar S, Chan M, Lee J, Wong MC, Yudi M, Morton JB, et al. Catheter-tissue contact force determines atrial electrogram characteristics before and lesion efficacy after antral pulmonary vein isolation in humans. J Cardiovasc Electrophysiol 2014; 25: 122-129, doi: 10.1111/jce.12293.
    » https://doi.org/10.1111/jce.12293
  • 9
    Stabile G, Solimene F, Calo L, Anselmino M, Castro A, Pratola C, et al. Catheter-tissue contact force values do not impact mid-term clinical outcome following pulmonary vein isolation in patients with paroxysmal atrial fibrillation. J Interv Card Electrophysiol 2015; 42: 21-26, doi: 10.1007/s10840-014-9947-2.
    » https://doi.org/10.1007/s10840-014-9947-2
  • 10
    Higgins JP, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med 2002; 21: 1539-1558, doi: 10.1002/sim.1186.
    » https://doi.org/10.1002/sim.1186
  • 11
    Akca F, Janse P, Theuns DA, Szili-Torok T. A prospective study on safety of catheter ablation procedures: contact force guided ablation could reduce the risk of cardiac perforation. Int J Cardiol 2015; 179: 441-448, doi: 10.1016/j.ijcard.2014.11.105.
    » https://doi.org/10.1016/j.ijcard.2014.11.105
  • 12
    Andrade JG, Monir G, Pollak SJ, Khairy P, Dubuc M, Roy D, et al. Pulmonary vein isolation using "contact force" ablation: the effect on dormant conduction and long-term freedom from recurrent atrial fibrillation - a prospective study. Heart Rhythm 2014; 11: 1919-1924, doi: 10.1016/j.hrthm.2014.07.033.
    » https://doi.org/10.1016/j.hrthm.2014.07.033
  • 13
    Jarman JW, Panikker S, DAS M, Wynn GJ, Ullah W, Kontogeorgis A, et al. Relationship between contact force sensing technology and medium-term outcome of atrial fibrillation ablation: a multicenter study of 600 patients. J Cardiovasc Electrophysiol 2015; 26: 378-384, doi: 10.1111/jce.12606.
    » https://doi.org/10.1111/jce.12606
  • 14
    Jourda F, Providencia R, Marijon E, Bouzeman A, Hireche H, Khoueiry Z, et al. Contact-force guided radiofrequency vs. second-generation balloon cryotherapy for pulmonary vein isolation in patients with paroxysmal atrial fibrillation - a prospective evaluation. Europace 2015; 17: 225-231, doi: 10.1093/europace/euu215.
    » https://doi.org/10.1093/europace/euu215
  • 15
    Marijon E, Fazaa S, Narayanan K, Guy-Moyat B, Bouzeman A, Providencia R, et al. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results. J Cardiovasc Electrophysiol 2014; 25: 130-137, doi: 10.1111/jce.12303.
    » https://doi.org/10.1111/jce.12303
  • 16
    Martinek M, Lemes C, Sigmund E, Derndorfer M, Aichinger J, Winter S, et al. Clinical impact of an open-irrigated radiofrequency catheter with direct force measurement on atrial fibrillation ablation. Pacing Clin Electrophysiol 2012; 35: 1312-1318, doi: 10.1111/j.1540-8159.2012.03503.x.
    » https://doi.org/10.1111/j.1540-8159.2012.03503.x
  • 17
    Sciarra L, Golia P, Natalizia A, de Ruvo E, Dottori S, Scara A, et al. Which is the best catheter to perform atrial fibrillation ablation? A comparison between standard ThermoCool, SmartTouch, and Surround Flow catheters. J Interv Card Electrophysiol 2014; 39: 193-200, doi: 10.1007/s10840-014-9874-2.
    » https://doi.org/10.1007/s10840-014-9874-2
  • 18
    Sigmund E, Puererfellner H, Derndorfer M, Kollias G, Winter S, Aichinger J, et al. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement - a case-matched comparison in 198 patients. Pacing Clin Electrophysiol 2015; 38: 201-208, doi: 10.1111/pace.12549.
    » https://doi.org/10.1111/pace.12549
  • 19
    Wakili R, Clauss S, Schmidt V, Ulbrich M, Hahnefeld A, Schussler F, et al. Impact of real-time contact force and impedance measurement in pulmonary vein isolation procedures for treatment of atrial fibrillation. Clin Res Cardiol 2014; 103: 97-106, doi: 10.1007/s00392-013-0625-7.
    » https://doi.org/10.1007/s00392-013-0625-7
  • 20
    Wutzler A, Huemer M, Parwani AS, Blaschke F, Haverkamp W, Boldt LH. Contact force mapping during catheter ablation for atrial fibrillation: procedural data and one-year follow-up. Arch Med Sci 2014; 10: 266-272, doi: 10.5114/aoms.2014.42578.
    » https://doi.org/10.5114/aoms.2014.42578
  • 21
    Kowalski M, Grimes MM, Perez FJ, Kenigsberg DN, Koneru J, Kasirajan V, et al. Histopathologic characterization of chronic radiofrequency ablation lesions for pulmonary vein isolation. J Am Coll Cardiol 2012; 59: 930-938, doi: 10.1016/j.jacc.2011.09.076.
    » https://doi.org/10.1016/j.jacc.2011.09.076
  • 22
    Shah D, Lambert H, Langenkamp A, Vanenkov Y, Leo G, Gentil-Baron P, et al. Catheter tip force required for mechanical perforation of porcine cardiac chambers. Europace 2011; 13: 277-283, doi: 10.1093/europace/euq403.
    » https://doi.org/10.1093/europace/euq403
  • 23
    Neuzil P, Reddy VY, Kautzner J, Petru J, Wichterle D, Shah D, et al. Electrical reconnection after pulmonary vein isolation is contingent on contact force during initial treatment: results from the EFFICAS I study. Circ Arrhythm Electrophysiol 2013; 6: 327-333, doi: 10.1161/CIRCEP.113.000374.
    » https://doi.org/10.1161/CIRCEP.113.000374
  • 24
    Reddy VY, Shah D, Kautzner J, Schmidt B, Saoudi N, Herrera C, et al. The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study. Heart Rhythm 2012; 9: 1789-1795, doi: 10.1016/j.hrthm.2012.07.016.
    » https://doi.org/10.1016/j.hrthm.2012.07.016
  • 25
    Okumura Y, Johnson SB, Bunch TJ, Henz BD, O'Brien CJ, Packer DL. A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. J Cardiovasc Electrophysiol 2008; 19: 632-640, doi: 10.1111/j.1540-8167.2008.01135.x.
    » https://doi.org/10.1111/j.1540-8167.2008.01135.x
  • 26
    Kumar S, Haqqani HM, Chan M, Lee J, Yudi M, Wong MC, et al. Predictive value of impedance changes for real-time contact force measurements during catheter ablation of atrial arrhythmias in humans. Heart Rhythm 2013; 10: 962-969, doi: 10.1016/j.hrthm.2013.03.022.
    » https://doi.org/10.1016/j.hrthm.2013.03.022
  • 27
    Yokoyama K, Nakagawa H, Shah DC, Lambert H, Leo G, Aeby N, et al. Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circ Arrhythm Electrophysiol 2008; 1: 354-362, doi: 10.1161/CIRCEP.108.803650.
    » https://doi.org/10.1161/CIRCEP.108.803650
  • 28
    Thiagalingam A, D'Avila A, Foley L, Guerrero JL, Lambert H, Leo G, et al. Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. J Cardiovasc Electrophysiol 2010; 21: 806-811, doi: 10.1111/j.1540-8167.2009.01693.x.
    » https://doi.org/10.1111/j.1540-8167.2009.01693.x

Publication Dates

  • Publication in this collection
    Mar 2016

History

  • Received
    2 Oct 2015
  • Accepted
    5 Nov 2015
Associação Brasileira de Divulgação Científica Av. Bandeirantes, 3900, 14049-900 Ribeirão Preto SP Brazil, Tel. / Fax: +55 16 3315-9120 - Ribeirão Preto - SP - Brazil
E-mail: bjournal@terra.com.br