Experiência brasileira com uso de marcapasso cardíaco artificial: resultados atuais obtidos pelo Registro Brasileiro de Marcapassos - RBM

 

Brazilian permanent cardiac pacemaker implantation experience: results obtained from the Brazilian Pacemaker Register's data

 

 

Roberto Costa; Maria Inês de Paula Leão

Médico Coordenador do Registro Brasileiro de Marcapassos - RBM

Endereço para correspondência

 

 


RESUMO

FUNDAMENTOS: Registro Brasileiro de Marcapassos (RBM) é uma base nacional de dados que tem por finalidade coletar e divulgar informações concernentes aos procedimentos cirúrgicos realizados em pacientes com estimulação cardíaca artificial permanente em todo o Brasil.
OBJETIVO: Apresentar os resultados obtidos no período de junho a dezembro de 1994.
CASUÍSTICA: No período de 1/6/94 a 31/12/94 foram recebidos 4696 formulários enviados por 130 hospitais e 287 médicos. Os procedimentos referiam-se a 3403 (72,5%) implantes iniciais a 1053 (22,4%) reoperações e 240 (5,1%) formulários exibiam este campo de cadastramento como dado não disponível.
RESULTADOS: Dos 3403 pacientes submetidos a implante inicial, 52,8% eram do sexo masculino e 73,7% eram da raça branca. Sintomas de hipofluxo cerebral justificaram o implante em 76,4% dos pacientes e a insuficiência cardíaca congestiva esteve presente em 85,0% dos casos, sendo que em 266 (7,9%) pacientes foi a causa principal para a operação. Os achados eletrocardiográficos predominantes foram: bloqueio atrioventricular total (57,7%); disfunção do nó sinusal (13,9%), bloqueio AV do 2º grau (13,8%); e flütter ou fibrilação atrial com baixa resposta ventricular (6,6%). A doença de Chagas foi a etiologia predominante (30,4%). Implante de marcapasso ventricular foi realizado em 83,6% dos pacientes e atrioventricular em 16,1%. Dos 1053 casos de reoperações informados, 50,4% ocorreram por problemas no gerador de pulsos. O tempo transcorrido entre o implante inicial e a cirurgia atual variou de 1 mês a 24 anos, com média de 7,1 anos. A susbstituição do gerador de pulsos foi informada em 917 pacientes (87,0% dos casos de reoperação); sendo a principal causa de troca o esgotamento por fim de vida em 65,7%. A substituição de eletrodo atrial foi relatada em 21 pacientes e de eletrodo ventricular em 203 pacientes.
CONCLUSÕES: A participação ativa de todos os membros do Departamento de Estimulação Cardíaca Artificial (DECA), com o apoio das empresas fornecedoras de marcapasso, e o respaldo do Ministério da Saúde permitiram que a implantação do RBM ocorresse da forma mais tranqüila possível. Hoje se dispõe de uma estatística nacional, volumosa e confiável, sobre a cirurgia de marcapasso no Brasil. As dificuldades enfrentadas estão relacionadas, principalmente, à estrutura individual de coleta de dados e a graus variados de familiaridade com o formulário, dificuldades estas que, com o tempo, tenderão a desaparecer.

Descritores: Marcapassos cardíacos, cirurgia, coleta de dados. Estimulação cardíaca artificial, marcapassos, coleta de dados. Arritmias cardíacas. Registro Brasileiro de Marcapassos - RBM.


ABSTRACT

BACKGROUND: Brazilian Pacemaker Registry (RBM) is a nationwide database to collect informations about all permanent pacemaker procedures performed in Brazil. It is a task force composed by Medical Society, Health Ministeryand Pacemaker Companies.
OBJECTIVE: To report the data obtained from June to December, 1994.
METHODS: From June 1 st to December 31,1994,4696 surgical procedures for permanent cardiac pacing were informed. These procedures were 3403 (72,5%) initial implantations, 1053 (22.4%) re-operations and in 240 cases this information was non-available.
RESULTS: From 3403 initial implantations informed, 52.8% were males and 73.7% Caucasians. Preoperative dizziness or syncopes were refered in 76.4% and congestive heart failure in 85.0% of the patients. EKG evaluation showed 3rd degree atrioventricular (AV) block in 57.7%, sick sinus syndrome in 13.9%, 2nd degree AV block in 13.8%, and high degree AV block and flutter or atrial fibrillation in 6.6% of patients. Chaga's disease was the prédominât ethiology (30.4%). Ventricular pacemakers were implanted in 83.6% of cases, atrioventricular in 16.1 % and atrial pacing in 0.3%. In the 1053 cases of re-operation, the interval between the initial implantation and the present procedure ranged from 1 month to 24 years (M=7.1 years). Pulse generator replacements were refered in 917 patients (87.0% of re-operation procedures), 65.7% of them at the end of life. Atrial lead replacement were performed in 21, and ventricular lead replacement in 203 patients.

Descriptors: Cardiac pacemakers, surgery, data collection . Cardiac stimulation, artificial, data collection. Cardiac arrithmias. Brazilian Pacemaker Registry - RBM.


 

 

Texto completo disponível apenas em PDF.

Full text available only in PDF format.

 

 

REFERÊNCIAS BIBLIOGRÁFICAS

1 COSTA, R. & LEÃO, M. I. P. - Implantação do Registro Brasileiro de Marcapassos. Rev. Bras. Marcapasso Arritmia, 7: 2-3, 1994.         [ Links ]

2 COSTA, R. & LEÃO, M. I. P. - Registro Brasileiro de Marcapassos. Rev. Bras. Marcapasso Arritmia, 6: 31-34, 1993.         [ Links ]

3 COSTA, R. & LEÃO, M. I. P. - Registro Brasileiro de Marcapassos: resultados preliminares. Rev. Bras. Marcapasso Arritmia, 7: 124-129, 1994.         [ Links ]

4 LEÃO, M. I. P.; COSTA, R.; LATINI, R. - Registro Brasileiro de Marcapassos: orientação para preenchimento do formulário. Rev. Bras. Marcapasso Arritmia, 7: 72-77, 1994.         [ Links ]

 

 

Endereço para correspondência:
Roberto Costa
Av. Brig. Luís Antônio, 4268
Cep [011] 01402-002. São Paulo, SP, Brasil
Tel.: [011] 887-8866. Fax: (011)885-3919

 

 

Trabalho realizado no Departamento de Estimulação Cardíaca Artificial - DECA, da Sociedade Brasileira de Cirurgia Cardiovascular. São Paulo, SP, Brasil.
Apresentado ao 22º Congresso Nacional de Cirurgia Cardíaca. Brasília, DF, 30 de março a 1º de abril, 1995.

Creative Commons License All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License