Immediate Outcomes of Aortic Valve Neocuspidization with Glutaraldehyde-treated Autologous Pericardium: a Multicenter Study

Abstract

Objective:

To determine the feasibility of aortic valve neocuspidization (AVNeo) with glutaraldehyde-treated autologous pericardium.

Methods:

One hundred and seventy (170) AVNeo (84 males/86 females) were performed from January 2017 through March 2019 in three centers. All the records were prospectively collected and retrospectively reviewed.

Results:

Most of the patients were older than 60 years and over 95% were operated for aortic stenosis. Preoperatively, pressure gradients were 69.9±21.3 mmHg for patients with aortic stenosis, and the surgical annular diameter was 21.0±2.0 mm for all patients. Effective orifice area (EOA) and indexed EOA (iEOA) averaged 0.7±0.3 cm2 and 0.4±0.2 cm2/m2 for patients with aortic stenosis before surgery, respectively. There was no conversion to prosthetic aortic valve replacement. Eight patients needed reoperation for bleeding, but no patient needed reoperation due to early infective endocarditis. There were five in-hospital deaths due to noncardiac cause. Compared to preoperative echocardiographic measurements, postoperative peak pressure gradient decreased significantly (-58.7±1.7 mmHg; P<0.001) and reached 11.2±5.6 mmHg, and mean pressure gradient also decreased significantly (-36.8±1.1 mmHg; P<0.001) and reached 6.0±3.5 mmHg. Accordingly, EOA and iEOA increased significantly 2.0 cm2 and 1.0 cm2/m2 (both P<0.001) to reach 2.7±0.6 cm2 and 1.4±0.3 cm2/m2 after surgery, respectively, with minimal significant aortic regurgitation (0.6% > mild).

Conclusion:

AVNeo is feasible and reproducible with good clinical results. Hemodynamically, AVNeo produces immediate postoperative low-pressure gradients, large EOA, and minimal regurgitation of the aortic valve. Further studies are necessary to evaluate mid- and long-term evolution.

Keywords:
Aortic Valve; Glutaral; Hospital Mortality; Feasibility Studies; Aortic Valve; Stenosis; Endocarditis; Bacterial; Echocardiography; Pericardium

Abbreviations, acronyms & symbols AR = Aortic regurgitation iEOA = Indexed effective orifice area AS = Aortic stenosis LAA = Left atrial appendage AVNeo = Aortic valve neocuspidization LVEF = Left ventricular ejection fraction AVR = Aortic valve replacement NYHA = New York Heart Association COPD = Chronic obstructive pulmonary disease PG = Pressure gradients EOA = Effective orifice area PHV = Prosthetic heart valves EuroSCORE = European System for Cardiac Operative Risk Evaluation PPM = Prosthesis-patient mismatch ICU = Intensive care unit SD = Standard deviation

INTRODUCTION

Aortic valve replacement (AVR) is the gold standard treatment for aortic valve diseases. However, most of prosthetic valves have a stent structure for suturing and fixing to the aortic valve annulus, which culminates in less mobility and decrease of the annulus size and increase in the postoperative pressure gradients. Furthermore, there are also cases that result in prosthesis-patient mismatch (PPM) both in surgical AVR[11 Sá MPBO, de Carvalho MMB, Sobral Filho DC, Cavalcanti LRP, Rayol SDC, Diniz RGS, et al. Surgical aortic valve replacement and patient-prosthesis mismatch: a meta-analysis of 108 182 patients. Eur J Cardiothorac Surg. 2019;56(1):44-54. doi:10.1093/ejcts/ezy466.
https://doi.org/10.1093/ejcts/ezy466...
,22 Rayol SDC, Sá MPBO, Cavalcanti LRP, Saragiotto FAS, Diniz RGS, Sá FBCAE, et al. Prosthesis-patient mismatch after surgical aortic valve replacement: neither uncommon nor harmless. Braz J Cardiovasc Surg. 2019;34(3):361-5. doi:10.21470/1678-9741-2019-0008.
https://doi.org/10.21470/1678-9741-2019-...
] and in transcatheter aortic valve implantation[33 Sá MPBO, Cavalcanti LRP, Sarargiotto FAS, Perazzo ÁM, Rayol SDC, Diniz RGS, et al. Impact of prosthesis-patient mismatch on 1-year outcomes after transcatheter aortic valve implantation: meta-analysis of 71,106 patients. Braz J Cardiovasc Surg. 2019;34(3):318-26. doi:10.21470/1678-9741-2019-0073.
https://doi.org/10.21470/1678-9741-2019-...
]. Additionally, in patients with a small aortic annulus, AVR may be difficult and aortic root enlargement is necessary to implant the largest possible prosthesis in order to avoid postoperative PPM[44 Sá MPBO, Carvalho MMB, Sobral Filho DC, Cavalcanti LRP, Diniz RGS, Rayol SC, et al. Impact of surgical aortic root enlargement on the outcomes of aortic valve replacement: a meta-analysis of 13 174 patients. Interact Cardiovasc Thorac Surg. 2019;29(1):74-82. doi:10.1093/icvts/ivy364.
https://doi.org/10.1093/icvts/ivy364...
].

It has been over 10 years since Ozaki et al.[55 Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Matsuyama T, et al. Aortic valve reconstruction using self-developed aortic valve plasty system in aortic valve disease. Interact Cardiovasc Thorac Surg. 2011;12(4):550-3. doi:10.1510/icvts.2010.253682.
https://doi.org/10.1510/icvts.2010.25368...
] started to perform the aortic valve neocuspidization (AVNeo) procedure, which is a surgical procedure consisting of the complete resection of the aortic cusps and the direct suture of the glutaraldehyde-treated autologous pericardium to the aortic annulus. This technique has been applied to various aortic valve pathologies with good mid-term results[55 Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Matsuyama T, et al. Aortic valve reconstruction using self-developed aortic valve plasty system in aortic valve disease. Interact Cardiovasc Thorac Surg. 2011;12(4):550-3. doi:10.1510/icvts.2010.253682.
https://doi.org/10.1510/icvts.2010.25368...

6 Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Takatoh M, et al. A total of 404 cases of aortic valve reconstruction with glutaraldehyde-treated autologous pericardium. J Thorac Cardiovasc Surg. 2014;147(1):301-6.

7 Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Takatoh M, et al. Aortic valve reconstruction using autologous pericardium for ages over 80 years. Asian Cardiovasc Thorac Ann. 2014;22(8):903-8. doi:10.1177/0218492314520748.
https://doi.org/10.1177/0218492314520748...

8 Ozaki S, Kawase I, Yamashita H, Uchida S, Takatoh M, Hagiwara S, et al. Aortic valve reconstruction using autologous pericardium for aortic stenosis. Circ J. 2015;79(7):1504-10. doi:10.1253/circj.CJ-14-1092.
https://doi.org/10.1253/circj.CJ-14-1092...
-99 Ozaki S, Kawase I, Yamashita H, Uchida S, Takatoh M, Kiyohara N. Midterm outcomes after aortic valve neocuspidization with glutaraldehyde-treated autologous pericardium. J Thorac Cardiovasc Surg. 2018;155(6):2379-87. doi:10.1016/j.jtcvs.2018.01.087.
https://doi.org/10.1016/j.jtcvs.2018.01....
].

In this study, we aimed to elucidate whether AVNeo is an acceptable and reproducible option by evaluating the immediate postoperative clinical and echocardiographic outcomes.

METHODS

Study Design

We carried out a prospective multicenter study in men and women with aortic valve diseases selected for AVNeo procedure. The study was approved by the local ethics committee of each participating institution. All patients who were willing to undergo AVNeo were evaluated for in-hospital clinical results. Detailed descriptions of the surgical techniques for AVNeo have been previously published[55 Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Matsuyama T, et al. Aortic valve reconstruction using self-developed aortic valve plasty system in aortic valve disease. Interact Cardiovasc Thorac Surg. 2011;12(4):550-3. doi:10.1510/icvts.2010.253682.
https://doi.org/10.1510/icvts.2010.25368...

6 Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Takatoh M, et al. A total of 404 cases of aortic valve reconstruction with glutaraldehyde-treated autologous pericardium. J Thorac Cardiovasc Surg. 2014;147(1):301-6.

7 Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Takatoh M, et al. Aortic valve reconstruction using autologous pericardium for ages over 80 years. Asian Cardiovasc Thorac Ann. 2014;22(8):903-8. doi:10.1177/0218492314520748.
https://doi.org/10.1177/0218492314520748...

8 Ozaki S, Kawase I, Yamashita H, Uchida S, Takatoh M, Hagiwara S, et al. Aortic valve reconstruction using autologous pericardium for aortic stenosis. Circ J. 2015;79(7):1504-10. doi:10.1253/circj.CJ-14-1092.
https://doi.org/10.1253/circj.CJ-14-1092...

9 Ozaki S, Kawase I, Yamashita H, Uchida S, Takatoh M, Kiyohara N. Midterm outcomes after aortic valve neocuspidization with glutaraldehyde-treated autologous pericardium. J Thorac Cardiovasc Surg. 2018;155(6):2379-87. doi:10.1016/j.jtcvs.2018.01.087.
https://doi.org/10.1016/j.jtcvs.2018.01....

10 Sá MPBO, Perazzo ÁM, Zhigalov K, Komarov R, Kadyraliev B, Enginoev S, et al. Aortic valve neocuspidization with glutaraldehyde-treated autologous pericardium (Ozaki procedure) - a promising surgical technique. Braz J Cardiovasc Surg. 2019;34(5):610-4. doi:10.21470/1678-9741-2019-0304.
https://doi.org/10.21470/1678-9741-2019-...
-1111 Komarov R, Chernov I, Enginoev S, Sá MPBO, Tarasov D. The Russian conduit - combining Bentall and Ozaki procedures for concomitant ascending aorta replacement and aortic valve neocuspidization. Braz J Cardiovasc Surg. 2019;34(5):618-23. doi:10.21470/1678-9741-2019-0329.
https://doi.org/10.21470/1678-9741-2019-...
]. Figures 1A, 1B, and 1C depict the main steps of the procedure.

Fig. 1A
Trimming of treated autologous pericardium.

Fig. 1B – Suture of the neocusp to the aortic annulus.

Fig. 1C – Aortic valve neocuspidization, a final view.


Echocardiographic Measurement

All patients underwent an echocardiographic evaluation of the aortic valve before and after the procedure. Peak and mean gradient were calculated by the modified Bernouilli equation. Effective orifice area (EOA) (cm2) was calculated with the continuity equation, with the use of stroke volume estimated by the left ventricular outflow tract diameter and velocity time integral within the left ventricular outflow tract. EOA was indexed (iEOA) by the body surface area (m2). Aortic regurgitation (AR) was graded using an integrative approach including structural, Doppler, quantitative, and qualitative parameters as recommended by the American Society of Echocardiography guidelines[1212 Zoghbi WA, Adams D, Bonow RO, Enriquez-Sarano M, Foster E, Grayburn PA, et al. Recommendations for noninvasive evaluation of native valvular regurgitation: a report from the American society of echocardiography developed in collaboration with the society for cardiovascular magnetic resonance. J Am Soc Echocardiogr. 2017;30(4):303-71. doi:10.1016/j.echo.2017.01.007.
https://doi.org/10.1016/j.echo.2017.01.0...
].

Endpoints

The primary endpoints were operative mortality and hemodynamic performance of AVNeo through echocardiographic evaluation, which was done during the hospital stay prior to discharge. Secondary endpoints were perioperative adverse events. All the patients were followed during their hospital stay.

Statistical Analysis

Continuous data are presented as means ± standard deviation (SD) after confirming normal distribution with Shapiro-Wilk test. Categorical variables are expressed as numbers and percentages. Between-group differences were evaluated using the independent Student’s t-test for normally distributed data and the Mann-Whitney U test for non-normally distributed data. Post vs. preprocedure variation was tested with Levene’s test and F-test. All analyses were performed using the IBM SPSS Statistics software (version 22.0, SPSS Inc., Chicago, Illinois, United States of America). P-values < 0.05 were considered to indicate a statistically significant difference.

RESULTS

Population

Among the patients, as shown in Table 1, there were 84 males and 86 females. Mean age was 64 (range, 23-86) years. Regarding the age distribution, most of the patients were older than 60 years.

Table 1
Patients' baseline characteristics and risk factors.

One hundred sixty-two (95%) patients had aortic stenosis (AS), three patients had AR, and five had early infective endocarditis. Fifty-two (31%) patients had bicuspid aortic valve. Preoperative echocardiography showed an average peak pressure gradient through the aortic valve of 69.9±21.3 mmHg for patients with AS and a surgical annular diameter of 21.0±2.0 mm for all patients. EOA and iEOA averaged 0.7±0.3 cm2 and 0.4±0.2 cm2/m2 before surgery, respectively.

Perioperative Clinical Events

A total of 94 (55%) patients had isolated AVNeo during the study period and the other 76 (45%) patients underwent concomitant procedures (Table 2). There was no conversion to classical prosthetic AVR. Mean cardiopulmonary bypass time was 112.2±38.9 min and aortic cross-clamp time was 86.2±22.7 min (Table 3). Eight patients needed reoperation for bleeding and two pacemaker implantations, but no patient needed reoperation due to early infective endocarditis. No thromboembolic events were recorded. There were five in-hospital deaths due to noncardiac cause.

Table 2
Operative data.
Table 3
Postoperative events.

Postoperative Echocardiography

Average peak and mean pressure gradients were 11.2±5.6 mmHg and 6.0±3.5 mmHg (mean±SD), respectively, after surgery. Thus, both decreased significantly (difference in means±standard error; peak gradient: -58.7±1.7 mmHg, P<0.001; and mean gradient: -36.8±1.1 mmHg, P<0.001) in comparison to preoperative measures (Figure 2A).

Fig. 2A
Comparison between pre- and postoperative echocardiographic data regarding maximum and mean pressure gradients (PG) ± standard deviation. AVNeo=aortic valve neocuspidization.

Fig. 2B – Comparison between pre- and postoperative echocardiographic data regarding effective orifice area (EOA) and indexed effective orifice area (iEOA). AVNeo=aortic valve neocuspidization


EOA and iEOA averaged 2.7±0.6 cm2 and 1.4±0.3 cm2/m2 after surgery, respectively. They significantly increased 2.0±0.1 cm2 and 1.0±0.1 cm2/m2, respectively, in comparison to preoperative measures (Figure 2B). Interestingly, we observed no PPM after the procedure (i.e., all iEOA > 0.85 cm2/m2). After AVNeo procedure, 38 (22.3%) patients had a mild AR and one (0.6%) had moderate AR. The postoperative echocardiographic views are presented in Figures 3A and 3B.

Fig. 3
Postoperative echocardiographic views in diastole (A) and systole (B).

DISCUSSION

The main finding of our study was that AVNeo is a feasible procedure with: 1) good perioperative outcome (2.9% operative death observed in a population with 2.6% expected operative death calculated by the European System for Cardiac Operative Risk Evaluation - EuroSCORE II), 2) low perioperative complications, and 3) excellent postoperative hemodynamic results (mean gradient post-AVNeo 6 mmHg, only 0.6% moderate AR).

Whereas AVR with prosthetic heart valves (PHV) has been considered the best option for treatment of aortic valve diseases, many unanswered questions surround intraoperative sizing and labeling of these prostheses, making optimal intraoperative selection challenging. According to a document[1313 Durko AP, Head SJ, Pibarot P, Atluri P, Bapat V, Cameron DE, et al. Characteristics of surgical prosthetic heart valves and problems around labeling: a document from the European association for cardio-thoracic surgery (EACTS)-the society of thoracic surgeons (STS)-American association for thoracic surgery (AATS) valve labelling task force. J Thorac Cardiovasc Surg. 2019;158(4):1041-54. doi:10.1016/j.jtcvs.2019.04.001.
https://doi.org/10.1016/j.jtcvs.2019.04....
] from the European Association for Cardio-Thoracic Surgery - EACTS, the Society of Thoracic Surgeons - STS, and the American Association for Thoracic Surgery - AATS Valve Labelling Task Force, these questions include: non-uniform or incomplete reporting of PHV materials and physical dimensions in the instructions for use; unclear definition of labeled valve size and inconsistencies between size dimensions and labeled valve size; non-uniform marking of PHV support structures; lack of robust information on in vivo hemodynamic performance and no information available regarding hemodynamic performance on package labels; lack of uniform tools backed by solid evidence to prevent PPM; and lack of good-quality, robust clinical data on PHV thrombogenicity.

This situation has received many calls for action, but no solution has been achieved so far. In this scenario, AVNeo arises as an attractive option owing to its low cost, universal indications without any limit in annulus size nor any need for anticoagulation, and its potentially excellent hemodynamics.

The most important findings of our study were, in addition to low operative mortality, the statistically significant decreases in pressure gradients and increase in EOA and iEOA after AVNeo. We would like to highlight the fact that these findings were consistent regardless of the annulus size. We also found no patient in the postoperative period with values of iEOA < 0.85cm2/m2, which means no postoperative PPM. These observations might underlie the maintenance of the characteristics of the aortic annulus due to the absence of stent structures.

Iida et al.[1414 Iida Y, Akiyama S, Shimura K, Fujii S, Hashimoto C, Mizuuchi S, et al. Comparison of aortic annulus dimensions after aortic valve neocuspidization with those of normal aortic valve using transthoracic echocardiography. Eur J Cardiothorac Surg. 2018;54(6):1081-4. doi:10.1093/ejcts/ezy190.
https://doi.org/10.1093/ejcts/ezy190...
] performed AVNeo for various aortic valve pathologies in 147 patients from December 2010 to October 2017. Of these patients, the aortic annulus dimensions were measured in 25 patients who underwent AVNeo for aortic valve disease as follow-up examination and compared with those measured in 15 patients who had normal aortic valves. No significant differences in the aortic annulus dimensions were observed between the patients who had undergone AVNeo and those who had normal aortic valves. The authors concluded that the movement of the aortic annulus after AVNeo is comparable with that of the aortic annulus of a normal aortic valve; thus, AVNeo can be regarded as a more physiological operation in that it maintains the characteristics of the aortic valve similar to those of a normal aortic valve (which in turn does not happen when patients undergo AVR).

Similarly, Yamamoto et al.[1515 Yamamoto Y, Iino K, Shintani Y, Kato H, Kimura K, Watanabe G, et al. Comparison of aortic annulus dimension after aortic valve neocuspidization with valve replacement and normal valve. Semin Thorac Cardiovasc Surg. 2017;29(2):143-9. doi:10.1053/j.semtcvs.2016.11.002.
https://doi.org/10.1053/j.semtcvs.2016.1...
] measured aortic annular dimensions using electrocardiography-gated multidetector computed tomography in 23 patients. The sample included eight patients who had undergone AVNeo, 10 patients with normal aortic valves, and five patients who had undergone AVR. Postoperative peak pressure gradients for the AVNeo and AVR groups were compared. No statistically significant differences in annulus variation were observed between patients who had undergone AVNeo and those with normal aortic valves. Annular area was larger during systole than during diastole in both groups. Postoperative peak pressure gradients were significantly lower in the AVNeo group than in the AVR group. The results of this study demonstrated that aortic annular dimensions after AVNeo are similar to the dimensions of normal aortic valves. Lower postoperative pressure gradients might underlie the observed differences.

In another study, Iida et al.[1616 Iida Y, Fujii S, Akiyama S, Sawa S. Early and mid-term results of isolated aortic valve neocuspidization in patients with aortic stenosis. Gen Thorac Cardiovasc Surg. 2018;66(11):648-52. doi:10.1007/s11748-018-0976-0.
https://doi.org/10.1007/s11748-018-0976-...
] performed AVNeo for AS in 57 patients from December 2010 to June 2017. Their mean age was 77.5±8.8 years. Preoperative echocardiography revealed an average peak pressure gradient of 89±32.9 mmHg that decreased to 22±10.7 mmHg one week after the procedure and 19.2±9.7 mmHg 20 months after the procedure. There were no conversions to AVR. There were two noncardiac-related deaths. Two patients underwent reoperation owing to infective endocarditis and recurrent AR. The mean follow-up period was 30.4±20.8 months. The freedom from reoperation rates was 98.1 and 95.3% at 12 and 81 months of follow-up, respectively.

Mourad et al.[1717 Mourad F, Shehada SE, Lubarski J, Serrano M, Demircioglu E, Wendt D, et al. Aortic valve construction using pericardial tissue: short-term single-centre outcomes. Interact Cardiovasc Thorac Surg. 2019;28(2):183-90. doi:10.1093/icvts/ivy230.
https://doi.org/10.1093/icvts/ivy230...
] carried out a prospective single-center study including 52 consecutive patients who underwent AVNeo between September 2015 and March 2017 using autologous pericardium in 16 patients. Most patients presented with AS or endocarditis. The mean age was 60±14 years. Early outcomes included one stroke, two patients needing short-term dialysis, and one death. During follow-up (mean 11.2±4.8 months), trace AR was observed in four patients; the mean pressure gradient was 6.8±2.9 mmHg. Three patients died later (of noncardiac reasons) and five patients needed reoperation due to endocarditis.

Krane et al.[1818 Krane M, Boehm J, Prinzing A, Lange R. Excellent hemodynamic results in patients undergoing the Ozaki procedure for aortic valve reconstruction within the first year. Thorac Cardiovasc Surg. 2019;67(S01): S1-S100. doi:10.1055/s-0039-1678893.
https://doi.org/10.1055/s-0039-1678893...
] operated on 77 patients undergoing AVNeo following the Ozaki procedure between October 2016 and August 2018. Their mean age was 54.9±16.5 years; AS was present in 84.4% and insufficiency in 15.6% of the patients. At 1.76-year follow-up, freedom from reoperation was 97.4%. Two patients (2.6%) presented with moderate to severe aortic insufficiency after the procedure. Both received a prosthetic AVR within the same hospital stay. At discharge, mean pressure gradient was 9.3±4.2 mmHg, which decreased to a mean aortic gradient of 1.6±3.4 mmHg at six to 12 months. The authors concluded that AVNeo following the Ozaki procedure revealed excellent early hemodynamic results in terms of EOA, pressure gradients, and PPM. As we can see, surgeons from all over the world have achieved good initial results by reproducing this surgical procedure in their own populations, which makes AVNeo seem even more promising.

Study Limitations

The major limitation of the current study is the lack of mid- and long-term follow-ups. We will continue to monitor the results for a longer follow-up period with a larger number of patients.

CONCLUSION

The AVNeo procedure is feasible and reproducible and it has shown good immediate results. Our findings show that aortic valve tricuspidization with glutaraldehyde-treated autologous pericardium produces immediate postoperative low-pressure gradients, large EOA, and minimal regurgitation of the aortic valve. Further studies are required to assess mid- and long-term evolution of the neo-aortic valve.

Authors' roles & responsibilities VA Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published IC Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published RK Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published YS Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published BK Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published SE Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published MT Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published AI Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published AB Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published FA Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published MAC Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published PP Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published MPBOS Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published AW Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published KZ Substantial contributions to the conception or design of the work; or the acquisition, analysis, or interpretation of data for the work; drafting the work or revising it critically for important intellectual content; agreement to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved; final approval of the version to be published

REFERENCES

  • 1
    Sá MPBO, de Carvalho MMB, Sobral Filho DC, Cavalcanti LRP, Rayol SDC, Diniz RGS, et al. Surgical aortic valve replacement and patient-prosthesis mismatch: a meta-analysis of 108 182 patients. Eur J Cardiothorac Surg. 2019;56(1):44-54. doi:10.1093/ejcts/ezy466.
    » https://doi.org/10.1093/ejcts/ezy466
  • 2
    Rayol SDC, Sá MPBO, Cavalcanti LRP, Saragiotto FAS, Diniz RGS, Sá FBCAE, et al. Prosthesis-patient mismatch after surgical aortic valve replacement: neither uncommon nor harmless. Braz J Cardiovasc Surg. 2019;34(3):361-5. doi:10.21470/1678-9741-2019-0008.
    » https://doi.org/10.21470/1678-9741-2019-0008
  • 3
    Sá MPBO, Cavalcanti LRP, Sarargiotto FAS, Perazzo ÁM, Rayol SDC, Diniz RGS, et al. Impact of prosthesis-patient mismatch on 1-year outcomes after transcatheter aortic valve implantation: meta-analysis of 71,106 patients. Braz J Cardiovasc Surg. 2019;34(3):318-26. doi:10.21470/1678-9741-2019-0073.
    » https://doi.org/10.21470/1678-9741-2019-0073
  • 4
    Sá MPBO, Carvalho MMB, Sobral Filho DC, Cavalcanti LRP, Diniz RGS, Rayol SC, et al. Impact of surgical aortic root enlargement on the outcomes of aortic valve replacement: a meta-analysis of 13 174 patients. Interact Cardiovasc Thorac Surg. 2019;29(1):74-82. doi:10.1093/icvts/ivy364.
    » https://doi.org/10.1093/icvts/ivy364
  • 5
    Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Matsuyama T, et al. Aortic valve reconstruction using self-developed aortic valve plasty system in aortic valve disease. Interact Cardiovasc Thorac Surg. 2011;12(4):550-3. doi:10.1510/icvts.2010.253682.
    » https://doi.org/10.1510/icvts.2010.253682
  • 6
    Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Takatoh M, et al. A total of 404 cases of aortic valve reconstruction with glutaraldehyde-treated autologous pericardium. J Thorac Cardiovasc Surg. 2014;147(1):301-6.
  • 7
    Ozaki S, Kawase I, Yamashita H, Uchida S, Nozawa Y, Takatoh M, et al. Aortic valve reconstruction using autologous pericardium for ages over 80 years. Asian Cardiovasc Thorac Ann. 2014;22(8):903-8. doi:10.1177/0218492314520748.
    » https://doi.org/10.1177/0218492314520748
  • 8
    Ozaki S, Kawase I, Yamashita H, Uchida S, Takatoh M, Hagiwara S, et al. Aortic valve reconstruction using autologous pericardium for aortic stenosis. Circ J. 2015;79(7):1504-10. doi:10.1253/circj.CJ-14-1092.
    » https://doi.org/10.1253/circj.CJ-14-1092
  • 9
    Ozaki S, Kawase I, Yamashita H, Uchida S, Takatoh M, Kiyohara N. Midterm outcomes after aortic valve neocuspidization with glutaraldehyde-treated autologous pericardium. J Thorac Cardiovasc Surg. 2018;155(6):2379-87. doi:10.1016/j.jtcvs.2018.01.087.
    » https://doi.org/10.1016/j.jtcvs.2018.01.087
  • 10
    Sá MPBO, Perazzo ÁM, Zhigalov K, Komarov R, Kadyraliev B, Enginoev S, et al. Aortic valve neocuspidization with glutaraldehyde-treated autologous pericardium (Ozaki procedure) - a promising surgical technique. Braz J Cardiovasc Surg. 2019;34(5):610-4. doi:10.21470/1678-9741-2019-0304.
    » https://doi.org/10.21470/1678-9741-2019-0304
  • 11
    Komarov R, Chernov I, Enginoev S, Sá MPBO, Tarasov D. The Russian conduit - combining Bentall and Ozaki procedures for concomitant ascending aorta replacement and aortic valve neocuspidization. Braz J Cardiovasc Surg. 2019;34(5):618-23. doi:10.21470/1678-9741-2019-0329.
    » https://doi.org/10.21470/1678-9741-2019-0329
  • 12
    Zoghbi WA, Adams D, Bonow RO, Enriquez-Sarano M, Foster E, Grayburn PA, et al. Recommendations for noninvasive evaluation of native valvular regurgitation: a report from the American society of echocardiography developed in collaboration with the society for cardiovascular magnetic resonance. J Am Soc Echocardiogr. 2017;30(4):303-71. doi:10.1016/j.echo.2017.01.007.
    » https://doi.org/10.1016/j.echo.2017.01.007
  • 13
    Durko AP, Head SJ, Pibarot P, Atluri P, Bapat V, Cameron DE, et al. Characteristics of surgical prosthetic heart valves and problems around labeling: a document from the European association for cardio-thoracic surgery (EACTS)-the society of thoracic surgeons (STS)-American association for thoracic surgery (AATS) valve labelling task force. J Thorac Cardiovasc Surg. 2019;158(4):1041-54. doi:10.1016/j.jtcvs.2019.04.001.
    » https://doi.org/10.1016/j.jtcvs.2019.04.001
  • 14
    Iida Y, Akiyama S, Shimura K, Fujii S, Hashimoto C, Mizuuchi S, et al. Comparison of aortic annulus dimensions after aortic valve neocuspidization with those of normal aortic valve using transthoracic echocardiography. Eur J Cardiothorac Surg. 2018;54(6):1081-4. doi:10.1093/ejcts/ezy190.
    » https://doi.org/10.1093/ejcts/ezy190
  • 15
    Yamamoto Y, Iino K, Shintani Y, Kato H, Kimura K, Watanabe G, et al. Comparison of aortic annulus dimension after aortic valve neocuspidization with valve replacement and normal valve. Semin Thorac Cardiovasc Surg. 2017;29(2):143-9. doi:10.1053/j.semtcvs.2016.11.002.
    » https://doi.org/10.1053/j.semtcvs.2016.11.002
  • 16
    Iida Y, Fujii S, Akiyama S, Sawa S. Early and mid-term results of isolated aortic valve neocuspidization in patients with aortic stenosis. Gen Thorac Cardiovasc Surg. 2018;66(11):648-52. doi:10.1007/s11748-018-0976-0.
    » https://doi.org/10.1007/s11748-018-0976-0
  • 17
    Mourad F, Shehada SE, Lubarski J, Serrano M, Demircioglu E, Wendt D, et al. Aortic valve construction using pericardial tissue: short-term single-centre outcomes. Interact Cardiovasc Thorac Surg. 2019;28(2):183-90. doi:10.1093/icvts/ivy230.
    » https://doi.org/10.1093/icvts/ivy230
  • 18
    Krane M, Boehm J, Prinzing A, Lange R. Excellent hemodynamic results in patients undergoing the Ozaki procedure for aortic valve reconstruction within the first year. Thorac Cardiovasc Surg. 2019;67(S01): S1-S100. doi:10.1055/s-0039-1678893.