Development of a new experimental model of intramuscular electrical stimulation of the diaphragm in rabbits

Desenvolvimento de um novo modelo experimental de estimulação elétrica intramuscular do diafragma em coelhos

Abstracts

PURPOSE: To develop an animal model of diaphragmatic electrical stimulation able to generate an appropriate ventilatory support through the direct implantation of electrodes in the diaphragm (electroventilation). METHODS: Six New Zealand female rabbits (2-3 kg) were placed on mechanical ventilation. Then, a laparotomy was performed in order to identify the motor points in each hemidiaphragm, followed by the implantation of the electrodes for diaphragmatic stimulation. We tested two types of electrodes according to the conduction of electrical stimulation: unipolar and bipolar. The electrodes were placed on different occasions in the same animals and tested with current intensities of 20, 26 and 32 mA. Each current intensity was repeated three times for 10 respiratory cycles with 1 minute interval between each cycle, and 5 minutes for new current intensity. We recorded the relationship between current intensity and inspiratory volume. RESULTS: The electrodes adequately stimulate the diaphragm and obtain inspired volumes using different intensity currents. The bipolar electrode generated inspiratory volumes as high as 4.5 times of baseline while the unipolar electrode reached up to 3.5 times of baseline. CONCLUSION: This model has proved to be effective for studying the performance of the diaphragm under different electrical stimulations using different set of electrodes.

Electric Stimulation; Diaphragm; Pulmonary Ventilation; Rabbits


OBJETIVO: Desenvolver um modelo animal de estimulação elétrica diafragmática capaz de gerar um suporte ventilatório adequado através da implantação direta de eletrodos no diafragma (eletroventilação). MÉTODOS: Seis coelhas da raça Nova Zelândia (2-3 kg) foram colocadas em ventilação mecânica. Através de uma laparotomia foram identificados os pontos motores em ambas as cúpulas diafragmáticas, seguido pela implantação dos eletrodos para estimulação diafragmática. Foram testados dois tipos de eletrodos de acordo com a condução do estímulo elétrico: unipolar e bipolar. Os eletrodos foram posicionados em diferentes ocasiões nos mesmos animais e testados com intensidades de corrente de 20, 26 e 32 mA, sendo cada uma delas repetidas três vezes durante 10 ciclos respiratórios com intervalo de 1 minuto entre cada ciclo e repouso de 5 minutos para nova intensidade de corrente. Foram registradas a relação entre a intensidade de corrente e o volume de ar inspirado. RESULTADOS: Os eletrodos estimularam adequadamente o diafragma com um volume de ar inspirado até 4,5 vezes maior no modo bipolar e de 3,5 vezes no modo unipolar em relação à linha de base. CONCLUSÃO: Este modelo demonstrou ser efetivo para o estudo do desempenho do diafragma sob diferentes estimulações elétricas e diferentes tipos de eletrodos.

Estimulação Elétrica; Diafragma; Ventilação Pulmonar; Coelhos


3 – ORIGINAL ARTICLE

MODELS, BIOLOGICAL

Development of a new experimental model of intramuscular electrical stimulation of the diaphragm in rabbits1 1 Research performed at Animal Experimentation Unit Research Center, airway and lung laboratory, Clinics Hospital of Porto Alegre (HCPA), RS, Brazil.

Desenvolvimento de um novo modelo experimental de estimulação elétrica intramuscular do diafragma em coelhos

Rodrigo Guellner GhediniI; Ane MargaritesII; Elaine Aparecida FelixIII; Rogério Gastal XavierIV; Cristiano Feijó AndradeV

IPhysiotherapist, Porto Alegre Santa Casa, Researcher at airway and lung laboratory, HCPA, Porto Alegre-RS, Brazil

IIPhysiotherapist, Porto Alegre Santa Casa, RS, Brazil

IIIPhD, Associate Professor of Anesthesiology, Department of Surgery, Federal University of Rio Grande do Sul (UFRGS), Porto Alegre-RS, Brazil

IVPhD, Associate Professor of Pneumology, Department of Internal Medicine, UFRGS, Porto Alegre-RS, Brazil

VPhD, Thoracic Surgeon, HCPA, Santo Antônio Children's Hospital, Porto Alegre-RS, Brazil

Correspondence Correspondence: Cristiano Feijó Andrade Hospital de Clínicas de Porto Alegre Serviço de Cirurgia Torácica Ramiro Barcelos, 2.350 90035-903 Porto Alegre – RS Brasil Phone: (55 51)3359-8684 Fax: (55 51)3359-8884 cristianofa@cirurgiatoracica.net

ABSTRACT

PURPOSE: To develop an animal model of diaphragmatic electrical stimulation able to generate an appropriate ventilatory support through the direct implantation of electrodes in the diaphragm (electroventilation).

METHODS: Six New Zealand female rabbits (2-3 kg) were placed on mechanical ventilation. Then, a laparotomy was performed in order to identify the motor points in each hemidiaphragm, followed by the implantation of the electrodes for diaphragmatic stimulation. We tested two types of electrodes according to the conduction of electrical stimulation: unipolar and bipolar. The electrodes were placed on different occasions in the same animals and tested with current intensities of 20, 26 and 32 mA. Each current intensity was repeated three times for 10 respiratory cycles with 1 minute interval between each cycle, and 5 minutes for new current intensity. We recorded the relationship between current intensity and inspiratory volume.

RESULTS: The electrodes adequately stimulate the diaphragm and obtain inspired volumes using different intensity currents. The bipolar electrode generated inspiratory volumes as high as 4.5 times of baseline while the unipolar electrode reached up to 3.5 times of baseline.

CONCLUSION: This model has proved to be effective for studying the performance of the diaphragm under different electrical stimulations using different set of electrodes.

Key words: Electric Stimulation. Diaphragm. Pulmonary Ventilation. Rabbits.

RESUMO

OBJETIVO: Desenvolver um modelo animal de estimulação elétrica diafragmática capaz de gerar um suporte ventilatório adequado através da implantação direta de eletrodos no diafragma (eletroventilação).

MÉTODOS: Seis coelhas da raça Nova Zelândia (2–3 kg) foram colocadas em ventilação mecânica. Através de uma laparotomia foram identificados os pontos motores em ambas as cúpulas diafragmáticas, seguido pela implantação dos eletrodos para estimulação diafragmática. Foram testados dois tipos de eletrodos de acordo com a condução do estímulo elétrico: unipolar e bipolar. Os eletrodos foram posicionados em diferentes ocasiões nos mesmos animais e testados com intensidades de corrente de 20, 26 e 32 mA, sendo cada uma delas repetidas três vezes durante 10 ciclos respiratórios com intervalo de 1 minuto entre cada ciclo e repouso de 5 minutos para nova intensidade de corrente. Foram registradas a relação entre a intensidade de corrente e o volume de ar inspirado.

RESULTADOS: Os eletrodos estimularam adequadamente o diafragma com um volume de ar inspirado até 4,5 vezes maior no modo bipolar e de 3,5 vezes no modo unipolar em relação à linha de base.

CONCLUSÃO: Este modelo demonstrou ser efetivo para o estudo do desempenho do diafragma sob diferentes estimulações elétricas e diferentes tipos de eletrodos.

Descritores: Estimulação Elétrica. Diafragma. Ventilação Pulmonar. Coelhos.

Introduction

Trauma is the leading cause of death worldwide among young people. In Brazil, it is believed that there are 130,000 deaths annually due to trauma and about 450,000 patients with severe sequelae1-2. Many trauma patients have different types of sequelae such as damage of the central nervous system or spinal cord injuries. Patients with spinal cord injuries due to high cervical trauma (C1-C5) have tetraplegia and impaired diaphragmatic function, losing their ability to breathe spontaneously. Other neuromuscular diseases such as Amyotrophic Lateral Sclerosis also lead to a progressive impairment of diaphragmatic function resulting in poor quality of life and the need for permanent ventilatory support. These incapacitated patients usually require hospitalization in specialized centers or in homes adequately prepared for this purpose3. The possibility of electrical activation of the diaphragm with electrodes placed on this muscle (electroventilation) is an alternative for individuals dependent on mechanical ventilation and without any expectation of weaning. Therefore, there is a general interest in studies to develop new methods and devices for diaphragmatic pacing in order to provide better quality of live for such patients.

In this study, we demonstrate the development of a new animal model by electrical stimulation of the diaphragm through electrodes implanted directly into this respiratory muscle.

Methods

Six New Zealand female rabbits were used, weighing between 2 and 3kg. All animals were treated according to the Ethical Code for Animal Experimentation from the World Health Organization16. The study was approved by the Committee on Ethics in Research of the Clinics Hospital of Porto Alegre under number 07-133. The animals received standard intramuscularly premedication with ketamine hydrochloride (20 mg/kg), midazolam (0.2 mg/kg) and atropine (0.025 mg/kg), followed by trichotomy and peripheral venous access. Anesthesia was induced with propophol (4 mg/kg), to be maintained with an intravenously continuous infusion of propophol (11 mg/kg/h) and remiphentanyl (0.3 ml/kg/min). The animals were intubated orally and placed on mechanical ventilation (Nikkei–Takaoka, Sao Paulo, Brazil). For flow signal and tracheal pressure acquisition, two single sensors were connected at the end of the endotracheal tube (Hans Rudolph - Kansas City - USA) and attached to a pressure system acquisition, MVR 16X (Universidad de la Republica, Montevideo, Uruguay).

After hemodynamic stabilization, the animal inspiratory volume was measured at baseline. A laparotomy was performed for visualization of the diaphragm (Figure 1A), followed by identification of motor points (Figure 1B). The electrodes were then fixed to the motor points (Figure 1C), and they were connected to a generator of electric current (Dualpex 961, Phrenics, Quark Medical, Sao Paulo, Brazil). They produced a depolarized current with the following characteristics: rectangular waveform, alternating symmetrical pulse, frequency of 25HZ and burst width of 0.07 milliseconds. The abdomen wall was then sutured leaving space only to the conducting wires (Figure 1D). In order to deliver an induced apnea, the minute volume was increased gradually to reach a CO2 blood concentration around 17 mmHg.



We evaluated the responses with two different sets of unipolar and bipolar electrodes. One single channel electrode (unipolar) was sutured on the muscular portion of each hemidiaphragm, promoting simultaneous contraction of both hemidiaphragms. The bipolar set consisted of one pair of electrodes implanted in the same hemidiaphragm which could generate independent contractions. To evaluate the performance from the unipolar and bipolar sets, we compared the relationship between different current intensities and the generated inspired volume respectively. The two set of electrodes were positioned in the same animal at different periods during the same experiment. The current intensities used were 20, 26, and 32 mA to be sustained for 10 respiratory cycles. Each different intensity records were made in three separate periods with one minute interval. The baseline inspiratory volume was assessed with the animal in supine position.

Results

In four animals we were unable to record the inspiratory volume due to unilateral or bilateral pneumothorax and/or fluid accumulation in the abdominal cavity which interfered in transmission of the electrical stimulation. In two animals, the unipolar (one channel) and bipolar (two channel) electrodes were able to adequately stimulate the diaphragm and obtain inspired volumes using current intensities of 20, 26 and 32 mA. The bipolar electrode generated inspiratory volumes as high as 4.5 times of the baseline while the unipolar electrode reached up to 3.5 times of baseline. Since the volumes obtained were much higher than the baseline, we then added to the protocol the current intensities of 10 and 16 mA seeking volumes closer to the physiological pattern. The results from two animals are shown in the Figure 2.


Discussion

The use of electrical currents to produce artificial ventilation has been described since the nineteenth century through the generation of action potentials either on the nerve and/or on the diaphragm or any other ventilatory muscles throughout different techniques4-7. Several methods comparing stimulation through the phrenic nerve and ventilatory muscles have been tested, showing similar results in the generation of transdiaphragmatic pressure and speed nerve latency7.

The first studies of electroventilation were performed by stimulating the phrenic nerve in the neck or its path into the chest, in order to trigger action potentials8. This technique has proven to be clinically effective, but due to the extensive mobility of the cervical region, there were significant limitations and risks especially on the phrenic nerve path, increasing the risk of mechanical nerve damage. The implantation of electrodes directly into the diaphragm muscle seems to be a feasible and reproducible technique for animal studies and clinical trials so far undertaken. Peterson et al.11 studying the positioning of monopolar electrodes on diaphragm of dogs observed that was possible the activation of the diaphragmatic muscle using intramuscular electrodes. In this study, they obtained lung volumes up 167% from baseline, without inducing any muscle fatigue11. In another study, this same group could demonstrate the safety and reliability of intramuscular diaphragm stimulation through monopolar electrodes. Using five dogs which received electrical stimulation for 61-183 days17, the authors did not observe any muscle tissue damage as a result of the presence of the electrode, and any negative side effects as well as on the heart or the lung function. Immunohistochemistry studies suggested transformation of the diaphragm muscle fibers to type I, which have better capacity to resist fatigue, leading to the conclusion that this technique would present minimal risk of phrenic injury in human patients13.

In one of the first electroventilation clinical descriptions, one 35 years-old male victim of high cervical trauma making dependable of permanent mechanical ventilation was submitted to a laparotomy for intramuscular insertion of monopolar electrodes in the diaphragm. After such procedure, it was possible to reach an inspired volume of 380 ml when the left hemidiaphragmatic was stimulated with a current strength of 25mA. Nevertheless, the volume of the right hemidiaphragm was less than 100ml, suggesting that the right electrode was not properly positioned on the motor point. After new surgical procedure with appropriate fixation of the electrodes, a period of 20 weeks was required for diaphragmatic training and adaptation, resulting in maximum inspired volume of 1.120ml with a current strength of 25mA15.

Over the years, there has been a major breakthrough in technological development of new devices for electroventilation, which provided an important step toward creation of diaphragmatic pacemakers, placed through electrodes implanted directly on the diaphragm by laparotomy9-10. Despite displaying similar physiological response to nerve phrenic stimulation, this intramuscular technique showed to be less difficult to proceed in terms of muscle preparation and implantation11-12.

The full activation of the diaphragm depends on the proper implantation of electrodes on the diaphragm motor points, which represents the muscle area where the branches of the phrenic nerve penetrate into the diaphragm. A minimal amount of electrical stimulation should be necessary to cause full muscle contraction13-15. Specific anatomical landmarks of the diaphragm are usually used to place the electrodes, since the phrenic nerve branches are not readily identified on the bottom edge of the diaphragm11.

Ventilation with the use of electrical currents is supposed to be more physiological than mechanical ventilation because patients can keep using their own inspiratory muscles to inflate the lungs. On the other hand, there are many technical difficulties applying this technique, since it is necessary the full control over the muscle intended to be stimulated without interfering other organs function.

In our experiments, one of the major problems remains the difficulty to achieve appropriate identification of the diaphragmatic motor points. Furthermore, it is still a challenge the design of a specific electrode for diaphragmatic stimulation since many epimysial electrodes are designed for intramuscular stimulation, but not particularly for the diaphragm which is much thinner and delicate in comparison to skeletal muscle18. In our opinion, the current generator is also an unsolved problem, because the ideal device should have the ability to generate a reliable electric current of low power consumption in order to provide satisfactory security, mobility and certain autonomy.

Conclusion

This model has proved to be effective for studying the performance of the diaphragm under different electrical stimulations using different set of electrodes.

References

1. Organization WH. WHO Statistical Information System (WHOSIS). 2009 [cited 2010 April 01]. Available from: http://www.who.int/whosis/indicators/WHS09_IndicatorCompendium_20090701.pdf

2. Saúde Md. Secretaria de Vigilância em Saúde. 2008-01 [updated 2008; cited 2010 April 22]. Available from: http://tabnet.datasus.gov.br/cgi/tabcgi.exe?idb2008/c09.def.

Received: March 10, 2010

Review: May 11, 2010

Accepted: June 15, 2010

Conflict of interest: none

Financial source: FIPE at HCPA

How to cite this article

Ghedini RG, Margarites A, Felix EA, Xavier RG, Andrade CF. Development of a new experimental model of intramuscular electrical stimulation of the diaphragm in rabbits. Acta Cir Bras. [serial on the Internet] 2010 Nov-Dec;25(6). Available from URL: http://www.scielo.br/acb

*Color figure available from www.scielo.br/acb

  • 3. McCluskey L. Amyotrophic lateral sclerosis: ethical issues from diagnosis to end of life. NeuroRehabilitation. 2007;22(6):463-72.
  • 4. Glenn WW, Holcomb WG, Gee JB, Rath R. Central hypoventilation; long-term ventilatory assistance by radiofrequency electrophrenic respiration. Ann Surg. 1970;172(4):755-73.
  • 5. Geddes LA, Voorhees WD, Babbs CF, Deford JA. Electroventilation. Am J Emerg Med. 1985;3(4):337-9.
  • 6. Geddes LA, Voorhees WD, Lagler R, Riscili C, Foster K, Bourland JD. Electrically produced artificial ventilation. Med Instrum. 1988;22(5):263-71.
  • 7. Geddes LA. Electroventilationa missed opportunity? Biomed Instrum Technol. 1998;32(4):401-14.
  • 8. Sarnoff SJ, Hardenbergh E, Whittenberger JL. Electrophrenic respiration. Am J Physiol. 1948;155(1):1-9.
  • 9. Aiyar H, Stellato TA, Onders RP, Mortimer JT. Laparoscopic implant instrument for the placement of intramuscular electrodes in the diaphragm. IEEE Trans Rehabil Eng. 1999;7(3):360-71.
  • 10. Onders RP, Elmo M, Khansarinia S, Bowman B, Yee J, Road J, Bass B, Dunkin B, Ingvarsson PE, Oddsdóttir M. Complete worldwide operative experience in laparoscopic diaphragm pacing: results and differences in spinal cord injured patients and amyotrophic lateral sclerosis patients. Surg Endosc. 2009;23(7):1433-40.
  • 11. Peterson DK, Nochomovitz M, DiMarco AF, Mortimer JT. Intramuscular electrical activation of the phrenic nerve. IEEE Trans Biomed Eng. 1986;33(3):342-51.
  • 12. DiMarco AF, Supinski GS, Budzinska K. Inspiratory muscle interaction in the generation of changes in airway pressure. J Appl Physiol. 1989;66(6):2573-8.
  • 13. Peterson DK, Nochomovitz ML, Stellato TA, Mortimer JT. Long-term intramuscular electrical activation of the phrenic nerve: safety and reliability. IEEE Trans Biomed Eng. 1994;41(12):1115-26.
  • 14. Dunn RB, Walter JS, Walsh J. Diaphragm and accessory respiratory muscle stimulation using intramuscular electrodes. Arch Phys Med Rehabil. 1995;76(3):266-71.
  • 15. DiMarco AF, Onders RP, Kowalski KE, Miller ME, Ferek S, Mortimer JT. Phrenic nerve pacing in a tetraplegic patient via intramuscular diaphragm electrodes. Am J Respir Crit Care Med. 2002;166(12 Pt 1):1604-6.
  • 16. Howard-Jones N. A CIOMS ethical code for animal experimentation. WHO Chron. 1985;39(2):51-6.
  • 17. Peterson DK, Nochomovitz ML, Stellato TA, Mortimer JT. Long-term intramuscular electrical activation of the phrenic nerve: efficacy as a ventilatory prosthesis. IEEE Trans Biomed Eng. 1994;41(12):1127-35.
  • 18. Schmit BD, Mortimer JT. The effects of epimysial electrode location on phrenic nerve recruitment and the relation between tidal volume and interpulse interval. IEEE Trans Rehabil Eng. 1999;7(2):150-8.
  • Correspondence:
    Cristiano Feijó Andrade
    Hospital de Clínicas de Porto Alegre
    Serviço de Cirurgia Torácica
    Ramiro Barcelos, 2.350
    90035-903 Porto Alegre – RS Brasil
    Phone: (55 51)3359-8684
    Fax: (55 51)3359-8884
  • 1
    Research performed at Animal Experimentation Unit Research Center, airway and lung laboratory, Clinics Hospital of Porto Alegre (HCPA), RS, Brazil.
  • Publication Dates

    • Publication in this collection
      19 Nov 2010
    • Date of issue
      Dec 2010

    History

    • Accepted
      15 June 2010
    • Reviewed
      11 May 2010
    • Received
      10 Mar 2010
    Sociedade Brasileira para o Desenvolvimento da Pesquisa em Cirurgia https://actacirbras.com.br/ - São Paulo - SP - Brazil
    E-mail: actacirbras@gmail.com