Acessibilidade / Reportar erro

Noise in a neonatal Intensive Care Unit: measurement and perception of professionals and parents

Abstracts

OBJECTIVE: To study the perception about environmental noise of professionals and parents of neonates assisted in a Neonatal Intensive Care Unit (NICU), and to compare the findings with the measured noise levels. METHODS: The perception of parents and professionals that work in the NICU in relation to the presence of noise was evaluated by a questionnaire. Sound levels in three rooms and in the corridor of that environment were registered 24 hours/day during 9 days by the Quest Q-400 Noise Dosimeter and analyzed by QuestSuíteMR software. Kruskal-Wallis e Mann-Whitney tests were used to compare the noise levels in different places, being significant p<0.05. RESULTS: The average noise levels in the intensive, intermediary care, isolation rooms and in the corridor of the unit were 64.8, 62.1, 63.8 and 61.9dBA, respectively (p<0.001). Health professionals qualified the noise as present and intense, but parents evaluated the noise as moderate. Health professionals judged their own behaviors as noisy, and parents believe that they do not contribute to the existent noise at the place. Health professionals believed that newborns and professionals who work in the NICU may be injured by the noise, but this was not true for parents. All groups considered possible to reduce noise. CONCLUSIONS: The opinion about noise intensity differed significantly among health professionals and parents. Health professionals were more coherent about NICU's measured and perceived noise levels.

noise; infant, newborn; noise measurement; neonatology


OBJETIVO: Conhecer a percepção dos profissionais atuantes em Unidade de Terapia Intensiva Neonatal e dos pais dos recém-nascidos internados sobre o ruído existente, além de compará-la aos níveis de ruído mensurados. MÉTODOS: Por meio de questionário, investigou-se a percepção dos profissionais que atuam na unidade, bem como a dos pais dos recém-nascidos internados, quanto ao ruído existente no ambiente. Os níveis sonoros das três salas e do corredor da unidade foram registrados 24 horas/dia, por nove dias em cada local, com dosímetro Quest 400, e analisados pelo software QuestSuíteMR. Para comparar os níveis de ruído nos diferentes locais, aplicaram-se os testes de Kruskal-Wallis e Mann-Whitney, sendo significante p<0,05. RESULTADOS: Nas salas de cuidados intensivos e intermediários, no isolamento e no corredor da unidade foram registrados níveis médios de ruído de 64,8, 62,1, 63,8 e 61,9dBA, respectivamente, sendo diferentes entre si (p<0,001). Os profissionais da saúde consideraram o ruído presente e intenso e os pais, moderado. Só os primeiros julgam ruidoso o próprio comportamento. Os profissionais da saúde acreditam que tanto os recém-nascidos quanto as pessoas que trabalham nesse ambiente podem ser prejudicados pelo ruído, enquanto os pais acreditam que não. Todos os grupos julgam possível reduzir tal ruído. CONCLUSÕES: A opinião sobre a intensidade do ruído diferiu entre os profissionais da saúde e os pais dos recém-nascidos. Houve maior concordância entre o nível de ruído mensurado e a percepção do mesmo pelos profissionais da saúde.

ruído; recém-nascido; medição de ruído; neonatologia


ARTIGO ORIGINAL

IMestre em Distúrbios da Comunicação Humana pela UFSM, Fonoaudióloga

IIDoutora em Distúrbios da Comunicação Humana pela Universidade Federal de São Paulo (Unifesp); Professora-associada do Curso de Fonoaudiologia e do Programa de Pós-Graduação em Distúrbios da Comunicação Humana da UFSM, Santa Maria, RS, Brasil

Endereço para correspondência

ABSTRACT

OBJECTIVE: To study the perception about environmental noise of professionals and parents of neonates assisted in a Neonatal Intensive Care Unit (NICU), and to compare the findings with the measured noise levels.

METHODS: The perception of parents and professionals that work in the NICU in relation to the presence of noise was evaluated by a questionnaire. Sound levels in three rooms and in the corridor of that environment were registered 24 hours/day during 9 days by the Quest Q-400 Noise Dosimeter and analyzed by QuestSuíteMR software. Kruskal-Wallis e Mann-Whitney tests were used to compare the noise levels in different places, being significant p<0.05.

RESULTS: The average noise levels in the intensive, intermediary care, isolation rooms and in the corridor of the unit were 64.8, 62.1, 63.8 and 61.9dBA, respectively (p<0.001). Health professionals qualified the noise as present and intense, but parents evaluated the noise as moderate. Health professionals judged their own behaviors as noisy, and parents believe that they do not contribute to the existent noise at the place. Health professionals believed that newborns and professionals who work in the NICU may be injured by the noise, but this was not true for parents. All groups considered possible to reduce noise.

CONCLUSIONS: The opinion about noise intensity differed significantly among health professionals and parents. Health professionals were more coherent about NICU's measured and perceived noise levels.

Key-words: noise; infant, newborn; noise measurement; neonatology.

Introduction

Neonatal intensive care units (NICU) use a large number of technological resources to increase the survival of their newborn (NB) patients by guaranteeing them the best available treatment and the fastest recovery time. However, many of these technologies produce high sound pressure levels (SPL) and make the intensive care environment noisy, contributing to the appearance of physiological and behavioral disorders in people exposed to it.

The most common sources of noise are: mechanical ventilators, heated cots, infusion pumps, monitors, incubators(1), alarms(1-3) and air-conditioning(4-6). High noise levels have also been reported as being caused by workers laughing and conversing(1-4,6,7), lack of care opening and closing ICU doors(2,4,5,8,9); moving incubators; dragging chairs(4) and carelessness when closing cupboards, drawers and trashcans(2,5).

Exposing NB to high levels of noise does not only risk causing hearing deficiency(10), but can also alter sleeping patterns and cause irritability(11,12), agitation, crying, tiredness, increased oxygen consumption and elevated heart rate(11), with a negative impact on the healing process.

Professionals working in the presence of high sound pressure levels can suffer tiredness(13), headaches, reduced ability to concentrate, irritability, hypertension(14), altered cardiac rhythm(14,15) hearing loss(14,16), mood swings(15,17), psychiatric disorders (neuroses, psychoses and hysteria)(17) and stress(15,17,18).

Parents of patients subjected to high noise levels can suffer physiological and behavioral stress and sensory overload and the noise can also make it difficult for them to interact with their infants(13).

Current standards recommend that daytime noise levels in hospitals be maintained below 45dBA(10,19-22) with a Lmax(1 second) of 65dBA(22), while at night the limit is 35dBA(19). However the recommended levels, whether from Brazilian or international standards, are often exceeded.

The objective of this study is to investigate the perceptions of professionals working in the NICU at the Hospital Universitário de Santa Maria (HUSM) of the noise levels in the unit and also the perceptions of the unit's patients' parents and to compare these perceptions with actual noise levels measured in the unit.

Methods

This is a quantitative, non-experimental, descriptive and exploratory study using data collected in the NICU at the HUSM between the 10th of November and the 20th of December of 2007. The project was approved by the research ethics committee at the Universidade Federal de Santa Maria. Management at the HUSM and the NICU also gave their consent to the research after analyzing the proposal.

The unit comprises one intensive care ward (six/eight heated cots and incubators); one intermediate care ward (eight standard cots) and one isolation ward (three heated cots).

The study investigated those professionals working at the unit and the parents of its patients who did not have hearing deficiencies considered severe enough to interfere with the study and who agreed to take part. Employees who were on holiday or on leave were not included and neither were parents who did not feel emotionally able to take part because of their children's condition. No sample size calculation was made because the intention was to research the perceptions of those health professionals and parents who were in the unit during the study period. The final study sample comprised 43 (61%) health professionals, six (9%) professionals from other areas and 21 (30%) parents of patients who were in the NICU at the HUSM during the study period. Professionals who were not health professionals (secretarial and general staff) were classified as "other professionals" for analysis.

The lead author administered a questionnaire containing eight open or closed questions on: the respondent's opinion of the acoustic environment in the HUSM NICU; major sources of noise; possible effects on professionals, parents and patients and possibilities for reducing the noise in the unit.

At the same time that the questionnaire data was collected, noise levels in the NICU were measured. Sound levels were measured in all three wards and in the unit's corridor, 24 hours a day for nine days in each location, using a Quest 400 noise dosimeter calibrated and monitored by an occupational health and safety technician. There was an interval of 1 day between ending measurement in one location and setting up in the next, during which time the data were uploaded to a computer. Therefore, approximately 216 hours (12,960 minutes) of sound level data were collected at each location, making a total of more than 51,800 minutes of measurements for analysis. These noise measurements were planned on the basis of the availability of the noise dosimeter and no sample size calculation was performed.

The dosimeter was positioned in each measurement location, as described below:

• Intensive Care Ward: approximately 1m from the ceiling, on top of a cupboard fixed to one of the side walls of the ward.

• Intermediate Care Ward: approximately 80cm from the ceiling, also on top of a cupboard fixed to one of the side walls of the ward, this time to the left of the door.

• Isolation Ward: 1m 20cm from the ceiling, on top of a rectangular light fitting on the wall to the right of the door - the light fitting did not make any noise that could interfere with measurement.

• Corridor: 15cm from the ceiling on top of a cupboard, very close to the door leading to the intensive care ward.

It was not possible to position the noise dosimeter in the center of the wards or of the corridor, as recommended in the literature, because the flow of people, the unit's routine and the positioning of furniture were all unfavorable.

Fixing the noise dosimeter approximately 1m from the ceiling was also ruled out because the physical conditions were not conducive to this and it was not possible to interrupt the unit's routine or to make loud noises while fitting an appropriate support for the meter.

The meter was programmed to operate from 40 to 140 dB. The exchange rate adopted was the Brazilian standard of q=5dB(23), which means that sound intensity doubles for every 5dB increase. The battery was changed and the meter calibrated every day. The meter was also calibrated before starting measurements at each new measurement location.

Estatística 7.0was used to analyze data from the questionnaire and, wherever possible (closed questions), the chi-square test and the difference between two independent proportions were used, to a significance level of 5.

The SPL measurements were analyzed by TST using QuestSuíteMR software.

Noise levels at different measurement locations were compared using the Kruskal Wallis nonparametric test and the Mann-Whitney U test, both to a significance level of 5%.

Wherever possible, the questionnaire results were related to the noise level measurements.

Results

Table 1 shows the results for the parents of patients and the professionals' perceptions of whether the unit was noisy and, if so, of the intensity of the noise. There was only a statistically significant difference between the health professionals' and the parents' perceptions. When asked about the sources of noise in the unit, 98% of the health professionals and 67% of the other professionals said they thought the noise was primarily generated by equipment. A large proportion of the parents (48%) believed that equipment was primarily responsible for creating the noise, followed by 29% who believed that the professionals were the primary cause of the noise in the NICU. There were statistically significant differences between the health professionals' opinions and the opinions of the other professionals (p=0.003), and also between the opinions of the health professionals and the parents' opinions (p<0.001).

Table 2 shows the perceptions of the professionals working in the NICU and of the parents of the unit's patients about their own behavior in the unit. The health professionals' opinions differed statistically from the other two groups. The noisy behaviors mentioned with greatest frequency were unavoidable handling and movements and conversation.

Table 3 lists the groups' opinions on whether the noise in the NICU could be harmful to professionals, patients or parents. There were statistically significant differences between the opinions of the health professionals and the opinions of the other two groups on whether the noise in the NICU could harm the neonates (p<0.001). The health professionals' opinion on whether exposure to the noise in the NICU could be harmful to professionals was also different to the other two groups.

When asked about the possibility of effects harmful to the parents, the majority of respondents did not believe that they would be affected by the noise in the HUSM NICU because they did not spend long enough in the NICU. No statistically significant differences were detected between the three groups.

When asked whether it would be possible to reduce the noise in the unit, the majority of the health professionals (98%), the other professionals (67%) and the parents (57%) believed it would be possible, with a statistically significant difference between the proportion of health professionals and the proportion of other professionals (p=0.003) and between the proportion of health professionals and the proportion of parents who believed it would be possible to reduce noise (p<0.001).

Chart 1 illustrates the opinions of the different groups on the sources of noise in the HUSM NICU, the possibility of harmful effects to infants and professionals and on the measures that could be taken to reduce the noise in the unit studied.


Fifteen of the professionals investigated worked on the morning shift, 10 during the afternoon, 16 on the night shift, and 2 during the morning and afternoon. Taking just those professionals who worked on the morning shift, 53% thought the noise level was moderate and 47% thought it was intense. All mentioned that the noise is primarily caused by equipment and the majority (73%) thought they contributed to the noise level. All believed that both patients and professionals exposed to these noise levels could suffer harmful effects, but the majority (60%) believed that patients' parents would not be affected. All stated that it would be possible to reduce the noise in the unit.

Of those working in the afternoon, 80% classified the noise as intense, 90% of them believed that the noise in the unit is primarily caused by equipment and 80% thought they contributed to the noise on their shift. All mentioned that the noise in the unit could be harmful to both patients and professionals, but half believed that patients' parents would not be affected. Almost all (90%) stated that it would be possible to reduce the noise in the unit.

Among those working on the night shift, 69% classed the noise as intense and 31% as moderate; all thought that the noise is primarily caused by equipment and more than 55% thought they contributed to the noise level in the unit. All believed that both patients and professionals exposed to this environment could suffer harmful effects, but the majority (87%) mentioned that patients' parents would not be affected, probably as a result of the short exposure time. All stated that it would be possible to reduce the noise in the unit.

The mean noise levels recorded in the intensive and intermediate care wards, the isolation ward and the corridor were 64.8dBA, 62.1dBA, 63.8dBA and 61.9dBA respectively. Figure 1 illustrates the maximum and minimum average levels (AvgL) by shift and by measurement location. It was observed that in all four measurement locations the highest maximum and minimum AvgL both occurred during the afternoon shift. It was also observed that higher noise levels were recorded in the intensive care and isolation wards and the greatest variability was recorded in the intermediate care ward.


Taking the entire dataset, the noise level variability was also greatest in the intermediate care ward and the difference when compared to the other locations was statistically significant (p<0.001), but the other locations did not differ significantly from each other (p=0.26).

When the measurement results were compared with the respondents' perceptions of noise levels, the health professionals had closest agreement with recorded levels, particularly those who worked on the afternoon shift. Furthermore, the noise variability recorded supports the respondents' perceptions about sources of the noise (air-conditioning and professionals).

Discussion

It was found that the mean noise at all four measurement locations exceeded 45dBA for the majority of the measurement period. This finding suggests that this is an environment with elevated sound pressure levels and is in agreement with the perceptions of the great majority of the people in the sample, who considered the NICU to be a noisy environment. The health professionals' perceptions most clearly illustrate the presence of excessive sound levels, which do not meet the recommendations of Brazilian or international standards for neonatal and hospital environments(10,19-22).

On the subject of the sources of noise, all three groups' opinions agree with a study by Scochi et al(1) which found that the principal cause of noise in NICUs is the equipment. These findings also support the results of a recent study(24), which found that a large proportion of the nursing professionals in a NICU at a medium-size hospital in the state of São Paulo believed that the noise in their unit was primarily caused by equipment.

When asked to specify noise sources, a large proportion of respondents mentioned air-conditioning systems, in common with many other studies of NICUs(4-6,25,26), which have observed a significant increase in noise in front of running air-conditioners. Studies conducted by Oliveira et al(26) and Zamberlan(5) detected increases of 17dBA to 3.5dBA in the background noise levels in NICUs when air-conditioners were switched on.

The professional team's activities was the second most mentioned source of noise, in agreement with research by Chen and Chang(27), Diniz et al(6) and Carvalho et al(9) with conversation(1-4,6,7) being one of the most often mentioned activities. In agreement with these studies, the present study found that, after equipment noise, conversations were the second ranked cause of noise in the NICU(24).

The health professionals' perceptions of what behaviors increase this noise were similar to those listed in other studies that have identified elevated SPL in the environments studied as being caused by conversation(1-4,6,7), emergency procedures and treatment processes(8,28).

With reference to the possible harmful effects on newborn infants exposed to the SPL in the HUSM NICU, the health professionals mentioned hearing deficiency, which agrees with a document published by the American Academy of Pediatrics(10), which warns that exposure to the noise levels in PICUs can cause cochlear damage, because immature auditory structures may be more susceptible to lesions from a combination of noise and other risk factors.

Other harmful effects mentioned by these respondents were similar to the findings of other studies that have listed sleep disturbances, irritability(11,12), behavioral disorders(11) and stress(12) as possible harmful effects to neonates exposed to high sound levels.

With reference to the harmful effects of the noise in the HUSM NICU on the professionals working in the unit, the effect most often mentioned by health professionals was stress, similar to what was observed by Morrison, Haas, Shaffner et al(15), the World Health Organization(17) and Topf(18). The next most common were irritability(14), hearing problems(14,16), headaches(14), and tiredness(13), all of which have been mentioned in other studies. Also agreeing with the literature, the other professionals mentioned stress(15,17,18), irritability(14) and hearing problems(14,16) as possible harmful effects.

A large proportion of the parents of NICU patients believed that the professionals were unaffected by the noise, in common with reports published by Morrison, Haas, Shaffner et al(15), the World Health Organization(17) and Topf(18).

Asked about the possibility of harmful effects on parents exposed to the NICU noise levels, all three groups were in agreement that parents would not be affected since they spent so little time in the unit.

With reference to measures needed to reduce sound levels, in common with Diniz et al(6), our sample stated that the air-conditioners needed replacing. Reeducation of the professional team was also suggested by health professionals and parents, similar to the literature(3,6,9,25,29). Both health professionals and other professionals also mentioned the need to replace equipment for quieter models, which was also suggested by Lichtig and Maki(25). Reducing the volume of equipment and its sensitivity, suggested by the other professionals, has also been mentioned in a number of other studies(9,30). In addition to these measures, the parents mentioned taking more care not to drag chairs and to open and close trashcan lids more quietly, which agrees with a study by Holsbach et al(4).

The most intense sound pressure levels were recorded in the intensive care ward and the isolation ward, which have more life-support equipment. It is believed that the sound levels are intensified by the alarms on machines which, due to the conversations that take place in these wards and the distance between incubators and monitoring points, are set to maximum volume, as was found by Holsbach et al(4). It was expected that the SPL would be lower in the intermediate care ward, which has a smaller number of these machines.

When the mean noise levels were analyzed by shift, it was observed that in all measurement locations the noise levels were at their most intense during the afternoon shift and at their lowest during the night shift, in common with results published by Ichisato(2) and Corrêa(31). However, even during the night the noise levels exceeded the recommendations(19).

When the measured noise levels were compared with the health professionals' perceptions, the greatest degree of agreement was observed with the afternoon shift, since 80% of them considered the noise level in the NICU to be intense, while workers on the other two shifts were split between those who thought noise was moderate and those who thought it was intense. Furthermore, more professionals on the afternoon shift believed that their own behavior contributed to the noise, indicating their awareness that their shift had elevated SPL.

Noise level variability was greatest in the intermediate care ward, with periods of silence contrasting with periods of intense noise, possibly caused by the professionals or by air-conditioning turning on, since the ward has no life-support machines.

One limitation of this study was the failure to fix the dosimeter in the center of each measurement location, one meter from the floor or one meter from the ceiling. This may have affected the results since it prevents equal capture of the sound pressure levels originating from all points within the location. Furthermore, the low number of professionals who were not health professionals prevented extrapolation of the results of this group. Notwithstanding, it was still considered important to present their perceptions and to compare them with those of the health professionals.

In conclusion, the predominant perception among the health professionals was that the HUSM NICU has intense noise levels. They thought that they contribute to increasing the noise and that both newborn infants and professionals exposed to the noise could be affected by it. Half of the other professionals considered the noise in the NICU to be moderate and believed that these levels could be harmful to both patients and professionals. The majority of them did not think their own behavior was noisy. The parents thought the noise in this NICU was moderate and that they did not contribute to the noise. They believed that these levels would not affect the professionals. Opinions were divided on whether the neonates would suffer harmful effects, with a majority thinking they would not. All groups thought it would be possible to reduce the noise. The health professionals' perceptions exhibited the greatest degree of agreement with actual measured noise levels.

The findings of this study validate those of other studies conducted in neonatal units, finding excessive sound levels that do not meet Brazilian or international standards and recommendations for hospital and neonatal environments.

It was observed that the health professionals questioned considered the NICU studied to be noisy, that they demonstrated knowledge of the effects of that noise on newborn infants and professionals and that they believed that they themselves contributed to the noise levels, in contrast with the other two groups. These findings indicate a need to implement noise reduction programs in order to avoid the unit's patients suffering from physiological, psychological and behavioral disorders, thereby aiding their development and promoting the recovery process, in addition to improving the physiological and psychological wellbeing of the professional team. It is to be hoped that the fact that the health professionals are aware of their contribution to production of these excessive noise levels will facilitate implementation of the necessary programs.

References

  • 1. Scochi CG, Riul MI, Garcia CF, Barradas LS, Pileggi SO. Cuidado individualizado ao pequeno prematuro: o ambiente sensorial em unidade de terapia intensiva neonatal. Acta Paul Enferm 2001;14:9-16.
  • 2. Ichisato SM. Ruído em unidade de cuidado intensivo neonatal de um hospital universitário de Ribeirão Preto - SP[tese de doutorado]. Ribeirão Preto (SP): USP; 2004.
  • 3. Kakehashi TY, Pinheiro EM, Pizzarro G, Guilherme A. Nível de ruído em unidade de terapia intensiva neonatal. Acta Paul Enferm 2007;20:404-9.
  • 4. Holsbach LR, de Conto JA, Godoy PC. Avaliação dos níveis de ruído em Unidades de Tratamento Intensivo. Memoriais II Congresso Latioamericano de Ingeniaría Biomédica; 2001 Mayo 23-25; Habana, Cuba.
  • 5. Zamberlan NE. Ruído na unidade de cuidado intensivo de um hospital universitário de Ribeirão Preto-SP [tese de mestrado]. Ribeirão Preto (SP): USP; 2006.
  • 6. Diniz IN, Gomes Jr WM, Araújo GW. Determinação dos níveis de ruído nas unidades de terapia intensiva de dois hospitais de Belo Horizonte, visando uma melhoria na qualidade de vida [homepage on the Internet]. VIII Congresso de Ecologia do Brasil 2008; Caxambu, Brasil [cited 2008 Jul 20]. Available from: http://www.seb-ecologia.org.br/viiiceb/pdf/433.pdf
  • 7. Chang YJ, Lin CH, Lin LH. Noise and related events in a neonatal intensive care unit. Acta Paediatr Taiwan 2001;42:212-7.
  • 8. Kent WD, Tan AK, Clarke MC, Bardell T. Excessive noise levels in the neonatal ICU: potential effects on auditory system development. J Otolaryngol 2002;31:355-60.
  • 9. Carvalho WB, Pedreira ML, Aguiar MA. Noise level in a pediatric intensive care unit. J Pediatr (Rio J) 2005;81:495-8.
  • 10. [No authors listed]. Noise: a hazard for the fetus and newborn. American Academy of Pediatrics. Pediatrics 1997;100:724-7.
  • 11. Tamez RN, Silva MJ. Impacto do ambiente da UTI neonatal no desenvolvimento neuromotor. In: Tamez RN, Silva MJ. Enfermagem na UTI neonatal: assistência ao recém-nascido de alto risco. Rio de Janeiro: Guanabara Koogan; 1999. p. 147-54.
  • 12. Carvalho M. A influência do ambiente da UTI neonatal na assistência a recém-nascido de risco [homepage on the Internet]. XXXI Congresso Brasileiro de Pediatria; 2000 Out; Fortaleza, Brasil [cited 2007 Jul 12]. Available from: http//www.paulomargotto.com.br/documentos/utineo.doc
  • 13. Silva RN. Construindo o ambiente físico humanizado de uma UTI neonatal [homepage on the Internet]. III Encontro Nacional de Triagem Auditiva Neonatal Universal [cited in 2008 June 15]. Available from: http://www.gatanu.org/atualidades/Palestras%20III%20Encontro/16%2014h30%20Ricardo%20Nunes.pdf
  • 14. Parente S, Loureiro R. Quality improvement in ICU - ICU noise pollution. Eur J Anaesthesiology 2001;18 (Suppl 21):5.
  • 15. Morrison WE, Haas EC, Shaffner DH, Garrett ES, Fackler JC. Noise, stress, and annoyance in a pediatric intensive care unit. Crit Care Med 2003;31:113-9.
  • 16. Leme OL. Comparative audiometric study among noise exposed and unexposed hospital staff. Rev Bras Otorrinolaringol 2001;67:837-43.
  • 17. World Health Organization [homepage on the Internet]. Adverse health effects of noise [cited 2008 May 19]. Available from: http://www.who.int/docstore/peh/noise/Comnoise-3.pdf
  • 18. Topf M. Hospital noise pollution: an environmental stress model to guide research and clinical interventions. J Adv Nurs 2000;31:520-8.
  • 19
    United States Environmental Protection Agency. Information on levels of environmental noise requisite to protect public health and welfare with an adequate margin of safety (Report nº. 550/9-74-004). Washington DC: Government Printing Office; 1974.
  • 20
    Brasil - Associação Brasileira de Normas Técnicas. Níveis de ruídos para conforto acústico. NBR 10152. Rio de Janeiro: Associação Brasileira de Normas Técnicas; 1987.
  • 21. Brasil - Associação Brasileira de Normas Técnicas. Avaliação do ruído ambiente em recinto de edificações visando o conforto dos usuários - procedimento. Rio de Janeiro: Associação Brasileira de Normas Técnicas; 2008.
  • 22. White RD. Recommended standards for newborn ICU Design. Report of the Sixth Census Conference on Newborn ICU Design; 2006 Jan 25-27; Orlando, USA.
  • 23. Araújo GM, Regazzi RD. Perícia e avaliação de ruído e calor: passo a passo - teoria e prática. 2nd ed. Rio de Janeiro: Impresso do Brasil; 2002.
  • 24. Palma JGC, Pimenta PO, Corrêa AL, Freire SC. Conhecimento da equipe de enfermagem sobre ruídos na Unidade de Terapia Intensiva Neonatal. Anais XI Encontro Latino-Americano de Iniciação Científica e VII Encontro Latino-Americano de Pós-Graduação-Universidade do Vale do Paraíba; 2007; São José dos Campos, Brasil. p. 1332-5.
  • 25. Lichtig L, Maki K. Study of environmental and incubators noise levels in an neonatal intensive care unit. Pediatria 1992;14:30-4.
  • 26. Oliveira PF, França DC, Mor R. O nível de ruído na unidade de terapia intensiva neonatal e seus efeitos. Rev Cefac 2003;5:367-72.
  • 27. Chen HF, Chang YJ. Noise distribution of an incubator with nebulizer at a neonatal intensive care unit in Southern Taiwan. J Nurs Res 2001;9:25-32.
  • 28. Rodarte MD, Scochi CG, Santos CB. O ruído das incubadoras de um hospital de Ribeirão Preto - SP. Pro-Fono 2003;15:297-306.
  • 29. Evans JB, Philbin MK. Facility and operations planning for quiet hospital nurseries. J Perinatol 2000;20 (8 pt 2):S105-12.
  • 30. Cmiel CA, Karr DM, Gasser DM, Oliphant LM, Neveau AJ. Noise control: a nursing teams approach to sleep promotion. Am J Nurs 2004;104:40-8.
  • 31. Corrêa AL. Ruído: níveis de pressão sonora captados no interior e exterior de incubadora em Unidades de Cuidados Intensivos Neonatal [tese de mestrado]. São José dos Campos (SP): Univap; 2005.
  • Noise in a neonatal Intensive Care Unit: measurement and perception of professionals and parents

    Fernanda Soares AurélioI; Tania Maria TochettoII
  • Publication Dates

    • Publication in this collection
      12 July 2010
    • Date of issue
      June 2010

    History

    • Received
      08 July 2009
    • Accepted
      29 Sept 2009
    Sociedade de Pediatria de São Paulo R. Maria Figueiredo, 595 - 10o andar, 04002-003 São Paulo - SP - Brasil, Tel./Fax: (11 55) 3284-0308; 3289-9809; 3284-0051 - São Paulo - SP - Brazil
    E-mail: rpp@spsp.org.br