Fourier transform infrared determination of imidacloprid in pesticide formulations

Abstracts

A simple method has been developed for Fourier Transform Infrared (FTIR) determination of Imidacloprid in pesticide formulations. Samples were diluted with CHCl3 and the FTIR spectra of samples and standards were obtained at a nominal resolution of 4 cm-1 from 4000 to 900 cm-1 accumulating 25 scans. Imidacloprid determination was based on the measure of either peak area from 1577 to 1567 cm-1 or peak height at 1572 cm-1, both corrected using a horizontal baseline defined at 1900 cm-1. The limits of detection achieved, of the order of 9 µg g-1, were appropriate for the determination of Imidacloprid in commercially available formulations. FTIR results were statistically comparable to those found by High Performance Liquid Chromatography (HPLC). The procedure developed reduces organic solvent consumption, per sample preparation, from 53 mL CH3CN required for HPLC to 2.5 mL CHCl3, and reduces waste generation also increasing the sample measurement frequency from 10 h-1 to 60 h-1 as compared with the HPLC-UV reference method.

Imidacloprid; pesticide formulations; FTIR; HPLC-UV


Foi desenvolvido um método simples usando Infra-Vermelho com Transformada de Fourier (IRFT) para a determinação de Imidaclopride em formulações pesticidas. As amostram foram diluídas com CHCl3 e os espectros de IRFT de amostras e padrões foram obtidos numa resolução nominal de 4 cm-1, entre 4000 e 900 cm-1, com acumulação de 25 espectros. A determinação de Imidaclopride foi feita medindo-se tanto a área do sinal obtido entre 1577 e 1567 cm-1 ou a altura do sinal a 1572 cm-1, ambos com correção de linha base horizontal a 1900 cm-1. Os limites de detecção atingidos da ordem de 9 µg g-1, foram apropriados para a determinação de Imidoclopride em formulações comercias disponíveis. Os resultados com IRFT forma estatisticamente comparáveis àqueles encontrados para Cromatografia Líquida de Alta Performance (HPLC). O procedimento desenvolvido reduz o consumo de solventes orgânicos por preparação de amostra, de 53 mL de CH3CN requeridos com HPLC para 2,5 mL de CHCl3, reduzindo o descarte gerado, e aumenta a freqüência de medidas de 10 para 60 amostras h-1, quando comparado ao método de referência usando HPLC-UV.


ARTICLE

Fourier transform infrared determination of imidacloprid in pesticide formulations

Guillermo Quintás; Sergio Armenta; Salvador Garrigues* * e-mail: salvador.garrigues@uv.es ; Miguel de la Guardia

Department of Analytical Chemistry, University of Valencia, Edifici Jeroni Muñoz, 50th Dr. Moliner, 46100 Burjassot, Valencia, Spain

ABSTRACT

A simple method has been developed for Fourier Transform Infrared (FTIR) determination of Imidacloprid in pesticide formulations. Samples were diluted with CHCl3 and the FTIR spectra of samples and standards were obtained at a nominal resolution of 4 cm-1 from 4000 to 900 cm-1 accumulating 25 scans. Imidacloprid determination was based on the measure of either peak area from 1577 to 1567 cm-1 or peak height at 1572 cm-1, both corrected using a horizontal baseline defined at 1900 cm-1. The limits of detection achieved, of the order of 9 µg g-1, were appropriate for the determination of Imidacloprid in commercially available formulations. FTIR results were statistically comparable to those found by High Performance Liquid Chromatography (HPLC). The procedure developed reduces organic solvent consumption, per sample preparation, from 53 mL CH3CN required for HPLC to 2.5 mL CHCl3, and reduces waste generation also increasing the sample measurement frequency from 10 h-1 to 60 h-1 as compared with the HPLC-UV reference method.

Keywords: Imidacloprid, pesticide formulations, FTIR, HPLC-UV

RESUMO

Foi desenvolvido um método simples usando Infra-Vermelho com Transformada de Fourier (IRFT) para a determinação de Imidaclopride em formulações pesticidas. As amostram foram diluídas com CHCl3 e os espectros de IRFT de amostras e padrões foram obtidos numa resolução nominal de 4 cm-1, entre 4000 e 900 cm-1, com acumulação de 25 espectros. A determinação de Imidaclopride foi feita medindo-se tanto a área do sinal obtido entre 1577 e 1567 cm-1 ou a altura do sinal a 1572 cm-1, ambos com correção de linha base horizontal a 1900 cm-1. Os limites de detecção atingidos da ordem de 9 µg g-1, foram apropriados para a determinação de Imidoclopride em formulações comercias disponíveis. Os resultados com IRFT forma estatisticamente comparáveis àqueles encontrados para Cromatografia Líquida de Alta Performance (HPLC). O procedimento desenvolvido reduz o consumo de solventes orgânicos por preparação de amostra, de 53 mL de CH3CN requeridos com HPLC para 2,5 mL de CHCl3, reduzindo o descarte gerado, e aumenta a freqüência de medidas de 10 para 60 amostras h-1, quando comparado ao método de referência usando HPLC-UV.

Introduction

Imidacloprid, 1-(6-chloro-3-pyridylmethyl)-N-nitroimidazolidin-2-ylideneamine, 1-[(6-chloro-3-pyridinyl)methyl]-N-nitro-2-imidazolidinimine, (CAS 13826-41-3), is a general use pesticide, classified by the USA Environmental Protection Agency (EPA) in both toxicity class II and class III.1 This pesticide is found in a variety of commercial insecticides as soluble concentrate, concentrated suspension and as wettable powder in different concentrations between 20% m/v and 70% m/m.2

Imidacloprid is a systemic, chloro-nicotinyl insecticide for soil, seed and foliar uses, which is employed for the control of sucking insects including rice hoppers, aphids, thrips, whiteflies, termites, turf insects, soil insects and some beetles. It is commonly used on rice, cereal, maize, potatoes, vegetables, sugar beets, fruit, cotton, hops and turf, and is especially systemic when used for seed or soil treatment. The insecticidal effect of Imidacloprid is achieved through bioaccumulation of acetylcholine in the target insects.3

Imidacloprid has been determined at trace levels in different environmental and food samples by high performance liquid chromatography with either mass spectrometry (HPLC-MS)4-6 or UV detection (HPLC-UV) using conventional dispersive spectrometers7-10 or diode array detectors,11,12 with Imidacloprid being determined in water,4,9 fruits and vegetables 5-7,10,12 and greenhouse air.11

Gas chromatography has been also employed for Imidacloprid determination in coffee,13 vegetables14 and water and soil15 using mass spectrometric detection (GC-MS).

Non-chromatographic methods developed for Imidacloprid determination were based on enzyme-linked immunosorbent assays (ELISA)16-18 and flow injection analysis with photochemically induced fluorescence detection (FIA-PIF).19

For the determination of Imidacloprid in formulated insecticides, the Collaborative International Analytical Council (CIPAC) recommended procedure20 is based on HPLC-UV and it seems to be that the most extended methodology for the analysis of pesticide formulations.21,22

Fourier Transform Infrared spectrometry (FTIR) provides highly useful information about the structure of analites and it can be used for both qualitative and quantitative analysis.23,24 Because of this we have developed a series of analytical methodologies for the FTIR determination of Fluometuron,25 Buprofezin,26 Folpet and Metalaxyl.27 However there is no analytical procedure proposed in the literature for the determination of Imidacloprid in pesticide formulations by FTIR.

One of the identity tests proposed by CIPAC20 is based on the comparison of FTIR spectra of samples or technical standards with that of pure standard in the spectral region between 1700 and 1400 cm-1 using KBr discs. However, the possibilities of this technique for the quantification of Imidacloprid were not explored.

The aims of this work are the development of a fast and environmentally friendly procedure, based on the use of FTIR, for the determination of Imidacloprid in commercial formulations which could be applied for quantitative purposes in both commercial formulations and technical products.

Experimental

Apparatus and reagents

A Nicolet (Madison, WI, USA) Magna 750 FTIR spectrometer, equipped with a temperature-stabilized deuterated tryglycine sulfate (DGTS) detector and controlled with Nicolet Omnic 2.1 (Madison, WI, USA), was employed for spectral measurements of CHCl3 solutions of Imidacloprid using a micro flow cell with a pathlength of 0.11 mm (Graseby-Specac, Orpinghton, UK) with ZnSe and CaF2 windows.

For introduction of samples and standards into the flow cell a Gilson Minipuls 2 peristaltic pump (Villiers-le-Bel, France) equipped with solvent resistant viton tubes of 3 mm od and 1 mm id was employed. The connection between the cell and the peristaltic pump was made using a 0.8 mm id PTFE tube.

A Hewlett-Packard HPLC Series 1050 High Performance Liquid Chromatograph, equipped with a Kromasil C-18, 5 µm particle diameter, 250 x 4.6 mm id column, and a variable wavelength UV-Vis detector, was also employed for the analysis of insecticide formulations, this methodology being employed as a reference for the validation of the FTIR measurements.

The Imidacloprid standard (99.9% m/m) was purchased from Fluka (Buchs, Switzerland). Analytical grade chloroform stabilized with 150 mg L-1 amylene and HPLC grade acetonitrile were supplied by Scharlau (Barcelona, Spain) and were employed for the preparation of samples and standards, also using Milli-Q grade water for the mobile phase.

Imidacloprid, 20% (m/v) soluble concentrate commercial formulations, were obtained directly from the Spanish market.

Reference procedure

Weigh accurately 40 mg of sample and dilute with CH3CN in a 25 mL volumetric flask, sonicating during 5 min in an ultrasound water bath. Dilute 1 mL of this solution to 25 mL and filter through a 0.22 µm nylon filter. Inject 20 µL of this latter solution in the chromatographic system using a 60:40 acetonitrile:water mobile phase at 1 mL min-1 Imidacloprid being determined in the isocratic mode by absorbance measurements at 285 nm. For the quantification of Imidacloprid, peak areas obtained at a retention time of 3.5 min for the samples were interpolated using external calibration established using standard solutions of Imidacloprid in acetonitrile over the concentration range from 3 to 30 µg mL-1.

Recommended procedure

Weigh accurately 20 mg of sample and dilute with 4 g chloroform using ultrasonic agitation for 5 min. Filter this solution through a 0.22 µm nylon filter and aspirate directly into the FTIR measurement cell by using a peristaltic pump. The FTIR spectra were obtained, from 4000 to 860 cm-1, in the stopped-flow mode at 4 cm-1 nominal resolution, accumulating 25 scans using, as a background, the cell filled with CHCl3.

The concentration of Imidacloprid in the samples was obtained by interpolation of peak areas from 1577 to 1567 cm-1 with an horizontal baseline fixed at 1900 cm-1 using external calibration established under the same conditions for Imidacloprid standards in a concentration range from 0.6 to 3.3 mg g-1.

Results and Discussion

FTIR spectra of Imidacloprid

Figure 1 shows the FTIR spectra, in the wavenumber region from 2000 to 950 cm-1, of a chloroform solution of Imidacloprid at 3.3 mg g-1, a sample solution containing 0.8 mg g-1 Imidacloprid and a technical product solution containing 4.0 mg g-1 of Imidacloprid.


The spectra of both the technical product and the pure standard solution in CHCl3 present the same absorption bands, corresponding to the Imidacloprid molecule. On the other hand, it can be seen that the presence of different excipients in the formulation provide additional bands which, to a large extend, produce spectral differences between the FTIR spectra of the samples and the standards. Despite that, the bands at 1572 and 1548 cm-1, due to the =CH- stretching of the pyridine group, are clearly observed in all cases and seem to be free from matrix interferences.

The -NO2 symmetric stretching band located at 1300 – 1250 cm-1, while the bands in the region between 1420 and 1500 cm-1, and that at 1109 cm-1, which are probably due to the aromatic =CH- and to the aryl-Cl stretching, respectively,28 seem to be affected by the formulation components and thus can not be employed for Imidacloprid determination without carrying out matrix removal.

Selection of the appropriate bands

As can be seen in Figure 1 the bands at 1572 and 1548 cm-1 are the bands of Imidacloprid which are clearly isolated and, apparently, do not overlap with any band of the different excipients in the formulated samples.

Thus, these bands were selected for Imidacloprid measurements. To obtain the best analytical performance of the FTIR determination of Imidacloprid, the use of peak height and peak area were evaluated, also using different baseline corrections.

Table 1 summarizes the main characteristics of the different regression lines found using the 1572 cm-1 and the 1548 cm-1 bands. As it can be seen, the band at 1548 cm-1 provides a reduced sensitivity and a high limit of detection, also involving a reduced linearity to the calibration lines, as compared with results found using the band at 1572 cm-1. For this latter band the regression lines had linear coefficients higher than 0.9999 showing the good linearity of the fits. The limits of detection obtained for the latter band ranged from 4 to 9 µg g-1, from 0.04 to 0.1 % m/v in the formulated pesticide, being thus appropriate for the determination of Imidacloprid in formulations. On the other hand the use of peak areas does not significantly improve the limits of detection despite the higher sensitivities found when peak areas is employed. Thus the use of either peak height at 1572 cm-1 or peak area from 1577 to 1567 cm-1, both corrected using an horizontal baseline defined at 1900 cm-1 are indicated, because these measurement conditions provide the lowest RSD values, and limits of detection of 9 and 6 µg g-1, respectively, which are equivalent to 0.1 and 0.07% m/v of Imidacloprid in the original sample when using a sampling mass of 20 mg.

In the selected conditions the regression lines obtained using an external calibration set were comparable with those found by standard addition, which were A1577-1567 = (0.045 ± 0.001) + (0.559 ± 0.004) CIMID (mg g-1), with a regression coefficient r2 = 0.9993, and H1757 = (0.052 ± 0.002) + (0.062 ± 0.001) CIMID (mg g-1), with a regression coefficient r2 = 0.9992, thus indicating the absence of matrix effects.

Measurement conditions

In order to select the most appropriate measurement conditions for the FTIR determination of Imidacloprid, the effect of using different nominal resolution values was studied and also that of the number of accumulated scans per spectrum on the signal to noise ratio of the Imidacloprid band at 1572 cm-1.

The effect of the nominal resolution on the spectral size and shape of the bands of Imidacloprid in CHCl3 is shown in Figure 2. From the inset the effect of the resolution on the peak height at 1572 cm-1, corrected with a horizontal baseline established at 1900 cm-1 is observed. The use of a high value of nominal resolution can not be recommended, because of the low sensitivity achieved and the deformation of the absorption bands, due to the low collection time required and the low noise obtained. A nominal resolution of 4 cm-1 was chosen, which provides good sensitivity, with a loss of only a 2.4% with respect to that found using 1 cm-1, and implies a 50% reduction of the acquisition time, thus increasing the sample throughput with a small sensitivity loss.


Figure 3 shows, for a nominal resolution of 4 cm-1, the effect of increasing the number of scans, from 5 to 50, on the noise measured as root mean square (RMS) of absorbance signals of pure CHCl3 in the region between 1577 and 1567 cm-1. In order to ensure a compromise between sample measurement frequency and signal to noise ratio, 25 scans were selected for further measurements.


Comparison between HPLC and FTIR methods

Table 2 shows the data found for the determination of Imidacloprid in three commercially available pesticide formulation samples, obtained by using both the FTIR procedure and the HPLC-UV reference method and, as can be seen, statistically comparable data were found by both procedures using either peak height or peak area measurements for the band at 1572 cm-1, thus showing that the accuracy of the developed procedure is comparable to that found by the CIPAC recommended strategy. Additionally a higher repeatability found on using FTIR measurements can be noticed compared to that found by HPLC, probably due to the use of the successive dilutions involved in the chromatographic procedure.

Table 3 summarizes the main analytical parameters found using the developed procedure compared with those of the reference HPLC-UV method. The low limit of detection achieved by HPLC-UV makes it appropriate for the determination of Imidacloprid at low concentration levels as, for example, in treated crops or waste waters, but the concentration level at which HPLC works also requires a high sample dilution for the analysis of pesticide formulations which increases waste generation and degrades the precision of the method.

The solvent volume employed for the measurement of a single sample was 2.5 mL of CHCl3 for FTIR determination and 53 mL of CH3CN for HPLC, it also being necessary, in this later case, the use of a mobile phase which increases waste generation in 60 mL h-1. So the FTIR procedure can be considered an environmentally friendly alternative to the use of HPLC-UV in the analysis of formulations. On the other hand, the sample throughput of the FTIR procedure was 60 samples h-1, compared to 10 h-1 for HPLC, thus increasing laboratory efficiency and providing faster methodology.

Conclusions

The FTIR procedure developed for the quantitative determination of Imidacloprid in pesticide formulations provides statistically comparable results to those found by using a reference HPLC procedure and clearly reduces sample handling by reducing the need of sample dilution resulting in a significant reduction of waste generation and offering a faster alternative method.

Acknowledgements

Authors acknowledge the financial support of the Generalitat Valenciana Project GV01-249 and the grant provided by the Laboratorio de Higiene Laboral y Ambiental of the Universitat de Valencia and Laboratorios Ibáñez to carry out this study. S. Armenta acknowledges the FPU Grant of the Ministerio de Educación, Cultura y Deporte de España (Ref. AP2002-1874).

Received: February 25, 2003

Published on the web: April 6, 2004

  • 1. Meister, R.T., ed.In Farm Chemicals Handbook 1994; Meister Publishing Company: Willoughby, OH, 1994.
  • 2. de Liñan, C.; Vademecum de Productos Fitosanitarios y Nutricionales, Ed. Agrotécnica S. L.: Madrid, 2000.
  • 3. Kidd, H.; James, D. R., eds. In The Agrochemicals Handbook, 3rd ed., Royal Society of Chemistry Information Services: Cambridge, UK, 1991.
  • 4. Hernandez, F.; Sancho, J. V.; Pozo, O.; Lara, A.; Pitarch, E.; J. Chromatogr. A 2001, 939, 1.
  • 5. Pous, X.; Ruiz, M. J.; Pico, Y.; Font, G.; Fresenius' J. Anal. Chem. 2001, 371, 182.
  • 6. Fernandez-Alba, A. R.; Tejedor, A.; Aguera, A.; Contreras, M.; Garrido, J.; J. AOAC. Int 2000, 83, 748.
  • 7. Dai, H.; Li, Y. J; Zhang, Y.; Fenxi Ceshi Xuebao 2002, 21, 70. (CA 136:305.493)
  • 8. He, J.; Chang, J. B.; Guo, R. Y.; Liu, P.; Lin, S. F.; Chen, R. F.; Wang, Q. A.; Sepu 2000, 18, 181.
  • 9. Guiberteau-Cabanillas, A.; Galeano-Diaz, T.; Mora-Diez, N.; Parrilla-Vazquez, P.; Salinas-Lopez, F.; Talanta 2001, 53, 943.
  • 10. Ishii, Y.; Kobori, I.; Araki, Y.; Kurogochi, S.; Iwaya, K.; Kagabu, S.; J. Agric. Food Chem 1994, 42, 2917.
  • 11. Garrido-Frenich, A.; Egea-Gonzalez, F. J.; Martinez-Vidal, J. L.; Parrilla-Vazquez, P.; Mateu-Sanchez, M. J.; J. Chromatogr. A 2000, 869, 497.
  • 12. Fernandez-Alba, A. R.; Valverde, A.; Aguera, A.; Contreras, M.; Chiron, S.; J. Chromatogr. A 1996, 721, 97.
  • 13. Lodevico, R. G.; Li, Q. X.; Anal. Lett 2002, 35, 315.
  • 14. Navalon, A.; Gonzalez-Casado, A.; El-Khattabi, R.; Vilchez, J. L.; Fernandez-Alba, A. R.; Analyst 1997, 122, 579.
  • 15. Vilchez, J. L.; El-Khattabi, R.; Fernandez, J.; Gonzalez-Casado, A.; Navalon, A.; J. Chromatogr. A 1996, 746, 289.
  • 16. Lee, J. K.; Ahn, K. C.; Park, O. S.; Kang, S. Y.; Hammock, B. D.; J. Agric. Food. Chem. 2001, 49, 2159.
  • 17. Wanatabe, S.; Ito, S.; Kamata, Y.; Omoda, N.; Yamazaki, T.; Munakata, H.; Kaneko, T.; Yuasa, Y.; Anal. Chim. Acta 2001, 427, 211.
  • 18. Li, K.; Li, Q. X.; J. Agric. Food. Chem. 2000, 48, 3378.
  • 19. Sarkar, M. A.; Biswas, P. K.; Roy, S.; Kole, R. K.; Chowdhury, A.; Bull. Environ. Contam. Toxicol 1999, 63, 604.
  • 20. Collaborative International Pesticides Analytical Council (CIPAC) Handbook; CIPAC Ltd; 1994, Volume H, 185.
  • 21. Li, L. S.; Fenxi. Shiyanshi 1998, 17, 60. (CA 130: 91.577)
  • 22. Macke, M. M.; J. AOAC Int. 1998, 81, 344.
  • 23. Cadet, F.; de la Guardia, M., In Quantitative Analysis, Infrared Encyclopedia of Analytical Chemistry Meyers, R. A., ed., John Wiley & Sons, Ltd: Chichester, 2000.
  • 24. Garrigues, S.; de la Guardia, M.; Infrared Quantitative Analysis; Encyclopedia of Analytical Chemistry, John Wiley & Sons, Ltd: Chichester, 2002, vol. 2.
  • 25. Armenta S.; Quintás G.; Moros J.; Garrigues S.; de la Guardia M.; Anal. Chim. Acta 2002, 468, 81.
  • 26. Quintás G.; Morales-Noé A.; Parrilla C.; Garrigues S.; de la Guardia M.; Vib. Spectroscopy 2003, 31, 63.
  • 27. Quintás, G.; Armenta, S.; Morales-Noé, A.; Garrigues, S.; de la Guardia, M.; Anal. Chim. Acta 2003, 480, 11.
  • 28. Lin-Vien, D.; Colthup, N.B.; Fateley, W.G.; Grasselli, J.G.; Infrared and Raman Characteristic Frequencies of Organic Molecules Academic Press: London, 1991, 296.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      29 June 2004
    • Date of issue
      Apr 2004

    History

    • Received
      25 Feb 2003
    • Accepted
      06 Apr 2004
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br