Acessibilidade / Reportar erro

Sweet potato tissue-epoxy resin composite biosensor for hydroquinone determination in photographic process wastewater

Abstracts

A biosensor of graphite-epoxy resin-sweet potato tissue (source of polyphenol oxidase) was developed in this work. The evaluated compositions of graphite-epoxy resin-vegetable tissue were varied from 30 to 80% (m/m) graphite; from 5 to 60% (m/m) epoxy resin and from 5 to 20% (m/m) tissue. The cyclic and linear sweep voltammograms were obtained at a scan rate between 25 and 300 mV s-1 in the pH range from 3 to 8 in 0.1 mol L-1 phosphate buffer solutions. Several phenolic compounds were determined by the proposed biosensor and hydroquinone was determined in wastewater from photographic process. Relative standard deviations of 0.7% and 3.3% were obtained employing the same biosensor (n=10) and three different biosensors (n=10), respectively. Sensitivity and selectivity were evaluated employing the biosensor for hydroquinone with a lifetime of at least 6 months (more than 500 determinations).

composite biosensor; hydroquinone; sweet potato tissue; wastewater


Um biossensor de grafite-resina epóxi-tecido de batata doce (fonte de polifenol oxidase) foi desenvolvido nesse trabalho. As composições avaliadas de grafite-resina epóxi-tecido vegetal variaram de 30 a 80% (m/m) de grafite, de 5 a 60% (m/m) de resina epóxi e de 5 a 20% (m/m) de tecido. Os voltamogramas de varredura cíclica e linear foram obtidos variando a velocidade de varredura de 25 a 300 mV s-1 em uma faixa de pH de 3 a 8 em tampão fosfato 0,1 mol L-1. Vários compostos fenólicos foram determinados utilizando o biossensor proposto e hidroquinona foi determinada em águas residuárias de processo fotográfico. Desvios padrão relativos de 0,7% e 3,3% foram obtidos empregando o mesmo biocompósito (n=10) e três diferentes biossensores construídos (n=3), respectivamente. Sensibilidade e seletividade foram avaliadas empregando o biossensor para hidroquinona com um tempo de vida superior a 6 meses (mais de 500 determinações).


ARTICLE

Sweet potato tissue-epoxy resin composite biosensor for hydroquinone determination in photographic process wastewater

Karina O. Lupetti; Giacomo Zanotto-Neto; Orlando Fatibello-Filho* * e-mail: bello@dq.ufscar.br

Departamento de Química, Universidade Federal de São Carlos, CP 676, 13560-970 São Carlos-SP, Brazil

ABSTRACT

A biosensor of graphite-epoxy resin-sweet potato tissue (source of polyphenol oxidase) was developed in this work. The evaluated compositions of graphite-epoxy resin-vegetable tissue were varied from 30 to 80% (m/m) graphite; from 5 to 60% (m/m) epoxy resin and from 5 to 20% (m/m) tissue. The cyclic and linear sweep voltammograms were obtained at a scan rate between 25 and 300 mV s-1 in the pH range from 3 to 8 in 0.1 mol L-1 phosphate buffer solutions. Several phenolic compounds were determined by the proposed biosensor and hydroquinone was determined in wastewater from photographic process. Relative standard deviations of 0.7% and 3.3% were obtained employing the same biosensor (n=10) and three different biosensors (n=10), respectively. Sensitivity and selectivity were evaluated employing the biosensor for hydroquinone with a lifetime of at least 6 months (more than 500 determinations).

Keywords: composite biosensor, hydroquinone, sweet potato tissue, wastewater

RESUMO

Um biossensor de grafite-resina epóxi-tecido de batata doce (fonte de polifenol oxidase) foi desenvolvido nesse trabalho. As composições avaliadas de grafite-resina epóxi-tecido vegetal variaram de 30 a 80% (m/m) de grafite, de 5 a 60% (m/m) de resina epóxi e de 5 a 20% (m/m) de tecido. Os voltamogramas de varredura cíclica e linear foram obtidos variando a velocidade de varredura de 25 a 300 mV s-1 em uma faixa de pH de 3 a 8 em tampão fosfato 0,1 mol L-1. Vários compostos fenólicos foram determinados utilizando o biossensor proposto e hidroquinona foi determinada em águas residuárias de processo fotográfico. Desvios padrão relativos de 0,7% e 3,3% foram obtidos empregando o mesmo biocompósito (n=10) e três diferentes biossensores construídos (n=3), respectivamente. Sensibilidade e seletividade foram avaliadas empregando o biossensor para hidroquinona com um tempo de vida superior a 6 meses (mais de 500 determinações).

Introduction

Nowadays, composite biosensors are efficiently used in different kinds of analytical determinations for several electroactive species, showing robustness and sensitivity. Several materials have been used to construct electrodes. Among them, carbon-based matrices are the most applied due to their high conductivity, whereas epoxy resins, silicone, polyurethane, metacrylate resin, Teflon, etc, can be employed as agglutinating agents (insulator materials). Graphite and epoxy resins are employed to construct rigid composites that can be used in aqueous or nonaqueous media,1 so the great advantage of graphite epoxy electrodes over carbon paste electrodes is their tolerance to organic solvents. This type of composite biosensor offers additional important advantages: it can be customized easily according to biocatalytical and/or electrochemical requirements; it presents low background currents, favoring a high signal-to-noise ratio, and lower detection limits and it offers a new active layer after removal of the outer surface, extending the lifetime of such electrodes.2

Vegetable tissues have been employed as source of several enzymes, among them, polyphenol oxidase (PPO) and peroxidase3,4 are applied to modify efficiently carbon paste electrodes for different kinds of analytes, mainly phenolic compounds. A review focusing the utility of vegetable tissue and crude extracts can be found in the literature.5

Hydroquinone is a phenolic compound that is important in a wide number of biological, cosmetic and industrial processes (dermatological cream, coal-tar production, paper manufacturing or in photographic developing). In the aquatic environment, it is considered an important xenobiotic micropollutant. Thus, hydroquinone and its reaction products have been determined by several analytical techniques in cosmetics, hair products and pharmaceutical preparations, in air samples and in different biological fluids due to its possible carcinogenic effect. The determination and quantification of this analyte may be done using different techniques, such as HPLC6,7 with different detectors, flow injection analysis,8 kinetic spectrophotometry,9,10 GC/MS,11,12 differential pulse voltammetry,13 the Belousov–Zhabotinski oscillating chemical reaction in a continuous-flow stirred-tank reactor14,15 or with biosensors.16,17 The detection limits found in the literature for hydroquinone vary in a wide range of concentrations depending on the technique used, the lowest limits are obtained by using electrochemical detection systems. Fang et al.18 describe a system of analysis by flow injection with amperometric detection that reaches a detection limit of 22 µg L-1 for hydroquinone. This limit descends to 0.05 µg L-1 with capillary electrophoresis or liquid chromatography with electrochemical detection.

There are several works19-23 describing the application of vegetable tissues as the source of PPO in different determination systems for phenolic compounds and catecholamines. Nevertheless, mainly pharmacological compounds were determined employing these materials. In addition, only carbon paste23 or Clark type24 electrodes have been reported as the basis for this type of enzymatic modification. The use of epoxy resin and graphite powder improves some characteristics of the sensor for the determination of phenolic compounds that are environmentally damaging.

In the present work, a suitable composite biosensor of sweet potato tissue- graphite-epoxy resin was developed for hydroquinone determination in photographic process wastewater.

Experimental

Apparatus

A knife-grinder from Tecnal TE 631 was employed to grind the dried sweet potato tissue into small particles and mesh sieves (Granutest) were used to select different particle sizes (0.106; 0.250; 0.350; 0.500 mm) for the composite biosensor construction.

All electrochemical experiments were carried out in a 15 mL thermostated glass cell at 25 ºC. A three electrodes assembly incorporating the biosensor as working electrode; an Ag/AgCl (3.0 mol L-1 KCl) reference and platinum auxiliary electrodes were used in all experiments. Cyclic voltammetry (CV) and linear sweep voltammetry (LSV) measurements were performed with an EG & G PAR, Model 273 A Potentiostat/Galvanostat analyzer controlled by a GPIB PCIIA interfacial program.

Reagents

All chemicals were of analytical-reagents grade and all solutions were prepared with water from a Millipore (Bedford, MA, USA) Milli-Q system (Model UV Plus Ultra-Low Organic Water).

The composite biosensor was prepared using an Acheson 38 graphite powder from Fisher, epoxy resin from Reforplás and PPO from sweet potato tissue. Healthy sweet potato (Ipomoea batatas Lam (L.)) was purchased from a local producer.

All phenolic compounds, such as catechol, hydroquinone, chloromethylphenol, 4-nitrophenol, 4-chlorophenol and phenol were purchased from Aldrich and 2.5×10-2 mol L-1 stock solutions were prepared in 0.1 mol L-1 phosphate buffer solution at pH 7.0.

Hydroquinone´s wastewater samples were obtained from a local photographic shop.

Biosensor preparation

The bicomposite was initially prepared by homogenizing the mass varying from 0.150 to 0.400 g of graphite powder and a specific mass (0.025g; 0.050g; 0.075g and 0.100g) of sweet potato tissue in a mortar for at least 20 min. After that, 0.025 to 0.300g of epoxy resin was subsequently added to this mixture in a mortar and mixed for at least 20 min to produce the final biocomposite, that was packed (330 mg) into the tip of a 1 mL insulin plastic syringe and a copper wire was used to provide the external electric contact.

Voltammetric measurements

Cyclic and linear sweep voltammetry measurements were performed in an unstirred and non-aerated 0.1 mol L-1 phosphate buffer solution at pH 7.0 and 25 ºC. The cyclic voltammograms were obtained by cycling the potential from 1.0 to –1.0 V vs. Ag/AgCl (3.0 mol L-1 KCl) reference electrode at a scan rate of 200 mV s-1, after successive additions of phenolic compounds or sample solutions. The linear sweep voltammograms were performed between 0.25 and –0.55 V vs. Ag/AgCl at a scan rate of 200 mV s-1.

Optimization of the biosensor response

Intending to achieve the best composition for the proposed composite biosensor some studies were performed. The composite biosensor response for 0.025 mol L-1 catechol solution in 0.1 mol L-1 phosphate buffer solution at pH 7.0 was initially studied for compositions varying from 30 to 80% (m/m) of graphite powder, from 5 to 60% (m/m) epoxy resin, and from 5 to 20% (m/m) vegetable tissue. Also, the effect of particle sizes of the dried sweet potato tissue was evaluated in the range from 0.106 mm to 0.500 mm.

The pH was varied from 3.0 to 8.0 and the effect on the composite biosensor response for 4.5 × 10-4 mol L-1 catechol solution was also investigated.

Initial stirring time varying from 30 to 150 s and scan rate from 25 to 300 mV s-1 have been also studied, employing the same catechol solution used for pH studies.

Recovery and repeatability studies

Aiming to evaluate the recovery, 4.95×10-4, 9.80×10-4 and 1.45×10-3 mol L-1 of hydroquinone were added to the sample of wastewater from three photographic process and the cathodic peak current was registered. In order to estimate the repeatability, ten reference solutions of 5.0×10-4 mol L-1 hydroquinone were measured in 0.1 mol L-1 phosphate buffer solution at pH 7.0.

Linearity and relative response studies

After the optimization studies, an analytical curve was plotted using the developed composite biosensor for hydroquinone solutions in the following concentrations: 2.5×10-4, 7.4×10-4, 1.2×10-3, 1.7×10-3, 2.1×10-3, 3.0×10-3 and 3.9×10-3 mol L-1.

Several phenolic compounds were determined by the proposed methodology, among them: catechol; hydroquinone; chloromethylphenol; 4-nitrophenol; 4-chlorophenol and phenol and a relative response relationship were obtained for this composite biosensor.

Hydroquinone determination and comparative method

Hydroquinone was chosen to be determined in wastewater from photographic process by the standard addition procedure, using linear sweep voltammetry measurements due to the sensibility of this technique, considering the analyte concentration in the samples.

A comparative methodology employing the redox titration of the analyte with 0.05 mol L-1 Ce4+ solution was performed to determine the hydroquinone concentration in the samples.25

Results and Discussion

The robustness and sensitivity are the most interesting characteristics of this kind of biosensor for phenolic compounds determination. Figure 1 presents the difference in the CV response between a plain carbon paste electrode and the proposed biosensor. In this study, cyclic voltammetric measurements were performed scanning the potential between 1.0 and –1.0 V vs. Ag/AgCl at a scan rate of 200 mV s-1 using both electrodes. The cyclic voltammograms A and C were obtained using both electrodes in 0.1 mol L-1 phosphate buffer solution, respectively. The cyclic voltammograms B and D were obtained using the plain carbon paste and the biosensor in a 2.34 ×10-3 mol L-1 catechol solution, respectively. As can be seen, the composite biosensor presents higher signal (higher cathodic current at –0.20 V) than the plain carbon paste in the presence of catechol solution.


Composite biosensor optimization

To obtain the highest signal to noise ratio for the composite biosensor with polyphenol oxidase, the effects of several variables were evaluated employing the proposed device. Table 1 summarizes the range over each variable was studied and the optimal value found in those studies.

The effect of graphite powder varying from 30 to 80% (m/m), epoxy resin from 5 to 60% (m/m), and tissue from 5 to 20% (m/m) on the biosensor response using 0.025 mol L-1 catechol solution at pH 7 was studied. The best composition obtained was 75:15:10% (m/m) graphite: vegetable tissue: epoxy resin, respectively. Besides, the diameter of the vegetable tissue particles was evaluated in the range from 0.106 mm to 0.500 mm and the higher signal to noise ratio was obtained with 0.350 mm particles size.

The highest signal for phenolic compounds was found at pH 7. In this study, a scan rate of 200 mV s-1 was chosen to achieve a higher cathodic peak current at –0.25 V. In addition, the effect of initial stirring time in an interval from 30 to 150 s on the biosensor, for 5.0×10-4 mol L-1 hydroquinone in 0.1 mol L-1 phosphate buffer solution at pH 7.0, was investigated. It was observed that 120 s was the best initial stirring time to determine this substrate.

Recovery and repeatability studies

Recoveries studies of 4.95×10-4, 9.80×10-4 and 1.45×10-3 mol L-1 of hydroquinone from 3 samples of a photographic process wastewater varied from 96.8 to 103% (Table 2) using the composite biosensor, suggesting that there are not matrix effects in hydroquinone determination.

After 10 additions of 5.00×10-4 mol L-1 hydroquinone in 0.1 mol L-1 phosphate buffer solution (pH 7.0) a relative standard deviation of 0.7% was obtained and the reproducibility studies between three composite biosensors showed only a slight variation of 3.3%.

Analytical characteristics and sample determination

The electrochemical process involves a chemical oxidation of hydroquinone to quinone promoted by the PPO enzyme and an electrochemical reverse reaction at -0.25 V.

The epoxy resin was used as the agglutinant agent to provide higher mechanical resistance to the electrode in basic and acidic media.

The optimum conditions established before, namely, the composite biosensor composition of 75:15:10% (m/m) of graphite: vegetable tissue: epoxy resin, 200 mV s-1 scan rate, 120 s of initial stirring time in 0.1 mol L-1 phosphate buffer solution at pH 7.0, were employed in the remaining determinations. Figure 2 shows the linear sweep voltammograms obtained with this composite biosensor using cathodic sweeping for: (A) 0.1 mol L-1 phosphate buffer solution, and for hydroquinone solutions at the following concentrations (B) 2.5×10-4, (C) 7.4×10-4, (D) 1.2×10-3, (E) 1.7×10-3, (F) 2.1×10-3, (G) 3.0×10-3 and (H) 3.9×10-3 mol L-1. The analytical curve obtained was rectilinear from 2.5×10-4 to 4.0×10-3 mol L-1 of hydroquinone (Ipc = 5.72 + 56883.6 [hydroquinone]); r= 0.9982, where Ipc is the cathodic peak current in mA and [hydroquinone] is the hydroquinone concentration in mol L-1, with a detection limit (three times the blank standard deviation/slope) of 3.8x10-5 mol L-1 hydroquinone.


Table 3 shows the relative response relationship of the composite biosensor to catechol, hydroquinone, chloromethylphenol, 4-chlorophenol, 4-nitrophenol and phenol. The electrode was more sensitive to catechol (100%) and hydroquinone (58%), followed by chloromethylphenol (24%) and 4-chlorophenol (16%), presenting no answer to 4-nitrophenol and phenol. The different responses between those phenolic compounds are due to the different substituent groups and their position. Therefore, higher signals were obtained to –OH groups in –ortho and –para position once they confer a higher reactivity characteristic to the compounds.

The proposed method was applied to the determination of hydroquinone in wastewater samples from photographic process. Table 4 presents the comparison between the values obtained using the proposed biosensor and a comparative method.25

The lifetime of this composite biosensor was at least 6 months performing 500 determinations with a decreasing of 30% of the initial activity. The initial activity can be reestablished by renewing the surface and keeping under refrigeration (4 ºC) after use.

Conclusions

The developed composite biosensor based on graphite-epoxy resin modified with PPO from sweet potato tissue was easily assembled and could be reliably used to determine hydroquinone in wastewater from photographic process samples. The robustness and sensitivity is a remarkable characteristic of this sensor. The results of the proposed method agreed at the 95% confidence level with those from a titration procedure.

Acknowledgments

Financial supports from FAPESP, CNPq and CAPES and scholarship granted by FAPESP to KOL (00/11512-2) are gratefully acknowledged.

Received: March 14, 2006

Published on the web: September 26, 2006

FAPESP helped in meeting the publication costs of this article.

  • 1. Serra, B.; Jiménez, S.; Mena, M. L.; Reviejo, A. J.; Pingarrón, J. M.; Biosens. Bioelectronic 2002, 17, 217.
  • 2. Önnerfjord, P.; Emnéus, J.; Marko-Varga, G.; Lo Gorton, G.; Biosens. Bioeletronics 1995, 10, 607.
  • 3. Fatibello-Filho, O.; Lupetti, K. O.; Vieira, I. C.; Talanta 2001, 55, 685.
  • 4. Vieira, I. C.; Lupetti, K. O.; Fatibello-Filho, O.; Anal. Lett. 2002, 35, 2221.
  • 5. Fatibello-Filho, O.; Vieira, I. C.; Quim. Nova 2002, 25, 455.
  • 6. Penner, N. A.; Nesterenko, P. N.; Analyst 2000, 125, 1249.
  • 7. Penner, N. A.; Nesterenko, P. N.; Rybalko, N. A.; J. Anal. Chem. 2001, 56, 934.
  • 8. Chen, G. N.; Liu, J. S.; Duan, J. P.; Chen, H. Q.; Talanta 2000, 53, 651.
  • 9. Afkhami, A.; Khatami, H. A.; J. Anal. Chem. 2001, 56, 429.
  • 10. Medien, H. A. A.; Zahran, A. A.; Spectrochim. Acta, Part A 2001, 57, 2505.
  • 11. MolnarPerl, I.; Tisza, S.; Koros, E.; KurinCsorgei, K. I.; Szalai, H. R. C.; J. High Resolut. Chromatogr 1995, 18, 749.
  • 12. Pryor ,W. A.; Stone, K.; Zang, L. Y.; Bermúdez, E.; Chem. Res. Toxicol. 1998, 11, 441.
  • 13. Carvalho, R. M.; Mello, C.; Kubota, L. T.; Anal. Chim. Acta 2000, 420, 109.
  • 14. Toledo, R.; Silva, M.; Khavrus, V. O.; Strizhak, P. E.; Analyst 2000, 125, 2118.
  • 15. Gao, J. Z.; Ren, J.; Yang, W.; Liu, X. H.; Yang, H.; Li, Q. Z.; Deng, H. L.; J. Elelectroanal. Chem 2002, 520, 157.
  • 16. Fatibello-Filho, O.; Vieira, I.C.; Fresenius J. Anal. Chem. 2000, 368, 338.
  • 17. Vieira, I. C.; Fatibello-Filho, O.; Talanta 2000, 52, 681.
  • 18. Fang, T.; Mc Grant, M.; Diamond, D.; Smyth, M. R.; Anal. Chim. Acta 1995, 305, 347.
  • 19. Wang, Z. P.; Zhang, Z. J.; Fu, Z. F.; Fang, L. Q.; Luo, W. F.; Chen, D. L.; Zhang, X.; Anal. Chim. Acta 2003, 494, 63.
  • 20. Lima, A. W. O.; Bora, P. S.; Ritcher, E. M.; Angnes, L.; J. Food Biochem. 2003, 27, 237.
  • 21. Li, B. X.; Zhang, Z. J.; Jin, Y.; Biosens. Biolelectron. 2002, 17, 585.
  • 22. Forzani, E. S.; Rivas, G. A.; Solis, V. M.; J. Electroanal. Chem 1997, 435, 77.
  • 23. Botre, F.; Mazzei, F.; Lanzi, M.; Lorenti, G.; Botre, C.; Anal. Chim. Acta 1991, 255, 59.
  • 24. Eggins, B. Biosensors: An introduction, Wiley & Teubner: New York, 1996.
  • 25. Berka, A.; Chem. Listy 1969, 63, 298.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      29 Jan 2007
    • Date of issue
      Dec 2006

    History

    • Received
      14 Mar 2006
    • Accepted
      26 Sept 2006
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br