Potentiometric sensor for furosemide determination in pharmaceuticals, urine, blood serum and bovine milk

Abstracts

The construction, evaluation and analytical application of a potentiometric sensor, namely, Pt|Hg|Hg2(Fur)2|Graphite, where Fur stands for 4-chloro-N-furfuryl-5-sulphamoyl-anthranilate ion, are described. This electrode has a linear dynamic range between 5.0 × 10-7-1.0 × 10-2 mol L-1 with a near-Nernstian slope of (-58.4± 0.9) mV decade-1 and a detection limit of 3.8 × 10-7 mol L-1. The potentiometric response is independent of the pH of the solution in the pH range 7.0-9.0. The electrode shows easy construction, low-cost, fast response time (within 10-20 s) and can be used for a period of 6 months without significant change in its performance characteristics. The proposed sensor displayed good selectivities over a variety of other anions (carboxylates and inorganic anions). Application of this potentiometric sensor for the furosemide determination in pharmaceuticals, urine, blood serum and commercial milk samples is reported.

4-cloro-N-furfuryl-5-sulphamoylanthranilate-sensitive electrode; potentiometry; pharmaceuticals; urine; human blood serum; bovine milk


A construção, avaliação e aplicação analítica de um sensor potenciométrico do tipo Pt|Hg|Hg2(Fur)2|Grafite, onde Fur significa íon 4-cloro-N-furfuril-5-sulfamoil-antranilato, são descritas. O eletrodo responde a Fur com sensibilidade de (-58,4 ± 0,9) mV década-1 no intervalo de 5,0 × 10-7-1,0 × 10-2 mol L-1, a pH 7,0-9,0 e com um limite de detecção de 3,8 × 10-7 mol L-1. O eletrodo é facilmente construído e de baixo custo, apresenta um rápido tempo de resposta (10-20 s) e pode ser usado por um período de 6 meses sem qualquer variação considerável nas suas características de desempenho. O eletrodo proposto mostrou boa seletividade na presença de várias substâncias, bem como na presença de alguns carboxilatos e ânions inorgânicos. O eletrodo foi aplicado com sucesso na determinação de furosemida em medicamentos, urina, soro sanguíneo e amostras comerciais de leite bovino.


ARTICLE

Potentiometric sensor for furosemide determination in pharmaceuticals, urine, blood serum and bovine milk

Alberto O. Santini; Helena R. Pezza; Rodrigo Sequinel; José L. Rufino; Leonardo Pezza* * e-mail: pezza@iq.unesp.br FAPESP helped in meeting the publication costs of this article.

Instituto de Química, Universidade Estadual Paulista, CP 355, 14801-970 Araraquara-SP, Brazil

ABSTRACT

The construction, evaluation and analytical application of a potentiometric sensor, namely, Pt|Hg|Hg2(Fur)2|Graphite, where Fur stands for 4-chloro-N-furfuryl-5-sulphamoyl-anthranilate ion, are described. This electrode has a linear dynamic range between 5.0 × 10-7-1.0 × 10-2 mol L-1 with a near-Nernstian slope of (-58.4± 0.9) mV decade-1 and a detection limit of 3.8 × 10-7 mol L-1. The potentiometric response is independent of the pH of the solution in the pH range 7.0-9.0. The electrode shows easy construction, low-cost, fast response time (within 10-20 s) and can be used for a period of 6 months without significant change in its performance characteristics. The proposed sensor displayed good selectivities over a variety of other anions (carboxylates and inorganic anions). Application of this potentiometric sensor for the furosemide determination in pharmaceuticals, urine, blood serum and commercial milk samples is reported.

Keywords: 4-cloro-N-furfuryl-5-sulphamoylanthranilate-sensitive electrode, potentiometry, pharmaceuticals, urine, human blood serum, bovine milk

RESUMO

A construção, avaliação e aplicação analítica de um sensor potenciométrico do tipo Pt|Hg|Hg2(Fur)2|Grafite, onde Fur significa íon 4-cloro-N-furfuril-5-sulfamoil-antranilato, são descritas. O eletrodo responde a Fur com sensibilidade de (-58,4 ± 0,9) mV década-1 no intervalo de 5,0 × 10-7-1,0 × 10-2 mol L-1, a pH 7,0-9,0 e com um limite de detecção de 3,8 × 10-7 mol L-1. O eletrodo é facilmente construído e de baixo custo, apresenta um rápido tempo de resposta (10-20 s) e pode ser usado por um período de 6 meses sem qualquer variação considerável nas suas características de desempenho. O eletrodo proposto mostrou boa seletividade na presença de várias substâncias, bem como na presença de alguns carboxilatos e ânions inorgânicos. O eletrodo foi aplicado com sucesso na determinação de furosemida em medicamentos, urina, soro sanguíneo e amostras comerciais de leite bovino.

Introduction

Furosemide or frusemide (HFur, 4-chloro-N-furfuryl-5-sulphamoyl-anthranilic acid) is formally a sulfonamide, an antibacterial agent (Figure 1). However, the intense and fast dieresis produced by this drug has extended its application as a powerful acidic diuretic for diverse treatments in humans and veterinary medicine. Furosemide is often classified as a loop diuretic due to its predominant action in the nephron, where the drug interferes with the tubular re-absorption of sodium on Henle's loop.1 The renal excretion of ions is not limited to sodium and chloride, but it may also influence potassium, magnesium, calcium and, to a lesser extent, hydrogen carbonate ions.2 In the clinical practice, the effects of furosemide are applied in the treatment of edema associated with pulmonary, cardiac, hepatic and renal disease, and of hypertension accompanied by fluid retention or impaired renal failure.3-7 A marked diuresis is also associated to loss of weight, a side effect that has been incorrectly abused in the sport practice to achieve acute weight losses before competition, usually where weight categories are involved. Intense and fast diuresis may also mask the ingestion of other doping agents by reducing their concentration in urine.8 For this reason, the Medical Commission of the International Olympic Committee banned the use of furosemide among other diuretics in 1986.9 Furosemide is approved for use in cattle for the treatment of physiological parturient edema of the mammary gland and associated structures.10 Consequently, harmful residue concentrations can be found in milk for human consumption.


The clinical interest and wide use of furosemide has promoted the development of several analytical methods that permit the determination of this diuretic drug in pharmaceuticals and biological samples (milk, urine, plasma and serum).11,12

These methods include spectrophotometry,13-18 variable-angle scanning fluorescence spectrometry,19 spectrofluorimetry,20,21 diffuse reflectance spectroscopy,22 potentiometry,23 voltammetry,24 chromatography,25-35 capillary electrophoresis,36-39 and flow-injection analysis.40,41 However, many of these methods are often time-consuming, technically demanding and require the use of costly, highly specialized instruments. Thus, there is an important demand for simple, low-cost, sensitive and rapid alternative methods for the determination of HFur in pharmaceuticals and biological samples.

Potentiometric methods with ion-selective electrodes (ISE's) have proved to be effective for the analysis of pharmaceuticals and biological samples, because these sensors offer the advantages of simple design, construction, and manipulation, reasonable selectivity, fast response time, applicability to colored and turbid solutions and possible interfacing with automated and computerized systems.42,43

To the best of our knowledge, there is a single report on the use of ion-selective potentiometric sensor for the determination of HFur whose application was focused in the analysis of pharmaceutical preparations.44 This electrode was constructed by incorporating the furosemide-aliquat ion pair complex into a poly (vinylchloride) (PVC) membrane. The electrode responded to HFur with sensitivity of -58.9 mV decade-1 over the range 1.00 × 10-2 to 1.50 × 10-4 mol L-1 at pH 9.6 (borate buffer). The proposed sensor exhibited a detection limit of 1.19 × 10-4 mol L-1 and had an operative life of 3 months.

The present work describes the development and application of a simple, and low-cost potentiometric 4-cloro-N-furfuryl-5-sulphamoyl-anthranilate ion sensor immobilized in a graphite matrix (Pt|Hg|Hg2(Fur)2|Graphite, where Fur stands for 4-cloro-N-furfuryl-5-sulphamoyl-anthranilate ion) for the determination of the loop diuretic furosemide in pharmaceuticals, urine, blood serum and commercial milk samples. The proposed sensor has the advantages of simplicity, versatility, fast response, fair stability and repeatability and low detection limit.

Experimental

Reagents

High purity deionized water (resistivity 18.2 MΩ cm) obtained by using a Milli-Q Plus system (Millipore Corp., Bedford, MA, USA) was used throughout. All reagents employed were of analytical grade and obtained from E. Merck (Darmstadt, Germany) except furosemide, which was supplied by Sigma (St. Louis, MO, USA).

Stock solution of the sodium salt of furosemide (NaFur) was prepared in a manner similar to a previously reported procedure for preparation of sodium naproxenate.45 This NaFur stock solution was analyzed by evaporating and drying to constant weight at 120 ºC.

Standardizations of carbonate-free sodium hydroxide, nitric acid and sodium nitrate solutions were performed as described elsewhere.46,47 Metallic mercury was purified according to a previously reported procedure.46 Hg2(Fur)2 was prepared by mixing, in aqueous solution, the corresponding nitrate with an excess of NaFur. The resulting precipitate was filtered through a sintered glass funnel, washed with deionized water until nitrate free, and then dried in a desiccator, over calcium chloride under reduced pressure, at room temperature, to constant mass. An orange powder was obtained as the final product.

Electrodes preparation and conditioning

The Hg2(Fur)2 indicator electrodes were prepared as follows: Hg2(Fur)2 (1.8 g) and metallic mercury (ca. 0.3 g) were mixed in an agate mortar and the material was crushed until a homogeneous solid was obtained. Pure powdered graphite (1 to 2 μm particle size, Aldrich, USA) was then added in different proportions (20-60% graphite, m/m) and the crushing process was continued until perfect homogenization was attained. Part of the resulting solids was transferred to a press mold, being compressed at 9 tons for about 6 min. The black pellets (external diameter 12 mm) obtained were fixed in the tip of a glass tube (external diameter 12 mm and length 8 cm) with a silicone-rubber glue ("Rhodiastic", Rhône-Poulenc, France) and allowed to dry for 48 h. Sufficient metallic mercury (ca. 0.6 g) was then introduced into each glass tube to produce a small pool on the inner pellet surface; electric contact was established through a platinum wire plunged into the mercury pool and a subsequent conductor cable. These electrodes are sealed, using the procedure previously reported by the authors.48-50

The electrodes had a geometric area of ca. 0.79 cm2. Previous studies of the authors involving the construction and caracterization of similar mercury (I)-carboxylate electrodes have demonstrated that the sensitivity and the linear concentration range for each sensor tested were practically independent of the electrode area. Similar behavior involving solid contact potassium-selective electrodes has been reported by Gyurcsányi et al.51

Elemental mercury and its compounds are potentially toxic (all forms of mercury can enter the body) and as such are considered to be hazardous wastes. For this reason, safety precautions were adopted in the manipulation and disposal of metallic mercury and mercury(I) compounds. All operations involving the use and manipulation of metallic mercury and mercury(I) compounds were conducted in a laboratory hood that has been evaluated for adequate face velocity and proper operation. During the handling of metallic mercury and mercury(I) compounds the operator has used safety glasses, appropriate protective clothing and latex or nitrile gloves. The metallic mercury and mercury(I) compounds were disposed in appropriate containers (marked appropriately) and removed periodically by a specialized disposal company. The safety precautions previously mentioned prevent or at least reduce considerably the operator's exposition to metallic mercury and mercury(I) compounds. Therefore, the proposed sensors do not offer significant risk to the operator's health and thus are recognized as safe.

When not in use, the electrode's pellets were kept immersed in a small volume of 0.010 mol L-1 NaFur solution (pH = 8.5) whose ionic strength (μ) was adjusted to 0.500 mol L-1 with a sodium nitrate solution. Before carrying out each experiment, the external surface of the aforementioned pellets were washed with deionized water and dried with absorbent paper.

Instruments

The electromotive force (emf) values were read to the nearest 0.1 mV with a Metrohm model 692 pH|ion meter (Metrohm Ltd., Herisau, Switzerland).

The reference electrode was a Metrohm Ag|AgCl double junction, model 6.0726.100. The pH of aqueous solutions was adjusted and monitored with the aid of a Metrohm pH electrode, model 6.0234.100. A thermostated titration cell (25.0 ± 0.1 ºC) was employed.

The standard procedure of the United States Pharmacopeia (USP) employed for the assay of furosemide in dosage forms utilizes a HPLC method.25

Chromatographic analysis were carried out on a Shimadzu model SPD-10A liquid chromatograph (Shimadzu Seisakusho, Kyoto, Japan), equipped with a LC-10 AS pump (Shimadzu), variable UV-Visible detector (model SR-10A, Shimadzu) set at 254 nm, gradient control (Waters, model 680; Waters Chromatography Div., Milford, MA, USA) and a "Rheodyne" 20 μL injector (Rheodyne, Inc., Berkeley, CA, USA). A stainless steel "Microsorb LC-18" analytical column (250 mm × 4.6 mm i.d., Varian, Walnut Creek, CA, USA) with 5 μm particle size packing material was used. Before injection the samples were filtered through a Millex unit (Millex-HV, 0.45 μm, Millipore). Chromatograms were recorded and the areas were measured with an integrator (Shimadzu, model C-R6A Chromatopac).

Volume measurements (± 0.001 mL) were performed using a Metrohm model 665 automatic burette.

All experiments were performed in a thermostated room, maintained at 25 ± 1 ºC.

Potentiometric cell

The following cell was used,

where Fur stands for 4-cloro-N-furfuryl-5-sulphamoyl-anthranilate ion and × was in the range 1.00 × 10-2-1.00 × 10-7 mol L-1. The ionic strength (μ) of the cell compartments was kept constant at 0.500 mol L-1. The outer compartment of the reference electrode was refilled periodically with fresh NaNO3 solution.

The performance of the Hg2(Fur)2 electrodes were assessed by measuring the emf of the aforementioned cell for 1.00 × 10-2 to 1.00 × 10-7 mol L-1 NaFur solutions. These solutions were freshly prepared by serial dilution of a 2.00 × 10-2 mol L-1 stock standard solution with deionized water, at constant pH (8.5 ± 0.1) and μ adjusted to 0.500 mol L-1 with NaNO3. The emf readings were obtained for solutions under stirring and recorded when they became stable. A typical calibration plot of an electrode with composition of 50% (graphite, m/m) is shown in Figure 2.


Determination of furosemide in pharmaceuticals

The analyzed products were purchased locally or directly from the manufacturers and all were tested prior to the listed expiration date. Six pharmaceutical formulations (tablets) containing furosemide (HFur) and other components were analyzed with the Fur-sensitive electrode.

Representative samples of finely ground tablets containing a quantity equivalent to about 25 mg of HFur was accurately weighed and placed in a glass vessel; 50 mL of 10-2 mol L-1 NaOH was added and magnetically stirred for 10 min. The resulting mixture was filtered and its ionic strength was adjusted to 0.500 mol L-1 with NaNO3 and the pH to 8.5 ± 0.1 with 10-2 mol L-1 HNO3 or 10-2 mol L-1 NaOH before volume completion. The resulting solution was quantitatively transferred to a 100 mL volumetric flask using deionized water (pH = 8.5 ± 0.1) for rinsing and volume completion. An aliquot of 20 mL is employed for analysis with the Fur-sensitive electrode.

Determination of furosemide in human urine samples

Drug-free human urine used in this study was obtained from healthy volunteers. The urine was centrifugated at 4000 × g, filtered through filter paper and kept frozen (-20 ºC) pending analysis.

The extraction procedure of HFur used here was a modification of the technique reported by Barroso et al.24 To a 5 mL of spiked human urine with different quantities of HFur, 5 mL of 1.0 mol L-1 H3PO4 was added and mixed for 60 s. The samples were then extracted with 10 mL of ethyl acetate, vortex mixed for 10 min and centrifuged at 10.000 rpm ( 8500 × g ) for 3 min. The organic phase was then transferred to an appropriate flask and evaporated in a rotary evaporator under reduced pressure to dryness. The residue was dissolved in 10 mL of 0.500 mol L-1 NaNO3 (pH = 8.5 ± 0.1) and an aliquot of 8 mL of the solution was employed for analysis with the Fur-sensitive electrode using the standard additions method (multiple addition method).

Determination of furosemide in human blood serum samples

Drug-free human serum used in this study was obtained from healthy volunteers. The serum was kept in a freezer at -20 ºC until analysis.

The extraction procedure of HFur used here was a modification of the technique reported by Abou-Auda et al.52 To a 5 mL of spiked human serum with different quantities of HFur, 5 mL of 2.0 mol L-1 phosphoric acid was added and mixed for 75 s. The samples were then extracted with 15 mL of ethyl acetate, vortex mixed for 5 min and centrifuged at 10.000 rpm (8500 × g) for 3 min. The organic phase was then transferred to an appropriate flask and evaporated in a rotary evaporator under reduced pressure to dryness. The residue was dissolved in 10 mL of 0.500 mol L-1 NaNO3 (pH = 8.5 ± 0.1) and an aliquot of 8 mL of the solution was employed for analysis with the Fur-sensitive electrode using the standard additions method (multiple addition method).

Determination of furosemide in commercial milk samples

Commercial milk samples were purchased in a local supermarket. The milk samples were refrigerated at 4 ºC if not analyzed immediately.

The extraction procedure of HFur adopted was a modification of the procedure reported by Guzmán et al.41 To 25.0 mL of skimmed milk samples spiked with 150 ng mL-1 of HFur, 5 mL of 2.0 mol L-1 phosphoric acid was added and mixed for 90 s. The samples were then extracted with 50 mL of ethyl acetate, vortex mixed for 3 min and centrifuged at 10.000 rpm (8500 × g) for 3 min. The organic phase was then transferred to an appropriate flask and evaporated in a rotary evaporator under reduced pressure to dryness. The residue was dissolved in 10 mL of 0.500 mol L-1 NaNO3 (pH = 8.5 ± 0.1) and an aliquot of 8 mL of the solution was employed for analysis with the Fur-sensitive electrode using the standard additions method (multiple addition method).

Results and Discussion

Effect of the graphite content in the composite on the electrode response

Different pellet compositions (20-60% graphite, m/m) were tried for use as sensors in the composed electrodes (Table 1). Experiments carried out as described under "Potentiometric Cell" led to the following linear relationship between the measured emf (E, in mV) and Fur ion concentration:

where Eº is the formal cell potential and S represents the Nernst coefficient (59.16 mV decade-1, at 25 ºC, for monovalent ions).

It was found that the Fur-sensitive electrode with composition of 50% (graphite, m/m) responds perfectly to the Fur with a S value of 58.4 mV decade-1 and the widest linear concentration range (5.0 × 10-7-1.0 × 10-2 mol L-1). For electrodes with compositions of 40 and 60% (graphite, m/m) Nernst coefficients of 55.1 and 56.8 mV decade-1 were observed but for a shorter linear concentration range (2.5 × 10-6-1.0 × 10-2 mol L-1). The electrode with composition of 20% (graphite, m/m) presents the lowest slope (45.2 mV decade-1) for a linear concentration range (1.0 × 10-5-1.0 × 10-2 mol L-1).

Potentiometric parameters and other features associated with the Hg2(Fur)2 electrode with 50% graphite (m/m) are given in Table 2. The calibration equation and the slope value (Table 2) show that the mentioned electrode provides a near-Nernstian response to the Fur ion in the range of 1.0 × 10-2 to 5.0 × 10-7 mol L-1. The detection limit, as determined from the intersection of the two extrapolated segments of the calibration graph (Figure 1), was 3.8 × 10-7 mol L-1.53 The sensor response displayed good stability and repeatability over the tests; the last mentioned feature is illustrated by the standard deviation values shown in Table 2.

For the reasons previously mentioned the Fur-sensitive electrode with composition of 50% (graphite, m/m) was chosen to perform all of the following studies.

Response time and lifetime of the electrode

For analytical applications, the response time and the lifetime of a sensor are of critical importance. According to IUPAC recommendations, the response time may be defined as the time between the addition of analyte to the sample solution and the time when a limiting potential has been reached.53

The response time of the electrode was tested by measuring the time required to achieve a steady state potential (within ± 0.2 mV min-1), for 1.0 × 10-2 to 1.0 × 10-6 mol L-1 NaFur solutions at pH 8.5.53 The electrode yielded steady potentials within 10 to 15 s at high concentrations (> 1.0 × 10-3 mol L-1) and about 20 s at concentrations near the detection limit. The experimental results show that the lifetime of the electrode was about 6 months, with a total of 620 determinations. During this period, the sensor was in daily use and was stored in 0.010 mol L-1 NaFur solution (pH = 8.5 and μ = 0.500 mol L-1 adjusted with NaNO3) when not in use. No significant change in working concentration range, slope and response time was observed during this period.

pH effect

The pH dependence of the potentials of the proposed electrode was tested over the pH range 4.0-10.0 for 1.0 × 10-3 and 1.0 × 10-4 mol L-1 Fur ion concentrations. The resulting solutions' pH(s) were adjusted with diluted HNO3 or NaOH solutions. As it can be seen in Figure 3, the potential response remains almost constant over the pH range 7.0-9.0 which can be taken as the working pH range of the electrode.


However, for pH values below 7.0, we observed the progressive precipitation of the free furosemide acid (pKa=3.9),54 and protonation of the secondary amino groups of the drug to form other species which are not detected by the electrode. For pH > 9.0, the hydroxide ion interferes with the electrode's response. In high pH media, probably OH- competes with Fur ion in the electrode process and alters the potentiometric response of the proposed sensor.

Electrode selectivity

The potentiometric selectivity coefficient, which reflects the relative response of a sensor for the primary ion over other ions, present in solution, is perhaps the most important characteristic of any ion sensitive sensor. The potentiometric selectivity coefficients for the Fur-sensitive electrode (KFur,M) were determined, for a number of anions (M), by the matched potential method (MPM).55-57 In this method, the selectivity coefficient is defined by the ratio of the activity of the primary ion relative to an interfering ion, when they generate identical potentials in the same reference solution. In the MPM method, both monovalent and divalent ions are treated in the same manner and the valence of the ions does not influence the selectivity coefficient. Furthermore, the MPM can be used with no regard to the electrode slopes being Nernstian or even linear.58

The MPM-selectivity coefficients (KFur,M) were determined under the following conditions: Initial reference solution (pH = 8.5) contains 0.500 mol L-1 NaNO3 as a supporting electrolyte and 1.0 × 10-5 mol L-1 of the primary ion (4-cloro-N-furfuryl-5-sulphamoyl-anthranilate). The selectivity coefficients were calculated from the concentration of the interfering ion (M), which induced the same amount of the potential change (Δemf = 15.0 mV) as that induced by increasing the concentration of primary ion. The resulting values of KFur,M are presented in Table 3.

The results comprised in the aforementioned Table 3 show that the selectivity of the Hg2(Fur)2 electrode towards all tested organic acid anions is good. No interference was noted for most of the common excipients used in commercial formulations (tablets) such as glucose, lactose, starch, talc, magnesium stearate, ethylcellulose, microcrystalline cellulose, croscarmellose sodium and silicon dioxide. Drugs commonly used in veterinary such as oxytetracycline (antibiotic) and methylthiouracil (thyreostat drug) do not interfere in the determination of furosemide in commercial milk samples.

Sulfate and borate has a very low selectivity coefficient (Table 3); no interference at all is caused by nitrate or perchlorate and they can therefore be used as background electrolytes or ionic strength adjusters for 4-cloro-N-furfuryl-5-sulphamoyl-anthranilate solutions before performing potentiometric measurements.

Some interference by chloride ion might be expected as shown in Table 3. In the dosage forms analyzed in this work (tablets), chloride ion is seldom found and hence the proposed sensor can be used for direct determination of HFur in these samples without previous extraction procedures. Concerning the biological samples(milk, urine and human serum) analyzed by the potentiometric sensor, it should be noted that analytical procedure adopted in this work is based on ethyl acetate extraction of HFur from acidified biological matrices followed by its reversion to the aqueous phase (0.500 mol L-1 NaNO3(aq); pH = 8.5 ± 0.1) as 4-cloro-N-furfuryl-5-sulphamoyl-anthranilate ion. The chloride content found in the last mentioned aqueous phase (which originates from the analyses of milk, urine or human blood serum) was always < 1 μg L-1, as analyzed by the mercury thiocyanate method.59 Therefore, the working procedure removes chloride interference.

Analytical application

In order to check the usefulness of the proposed sensor for resolving real samples, we addressed the determination of furosemide in pharmaceuticals and in complex samples as urine, blood serum and bovine milk using a standard additions method (multiple addition method).60

Table 4 shows statistical analysis of the results obtained by using the presently proposed Fur-sensitive electrode and the official method of USP25 for HFur determination in pharmaceuticals. In all cases, the calculated F and t values did not exceed the theoretical values, indicating that there is no significant difference between either methods in concerning accuracy (t-test) and precision (F-test).

In humans, HFur is mainly metabolized into acyl glucuronides.21,61,62 In serum and urine, furosemide is the predominant species.21,61 Taking into account the role of the 4-chloro-N-furfuryl-5-sulphamoyl-anthranilate ion (Fur) in the electrode response, along with the low concentration of acyl glucuronides in serum and urine,21,61 no interference from these metabolites is expected in the determination of HFur in urine and human blood serum by the proposed sensor.

Experiments were performed to determine the feasibility of using the Fur- sensitive electrode for determination of HFur in urine and human serum samples. Urine and serum samples previously spiked with HFur at four different concentrations were treated and analyzed as described under "Determination of Furosemide in Human Urine Samples" and "Determination of Furosemide in Human Blood Serum Samples", respectively. The amounts of HFur added to each biological fluid (urine and human blood serum) were those usually found after administration of a pharmaceutical dose.21,28,30-32,63,64 Commercial skimmed milk samples were spiked with HFur at a concentration level of 150 ng mL-1 and treated following the procedure described under "Determination of Furosemide in Commercial Milk Samples". This aforementioned concentration is the mean concentration of diurethic that can be detected in the next 8 h after administration of the usual 500 mg dosage of drug.29,41

The recovery values of HFur in urine, serum and milk samples found by application of the proposed potentiometric sensor and the comparative chromatographic methods29,52 are given in Tables 5, 6 and 7, respectively. Satisfactory recovery values for HFur in urine, blood serum and commercial milk samples were obtained indicating that the extraction procedures of the proposed method provided adequate selectivity and sensitivity to process the selected samples. Therefore, the proposed sensor can be used for the determination of HFur in pharmaceuticals, urine, blood serum and commercial milk samples.

Conclusions

The results obtained in the present work demonstrate that the potentiometric method employing a Hg2 (Fur)2 electrode immobilized in a graphite matrix may provide an attractive alternative for the determination of furosemide.

The proposed potentiometric sensor is easy to prepare, exhibits long lifetime, shows high sensitivity and wide dynamic range. Good selectivity, very low detection limit, rapid response and low-cost of fabrication make this electrode suitable for analysis of HFur in pharmaceuticals, urine, blood serum and commercial milk samples. The electrode developed in this laboratory is superior (especially concerning lifetime, sensitivity and versatility) as compared with the furosemide selective electrode described in the literature.44

Acknowledgments

We would like to thank FAPESP, CNPq, CAPES and FUNDUNESP Foundations (Brazil), for financial support.

Received: May 13, 2008

Web Release Date: November 12, 2008

  • 1. Delgado, J. N.; Remers, W. A.; Textbook of Organic Medicinal and Pharmaceutical Chemistry, 9th ed., J. B. Lippincot: Philadelphia, 1991.
  • 2. Foye, W. O.; Lemke, T. L.; Williams, D. A.; Principles of Medicinal Chemistry, 4th ed., Williams and Wilkins: Philadelphia, 1995.
  • 3. Shinto, R. A.; Light, R. W.; Am. J. Med. 1990, 88, 230.
  • 4. Modell, W.; Drugs of Choice, C.V. Mosby: Toronto, 1984.
  • 5. Reynolds, E. F.; Martindale: The Extra Pharmacopoeia, 28th ed., The Pharmaceutical Press: London, 1982.
  • 6. Leary, W. P.; Asmal, A. C.; Curr. Ther. Res. Clin. Exp 1980, 28, 549.
  • 7. Kristensen, B. O.; Show, J.; Lancet 1980, 2, 699.
  • 8. Ventura, R.; Segura, J.; J. Chromatogr., B 1996, 687, 127.
  • 9. International Olympic Committee; Medical Commission, International Olympic Charter against Doping in Sport, IOC: Lausanne, 1990.
  • 10. Code of Federal Regulations, Food and Drugs, Parts 522.1010, c(3); U.S. GPO: Washington, 1991.
  • 11. Ruiz-Angel, M. J.; Berthod, A.; Carda-Broch, S.; Álvarez-Coque, M. C. G.; Separation & Purification Reviews 2006, 35, 39.
  • 12. Shaikh, B. In Veterinary Drug Residues-Diuretic Drugs Used in Food Producing Animals; Moats, W. A.; Medina, M. B., eds.; ACS Symposium Series 636, American Chemical Society: Washington, 1996.
  • 13. British Pharmacopoeia, HM Stationary Office: London, 1998.
  • 14. United States Pharmacopeia, USP 23: Rockville, 1995.
  • 15. Issopoulos, P. B.; Fresenius Z. Anal. Chem 1989, 334, 554.
  • 16. Zivanovic, L.; Agatonovic, S.; Radulovic, D.; Mikrochim. Acta 1990, 1, 49.
  • 17. Sastry, C. S. P.; Prasad, T. N. V.; Sastry, B. S.; Rao, E. V.; Analyst 1988, 113, 225.
  • 18. Ferraro, M.; Castellano, P.; Kaufman, T.; J. Pharm. Biomed. Anal 2001, 26, 443.
  • 19. García-Sánchez, F.; Fernández-Gutiérrez, A.; Cruces-Blanco, C.; Anal. Chim. Acta 1995, 306, 313.
  • 20. Luis, M. L.; Fraga, J. M. G.; Jiménez, A. I.; Jiménez, F.; Hernández, O.; Arias, J. J.; Talanta 2004, 62, 307.
  • 21. Ioannou, P. C.; Rusakova, N. V.; Andrikopoulou, D. A.; Glynoy, K. M.; Tzompanaki, G. M.; Analyst 1998, 123, 2839.
  • 22. Gotardo, M. A.; Gigante, A. C.; Pezza, L.; Pezza, H. R.; Talanta 2004, 64, 361.
  • 23. Nicolic, K. I.; Medenica, M.; Acta Pharm. Jugosl 1990, 40, 521.
  • 24. Barroso, M. B.; Alonso, R. M.; Jiménez, R. M.; Anal. Chim. Acta 1995, 305, 332.
  • 25. United States Pharmacopeia, USP 24: Rockville, 2000.
  • 26. Abdel-Hamid, M.; Il Farmaco 2000, 55, 448.
  • 27. El-Saharty, Y. S.; J. Pharm. Biomed. Anal. 2003, 33, 699.
  • 28. Baranowskaa, I.; Markowski, P.; Baranowski, J.; Anal. Chim. Acta 2006, 570, 46.
  • 29. Shaikh, B.; J. Agric. Food Chem. 1995, 43, 2117.
  • 30. Carda-Broch, S.; Esteve-Romero, J.; Ruiz-Angel, M. J.; Alvarez-Coque, M. C. G.; Analyst 2002, 127, 29.
  • 31. Barroso, M. B.; Alonso, R. M.; Jiménez, R. M.; Ortiz, E.; J. Chromatogr., B 1996, 675, 303.
  • 32. Nava-Ocampo, A. A.; Velázquez-Armenta, E. Y.; Reyes-Pérez, H.; Ramirez-Lopez, E.; Ponce-Monter, H.; J. Chromatogr., B 1999, 730, 49.
  • 33. Tsai, F. Y.; Lui, L. F.; Chang, B.; J. Pharm. Biomed. Anal. 1991, 9, 1069.
  • 34. Saugy, M.; Meuwly, P.; Munafo, A.; Rivier, L.; J. Chromatogr 1991, 564, 567.
  • 35. Hagedorn, H. W.; Schulz, R.; J. Anal. Toxicol 1992, 16, 194.
  • 36. Riekkola, M. L.; Jumppanen, J. H.; J. Chromatogr., A 1996, 735, 151.
  • 37. Luis, M. L.; Corujedo, S.; Blanco, D.; Fraga, J. M. G.; Jiménez, A. I.; Jiménez, F.; Arias, J. J.; Talanta 2002, 57, 223.
  • 38. Soetebeer, U. B.; Shierenberg, M. O.; Moller, J. G.; Shulz, H.; Grunefeld, G.; Andresen, P.; Blaschke, G.; J. Chromatogr., A 2000, 895, 147.
  • 39. Caslavska, J.; Thormann, W.; J. Chromatogr., B 2002, 770, 207.
  • 40. Rao, Y.; Zhang, X.; Luo, G.; Baeyens, W. R. G.; Anal. Chim. Acta 1999, 396, 273.
  • 41. Guzmán, A.; Aguí, L.; Pedrero, M.; Yánez-Sedeño, P.; Pingarrón, J. M.; J. Pharm. Biomed. Anal 2003, 33, 923.
  • 42. Stefan, R. I.; Baiulescu, G. E.; Aboul-Enein, H. Y.; Crit. Rev. Anal. Chem. 1997, 27, 307.
  • 43. Lewenstam, A.; Maj-Zurawska, M.; Hulanicki, A.; Electroanalysis 1991, 3, 727.
  • 44. Dias, I. L. T.; Neto, G. O.; Vendramini, D. C.; Sommer, C.; Martins, J. L. S.; Kubota, L. T.; Anal. Lett 2004, 37, 35.
  • 45. Valsami, G. N.; Macheras, P. E.; Koupparis, M. A.; Analyst 1989, 114, 387.
  • 46. Pezza, L.; Molina, M.; Moraes, M.; Melios, C. B.; Tognolli, J. O.; Talanta 1996, 43, 1689.
  • 47. Brito, M. S. L.; Angnes, L.; Brett, C. M. A.; Gutz, I. G. R.; J. Electroanal. Chem 1999, 468, 150.
  • 48. Peres, A. M.; Moraes, M.; Pezza, L.; Pezza, H. R.; Melios, C. B.; Microchem. J 1998, 60, 184.
  • 49. Ferreira, V. J. F.; Cavalheiro, A. C. V.; Fagnani, E.; Moraes, M.; Pezza, L.; Pezza, H. R.; Melios, C. B.; Anal. Sci 1999, 15, 249.
  • 50. Pezza, L.; Santini, A. O.; Pezza, H. R.; Melios, C. B.; Ferreira, V. J. F.; Nasser, A. L. M.; Anal. Chim. Acta 2001, 433, 281.
  • 51. Gyurcsányi, R. E.; Nybäck, Ann-Sofi; Tóth, K.; Nagy, G.; Ivaska, A.; Analyst 1998, 123, 1339.
  • 52. Abou-Auda, H.; Al-Yamani, M. J.; Morad, A. M.; Bawazir, S. A.; Khan, S. Z.; Al-Khamis, K. I.; J. Chromatogr., B 1998, 710, 121.
  • 53. Buck, R. P.; Lindner, E.; Pure Appl. Chem 1994, 66, 2527.
  • 54. Benet, L. Z.; Massound, M.; Gambetogoho J. G.; Pharmacokinetic Basis of Drug Treatment, Raven Press: New York, 1984.
  • 55. Gadzekpo, V. P. Y.; Christian, G. D.; Anal. Chim. Acta 1984, 164, 279.
  • 56. Christian, G. D.; Analyst 1994, 119, 2309.
  • 57. Umezawa, Y.; Umezawa, K.; Sato, H.; Pure Appl. Chem 1995, 67, 507.
  • 58. Bakker, E.; Bühlmann, P.; Pretsch, E.; Chem. Rev 1997, 97, 3083.
  • 59. Williams, W. J.; Handbook of Anion Determination, Butterworths: London, 1979.
  • 60. Li, G.; Polk, B. J.; Meazell, L. A.; Hatchett, D. W.; J. Chem. Educ 2000, 77, 1049.
  • 61. Gilman, A. G.; Goodman, L. S.; Gilman, A.; The Pharmacological Basis of Therapeutics, 8th ed., Macmillan: New York, 1991.
  • 62. McEvoy, G. K.; McQuarrie, G. M.; DiPietro, H. J.; American Hospital Formulary Service Monograph, Furosemide USP (40:28), American Hospital Formulary Service; American Society of Hospital Pharmacists: Bethesda, 1987.
  • 63. Moffat, A. C.; Clarke's Isolation and Identification of Drugs in Pharmaceutics, Body Fluids, and Post-Mortem Material, 2nd ed., The Pharmaceutical Press: London, 1986.
  • 64. Winek, C. L.; Wahba, W. W.; Winek Jr., C.L.; Balzer, T. W.; Foren. Sci. Intern. 2001, 122, 107.
  • *
    e-mail:
    FAPESP helped in meeting the publication costs of this article.
  • Publication Dates

    • Publication in this collection
      16 Jan 2009
    • Date of issue
      2009

    History

    • Received
      13 May 2008
    • Accepted
      12 Nov 2008
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br