Validation of a method using solid phase extraction and liquid chromatography for the determination of pesticide residues in groundwaters

Abstracts

A method is described for the determination of the pesticides carbofuran, clomazone, 2,4-D and tebuconazole in groundwaters. The method involves solid phase extraction (SPE) with C18 cartridges and quantification by high performance liquid chromatography with diode array detector (HPLC-DAD). After the optimization of the extraction and separation parameters, the method was validated by evaluating the analytical curve, linearity, limits of detection and quantification, precision and accuracy (recovery). The method presents an average recovery of 87.9% and 96.9%, in repeatability and intermediate precision conditions, respectively, with adequate precision (RSD from 0.8 to 20.7%), for all compounds. The method will be applied to determine pesticides in groundwater samples with limit of quantification of 0.2 μg L-1.

pesticides; groundwater; HPLC-DAD; SPE; validation


Neste trabalho, um método para determinação dos agrotóxicos carbofurano, clomazona, 2,4-D e tebuconazol em águas subterrâneas é descrito. O método utiliza a Extração em Fase Sólida (EFS) com cartuchos de C18 e quantificação por Cromatografia Líquida de Alta eficiência com Detector de Arranjo de Diodos (CLAE-DAD). Após a otimização dos parâmetros de extração e separação dos compostos, o método foi validado avaliando-se curva analítica, linearidade, limites de detecção e quantificação, precisão e exatidão (recuperação). O método apresentou recuperações médias de 87,9% e 96,9%, para a repetibilidade e precisão intermediária, respectivamente, com RSD de 0,8 a 20,7% para todos os compostos. O método será empregado na determinação de agrotóxicos em águas subterrâneas com um limite de quantificação de 0,2 μg L-1.


ARTICLE

Validation of a method using solid phase extraction and liquid chromatography for the determination of pesticide residues in groundwaters

Sergiane S. Caldas; Adriana Demoliner; Ednei G. Primel* * e-mail: dqmednei@furg.br

Escola de Química e Alimentos, Programa de Pós-Graduação em Química Tecnológica e Ambiental, Universidade Federal do Rio Grande - FURG, Av Itália, km 8 s/n, 96201-900 Rio Grande-RS, Brazil

ABSTRACT

A method is described for the determination of the pesticides carbofuran, clomazone, 2,4-D and tebuconazole in groundwaters. The method involves solid phase extraction (SPE) with C18 cartridges and quantification by high performance liquid chromatography with diode array detector (HPLC-DAD). After the optimization of the extraction and separation parameters, the method was validated by evaluating the analytical curve, linearity, limits of detection and quantification, precision and accuracy (recovery). The method presents an average recovery of 87.9% and 96.9%, in repeatability and intermediate precision conditions, respectively, with adequate precision (RSD from 0.8 to 20.7%), for all compounds. The method will be applied to determine pesticides in groundwater samples with limit of quantification of 0.2 μg L-1.

Keywords: pesticides, groundwater, HPLC-DAD, SPE, validation

RESUMO

Neste trabalho, um método para determinação dos agrotóxicos carbofurano, clomazona, 2,4-D e tebuconazol em águas subterrâneas é descrito. O método utiliza a Extração em Fase Sólida (EFS) com cartuchos de C18 e quantificação por Cromatografia Líquida de Alta eficiência com Detector de Arranjo de Diodos (CLAE-DAD). Após a otimização dos parâmetros de extração e separação dos compostos, o método foi validado avaliando-se curva analítica, linearidade, limites de detecção e quantificação, precisão e exatidão (recuperação). O método apresentou recuperações médias de 87,9% e 96,9%, para a repetibilidade e precisão intermediária, respectivamente, com RSD de 0,8 a 20,7% para todos os compostos. O método será empregado na determinação de agrotóxicos em águas subterrâneas com um limite de quantificação de 0,2 μg L-1.

Introduction

A group of artificially synthesized substances, called pesticides, has been used in agriculture to control pests and to increase production.1 These substances protect the agricultural crops, but overuse and incorrect use can pose risks to human health and the environment.2,3

Brazil ranks eighth in pesticide use in the world,4 and Rio Grande do Sul State is responsible for 10.4% of pesticide consumption in Brazil.5 Many researchers have analyzed the residue of pesticides in water samples.6-8 The pesticides 2,4-D, clomazone, tebuconazole and carbofuran are often used in the regional agriculture of Rio Grande do Sul.

The increase in the amount and variety of products applied to agriculture makes it necessary to monitor residues in the environment;9 therefore, the analysis of pesticides has received increasing attention in the last few decades.

Groundwater represents an important source of drinking water supply in many countries, and the use of pesticides in agricultural areas may lead to its contamination by drift, runoff, drainage and leaching.10 The groundwater contamination by pesticides has been documented in many papers around the world.10-14

For many years, the pesticide analysis in environment samples has attracted the attention due to the wide use of such compounds, as well as its environmental impact. The European Union establishes rigid limits for pesticides in water destined to human consumption, establishing 0.1 μg L-1 for individual pesticides and 0.5 μg L-1 for the sum of all pesticides. In the case of surface water that will be treated for human consumption, the maximum limits allowed are 1 μg L-1 for individual pesticides and 5 μg L-1 for the total pesticides.15 In Brazil, Law No. 518, issued by the Ministry of Health, establishes the maximum values for drinking waters and Resolution No. 357, issued by the National Council of Environment, establishes the concentration accepted for each individual pesticide in water bodies.16,17 The Brazilian legislation does not include the pesticides carbofuran, clomazone and tebuconazole, and for 2,4-D the maximum value accepted is 30 μg L-1 for drinking waters and 4 μg L-1 accepted for water bodies.

Due to the low detection levels required by regulatory bodies and the complex nature of the matrices in which the target compounds are present, efficient sample preparation and trace-level detection and identification are important aspects of analytical methods.18 However, in order to reach the low concentration levels permitted in potable water, a preliminary concentration step is required before the chromatographic analysis.14

Pesticide in water samples are usually enriched by liquid-liquid extraction (LLE)19,20 or solid phase extraction (SPE).10,21 The conventional liquid-liquid extraction often needs large amounts of toxic solvent and a time-consuming procedure,22 while SPE is simple, fast and consumes very small volumes of high purity solvents.20 Nowadays, SPE is the most appropriate method for the enrichment of pesticides in aqueous matrices.18,23

Gas chromatography (GC) and high performance liquid chromatography (HPLC) coupled with various detection systems are the most powerful tools for the analysis of pesticides in water.14,24-28 However, many classes of pesticide that have polar characteristics, low volatility or thermal instability, cannot be analyzed directly by GC and require special conditions such as derivatisation procedures. Liquid chromatography (LC) is the preferred approach for these polar and thermally labile pesticides, with a conventional UV detection or diode array detection (DAD).29 The pesticides carbofuran, 2,4-D, clomazone and tebuconazole (Figure 1) have been studied by many authors by HPLC.28,30,31


The use of liquid chromatography combined with mass spectrometry (LC-MS-MS) has been proposed for determining some of these pesticides.32-34 These methods are much more specific and sensitive analytical techniques, but they are not affordable in most research laboratories because of the high cost of the equipment.

To guarantee that an analytical procedure gives reliable, exact and interpretable information about a sample, it must be validated. Therefore, diverse analytical methods for determination of pesticide residues have been developed and validated;20,33,35 they describe definitions, procedures, parameters and strategies for validation. The analytical parameters normally found for validation of separation methods are: selectivity, linearity and range, precision, accuracy, limit of detection, limit of quantification and robustness.36

This paper reports a simple, relatively fast, and efficient SPE and HPLC-DAD method which was developed for the determination of carbofuran, clomazone, 2,4-D and tebuconazole in groundwaters. To obtain efficient pre-concentration with good precision and recovery, a C18 solid phase extraction system was applied. The method was validated and the parameters involved in the validation were calibration, linearity and range, limit of detection (LOD) and quantification (LOQ), precision (repeatability and intermediate precision), and accuracy (recovery). The proposed method will be used for the determination of pesticide residues in groundwaters, in an agricultural area in Rio Grande, Rio Grande do Sul State, Brazil.

Experimental

Chemicals

Carbofuran, clomazone, tebuconazole and 2,4-D analytical standard (purity > 99%) were supplied by Sigma Aldrich (São Paulo, Brazil). HPLC grade methanol and acetonitrile were supplied by Mallinckrodt (Phillisburg, NJ, USA). Phosphoric acid (85%) of analytical grade, by Merck (Darmstadt, Germany). Water was purified with a Direct-Q UV3® (resistivity 18.2 MΩ cm, Millipore, USA) water purification system (Millipore, Bedford, MA, USA). The SPE extraction cartridges were octadecylsilane (Strata C18-E) with 200 mg of phase in cartridges and another polymeric Strata-X (copolymer of styrene-divinylbenzene with modified surface) with 200 mg of phase, both from Phenomenex (Torrance, CA, USA).

Equipment

To determine the optimum conditions of separation, two columns were tested, Spherisorb ODS2 5 μm 80 Å (150 × 4.6 mm), Waters, and Hyperclone BDS C18 5 μm 130 Å (250 × 4.6 mm), Phenomenex.

Different mobile phases comprising several combinations of methanol, acetonitrile and purified water were tested to provide sufficient resolution. The pH of the mobile phase was adjusted with the use of a Hanna pH 21, pHmeter.

Separation was performed using an HPLC apparatus consisting of a column Hyperclone BDS C18 5 μm 130 Å (250 × 4.6 mm I.D), from Phenomenex, a Waters 600 pump model, associated with a Waters 2996 Photodiode Array Detector, Rheodyne 20 μL loop injector, connected to an Empower PDA software for data acquisition.

The analytical column was conditioned by passing the mobile phase through it for 1 h at a flow-rate of 0.8 mL min-1 and the analytical column was operated at ambient temperature. The flow-rate was set at 0.8 mL min-1 and quantification was carried out with DAD detection at 220.3 nm.

The identification of the four substances in the samples was accomplished on the basis of their retention times and by comparison between the DAD spectrum of the compounds in the standard solutions and the DAD spectrum of the detected peak in the sample.

Preparation of solutions and mobile phases

Individual pesticide stock solutions containing 1000 mg L-1 of the target compounds were prepared in methanol and stored at -18 ºC. Intermediate working standards mixtures in methanol, containing 100 mg L-1 for each pesticide were prepared and used to prepare the working standard solutions containing 10, 7.5, 5.0, 2.5, 1.25, 0.62, 0.31, 0.15, 0.07 and 0.05 mg L-1, that was used for spiking samples and to prepare the analytical curves. Working standard solutions were prepared monthly, while the dilutions used for the analytical curves were prepared daily.

The mobile phase consisted of acetonitrile and purified water adjusted to pH 3.0 with phosphoric acid. The mobile phases were degassed for 30 min in an ultrasonic bath before use.

Solid phase extraction procedure

The optimization of the extraction procedure included the type of the adsorbent. Two different sorbents were tested: octadecylsilane (Strata C18-E) with 200 mg of phase in cartridges and an average particle size of 55 μm and another polymeric Strata-X (copolymer of styrene-divinylbenzene with modified surface) containing 200 mg of phase in cartridges of 3 mL and an average particle size of 33 μm. All the experiments were repeated at least in triplicate.

A 250 mL volume of tap water sample was fortified by the addition of an established volume of solution with 10.0 mg L-1 of each pesticide, resulting in four levels of fortification 0.2, 1.25, 5.0 and 10.0 μg L-1. Before sample application, the SPE column was conditioned by 3 mL of methanol, 3 mL of purified water and 3 mL of purified water at pH 3.0, acidified with phosphoric acid 1:1 (v/v). After the conditioning step, aliquots of 250 mL of water samples, acidified at pH 3.0 with phosphoric acid (to increase the pesticide retention in the SPE cartridge) were loaded through the cartridges with a flow rate of 6 mL min-1. After that, the analytes were eluted with 1 mL (2 × 500 μL) of methanol, the volume was adjusted in 1 mL and injected into the chromatographic system.

Method validation

A validation process was carried out to confirm the suitability of the method for the intended use. The concepts of validation continue evolving and are constantly under consideration by regulating agencies.37 Once the best conditions for the analysis of the pesticides were defined, the validation of the method was carried out according to parameters described as follows.

Analytical curves and linearity

Linearity corresponds to the capacity of the method to supply results directly proportional to the concentration of the substance under investigation, within one determined application range.38,39 Range is the interval between the upper and the lower levels of analyte that have been demonstrated to be determined with precision, accuracy and linearity using the method as written.39 The linearity of a method can be observed by the equation of the linear regression (y = ax + b). The results should not show a significant deviation from linearity, which is taken to mean that the correlation coefficient r > 0.99. The analytical curves and linearity of the detector response for the test compounds was evaluated by injecting a total of ten calibration working standard solutions in the concentration levels 0.05, 0.07, 0.15, 0.31, 0.62, 1.25, 2.5, 5.0, 7.5 and 10.0 mg L-1 with three replicate injections per concentration.

Limit of detection (LOD) and quantification (LOQ)

LOD is the lowest concentration of analyte that can be detected and reliably distinguished from zero (or the noise level of the system), but not necessarily quantified; the concentration at which a measured value is larger than the uncertainty associated with it,40 and the limit of quantification (LOQ) is the lowest solute concentration that can be determined with acceptable precision and accuracy, under the stated experimental conditions. It is also expressed in concentration units. In this study, LOD and LOQ were determined considering the LOD as 3 times the baseline noise and the LOQ as the concentration that produced a signal 10 times the baseline noise, in a time close to the retention time of the analyte.36

Precision and accuracy

Precision represents the dispersion of results between independent assays, repetition of the same samples, similar samples or standards, under definite conditions.41,42

The precision is evaluated in terms of repeatability (RSDr) and intermediate precision (RSDip) and generally is expressed as relative standard deviation (RSD).37,43

Precision (repeatability) is an important criterion for evaluating the analytical method performance. It is the degree of agreement among individual test results when the procedure is applied repeatedly. The precision is usually expressed as the relative standard deviation (RSD).44

The precision in terms of repeatability was obtained by carrying out the extraction and analysis of fortified samples. Each spike level was extracted in three replicates and each extract injected three times. To evaluate the intermediate precision of the method, different days and operators were used, only for spiked levels of 5.0 and 10.0 μg L-1.

To evaluate the instrument precision, intra-day and inter-day precision was calculated by measuring the areas of the peaks obtained from three injections of three different concentrations in the same day, and in different days and operators, respectively.

The accuracy of the analytical method is the proximity of the results obtained for the method under investigation in relation to the true value. In pesticide residue analysis, the accuracy normally is evaluated by carrying out recovery assays.37,44 Recovery was determined at four concentrations, in accordance with the recommendations of the International Conference on Harmonisation (ICH) and Brazilian National Agency of Sanitary Vigilance (ANVISA), by adding known amounts of the reference substance at the beginning of the process. Analyses were carried out in three replicates of "blank" surface water samples spiked at four different levels (0.2, 1.25, 5.0 and 10.0 μg L-1). The accuracy of the method should be within 70 and 120% at all concentrations with RSD < 20%.37

Results and Discussion

Optimization of the chromatographic parameters and HPLC-DAD procedure

The optimum conditions of separation were better with the column Hyperclone BDS C18 5 μm 130 Å (250 × 4.6 mm) from Phenomenex because, in this case, a good peak shape and acceptable retention times were obtained. The water pH was adjusted to 3.0, because this pH provided optimum selectivity.

Several mixtures of mobile phases were tested. The mixtures with greater proportion of methanol, such as methanol:water (70:30, v/v) were not satisfactory because some substances were detected in a short time, before 5 min. This is not very good because the samples contain a relatively high concentration of anions and humic and fulvic acids that produce a high DAD response in the early part of the chromatogram. However, in proportions such as methanol:water (40:60, v/v), with a lower percentage of methanol, some compounds were detected in a long time, the same that had occurred with using a mixture of methanol, acetonitrile and water (30:24:46, v/v/v), where the results were not satisfactory because the pesticide tebuconazole was detected in a time longer than 20 min. Since run time is very important, we tried to find a shorter run time.

Unsatisfactory resolution was obtained for carbofuran and 2,4-D when the mobile phase was acetonitrile:water (60:40, v/v) pH 3.0, because co-eluting substances and the peaks were not ideal.

The gradient elution of the mobile phase was not satisfactory and we used an isocratic elution with a programation of the flow rate of the mobile phases, where the peaks were well resolved within acceptable retention time. The effective separation of the peaks in the chromatogram was achieved when the mobile phase composition were acetonitrile:water (52:48, v/v), pH 3.0 acidified with H3PO4 1:1; using a programming of the flow rate of 0.8 mL min-1 to 8 min, increasing for 1.2 mL min-1 to 14 min and coming back for 0.8 mL min-1 to 15 min.

When we used a programation of a flow rate of mobile phases, the run time was increased but the peaks were not co-eluting, resulting in the most satisfactory resolution for the four substances.

Method validation

Data of analytical curves were performed using ten different concentrations ranged between 0.05 and 10.0 mg L-1 of each substance with three replicate injections per concentration. Linear relationships between the ratios of the peak area signals and the corresponding concentrations were observed. The parameters of the analytical curves with the correlation coefficients are presented in Table 1. The values of the correlation coefficients were in accordance with the acceptable values of the ANVISA and Instituto Nacional de Metrologia, Normalização e Qualidade Industrial (INMETRO).41,42

By comparison of the response with the baseline noise, the LOD for the mixture of pesticides were between 0.01 and 0.02 mg L-1 and the LOQ were between 0.04 and 0.05 mg L-1. After the 250-fold SPE pre-concentration step the effective LOQ for the samples was 0.2 μg L-1, value that was below the maximum residue limit established by the European Union for the sum of all pesticides, and the values established for some pesticides in Brazilian legislation.15-17

Solid phase extraction prior to chromatographic determination was used in order to achieve a more sensitive method for the quantification of these pesticides. To accomplish this, samples of tap water were spiked with the pesticides at a concentration level range between 0.2 to 10.0 μg L-1.

The results from the optimization of the extraction procedure are shown in Table 2. Recoveries for all the compounds resulted in values ranged between 60.3% and 107.7%, using Strata X C18-E cartridges with good reproducibility RSD less than 20.7%, in the range of the acceptable values.37 In consonance with other studies the extraction of 2,4 D and clomazone showed good recoveries with C18.7,25

The recoveries using polymeric cartridges were not satisfactory, mainly the ones for 2,4-D and clomazone, which present recoveries lower than 40%.

The method precision was measured by comparing the standard deviation of the response for nine injections of the four different calibration standard solutions (0.2, 1.25, 5.0 and 10.0 mg L-1), evaluated in terms of repeatability (RSDr) and for two different calibration standard solutions (5.0 and 10.0 mg L-1) in terms of intermediate precision (RSDip). Table 2 summarizes the results of recovery and precision of the developed method, which presented satisfactory values.

Intra-day precision was calculated by measuring the areas of the peaks obtained from three injections of three analytical standard solutions (0.625, 2.5 and 7.5 mg L-1) in the same day, and inter-day values were taken in 10 days over a 2 month period. The precision was expressed as the percentage of the relative standard deviation (RSD). The values for the three different concentrations were ranged from 0.7-4.6% and 1.2-8.6% for intra- and inter-day values, respectively.

Applicability of the method

After the optimization and validation, the method was applied to real samples, to evaluate its applicability. The samples were collected in ten wells, located in an agricultural area in Rio Grande, the same region where a monitoring study of carbofuran, 2,4-D, clomazone and tebuconazole will be implemented for the period of one year. The groundwater samples were analyzed with the methodology described in this paper and the samples that gave negative findings for pesticides were spiked with 5.0 μg L-1 of all compounds and analyzed according to the procedure described.

The method was also successfully used for the determination of four pesticides in well water samples, demonstrating the usefulness of the developed method. The chromatogram of well water blank and spiked well water samples can be observed in Figure 2, with their identification spectra of the detected compounds.


Conclusions

The proposed HPLC-DAD method has been evaluated in terms of linearity, precision and accuracy, in a concentration range of 0.05 to 10.0 mg L-1, with a correlation coefficient higher than 0.9991. The method presented an average recovery of 87.9% and 96.9%, in repeatability and intermediate precision conditions for all compounds. It offers good accuracy and precision to determine pesticides in groundwater. The short analytical run time of 15.0 min leads to an effective cost and fast chromatographic procedure. Thus, the proposed methodology is rapid and selective with a simple sample preparation procedure that could be used for the convenient and effective determination of pesticide residue in groundwater samples.

Acknowledgments

The authors acknowledge CNPq, FINEP and FAPERGS for the financial support.

Received: March 12, 2008

Web Release Date: November 26, 2008

  • 1. Acero, J. L.; Benítez, F. J.; Real, F. J.; González, M.; J. Hazard. Mater. 2008, 153, 320.
  • 2. Spadotto, C. A.; Gomes, M. A. F.; Luchini, L. C.; Andréa, M. M.; Monitoramento do Risco Ambiental de Agrotóxicos: Príncipios e Recomendações, Embrapa Meio Ambiente: Jaguariúna, 2004, p. 29.
  • 3. Coutinho, C. F. B.; Tanimoto, S. T.; Galli, A.; Garbellini, G. S.; Takayama, M.; Amaral, R. B.; Mazo, L. H.; Avaca, L. A.; Machado, S. A. S.; Pesticidas: Revista de Ecotoxicologia e Meio Ambiente 2005, 15, 65.
  • 4
    SINDAG; Sindicato Nacional da Indústria de Produtos para Defesa Agrícola; 2002, available at http://www.sindag.com.br, accessed in July 2007.
    » link
  • 5
    SINDAG; Sindicato Nacional da Indústria de Produtos para Defesa Agrícola; 2006, available at http://www.sindag.com.br, accessed in August 2007.
    » link
  • 6. Filizola, H. F.; Ferracini, V. L.; Sans, L. M. A.; Gomes, M. A. F.; Ferreira, C. J. A.; Pesq. Agropec. Bras. 2002, 37, 659.
  • 7. Primel, E. G.; Zanella, R.; Kurz, M. H. S.; Gonçalves, F. F.; Machado, S. O.; Marchezan, E.; Quim. Nova 2005, 28, 605.
  • 8. Raposo Júnior, J. L.; Ré-Poppi, N.; Talanta 2007, 721, 1833.
  • 9. Brito, N. M.; Amarante Jr., O. P.; Santos, R. A. T. C. R.; Ribeiro, M. L.; Pesticidas: Revista de Ecotoxicologia e Meio Ambiente 2001, 11, 93.
  • 10. Cerejeira, M. J.; Viana, P.; Batista, S.; Pereira, T.; Silva, E.; Valério, M. J.; Silva, A.; Ferreira, M.; Silva-Fernandes, A. M.; Water Res. 2003, 37, 1055.
  • 11. Battaglin, W. A.; Furlong, E. T.; Burkhardt, M. R.; Peter, C. J.; Sci. Total Environ. 2000, 248, 123.
  • 12. Carabias-Martínez, R.; Rodríguez-Gonzalo, E.; Herrero-Hernández, E.; Sánchez-San Román, F. J.; Flores, M. G. P.; J. Chromatogr., A 2002, 950, 157.
  • 13. Guzzella, L.; Pozzoni, F.; Giuliano, G.; Environ. Pollut. 2006, 142, 344.
  • 14. D'Archivio, A. A.; Fanelli, M.; Mazzeo, P.; Ruggieri, F.; Talanta 2007, 71, 25.
  • 15. Alda, M. L.; Kuster, M.; Barceló, D.; Mass Spectrom. Rev. 2006, 25, 900.
  • 16. Ministério da Saúde. Portaria No. 518, de 25 de março de 2004. Diário Oficial da União, 26 março. Ministry of Health, Brazil.
  • 17. Ministério do Meio Ambiente. Conselho Nacional do Meio Ambiente (CONAMA). Resolução no 357, de 17 março de 2005. Classificação dos corpos de água e diretrizes ambientais para o seu enquadramento bem como estabelece condições e padrões de lançamento de efluentes. Diário Oficial da União, 18 março 2005. Ministry of Environment, National Council of Environment, Brazil.
  • 18. Picó, Y.; Fernández, M.; Ruiz, M. J.; Font, G.; J. Biochem. Biophys. Methods 2007, 70, 117.
  • 19. Ferracini, V. L.; Queiroz, S. C. N.; Gomes, M. A. F.; Santos, G. L.; Quim. Nova 2005, 28, 380.
  • 20. Brondi, S. H. G.; Lanças, F. M.; J. Braz. Chem. Soc. 2005, 16, 650.
  • 21. Azevedo, D. A.; Gerchon, E.; Reis, E. O.; J. Braz. Chem. Soc. 2004, 15, 292.
  • 22. Ahmadi, F.; Assadi, Y.; Milani Hosseini, S. M. R.; Rezaee, M.; J. Chromatogr., A 2006, 1101, 307.
  • 23. Bruzzoniti, M. C.; Sarzanini, C.; Costantino, G.; Fungi, M.; Anal. Chim. Acta 2006, 578, 241.
  • 24. Ahmed, F. E.; Trends Anal. Chem. 2001, 20, 649.
  • 25. Palma, G.; Sánchez, A.; Olave, Y.; Encina, F.; Palma, R.; Barra, R.; Chemosphere 2004, 57, 763.
  • 26. Sánchez-Ortega, A.; Sampedro, M. C.; Unceta, N.; Goicolea, M. A.; Barrio, R. J.; J. Chromatogr., A 2005, 1094, 70.
  • 27. He, Y.; Lee, H. K.; J. Chromatogr., A 2006, 1122, 7.
  • 28. Tran, A. T. K.; Hyne, R. V.; Doble, P.; Chemosphere 2007, 67, 944.
  • 29. Carabias-Martínez, R.; Rodríguez-Gonzalo, E.; Herrero-Hernández, E.; Hernández-Méndez, J.; Anal. Chim. Acta 2004, 517, 71.
  • 30. Miyauchia, T.; Moria, M.; Itob, K.; J. Chromatogr., A 2005, 1063, 137.
  • 31. Cid, A. P.; Del Mundo, F. R.; Espino, M. P. B.; Philippine Agric. Scientist 2006, 89, 71.
  • 32. Sannino, A.; Bolzoni, L.; Bandini, M.; J. Chromatogr., A 2004, 1036, 161.
  • 33. Rodrigues, A. M.; Ferreira, V.; Cardoso, V. V.; Ferreira, E.; Benoliel, M. J.; J. Chromatogr., A 2007, 1150, 267.
  • 34. Pirard, C.; Widart, J.; Nguyen, B. K.; Deleuze, C.; Heudt, L.; Haubruge, E.; De Pauw, E.; Focant, J. F.; J. Chromatogr., A 2007, 1152, 116.
  • 35. Beceiro-González, E.; Concha-Graña, E.; Guimaraes, A.; Gonçalves, C.; Muniategui-Lorenzo, S.; Alpendurada, M. F.; J. Chromatogr., A 2007, 1141, 165.
  • 36. Ribani, M.; Collins, C. H.; Bottoli, C. B. G.; J. Chromatogr., A 2007, 1156, 201.
  • 37. Ribani, M.; Bottoli, C. B. G.; Collins, C. H.; Jardim, I. C. S. F.; Melo, L. F. C.; Quim. Nova 2004, 27, 771.
  • 38. International Conference on Harmonisation (ICH), Validation of Analytical Procedures: Definitions and Terminology, Q2A (CPMP/ICH/381/95), 1995.
  • 39. Shabir, G.A.; J. Chromatogr., A 2003, 987, 57.
  • 40. González, A. G.; Herrador, M. A.; TrAC, Trends Anal. Chem. 2007, 26, 227.
  • 41. Instituto Nacional de Metrologia, Normalização e Qualidade Industrial (INMETRO); Orientações sobre Validacão de Métodos de Ensaios Químicos, DOQ-CGCRE-008, 2003.
  • 42
    Agência Nacional de Vigilância Sanitária (ANVISA); Guia para Validação de Métodos Analíticos e Bioanalíticos, Resolução RE No. 899, 2003. -National Health Surveillance Agency, Brazil.
  • 43. Zanella, R.; Primel, E. G.; Gonçalves, F. F.; Martins, A. F.; J. Chromatogr., A 2000, 904, 257.
  • 44. International Conference on Harmonisation (ICH), Validation of Analytical: Methodology, Q2B (CPMP/ICH/281/95), 1996.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      14 Jan 2009
    • Date of issue
      2009

    History

    • Accepted
      26 Nov 2008
    • Received
      12 Mar 2008
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br