Evaluation of Graphene for Dispersive Solid-Phase Extraction of Triazine and Neonicotine Pesticides from Environmental Water

Abstracts

In this work, graphene was used for dispersive solid-phase extraction of 11 triazine herbicides and 5 neonicotine insecticides from water. The extraction conditions such as the amount of graphene, the desorption solvent and the solution pH were optimized. Under the optimal conditions, high efficient enrichment was achieved for the quantitative analysis of the pesticides. Triazines and neonicotines were determined by gas chromatograph-mass spectrometer (GC-MS) and high performance liquid chromatography electrospray tandem mass spectrometry (LC-MS/MS), respectively. The results indicated that graphene was an excellent adsorbent for the adsorption of pesticides. The recoveries of the pesticides were 83.0-108.9% with relative standard deviation (RSD) between 2.4% and 12.3%. The limits of detection (LODs) ranged from 0.03 μg L-1 to 0.40 μg L-1. Finally, the proposed method was applied for real sample analysis, such as tap and river water.

graphene; triazines; neonicotines; fast extraction


Neste trabalho, grafeno foi usado na extração em fase sólida dispersiva de 11 herbicidas triazina e 5 inseticidas neonicotina da água. As condições de extração como quantidade de grafeno, solvente de dessorção e pH da solução foram otimizados. Nas condições otimizadas, um enriquecimento de alta eficiência foi alcançado na análise quantitativa dos pesticidas. Triazinas e neonicotinas foram determinadas por espectrometria de massas-cromatografia gasosa (GC-MS) e cromatografia líquida de alta eficiência acoplada à espectrometria de massas (LC-MS/MS), respectivamente. Os resultados indicaram que o grafeno foi um excelente adsorvente na adsorção de pesticidas. A recuperação dos pesticidas foi 83,0-108,9% com desvio padrão relativo (RSD) entre 2,4% e 12,3%. Os limites de detecção (LODs) variaram entre 0,03 μg L-1 e 0,40 μg L-1. Finalmente, o método proposto foi aplicado na análise de amostras reais, como água de torneira e de rio.


Introduction

In recent years, herbicides and insecticides have been broadly used for increasing agricultural production and quality. For example, triazines were important herbicides used in weed control and neonicotine insecticides have been widely used due to their low toxicity and high activity against insects.1Nagaraju, D.; Huang, S. D.; J. Chromatogr. A 2007, 1161, 89.,2Zhou, Q. X.; Ding, Y. J.; Xiao, J. P.; Anal. Bioanal. Chem. 2006, 385, 1520. However, the abuse of these compounds has resulted in contamination of surface water, groundwater, soil and air.3Fernandez-Alvarez, M.; Llompart, M.; Lamas, J. P.; Lores, M.; Garcia-Jares, C.; Cela, R.; Dagnac, T.; J. Chromatogr. A 2009, 1216, 2832. To assure water safety and quality, it is of paramount importance for developing sensitive, selective, rapid and cost-effective methods for monitoring the presence of pesticides.

Pretreatment is required for trace analysis of pesticides in water samples. The most commonly used pretreatment methods for water sample were solid-phase extraction (SPE),4Font, G.; Manes, J.; Molto, J. C.; Pico, Y.; J. Chromatogr. A 1993, 642, 135.,5Molto, J. C.; Pico, Y.; Font, G.; Manes, J.; J. Chromatogr. A 1991, 555, 137. matrix solid-phase dispersion extraction (MSPE),6Guan, S. X.; Yu, Z. G.; Yu, H. N.; Song, C. H.; Song, Z. Q.; Qin, Z.; Chromatographia 2011, 73, 33. and solid-phase micro extraction (SPME).7Eisert, R.; Pawliszyn, J.; J. Chromatogr. A 1997, 776, 293. For all these methods, the adsorbent materials, which determine the selectivity and sensitivity of the methods, are crucial components. The most widely used adsorbent are C18 chemically-bonded to silica,8Somsen, G. W.; Jagt, I.; Gooijer, C.; Velthorst, N. H.; Brinkman, U. A. T.; Visser, T.; J. Chromatogr. A 1996, 756, 145.

Pocurull, E.; Marce, R. M.; Borrull, F.; Chromatographia 1995, 41, 521.
-1010 Chiron, S.; Alba, A. F.; Barcelo, D.; Environ. Sci. Technol. 1993, 27, 2352. carbon black,1111 Sabik, H.; Jeannot, R.; Sauvard, E.; Analysis 2000, 28, 835.,1212 DiCorcia, A.; Bellioni, A.; Madbouly, M. D.; Marchese, S.; J. Chromatogr. A 1996, 733, 383. and polymeric resins.1313 Puig, D.; Barcelo, D.; Anal. Chim. Acta 1995, 311, 63.

14 Pocurull, E.; Marce, R. M.; Borrull, F.; J. Chromatogr. A 1996, 738, 1.

15 Marce, R. M.; Prosen, H.; Crespo, C.; Calull, M.; Borrull, F.; Brinkman, U. A. T.; J. Chromatogr. A 1995, 696, 63.
-1616 Liska, I.; Brouwer, E. R.; Ostheimer, A. G. L.; Lingeman, H.; Brinkman, U. A. T.; Geerdink, R. B.; Mulder, W. H.; Int. J. Environ. Anal. Chem. 1992, 47, 267. Carbon nano-materials represent a novel class of adsorbent with large specific surface area, including fullerenes,1717 Cheng, X. K.; Kan, A. T.; Tomson, M. B.; J. Chem. Eng. Data 2004, 49, 675.,1818 Ballesteros, E.; Gallego, M.; Valcarcel, M.; J. Chromatogr. A 2000, 869, 101. carbon nanotubes (CNTs),6Guan, S. X.; Yu, Z. G.; Yu, H. N.; Song, C. H.; Song, Z. Q.; Qin, Z.; Chromatographia 2011, 73, 33.,1919 Liu, Q.; Shi, J. B.; Zeng, L. X.; Wang, T.; Cai, Y. Q.; Jiang, G. B.; J. Chromatogr. A 2011, 1218, 197.,2020 Ravelo-Perez, L. M.; Herrera-Herrera, A. V.; Hernandez‑Borges, J.; Rodriguez-Delgado, M. A.; J. Chromatogr. A 2010, 1217, 2618. and graphene.1919 Liu, Q.; Shi, J. B.; Zeng, L. X.; Wang, T.; Cai, Y. Q.; Jiang, G. B.; J. Chromatogr. A 2011, 1218, 197. Graphene has attracted tremendous attention since it was discovered in 2004. It has a large delocalized π-electron system and high theoretical specific surface area (about 2630 m2 g-1),2121 Stoller, M. D.; Park, S.; Zhu, Y.; An, J.; Ruoff, R. S.; Nano Lett. 2008, 8, 3498. which make it very attractive as an adsorbent material for the adsorb of both non-polar and polar compounds.2222 Zhang, Y. W.; Ma, H. L.; Peng, J.; Zhai, M. L.; Yu, Z. Z.; J. Mater. Sci. 2013, 48, 1883.

23 Pei, Z. G.; Li, L. Y.; Sun, L. X.; Zhang, S. Z.; Shan, X. Q.; Yang, S.; Wen, B.; Carbon 2013, 51, 156.

24 Maliyekkal, S. M.; Sreeprasad, T. S.; Krishnan, D.; Kouser, S.; Mishra, A. K.; Waghmare, U. V.; Pradeep, T.; Small 2013, 9, 273.
-2525 Xu, J.; Wang, L.; Zhu, Y. F.; Langmuir 2012, 28, 8418. Besides, graphene and graphene-based materials also have been reported with good performance for the adsorption of heavy metals,2626 Chandra, V.; Park, J.; Chun, Y.; Lee, J. W.; Hwang, I. C.; Kim, K. S.; ACS Nano 2010, 4, 3979.

27 Hao, L. Y.; Song, H. J.; Zhang, L. C.; Wan, X. Y.; Tang, Y. R.; Lv, Y.; J. Colloid Interface Sci. 2012, 369, 381.

28 Kyzas, G. Z.; Travlou, N. A.; Deliyanni, E. A.; Colloids Surf., B 2014, 113, 467.
-2929 Pastewka, L.; Malola, S.; Moseler, M.; Koskinen, P.; J. Power Sources 2013, 239, 321. organic dyes,3030 Wang, C.; Feng, C.; Gao, Y. J.; Ma, X. X.; Wu, Q. H.; Wang, Z.; Chem. Eng. J. 2011, 173, 92.

31 Travlou, N. E.; Kyzas, G. Z.; Lazaridis, N. K.; Deliyanni, E. A.; Chem. Eng. J. 2013, 217, 256.

32 Kyzas, G. Z.; Travlou, N. A.; Kalogirou, O.; Deliyanni, E. A.; Materials 2013, 6, 1360.
-3333 He, G. Y.; Zhang, J. G.; Zhang, Y.; Chen, H. Q.; Wang, X.; J. Dispersion Sci. Technol. 2013, 34, 1223. sulfonamide antibiotics,3434 Luo, Y. B.; Shi, Z. G.; Gao, Q.; Feng, Y. Q.; J. Chromatogr. A 2011, 1218, 1353. and pesticides,3535 Zhao, G. Y.; Song, S. J.; Wang, C.; Wu, Q. H.; Wang, Z.; Anal. Chim. Acta 2011, 708, 155.,3636 Zhang, S.; Du, Z.; Li, G.; Talanta 2013, 115, 32. and phthalic acid esters (PAEs).3737 Wu, X. L.; Hong, H. J.; Liu, X. T.; Guan, W. B.; Meng, L. X.; Ye, Y.; Ma, Y. Q.; Sci. Total Environ. 2013, 444, 224.

In this work, graphene was successfully used for effective dispersive solid-phase extraction of 11 triazines and 5 neonicotines from environmental water. The two chemical classes of pesticide were determined by gas chromatograph-mass spectrometer (GC-MS) and high performance liquid chromatography electrospray tandem mass spectrometry (LC-MS/MS), respectively. The method was faster and simpler compared with other methods. Furthermore, the amount of adsorbent used was much lower than other methods.

Experimental

Reagents and solutions

Standards of all pesticides and triphenyl phosphate (TPP) (99%) used as internal standard with purity higher than 98% were purchased from J&K Scientific Ltd (Beijing, China) and were maintained at 4 ºC in the dark.

Chromatographic grade acetonitrile, acetone, n-hexane, methanol and ethyl acetate were purchased from MREDA Co., Ltd (Beijing, China). Analytical reagent grade anhydrous sodium chloride (NaCl), potassium permanganate (KMnO4), hydrazine hydrate (85%), ammonia solution (25%), sulphuric acid (98%), hydrochloric acid (38%) and 20% hydrogen peroxide (H2O2) were purchased from Beijing Chemical Works (Beijing, China). Ultrapure water was obtained from a Milli-Q water purification system (Millipore, Billerica, MA, USA). Expandable graphite was obtained from Qingdao Hensen Graphite Co., Ltd (Qingdao, China).

Tap water samples were collected from a water tap in our lab (Beijing, China). River water samples were collected from the top layer (0-50 cm) of Jingmi River and Xiaojiahe River (Beijing, China). All samples were cooled in a refrigerator (4 ºC) during transport to the laboratory.

Preparation of graphene

Graphene oxide (GO) was synthesized from expandable graphite using a modified Hummers method,3838 Hummers, W. S.; Offeman, R. E.; J. Am. Chem. Soc. 1958, 80, 1339. and then was reduced via chemical reduction to prepare graphene,3939 Stankovich, S.; Dikin, D. A.; Piner, R. D.; Kohlhaas, K. A.; Kleinhammes, A.; Jia, Y.; Wu, Y.; Nguyen, S. T.; Ruoff, R. S.; Carbon 2007, 45, 1558. which was described detailed in our previous work.3737 Wu, X. L.; Hong, H. J.; Liu, X. T.; Guan, W. B.; Meng, L. X.; Ye, Y.; Ma, Y. Q.; Sci. Total Environ. 2013, 444, 224. Briefly, GO was reduced to graphene by hydrazine hydrate in an oil bath at 80 ºC for 24 h, then was purified with ultrapure water. The obtained graphene was dispersed in water with a concentration of 3 mg mL-1, which was determined by freeze drying.

Instrumentation

The analysis of neonicotines was achieved using an Agilent 1200 HPLC series and an Agilent 6410B triple-quadrupole mass spectrometer equipped with an electrospray (ESI) ionization interface (Agilent Technologies, USA). For instrument control, masshunter workstation software data acquisition for triple quad B.02.01 (B2043.12) and qualitative analysis version B.03.01/build 3.1.346.0 were used for data acquisition and processing. A reversed phase ZORBAX SB-C18 column (1.8 μm particle size, 2.1 mm × 50 mm) from Agilent technologies was employed for HPLC separation at 30 ºC. The mobile phase consisted of methanol and water (1:3, v/v). The flow rate was 0.3 mL min-1. The source temperature was 100 ºC and desolvation gas temperature was 300 ºC. The nebulizer gas (N2) were set at 10.0 L min-1 and 35.0 psi. The quantitative parameters were showed in Table 1.

Table 1
HPLC-MS/MS analysis of 5 neonicotine pesticides

The analysis of triazines was performed by an Agilent 6890N GC-5975B MSD (Agilent Technologies, USA). The separation was achieved on a fused silica capillary column (HP-5, 30 m × 0.25 mm i.d., with 0.25 μm film thickness). The column temperature was programmed at 120 ºC for 1 min initially. Then raised to 175 ºC at rate of 35 ºC min-1 and held for 1 min. Afterwards, it was increased to 215 ºC at a rate of 35 ºC min-1 and held for 1 min. Finally, increased to 260 ºC at a rate of 35 ºC min-1, and kept for 5 min. The injector temperature, ion source temperature and quadrupole temperature were 290 ºC, 230 ºC and 280 ºC, respectively. Helium of high purity flow rate: 1 mL min-1. The retention times, quantitative ions and qualitative ions of 11 triazines were shown in Table 2.

Table 2
Retention time, qualitative ions and quantitative ions for GC/MS analysis of 11 triazine pesticides

Procedure for the extraction of the pesticides

Graphene dispersive-SPE

Firstly, an appropriate amount of graphene was added into 40 mL water sample in a centrifuge tube. Secondly, to trap the analytes, the mixture was vortexed for 1 min. Subsequently, graphene was isolated from the solution by centrifugation (3800 r min-1, 5 min). The supernatant was discarded. Finally, the target compounds were adsorbed on graphene. In this procedure, the amount of graphene and the pH of the solution were tested, which were shown in Table 3.

Table 3
Conditions studied to optimize the extraction procedure

The fast desorption procedure

In this procedure, desorption solvent was added and vortexed for 1 min to desorb the target analytes. In the meantime, NaCl (2 g) was used for salting out. Subsequently, the mixture was centrifuged for 5 min (3800 r min-1). Finally, in order to avoid extracting the bottom graphene as far as possible and concentrate the analytes, 4 mL of the upper organic phase was concentrated by evaporator and dissolved in 1 mL acetonitrile, then was transferred into a 1.5 mL vial for determination. In this procedure, the types of desorption solvent and the volume of desorption were optimized (Table 3).

Results and Discussion

In order to obtain optimized extraction conditions, the adsorption and desorption conditions, including the amount of graphene, the desorption solvent and the pH of the solution were tested. The pesticides were all spiked at 1 μg L-1 with 5 replicates.

The amount of graphene

The amount of adsorbent, the key factor of the extraction procedure, had a significant effect on the extraction efficiency. The graphene used in this work was dispersed in ultrapure water with a final concentration of 3 mg mL-1. Different volumes of graphene were tested for the two chemical classes of pesticides in neutral solution, and 10 mL of acetonitrile was used for desorption.

4 mL, 5 mL, 6 mL, 7 mL and 8 mL of graphene were tested for adsorption of neonicotines. The results shown in Figure 1a indicated that the recoveries of 5 neonicotines were the hightest when 6 mL of graphene was used. However, more graphene results in lower recoveries. It may be because the recoveries of the analytes were affected by both graphene and eluting solvent, which means that when the amount of graphene was increased and the eluting solvent remained unchanged, the recoveries of some pesticides may have decreased.

Figure 1
Selection of the amount of graphene (a) neonicotine pesticides and (b) triazine pesticides.

Besides, 1 mL, 2 mL, 3 mL, 4 mL and 5 mL of graphene were used for the adsorption of trazines. The results (Figure 1b) showed that the recoveries of most pesticides increased with the increase of the amount of graphene. With the recoveries of pesticides and the amount of graphene taken into account, 3 mL of graphene was chosen for the adsorption of trazines.

Selection of desorption solvent

The desorption solvent is another prime factor that affects the extraction efficiency. 10 mL of different desorption solvents were used for desorption in neutral solution, without adjusting, with optimized amount of graphene.

Acetonitrile, ethyl acetate, acetone, methanol were tested for 5 neonicotines. It indicated that when acetonitrile was used for desorption of the 5 neonicotine insecticides, the recoveries (65.8-81.4%) were higher than the other 3 solvents (Figure 2a). The recoveries of 5 neonicotines were 50.6-77.9% when acetone was used. Ethyl acetate and methanol were not chosen because of the relatively low recoveries (26.9-58.8%), therefore, acetonitrile was chosen.

Figure 2
Selection of desorption solvent (a) neonicotine pesticides and (b) triazine pesticides.

Triazine pesticides are planar analytes, which can easily adsorb on graphene. Consequently, n-hexane with relatively weak polarity and benzene with strong elution ability were chosen besides acetonitrile and ethyl acetate. The results were showed in Figure 2b. When n-hexane was used, the recoveries of 11 trazine herbicides were 0.0-11.9%. For benzene, the recoveries were between 0.0% and 96.8%. Benzene and n-hexane are water-insoluble and the polarity of both was too weak to elute the analytes completely, which result in low recoveries of the analytes. The recoveries of 11 trazine herbicides were 72.7-106.6% when acetonitrile and ethyl acetate were used. Thus, acetonitrile was chosen for relative higher recoveries.

Therefore, acetonitrile was chosen as the final desorption solvent for both trazines and neonicotines.

Selection of desorption solvent volume

Besides the types of desorption solvent, the volume of the solvent can also influence the desorption efficiencies. Acetonitrile was used for desorption in neutral solution with optimized amount of graphene.

4 mL, 6 mL, 7 mL, 8 mL and 10 mL of acetonitrile were tested for the desorption of neonicotine pesticides (Figure 3a). 6 mL, 8 mL, 10 mL, 12 mL and 14 mL of acetonitrile were tested for the desorption of trazines (Figure 3b). The recoveries showed a growing trend initially with the increase of desorption solvent, and then remained almost unchanged. In order to obtain higher desorption efficiencies and save solvent, 8 mL of acetonitrile was chosen as the final volume for both trazines and neonicotines.

Figure 3
Selection of the eluent volume (a) neonicotine pesticides and (b) triazine pesticides.

Selection of solution pH

Solution pH plays another important role for the adsorption of the analytes by affecting both the existing forms of the target compounds and the charge species and density on the sorbent surface.3232 Kyzas, G. Z.; Travlou, N. A.; Kalogirou, O.; Deliyanni, E. A.; Materials 2013, 6, 1360. The solution pH was adjusted by adding hydrochloric acid or sodium hydroxide solution. In this work, the sample pH was also investigated (Figure 4). The set points were pH 3, 5, 7, 9, 11 for trazines and 4, 6, 7, 8, 10 for neonicotines. The results showed that the solution pH has no significantly influence on the recoveries of neonicotines. In the acidic conditions (pH 3-5), the recoveries of some trazines were unsatisfactory, for example, the recoveries of atrazine and cyprazine were 41.3% and 44.7%, respectively. However, in neutral (pH 7) or alkaline (pH 9-11) conditions, the recoveries were all more than 70%. In conclusion, graphene possess satisfactory absorption performance for both trazines and neonicotines in neutral water system. As most of the environmental water samples were close to neutralization, the extraction procedure was taken without adjusting of pH value.

Figure 4
Selection of the solution pH (a) neonicotine pesticides and (b) triazine pesticides.

Validation of the method

Based on the above optimal conditions, one simple and rapid method for the determination of trazines and neonicotines was developed. The linear range, the limit of detection (LOD), the limit of quantitation (LOQ) and the precision of the method were all investigated (Table 4). The linear range for the analytes was in the final extract.

Table 4
Linear range, coefficient of determination (R2), LOD and LOQ for all pesticides

The LODs, which were calculated based on the ratio of signal to noise (S/N = 3), were 0.03-0.4 μg L-1. Calibration curves were established over the range 0.5-100 μg L-1 for all pesticides and satisfactory coefficient of determination (R2) (0.9979-0.9998) were obtained. The repeatability of the method was carried out by five parallel experiments spiked at 0.5, 1 and 5 μg L-1 for neonicotines and 0.5, 5 and 10 μg L-1 for trazines. The recoveries (Table 5) of all pesticides were 83.0-108.9%, with the relative standard deviation (RSD) between 2.4% and 12.3%, which indicate that graphene has an outstanding adsorption capacity for all targets.

Table 5
Repeatability study of the method in purified water (n = 5)

Comparison with SPE

The most commonly used method for the pre-concentration of trazines and neonicotines is SPE.2Zhou, Q. X.; Ding, Y. J.; Xiao, J. P.; Anal. Bioanal. Chem. 2006, 385, 1520.,4040 Baskaran, S.; Kookana, R. S.; Naidu, R.; J. Chromatogr. A 1997, 787, 271.

41 DiCorcia, A.; Crescenzi, C.; Guerriero, E.; Samperi, R.; Environ. Sci. Technol. 1997, 31, 1658.

42 Zhou, Q. X.; Xiao, J. P.; Wang, W. D.; Liu, G. G.; Shi, Q. Z.; Wang, J. H.; Talanta 2006, 68, 1309.

43 Zhou, Q. X.; Ding, Y. J.; Xia, J. P.; Chromatographia 2007, 65, 25.

44 Zhou, Q. X.; Xiao, J. P.; Ding, Y. J.; Anal. Chim. Acta 2007, 602, 223.

45 Min, G.; Wang, S.; Zhu, H. P.; Fang, G. Z.; Zhang, Y.; Sci. Total Environ. 2008, 396, 79.
-4646 Zhao, R. S.; Yuan, J. P.; Jiang, T.; Shi, J. B.; Cheng, C. G.; Talanta 2008, 76, 956. Listed in Table 6 is the comparison of this method with SPE procedure for the enrichment of the two chemical classes of pesticide. Although the method developed in our current study did not give prominence to low LODs, it was outstanding in other aspects.

Table 6
Comparation of this method with other methods

Firstly, the water volume needed for analysis in this method was only 40 mL, which was lower than SPE listed in Table 6, which were more than 100 mL.4040 Baskaran, S.; Kookana, R. S.; Naidu, R.; J. Chromatogr. A 1997, 787, 271.,4545 Min, G.; Wang, S.; Zhu, H. P.; Fang, G. Z.; Zhang, Y.; Sci. Total Environ. 2008, 396, 79.,4646 Zhao, R. S.; Yuan, J. P.; Jiang, T.; Shi, J. B.; Cheng, C. G.; Talanta 2008, 76, 956.

Secondly, the amount of adsorbent was much lower than other methods. In this method, 3 mL of graphene (3 mg mL-1) was applied for adsorbing trazines and 6 mL for neonicotines, indicating that 0.018 g of graphene can effectively extract these two chemical classes of pesticide. However, other adsorbents used in SPE procedure were more than 0.1 g.4040 Baskaran, S.; Kookana, R. S.; Naidu, R.; J. Chromatogr. A 1997, 787, 271.

41 DiCorcia, A.; Crescenzi, C.; Guerriero, E.; Samperi, R.; Environ. Sci. Technol. 1997, 31, 1658.

42 Zhou, Q. X.; Xiao, J. P.; Wang, W. D.; Liu, G. G.; Shi, Q. Z.; Wang, J. H.; Talanta 2006, 68, 1309.

43 Zhou, Q. X.; Ding, Y. J.; Xia, J. P.; Chromatographia 2007, 65, 25.

44 Zhou, Q. X.; Xiao, J. P.; Ding, Y. J.; Anal. Chim. Acta 2007, 602, 223.

45 Min, G.; Wang, S.; Zhu, H. P.; Fang, G. Z.; Zhang, Y.; Sci. Total Environ. 2008, 396, 79.
-4646 Zhao, R. S.; Yuan, J. P.; Jiang, T.; Shi, J. B.; Cheng, C. G.; Talanta 2008, 76, 956.

Thirdly, as for solvent consuming, only 8 mL of acetonitrile was used for desorption in this work. However, for SPE methods, the cartridges should be conditioned with appropriate solvent prior to preconcentration procedure. Besides, the parameters including the kind, the volume, and the flow rate of the solvent should be optimized, which were verbose and expensive.

Fourthly, compared to previously reported SPE methods, our method was a fast and effective extraction method. The extraction procedure including two main steps: fast adsorption and fast desorption procedure, which was completed within 20 min. It was time-saving than SPE methods (more than 20 min).4040 Baskaran, S.; Kookana, R. S.; Naidu, R.; J. Chromatogr. A 1997, 787, 271.,4141 DiCorcia, A.; Crescenzi, C.; Guerriero, E.; Samperi, R.; Environ. Sci. Technol. 1997, 31, 1658.

Analysis of environmental water samples

The repeatability study was carried out by five parallel experiments spiked at 2 μg L-1 for both tap water and river water. The recoveries were 91.1-102.2%, and the RSDs were all less than 13.0%. This indicated that the water matrixes have no tremendous influence on the recoveries of the pesticides. The proposed method was applied for the analysis of the pesticides in real environmental water samples including tap and river water samples collected in Beijing (shown in Table 7).

Table 7
Recoveries (spiked at 2 μg L-1, n = 5) of the two chemical classes of pesticide in tap water, river water and real sample analysis

The detection of acetamiprid was 0.2 μg L-1 in Jingmi River and the detection of atrazine was 100 μg L-1 in Xiaojia River. Other pesticides were not found in river samples. Besides, none of the pesticides were found in tap water. The typical chromatograms of the extracted pesticides were showed in Figure 5-6.

Figure 5
Typical chromatograms of (a) 5 neonicotines spiked at 2 μg L-1 in water sample, (b) tap water sample and (c) Jingmi River water sample. 1, thiamethoxam; 2, imidacloprid; 3, imidaclothiz; 4, acetamiprid; 5, thiacloprid.
Figure 6
Total ion chromatogram (TIC) of (a) 11 triazine pesticides spiked at 50 μg L-1 in water sample, (b) tap water sample, and (c) Xiaojia River water sample.

Conclusions

In this study, graphene was used as an adsorbent for effective enrichment of neonicotines and triazines from water. Under the optimal conditions, a fast extraction method including fast adsorption and fast desorption procedure was established. This rapid and simple method has many advantages over other preparation methods, such as less use amount of adsorbent and eluent, easy-operated and time-saving. The results indicated that graphene was an excellent adsorbent material and has a wide application in pesticides analysis.

Acknowledgements

This work was supported by Beijing Municipal Natural Science Foundation (No. KZ201210020019) and National Science Foundation for Fostering Talents in Basic Research of China (No. J1210064).

References

  • 1
    Nagaraju, D.; Huang, S. D.; J. Chromatogr. A 2007, 1161, 89.
  • 2
    Zhou, Q. X.; Ding, Y. J.; Xiao, J. P.; Anal. Bioanal. Chem. 2006, 385, 1520.
  • 3
    Fernandez-Alvarez, M.; Llompart, M.; Lamas, J. P.; Lores, M.; Garcia-Jares, C.; Cela, R.; Dagnac, T.; J. Chromatogr. A 2009, 1216, 2832.
  • 4
    Font, G.; Manes, J.; Molto, J. C.; Pico, Y.; J. Chromatogr. A 1993, 642, 135.
  • 5
    Molto, J. C.; Pico, Y.; Font, G.; Manes, J.; J. Chromatogr. A 1991, 555, 137.
  • 6
    Guan, S. X.; Yu, Z. G.; Yu, H. N.; Song, C. H.; Song, Z. Q.; Qin, Z.; Chromatographia 2011, 73, 33.
  • 7
    Eisert, R.; Pawliszyn, J.; J. Chromatogr. A 1997, 776, 293.
  • 8
    Somsen, G. W.; Jagt, I.; Gooijer, C.; Velthorst, N. H.; Brinkman, U. A. T.; Visser, T.; J. Chromatogr. A 1996, 756, 145.
  • 9
    Pocurull, E.; Marce, R. M.; Borrull, F.; Chromatographia 1995, 41, 521.
  • 10
    Chiron, S.; Alba, A. F.; Barcelo, D.; Environ. Sci. Technol. 1993, 27, 2352.
  • 11
    Sabik, H.; Jeannot, R.; Sauvard, E.; Analysis 2000, 28, 835.
  • 12
    DiCorcia, A.; Bellioni, A.; Madbouly, M. D.; Marchese, S.; J. Chromatogr. A 1996, 733, 383.
  • 13
    Puig, D.; Barcelo, D.; Anal. Chim. Acta 1995, 311, 63.
  • 14
    Pocurull, E.; Marce, R. M.; Borrull, F.; J. Chromatogr. A 1996, 738, 1.
  • 15
    Marce, R. M.; Prosen, H.; Crespo, C.; Calull, M.; Borrull, F.; Brinkman, U. A. T.; J. Chromatogr. A 1995, 696, 63.
  • 16
    Liska, I.; Brouwer, E. R.; Ostheimer, A. G. L.; Lingeman, H.; Brinkman, U. A. T.; Geerdink, R. B.; Mulder, W. H.; Int. J. Environ. Anal. Chem. 1992, 47, 267.
  • 17
    Cheng, X. K.; Kan, A. T.; Tomson, M. B.; J. Chem. Eng. Data 2004, 49, 675.
  • 18
    Ballesteros, E.; Gallego, M.; Valcarcel, M.; J. Chromatogr. A 2000, 869, 101.
  • 19
    Liu, Q.; Shi, J. B.; Zeng, L. X.; Wang, T.; Cai, Y. Q.; Jiang, G. B.; J. Chromatogr. A 2011, 1218, 197.
  • 20
    Ravelo-Perez, L. M.; Herrera-Herrera, A. V.; Hernandez‑Borges, J.; Rodriguez-Delgado, M. A.; J. Chromatogr. A 2010, 1217, 2618.
  • 21
    Stoller, M. D.; Park, S.; Zhu, Y.; An, J.; Ruoff, R. S.; Nano Lett. 2008, 8, 3498.
  • 22
    Zhang, Y. W.; Ma, H. L.; Peng, J.; Zhai, M. L.; Yu, Z. Z.; J. Mater. Sci. 2013, 48, 1883.
  • 23
    Pei, Z. G.; Li, L. Y.; Sun, L. X.; Zhang, S. Z.; Shan, X. Q.; Yang, S.; Wen, B.; Carbon 2013, 51, 156.
  • 24
    Maliyekkal, S. M.; Sreeprasad, T. S.; Krishnan, D.; Kouser, S.; Mishra, A. K.; Waghmare, U. V.; Pradeep, T.; Small 2013, 9, 273.
  • 25
    Xu, J.; Wang, L.; Zhu, Y. F.; Langmuir 2012, 28, 8418.
  • 26
    Chandra, V.; Park, J.; Chun, Y.; Lee, J. W.; Hwang, I. C.; Kim, K. S.; ACS Nano 2010, 4, 3979.
  • 27
    Hao, L. Y.; Song, H. J.; Zhang, L. C.; Wan, X. Y.; Tang, Y. R.; Lv, Y.; J. Colloid Interface Sci. 2012, 369, 381.
  • 28
    Kyzas, G. Z.; Travlou, N. A.; Deliyanni, E. A.; Colloids Surf., B 2014, 113, 467.
  • 29
    Pastewka, L.; Malola, S.; Moseler, M.; Koskinen, P.; J. Power Sources 2013, 239, 321.
  • 30
    Wang, C.; Feng, C.; Gao, Y. J.; Ma, X. X.; Wu, Q. H.; Wang, Z.; Chem. Eng. J. 2011, 173, 92.
  • 31
    Travlou, N. E.; Kyzas, G. Z.; Lazaridis, N. K.; Deliyanni, E. A.; Chem. Eng. J. 2013, 217, 256.
  • 32
    Kyzas, G. Z.; Travlou, N. A.; Kalogirou, O.; Deliyanni, E. A.; Materials 2013, 6, 1360.
  • 33
    He, G. Y.; Zhang, J. G.; Zhang, Y.; Chen, H. Q.; Wang, X.; J. Dispersion Sci. Technol 2013, 34, 1223.
  • 34
    Luo, Y. B.; Shi, Z. G.; Gao, Q.; Feng, Y. Q.; J. Chromatogr. A 2011, 1218, 1353.
  • 35
    Zhao, G. Y.; Song, S. J.; Wang, C.; Wu, Q. H.; Wang, Z.; Anal. Chim. Acta 2011, 708, 155.
  • 36
    Zhang, S.; Du, Z.; Li, G.; Talanta 2013, 115, 32.
  • 37
    Wu, X. L.; Hong, H. J.; Liu, X. T.; Guan, W. B.; Meng, L. X.; Ye, Y.; Ma, Y. Q.; Sci. Total Environ. 2013, 444, 224.
  • 38
    Hummers, W. S.; Offeman, R. E.; J. Am. Chem. Soc. 1958, 80, 1339.
  • 39
    Stankovich, S.; Dikin, D. A.; Piner, R. D.; Kohlhaas, K. A.; Kleinhammes, A.; Jia, Y.; Wu, Y.; Nguyen, S. T.; Ruoff, R. S.; Carbon 2007, 45, 1558.
  • 40
    Baskaran, S.; Kookana, R. S.; Naidu, R.; J. Chromatogr. A 1997, 787, 271.
  • 41
    DiCorcia, A.; Crescenzi, C.; Guerriero, E.; Samperi, R.; Environ. Sci. Technol. 1997, 31, 1658.
  • 42
    Zhou, Q. X.; Xiao, J. P.; Wang, W. D.; Liu, G. G.; Shi, Q. Z.; Wang, J. H.; Talanta 2006, 68, 1309.
  • 43
    Zhou, Q. X.; Ding, Y. J.; Xia, J. P.; Chromatographia 2007, 65, 25.
  • 44
    Zhou, Q. X.; Xiao, J. P.; Ding, Y. J.; Anal. Chim. Acta 2007, 602, 223.
  • 45
    Min, G.; Wang, S.; Zhu, H. P.; Fang, G. Z.; Zhang, Y.; Sci. Total Environ. 2008, 396, 79.
  • 46
    Zhao, R. S.; Yuan, J. P.; Jiang, T.; Shi, J. B.; Cheng, C. G.; Talanta 2008, 76, 956.

Publication Dates

  • Publication in this collection
    Jan 2015

History

  • Received
    04 July 2014
  • Accepted
    30 Sept 2014
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br