Characterization of Polylactide-Stabilized Gold Nanoparticles and Its Application in the Fabrication of Electrochemical DNA Biosensors

Abstract

In this work, two different approaches to gold nanoparticles (AuNPs) have been explored for the modification of screen-printed electrode based on AuNPs and AuNPs stabilized with polylactic acid (PLA). The modified substrate has been characterized using UV-Vis spectroscopy, X-ray diffraction (XRD), transmission electron microscopy (TEM), field emission scanning electron microscopy (FESEM) and cyclic voltammetry. Both synthesized AuNPs were studied in terms of stability, sensitivity and reproducibility to enhance the sensing capability of modified electrodes. The PLA-stabilized AuNPs form strong structured nanoparticles and stabilize in aqueous solution. A larger active surface area (0.41 cm2) and lower charge transfer resistance (Rct) value were reported in the modification of sensing material with PLA-stabilized AuNPs, which resulted in enhancement of sensitivity. Therefore, AuNPs in PLA can be used as a potential alternative modifier for sensing chemicals and biomolecules in electrochemical sensors.

Keywords:
electrochemical DNA biosensor; screen-printed carbon electrode; polylactic acid; gold nanoparticles


Introduction

Research in the development of deoxyribonucleic acid (DNA) biosensor based on the hybridization process has enormously increased in recent years due to its importance towards various applications including environmental monitoring,11 Xu, H. B.; Ye, R. F.; Yang, S. Y.; Li, R.; Yang, X.; Chin. Chem. Lett. 2014, 25, 29.,22 Li, M.; Gou, H.; Al Ogaidi, I.; Wu, N.; ACS Sustainable Chem. Eng. 2013, 1, 713. clinical diagnosis,33 Wang, Z.; Fan, Y.; Chen, J.; Guo, Y.; Wu, W.; He, Y.; Xu, L.; Fu, F.; Electrophoresis 2013, 34, 2177.,44 Hong, C. Y.; Chen, X.; Liu, T.; Li, J.; Yang, H. H.; Chen, J. H.; Chen, G. N.; Biosens. Bioelectron. 2013, 50, 132. gene analysis,55 Zhao, H.; Lu, Z. H.; Chin. Chem. Lett. 2014, 25, 1559.,66 Luo, C.; Tang, H.; Cheng, W.; Yan, L.; Zhang, D.; Ju, H.; Ding, S.; Biosens. Bioelectron. 2013, 48, 132. and food safety.77 Vetrone, S. A.; Huarng, M. C.; Alocilja, E. C.; Sensors 2012, 12, 10487. A lot of methods have been explored to determine DNA such as fluorometry,88 Bai, W.; Zheng, H.; Long, Y.; Mao, X.; Gao, M.; Zhang, L.; Anal. Sci. 2011, 27, 243. chromatography,99 Pereira, L. R.; Prazeres, D. M. F.; Mateus, M.; J. Sep. Sci. 2010, 33, 1175. spectrophotometry,1010 Lin, H.; Li, M. X.; Liu, F.; Mihailovic, D.; Chin. Chem. Lett. 2014, 25, 645. chemiluminescence1111 Luo, M.; Chen, X.; Zhou, G.; Xiang, X.; Chen, L.; Ji, X.; He, Z.; Chem. Commun. 2012, 48, 1126. and electrochemical methods.1212 Lazerges, M.; Bedioui, F.; Anal. Bioanal. Chem. 2013, 405, 3705. Electrochemical methods gained much attention among other methods due to their high sensitivity, low cost instrumentation, fast response, simplicity and portability in detection.1313 Liu, A.; Wang, K.; Weng, S.; Lei, Y.; Lin, L.; Chen, W.; Lin, X.; Chen, Y.; TrAC, Trends Anal. Chem. 2012, 37, 101.

The sensitivity of electrochemical DNA sensors highly depends on the signal amplification; thus, suitable sensing materials must be carefully applied.1414 Shi, A.; Wang, J.; Han, X.; Fang, X.; Zhang, Y.; Sens. Actuators, B 2014, 200, 206. Unique properties of nanomaterials have contributed to unlimited options of sensing materials. Interestingly, gold nanoparticles (AuNPs), as one of the widely used nanomaterials, differ from their bulk counterparts, and have various interesting spectroscopic, electronic and chemical properties, including their high surface-to-volume ratio, nanometer size and ability to couple with electromagnetic wave or with surface plasmon of neighboring metal particles.1515 Daniel, M. C.; Astruc, D.; Chem. Rev. 2004, 104, 293. Hence, they offer an open biological assembly platform to enhance the electron-transfer process between biological molecules and substrate.1616 Karhanek, M.; Kemp, J. T.; Pourmand, N.; Davis, R. W.; Webb, C. D.; Nano Lett. 2005, 5, 403. Organic linker molecules such as thiols and polymers are commonly used to fabricate AuNPs-modified electrodes by assembling AuNPs on electrode surfaces.

Polylactic acid (PLA) is one of the biodegradable polymers that have recently gained attention in bioanalysis areas due to their special characteristics including subcellular size, biocompatibility as well as controlled and sustained release properties.1717 Babu, A.; Templeton, A. K.; Munshi, A.; Ramesh, R.; J. Nanomater. 2013, 2013, 1.,1818 Pridgen, E. M.; Langer, R.; Farokhzad, O. C.; Nanomedicine 2007, 2, 669. Highly stable PLA-AuNPs have been proven to significantly improve both surface area and particle size of modified electrodes.1919 Chirea, M.; García-Morales, V.; Manzanares, J. A.; Pereira, C.; Gulaboski, R.; Silva, F.; J. Phys. Chem. B. 2005, 109, 21808. Earlier research has reported that stabilization of AuNPs takes place between the carboxylic groups of polymer matrix.2020 Chen, H. Y.; Huang, B. H.; Lin, C. C.; Macromolecules 2005, 38, 5400. Practically, the application of screen-printed carbon electrode (SPCE) as a disposable sensor for detection of DNA hybridization is attractive since it allows low-cost electrochemical biosensor. PLA-stabilized AuNPs modified electrodes have been described previously as successful and practical sensor identifying AsIII and sulfide in water.2121 Song, Y. S.; Muthuraman, G.; Zen, J. M.; Electrochem. Commun. 2006, 8, 1369.,2222 Song, Y. S.; Muthuraman, G.; Chen, Y. Z.; Lin, C. C.; Zen, J. M.; Electroanalysis 2006, 18, 1763. Although the same electrode modification was extended in our study, we used a different approach for preparation of PLA-stabilized AuNPs-based electrochemical sensor. To the best of our knowledge, there is no record on comparative study between AuNPs and PLA-stabilized AuNPs, and their properties as modified electrodes. Therefore, this work reports the characteristics of AuNPs and PLA-stabilized AuNPs as a modified substrate for construction of electrochemical sensors. Properties of modified electrode and its preliminary application in DNA biosensors were also described in detail.

Experimental

Materials and instruments

PLA with the molecular weight of 320,000 Da was used for stabilization of AuNPs. CHCl3 was purchased from Merck and used as a solvent. HAuCl4.3H2O and sodium citrate were purchased from Sigma-Aldrich. All reagents were of analytical grade and used without further purification. µAutolab III voltammetric analyzer (Eco-Chemie) incorporated with general purpose electrochemical system (GPES) 4.9 software was used to analyze samples. The characterization of AuNPs was studied using UV-Vis spectrophotometer (Shimadzu UV-Vis-1650), X-ray diffraction (XRD) (Philips PW 3040/60 MPD X'PERT Pro Analytical), transmission electron microscopy (TEM) (Hitachi H-7100) and field-emission scanning electron microscopy (FESEM) with energy dispersive X-ray (EDX) analyzer (JEOL JSM 7600F).

Preparation and characterization of AuNPs

AuNPs were synthesized according to the method reported by Nasir and Nur.2323 Nasir, S.; Nur, H.; J. Fund. Sci. 2008, 4, 245. Briefly, 10 mL of sodium citrate solution (38.8 mmol L-1) were added to 100 mL of aqueous chloroauric solution (1 mmol L-1) under boiling. The color of solution first changed from yellow to blackish and finally to dark ruby red after about 10 min. The solution was then cooled to ambient temperature and kept in fridge.

Preparation and characterization of PLA-stabilized AuNPs

Prior to preparation of PLA sheet, PLA pellets were placed in stainless steel mould, preheated for melting at 190 ºC for 10 min and then pressed under pressure of 2.2 MPa for 1 min. The melted PLA was cooled and ready for usage. About 0.68 mg of PLA sheets were cut into small pieces, dissolved in 5 mL of chloroform and stirred vigorously in room temperature, and then 10 mL of AuNPs solution were added to the dissolved PLA. The mixture was stirred in room temperature until completely homogeneous. The characterization of PLA-stabilized AuNPs was determined with the same instruments as described in AuNPs characterization.

Preliminary study of biosensor fabrication

A volume of 25 µL of homogenous AuNPs and PLA-stabilized AuNPs solution was pipetted on SPCE and dried at ambient temperature for 24 h prior to use. For preliminary study, the effect of six different types of supporting electrolytes at 0.1 mol L-1 including phosphate buffer, potassium chloride, potassium dihydrogen phosphate (KH2PO4), phosphate-buffered saline (PBS), Tris-EDTA and Tris-HCl were studied on the sensitivity of electrochemical biosensor. The effect of pH was studied in the range of 2 to 12. All experiments were recorded by cyclic voltammetry (CV) in the potential range of -0.8 to 0.3 V for 3 cycles. The performance of the developed biosensor was investigated by CV analysis in the presence of 20 µmol L-1 methylene blue (MB).

Results and Discussions

Characterization of AuNPs and PLA-stabilized AuNPs

The formation of AuNPs was indicated by the color change of aqueous solution from light yellow to dark ruby red in the presence of sodium citrate. According to Supplementary Information Figure S1, absorption peaks of AuNPs and PLA-stabilized AuNPs solutions are in the wavelength range of 520 to 540 nm, showing the presence of AuNPs in the solution. The absorbance of AuNPs and PLA-stabilized AuNPs appears at wavelengths ranging from 500 to 600 nm depending on their particle size.2424 Seoudi, R.; Fouda, A. A.; Elmenshawy, D. A.; Phys. B 2010, 405, 906. Figure 1 shows XRD pattern of synthesized AuNPs and PLA-stabilized AuNPs. The strong diffraction peaks with 2θ values located at 38.2, 44.4, 64.7 and 77.7º correspond to Bragg's reflection from (111), (200), (220) and (311) planes of the face-centered cubic (fcc) metal gold nanostructures. These results correspond with previous reports on AuNPs by Rastogi and Arunachalam.2525 Rastogi, L.; Arunachalam, J.; Adv. Mater. Lett. 2013, 4, 548. The broadening of XRD peaks were indicated on formation of nanosized PLA-stabilized AuNPs.

Figure 1
X-Ray diffraction spectra of AuNPs and PLA-stabilized AuNPs.

Furthermore, all diffraction peaks of AuNPs appeared on the PLA-stabilized AuNPs indicating that the structure of AuNPs is stable in the presence of polylactide. Figure 1 shows some new diffraction peaks for PLA-stabilized AuNPs at 2θ regions of 32º, which indirectly indicate the good crystallization of polymers (PLA) on the surface of AuNPs. The crystalline sizes of AuNPs in Table 1 were determined from (111) plane by the application of Scherrer's equation as:

(1)

Table 1
Crystal size of AuNPs and PLA-stabilized AuNPs

where L is the crystallites mean dimension, k is a constant equal to 0.94, λ is Cu Kα radiation wavelength (0.154 nm), β is the full width at half maximum (FWHM) of diffraction peak and θ is the diffraction angle. The calculated crystalline sizes of AuNPs are 32 and 27 nm for AuNPs and PLA-stabilized AuNPs samples, respectively. It was observed that FWHM is inversely proportional to the crystalline size. The smaller size of PLA-stabilized AuNPs might be due to the compression of AuNPs by a fairly thin PLA layer on the surface. These results clearly indicate that the PLA-stabilized AuNPs form strong structured nanoparticles and stabilize in aqueous solution.

The size and morphology of AuNPs and PLA-stabilized AuNPs were further studied by TEM. The TEM images and particle size frequency distribution curve of AuNPs and PLA-stabilized AuNPs are presented in Figure 2, which exhibit spherical and oval shapes. The shapes displayed in Figure 2a are largely spherical and highly dispersed as compared to those in Figure 2b. The nanostructure shapes of Figure 2b are clearly outlined, although the shapes are widely agglomerated. However, these results disagreed with the crystallite sizes found in the XRD results, as they are an estimate value calculated using Scherrer's equation, whereas TEM directly measures the actual particle size. Nanda et al.2626 Nanda, J.; Sapra, S.; Sarma, D. D.; Chandrasekharan, N.; Hodes, G.; Chem. Mater. 2000, 12, 1018. found that the calculated particle size by TEM technique has larger size due to large amount of several crystalline domains oriented in different directions inside a single crystallite. Hence, the sensitivity of XRD to the dimensions of such individual local crystalline domains within a single nanocrystallite projected smaller particle size. In line with that, Rath et al.2727 Rath, C.; Sahu, K. K.; Kulkarni, S. D.; Anand, S.; Date, S. K.; Das, R. P.; Mishra, N. C.; Appl. Phys. Lett. 1999, 75, 4171. also found the smaller size estimated in XRD as compared to TEM, which indicated that particles were polycrystalline as they contained an abundance of smaller crystallites (subparticles) inside them. In addition, Corbierre et al.2828 Corbierre, M. K.; Cameron, N. S.; Lennox, R. B.; Langmuir 2004, 20, 2867. discovered that polymer-stabilized gold nanoparticles have larger diameters and polydispersities.

Figure 2
TEM images of (a) AuNPs and (b) PLA-stabilized AuNPs. Frequency diameter distributions for (c) AuNPs and (d) PLA-stabilized AuNPs. SD: standard deviation.

Figure 3 shows FESEM images and EDX spectra of SPCE surface (5 mm diameter) before and after modification with AuNPs and PLA-stabilized AuNPs. It is clear that the modification of SPCE with PLA-stabilized AuNPs has greater density and weight percentage as well as larger surface coverage of nanoparticles.

Figure 3
FESEM images and EDX spectra of (a) bare SPCE, (b) AuNPs modified SPCE and (c) PLA-stabilized AuNPs modified SPCE.

Electrochemical characterization of PLA-stabilized AuNPs modified SPCE

The electrochemical performance of modified electrode was evaluated using potassium ferrocyanide solution by CV. Due to the fast electrochemical reaction and fine reversibility of ferrocyanide ion ([Fe(CN)6]3-/4-), it has been used for electrochemical study of SPCE after modification with AuNPs in aqueous and organic buffer solutions.

Supporting electrolyte has important role in electrochemical biosensors as it affects the activation of DNA in aqueous solution without interrupting DNA structure. Proper choice of buffer must be studied carefully as it is correlated with the rate of charge transfer. CV of [Fe(CN)6]3-/4- ion at modified SPCE surface was significantly studied at different kinds of supporting electrolytes. Based on Figure S2, PBS (0.1 mol L-1) has higher sensitivity and was selected as working electrolyte for DNA sensors based on PLA-stabilized AuNPs. PBS is stable at various temperatures, is non-toxic to DNA, has microbial resistance and can be stored for several weeks.

pH of solution plays an important role in providing appropriate environment for DNA to operate efficiently. However, DNA molecules may denature at high pH values. From the results shown in Figure S3, by increasing pH from 2.0 to 7.0, the anodic peak current of [Fe(CN)6]3-/4- ion was increased and reached the highest at pH 7.0, so it was chosen as the optimum working pH for PLA-stabilized AuNPs modified DNA sensors.

Effect of scan rate

Based on the CV obtained (Figure 4), the cathodic peak current of [Fe(CN)6]3-/4- ion significantly increased in the scan rate of 0.01 to 1.0 V s-1. Lower scan rate shows more reversible voltammograms with smaller charging current. Hence, scan rate of 0.1 V s-1 was chosen to be applied in the electrochemical study as the electrochemical reaction was reversible.

Figure 4
Cyclic voltammograms of [Fe(CN)6]3-/4- at different scan rates: 1.0, 0.5, 0.1, 0.05 and 0.01 V s-1 at potential range of 1 to -1 V.

Properties of modified electrode

The property of PLA-stabilized AuNPs modifier was assessed by calculating the active surface area using Randles-Sevcik equation as:

(2)

where ipc is the cathodic peak current, n is the number of electrons participating in the redox reaction, A is the activate surface area of electrode (cm2), D is diffusion coefficient of K3Fe(CN)6 in solution (7.6 × 10-6 cm2 s-1), C is the concentration of K3Fe(CN)6 (mol L-1) and ν is scan rate (V s-1). The active surface area was calculated as 0.26 and 0.31 cm2 for bare SPCE and SPCE/AuNPs, respectively, while it was 0.41 cm2 after modification with PLA-stabilized AuNPs. In fact, the active surface area for modified PLA-stabilized AuNPs electrode was higher (Figure 5). This result implies that the application of PLA-stabilized AuNPs as electrochemical sensing material can improve the active surface area better than AuNPs. This result is consistent with previous work of polymer-embedded gold nanoparticles, which demonstrates that higher active surface area showed better electron-transfer process, resulting in higher sensitivity.2929 Manivannan, S.; Ramaraj, R.; Pure Appl. Chem. 2011, 83, 2041. The electrochemical reaction of modified SPCE is under diffusion control process, which was indicated by the dependence of reduction peak current of K3[Fe(CN)6]3-/4- on square root of scan rate (ν1/2).

Figure 5
Cyclic voltammograms and plot of oxidation peak currents against square root of scan rate for (a) bare SPCE, (b) SPCE/AuNPs and (c) SPCE/PLA-stabilized AuNPs in 1.0 mmol L-1 K3[Fe(CN)6]. The scan rates were 50, 100, 150, 200, 250, 300 and 350 mV s-1.

Electrochemical impedance spectroscopy (EIS) (Figure 6) shows that the charge transfer resistance value of the bare SPCE is 1932 Ω, while after modification with AuNPs (SPCE/AuNPs) and PLA-stabilized AuNPs, the resistance decreased to 1730 and 1444 Ω, respectively. The EIS results are in agreement with the CV measurements as shown in Figure S4, which clearly exhibits the increase of the peak currents after modification with AuNPs and PLA-stabilized AuNPs. This indicates that the higher sensitivity is inversely proportional to the lower charge transfer resistance.3030 Wu, X.; Jiang, H.; Zheng, J.; Wang, X.; Gu, Z.; Chen, C.; J. Electroanal. Chem. 2011, 656, 174.

Figure 6
EIS study of modified electrodes in 1.0 mmol L-1 K3[Fe(CN)6]. Frequency range: 0.1-105 Hz, amplitude: 5 mV.

The repeatability of constructed electrodes was investigated by repetitive recording of CV of K3[Fe(CN)6]3-/4- for six cycles. The relative standard deviations (RSD, %) were determined as 4.26 and 9.97% for SPCE/PLA-stabilized AuNPs and SPCE/AuNPs, respectively, showing that the fabricated electrochemical sensor based on PLA-stabilized gold nanoparticles is more reusable and can be used for several times. Moreover, six modified electrodes were fabricated independently with the same procedure, which gave a satisfactory RSD value of 4.05% on the oxidation peak current of K3[Fe(CN)6]3-/4-, while it was 9.99% for SPCE/AuNPs, showing better reproducibility for PLA-stabilized AuNPs (Table 2). The modified substrate based on PLA-AuNPs can be stored at room temperature for a long time and can be used for recording 20 CV with 18% signal reduction while SPCE/AuNPs substrate has lower stability (38% reduction in peak current after 20 cycles).

Table 2
Characteristics of modified electrodes based on AuNPs and PLA-stabilized AuNPs

Fabrication of electrochemical DNA biosensors based on PLA-AuNPs and AuNPs modified SPCE

Figure 7 shows the fabrication process of PLA-AuNPs and AuNPs modified SPCE. The working electrode substrate on SPCE was drop cast with 25 µL of homogenous AuNPs and PLA-stabilized AuNPs solution, then dried in ambient temperature for 24 h. The modified SPCE was denoted as AuNPs/SPCE and PLA-AuNPs/SPCE. Self-assembled monolayer technique based on thiol-AuNPs bonding process was used for bonding probe single-stranded DNA (ssDNA) on the surface of AuNPs and PLA-AuNPs modified electrode by bonding AuNPs to thiol groups. For hybridization process, the complementary target DNA was introduced into the probe DNA-AuNPs/SPCE and DNA/PLA-AuNPs/SPCE that converted into electrochemical signal using MB as a redox indicator, which was able to incorporate at DNA surface. The performance of fabricated biosensor was evaluated using PBS (pH 6.8) at 100 mV s-1 by CV after 30 min of incubation time for MB.

Figure 7
Fabrication process of PLA-AuNPs and AuNPs modified screen-printed carbon electrode for electrochemical biosensor.

In this study, performance of the developed biosensor was further investigated in the presence of MB as a redox label, which is widely known to have an affinity towards DNA. In general, the degree of MB immersion time was studied in the range of 1.0 to 180 min. The immersion time of 30 min was chosen as the optimum duration to enhance sensitivity. There are two possible mechanisms of MB interaction, namely with an interaction between guanine bases which are mainly in ssDNA, or by an electrostatic interaction with the negatively charged phosphate groups.3131 Kara, P.; Kerman, K.; Ozkan, D.; Meric, B.; Erdem, A.; Ozkan, Z.; Ozsoz, M.; Electrochem. Commun. 2002, 4, 705. The advantage of MB utilization in the electrochemical reaction happens at comparatively low potential where the interferences from other electroactive species and background can be minimized compared to detection with label-free electrochemical method.

Figure 8 indicates representative CV peak current obtained at PLA-AuNPs and AuNPs modified SPCE coated with the same probe DNA concentration and complementary target DNA. In order to get higher sensitivity by CV, SPCE was pre-treated using 0.5 mol L-1 acetate buffer solution (ABS) and 20 mmol L-1 NaCl (pH 4.53), following the modified method from Issa et al.3232 Issa, R.; Hamdan, N. A.; Noh, M. F. M.; Biotechnology 2010, 9, 304. prior to measurement throughout the experiment. An increment in MB peak current happened after immobilization of thiolated single stranded DNA onto PLA-AuNPs and AuNPs modified electrode. Nevertheless, the peak current of MB significantly decreased about 1-fold lower upon hybridization with target DNA. This finding is in correlation with Khezrian et al.3333 Khezrian, S.; Salimi, A.; Teymourian, H.; Hallaj, R.; Biosens. Bioelectron. 2013, 43, 218. and Mao et al. ,3434 Mao, K.; Wu, D.; Li, Y.; Ma, H.; Ni, Z.; Yu, H.; Luo, C.; Wei, Q.; Du, B.; Anal. Biochem. 2012, 422, 22. who also reported high MB accumulation on double-stranded DNA surface. This result shows that the detection of target DNA by the fabricated electrochemical DNA biosensor was enabled by MB utilization as redox indicator. The electrode modification with PLA-stabilized AuNPs shows higher sensitivity and stability in comparison with AuNPs alone on the surface of SPCE.

Figure 8
Peak current of DNA detection using fabricated biosensor after 30 min incubation in 20 µmol L-1 MB in 0.1 mol L-1 PBS (pH 6.8) at 100 mV s-1.

Conclusions

In this paper, two different gold nanoparticles were synthesized as modified substrates for using in electrochemical DNA biosensors. AuNPs stabilized with polylactic acid were found to be more stable and exhibited better characteristics as a modifier. The SPCE/PLA-stabilized AuNPs increases the sensitivity due to the larger active surface area produced and lower Rct value. The stability, reproducibility and repeatability of SPCE/PLA-stabilized AuNPs revealed satisfying results. The preliminary study shows that PLA-stablized gold nanoparticles is a reliable substrate for binding thiolated DNA oligumers in electrochemical DNA biosensors. Hence, the application of PLA-stabilized AuNPs has great potential to be applied as modifier for further sensor fabrication.

Supplementary Information

Supplementary information is available free of charge at http://jbcs.sbq.org.br as PDF file.

Acknowledgments

The authors gratefully acknowledged the financial support of Universiti Putra Malaysia and all available facilities in the Department of Chemistry, Faculty of Science.

References

  • 1
    Xu, H. B.; Ye, R. F.; Yang, S. Y.; Li, R.; Yang, X.; Chin. Chem. Lett 2014, 25, 29.
  • 2
    Li, M.; Gou, H.; Al Ogaidi, I.; Wu, N.; ACS Sustainable Chem. Eng 2013, 1, 713.
  • 3
    Wang, Z.; Fan, Y.; Chen, J.; Guo, Y.; Wu, W.; He, Y.; Xu, L.; Fu, F.; Electrophoresis 2013, 34, 2177.
  • 4
    Hong, C. Y.; Chen, X.; Liu, T.; Li, J.; Yang, H. H.; Chen, J. H.; Chen, G. N.; Biosens. Bioelectron 2013, 50, 132.
  • 5
    Zhao, H.; Lu, Z. H.; Chin. Chem. Lett 2014, 25, 1559.
  • 6
    Luo, C.; Tang, H.; Cheng, W.; Yan, L.; Zhang, D.; Ju, H.; Ding, S.; Biosens. Bioelectron 2013, 48, 132.
  • 7
    Vetrone, S. A.; Huarng, M. C.; Alocilja, E. C.; Sensors 2012, 12, 10487.
  • 8
    Bai, W.; Zheng, H.; Long, Y.; Mao, X.; Gao, M.; Zhang, L.; Anal. Sci 2011, 27, 243.
  • 9
    Pereira, L. R.; Prazeres, D. M. F.; Mateus, M.; J. Sep. Sci 2010, 33, 1175.
  • 10
    Lin, H.; Li, M. X.; Liu, F.; Mihailovic, D.; Chin. Chem. Lett 2014, 25, 645.
  • 11
    Luo, M.; Chen, X.; Zhou, G.; Xiang, X.; Chen, L.; Ji, X.; He, Z.; Chem. Commun 2012, 48, 1126.
  • 12
    Lazerges, M.; Bedioui, F.; Anal. Bioanal. Chem 2013, 405, 3705.
  • 13
    Liu, A.; Wang, K.; Weng, S.; Lei, Y.; Lin, L.; Chen, W.; Lin, X.; Chen, Y.; TrAC, Trends Anal. Chem 2012, 37, 101.
  • 14
    Shi, A.; Wang, J.; Han, X.; Fang, X.; Zhang, Y.; Sens. Actuators, B 2014, 200, 206.
  • 15
    Daniel, M. C.; Astruc, D.; Chem. Rev 2004, 104, 293.
  • 16
    Karhanek, M.; Kemp, J. T.; Pourmand, N.; Davis, R. W.; Webb, C. D.; Nano Lett 2005, 5, 403.
  • 17
    Babu, A.; Templeton, A. K.; Munshi, A.; Ramesh, R.; J. Nanomater 2013, 2013, 1.
  • 18
    Pridgen, E. M.; Langer, R.; Farokhzad, O. C.; Nanomedicine 2007, 2, 669.
  • 19
    Chirea, M.; García-Morales, V.; Manzanares, J. A.; Pereira, C.; Gulaboski, R.; Silva, F.; J. Phys. Chem. B 2005, 109, 21808.
  • 20
    Chen, H. Y.; Huang, B. H.; Lin, C. C.; Macromolecules 2005, 38, 5400.
  • 21
    Song, Y. S.; Muthuraman, G.; Zen, J. M.; Electrochem. Commun 2006, 8, 1369.
  • 22
    Song, Y. S.; Muthuraman, G.; Chen, Y. Z.; Lin, C. C.; Zen, J. M.; Electroanalysis 2006, 18, 1763.
  • 23
    Nasir, S.; Nur, H.; J. Fund. Sci. 2008, 4, 245.
  • 24
    Seoudi, R.; Fouda, A. A.; Elmenshawy, D. A.; Phys. B 2010, 405, 906.
  • 25
    Rastogi, L.; Arunachalam, J.; Adv. Mater. Lett 2013, 4, 548.
  • 26
    Nanda, J.; Sapra, S.; Sarma, D. D.; Chandrasekharan, N.; Hodes, G.; Chem. Mater 2000, 12, 1018.
  • 27
    Rath, C.; Sahu, K. K.; Kulkarni, S. D.; Anand, S.; Date, S. K.; Das, R. P.; Mishra, N. C.; Appl. Phys. Lett 1999, 75, 4171.
  • 28
    Corbierre, M. K.; Cameron, N. S.; Lennox, R. B.; Langmuir 2004, 20, 2867.
  • 29
    Manivannan, S.; Ramaraj, R.; Pure Appl. Chem 2011, 83, 2041.
  • 30
    Wu, X.; Jiang, H.; Zheng, J.; Wang, X.; Gu, Z.; Chen, C.; J. Electroanal. Chem 2011, 656, 174.
  • 31
    Kara, P.; Kerman, K.; Ozkan, D.; Meric, B.; Erdem, A.; Ozkan, Z.; Ozsoz, M.; Electrochem. Commun 2002, 4, 705.
  • 32
    Issa, R.; Hamdan, N. A.; Noh, M. F. M.; Biotechnology 2010, 9, 304.
  • 33
    Khezrian, S.; Salimi, A.; Teymourian, H.; Hallaj, R.; Biosens. Bioelectron 2013, 43, 218.
  • 34
    Mao, K.; Wu, D.; Li, Y.; Ma, H.; Ni, Z.; Yu, H.; Luo, C.; Wei, Q.; Du, B.; Anal. Biochem 2012, 422, 22.