A Systematic Procedure to Develop a Capillary Electrophoresis Method Using a Minimal Experimental Data

Abstract

The choice of an appropriate background electrolyte (BGE) and its components for capillary electrophoresis analysis is the main step in capillary electrophoresis method development. The use of an inadequate co-ion component could lead to asymmetrical peaks and selecting an inappropriate counter-ion could affect the buffer capacity and the pH of the BGE, leading to unreliable analysis. In this paper, we describe a systematic procedure for the development of a capillary electrophoresis method, based on the effect of varying pH on the ion effective mobility, to optimize the BGE composition. The method was applied to the separation of L-ascorbic acid in different samples. The optimized background electrolyte composition was 40 mmol L-1 tris(hydroxymethyl)aminomethane and 20 mmol L-1 2-morpholinoethanesulfonic acid, at pH 8.1. Sorbic acid was used as the internal standard and separation was carried out in a fused-silica capillary (32 cm total length and 8.5 cm effective length, 50 µm inner diameter), with a short-end-injection configuration and direct ultraviolet (UV) detection at 266 nm. The separation was performed in 26 s. The method shows good linearity (R2 > 0.999), excellent values for inter-day and intra-day precision and good recovery (in the range of 94-107%). The values obtained for limit of detection (LOD) and limit of quantification (LOQ) were 0.14 and 41 mg L-1, respectively. The systematic procedure applied shows to be a very useful tool for the first step method development for capillary electrophoresis.

Keywords:
capillary electrophoresis; effective mobility curves; optimization; L-ascorbic acid


Introduction

Capillary zone electrophoresis (CZE) is the simplest and most widely used form of capillary electrophoresis (CE).11 Terabe, S.; Otsuka, K.; Nishi, H.; J. Chromatogr. A 1994, 666, 295. In this mode, a capillary is filled with an electrolyte (run buffer), the sample is introduced in the inlet and an electrical field is applied. Separations are accomplished through the movement of ions in the electric field. This movement is governed by the electrophoretic mobility of the ion (µe), this mobility being defined as the average velocity in which an ion moves under the influence of an applied potential field (normalized to the electric field strength).22 Landers, J. P.; Handbook of Capillary and Microchip Electrophoresis and Associated Microtechniques, 3rd ed.; CRC Press: Boca Raton, 2008. The difficulty, however, lies with determining the optimal conditions of the background electrolyte (BGE), which the charge-to-size ratio of the analytes is appreciably different from one another.33 Johns, K. F.; Breadmore, M. C.; Brudo, R.; Haddad, P. R.; Electrophoresis 2009, 30, 839.

The choice of buffer is critical to obtaining the successful separation of analytes by CZE. Once the optimal wavelength for detection has been established, a buffer that does not interfere with the ability to detect the analytes of interest must be selected, maintaining the solubility of the analytes, the buffering capacity through the analysis, and producing the desired separation. Since a variation in the pH of the BGE could affect either the electroosmotic flow (EOF) or effective mobility (meff) of the solute, successful separations are usually obtained when the conditions optimize both the EOF and the solute mobility properties.22 Landers, J. P.; Handbook of Capillary and Microchip Electrophoresis and Associated Microtechniques, 3rd ed.; CRC Press: Boca Raton, 2008.

A good approach is to start with the use of a simple and effective tool: the effective mobility vs. pH curve, that can be constructed knowing µ0 and the Ka values for each compound, and can provide some important information, such as the pH range in which separation can occur, that is, where the effective mobility of each analyte is different. By considering that good peak, symmetry is usually obtained when the analyte mobility matches the co-ion mobility, an appropriate co-ion can also be selected from the inspection of the mobility vs. pH curves. The applicability of these curves has been already demonstrated for different kind of matrix as soft drinks and sweeteners,44 Vistuba, J. P.; Dolzan, M. D.; Vitali, L.; de Oliveira , M. A. L.; Micke, G. A.; J. Chromatogr. A 2015, 1396, 148. biodiesel,55 Piovezan, M.; Costa, A. C. O.; Jager, A. V.; de Oliveira, M. A. L.; Micke, G. A.; Anal. Chim. Acta 2010, 673, 200. biological samples,66 Vitali, L.; Favere, V. T.; Micke, G. A.; J. Chromatogr. A 2011, 1218, 2327. etc.

L-Ascorbic acid (L-AA, vitamin C) is a natural antioxidant present in food and biological systems with important nutritional benefits for human health.77 Meister, A.; Biochem. Pharmacol. (Amsterdam, Neth.) 1992, 44, 1905. Fruit juices, pharmaceutical formulations and vitamin supplements (e.g., tablets and powdered juices containing L-AA) are sources of vitamin C. The determination of L-AA in these samples is very important with regard to the standardization of nutrient composition and to monitoring the L-AA stability during processing and storage.

Several different analytical methods have been employed to study L-AA including cyclic voltammetric,88 Nasirizadeh, N.; Shekari, Z.; Tabatabaee, M.; Ghaani, M.; J. Braz. Chem. Soc. 2015, 26, 713. chemiluminescence,99 Alwarthan, A. A.; Analyst (Cambridge, U. K.) 1993, 118, 639. high-performance liquid chromatography (HPLC),1010 Bottegal, M.; Lang, L.; Miller, M.; McCord, B.; Rapid Commun. Mass Spectrom. 2010, 24, 1377. fluorimetry,1111 Park, H. W.; Alam, S. M.; Lee, S. H.; Karim, M. M.; Wabaidur, S. M.; Kang, M.; Choi, J. H.; Luminescence 2009, 24, 367. microchip electrophoresis1212 Fischer, D. J.; Hulvey, M. K.; Regel, A. R.; Lunte, S. M.; Electrophoresis 2009, 30, 3324.,1313 Zhao, S. L.; Huang, Y.; Liu, Y. M.; J. Chromatogr. A 2009, 1216, 6746. and capillary electrophoresis.1414 Dong, S. Q.; Zhang, S.; Cheng, X.; He, P. G.; Wang, Q. J.; Fang, Y. Z.; J. Chromatogr. A 2007, 1161, 327.

15 Wang, X. L.; Li, L. J.; Li, Z. Y.; Wang, J.; Fu, H. Y.; Chen, Z. Z.; J. Electroanal. Chem. 2014, 712, 139.
-1616 Marra, M. C.; Silva, P. L.; Munoz, R. A. A.; Ritcher, E. M.; J. Braz. Chem. Soc. 2014, 25, 913. The CE technique offers attractive advantages including simple sample preparation, low consumption of chemical reagents, good resolution, reduced residue generation and reduced analysis time. Several previous publications report methods using capillary zone electrophoresis and a basic borate buffer (pH 8.0-9.2) as the BGE for the determination of L-AA in pharmaceutical formulations,1616 Marra, M. C.; Silva, P. L.; Munoz, R. A. A.; Ritcher, E. M.; J. Braz. Chem. Soc. 2014, 25, 913. beverages,1717 Tang, Y. J.; Wu, M. J.; Talanta 2005, 65, 794.,1818 Choi, O. K.; Jo, J. S.; J. Chromatogr. A 1997, 781, 435. vegetables,1919 Fukushi, K.; Takeda, S.; Wakida, S.; Yamane, M.; Higashi, K.; Hiiro, K.; J. Chromatogr. A 1997, 772, 313.,2020 Davey, M. W.; Bauw, G.; VanMontagu, M.; Anal. Biochem. 1996, 239, 8. fruits2121 Galiana-Balaguer, L.; Rosello, S.; Herrero-Martinez, J. M.; Maquieira, A.; Nuez, F.; Anal. Biochem. 2001, 296, 218 and biological fluids.2222 Zinellu, A.; Carru, C.; Sotgia, S.; Deiana, L.; Anal. Biochem. 2004, 330, 298.

In this work, a simple systematic procedure based on effective mobility vs. pH curves is shown aiming a method development with minimal experimental data. The procedure was applied in the development of a fast method for the determination of L-AA vitamin supplements (liquid and effervescent tablet preparations) and in orange juice.

Experimental

All experiments were performed on an Agilent Technologies HP3DCE instrument (Palo Alto, CA, USA), equipped with a diode array detector set at 266 nm. The measurements were performed at 25 °C in an uncoated fused-silica capillary 32 cm × 50 µm i.d. × 375 µm o.d. (effective length 8.5 cm) obtained from Polymicro Technologies (Phoenix, AZ, USA). Between runs, the capillary was rinsed for 1 min with running buffer. Standard solutions and samples were introduced from the outlet capillary extremity and injected hydrodynamically at 50 mbar for 4 s (50 mbar = 4996.2 Pa) with negative pressure. The applied separation voltage was 30 kV, with positive polarity (anode) on the injection side. Data acquisition and treatment were performed with HP ChemStation software (Palo Alto, CA, USA).

Reagents and solutions

All chemicals used in the buffer preparation were of analytical reagent grade. Chromatographic grade acetonitrile (ACN), methanol (MeOH) and sodium hydroxide were obtained from Merck (Darmstadt, Germany), tris(hydroxymethyl)aminomethane (Tris), 2-morpholinoethanesulfonic acid (MES), L-ascorbic acid and sorbic acid (SA) were purchased from Sigma-Aldrich (St. Louis, MO, USA). Deionized water (Milli-Q deionizer, Millipore, Bedford, MA, USA) was used to prepare the solutions. L-ascorbic acid standard stock solution (1000 mg L-1) was prepared in deionized water daily. Sorbic acid (internal standard, IS) solution (200 mg L-1) was used. Stock solutions of Tris and MES, both of 100 mmol L-1, were used to prepare the background electrolyte. The optimized electrolyte composition was 40 mmol L-1 Tris and 20 mmol L-1 MES, at pH 8.1.

Samples

The samples were purchased from local supermarkets and subjected to sample pretreatment before injection: (i) for effervescent tablets vitamin C, the tablets were macerated to homogenize the sample and 3.551 g was dissolved in 100 mL of deionized water; (ii) for liquid vitamin C, 100 µL of the liquid sample was dissolved in 100 mL of deionized water; and (iii) for orange juice, the juice was extracted from the fruit by pressure and diluted 1:1 (v/v) with deionized water.

A 100 µL aliquot of each sample solution was transferred to a 1.5 mL microtube and 200 µL of deionized water, 100 µL of internal standard (sorbic acid 200 mg L-1) and 600 µL of ACN were added. The tube was centrifuged at 2,000 rpm for 5 min. After centrifugation, 500 µL of the upper phase was transferred to a vial. Each sample was injected into the CE equipment.

Results and Discussion

Method development

The steps of the method development are shown in Figure 1 and can be described as follows.2323 Goncalves, L. C.; Micke, G. A.; J. Chromatogr. A 2007, 1154, 477.,2424 Costa, A. C. O.; da Costa, J. L.; Tonin, F. G.; Tavares, M. F. M.; Micke, G. A.; J. Chromatogr. A 2007, 1171, 140. The first step is to construct the effective mobility (µeff) vs. pH. In the electrophoretic separation of compounds that present weak acid character, at least two species should be considered: the non-ionized acid molecule (with zero mobility) and its conjugated base, an anion. The concept of effective mobility is used to describe the migration behavior of such curves weak electrolytes. By definition, any substance presented in solution as different species, related each other by rapid acid-base equilibrium, migrates in the electric field as a single solute, exhibiting an effective mobility calculated by the following equation:

(1)

Figure 1
Flowchart showing the generic steps for method development.

where α is the molar fraction or distribution function and µ0 is the ionic mobility of each individual species. The values for the absolute mobility and pKa used to construct the curves were obtained from a database compiled by Hirokawa et al.2525 Hirokawa, T.; Nishino, M.; Aoki, N.; Kiso, Y.; Sawamoto, Y.; Yagi, T.; Akiyama, J.; J. Chromatogr. A 1983, 271, D1.

Since the SA is a monoprotic acid and L-AA is monoprotic till pH 9, the distribution function can be expressed in terms of the dissociation constant and the medium acidity (equation 2):

(2)

From these equations and the data compiled by Hirokawa et al.,2525 Hirokawa, T.; Nishino, M.; Aoki, N.; Kiso, Y.; Sawamoto, Y.; Yagi, T.; Akiyama, J.; J. Chromatogr. A 1983, 271, D1. effective mobility vs. pH curves could be constructed (Figure 2).

Figure 2
Effective mobility vs. pH curve for ascorbic acid, sorbic acid (internal standard) and 2-morpholinoethanesulfonic acid (MES, co-ion). Enclosed area represents the best region to choose the optimal pH.

Determination of pH and operation mode BGE composition

The inspection of such curves gives a preliminary assessment of an appropriate pH range where the separation is likely to occur, i.e., where relative differences of mobility are maximized. As it can be inferred from inspection of the Figure 2, above pH 3 the separation of the analytes can be approached experimentally. However, the optimal pH value for the method was determined in the range of 7-9 (enclosed area Figure 2) once; at this pH range, both L-AA and SA will be fully ionized, which means that µeff is higher and differences of mobility, in this case, are maximized.

At this pH, the mode of operation could be co-electroosmotic, which means that analytes and EOF are migrating toward the anode, condition that could be achieved by adding a cationic EOF modifier on the BGE. Another option is to perform the analysis on counter-electroosmotic mode. In this mode, two situations are possible: (i) µeff of the analytes are higher than the EOF mobility, ensuring the anion migration past the detector (i.e., inorganic anions); or (ii) the EOF mobility is higher than the µeff, ensuring that the analytes will be carried toward the detector. Since at the pH range selected the EOF mobility has a high magnitude (40-60 × 10-5 cm2 V-1 S-1),2626 Vitali, L.; Della Betta, F.; Costa, A. C. O.; Vaz, F. A. S.; Oliveira, M. A. L.; Vistuba, J. P.; Favere, V. T.; Micke, G. A.; Talanta 2014, 123, 45. the counter-electroosmotic mode was selected.

Determination of BGE composition

In the selection of an appropriate background electrolyte, several characteristics should be considered, including good buffering capacity and adequate co- and counter-ions, to minimize peak asymmetry and anomalous dispersion effects.2424 Costa, A. C. O.; da Costa, J. L.; Tonin, F. G.; Tavares, M. F. M.; Micke, G. A.; J. Chromatogr. A 2007, 1171, 140. Several important parameters are readily obtained from PeakMaster,2727 Gas, B.; Jaros, M.; Hruska, V.; Zuskova, I.; Stedry, M.; Lc GC Eur. 2005, 18, 282. which allows determination of the electromigration dispersion (EMD), system zone phenomena, effective mobility, transference ratio and analyte parameters.

The first BGE parameter to be selected in this proposed procedure is the co-ion type. As is well established for analysis performed with CZE system using a absorbance detector, good peak symmetry is usually obtained when the analyte mobility matches the co-ion mobility. Due to the similarity of the effective mobility between MES and L-AA and SA, MES was chosen as BGE co-ion. Since MES has no buffering capacity in the pH range selected, Tris (with a pKa of 8.15) was selected as the BGE counter-ion. Once the co- and counter-ion had been selected, other BGE parameters were calculated using the PeakMaster software.2727 Gas, B.; Jaros, M.; Hruska, V.; Zuskova, I.; Stedry, M.; Lc GC Eur. 2005, 18, 282. The optimized running electrolyte was composed of 40 mmol L-1 Tris and 20 mmol L-1 MES, at pH 8.1.

Choice of internal standard

The selection of the internal standard was based on the following criteria: the compound is fully dissociated at the chosen pH, which leads to a more robust and reproducible method since its µeff will not vary with any pH change; and the effective mobility is close to the mobility of the co-ion but different to the effective mobility of L-AA. Sorbic acid was the compound selected because it satisfies the established criteria (Figure 2), and has a high value for the absorptivity coefficient at 266 nm (λmax of L-AA). Figure 2 shows the effective mobility vs. pH curve for SA. For this selection, a PeakMaster database was used and it was possible to sort the database components according to their ionic mobility and choose analyte with suitable properties.

Solvent addition effect

In order to remove macromolecules, such as proteins and polysaccharides, which may be present in the samples, MeOH or ACN was added on the sample to promote the precipitation of these compounds, since they can interact with the capillary walls and alter the EOF.2828 Ragozina, N. Y.; Putz, M.; Heissler, S.; Faubel, W.; Pyell, U.; Anal. Chem. (Washington, DC, U. S.) 2004, 76, 3804.

The addition of solvent increases the plate number due to a decrease in the conductivity of the sample plug, favoring the occurrence of normal stacking.2929 Quirino, J. P. ; Terabe, S.; J. Chromatogr. A 2000, 902, 119. Figure 3 shows the plate number per meter (N m-1) for the L-AA when 0-80% of MeOH or ACN was added on the sample. The effect of ACN (dielectric constant 37.5 at 20 °C and viscosity 0.38 cP at 20 °C) was greater than that of MeOH (dielectric constant 32.7 at 20 ºC, viscosity 0.59 cP at 20 °C). The higher efficiency achieved with the addition of ACN does not appear to be related only to the dielectric constant value, but also to a change in the electrophoretic velocity caused by a change in viscosity, since the sample ions move from the low viscosity zone (ACN) to the BGE zone, which has no organic solvent.2929 Quirino, J. P. ; Terabe, S.; J. Chromatogr. A 2000, 902, 119.

Figure 3
Effect of addition of the organic modifiers on the sample on the plate number per meter.

The best efficiency was obtained using 80% of ACN, however, with this amount of solvent current breakdown occurred in several runs, probably due to the generation of bubbles from solvent evaporation in the plug sample caused by the high electric field in the plug. An amount of 60% (v/v), which provided high efficiency and macromolecule precipitation without current breakdown, was therefore used in the sample preparation.

Electrophoretic separation

To perform the electrophoretic separations, all samples were submitted to the procedure described above (Samples sub-section). The results for the separation of working standards and samples, under optimized conditions, are shown below (Figures 4a- c), where the fast separation of L-AA and SA (< 30 s) can be noted. This was possible due to the difference in the effective mobility values and performing the injection at short-end (Ldet 8.5 cm), the polarity was reversed that means positive polarity at the injection side, with a high electrical field (937.5 V cm-1). Since high resolutions (Rs) could be applied between multiple injections of L-AA and SA (Rs 2.05), the total analysis time was further reduced, leading to an increase in the analytical frequency.

Figure 4
Electropherogram of L-ascorbic acid (AA) and sorbic acid (SA) in food samples. Orange juice (a); vitamin C (b); and working standard solution of 50 mg L-1 L-AA, 20 mg L-1 sorbic acid (c). Electrolyte system composed of 40 mmol L-1 Tris and 20 mmol L-1 2-morpholinoethanesulfonic acid (MES, pH 8.1). Other conditions: short-end capillary hydrodynamic injection (-50 mbar, 4 s), 30 kV applied voltage, 25 °C, direct detection at 266 nm. Peak legend: (i) L-AA; (ii) sorbic acid; #: unknown peak.

Method validation

The evaluation of the proposed method optimized for the determination of L-AA was performed in accordance with the US Food and Drug Administration (FDA),3030 Vitali, L.; Valese, A. C.; Azevedo, M. S.; Gonzaga, L. V.; Costa, A. C. O.; Piovezan, M.; Vistuba, J. P.; Micke, G. A.; Talanta 2013, 106, 181.,3131 Zimmer, D.; Bioanalysis 2014, 6, 13. using the validation parameters: linearity, recovery, repeatability, precision, LOD, LOQ and selectivity. The results are shown in Table 1. In all cases, 20 mg L-1 of internal standard was added.

Table 1
Figures of merit for the proposed method

In order to obtain quantification models through the use of ordinary least squares regression, external calibration curves with internal standard addition employing six levels of L-ascorbic acid (1-50 mg L-1) were constructed with authentic replicates (n = 3). The linear mode showed a significance of F > 25000 and a good correlation coefficient (R2 > 0.999) were also achieved.

The precision was determined in order to evaluate the relative standard deviation (RSD) for the corrected peak area (ratio, analyte area / IS area) and the migration time of L-AA using a standard solution of 20 mg L-1 sorbic acid. Instrumental precision was established through seven consecutive injections of a standard solution. Repeatability values for the migration time and peak area ratio were better than 0.3 and 1.5%. Intra-day precision was established through three independent preparations of samples and injection in triplicate. The RSD values for the migration time and peak area were better than 0.5 and 1.4%, respectively. Inter-day precision was established through three injections of a standard solution, on three different days. The values obtained for peak area and migration time were better than 1.2 and 1.5%, respectively. The values obtained for the precision were appropriate, considering the complexity of the sample matrix. The limits of detection and quantification obtained are considered to be suitable for the concentrations of L-AA in all samples.

The matrix effect of the proposed method was evaluated using the method of standard addition, due to difficulties associated with obtaining a matrix free of the active ingredient. The slopes of the standard addition curves (0.0218 ± 0.0009) and external standard calibration curves (0.0209 ± 0.0007) were compared.

Table 2
Verification on the selectivity of the method for the determination of L-ascorbic acid in commercial products using the standard addition method

The similarity of these slopes (the ratio between the slopes of the calibration curves was around unity) indicates that matrix did not influence the method.2626 Vitali, L.; Della Betta, F.; Costa, A. C. O.; Vaz, F. A. S.; Oliveira, M. A. L.; Vistuba, J. P.; Favere, V. T.; Micke, G. A.; Talanta 2014, 123, 45. Moreover, the recovery results showed good agreement (92-107% recovery) with the reference values. Thus, the method described herein showed appropriate selectivity for L-AA in natural and synthetic samples.

Application to real samples

To demonstrate the applicability of the electrophoretic procedure developed, some samples were analyzed, including a sample of orange juice. The electropherograms for the real samples and standards are shown in Figure 4. Each sample was prepared in duplicate and injected in triplicate. A notable observation from this figure is that even though this is a real sample, the electropherogram is clear, which demonstrates that the clean-up step with the addition of ACN was efficient. The results for the quantification of L-AA in the samples and the value informed on the label are shown in Table 3.

Table 3
Results for the determination of L-ascorbic acid in commercial samples using the optimized capillary zone electrophoresis (CZE)-ultraviolet (UV) method

The values informed and determined are matching for orange juice and effervescent tablets, which indicates the applicability of the method for analysis of L-AA. The difference between values found for vitamin C liquid and the declared value can be related to a several possibilities as bad packaging, exposure to sun light and so on, that could affect the stability of the L-AA.

Conclusions

In this study, an electrolyte composition for the determination of L-AA was optimized using a systematic procedure. Satisfactory precision data were obtained using sorbate as the internal standard. The proposed method was found to be simple, fast and reliable, and baseline resolution was achieved within 26 s. The optimized procedure allows the determination of L-AA in different types of samples. The proposed steps for the systematic development of a CE method represent a convenient approach to the selection of the co-ion, operation mode and internal standard and may be considered for application in routine optimization procedures.

Acknowledgments

The authors wish to thank the Conselho Nacional de Pesquisa e Desenvolvimento (CNPq, Brazil), Instituto Nacional de Ciência e Tecnologia de Catálise (INCT-catálise) and Farma Service Bioextract for financial support.

References

  • 1
    Terabe, S.; Otsuka, K.; Nishi, H.; J. Chromatogr. A 1994, 666, 295.
  • 2
    Landers, J. P.; Handbook of Capillary and Microchip Electrophoresis and Associated Microtechniques, 3rd ed.; CRC Press: Boca Raton, 2008.
  • 3
    Johns, K. F.; Breadmore, M. C.; Brudo, R.; Haddad, P. R.; Electrophoresis 2009, 30, 839.
  • 4
    Vistuba, J. P.; Dolzan, M. D.; Vitali, L.; de Oliveira , M. A. L.; Micke, G. A.; J. Chromatogr. A 2015, 1396, 148.
  • 5
    Piovezan, M.; Costa, A. C. O.; Jager, A. V.; de Oliveira, M. A. L.; Micke, G. A.; Anal. Chim. Acta 2010, 673, 200.
  • 6
    Vitali, L.; Favere, V. T.; Micke, G. A.; J. Chromatogr. A 2011, 1218, 2327.
  • 7
    Meister, A.; Biochem. Pharmacol. (Amsterdam, Neth.) 1992, 44, 1905.
  • 8
    Nasirizadeh, N.; Shekari, Z.; Tabatabaee, M.; Ghaani, M.; J. Braz. Chem. Soc. 2015, 26, 713.
  • 9
    Alwarthan, A. A.; Analyst (Cambridge, U. K.) 1993, 118, 639.
  • 10
    Bottegal, M.; Lang, L.; Miller, M.; McCord, B.; Rapid Commun. Mass Spectrom. 2010, 24, 1377.
  • 11
    Park, H. W.; Alam, S. M.; Lee, S. H.; Karim, M. M.; Wabaidur, S. M.; Kang, M.; Choi, J. H.; Luminescence 2009, 24, 367.
  • 12
    Fischer, D. J.; Hulvey, M. K.; Regel, A. R.; Lunte, S. M.; Electrophoresis 2009, 30, 3324.
  • 13
    Zhao, S. L.; Huang, Y.; Liu, Y. M.; J. Chromatogr. A 2009, 1216, 6746.
  • 14
    Dong, S. Q.; Zhang, S.; Cheng, X.; He, P. G.; Wang, Q. J.; Fang, Y. Z.; J. Chromatogr. A 2007, 1161, 327.
  • 15
    Wang, X. L.; Li, L. J.; Li, Z. Y.; Wang, J.; Fu, H. Y.; Chen, Z. Z.; J. Electroanal. Chem. 2014, 712, 139.
  • 16
    Marra, M. C.; Silva, P. L.; Munoz, R. A. A.; Ritcher, E. M.; J. Braz. Chem. Soc. 2014, 25, 913.
  • 17
    Tang, Y. J.; Wu, M. J.; Talanta 2005, 65, 794.
  • 18
    Choi, O. K.; Jo, J. S.; J. Chromatogr. A 1997, 781, 435.
  • 19
    Fukushi, K.; Takeda, S.; Wakida, S.; Yamane, M.; Higashi, K.; Hiiro, K.; J. Chromatogr. A 1997, 772, 313.
  • 20
    Davey, M. W.; Bauw, G.; VanMontagu, M.; Anal. Biochem. 1996, 239, 8.
  • 21
    Galiana-Balaguer, L.; Rosello, S.; Herrero-Martinez, J. M.; Maquieira, A.; Nuez, F.; Anal. Biochem. 2001, 296, 218
  • 22
    Zinellu, A.; Carru, C.; Sotgia, S.; Deiana, L.; Anal. Biochem. 2004, 330, 298.
  • 23
    Goncalves, L. C.; Micke, G. A.; J. Chromatogr. A 2007, 1154, 477.
  • 24
    Costa, A. C. O.; da Costa, J. L.; Tonin, F. G.; Tavares, M. F. M.; Micke, G. A.; J. Chromatogr. A 2007, 1171, 140.
  • 25
    Hirokawa, T.; Nishino, M.; Aoki, N.; Kiso, Y.; Sawamoto, Y.; Yagi, T.; Akiyama, J.; J. Chromatogr. A 1983, 271, D1.
  • 26
    Vitali, L.; Della Betta, F.; Costa, A. C. O.; Vaz, F. A. S.; Oliveira, M. A. L.; Vistuba, J. P.; Favere, V. T.; Micke, G. A.; Talanta 2014, 123, 45.
  • 27
    Gas, B.; Jaros, M.; Hruska, V.; Zuskova, I.; Stedry, M.; Lc GC Eur 2005, 18, 282.
  • 28
    Ragozina, N. Y.; Putz, M.; Heissler, S.; Faubel, W.; Pyell, U.; Anal. Chem. (Washington, DC, U. S.) 2004, 76, 3804.
  • 29
    Quirino, J. P. ; Terabe, S.; J. Chromatogr. A 2000, 902, 119.
  • 30
    Vitali, L.; Valese, A. C.; Azevedo, M. S.; Gonzaga, L. V.; Costa, A. C. O.; Piovezan, M.; Vistuba, J. P.; Micke, G. A.; Talanta 2013, 106, 181.
  • 31
    Zimmer, D.; Bioanalysis 2014, 6, 13.

Publication Dates

  • Publication in this collection
    Nov 2016

History

  • Received
    12 Nov 2015
  • Accepted
    21 Mar 2016
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br