Organophosphorus Pesticide in Sapodilla (Manilkara zapota) Fruit

Abstract

A method based on QuEChERS (quick, easy, cheap, effective, rugged, and safe) extraction and gas chromatography-mass spectrometry (GC-MS) detection was described for the pesticides dichlorvos, disulfoton, ethoprophos, parathion methyl, fenchlorphos, chlorpyrifos, azinphos methyl and prothiofos in sapodilla. For all compounds studies, it was observed a strong matrix effect showing the need to use matrix matched calibration strategy. Method was validated, and good linearity (R > 0.99) was obtained for all pesticides studied with limits of detection (LODs) and quantification (LOQs) ranging from 0.01 to 0.06 mg kg-1 and 0.03 to 0.2 mg kg-1, respectively. Recovery studies were performed at different levels (0.08, 0.10, 0.14, 0.20, 0.35 and 1.17 mg kg-1) and showed good results (between 70 and 120% with relative standard deviation (RSD) < 20%). A statistical test was applied to the coefficients of the analytical curves obtained in the sapodilla matrix. Analyses of commercial samples showed chlorpyrifos were detected in about 70 and 33% for fruit and pulps samples, respectively. It should be noted that chlorpyrifos is not permitted in sapodilla crops by ANVISA and EC guidelines.

Keywords:
QuEChERS; validation; matrix effect; GC-MS


Introduction

The sapodilla (Manilkara zapota) is a fruit native to southern Mexico and Central America which had good adaptability to Brazil, mainly in the northeastern part of the country, due to the favorable climatic conditions.11 https://ww2.ibge.gov.br/home/estatistica/pesquisas/pesquisa_resultados.php?id_pesquisa=44, accessed in May 2018.
https://ww2.ibge.gov.br/home/estatistica...
,22 Silva Junior, J. F.; Bezerra, J. E. F.; Ledermen, I. E.; Moura, R. J. M.; Rev. Bras. Frutic. 2014, 36, 86.

Considered to be the ‘fruit honey of the tropics’ by the Brazilian Agricultural Research Company (Embrapa), consumer’s consumption of sapodilla has increased in the last years due to its nutritional benefits along with the pleasant sensory characteristics such as its aroma and flavor.33 de Miranda, M. R. A.; Filgueiras, H. A. C.; Alves, R. E.; Soares, A. A.; Benbadis, A. K.; Rev. Ciênc. Agron. 2008, 39, 4. It is commonly used as a raw material for ice cream, pudding, candy, pulp, besides being quite consumed in its nature form. Among all considered exotic fruits in Brazil, it is considered to be the one with the greatest potential for economic exploration.33 de Miranda, M. R. A.; Filgueiras, H. A. C.; Alves, R. E.; Soares, A. A.; Benbadis, A. K.; Rev. Ciênc. Agron. 2008, 39, 4.

However, it can also be a source of toxic substances due to the application of pesticides in the crops.44 Phan, K. T. K. P.; Phan, H. T.; Brennan, C. S.; Phimolsiripol, Y.; Int. J. Food Sci. Technol. 2017, 52, 2127. In 2010, the application of pesticides in Brazil increased twice the world average.55 http://www.abrasco.org.br/dossieagrotoxicos/wp-content/uploads/2013/10/DossieAbrasco_2015_web.pdf, accessed in May 2018.
http://www.abrasco.org.br/dossieagrotoxi...
The deleterious action of pesticides can cause a range of clinical manifestations, such as nausea, lack of appetite, headaches, allergies and can be also linked to different forms of cancer, genetic alterations and neurological effects.66 Bakirci, G. T. D.; Dilek, B. Y. A.; Bakirci, F.; Ötles, S.; Food Chem. 2014, 160, 379.,77 Akoto, O.; Gavor, S.; Appah, M. K.; Apau, J.; Environ. Monit. Assess. 2015, 187, 236.

In Brazil, there are over 800 compounds belonging to more than 100 different chemical classes applied as pesticides in different crops.88 Paz, M.; Correia-Sá, L.; Becker, H.; Longhinitti, E.; Domingues, V. F.; Food Control 2015, 54, 374. According to the last monitoring of the Brazilian National Health Surveillance Agency (ANVISA),99 http://portal.anvisa.gov.br/programa-de-analise-de-registro-de-agrotoxicos-para, accessed in May 2018.
http://portal.anvisa.gov.br/programa-de-...
organophosphorus (OPPs) pesticides were detected in a wide variety of fruits and vegetables.

They are commonly used as insecticides, acaricides, nematicides and fungicides.1010 Savoy, V. L. T.; Biológico 2011, 73, 91. Due to their low cost, high efficiency and wide usage spectrum, they are the most extensively used group of pesticide.1111 Tang, Q.; J. Sep. Sci. 2014, 37, 820.,1212 Seebunrueng, K.; Santaladchaiyakit, Y.; Srijaranai, S.; Talanta 2015, 132, 769. Therefore, it is necessary to develop efficient and reliable analytical procedures for the determination of OPPs pesticides residues in foods, mainly in unusual fruits such as sapodilla since no legislation have been implemented for those products.

Since the majority of the OPPs are volatile and thermally stable, they are amenable to gas chromatography (GC) analysis.1313 Sapahin, H. A.; Makahleh, A.; Saad, B.; Arabian J. Chem. 2015, DOI: 10.1016/j.arabjc.2014.12.001.
https://doi.org/10.1016/j.arabjc.2014.12...
Gas chromatography coupled to mass spectrometry (MS) with electron impact ionization has frequently been applied for the analysis of pesticide residues in different matrices with a high efficiency.1414 Guedes, J. A. C.; Silva, R. O.; Lima, C. G.; Milhome, M. A. L.; Nascimento, R. F.; Food Chem. 2016, 199, 380.,1515 Sousa, J. S.; Carius, C. R.; Andrade, G. A.; Lima, C. G.; Lima, L. K.; Milhome, M. A. L.; Nascimento, R. F.; Food Chem. 2013, 141, 2675. The single quadrupole analyzer is the most common instrument used in the coupling of GC-MS due to its low cost and ability to work in the selective ion monitoring mode (SIM) or extracted ion (EI) monitoring mode in order to increase the selectivity and sensitivity in addition to allow the possibility of confirmation of the investigated compound.1616 Pavia, D. L.; Lampman, G. M.; Kriz, G. S.; Vyvyan, J. A.; Introduction to Spectroscopy, 5th ed.; Cengage Learning: Stamford, USA, 2015. The advantage of the EI over the SIM is that there is no necessity in performing another chromatographic run to monitor specific m/z ratios corresponding the analyte of interest, once the EI mode allows selecting some values of m/z within the monitored range in total ion chromatogram (TIC) mode, and only the chromatogram containing them appears on the screen.

Since fruits have very complex matrices and generally the concentrations of pesticides are in ppb levels, a sample preparation step is required prior to instrumental analysis.1717 Christia, C.; Bizani, E.; Christophoridis, C.; Fytianos, K.; Environ. Sci. Pollut. Res. 2015, 22, 13167. Current trends prioritize the development of procedures for sample preparation that comply with the principles of green chemistry, e.g., decreasing the organic solvent consumption. Anastassiades et al.1818 Anastassiades, M.; Lehotay, S. J. S.; Tajnbaher, D.; Schenck, F. J.; J. AOAC Int. 2003, 86, 412. introduced a new method for the extraction of pesticides, which was mainly applicable to fruits and vegetables to overcome the practical limitations of multiresidue extraction methods that are generally time-consuming and laborious. This method is described as quick, easy, cheap, effective, rugged and safe (QuEChERS).1818 Anastassiades, M.; Lehotay, S. J. S.; Tajnbaher, D.; Schenck, F. J.; J. AOAC Int. 2003, 86, 412. In comparison with other techniques, the QuEChERS method has proven to have several advantages due to it giving excellent recoveries, involving less time and less solvent consumption.1919 Zhang, L.; Liu, S.; Cui, X.; Pan, C.; Zhang, A.; Chen, F.; Cent. Eur. J. Chem. 2012, 10, 900.

20 Vidal, J. L. M.; Frenich, A. G.; Methods in Biotechnology: Pesticide Protocols, vol. 19, 1st ed.; Humana Press: Totowa, NJ, USA, 2005.

21 González-Curbelo, M. A.; Lehotay, S. J.; Hernández-Borges, J.; Rodriguez-Delgado, M. A.; J. Chromatogr. A 2014, 1358, 75.

22 Li, Y. F.; Qiao, L.; Li, F.; Ding, Y.; Yang, Z.; Wang, M.; J. Chromatogr. A 2014, 1361, 77.
-2323 https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf, accessed in February 2018.
https://ec.europa.eu/food/sites/food/fil...

As far as the authors know, an analytical methodology for the determination of pesticide residues in sapodilla has not been reported. The purpose of the present work was to develop and validate a methodology for the determination for eight OPPs (dichlorvos, ethoprophos, disulfoton, parathion-methyl, fenchlorphos, chlorpyrifos, prothiofos and azinphos-methyl) in sapodilla samples using the QuEChERS/GC-MS method, as well as to determine the matrix effects for the evaluated analytes during the analyses of sapodilla.

Experimental

Chemicals and reagents

Acetonitrile 99.9% (v v-1), methanol 99.97% (v v-1), n-hexane 99.97% (v v-1), anhydrous magnesium sulfate P.A., sodium chloride P.A. and trisodium citrate dihydrate P.A. were purchased from Vetec (Brazil). The sodium hydrogen citrate sesquihydrate P.A. was obtained from Sigma-Aldrich (Brazil). Primary secondary amine (PSA) and graphitized carbon black (GCB) were purchased from Supelco (USA). Water was purified by a Milli-Q (Millipore, Academic, USA) system (18.2 MΩ cm).

The standard solution (2000 mg L-1 in n-hexane) for eight OPPs pesticides (dichlorvos, disulfoton, ethoprophos, parathion methyl, fenchlorphos, chlorpyrifos, methyl azinphos and prothiofos) was purchased from Sigma-Aldrich (Brazil). A stock standard solution (10 mg L-1) was prepared in n-hexane. To determine the linear range of each analyte were obtained work standard solutions of 5, 10, 15, 25, 50, 75, 100, 150, 200, 250, 500, 1000, 1500, 2000 and 2500 µg L-1; however, the analytical curves (concentration × area) in the solvent and matrix, for each compound, were constructed using five concentration levels starting from their limits of quantification (LOQs). All standards were stored under refrigeration at 4 ºC until use.

Sample collection

Sapodilla’s samples were collected randomly in ten different supermarkets of the city of Fortaleza, Ceará, Brazil. Samples (4 fruits from each supermarket) were crushed, homogenized using an industrial blender, placed in amber glass bottles, sealed and stored under refrigeration at 4 ºC until use. Samples were named S1, S2, S3, S4, S5, S6, S7, S8, S9, and S10, respectively, for each supermarket.

Commercial pulps of three different brands were also collected. They were also placed in amber glass bottles, sealed and stored under refrigeration at 4 ºC. Samples were named B1, B2, and B3, respectively, for each brand.

QuEChERS extraction

The samples of fruit and commercial pulp of sapodilla were analyzed according to the QuEChERS method.2424 Prestes, O. D.; Adaime, M. B.; Zanella, R.; Sci. Chromatogr. 2011, 3, 51. Basically, a portion of 10.0 g of sample was weighed into a 50.0 mL polypropylene centrifuge tube. Ten milliliters of acetonitrile was added, and the resulting solution was shaken by vortex mixing (1 min) to ensure that the solvent interacted well with the matrix. Next, the QuEChERS extraction sorbents (4.0 g MgSO4, 1.0 g NaCl, 1.0 g trisodium citrate dihydrate and 0.5 g disodium hydrogen citrate sesquihydrate) were added, and the final mixture was also shaken by vortex mixing (2 min) to prevent salt agglomeration. Afterwards, the extracts were then centrifuged for 10 min at 1008 × g (centrifuge Kindly, KC5, Brazil). A volume of 1.0 mL of the extracts was transferred to a 15.0 mL polypropylene centrifuge tube containing clean-up sorbents (150.0 mg of magnesium sulfate, 150.0 mg of PSA and 30.0 mg of GCB). The tubes were capped tightly and vortexed for 1 min, followed by centrifugation at 1008 × g for 10 min. The extract after clean-up was transferred to an Eppendorf tube for complete solvent evaporation in a vacuum centrifuge (SpeedVac, Thermo, Brazil). Then, the residue obtained was reconstituted in 1.0 mL of n-hexane, the same extract volume there was previously the evaporation step.

Gas chromatography-mass spectrometry (GC-MS)

A gas chromatograph (Agilent Technologies, model 7890B, USA) coupled to a mass spectrometer (Agilent Technologies, model 5977A, USA) was used in the development of the multiresidue method. The separation of the pesticides was performed using an HP5-S ms (30 m × 0.25 mm i.d. × 0.25 µm thickness) capillary column and helium (99.99%) carrier gas at a constant flow of 1 mL min-1.

The injection temperature (IT) was set at 240 ºC, and a 5 µL volume was injected in the splitless mode. The oven temperature was programmed as follows: initial temperature of 100 ºC for 1 min, 12 ºC min-1 rate to 165 ºC, then 10 ºC min-1 rate to 300 ºC, and holding at 300 ºC for 2 min, with a total run time 21.9 min. The mass spectrometer conditions were set as follows: ionization mode, electron impact (EI), 70 eV; ion source temperature, 240 ºC; and transfer line temperature, 240 ºC.

Method validation

The analytical method was validated following the SANTE/12571/20172323 https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf, accessed in February 2018.
https://ec.europa.eu/food/sites/food/fil...
and ANVISA2525 http://portal.anvisa.gov.br/documents/10181/2721567/RDC_166_201_COMP.pdf/d5fb92b3-6c6b-4130-8670-4e3263763401/, accessed in May 2018.
http://portal.anvisa.gov.br/documents/10...
guidelines. The validation parameters, including the selectivity, linearity, matrix effect, limit of detection (LOD), limit of quantification (LOQ), accuracy and precision (instrument and intra-assay precision), were determined.

The analyses were performed in the total ion chromatogram (TIC) mode, monitoring a wide range of ions (40-400 m/z). This way was made use of the extracted ion chromatogram (EI) tool, selecting the main fragments of the mass spectrum of each analyte (one ion of quantification and two of identification) to obtain a chromatogram only for the compound of interest, thus improving the sensitivity and selectivity. Table 1 shows the retention times of pesticides and their fragments used quantitatively and qualitatively to obtain the extracted ion chromatogram of each one. The fragments were selected in agreement with the base peak, the second most intense fragment and the one corresponding the molecular ion for each pesticide evaluated.

Table 1
Retention times and fragments used to obtain EI of each pesticide

Analytical standards ranging from 5 to 2500 µg L-1 were injected into the chromatographic system to obtain the analytical curve in the solvent. However, to construct the curve in the matrix, the sample extracts obtained after the evaporation of the acetonitrile in a vacuum centrifuge (SpeedVac, Brazil) were resuspended with each one of the analytical standards already prepared for subsequent injection into the system.

A good linear fit was found for the concentration range adopted, as it was found to approach a correlation coefficient (R) value of one. Statistical significance tests, using the Student’s t-test, of the calibration parameters (linear, a, and angular, b, coefficients) were applied in order to verify the adjustment of the analytical curve by the following equations:2626 Harris, D. C.; Quantitative Chemical Analysis, 8th ed.; Freeman and Company: New York, 2013.

(1) s a 2 = s y 2 x i 2 D
(2) s b 2 = s y 2 n D
(3) s y 2 = d i 2 n 2
(4) t calc , b = b 1 s b
(5) t calc , a = a 0 s a
(6) D = x i 2 x i x i n

where sa2, sb2 and sy2 are the square of the standard deviations of the linear coefficient, angular coefficient and y-axis, respectively; xi are individual values on the x-axis; n is the total number of points on the analytical curve; di is the vertical deviation of each point; D is the determinant; and tcalc is the calculated t-test. If tcalc for the parameters is greater than tabulated critical value (tcrit, confidence level 95%), the parameter is significant and must be kept in the equation. Otherwise, the parameter is considered insignificant and should be removed from the linear equation (y = a + bx).2727 Ligiero, C. B. P.; Reis, L. A.; Parrilha, G. L.; Baptista Filho, M.; Canela, M. C.; Quim. Nova 2009, 32, 1338.

The LODs and LOQs were estimated from successive injection of dilute solutions of the lowest concentration that resulted in signal/noise ratios of three and ten, respectively. Accuracy was demonstrated by the recovery experiments through analyzing samples spiked with pesticides at 0.08, 0.1, 0.14, 0.17, 0.35 and 1.17 mg kg-1, performed by triplicate assay and duplicate injection. The instrument precision has been verified by the repeatability method through calculating the relative standard deviation (RSD, in percentage) of areas obtained after ten injections of the 1 mg L-1 standard solution into the chromatographic system, and the intra-assay precision through calculating the RSD of the recovery experiments performed in triplicate for each spiked level.

Matrix effect

The evaluation of the matrix effect caused by the sapodilla extract was performed by comparing the angular coefficients obtained by the curves in the solvent and in the matrix according to the following equation:

(7) Matrix effect % = b m b s 100 b s

where bm and bs are the angular coefficients of the curve in the matrix and in the solvent, respectively, not considering the existence of a matrix effect for relations within the range of -20 to 20%.2323 https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf, accessed in February 2018.
https://ec.europa.eu/food/sites/food/fil...
,2828 Silva, C. M. S.; Habermann, G.; Marchi, M. R. R.; Zocolo, G. J.; Braz. Soc. Plant Physiol. 2012, 24, 223.

Calculation of analyte concentration in the sample in mg kg-1

After the injection of a sample previously prepared by the QuEChERS method and data processing via an analytical curve, concentration results are obtained in µg L-1 (weight/volume) (primary extract concentration). Thus, to obtain the estimated analyte concentration in the sample in mg kg-1 (weight/weight), the following equation was used:

(8) C F = C ext μ g L 1 × V L m a kg × 1000 mg kg 1

where CF is the analyte concentration in the sample, Cext is the analyte concentration in the extract (obtained from the analytical curve), V is the extract volume, ma is the mass of the weighed sample and 1000 is the conversion factor from µg to mg. This equation was used to estimate the concentration values of the analytes in commercial samples as well as the LOD and LOQ values and spiked levels.

Results and Discussion

Chromatographic analysis by GC-MS

The chromatographic conditions described previously for the GC-MS system allowed an efficient resolution for the eight OPPs pesticides. Figure 1a shows the separation profile of the analytes contained in a mixed standard solution (1 mg L-1) in the total ion chromatogram (TIC) mode, where total elution occurred in less than 20 min. Despite the low intensity relative to the azinphos methyl and the formation of two peaks that indicate dichlorvos, the compounds showed good sensitivity and selectivity.

Figure 1
(a) Chromatogram obtained using an IT of 240 ºC after the injection of 5 μL of the standard solution of 1 mg L-1 in the TIC model (retention times: 4.8 min (dichlorvos); 9.1 min (ethoprophos); 10.9 min (disulfoton); 11.8 min (prathion methyl); 12 min (fenchlorphos); 12.7 min (chlorpyriphos); 14.3 min (prothiofos) and 17.5 min (azinphos methyl)); (b) total ion chromatogram of the 1.0 mg L-1 standard solution of pesticides in the sapodilla matrix; (c) ethoprophos extracted ion chromatogram; and (d) ethoprophos mass spectrum.

Selectivity

The use of modern mass spectrometry detectors allows the chromatogram of the compound of interest to be extracted (extracted ion chromatogram, EI) from the total ion chromatogram (TIC). Figures 1b, 1c and 1d show that when using the EI tool, the method is selective for the pesticides and shows improved sensitivity (for example, ethoprophos). This tool was used for the construction of analytical curves in the matrix and for qualitative and quantitative analysis in real samples.

Matrix effect and linearity

The calibration curve in the presence of matrix was performed and compared with the one obtained in the presence of pure standards in solvent. The results (Figure 2a) indicate that the sapodilla matrix strongly influences the analytical signals of the compounds examined. Figure 2b shows the analytical curves for ethoprophos using solvent and matrix, illustrating the high occurrence of a matrix effect in sapodilla.

Figure 2
(a) Matrix effect of the eight pesticides in the sapodilla matrix analyzed by GC-MS; (b) superposition of solvent and matrix curves for ethoprophos.

It has been shown that the intensity of this effect is dependent on the physicochemical properties of pesticides such as polarity, molecular weight, thermal stability, temperature of boiling analytes, etc.2929 Pinho, G. P.; Neves, A. A.; Queiroz, M. E. L. R.; Silvério, F. O.; Quim. Nova 2009, 32, 987.,3030 Pinho, G. P.; Silvério, F. O.; Neves, A. A.; Queiroz, M. E. L. R.; Quim. Nova 2012, 35, 738. Negative matrix effect is observed due to degradation of the pesticide in injector or analyte adsorption on the active sites (free silanols groups) from liner. This effect causes decreasing of the analyte flux transferred to the column and consequently the decreasing of the detector signal.3131 Erney, D. R.; Gillespie, A. M.; Gilvydis, D. M.; Poole, C. F.; J. Chromatogr. A 1993, 638, 57. When the sample matrix is injected, there is a competitive process between the analyte and other matrix compounds to be adsorbed on liner active sites, which favors the analyte availability to the column and consequently bring an increase of the signal.2929 Pinho, G. P.; Neves, A. A.; Queiroz, M. E. L. R.; Silvério, F. O.; Quim. Nova 2009, 32, 987.

30 Pinho, G. P.; Silvério, F. O.; Neves, A. A.; Queiroz, M. E. L. R.; Quim. Nova 2012, 35, 738.
-3131 Erney, D. R.; Gillespie, A. M.; Gilvydis, D. M.; Poole, C. F.; J. Chromatogr. A 1993, 638, 57. Thus, the comparative responses between the analyte presented in the standard solution and in the sample matrix are characterized by an response overestimation. Probably these discussed aspects justify the positive matrix effect observed for most pesticides studied, except for dichlorvos (Figure 2a).

Additionally, the sample nature can also influence the intensity of the matrix effect. Compounds such as lipids, pigments and other components with high molar mass are present in the sapodilla fruit.22 Silva Junior, J. F.; Bezerra, J. E. F.; Ledermen, I. E.; Moura, R. J. M.; Rev. Bras. Frutic. 2014, 36, 86.,33 de Miranda, M. R. A.; Filgueiras, H. A. C.; Alves, R. E.; Soares, A. A.; Benbadis, A. K.; Rev. Ciênc. Agron. 2008, 39, 4. These compounds can cause the observed matrix effect for dichlorvos, which is the most hydrophilic (Kow = 79.4) among the pesticides studied. Probably it is due to its interaction with high polarity compounds contained in the sample promoting a suppression in its signal (Figure 2a).3232 http://sitem.herts.ac.uk/aeru/ppdb/en/index.htm, accessed in May 2018.
http://sitem.herts.ac.uk/aeru/ppdb/en/in...
According to Hajslová et al.,3333 Hajslová, J.; Holadova, K.; Kocourek, V.; Poustka, J.; Godula, M.; Cuhra, P.; Kempny, M.; J. Chromatogr. A 1998, 800, 283. normally polar compounds such as those having P=O groups, like organophosphorus, are more susceptible to matrix effects.

Several authors have noticed matrix effect in the analysis of pesticides in different food matrices by GC-MS. Guedes et al.1414 Guedes, J. A. C.; Silva, R. O.; Lima, C. G.; Milhome, M. A. L.; Nascimento, R. F.; Food Chem. 2016, 199, 380. analyzed twenty one multiclass pesticides in guava samples by GC-MS and found that about 38% of them presented a positive matrix effect. Martins et al.3434 Martins, F. I. C. C.; Barbosa, P. G. A.; Zocolo, G. J.; Nascimento, R. F.; Chromatographia 2018, 81, 677. also observed matrix effect in the analysis of twelve multiclass pesticides in mango by GC-MS where the pesticide Y-clordano presented the strongest matrix effect with signal overestimation of almost 250%.

This proposed method showed good linearity, expressed by the values of R near 1.00, with the lowest value being 0.993 (ethoprophos curves in the matrix), and all of the analytes were in accordance with the guidelines of ANVISA2525 http://portal.anvisa.gov.br/documents/10181/2721567/RDC_166_201_COMP.pdf/d5fb92b3-6c6b-4130-8670-4e3263763401/, accessed in May 2018.
http://portal.anvisa.gov.br/documents/10...
that recommend values above 0.99. However, ideally the regression line should pass through the origin,2727 Ligiero, C. B. P.; Reis, L. A.; Parrilha, G. L.; Baptista Filho, M.; Canela, M. C.; Quim. Nova 2009, 32, 1338.,3535 Gomes, C. L.; Lima, A. C. A.; Barbosa, P. G. A.; Costa, R. S.; Melo, D. Q.; Nascimento, R. F.; J. Braz. Chem. Soc. 2017, 28, 2136. but this fact is rarely observed. Therefore, the adjustment of the analytical curve by statistical tests to minimize quantitative errors must be taken into account. For Guedes et al.,1414 Guedes, J. A. C.; Silva, R. O.; Lima, C. G.; Milhome, M. A. L.; Nascimento, R. F.; Food Chem. 2016, 199, 380. analytical curves with a high positive value of the linear coefficient indicate contamination process (background noise, contamination in the sample preparation, etc.) and negative values generally indicate analyte losses (analyte consumption, adsorption on chromatographic columns, etc.).

Thus, the t-test was used for the matrix analytical curve coefficients, a procedure rarely taken into account in the literature, in order to reduce errors during the quantification of pesticide residues. Table 2 shows the results of the statistical tests of the parameters of the analytical curves prepared in the matrix as well as the values of R, linear range obtained for each pesticide and final curves that should be used for the quantification of pesticides in commercial samples. The linear coefficients of dichlorvos, fenchlorphos, chlorpyrifos and prothiofos curves did not show statistical significance (tcalc < tcrit), and they were removed from the line equation in order to achieve more accurate results.

Table 2
Linearity of different pesticides in sapodilla matrix

Sensitivity, accuracy and precision

LODs and LOQs ranged from 0.01 to 0.06 mg kg-1 and 0.03 to 0.20 mg kg-1, respectively (Table 3). The compounds chlorpyrifos and prothiofos were shown to be the most sensitive pesticides. None of the pesticides studied have been permitted for use in the cultivation of sapodilla by ANVISA,2525 http://portal.anvisa.gov.br/documents/10181/2721567/RDC_166_201_COMP.pdf/d5fb92b3-6c6b-4130-8670-4e3263763401/, accessed in May 2018.
http://portal.anvisa.gov.br/documents/10...
and consequently, there are no maximum residue limits (MRLs) available for these pesticides in sapodilla fruit. However, MRLs is available, through ANVISA2525 http://portal.anvisa.gov.br/documents/10181/2721567/RDC_166_201_COMP.pdf/d5fb92b3-6c6b-4130-8670-4e3263763401/, accessed in May 2018.
http://portal.anvisa.gov.br/documents/10...
or European Commision (EC),2323 https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf, accessed in February 2018.
https://ec.europa.eu/food/sites/food/fil...
for OPPs pesticides in several food materials as also presented in Table 3.

Table 3
LODs, LOQs and MRLs of different pesticides in sapodilla matrix

Recovery studies were assessed at six levels of intentional contamination (0.08 to 1.17 mg kg-1), and they ranged from 42.3 to 170.76% with an RSD of 0.8 to 34.6% (Table 4). However, all pesticides had at least three concentration levels (low, medium and high) with the recovery percentages being in accordance with document SANTE/11813/20172323 https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf, accessed in February 2018.
https://ec.europa.eu/food/sites/food/fil...
establishing acceptance limits of 70-120% and RSD < 20%. Furthermore, it is important to consider that the accuracy is dependent on the concentration level of the analyte in the sample. In cases where the analyte is at trace levels and where there is a complex matrix, higher intervals of recovery percentages can be accepted (50 to 120%).3636 Ribani, M.; Bottoli, C. B. G.; Collins, C. H.; Jardim, I. C. S. F.; Melo, L. F. C.; Quim. Nova 2004, 27, 771. Thus, the results of the recoveries of pesticides in sapodilla can be considered to be satisfactory.

Table 4
Accuracies in terms of recovery with its RSDs (intra-assay precision) and repeatability precisions

The instrument precision evaluated in terms of the repeatability, and the intra-assay precision (RSDs of the recovery experiments), also shown in Table 4, are in agreement to the SANTE/11813/2017 guidelines2323 https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf, accessed in February 2018.
https://ec.europa.eu/food/sites/food/fil...
for most pesticide, that establishes RSDs less than 20% for multiresidue methods. Only azinphos methyl presented unsatisfactory instrument precision (21.68%) when considering those guidelines.

Analysis of commercial sapodilla samples

The QuEChERS/GC-MS method was applied for the determination of pesticide residues in ten sapodilla fruit samples sold in supermarkets in different areas (S1, S2, S3, S4, S5, S6, S7, S8, S9 and S10) and commercial pulps samples of three different brands (B1, B2 and B3).

From all pesticides studied, only chlorpyrifos was detected. This pesticide was presented in 70% of the fruit samples evaluated and in one of the three brands of commercial pulp. Among the OPPs pesticides, this is one of the most known active ingredients used in the farming community due to its wide spectrum utilization (formicide, acaricide, insecticide), and thus, it is indicated by legislation for application in the cultivation of various crops.3636 Ribani, M.; Bottoli, C. B. G.; Collins, C. H.; Jardim, I. C. S. F.; Melo, L. F. C.; Quim. Nova 2004, 27, 771. Nevertheless, the compound is not permitted for use in the cultivation of sapodilla, and thus, it should not be present in it.2525 http://portal.anvisa.gov.br/documents/10181/2721567/RDC_166_201_COMP.pdf/d5fb92b3-6c6b-4130-8670-4e3263763401/, accessed in May 2018.
http://portal.anvisa.gov.br/documents/10...

However, the results presented in Table 5 show that when compared with the MRLs of other samples, the levels obtained in the samples S7, S10 and B3 are a little irregular, and that one obtained at supermarket S6 showed a fairly significant level (0.63 mg kg-1).

Table 5
Results of analyzes of chlorpyrifos in samples of sapodilla fruits (S) and pulps (B) and similarity

Comparing the value of chlorpyrifos found in S6 with the acceptable daily intake (ADI) set by EC (0.001 mg kg-1 b.w.),3737 www.ec.europa.eu/food/plant/protection/resources/qualcontrol_en.pdf, accessed in May 2018.
www.ec.europa.eu/food/plant/protection/r...
a person of 50 kg can consume up to 0.05 mg of this pesticide per day. Knowing that, on average, sapodilla fruit sold in that supermarket has a mass of approximately 0.080 kg; it is estimated that one fruit from the supermarket S6 would have enough chlorpyrifos residue to exceed the daily limit acceptable for this person. It is known that this pesticide has a level II toxicological classification, and in mammals, it acts by inhibiting the cholinesterase enzyme causing neurotoxic effects.3838 Venugopal, N. V. S.; Sumalatha, B.; Bonthula, S. R.; Orient. J. Chem. 2012, 28, 979. Table 5 also shows the similarity of the chlorpyrifos mass spectrum obtained from the sample with the mass spectrum of its standard, where the samples S6, S7 and S10 have gotten more than 90% of possibility to really be the analyte.

The presence of pesticide residues in sapodilla samples has not been approached in the literature yet. The results indicate the need for studies to be expanded in the referred fruit plus the inclusion of this culture in the evaluations of national agencies such as ANVISA and INMETRO.

Conclusions

The QuEChERS/GC-MS method was a suitable technique for qualitative and quantitative analysis of organophosphorus pesticides in a sapodilla matrix. The validation parameters (selectivity, linearity, LODs, LOQs, accuracy and precision) proved to be satisfactory. The statistical evaluation of the linearity of the analytical curves indicates the insignificance of the linear coefficient for the compounds dichlorvos, fenchlorphos, chlorpyrifos and prothiofos. A strong matrix effect was observed for the eight OPPs compounds evaluated in this study, where the higher value was obtained for parathion methyl. The analyses of commercial samples demonstrated the increased use of commercial products containing the active ingredient chlorpyrifos by some farmers in the cultivation of sapodilla. This method can be efficiently applied to increase the scope of analysis of monitoring programs and improve food security.

Acknowledgments

The authors gratefully acknowledge the Coordination for the Improvement of Higher Education Personnel (CAPES) for financial support of this research and all of the infrastructure granted by the Trace Analysis Laboratory (LAT) of the Federal University of Ceará (UFC) and Multi-User Laboratory of Chemistry of Natural Products (LMQPN) of the Brazilian Company of Agricultural Research (Embrapa).

References

  • 1
    https://ww2.ibge.gov.br/home/estatistica/pesquisas/pesquisa_resultados.php?id_pesquisa=44, accessed in May 2018.
    » https://ww2.ibge.gov.br/home/estatistica/pesquisas/pesquisa_resultados.php?id_pesquisa=44
  • 2
    Silva Junior, J. F.; Bezerra, J. E. F.; Ledermen, I. E.; Moura, R. J. M.; Rev. Bras. Frutic. 2014, 36, 86.
  • 3
    de Miranda, M. R. A.; Filgueiras, H. A. C.; Alves, R. E.; Soares, A. A.; Benbadis, A. K.; Rev. Ciênc. Agron. 2008, 39, 4.
  • 4
    Phan, K. T. K. P.; Phan, H. T.; Brennan, C. S.; Phimolsiripol, Y.; Int. J. Food Sci. Technol. 2017, 52, 2127.
  • 5
    http://www.abrasco.org.br/dossieagrotoxicos/wp-content/uploads/2013/10/DossieAbrasco_2015_web.pdf, accessed in May 2018.
    » http://www.abrasco.org.br/dossieagrotoxicos/wp-content/uploads/2013/10/DossieAbrasco_2015_web.pdf
  • 6
    Bakirci, G. T. D.; Dilek, B. Y. A.; Bakirci, F.; Ötles, S.; Food Chem. 2014, 160, 379.
  • 7
    Akoto, O.; Gavor, S.; Appah, M. K.; Apau, J.; Environ. Monit. Assess. 2015, 187, 236.
  • 8
    Paz, M.; Correia-Sá, L.; Becker, H.; Longhinitti, E.; Domingues, V. F.; Food Control 2015, 54, 374.
  • 9
    http://portal.anvisa.gov.br/programa-de-analise-de-registro-de-agrotoxicos-para, accessed in May 2018.
    » http://portal.anvisa.gov.br/programa-de-analise-de-registro-de-agrotoxicos-para
  • 10
    Savoy, V. L. T.; Biológico 2011, 73, 91.
  • 11
    Tang, Q.; J. Sep. Sci. 2014, 37, 820.
  • 12
    Seebunrueng, K.; Santaladchaiyakit, Y.; Srijaranai, S.; Talanta 2015, 132, 769.
  • 13
    Sapahin, H. A.; Makahleh, A.; Saad, B.; Arabian J. Chem. 2015, DOI: 10.1016/j.arabjc.2014.12.001.
    » https://doi.org/10.1016/j.arabjc.2014.12.001.
  • 14
    Guedes, J. A. C.; Silva, R. O.; Lima, C. G.; Milhome, M. A. L.; Nascimento, R. F.; Food Chem. 2016, 199, 380.
  • 15
    Sousa, J. S.; Carius, C. R.; Andrade, G. A.; Lima, C. G.; Lima, L. K.; Milhome, M. A. L.; Nascimento, R. F.; Food Chem. 2013, 141, 2675.
  • 16
    Pavia, D. L.; Lampman, G. M.; Kriz, G. S.; Vyvyan, J. A.; Introduction to Spectroscopy, 5th ed.; Cengage Learning: Stamford, USA, 2015.
  • 17
    Christia, C.; Bizani, E.; Christophoridis, C.; Fytianos, K.; Environ. Sci. Pollut. Res. 2015, 22, 13167.
  • 18
    Anastassiades, M.; Lehotay, S. J. S.; Tajnbaher, D.; Schenck, F. J.; J. AOAC Int. 2003, 86, 412.
  • 19
    Zhang, L.; Liu, S.; Cui, X.; Pan, C.; Zhang, A.; Chen, F.; Cent. Eur. J. Chem. 2012, 10, 900.
  • 20
    Vidal, J. L. M.; Frenich, A. G.; Methods in Biotechnology: Pesticide Protocols, vol. 19, 1st ed.; Humana Press: Totowa, NJ, USA, 2005.
  • 21
    González-Curbelo, M. A.; Lehotay, S. J.; Hernández-Borges, J.; Rodriguez-Delgado, M. A.; J. Chromatogr. A 2014, 1358, 75.
  • 22
    Li, Y. F.; Qiao, L.; Li, F.; Ding, Y.; Yang, Z.; Wang, M.; J. Chromatogr. A 2014, 1361, 77.
  • 23
    https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf, accessed in February 2018.
    » https://ec.europa.eu/food/sites/food/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf
  • 24
    Prestes, O. D.; Adaime, M. B.; Zanella, R.; Sci. Chromatogr. 2011, 3, 51.
  • 25
    http://portal.anvisa.gov.br/documents/10181/2721567/RDC_166_201_COMP.pdf/d5fb92b3-6c6b-4130-8670-4e3263763401/, accessed in May 2018.
    » http://portal.anvisa.gov.br/documents/10181/2721567/RDC_166_201_COMP.pdf/d5fb92b3-6c6b-4130-8670-4e3263763401/
  • 26
    Harris, D. C.; Quantitative Chemical Analysis, 8th ed.; Freeman and Company: New York, 2013.
  • 27
    Ligiero, C. B. P.; Reis, L. A.; Parrilha, G. L.; Baptista Filho, M.; Canela, M. C.; Quim. Nova 2009, 32, 1338.
  • 28
    Silva, C. M. S.; Habermann, G.; Marchi, M. R. R.; Zocolo, G. J.; Braz. Soc. Plant Physiol. 2012, 24, 223.
  • 29
    Pinho, G. P.; Neves, A. A.; Queiroz, M. E. L. R.; Silvério, F. O.; Quim. Nova 2009, 32, 987.
  • 30
    Pinho, G. P.; Silvério, F. O.; Neves, A. A.; Queiroz, M. E. L. R.; Quim. Nova 2012, 35, 738.
  • 31
    Erney, D. R.; Gillespie, A. M.; Gilvydis, D. M.; Poole, C. F.; J. Chromatogr. A 1993, 638, 57.
  • 32
    http://sitem.herts.ac.uk/aeru/ppdb/en/index.htm, accessed in May 2018.
    » http://sitem.herts.ac.uk/aeru/ppdb/en/index.htm
  • 33
    Hajslová, J.; Holadova, K.; Kocourek, V.; Poustka, J.; Godula, M.; Cuhra, P.; Kempny, M.; J. Chromatogr. A 1998, 800, 283.
  • 34
    Martins, F. I. C. C.; Barbosa, P. G. A.; Zocolo, G. J.; Nascimento, R. F.; Chromatographia 2018, 81, 677.
  • 35
    Gomes, C. L.; Lima, A. C. A.; Barbosa, P. G. A.; Costa, R. S.; Melo, D. Q.; Nascimento, R. F.; J. Braz. Chem. Soc. 2017, 28, 2136.
  • 36
    Ribani, M.; Bottoli, C. B. G.; Collins, C. H.; Jardim, I. C. S. F.; Melo, L. F. C.; Quim. Nova 2004, 27, 771.
  • 37
    www.ec.europa.eu/food/plant/protection/resources/qualcontrol_en.pdf, accessed in May 2018.
    » www.ec.europa.eu/food/plant/protection/resources/qualcontrol_en.pdf
  • 38
    Venugopal, N. V. S.; Sumalatha, B.; Bonthula, S. R.; Orient. J. Chem. 2012, 28, 979.

Publication Dates

  • Publication in this collection
    Oct 2018

History

  • Received
    17 Jan 2018
  • Accepted
    14 May 2018