Determination of Parabens in Breast Milk Samples by Dispersive Liquid-Liquid Microextraction (DLLME) and Ultra-High-Performance Liquid Chromatography Tandem Mass Spectrometry

Abstract

A reliable method was developed and validated based on dispersive liquid-liquid microextraction (DLLME) and ultra-high-performance liquid chromatography coupled to tandem mass spectrometry (UHPLC-MS/MS) to determine simultaneously methylparaben (MePB), ethylparaben (EtPB), propylparaben (PrPB), and butylparaben (BuPB) in breast milk samples. A central composite design 24 with four independent variables and five levels of variance was used to optimize the DLLME conditions. The DLLME/UHPLC-MS/MS method was linear from 5 (lower limit of quantification, LLOQ) to 100 ng mL-1 (upper limit of quantification, ULOQ). The inter-assay and intra-assay precision values presented coefficients of variation (CVs) lower than 15%, and the relative standard error of the inter- and intra-assay accuracy values was lower than 10.4%. The DLLME/UHPLC-MS/MS method was successfully applied to determine parabens in human milk samples obtained from sixteen volunteer lactating mothers.

Keywords:
parabens; breast milk sample; dispersive liquid-liquid microextraction; ultra-high-performance liquid chromatography tandem mass spectrometry


Introduction

The alkyl esters of p-hydroxybenzoic acid known as parabens are widely used as antimicrobial preservatives, especially to prevent mold and yeast growth in cosmetic products and pharmaceuticals; they are also employed in food and beverage processing. Concerning their chemical structure, parabens contain different alkyl groups, including methylparaben (MePB), ethylparaben (EtPB), propylparaben (PrPB), isopropylparaben (iPrPB), butylparaben (BuPB), isobutylparaben (iBuPB), and benzylparaben (BePB).11 Błędzka, D.; Gromadzińska, J.; Wąsowicz, W.; Environ. Int. 2014, 67, 27.

Given the large use of parabens as preservatives, human exposure to these compounds is widespread. Once in contact with humans, parabens are absorbed through the skin or via the gastrointestinal tract after oral intake. Oral and dermal administration of these compounds leads to their hydrolysis into the harmless p-hydroxybenzoic acid and the respective glucuronic and sulfuric acid conjugates.22 Jardim, V. C.; Melo, L. P.; Domingues, D. S.; Queiroz, M. E. C.; J. Chromatogr. B 2015, 974, 35. Although parabens are still considered safe, the potential endocrine disrupting effects of the exposure to high levels of these compounds have raised health concerns.33 Frederiksen, H.; Jørgensen, N.; Andersson, A.-M.; J. Exposure Sci. Environ. Epidemiol. 2011, 21, 262. High concentrations of parabens in the breast tissue might be associated with breast cancer. Parabens can mimic estrogen action stimulating the growth of MCF7 cells, which are related to the development of breast cancer.44 Vela-Soria, F.; Ballesteros, O.; Zafra-Gómez, A.; Ballesteros, L.; Navalón, A.; Talanta 2014, 129, 209.,55 Byford, J. R.; Shaw, L. E.; Drew, M. G. B.; Pope, G. S.; Sauer, M. J.; Darbre, P. D.; J. Steroid Biochem. Mol. Biol. 2002, 80, 49.

Parabens at trace levels have been determined in several biological matrices, such as urine,22 Jardim, V. C.; Melo, L. P.; Domingues, D. S.; Queiroz, M. E. C.; J. Chromatogr. B 2015, 974, 35.,66 Carrasco-Correa, E. J.; Vela-Soria, F.; Ballesteros, O.; Ramis-Ramos, G.; Herrero-Martínez, J. M.; J. Chromatogr. A 2015, 1379, 65. breast milk,77 Rodríguez-Gómez, R.; Jiménez-Díaz, I.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Talanta 2014, 130, 561.

8 Rodríguez-Gómez, R.; Roldán-Pijuán, M.; Lucena, R.; Cárdenas, S.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26.

9 Souza, I. D.; Melo, L. P.; Jardim, I. C. S. F.; Monteiro, J. C. S.; Nakano, A. M. S.; Queiroz, M. E. C.; Anal. Chim. Acta 2016, 932, 49.

10 Ye, X.; Bishop, A. M.; Needham, L. L.; Calafat, A. M.; Anal. Chim. Acta 2008, 622, 150.
-1111 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538. plasma,1212 Suh, J. H.; Lee, Y. Y.; Lee, H. J.; Kang, M.; Hur, Y.; Lee, S. N.; Yang, D.-H.; Han, S. B.; J. Pharm. Biomed. Anal. 2013, 75, 214. serum,1313 Vela-Soria, F.; Ballesteros, O.; Rodríguez, I.; Zafra-Gómez, A.; Ballesteros, L.; Cela, R.; Navalón, A.; Anal. Bioanal. Chem. 2013, 405, 7259. amniotic fluid,1414 Frederiksen, H.; Taxvig, C.; Hass, U.; Vinggaard, A. M.; Nellemann, C.; Toxicol. Sci. 2008, 106, 376. and placental tissue.1515 Jiménez-Díaz, I.; Vela-Soria, F.; Zafra-Gómez, A.; Navalón, A.; Ballesteros, O.; Navea, N.; Fernández, M. F.; Olea, N.; Vílchez, J. L.; Talanta 2011, 84, 702.,1616 Vela-Soria, F.; Rodríguez, I.; Ballesteros, O.; Zafra-Gómez, A.; Ballesteros, L.; Cela, R.; Navalón, A.; J. Chromatogr. A 2014, 1371, 39. As known, lipophilic substances are mainly excreted in breast milk, and consequently, that breast milk is the major way through which breastfed infants are exposed to environmental chemicals. Breastfeeding women may be exposing their children to harmful levels of parabens in the early stages of their life when the babies are particularly vulnerable. Therefore, analysis of parabens in breast milk samples has become a topic of great scientific interest.

Pretreatment of biological samples is an important step of bioanalysis, but it is also the most difficult aspect prior the liquid chromatography (LC) separation and tandem mass spectrometry (MS/MS) detection for the target analytes. Indeed, macromolecules present in the biological samples could co-elute with the target analytes, to impair chromatographic performance, increase column backpressure,1717 Chaves, A. R.; Leandro, F. Z.; Carris, J. A.; Queiroz, M. E. C.; J. Chromatogr. B 2010, 878, 2123. and suppress electrospray ionization (ESI) during LC-MS/MS analysis.1818 Trufelli, H.; Palma, P.; Famiglini, G.; Cappiello, A.; Mass Spectrom. Rev. 2011, 30, 491.

Different sample preparation techniques combined with LC-MS/MS have been used to determine parabens in breast milk samples, including conventional techniques like online solid-phase extraction (online SPE),1010 Ye, X.; Bishop, A. M.; Needham, L. L.; Calafat, A. M.; Anal. Chim. Acta 2008, 622, 150. molecularly imprinted micro-disc solid-phase extraction (MISPE),1111 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538. stir-membrane solid-liquid-liquid microextraction (SM-SLLME),88 Rodríguez-Gómez, R.; Roldán-Pijuán, M.; Lucena, R.; Cárdenas, S.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26. in-tube solid-phase microextraction (in-tube SPME),99 Souza, I. D.; Melo, L. P.; Jardim, I. C. S. F.; Monteiro, J. C. S.; Nakano, A. M. S.; Queiroz, M. E. C.; Anal. Chim. Acta 2016, 932, 49. and quick, easy, cheap, effective, rugged and safe extraction (QuEChERS).77 Rodríguez-Gómez, R.; Jiménez-Díaz, I.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Talanta 2014, 130, 561.

Rezaee et al.1919 Rezaee, M.; Assadi, Y.; Milani Hosseini, M.-R.; Aghaee, E.; Ahmadi, F.; Berijani, S.; J. Chromatogr. A 2006, 1116, 1. introduced the dispersive liquid-liquid microextraction (DLLME) in 2006. DLLME is a powerful pre-concentration technique that uses a ternary solvent system. DLLME has received special attention because it is a simple and inexpensive procedure, provides high recovery and enrichment factors, requires short extraction time, and is compatible with various matrices.66 Carrasco-Correa, E. J.; Vela-Soria, F.; Ballesteros, O.; Ramis-Ramos, G.; Herrero-Martínez, J. M.; J. Chromatogr. A 2015, 1379, 65.,88 Rodríguez-Gómez, R.; Roldán-Pijuán, M.; Lucena, R.; Cárdenas, S.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26.,1212 Suh, J. H.; Lee, Y. Y.; Lee, H. J.; Kang, M.; Hur, Y.; Lee, S. N.; Yang, D.-H.; Han, S. B.; J. Pharm. Biomed. Anal. 2013, 75, 214.,2020 Saraji, M.; Khalili Boroujeni, M.; Hajialiakbari Bidgoli, A. A.; Anal. Bioanal. Chem. 2011, 400, 2149.

21 Asensio-Ramos, M.; Hernández-Borges, J.; Borges-Miquel, T. M.; Rodríguez-Delgado, M. Á.; J. Chromatogr. A 2011, 1218, 4808.
-2222 Han, Y.; Jia, X.; Liu, X.; Duan, T.; Chen, H.; Chromatographia 2010, 72, 351. Several factors such as the physical properties and volume of the extraction and the disperser solvents, extraction pH, and ionic strength, can influence DLLME efficiency.2323 Saraji, M.; Boroujeni, M. K.; Anal. Bioanal. Chem. 2014, 406, 2027.

In the present study, a powerful, fast and reliable method was developed applying DLLME and ultra-high-performance liquid chromatography coupled to tandem mass spectrometry (UHLPC-MS/MS) to extract and determine simultaneously four parabens in breast milk sample. In addition, an experimental design was applied to optimize the extraction of parabens carried out through DLLME. Lastly, the method herein was successfully applied to determine parabens in real samples collected from volunteer lactating mothers.

Experimental

Reagents and analytical standards

The methylparaben, ethylparaben, propylparaben, butylparaben, and propyl 4-hydroxybenzoate-ring-13C6 (internal standard (IS)) analytical standards were acquired from Sigma-Aldrich (St. Louis, USA).

The standard working solutions of the four parabens and the internal standard were prepared by diluting stock solutions of the parabens (1 mg mL-1) in methanol. The resulting solutions were stable at -4 ºC for 45 days.

Breast milk samples

Breast milk samples were collected from volunteer lactating mothers assisted at the Reference Center for Women's Health of the Nursing School of Ribeirão Preto, University of São Paulo (USP), Brazil. The protocol used to collect milk samples prevented sample contamination. Briefly, the breast was cleaned with deionized water, and the breast milk was hand expressed. Human milk samples were collected in glass jars with lids placed below the areola. Then, the milk samples were stored in a freezer at -20 ºC until use. The Ethics Committee of the University of São Paulo in Ribeirão Preto, Brazil, approved this study (protocol No. 1397/2011).

Ultra-high-performance liquid chromatography coupled to tandem mass spectrometry (UHPLC-MS/MS) conditions

The UHPLC-MS/MS system consisted of a Waters ACQUITY UHPLC (Waters Corporation, Milford, MA, USA) equipped with an ACQUITY UHPLC binary solvent system and an ACQUITY UHPLC sample manager. The UHPLC was coupled to the XEVO TQD mass spectrometer (tandem triple quadrupole) with electrospray ionization (ESI).

Chromatographic separations were conducted on a Kinetex C18 column (100 mm × 2.1 mm × 1.7 m, Phenomenex, USA) at 40 ºC. The mobile phase consisted of acetonitrile/water (40:60, v/v) at a flow rate of 0.2 mL min-1 in the isocratic mode. Data were acquired in the selected reaction monitoring (SRM) in the negative mode. Two specific transitions were optimized for each analyte, to increase the selectivity and the reliability of the method (Table 1). The first SRM transition was used for quantification, whereas the second was used for qualitative identification. The source and operating parameters were optimized as follows: capillary voltage, 3.0 kV; ion source temperature, 150 ºC; desolvation gas flow rate, 700 L h-1 (N2, 99.9%), at 500 ºC; cone gas flow of 20 L h-1 (N2, 99.9%) and argon as the collision gas.

Table 1
MS/MS transition, optimal declustering potential (DP), optimal collision energy (CE), and retention time (RT) for the studied parabens

Optimization of the DLLME/LC-MS/MS procedure

Commercial samples of liquid infant milk formula (NAN, 1 g) without parabens were used as blank samples. Before extraction, 1 g of the blank milk sample spiked with the IS (60 ng mL-1) and the standard solutions of the parabens (50 µL of 100-2000 ng mL-1 solutions) was diluted with acetonitrile (1 mL, disperser solvent) to precipitate the proteins. After centrifugation in a conic tube, 1700 µL of the supernatant was diluted with 10 mL of ammonium acetate solution (pH 6), to optimize the disperser (acetonitrile, methanol, ethanol, or acetone; 0.6 mL) and extraction solvents (dichloromethane, chlorobenzene, or trichloromethane; 500 µL).

Afterward, the independent variables extraction solvent volume (dichloromethane, 200-800 µL), disperser solvent volume (acetonitrile, 0.5-2.5 mL), ionic strength (NaCl, 0-20% m/v), and pH (4-8) were optimized according to the central composite design 24 at 5 levels (Table 2). The evaluated dependent variable (response) was the area of the target parabens. Twenty-eight randomized experiments were carried out; sixteen at the factorial points (-1, +1); eight at the axial points (-2, +2); and four at the central point (0, 0).

Table 2
Independent variables and central composite design of 24 varying levels

Analytical validation of the DLLME/UHPLC-MS/MS method

The DLLME/UHPLC-MS/MS method was fully validated in accordance with the requirements of European Medicines Agency (EMA)2424 European Medicines Agency (EMA); Reflection Paper on the Use of Methyl- and Propylparaben as Excipients in Human Medicinal Products for Oral Use; EMA: London, UK, 2015. Available at http://www.ema.europa.eu/docs/en_GB/document_library/Scientific_guideline/2015/11/WC500196733.pdf, accessed in August 2018.
http://www.ema.europa.eu/docs/en_GB/docu...
and US Food and Drug Administration (FDA)2525 http://www.fda.gov/cosmetics/productsingredients/ingredients/ucm128042.htm, accessed in August 2018.
http://www.fda.gov/cosmetics/productsing...
guidance documents for bioanalytical method validation.

Additionally, the matrix effect was assessed by post-column infusion. For this assay, a standard solution containing the target parabens (100 ng mL-1 in methanol) was continuously infused via mass spectrometer at a flow of 10 µL min-1. Then, 10 µL of a blank milk sample (DLLME extracted) was injected into the LC, and the analytical signal of the analytes was monitored throughout the chromatographic run.

Results and Discussion

Optimization of the DLLME conditions

The appropriate mixture of extraction and disperser solvents can be considered as the most important parameter for the DLLME procedure.2626 Zgoła-Grześkowiak, A.; Grześkowiak, T.; TrAC, Trends Anal. Chem. 2011, 30, 1382. Therefore, it is extremely important to evaluate suitable solvents through screening tests prior to the optimization of other parameters.

All the combinations based on acetonitrile, methanol, ethanol, and acetone (disperser solvents), and dichloromethane, trichloromethane, and chlorobenzene (extraction solvents) were evaluated. Based on the experimental results, the highest recovery rates were obtained for acetonitrile as disperser solvent and dichloromethane as extraction solvent (data not shown).

Both acetonitrile (miscible with both aqueous solvents and extraction solvents) and dichloromethane (water-immiscible extraction solvent) were rapidly injected (in sequence) into an aqueous sample containing the analytes of interest. Therefore, a cloudy solution containing fine droplets of the extraction solvent was formed. Then, the analytes were extracted from the aqueous phase to the extraction phase due the high superficial area generated between the aqueous phase and the extraction solvent.

In general, chemometric techniques require fewer laboratory tests than univariate approaches and can provide the optimal conditions for a particular analytical procedure, considering the influence as well as the interaction among the independent variables studied.2727 Kumar, N.; Bansal, A.; Sarma, G. S.; Rawal, R. K.; Talanta 2014, 123, 186. Hence, a central composite design 24 was performed in order to deeply study the four independent variables: extraction solvent volume (dichloromethane, µL), disperser solvent volume (acetonitrile, mL), ionic strength (NaCl, % m/v), and pH on DLLME parabens extraction efficiency.

Through the 28 experiments performed, a predictive response model was obtained for each paraben under study. The influence on the response (Y) of the independent variables (extraction solvent volume (x1), disperser solvent volume (x2), ionic strength (x3), and pH (x4)) as well as their interaction, were studied.

The model presented coefficient of determination (R2) of 0.711, 0.651, 0.773, and 0.813 for MePB, EtPB, PrPB, and BuPB, respectively. These values implied that the mathematical models were able to explain 71.1, 65.1, 77.3, and 81.3% of the experimental variations that occurred throughout the analytical process, respectively.

Considering the complexity of the studied matrix (breast milk), and that the technical DLLME work involved micro volumes of solvents, the R2 values might be considered satisfactory. Moreover, the lack-of-fit test was not significant at the 5% level (p < 0.05), indicating that the models were well adjusted.

Using the Derringer and Suich's desirability function approach that can maximize the overall desirability (multiple responses) based on the controllable factors, it was possible to obtain the best compromised condition for the simultaneous analysis of all the targeted parabens (Figure 1).2828 Costa, N. R.; Lourenço, J.; Pereira, Z. L.; Chemom. Intell. Lab. Syst. 2011, 107, 234.

Figure 1
Desirability parameters estimated in the optimal conditions of extraction for all the investigated parabens.

The desirability function indicated that the optimal extraction conditions were: (-2) 800 µL of dichloromethane; (-1) 1 mL of acetonitrile; (-2) 0% added with NaCl, and (-2) pH 4.0. The response surface plots obtained in the optimal conditions for all the parabens was displayed in Figure 2.

Figure 2
Response surfaces (Derringer and Suich) for all the evaluated parabens. (a) Disperser solvent × extractor solvent; (b) ionic strength NaCl × extractor solvent; (c) ionic strength NaCl × disperser solvent; (d) pH × extractor solvent; (e) pH × disperser solvent; (f) pH × ionic strength NaCl.

The extraction solvent volume is directly related to the extraction efficiency (Figures 2a, 2c and 2d). Based on the range estimated for the extraction solvent volume, the best extraction efficiency was obtained for 800 µL of dichloromethane.

The disperser solvent volume directly affects the formation of the cloudy solution (water/disperser solvent/extracting solvent), the degree of extraction solvent dispersion in the aqueous phase,2929 Rezaee, M.; Yamini, Y.; Faraji, M.; J. Chromatogr. A 2010, 1217, 2342. and the extraction efficiency.1919 Rezaee, M.; Assadi, Y.; Milani Hosseini, M.-R.; Aghaee, E.; Ahmadi, F.; Berijani, S.; J. Chromatogr. A 2006, 1116, 1. Large volumes of acetonitrile (1 mL) increased the solubility of the parabens in water, thereby decreasing the extraction efficiency (Figures 2b, 2c and 2e). The extraction carried out with 1 mL of acetonitrile presented the highest signal for the analytes.

According to Figures 2a, 2b and 2f, the responses (areas of the parabens) decreased with increasing concentration of NaCl. The salt concentration seems to decrease the solubility of acetonitrile in water phase impairing the cloudy state.3030 Alshana, U.; Ertaş, N.; Göğer, N. G.; Food Chem. 2015, 181, 1. Furthermore, the residual sodium might provide ion suppression decreasing the analytical signal in MS/MS.1717 Chaves, A. R.; Leandro, F. Z.; Carris, J. A.; Queiroz, M. E. C.; J. Chromatogr. B 2010, 878, 2123. Thus, no addition of NaCl was selected.

The evaluated pH ranged from 4.0 to 8.0. Values greater than 8.0 were not feasible because benzoic acid esters underwent hydrolysis.3131 Angelov, T.; Vlasenko, A.; Tashkov, W.; J. Liq. Chromatogr. Relat. Technol. 2007, 31, 188. Two processes occur simultaneously in the parabens molecules. The first process refers to the protonation of oxygen atom in the ester group (C=O), whereas the second process concerns deprotonation of the hydroxyl group.3131 Angelov, T.; Vlasenko, A.; Tashkov, W.; J. Liq. Chromatogr. Relat. Technol. 2007, 31, 188.

Protonation is relatively slow and strongly depends on the size of the ester aliphatic chain. Longer aliphatic chains increase the strength of the positive inductive effect, thereby raising the electron density on oxygen (C=O). Protonation usually takes place at pH below 3.0; partial deprotonation occurs at pH between 3.5 and 6.5; and molecules are completely deprotonated and ionized with a negative net charge at pH above 6.5.3131 Angelov, T.; Vlasenko, A.; Tashkov, W.; J. Liq. Chromatogr. Relat. Technol. 2007, 31, 188.

In pH 4, more efficient extraction of the parabens was obtained because the analytes were in their non-ionized form, and thus, increasing their solubility in the organic solvent (Figures 2d, 2e and 2f).

In summary, the conditions for the DLLME extraction for parabens from breast milk were optimized as follows: addition of 1 mL of acetonitrile to 1 g of breast milk sample to promote protein precipitation, followed by dilution of the supernatant with 10 mL of an ammonium acetate solution (pH 4 and 0% NaCl). Next, a mixture of acetonitrile (1 mL, disperser solvent) and dichloromethane (800 µL, extraction solvent) was used to extract the analytes from the aqueous phase.

Then, 750 µL of the organic phase was collected and transferred to an Eppendorf flask and dried. The dry extract was reconstituted in 80 µL of mobile phase (water/acetonitrile, 60:40 v/v) and injected into the LC-MS/MS system.

Analytical validation

Based on our assays, the parabens were stable in bench conditions (8 h) at controlled room temperature (25 ºC). According to the literature,1111 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538. parabens are stable in milk samples at -20 ºC for 45 days.

The proposed method was linear with coefficient of determination higher than 0.99. In addition, all the analytical curves showed p-value greater than 0.05 for the lack-of-fit test, which indicated that the linear model for the analytical calibration curves were well adjusted. Table 3 presented the linear regression equations, R2 values, and p-values obtained for each paraben.

Table 3
Linearity of the DLLME/UHPLC-MS/MS method for the determination of parabens in breast milk samples (n = 6)

For the accuracy, the residual standard error (RSE) values ranged from -19 (lower limit of quantification (LLOQ)) to 10.4% (intra-assay accuracy), and from -8.2 to 10.4% (inter-assay accuracy), while for precision, the assays presented coefficient of variation (CV) values ranging from 1.3 to 16.6% (LLOQ) (intra-assay precision) and 0.4 to 11.3% (inter-assay precision) (Table 4).

Table 4
Precision and accuracy (intra- and inter-assay) of the DLLME/UHPLC-MS/MS method for determination of parabens in breast milk samples

The matrix effect was evaluated by comparing the angular coefficients of six calibration curves plotted with three blank milk samples and three breast milk samples obtained from volunteer lactating mothers. The p-values obtained in these calibration curves were 0.2325, 0.1908, 0.1698, and 0.2079 for MePB, EtPB, PrPB, and BuPB, respectively. According to the t-test, there was no evidence (0.05 significance level) that the slopes were different.

Therefore, the results demonstrated that the calibration curves in commercial (blank) and breast (volunteers) milk samples were similar. Moreover, the similarities among the angular coefficients confirmed the minimum matrix effect of breast milk and the reliability of the proposed method. The post-column infusion also demonstrated that the biological matrix did not affect the ionization of the analytes significantly.

According to the chromatographic profile (Figure 3), the endogenous compound eluted around 1.6 min was related to methylparaben in the breast milk sample. In our study, it was not possible to obtain a breast milk sample completely free of parabens, thus, analytical curve was constructed using commercial milk samples spiked with the analytes. For the other parabens, the analytical signal did not enhance significantly.

Figure 3
Post-column infusion chromatogram of parabens and a single injection of a breast milk sample into the LC (superimposed).

The selectivity of the DLLME/UHPLC-MS/MS method was demonstrated through the chromatograms of the blank milk samples and blank milk samples spiked with the parabens at the concentration of 5 ng mL-1 (Figure 4). In these chromatograms, the interfering peaks did not co-elute with the parabens.

Figure 4
DLLME/UHPLC-MS/MS chromatograms (SRM) of a parabens-free milk sample (overwritten) and of parabens-free milk samples spiked with parabens (5 ng mL-1) and internal standard (250 ng mL-1).

Method comparison

Considering the analytical linear ranges of the methods and the concentration ranges of the parabens determined in different milk samples, Table 5 compares the results obtained in this work with those described in literature.

Table 5
DLLME method compared with other analytical methods used to determine parabens in milk samples

The DLLME/UHPLC-MS/MS developed method stands out because it uses smaller sample volume (as compared to QuEChERS77 Rodríguez-Gómez, R.; Jiménez-Díaz, I.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Talanta 2014, 130, 561. and SM-SLLME88 Rodríguez-Gómez, R.; Roldán-Pijuán, M.; Lucena, R.; Cárdenas, S.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26.) and requires shorter run time (as compared to online SPE).1010 Ye, X.; Bishop, A. M.; Needham, L. L.; Calafat, A. M.; Anal. Chim. Acta 2008, 622, 150. Compared to the method developed by Melo and Queiroz,1111 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538. which involved selective molecularly imprinted micro-disc solid-phase extraction (MISPE), the present work afforded lower LLOQ values (EtPB and PrPB). In addition, Melo and Queiroz1111 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538. did not analyze BuPB.

Regarding practical aspects involved in the sample preparation steps, DLLME is inexpensive, provides high extraction efficiency, is simple to operate, and produces smaller volume of waste solvent as compared to QuEChERS,77 Rodríguez-Gómez, R.; Jiménez-Díaz, I.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Talanta 2014, 130, 561. online SPE,1010 Ye, X.; Bishop, A. M.; Needham, L. L.; Calafat, A. M.; Anal. Chim. Acta 2008, 622, 150. MISPE,1111 Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538. and in-tube SPME.99 Souza, I. D.; Melo, L. P.; Jardim, I. C. S. F.; Monteiro, J. C. S.; Nakano, A. M. S.; Queiroz, M. E. C.; Anal. Chim. Acta 2016, 932, 49.

The DLLME method described by Vela-Soria et al.3737 Vela-Soria, F.; Jiménez-Díaz, I.; Díaz, C.; Pérez, J.; Iribarne-Durán, L. M.; Serrano-López, L.; Arrebola, J. P.; Fernández, M. F.; Olea, N.; Bioanalysis 2016, 8, 1777. to determine endocrine-disrupting chemicals in milk samples, presented lower LLOQ values. Probably, these results are related to the higher sensibility of the MS/MS detector. Furthermore, the proposed method herein presented LLOQ values adequate to determine parabens in breast milk samples, studied deeply the effects of DLLME factors through chemometric experimental design, and presented short analysis time.

Concentration of parabens in breast milk samples from volunteer lactating mothers

To demonstrate the applicability of the DLLME/UHPLC-MS/MS method, fifteen breast milk samples from volunteer lactating mothers were analyzed (Table 6).

Table 6
Concentrations of parabens in milk samples collected from postpartum women (n = 3)

Methyl and ethyl parabens were presented in all the samples with arithmetic mean concentrations of 24.51 and 12.07 ng mL-1, respectively. Propylparaben emerged in the majority of the samples, however, its concentration was lower than 5 ng mL-1 (LLOQ). Methyl and ethylparabens were detected at higher concentrations in breast milk samples because these two preservatives are frequently used in pharmaceutical, food, and cosmetic products. The results pointed out to the mothers' high exposure to these compounds since all the volunteers used cosmetics regularly.

The concentration range of parabens in breast milk determined in this study agreed with data of a previous study by our group,99 Souza, I. D.; Melo, L. P.; Jardim, I. C. S. F.; Monteiro, J. C. S.; Nakano, A. M. S.; Queiroz, M. E. C.; Anal. Chim. Acta 2016, 932, 49. conducted with 16 lactating mothers in Brazil. Using different sample preparation techniques, Rodríguez-Gómez et al.77 Rodríguez-Gómez, R.; Jiménez-Díaz, I.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Talanta 2014, 130, 561.,88 Rodríguez-Gómez, R.; Roldán-Pijuán, M.; Lucena, R.; Cárdenas, S.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26.,3333 Rodríguez-Gómez, R.; Zafra-Gómez, A.; Camino-Sánchez, F. J.; Ballesteros, O.; Navalón, A.; J. Chromatogr. A 2014, 1349, 69.,3434 Rodríguez-Gómez, R.; Dorival-García, N.; Zafra-Gómez, A.; Camino-Sánchez, F. J.; Ballesteros, O.; Navalón, A.; J. Chromatogr. B 2015, 992, 47. developed analytical methods and determined parabens in human milk samples with concentrations ranging from 0.1 to 7.5 ng mL-1, 0.5 to 37.0 ng mL-1, 0.6 to 43.5 ng mL-1 and 1.02 to 18.1 ng mL-1, respectively. Studies also conducted in Spain3737 Vela-Soria, F.; Jiménez-Díaz, I.; Díaz, C.; Pérez, J.; Iribarne-Durán, L. M.; Serrano-López, L.; Arrebola, J. P.; Fernández, M. F.; Olea, N.; Bioanalysis 2016, 8, 1777.

38 Azzouz, A.; Rascón, A. J.; Ballesteros, E.; J. Pharm. Biomed. Anal. 2016, 119, 16.
-3939 Vela-Soria, F.; Iribarne-Durán, L. M.; Mustieles, V.; Jiménez-Díaz, I.; Fernández, M. F.; Olea, N.; J. Chromatogr. A 2018, 1546, 1. determined parabens in a range from 0.12 to 8.1 ng mL-1, 0.9 to 21 ng mL-1 and from 0.7 to 18.6 ng mL-1, respectively. Hines et al.3535 Hines, E. P.; Mendola, P.; von Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120. determined parabens levels from 0.5 to 2.3 ng mL-1. In the United States,1010 Ye, X.; Bishop, A. M.; Needham, L. L.; Calafat, A. M.; Anal. Chim. Acta 2008, 622, 150. the average reported concentrations of parabens were 0.3-3.0 ng mL-1. Schlumpf et al.3232 Schlumpf, M.; Kypke, K.; Wittassek, M.; Angerer, J.; Mascher, H.; Mascher, D.; Vökt, C.; Birchler, M.; Lichtensteiger, W.; Chemosphere 2010, 81, 1171. established the presence of MePB, EtPB and PrPB in 26, 15 and 11% of samples, respectively. A study from Turkey3030 Alshana, U.; Ertaş, N.; Göğer, N. G.; Food Chem. 2015, 181, 1. detected parabens in milk samples in a concentration range from 300 to 6000 ng mL-1. In Iran,3636 Asiabi, H.; Yamini, Y.; Seidi, S.; Esrafili, A.; Rezaei, F.; J. Chromatogr. A 2015, 1397, 19.,4040 Fotouhi, M.; Seidi, S.; Shanehsaz, M.; Naseri, M. T.; J. Chromatogr. A 2017, 1504, 17. parabens levels were determined in a range from 0.792 to 1.180 ng mL-1 and 51.2 to 3270.3 ng mL-1, respectively. Fisher et al.4141 Fisher, M.; MacPherson, S.; Braun, J. M.; Hauser, R.; Walker, M.; Feeley, M.; Mallick, R.; Bérubé, R.; Arbuckle, T. E.; Environ. Sci. Technol. 2017, 51, 4009. determined parabens levels in milk samples in a range from 0.0023 to 16.325 ng mL-1.

These studies show that parabens are present in breast milk and these compounds are transferred to babies, not only during pregnancy period, but it extends throughout breast feeding. The literature4242 Kirchhof, M. G.; de Gannes, G. C.; Skin Ther. Lett. 2013, 18, 5. has suggested that the use and exposure to parabens were harmful, however, studies evaluating the effects of these compounds in long periods are insufficient, yet. The studied subset (demographics, age, gender, and other factors) and the number of collected samples constitute its limitations.

Conclusions

Through central composite design, the DLLME extraction variables (extraction solvent volume, disperser solvent volume, addition of NaCl, and extraction pH) were optimized obtaining the highest recovery of parabens from breast milk samples.

The DLLME extraction technique provided high enrichment factors for parabens analyzed in breast milk samples, minimal amount of sample and small volumes of organic solvent. Low cost and simple operation requirements made the present technique noteworthy.

Furthermore, the DLLME/UHPLC-MS/MS method was successfully validated based on the current international guidelines of EMA and FDA.

Lastly, the application of DLLME/UHPLC-MS/MS method to analyze real breast milk samples obtained from 15 volunteer lactating mothers provided the assess of the extent to which lactating women are exposed to parabens. MePB and EtPB emerged in all samples with arithmetic mean concentrations of 24.51 and 12.07 ng mL-1, respectively. These results enforce the real common exposure of women in the postpartum period to parabens.

Acknowledgments

This work was supported by grants from Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP, process 2015/04060-3, 2015/07619-1 and 2017/02147-0).

References

  • 1
    Błędzka, D.; Gromadzińska, J.; Wąsowicz, W.; Environ. Int. 2014, 67, 27.
  • 2
    Jardim, V. C.; Melo, L. P.; Domingues, D. S.; Queiroz, M. E. C.; J. Chromatogr. B 2015, 974, 35.
  • 3
    Frederiksen, H.; Jørgensen, N.; Andersson, A.-M.; J. Exposure Sci. Environ. Epidemiol. 2011, 21, 262.
  • 4
    Vela-Soria, F.; Ballesteros, O.; Zafra-Gómez, A.; Ballesteros, L.; Navalón, A.; Talanta 2014, 129, 209.
  • 5
    Byford, J. R.; Shaw, L. E.; Drew, M. G. B.; Pope, G. S.; Sauer, M. J.; Darbre, P. D.; J. Steroid Biochem. Mol. Biol. 2002, 80, 49.
  • 6
    Carrasco-Correa, E. J.; Vela-Soria, F.; Ballesteros, O.; Ramis-Ramos, G.; Herrero-Martínez, J. M.; J. Chromatogr. A 2015, 1379, 65.
  • 7
    Rodríguez-Gómez, R.; Jiménez-Díaz, I.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Talanta 2014, 130, 561.
  • 8
    Rodríguez-Gómez, R.; Roldán-Pijuán, M.; Lucena, R.; Cárdenas, S.; Zafra-Gómez, A.; Ballesteros, O.; Navalón, A.; Valcárcel, M.; J. Chromatogr. A 2014, 1354, 26.
  • 9
    Souza, I. D.; Melo, L. P.; Jardim, I. C. S. F.; Monteiro, J. C. S.; Nakano, A. M. S.; Queiroz, M. E. C.; Anal. Chim. Acta 2016, 932, 49.
  • 10
    Ye, X.; Bishop, A. M.; Needham, L. L.; Calafat, A. M.; Anal. Chim. Acta 2008, 622, 150.
  • 11
    Melo, L. P.; Queiroz, M. E. C.; Anal. Methods 2013, 5, 3538.
  • 12
    Suh, J. H.; Lee, Y. Y.; Lee, H. J.; Kang, M.; Hur, Y.; Lee, S. N.; Yang, D.-H.; Han, S. B.; J. Pharm. Biomed. Anal. 2013, 75, 214.
  • 13
    Vela-Soria, F.; Ballesteros, O.; Rodríguez, I.; Zafra-Gómez, A.; Ballesteros, L.; Cela, R.; Navalón, A.; Anal. Bioanal. Chem. 2013, 405, 7259.
  • 14
    Frederiksen, H.; Taxvig, C.; Hass, U.; Vinggaard, A. M.; Nellemann, C.; Toxicol. Sci. 2008, 106, 376.
  • 15
    Jiménez-Díaz, I.; Vela-Soria, F.; Zafra-Gómez, A.; Navalón, A.; Ballesteros, O.; Navea, N.; Fernández, M. F.; Olea, N.; Vílchez, J. L.; Talanta 2011, 84, 702.
  • 16
    Vela-Soria, F.; Rodríguez, I.; Ballesteros, O.; Zafra-Gómez, A.; Ballesteros, L.; Cela, R.; Navalón, A.; J. Chromatogr. A 2014, 1371, 39.
  • 17
    Chaves, A. R.; Leandro, F. Z.; Carris, J. A.; Queiroz, M. E. C.; J. Chromatogr. B 2010, 878, 2123.
  • 18
    Trufelli, H.; Palma, P.; Famiglini, G.; Cappiello, A.; Mass Spectrom. Rev. 2011, 30, 491.
  • 19
    Rezaee, M.; Assadi, Y.; Milani Hosseini, M.-R.; Aghaee, E.; Ahmadi, F.; Berijani, S.; J. Chromatogr. A 2006, 1116, 1.
  • 20
    Saraji, M.; Khalili Boroujeni, M.; Hajialiakbari Bidgoli, A. A.; Anal. Bioanal. Chem. 2011, 400, 2149.
  • 21
    Asensio-Ramos, M.; Hernández-Borges, J.; Borges-Miquel, T. M.; Rodríguez-Delgado, M. Á.; J. Chromatogr. A 2011, 1218, 4808.
  • 22
    Han, Y.; Jia, X.; Liu, X.; Duan, T.; Chen, H.; Chromatographia 2010, 72, 351.
  • 23
    Saraji, M.; Boroujeni, M. K.; Anal. Bioanal. Chem. 2014, 406, 2027.
  • 24
    European Medicines Agency (EMA); Reflection Paper on the Use of Methyl- and Propylparaben as Excipients in Human Medicinal Products for Oral Use; EMA: London, UK, 2015. Available at http://www.ema.europa.eu/docs/en_GB/document_library/Scientific_guideline/2015/11/WC500196733.pdf, accessed in August 2018.
    » http://www.ema.europa.eu/docs/en_GB/document_library/Scientific_guideline/2015/11/WC500196733.pdf
  • 25
    http://www.fda.gov/cosmetics/productsingredients/ingredients/ucm128042.htm, accessed in August 2018.
    » http://www.fda.gov/cosmetics/productsingredients/ingredients/ucm128042.htm
  • 26
    Zgoła-Grześkowiak, A.; Grześkowiak, T.; TrAC, Trends Anal. Chem. 2011, 30, 1382.
  • 27
    Kumar, N.; Bansal, A.; Sarma, G. S.; Rawal, R. K.; Talanta 2014, 123, 186.
  • 28
    Costa, N. R.; Lourenço, J.; Pereira, Z. L.; Chemom. Intell. Lab. Syst. 2011, 107, 234.
  • 29
    Rezaee, M.; Yamini, Y.; Faraji, M.; J. Chromatogr. A 2010, 1217, 2342.
  • 30
    Alshana, U.; Ertaş, N.; Göğer, N. G.; Food Chem. 2015, 181, 1.
  • 31
    Angelov, T.; Vlasenko, A.; Tashkov, W.; J. Liq. Chromatogr. Relat. Technol. 2007, 31, 188.
  • 32
    Schlumpf, M.; Kypke, K.; Wittassek, M.; Angerer, J.; Mascher, H.; Mascher, D.; Vökt, C.; Birchler, M.; Lichtensteiger, W.; Chemosphere 2010, 81, 1171.
  • 33
    Rodríguez-Gómez, R.; Zafra-Gómez, A.; Camino-Sánchez, F. J.; Ballesteros, O.; Navalón, A.; J. Chromatogr. A 2014, 1349, 69.
  • 34
    Rodríguez-Gómez, R.; Dorival-García, N.; Zafra-Gómez, A.; Camino-Sánchez, F. J.; Ballesteros, O.; Navalón, A.; J. Chromatogr. B 2015, 992, 47.
  • 35
    Hines, E. P.; Mendola, P.; von Ehrenstein, O. S.; Ye, X.; Calafat, A. M.; Fenton, S. E.; Reprod. Toxicol. 2015, 54, 120.
  • 36
    Asiabi, H.; Yamini, Y.; Seidi, S.; Esrafili, A.; Rezaei, F.; J. Chromatogr. A 2015, 1397, 19.
  • 37
    Vela-Soria, F.; Jiménez-Díaz, I.; Díaz, C.; Pérez, J.; Iribarne-Durán, L. M.; Serrano-López, L.; Arrebola, J. P.; Fernández, M. F.; Olea, N.; Bioanalysis 2016, 8, 1777.
  • 38
    Azzouz, A.; Rascón, A. J.; Ballesteros, E.; J. Pharm. Biomed. Anal. 2016, 119, 16.
  • 39
    Vela-Soria, F.; Iribarne-Durán, L. M.; Mustieles, V.; Jiménez-Díaz, I.; Fernández, M. F.; Olea, N.; J. Chromatogr. A 2018, 1546, 1.
  • 40
    Fotouhi, M.; Seidi, S.; Shanehsaz, M.; Naseri, M. T.; J. Chromatogr. A 2017, 1504, 17.
  • 41
    Fisher, M.; MacPherson, S.; Braun, J. M.; Hauser, R.; Walker, M.; Feeley, M.; Mallick, R.; Bérubé, R.; Arbuckle, T. E.; Environ. Sci. Technol. 2017, 51, 4009.