vol.30 issue2Controlled Dehydration of Fe(OH)3 to Fe2O3: Developing Mesopores with Complexing Iron Species for the Adsorption of β-Lactam AntibioticsPotential Slow Release Fertilizers and Acid Soil Conditioners Obtained by One-Pot Mechanochemical Activation of Chrysotile:Cement Roofing Sheets with K2HPO4 author indexsubject indexarticles search
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

Share


Journal of the Brazilian Chemical Society

Print version ISSN 0103-5053On-line version ISSN 1678-4790

J. Braz. Chem. Soc. vol.30 no.2 São Paulo Feb. 2019

http://dx.doi.org/10.21577/0103-5053.20180180 

Articles

Fast Determination of Iodine Number of Biodiesel Using Capillary Zone Electrophoresis with Multi- and Single-Point Calibration

Alysson V. F. Sakoa 

Gustavo A. Micke*  a 
http://orcid.org/0000-0002-5518-624X

aDepartamento de Química, Universidade Federal de Santa Catarina, 88040-900 Florianópolis-SC, Brazil

ABSTRACT

A new and fast method was developed to determine the iodine number of biodiesel samples by capillary zone electrophoresis (CZE). In this procedure, the biodiesel sample reacts with Wijs solution in the presence of a catalyst and all the remaining iodine monochloride is converted to the iodide ion, which is quantified by CZE. Only small amounts of sample and reactant are required and µL level amounts of chloroform and Wijs solution. The reaction time was reduced to 3 min and the total analysis time is less than 15 min. The simplicity and speed of the method can be further enhanced by using single-point calibration. The limits of quantification (LOQ) and determination (LOD) values were better than 0.6 and 0.18 g per 100 g of biodiesel. Instrumental, intraday, and interday precisions were better than 1.13, 1.04 and 6.02% relative standard deviation (RSD). The accuracy was statistically verified by comparison with the standard method EN 14111.

Keywords: iodine number; biodiesel; capillary zone electrophoresis; Wijs

Introduction

Biodiesel is an important alternative energy source. Compared to mineral diesel, biodiesel has several advantages, such as renewability, biodegradability, no toxic effects on the environment,1,2 compatibility with diesel engines2,3 and it usually contains no sulfur or aromatic compounds.1,4 Biodiesel is produced by a transesterification reaction, where fatty acids are released from triacylglycerols and these react with alcohols to form the corresponding esters.1,2,5,6

It is possible to produce biodiesel through vegetable6 or animal7 sources. All vegetable oils contain oleic (18:1) and linoleic (18:1(n-19)) fatty acids. Some vegetables, for instance, soy, linseed and canola, contain the α-linoleic fatty acid (18:3(n-9)).8 Therefore, it is possible to produce biodiesel with a high unsaturated content using vegetable oils. Biodiesels obtained from sunflower, rubber seed, jatropha and cotton seed, for example, may contain more than 70 wt.% of unsaturated fatty acid methyl esters.9 Unsaturated compounds are also presented in the biodiesel produced from animal sources. Tallow biodiesel, for example, has a calculated degree of unsaturation of 1.5, but this is low compared to vegetable biodiesels, such as linseed (3.2) and safflower (2.7).7

There is some concern regarding the degree of unsaturation of biodiesel. At high temperatures, polymerization can occur, thickening the fuel.9 Consequently, the lubrication in the injection nozzle and the combustion cylinder decreases and this affects the engine performance.5,10,11 The engine performance is not always affected by the degree of unsaturation of the biodiesel; however, an increase in this parameter can increase the peak heat release rate, maximum pressure gradient and peak in-cylinder bulk gas-averaged temperature in the engine.12 Also, biodiesel samples with a higher degree of unsaturation have exhibited higher NOx emission profiles.12,13

The degree of unsaturation is evaluated by the iodine number (IN), which is a biodiesel quality parameter and indirectly represents the biodiesel polymerization capacity.5 The IN was also correlated to the oxidation stability. However, this correlation has been questioned because there are natural antioxidants and free fatty acids that are not measured by the IN, but which greatly influence the oxidation stability of oils and biodiesel.14 Therefore, two oils with the same IN may have different oxidation stability.15

European Standards (ENs) and documents published by the American Society for Testing Materials (ASTM) have specified the quality parameters of biodiesel.2,5 The IN is established in the standard methods Deutsches Institut für Normung (DIN) EN 14111,16 EN 142144 and EN 14213,4 with a limit of 120 g of I2 per 100 g of biodiesel.4 The Brazilian National Agency of Petroleum, Natural Gas and Biofuels (Agência Nacional do Petróleo, Gás Natural e Biocombustíveis, ANP) only stipulates that the IN value should be reported.17 The method EN 14111 is based on the reaction between ICl and unsaturated long chain alkyl esters. This procedure is commonly known as the Wijs method.16 Alternative methods have been developed to determine the IN of biodiesel samples based on potentiometry titration,5 gas chromatography,181H nuclear magnetic resonance (NMR) spectroscopy,6 spot test exploiting of digital images19 and a spectrophotometric flow-based procedure.20

In this context, a new capillary zone electrophoresis (CZE) method is proposed herein for the determination of the IN of biodiesel samples. The principles of the Wijs method, with adaptations, were applied to obtain a fast, simple, and green analysis method. Detailed information on the method is provided in the following sections.

Experimental

Chemicals, reagents and samples

Analytical grade chemicals were obtained. Sodium nitrate solution (1000 mg L-1), potassium iodide solution (1000 mg L-1 I), iodine monochloride (1 mol L-1 ICl in acetic acid, Wijs solution), 6-aminocaproic acid, and sodium acetate were obtained from Sigma-Aldrich. Glacial acetic acid and HCl were obtained from Synth. Sodium thiosulfate (2 mol L-1 Na2S2O3) was obtained from Specsol. Biodiesel samples were obtained from 5 different production plants and labeled as A, B, C, D and E. INs of two biodiesel samples (A and E) were quantified according to the standard method DIN EN 1411116 by different laboratories. Sample A was evaluated by 12 laboratories and sample E by one laboratory.

Instrumentation and system configuration

The method was carried out on an Agilent Technologies capillary electrophoresis system (model 7100 CE). Two wavelengths were observed in the proposed method: 230 nm for I ion (analyte) and 210 nm for NO3 (internal standard). The applied separation voltage was –30 kV (negative on the injection side). The cartridge temperature was controlled at 25 ºC. Hydrodynamic injection was performed for 5 s at 50 mbar. A silica capillary (48.5 cm × 50 µm i.d.) with an external coat of polyimide was selected. Before the first use, the capillary was conditioned with 1 mol L-1 NaOH for 30 min and deionized water for 30 min. Before the first day of use, the capillary was conditioned with 1 mol L-1 NaOH, deionized water and background electrolyte (BGE), with 5 min for each step. Between runs, capillary preconditioning was performed by injecting the BGE for 1 min using the flush command (ca. 950 mbar) of the Chemstation software.

BGE development and preparation

The BGE composition was studied using the Peakmaster 5.3 freeware.21 The selected BGE components were 6-aminocaproic acid and HCl. Simulations using 6-aminocaproic acid, ranging from 20 to 100 mmol L-1, and 10 mmol L-1 HCl were performed. The optimized BGE for this method was composed of 50 mmol L-1 6-aminocaproic acid adjusted to pH 5.02 with HCl.

Catalyst preparation

The catalyst was developed prior to the experiment and it is composed of sodium acetate and glacial acetic acid.22 For this method, the catalyst solution was prepared with approximately 1 g of sodium acetate dissolved in 5000 µL of glacial acetic acid.

Sample preparation procedure and Wijs reagent quantification

Approximately 50 mg of the biodiesel samples were placed in a volumetric flask (25 mL) and 100 µL of the catalyst solution were added to the sample. In the next step, 400 µL of 1 mol L-1 ICl dissolved in acetic acid were added. The reaction was carried out in the dark, with occasional stirring, for at least 3 min. Next, 1000 µL of 2 mol L-1 sodium thiosulfate were added. The solution was stirred until all of the dark brown drops had disappeared. The addition of some deionized water in this step helps to accelerate the process. The flask volume was completed with deionized water to the meniscus and 500 µL of the solution were placed in an Eppendorf tube. To perform the liquid-liquid extraction, 200 µL of chloroform were also added to the Eppendorf tube. The tube was vortexed for 10 s and centrifuged for 60 s at 14500 rpm. After centrifugation, 50 µL of the supernatant were diluted with 50 µL of 1000 mg L-1 nitrate solution and 900 µL of deionized water, and 700 µL of the resulting solution were then transferred to a vial for injection into the CE equipment. The Wijs solution was also quantified in triplicate by the same procedure described above, but without the biodiesel sample. The calculated average concentration was used to determine the ICl added initially to the flasks for the calculation of the IN of the biodiesel samples.

Validation of the multi-point calibration and single-point calibration methods

The validation was performed for conventional multi-point calibration (MPC) and single-point calibration (SPC).23,24 The method parameters of specificity, linearity, linear range, limits of quantification (LOQ) and detection (LOD), accuracy, intraday, interday and instrumental precision25 were evaluated. Calibration curves were prepared in triplicate in the range of 10 to 90 mg L-1 of I. The linearity of the method was observed from the Fcalculated and r2 values.23 External calibration curves were prepared with 50 mg L-1 NO3 (internal standard) and 4 mmol L-1 Na2S2O3.

Instrumental precision was evaluated injecting repeatedly (n = 10) a 40 mg L-1 iodide solution. Intraday precision was evaluated through the preparation of each of the calibration curve solutions (n = 8) in triplicate. Interday precision was evaluated by preparing the calibration curve solutions on three different days. The results for the precision parameters are given in percentage relative standard deviation (% RSD) values.

Accuracy was evaluated comparing statistically (t-test) the values obtained for the IN according to the standard method (DIN EN 14111)16 for two different samples. LOQ and LOD for the proposed method were calculated considering signal-to-noise (S/N) ratios of 10:1 and 3:1 for MPC and calculated according to Olivieri23 for SPC.

Results and Discussion

An overview of the method

The Wijs reaction involves an electrophilic addition where the double bond is cleaved and both I and Cl are added to unsaturated fatty acid chains.26 In the conventional Wijs method, the ICl is added in excess in relation to the sample. After the reaction of ICl with the unsaturations, I is added to convert the remaining ICl to I2 and Cl, and in the final step I2 is titrated with S2O32– in the presence of starch. The titration ends when the blue color disappears, and it is possible to evaluate the IN by the difference between the initial amount of ICl added in excess, subtracted from the remaining ICl, determined by the titration.5 In the proposed procedure, the Wijs method was used with some adaptations. First, a catalyst was added to accelerate the reaction from 30 to 3 min.22 The volume and catalyst concentrations were prepared considering that the sodium acetate concentration is approximately 3-5% m/m during the reaction step. A good strategy is to convert the remaining ICl in the solution to I and Cl by adding an excess of S2O32–, and then analyze the I by CZE-UV. Data on half reactions involving ICl, S2O32– and I2, and their standard reduction potentials27 provide information on the global reaction. The calculated global standard reduction potential for the direct reaction between ICl and S2O32– is positive (Table 1) and, therefore, the reaction is spontaneous.28

Table 1 Half-reactions, standard reduction potentials and global standard reaction for the proposed direct reaction between ICl and S2O32– 

Semi-reactiona Standard reduction potentiala / V
2IC1s+2eI2+2C1 E0red = +1.22
S2O32+3H2O2H2SO3+3H++4e E0red = –0.40
I2aq+2e2I E0red = +0.62
Global reaction Global standard potential / V
2IC1s+S2O32+3H2O2H2SO3+3H++2I+2C1 E0global red = +1.44

aData adapted from Harris.27 E0red: standard reduction potential; E0global red: global standard reduction potential.

Figure 1 shows a simplified model of the method and an electropherogram of sample A. It is possible to determine the IN from the difference between the iodine from the Wijs solution added initially and the remaining I quantified by CZE-UV.

Figure 1 Simplified model for the proposed method. Electropherogram of biodiesel sample A observed at 230 nm. CE configuration: fused-silica capillary of 48.5 cm × 50 µm i.d.; hydrodynamic injection: 50 mbar for 5 s; separation voltage: –30 kV; cartridge temperature: 25 ºC; BGE composition: 50 mmol L-1 6-aminocaproic acid adjusted to pH 5.02 with HCl. Detailed information of the method is described in the Experimental section. 

Background electrolyte optimization

Peakmaster freeware21 is a powerful tool to simulate peak separation in electrophoresis and it contains the values for the pKa and ionic mobility (µionic) of several organic and inorganic compounds.29 The Peakmaster data in this section were obtained from the freeware and the strategy for the BGE development is described below.

Efficient peaks are obtained when the effective mobility (µeff) of an analyte and its co-ion are similar.29,30 Under this condition, the electromigration dispersion (EMD) of the analyte is reduced to almost zero. Therefore, Cl is a good choice of co-ion for I because their µionic values are –79.1 × 10-9 and –79.6 × 10-9 m2 V-1 s-1, respectively. Since HCl and HI have the same pKa, their µeff values are approximately equivalent, independently of the pH. Also, Cl does not show UV absorbance at the analysis wavelength. Nitrate was selected as the internal standard because of its µionic value (74.1 × 10-9 m2 V-1 s-1) and UV absorbance. There are S2O32– ions present, so these were also considered. The S2O32– is in excess in the solution and, thus, the associated broad peak was expected to hinder the separation. H2S2O3 has two pKa values: 0.6 and 1.72. If the pH is greater than 2.72, the S2O32– form represents practically 100%. The µeff value for anionic species is related to the fraction of its ionized forms,31 therefore changes in the BGE pH may change the effective mobility of the analyte. Thus, a pH higher than 2.72 was used so that the µeff of the S2O32– ion would be more constant.

In this study, 6-aminocaproic acid was selected because it does not absorb at the wavelength of analysis and it is an amino acid with two pKa values: 4.43 and 10.75.32 The total charge of 6-aminocaproic acid is 0 or +1 if the pH is lower than 9.75, because the amino group is still protonated. Therefore, a 6-aminocaproic acid ion will not act as a co-ion for negative analytes until the pH is higher than 9.75, when the amino group starts to deprotonate. When there is more than one co-ion for the analyte in the BGE, system peaks that can interfere in the analysis are also present.33

The buffer capacity of the BGE in the acid range is between 3.43 and 5.43. To keep the ionic strength low, only 10 mmol L-1 HCl was used. The buffer capacity increases from 11.81 to 19.07 mmol L-1 when the 6-aminocaproic acid is added at concentrations ranging from 20 to 50 mmol L-1, and at 100 mmol L-1 the buffer capacity does not increase significantly (21.541 mmol L-1). The EMD values for I, NO3 and S2O32– show no significant variation in the range studied and are all close to zero. Therefore, 10 mmol L-1 HCl and 50 mmol L-1 6-aminocaproic acids were chosen for the BGE composition. However, it can be observed in Figure 1 that, for S2O32– and NO3, the experimental peaks are relatively broad in comparison with simulations. Therefore, the experimental and simulated EMDs are different. The S2O32– ion is in excess then the peak broadening is expected. The peak broadening for NO3 is not a problem because it still has an intense UV absorption signal in 210 nm for the concentration utilized. Table S1 in the Supplementary Information (SI) section contains the data obtained from the simulations carried out in Peakmaster.

Validation of the proposed method and sample analysis

The results of the method validation are reported in Table 2. Equation 1 was used to evaluate the linearity of the method for MPC and SPC.24,35 The Fcalculated value was lower than Fcritical = 2.74 (Table S2, SI section), which means that there is no lack of fit for the regression models. To obtain the response factor (Rf) for SPC, equation 2 was plotted using the average responses ( Ari/AriNO3 ), for each concentration level i of I (Ci), with a constant concentration of NO3 (CiNO3-). The average peak areas of I and NO3 were abbreviated as Ari and AriNO3 , respectively. The response factor was represented by Rf. Based on statistical analysis,34 the linear coefficient for SPC was neglected.

Fcalculated=Σi=1pmiyiŷi2/p2Σi=1pΣj=hmiyijyi2/mp (1)

AriAriNO3=RfCiCiNO3 (2)

Table 2 Validation parameters for MPC and SPC methods 

Parameter Multi-point calibration (MPC) Single-point calibration (SPC)
Linear regression equation Ar1/ArNO3=0.0125C1+0.0062 Ar1/ArNO3=0.6239C1/CNO3
Linear range / (mg L-1) 5.0-90.0
Linearity Fcalculated=0.296<F0.05,5,14=Fcritical=2.958;r2=0.9998
Instrumental precision / (% RSD) 1.13
Intraday precision / (% RSD) 0.036-1.04
Interday precision / (% RSD) 1.50-6.02
LOQb / (g I2per 100g of biodiesel) 0.60 0.165
LODc / (g I2per 100g of biodiesel) 0.18 0.055
Accuracyd Accurate with 95% of confidence (t-test)
reference sample A:
tcalculated = −0.87
(tcritical = 2.16)
reference sample E:
tcalculated = 2.70
(tcritical = 2.78)
reference sample A:
tcalculated = −1.72
(tcritical = 2.16)
reference sample E:
tcalculated = 2.39
(tcritical = 2.78)

aSince –tα / 2,n – 2 < b0 / s0 < +tα / 2,n – 2 (where α is the confidence level; b0 is the intercept from the linear regression; and s0 is the standard error of the intercept from the linear regression), the null hypothesis (intercept = 0) is accepted and single-point calibration is possible;34

bconsidering the sample mass and dilution in the “Sample preparation procedure and Wijs reagent quantification” sub-section. Based on S/N ratio of 10:1 for MPC, and based on equations 2 and 4 for SPC. If converted to ICl values, limits of quantification (LOQs) are 0.38 and 0.14 g per 100 g of biodiesel for MPC and SPC, respectively;

cconsidering the sample mass and dilution in the “Sample preparation procedure and Wijs reagent quantification” sub-section. Based on S/N ratio of 3:1 for external calibration curve, and based on equations 3 and 4 for SPC. If converted to ICl values, limits of detection (LODs) are 0.12 and 0.05 g per 100 g of biodiesel for MPC and SPC, respectively;

dreference sample A was analyzed by 12 laboratories and reference sample E by 1 laboratory using the DIN EN 1411116 standard method. Fcalculated and Fcritical: calculated and tabulated (Fisher-Snedecor distribution) F-test values; r2: coefficient of determination; RSD: relative standard deviation; tcalculated and tcritical: calculated and tabulated Student’s t-values.

From equation 1, p is the total number of calibration levels; 1 is the specific calibration level; mi, yi and ŷi are the number of replicates, average response and estimated response for a specific calibration level, respectively; yij is the experimental response for a specific calibration level and replicate.

The precision parameters were very satisfactory, with the instrumental and intraday precisions being better than 1.13% RSD and interday precision ranging from 1.50 to 6.02% RSD. The LOQ and LOD of the proposed method had different values for MPC and SPC. For SPC, equations 3, 4 and 5 were applied as recommended.23 Despite the differences between the LOQs and LODs for MPC and SPC, in both cases they were good enough to cover the real range of interest, since a biodiesel sample will rarely have IN values lower than 1.

LOD=3.3Sy/xA1+h0+1/I (3)

LOQ=10Sy/xA1+h0+1/I (4)

h0=Ccal2Σi=1ICiCcal2 (5)

where Sy / x is the residual standard deviation; A is the slope of the linear regression of equation 2; h0 is the leverage for the blank sample; I is the total calibration solutions prepared (I = mip); and Ccal is the average concentration of calibration curve levels Ccal=i=1pCi/p .

The t-test results verified the accuracy of the method with a 95% confidence level using the MPC and SPC models. The calculated IN for the samples using MPC and SPC (Table 3) ranged from 57.25 g of I2 per 100 g of sample to values close to 131.60 g of I2 per 100 g of sample. Four of the five samples were close to the maximum IN limit stipulated in the EN standards. Figure 1 shows an electropherogram of sample A at 230 nm. The method is also specific for the analyte, which is in agreement with the peak purity of I provided by the Chemstation software and accuracy results.

Table 3 IN found using the proposed method for five different biodiesel samples calculated using MPC and SPC methods 

Biodiesel sample Iodine number / (g I2per 100 g of sample)
Multi-point calibration (MPC) Single-point calibration (SPC)
Aa 132 ± 1 131 ± 1
B 126 ± 3 126 ± 3
C 128 ± 5 128 ± 5
D 124 ± 5 124 ± 5
Eb 58 ± 3 57 ± 3

aReference method: DIN EN 14111,16 quantified by 12 laboratories. Reference value: 130 ± 2 g I2 per 100 g of sample;

breference method: DIN EN 14111,16 quantified by 1 laboratory. Reference value: 62 ± 1 g I2 per 100 g of sample.

The Wijs solution was also quantified using the proposed method since its concentration in the specification ranged from 0.8 to 1.2 mol L-1. The concentration determined using the proposed method was 1.04 ± 0.09 mol L-1. This step is important to increase the precision and accuracy of the reagent. Also, since the Wijs solution passes through the same steps in the quantification of biodiesel samples, this is a manner of compensating for errors associated with the pipetting or dilution of the solutions.

Statistical study of SPC in comparison with MPC

A statistical study was conducted to compare the quantification by Rf and MPC. As previously noted, Rf is used in SPC and it can be obtained from the slope of the linear regression using equation 2. However, SPC is used to avoid the need for the preparation of a calibration curve and reduce the number of experiments. Therefore, the injection of a single level i of the calibration curve should represent the total group of levels and the same sample concentrations as MPC should be found. In other words, instead of using the Rf measured through the linear regression, it is possible to calculate the response factor for each i calibration level, using equation 2, directly through the responses and known Ci concentrations. This last calculation provides the p response factors, and each response factor is an . In SPC, the aim is to find the i concentration levels where leads to the same sample concentration as MPC. Thus, the t-test was applied to verify the levels where gives statistically equivalent concentrations for the samples in relation to MPC. Figure 2 shows the t-value calculated for each level i. The values are closer to the Rf value when concentrations between 20 and 90 mg L-1 of I are used in the MPC, because the tcalculated is located within the accepted range (–2.77 to + 2.77). This does not mean that it is not possible to quantify diluted samples where the response is equivalent to a calibration point of 10 mg L-1 of I or lower. The results in Figure 2 indicate that, to measure Rfiapproximately equal to Rf, the analyst should choose a calibration point in the range of 20 to 90 mg L-1 of I. For better results, concentration levels where tcalculated is approximately equal to 0 are the best choice to perform the SPC, which occurs between 30 to 90 mg L-1 of I.

Figure 2 Comparison between MPC and SPC using t-test with the response factors for each i level (Rfi). Between the accepted t-zone, Rf is approximately equal to Rf and SPC is statistically equivalent to the MPC results for biodiesel sample quantifications. 

Advantages of CZE-UV for IN determination

There are several advantages to using CZE to measure the IN. In comparison to visual or coulometric titration methods, CZE has no visual error, low consumption of solvent, reagents and sample (from mL to µL) and better analytical frequency and specificity. The CZE-UV system has low cost operation compared to gas chromatography and 1H NMR spectroscopy. Also, the CE equipment has a powerful versatility and can be used to perform important analysis for the quality control of biodiesel, for instance, the determination of inorganic cations,36,37 anions37 and free and total glycerol.38

Conclusions

The proposed method showed good performance in the determination of the IN of biodiesel samples by CZE-UV. In comparison with EN standard methods, the use of Wijs solution and chloroform was reduced from mL to µL levels. The sample preparation is easy, simple and fast. The catalyst shows good performance and the reaction time was reduced from 30 to 3 min. Also, the method has clear advantages in terms of operational costs, reagent and solvent consumption, waste generation and analytical frequency in comparison with other published methods. Both calibration methods are accurate at the 95% confidence level and statistically equivalent to the standard method EN 14111. The statistical studies verified the applicability of the SPC model, which can be used for faster IN determination compared to MPC, but shows the same performance if the recommended considerations are taken into account.

Supplementary Information

Supplementary information is available free of charge at http://jbcs.sbq.org.br as PDF file.

Acknowledgments

The authors would like to thank INCT-Catálise, Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) and Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) for the scientific incentives and financial support.

References

1 Peiter, A. S.; Lins, P. V. S.; Meili, L.; Soletti, J. I.; Carvalho, S. H. V.; Pimentel, W. R. O.; Meneghetti, S. M. P.; J. King Saud Univ., Sci., in press, DOI: 10.1016/j.jksus.2018.01.010. [ Links ]

2 Thoai, D. N.; Kumar, A.; Prasertsit, K.; Tongurai, C.; Energy Procedia 2017, 138, 544. [ Links ]

3 Efe, Ş.; Ceviz, M. A.; Temur, H.; Renewable Energy 2018, 119, 142. [ Links ]

4 Saluja, R. K.; Kumar, V.; Sham, R.; Renewable Sustainable Energy Rev. 2016, 62, 866. [ Links ]

5 Tubino, M.; Aricetti, J. A.; Fuel 2013, 103, 1158. [ Links ]

6 Shimamoto, G. G.; Tubino, M.; Talanta 2018, 179, 816. [ Links ]

7 Balakrishnan, A.; Parthasarathy, R. N.; Gollahalli, S. R.; Fuel 2016, 182, 798. [ Links ]

8 Dijkstra, A. J. In Encyclopedia of Food and Health; Caballero, B.; Finglas, P. M.; Toldrá, F., eds.; Academic Press: Oxford, 2016, p. 357-364. [ Links ]

9 Gopinath, A.; Puhan, S.; Govindan, N.; Int. J. Energy Environ. 2010, 1, 411. [ Links ]

10 Leung, D. Y. C.; Koo, B. C. P.; Guo, Y.; Bioresour. Technol. 2006, 97, 250. [ Links ]

11 Kalayasiri, P.; Jeyashoke, N.; Krisnangkura, K.; J. Am. Oil Chem. Soc. 1996, 73, 471. [ Links ]

12 Benjumea, P.; Agudelo, J. R.; Agudelo, A. F.; Energy Fuels 2011, 25, 77. [ Links ]

13 Altun, Ş.; Fuel 2014, 117, 450. [ Links ]

14 Lapuerta, M.; Rodríguez-Fernández, J.; de Mora, E. F.; Energy Policy 2009, 37, 4337. [ Links ]

15 Knothe, G.; J. Am. Oil Chem. Soc. 2002, 79, 847. [ Links ]

16 En 14111: Fat and Oil Derivatives - Fatty Acid Methyl Esters (FAME), Determination of Iodine Value, European Committee for Standardization, Berlin, 2003. [ Links ]

17 Agência Nacional do Petróleo, Gás Natural e Biocombustíveis (ANP); Resolução ANP No. 7, de 19.3.2008-DOU 20.3.2008. [ Links ]

18 Bondioli, P.; Della Bella, L.; Gallonzelli, A.; Riv. Ital. Delle Sostanze Grasse 2011, 88, 77. [ Links ]

19 Soares, S.; Lima, M. J. A.; Rocha, F. R. P.; Microchem. J. 2017, 133, 195. [ Links ]

20 Pereira, A. C.; Rocha, F. R. P.; Anal. Chim. Acta 2014, 829, 28. [ Links ]

21 JaroŠ, M.; HruŠka, V.; Štědrý, M.; Zusková, I.; GaŠ, B.; Electrophoresis 2004, 25, 3080. [ Links ]

22 Makino, A.; US pat. 4297106A 1980. [ Links ]

23 Olivieri, A. C.; Anal. Chim. Acta 2015, 868, 10. [ Links ]

24 Danzer, K.; Currie, L. A.; Pure Appl. Chem. 1998, 70. [ Links ]

25 Taverniers, I.; De Loose, M.; Van Bockstaele, E.; TrAC, Trends Anal. Chem. 2004, 23, 535. [ Links ]

26 De La Mare, P. B. D.; Bolton, R.; Electrophilic Additions to Unsaturated Systems, 2nd ed.; Elsevier Science: Amsterdam, 2013. [ Links ]

27 Harris, D. C.; Quantitative Chemical Analysis, 7th ed.; W. H. Freeman: New York, 2007. [ Links ]

28 Bratsch, S. G.; J. Phys. Chem. Ref. Data 1989, 18, 1. [ Links ]

29 BeneŠ, M.; Svobodová, J.; HruŠka, V.; Dvořák, M.; Zusková, I.; Gaš, B.; J. Chromatogr. A 2012, 1267, 109. [ Links ]

30 Shallan, A.; Guijt, R.; Breadmore, M. In Encyclopedia of Forensic Sciences, 2nd ed.; Siegel, J. A.; Saukko, P. J.; Houck, M. M., eds.; Academic Press: Waltham, 2013, p. 549-559. [ Links ]

31 Dolzan, M. D.; Spudeit, D. A.; Azevedo, M. S.; Costa, A. C. O.; de Oliveira, M. A. L.; Micke, G. A.; Anal. Methods 2013, 5, 6023. [ Links ]

32 O’Neil, M. J.; The Merck Index: An Encyclopedia of Chemicals, Drugs, and Biologicals; Merck & Co.: Whitehouse Station, 2013. [ Links ]

33 Štědrý, M.; JaroŠ, M.; GaŠ, B.; J. Chromatogr. A 2002, 960, 187. [ Links ]

34 Bánfai, B.; Kemény, S.; J. Chemom. 2012, 26, 117. [ Links ]

35 Faria, A. F.; de Souza, M. V. N.; de Oliveira, M. A. L.; J. Braz. Chem. Soc. 2008, 19, 389. [ Links ]

36 Piovezan, M.; Costa, A. C. O.; Jager, A. V.; de Oliveira, M. A. L.; Micke, G. A.; Anal. Chim. Acta 2010, 673, 200. [ Links ]

37 Nogueira, T.; do Lago, C. L.; Microchem. J. 2011, 99, 267. [ Links ]

38 Gonçalves Filho, L. C.; Micke, G. A.; J. Chromatogr. A 2007, 1154, 477. [ Links ]

Received: July 5, 2018; Accepted: September 17, 2018

Creative Commons License This is an Open Access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.