Automatized Separation of Fractions from Petroleum Based on Spectrophotometric Signal Derivative Using Open-Source Hardware for the Determination of Ni and V Linked to Porphyrins

Abstract

An automated method based on open-source hardware was created for the separation of porphyrins fractions from crude oil, an essential step for the quantification of V and Ni as metalloporphyrins. Preparative liquid chromatography was implemented with a reverse phase gradient from MeOH to toluene. The system allows three solvents and fractions changes by using quaternary valves connected to an Arduino microcontroller, programmed on the basis of the absorbance signal measured and its first derivative. Additionally, the method was applied to crude oils (Venezuelan National Institute of Standards and Technology (NIST) 8505 and from a marine reservoir). V and Ni were determined by inductively coupled plasma mass spectrometry (ICP-MS) and the V-porphyrin profile was determined by high performance liquid chromatography (HPLC)-ICP-MS. The concentrations of the metals grow in the last eluting toluene fractions, associated with asphaltenes. On the other hand, V-porphyrins in the first eluting MeOH fractions showed profiles of labile compounds, with a higher percentage in the Brazilian crude oil (4.9 mg L-1, corresponding to 18% of total V), in comparison to the one from the biodegraded NIST 8505 (59 mg L-1, corresponding to 7% of total V). The overall methodology has a good reproducibility and provides information on elements concentrations associated with porphyrin class compounds.

Keywords:
automated separation; V and Ni porphyrins; absorbance derivative; ICP-MS; HPLC-ICP-MS; open-source scientific hardware


Introduction

Vanadium and nickel occurrence and stability in crude oils result from their complexation with porphyrins, which are tetrapyrrolic structures with high carbon number (> C20) that vary with origin and oil maturity.11 Ocampo, R.; Callot, H. J.; Albrecht, P.; Kintzinger, J. P.; Tetrahedron Lett. 1984, 25, 2589.

2 Filby, R. H.; Van Berkel, G. J.; ACS Symp. Ser. 1987, 344, 2.
-33 Ocampo, R.; Riva, A.; Trendel, J. M.; Riolo, J.; Callot, H. J.; Albrecht, P.; Energy Fuels 1993, 7, 191. The total V and Ni contents in crude oils are commonly determined in the industry with different objectives, such as in geochemical studies44 Doukkali, A.; Saoiabi, A.; Zrineh, A.; Hamad, M.; Ferhat, M.; Barbe, J. M.; Guilard, R.; Fuel 2002, 81, 467.

5 Galarraga, F.; Reategui, K.; Martïnez, A.; Martínez, M.; Llamas, J. F.; Márquez, G.; J. Pet. Sci. Eng. 2008, 61, 9.

6 Escobar, M.; Márquez, G.; Azuaje, V.; Da silva, A.; Tocco, R.; Fuel 2012, 97, 186.

7 Akinlua, A.; Sigedle, A.; Buthelezi, T.; Fadipe, O. A.; Mar. Pet. Geol. 2015, 59, 286.
-88 López, L.; Lo Mónaco, S.; Org. Geochem. 2017, 104, 53. or for petroleum refining, in order to prevent catalyst fouling.99 Panariti, N.; Del Bianco, A.; Del Piero, G.; Marchionna, M.; Appl. Catal., A 2000, 204, 203.,1010 Ali, M. F.; Abbas, S.; Fuel Process. Technol. 2006, 87, 573. Accurate and reproducible results are achieved with atomic spectrometries, and authors have reported good analytical performances and faster analysis with the direct introduction of the samples diluted in solvents in inductively coupled plasma mass spectrometry (ICP-MS), instead of their microwave assisted acid decomposition.1111 Duyck, C.; Miekeley, N.; Porto da Silveira, C. L.; Aucélio, R. Q.; Campos, R. C.; Grinberg, P.; Brandão, G. P.; Spectrochim. Acta, Part B 2007, 62, 939.

12 Sánchez, R.; Todolí, J. L.; Lienemann, C.-P.; Mermet, J.-M.; Spectrochim. Acta, Part B 2013, 88, 104.
-1313 Poirier, L.; Nelson, J.; Leong, D.; Berhane, L.; Hajdu, P.; Lopez-Linares, F.; Energy Fuels 2016, 30, 3783.

Recently, the growing interest for the hydroprocessing of heavy crude oils and distillation residues containing high metalloporphyrin contents have urged a better understanding of the structure and distribution of these compounds in the distillation fractions of petroleum, in order to predict demetalization strategies.1414 Rana, M. S.; Sámano, V.; Ancheyta, J.; Diaz, J. A. I.; Fuel 2007, 86, 1216.

15 Rana, M. S.; Ancheyta, J.; Sahoo, S. K.; Rayo, P.; Catal. Today 2014, 220-222, 97.
-1616 Jenifer, A. C.; Sharon, P.; Prakash, A.; Sande, P. C.; Energy Fuels 2015, 29, 7743. The fractionation of petroleum in a laboratory scale is usually performed by chromatographic techniques, such as preparative liquid chromatography (LC) and thin layer chromatography (TLC). The separation process is based on compounds elution in solvents or solvent mixtures with increasing polarity, and the obtained fractions are saturated hydrocarbons (S), aromatics and N, S, O compounds (A), resin (R) and asphaltenes (A), the latter being the heaviest, with structural similarity to coke. The results are expressed as mass percentage and the overall methodology is known as SARA.1717 Fan, T.; Buckley, J. S.; Energy Fuels 2002, 16, 1571.

18 Speight, J. G.; Oil Gas Sci. Technol. 2004, 59, 467.
-1919 Kharrat, A. M.; Zacharia, J.; Cherian, V. J.; Anyatonwu, A.; Energy Fuels 2007, 21, 3618. The American Society for Testing and Materials (ASTM) method for SARA2020 ASTM D2007: Standard Test Method for Characteristic Groups in Rubber Extender and Processing Oils and Other Petroleum-Derived Oils by the Clay-Gel Absorption Chromatographic Method, Philadelphia, 2016. is based on LC, the oil sample is deasphaltened and further deposited on a sorbent (silica gel or alumina) and separated into its SAR fractions.

Authors have observed that the metalloporphyrins are mostly concentrated in the resin and asphaltenes (RA) fractions, associated with high molecular weight macrocycles, which are poorly separated when using size exclusion chromatography (SEC) or SEC-ICP-MS.2121 Merdrignac, I.; Espinat, D.; Oil Gas Sci. Technol. 2007, 62, 7.

22 Caumette, G.; Lienemann, C.-P.; Merdrignac, I.; Bouyssiere, B.; Lobinski, R.; J. Anal. At. Spectrom. 2010, 25, 1123.

23 Barbier, J.; Marques, J.; Caumette, G.; Merdrignac, I.; Bouyssiere, B.; Lobinski, R.; Lienemann, C.-P.; Fuel Process. Technol. 2014, 119, 185.
-2424 Vargas, V.; Castillo, J.; Torres, R. O.; Bouyssiere, B.; Lienemann, C.-P.; Fuel Process. Technol. 2017, 162, 37. Conversely, V and Ni porphyrins in the aromatics fraction are more labile, due to the presence of small polar molecules that act as solvents, increasing mobility.2525 Cui, Q.; Nakabayashi, K.; Ma, X.; Miyawaki, J.; Ideta, K.; Tennichi, Y.; Ueda, M.; Al-Mutairi, A.; Marafi, A. M. J.; Park, J.-I.; Yoon, S.-H.; Mochida, I.; Energy Fuels 2017, 31, 4748. The lability of those smaller and more polar compounds allow to employ a reverse phase separation, which results in improved resolutions, when compared to TLC or SEC. In this case, the separation is based on the difference between the less polar Ni2+-tetrapyrrole planar complexes and the more polar VO2+ squared pyramidal forms.2626 Duyck, C.; Miekeley, N.; Fonseca, T. C. O.; Szatmari, P.; Santos Neto, E. V.; J. Braz. Chem. Soc. 2008, 19, 978.

27 Dechaine, G. P.; Gray, M. R.; Energy Fuels 2010, 24, 2795.
-2828 Rodgers, R. P.; Hendrickson, C. L.; Emmett, M. R.; Marshall, A. G.; Greaney, M.; Qian, K.; Can. J. Chem. 2001, 79, 546. Besides, V and Ni metalloporphyrins can have a high number of substitutions, from aliphatic chains, such as in deoxophylloerythroetioporphyrin (DPEP) and etioporphyrin families, to aromatic or carboxyl groups. Then, adaptations of conventional SARA and TLC methods were necessary, in order to obtain cleaner fractions of neutral, polar and acidic metalloporphyrins, and improve their ionization process in Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR-MS).2828 Rodgers, R. P.; Hendrickson, C. L.; Emmett, M. R.; Marshall, A. G.; Greaney, M.; Qian, K.; Can. J. Chem. 2001, 79, 546.

29 Zhao, X.; Liu, Y.; Xu, C.; Yan, Y.; Zhang, Y.; Zhang, Q.; Zhao, S.; Chung, K.; Gray, M. R.; Shi, Q.; Energy Fuels 2013, 27, 2874.

30 Kim, E.; Cho, E.; Moon, S.; Park, J.-I.; Kim, S.; Energy Fuels 2016, 30, 2932.
-3131 Putman, J. C.; Rowland, S. M.; Corilo, Y. E.; McKenna, A. M.; Anal. Chem. 2014, 86, 10708.

Additionally, UV-Vis spectrophotometry can be used to control the fractionation of metalloporhyrins,3232 Sundararaman, P.; Anal. Chem. 1985, 57, 2204. because they exhibit specific absorption bands, one being the Soret band (around 400 nm), with a high absorptivity due to π-π* transition of the porphyrin nucleus.3333 Foster, N. S.; Day, J. W.; Filby, R. H.; Alford, A.; Rogers, D.; Org. Geochem. 2002, 33, 907. Automated versions of this technique are attractive, because they require a minimum of operation, reduce solvent consumption, improve precision and allow easily multiple steps.3434 Grizzle, P. L.; Sablotny, D. M.; Anal. Chem. 1986, 58, 2389.,3535 Boysen, R. B.; Schabron, J. F.; Energy Fuels 2013, 27, 4654.

Automation and maker/do-it-yourself (DIY) culture will play a key role in the so-called fourth industrial revolution (Industry 4.0). As such, automation in flow chemistry will surely contribute for such important strategic movement, by applying the open-source concept.3636 Pearce, J. M.; Science 2012, 337, 1303. We propose an automated method by using open-source hardware for petroleum fractionation based on the porphyrin absorption derivative at 400 nm. The signal derivatives are commonly employed in real time data processing for peak detection and quantification, particularly in the case of complex mixtures that result in peaks superposition.3737 Wittbrodt, B. T.; Squires, D. A.; Walbeck, J.; Campbell, E.; Campbell, W. H.; Pearce, J. M.; PLoS One 2015, 10, e0134989.,3838 Kelley, C.; Krolick, A.; Brunner, L.; Burklund, A.; Kahn, D.; Ball, W.; Weber-Shirk, M.; Sensors 2014, 14, 7142. However, the objective was to effectively separate absorbing fractions without further peak processing, therefore, a simple mathematical approach, based on the first and fourth derivatives and without smoothing algorithms, is employed. National Instruments LabVIEW™3939 National Instruments; LabVIEWversion 14; National Instruments, USA, 2015. is used to program an Arduino microcontroller for the control of valves and the separation of the metalloporphyrin fractions eluted with a reverse gradient, from methanol to toluene. Total V and Ni in these fractions are determined by ICP-MS after dilution in xylene and V linked to porphyrins is determined by high performance liquid chromatography (HPLC)-ICP-MS. The proposed automatized methodology has the advantage of improving the porphyrins class separation, when comparing to the automatized methods employing TLC, in order to quantify V and Ni linked to metalloporphyrins.

Experimental

Samples, reagents and standards

A crude oil (O-1) with American Petroleum Institute (API) degree of 21.9, density of 0.9187 g cm-3 and viscosity at 40 ºC of 1.94 cm2 s-1, from an onshore reservoir and used as an internal reference in our laboratory,4040 Ford, S.; J. Pharm. Biomed. Anal. 2003, 33, 563. was used for the development of the methodology, which was also applied to a Venezuelan certified crude oil, National Institute of Standards and Technology (NIST) 8505 and a crude oil A of marine influence. The oils (0.5 g) were pre-fractionated on a gravimetric glass column (50 × 2.5 cm2), using a gradient elution of heptane-toluene-dichloromethane-methanol/chloroform/water (65/35/4, v/v/v).4141 Wandekoken, F. G.; Duyck, C. B.; Fonseca, T. C. O.; Saint’Pierre, T. D.; Spectrochim. Acta, Part B 2016, 119, 1. Two fractions were obtained: (i) paraffins, aromatics and polar compounds (PAP); and (ii) RA, with mass recoveries higher than 85%. This first separation was necessary, in order to minimize the contamination of the RA fraction during the automated separation. All solvents used for fractionation were of HPLC grade (J.T. Baker). Xylene (analytical grade, Vetec, Sigma-Aldrich) was used for the dilution of the crude oils, fractions and the calibration solutions, prepared with S-21 multielemental organic standard (SCP Science) in a concentration range from 5 to 80 ng g-1, for Ni and V determination by ICP-MS.4242 de Souza, J. R.; Duyck, C. B.; Fonseca, T. C. O.; Saint’Pierre, T. D.; J. Anal. At. Spectrom. 2012, 27, 1280. A standard of vanadyl octaethylporphine (Aldrich) was used for detector calibration. The NIST standard reference materials 1084a, certified in Ni and V, and 8505, certified in V, were used for accuracy assessment of total concentrations in fractions. The HPLC-ICP-MS method for V determination in the labile metalloporphyrins fractions was applied to the samples.4141 Wandekoken, F. G.; Duyck, C. B.; Fonseca, T. C. O.; Saint’Pierre, T. D.; Spectrochim. Acta, Part B 2016, 119, 1.

System assembly

The overall procedure can be divided into three subsystems: (i) injection and solvents pumping; (ii) detection; and (iii) fractions separation, all connected to an Arduino Uno SMD R3 microcontroller with an ATmega16U2 USB-to-serial converter and a National Instruments LabVIEW™3939 National Instruments; LabVIEWversion 14; National Instruments, USA, 2015. interface program.

A glass demountable column (25 × 2.5 cm2; Supelco/Sigma-Aldrich) filled with about 30 g of silica gel (63-200 mm, Merck), activated at 150 ºC, is used for the automated fractionation of the PAP or RA fractions by liquid chromatography and the gel is discarded at the end of the process. The sample and the solvents are pumped on the column with a peristaltic pump (Minipuls 3, Gilson). Two valves (12 V, Rheodyne) with 4 inlets and 1 outlet are used in the process, one for injection and solvents selection, and the other for fraction separation. The column is connected to a 0.45 mL quartz flow cell (Hellma) with a 1 cm optical path length, where absorption occurs. All connections and tubing are made of Teflon or polyetheretherketone (PEEK).

The detection subsystem consists of a source and photovoltaic cell detector, both fixed and carefully aligned on the optical path of the flow cell in a homemade support, where it can be inserted or removed. The source is a 400 ± 5 nm light-emitting diode (LED) with a high brightness (12 mW), 3.7 V forward voltage at 20 mA (L3-0-U5TH15-1, LED Supply) and a 15º irradiation half angle with a low reverse current (10 µA at 5 V). The detector is a negative-positive-negative (NPN) silicon phototransistor (TIL 78, Texas Instruments) packed in a LED, operating at minimum of 1 mA light current and 5 V. The phototransistor voltage was amplified with rotatory potentiometers (100 and 2 kΩ, ± 3%, P47221B Amphenol), at a constant gain. This open-source hardware assembly has been successfully used by authors in photometric circuits.3737 Wittbrodt, B. T.; Squires, D. A.; Walbeck, J.; Campbell, E.; Campbell, W. H.; Pearce, J. M.; PLoS One 2015, 10, e0134989.,3838 Kelley, C.; Krolick, A.; Brunner, L.; Burklund, A.; Kahn, D.; Ball, W.; Weber-Shirk, M.; Sensors 2014, 14, 7142.,4343 González, P.; Pérez, N.; Knochen, M.; Quim. Nova 2016, 39, 305. Details of the circuit protoboard, circuit schematics and Arduino program, as well as the LabVIEW™3939 National Instruments; LabVIEWversion 14; National Instruments, USA, 2015. program interface (.txt) are available in the Supplementary Information section.

Automated separation of porphyrins fractions

The PAP fraction (about 0.4 g) is dissolved in about 1 mL of toluene and automatically injected at 1.4 mL min-1, corresponding to the position 1 of the valves 1 and 2 (the simplified scheme is shown in Figure 1). Then, the valve 1 is switched to position 2 and the peristaltic pump is adjusted to 1.2 mL min-1 for methanol elution, data acquisition begins and the non-absorbing fraction is separated. When the beginning of the first peak is detected, the valve 2 is automatically switched to position 2, for the separation of the first absorbing fraction (fraction a). At the end of fraction a elution, the valves 1 and 2 are switched to position 3, and the 50% (v/v) methanol/toluene mixture is pumped into the column, whereas the second absorbing fraction (fraction b) is separated. Again, at the end of fraction b elution, both valves are changed to position 4, for toluene pumping and separation of the third absorbing fraction (fraction c). At the end of the third peak, data acquisition is maintained for 5 min because a fourth peak can occur, depending on the original crude oil. Then, the elution and program are stopped. The program also has a time threshold at 90 min, when it is stopped if the third peak has not yet been detected. The overall sequence is listed in Table 1.

Figure 1
Scheme of the automated fractionation system and positions adopted by the two valves.

Table 1
Conditions of absorbance variation for peak definition and valves change, defined by the automation program. The repetition of dA threshold settled by varpeak (|dA|) is listed with the number of time intervals

Results and Discussion

System optimization and data acquisition

The calibration of the detection subsystem was performed with the vanadyl octaethylporphine standard and the response was linear in a concentration range from 0.3 to 2 µmol L-1 (r2 = 0.91). A molar absorptivity of 403 mmol L-1 cm-1 was obtained from the inclination of the curve, which was similar to another published result (364 mM-1 cm-1) obtained with a spectrophotometer.3333 Foster, N. S.; Day, J. W.; Filby, R. H.; Alford, A.; Rogers, D.; Org. Geochem. 2002, 33, 907. The overall detection system presents a good sensitivity, with a dark current of 0.01 V.

The transmittance is the ratio of the irradiance power when no absorbing compound is eluted (P0) to the one in the presence of absorbing compounds (P). At constant gain, the transmittance can be simplified as the ratio of the corresponding voltages: V0 / V. Then, the absorbance of the eluent is calculated as:

(1) A = log V / V 0

The program calculated the first derivative (dA / dt) from average values of the measured absorbance (A) during a settled time interval of dt = 10 s, which is a compromise between the peaks circumvolution and the baseline fluctuations. The derivative is calculated as:

(2) dA / dt n + 1 = A n + 1 A n / dt

where dA = An + 1 – An is the absorbance variation of the eluent within the time interval. Since only dA varies, its values are used for comparison with threshold values, in order to define the elution of absorbing compounds, and to send the information to the Arduino, which controls the system. The variations of A and dA / dt during the separation of about 0.3 g of PAP fraction obtained from crude oil O-1 can be seen in Figure 2. In this case, the cuts of the fractions (a, b, c) are adjacent, because these were used for V-porphyrin analysis by HPLC-ICP-MS. The first peak was determined between 50.8 and 93.8 min, the second one between 147 and 212.2 min, and the third one between 256.2 and 285.2 min.

Figure 2
Variation of absorbance (400 nm) and its derivative (dA / dt) as a function of time, with the fractions cuts. Details of the fourth derivative (d4A / dt4) are shown at the beginning (B) and end (E) of the first peak.

Conditions and control of the valves

The program uses a peak counter variable, named varpeak, which establishes conditions for dA absolute values (|dA|) for the definition of a peak. In parallel, a Boolean variable, named varBool, tests the conditions imposed by varpeak and the change of varBool from “false” to “true” happens when |dA| is significantly above or below a threshold value imposed by varpeak, indicating the beginning or the end of a peak. The program starts to test |dA| for varBool after 15 min of elution. It sends commands to the Arduino by the serial port when any of these thresholds are crossed, in order to control the valves. The threshold values and conditions are listed in Table 1 and were determined with precision after a first elution of the PAP fraction. Considering that the noise function (f(x)) is oscillatory and results from the sinusoidal current function, when the absorbance function (g(x)) depends on Beer’s Law, then according to Leibniz’s equation, the derivative of nth order is:

(3) f x × g x n = i = 0 n n i f n i x × g i x

Admitting that the k coefficient in the noise sinusoidal function f(x) = sin(kx) is below unity, then the nth derivative:

(4) k n sin kx + n π 2

tends to zero and lowers the corresponding terms in the Leibniz equation. Usually, the fourth derivative is satisfactory for noise attenuation in signal processing, and the beginning and end of peaks is visible when d4A / dt4 is different from the baseline such as in the Figure 2 details of the first peak elution. Then, the threshold values are the corresponding |dA| values found at the B (|dA| = 0.01) and E (|dA| = 0.001) points.

As it can be inferred from the numbers of peaks in the d4A / dt4 plot as a function of time, there is a plurality of eluting compounds, considering that porphyrins without V and Ni, as well as other aromatic compounds, can absorb at 400 nm and the fractions did not contain exclusively these metalloporphyrins. After complete separation, the program shuts down the system and generates a .txt archive that can be used for the A = f(t) and derivatives graphs, as exemplified in Figure 2.

Determination of Ni and V in fractions by ICP-MS

Total V and Ni concentrations in the original samples and in the porphyrins fractions were determined by ICP-MS after dilution in xylene. The solutions were introduced into the instrument at a 0.1 mL min-1 aspiration rate with a micronebulizer (PFA-100, Elemental Scientific) coupled to a cyclonic spray chamber refrigerated at 5 ºC (ISOMIST™, Glass Expansion) and the ICP-MS was operated at a forward power of 1.3 kW, Ar flows of 17 (plasma), 1.2 (auxiliary) and 0.4 L min-1 (nebulizer) with addition of 0.1 L min-1 O2 (nebulizer). The analysis of NIST 1084a, certified in V (95.9 ± 9.4 mg kg-1) and Ni (99.7 ± 1.6 mg kg-1), resulted in recoveries between 90 and 95%, respectively, and NIST 8505, certified in V (390 ± 10 mg kg-1), resulted in a recovery of 103%. Limits of detection (LOD) were calculated as 3 times the standard deviation of 10 measurements of signals in xylene divided by the slopes of the calibration curves, and the values were 0.4 (51V) and 0.1 µg kg-1 (60Ni). The LODs were of the same order or better than those reported in previous works.2222 Caumette, G.; Lienemann, C.-P.; Merdrignac, I.; Bouyssiere, B.; Lobinski, R.; J. Anal. At. Spectrom. 2010, 25, 1123.,4242 de Souza, J. R.; Duyck, C. B.; Fonseca, T. C. O.; Saint’Pierre, T. D.; J. Anal. At. Spectrom. 2012, 27, 1280.

The samples O-1 and NIST 8505 were previously separated into their PAP and RA fractions and the PAP fractions were separated with the automatized method for the quantification of V and Ni in fractions a, b and c (PAP) and RA. The reproducibility of the automatized separation methodology was evaluated through the standard deviations of the quadruplicate of O-1 PAP fractions and the results, all lower than 10%, are presented in Table 2. These results show the potentiality of the automatized method in terms of reproducible data, considering all the uncertainties associated with a chromatographic fractionation. In the case of NIST 8505, the reproducibility study was not carried out, due to the low amount of sample available in the lab and, therefore, the results are reported for one replicate.

Table 2
V and Ni concentrations in porphyrin fractions (a to c), and in the RA fractions of crude oil samples O-1 and NIST 8505, separated automatically

From the point of view of the oils compositions, the NIST 8505 certified crude oil showed higher metalloporphyrins content than crude oil O-1. This difference can be attributed to the higher biodegradation stage of the Venezuelan NIST 8505 oil, in comparison to the Brazilian O-1 oil. In addition, the elemental mass balances were calculated as the sum of the quantities of the element (concentration × mass) in fractions a, b, c (if analyzed) and RA, divided by its quantity in the oil. The values were 83% for both V and Ni in sample O-1 and 64 (V) and 67% (Ni) in NIST 8505. It can also be noted that, in the case of O-1, V in the labile fraction (fraction a) represents 15% of the element quantity in the oil, whereas this percentage lowers to 5% in the case of NIST 8505. As it can be seen, the lability of vanadyl porphyrins decreases with increasing biodegradation of the sample, with a progressive incorporation of these compounds into late-eluting, low mobility fractions, such as fractions b and c.

In addition, fraction c from O-1 PAP and fraction RA from O-1 were gathered and again submitted to automated fractionation, as one replicate (Table 2). Similarly to the PAP fractionation, three fractions (d to f) were obtained, with V and Ni concentrations higher than in the original sample. Consequently, these fractions are consistent with an asphaltenic character and should not have significant quantities of labile compounds.

Determination of V-porphyrins by HPLC-ICP-MS

The interest of using a reversed phase elution program is in the selectivity obtained for the separation of labile V-porphyrins, which are readily soluble in methanol. Qualitative determination of labile V-porphyrins was performed by analyzing fraction a obtained from duplicates of PAP of O-1 and NIST 8505, and from one PAP replicate of crude oil sample A, employing HPLC-ICP-MS. Fraction d from the resin fractionation of O-1 was also analyzed.

As it can be seen in Figure 3, the improvement in resolution regarding conventional SARA and SEC separation allows to distinguish between the V-porphyrinic samples profiles.2222 Caumette, G.; Lienemann, C.-P.; Merdrignac, I.; Bouyssiere, B.; Lobinski, R.; J. Anal. At. Spectrom. 2010, 25, 1123.,4444 Duyck, C. B.; Saint’Pierre, T. D.; Miekeley, N.; da Fonseca, T. C. O.; Szatmari, P.; Spectrochim. Acta, Part B 2011, 66, 362. The V chromatograms of the labile fraction obtained for crude oils from different origins and biodegradation levels by HPLC-ICP-MS with two RP18 columns in series, using the methodology described in Wandekoken et al.,4141 Wandekoken, F. G.; Duyck, C. B.; Fonseca, T. C. O.; Saint’Pierre, T. D.; Spectrochim. Acta, Part B 2016, 119, 1. show that the complexity of V-porphyrins rise with the biodegradation level (O-1 < NIST 8505). There is a profile difference between O-1 and A, as well, from terrestrial and marine reservoirs, possibly related to the difference in bacterial activity. In the case of the O-1 RA fractionation (Figure 3), early elution of the V chromatogram suggests acid groups. The chromatograms obtained for the other fractions show unresolved peaks, which are a general profile observed for macromolecules such as asphaltenes.

Figure 3
V linked to porphyrins in the labile fractions (fractions a and d) determined by HPLC-ICP-MS (HPLC: RP18 150 × 4.6 and 125 × 4 mm2 in series, 0.6 mL min-1 methanol elution). Black line: 51V ICP-MS signal; gray dashed line: 400 nm HPLC photodiode array signal.

Conclusions

The automated separation of metalloporphyrins fractions from crude oils was successfully performed on the basis of the absorbance derivatives at the Soret band of 400 nm, for the control of the separation. The threshold values of the derivative dA / dt, imposed by the peak variable, were correctly defined for peaks separation, as observed by the significant change of the fourth derivative, d4A / dt4, from the attenuated baseline. A simple approach, employing two variables and absorbance derivative, could be used for the testing of the conditions of peak elution and the quantification of V and Ni by ICP-MS, with acceptable elemental mass balances. The reversed solvent program, from methanol to toluene, was employed for mobility comparison of V-porphyrins determined in the first eluting fraction by HPLC-ICP-MS obtained from the investigated crude oils and rise with the biodegradation level of the oil and from marine to onshore reservoirs. The overall methodology was simple to operate, consumed less solvent than the manual separation and generated reproducible fractions for quantitative results improvement. This method is versatile and can be adjusted for other fractionations based on absorbing chromophores, and we are currently working on an adapted version of the Labview program on open source software.

Acknowledgments

The authors thank the Brazilian Petroleum Company (Petrobras) for the funding and in particular Flávio Albuquerque, Marco Antônio Teixeira and Rosana Cardoso, for the valuable contributions in porphyrins separation. Also, the authors thank the Brazilian National Research Council (CNPq) for the grants.

Supplementary Information

Supplementary data (scheme for circuit protoboard, circuit schematics, code for Arduino (.ino) and LabVIEW™) are available free of charge at http://jbcs.sbq.org.br.

https://minio.scielo.br/documentstore/1678-4790/D5BYcRk4JHphjDYCqj35ShS/a574dbd4977460d386e6fbf802512d33de597afe.pdf

References

  • 1
    Ocampo, R.; Callot, H. J.; Albrecht, P.; Kintzinger, J. P.; Tetrahedron Lett. 1984, 25, 2589.
  • 2
    Filby, R. H.; Van Berkel, G. J.; ACS Symp. Ser. 1987, 344, 2.
  • 3
    Ocampo, R.; Riva, A.; Trendel, J. M.; Riolo, J.; Callot, H. J.; Albrecht, P.; Energy Fuels 1993, 7, 191.
  • 4
    Doukkali, A.; Saoiabi, A.; Zrineh, A.; Hamad, M.; Ferhat, M.; Barbe, J. M.; Guilard, R.; Fuel 2002, 81, 467.
  • 5
    Galarraga, F.; Reategui, K.; Martïnez, A.; Martínez, M.; Llamas, J. F.; Márquez, G.; J. Pet. Sci. Eng. 2008, 61, 9.
  • 6
    Escobar, M.; Márquez, G.; Azuaje, V.; Da silva, A.; Tocco, R.; Fuel 2012, 97, 186.
  • 7
    Akinlua, A.; Sigedle, A.; Buthelezi, T.; Fadipe, O. A.; Mar. Pet. Geol. 2015, 59, 286.
  • 8
    López, L.; Lo Mónaco, S.; Org. Geochem. 2017, 104, 53.
  • 9
    Panariti, N.; Del Bianco, A.; Del Piero, G.; Marchionna, M.; Appl. Catal., A 2000, 204, 203.
  • 10
    Ali, M. F.; Abbas, S.; Fuel Process. Technol. 2006, 87, 573.
  • 11
    Duyck, C.; Miekeley, N.; Porto da Silveira, C. L.; Aucélio, R. Q.; Campos, R. C.; Grinberg, P.; Brandão, G. P.; Spectrochim. Acta, Part B 2007, 62, 939.
  • 12
    Sánchez, R.; Todolí, J. L.; Lienemann, C.-P.; Mermet, J.-M.; Spectrochim. Acta, Part B 2013, 88, 104.
  • 13
    Poirier, L.; Nelson, J.; Leong, D.; Berhane, L.; Hajdu, P.; Lopez-Linares, F.; Energy Fuels 2016, 30, 3783.
  • 14
    Rana, M. S.; Sámano, V.; Ancheyta, J.; Diaz, J. A. I.; Fuel 2007, 86, 1216.
  • 15
    Rana, M. S.; Ancheyta, J.; Sahoo, S. K.; Rayo, P.; Catal. Today 2014, 220-222, 97.
  • 16
    Jenifer, A. C.; Sharon, P.; Prakash, A.; Sande, P. C.; Energy Fuels 2015, 29, 7743.
  • 17
    Fan, T.; Buckley, J. S.; Energy Fuels 2002, 16, 1571.
  • 18
    Speight, J. G.; Oil Gas Sci. Technol. 2004, 59, 467.
  • 19
    Kharrat, A. M.; Zacharia, J.; Cherian, V. J.; Anyatonwu, A.; Energy Fuels 2007, 21, 3618.
  • 20
    ASTM D2007: Standard Test Method for Characteristic Groups in Rubber Extender and Processing Oils and Other Petroleum-Derived Oils by the Clay-Gel Absorption Chromatographic Method, Philadelphia, 2016.
  • 21
    Merdrignac, I.; Espinat, D.; Oil Gas Sci. Technol. 2007, 62, 7.
  • 22
    Caumette, G.; Lienemann, C.-P.; Merdrignac, I.; Bouyssiere, B.; Lobinski, R.; J. Anal. At. Spectrom. 2010, 25, 1123.
  • 23
    Barbier, J.; Marques, J.; Caumette, G.; Merdrignac, I.; Bouyssiere, B.; Lobinski, R.; Lienemann, C.-P.; Fuel Process. Technol. 2014, 119, 185.
  • 24
    Vargas, V.; Castillo, J.; Torres, R. O.; Bouyssiere, B.; Lienemann, C.-P.; Fuel Process. Technol. 2017, 162, 37.
  • 25
    Cui, Q.; Nakabayashi, K.; Ma, X.; Miyawaki, J.; Ideta, K.; Tennichi, Y.; Ueda, M.; Al-Mutairi, A.; Marafi, A. M. J.; Park, J.-I.; Yoon, S.-H.; Mochida, I.; Energy Fuels 2017, 31, 4748.
  • 26
    Duyck, C.; Miekeley, N.; Fonseca, T. C. O.; Szatmari, P.; Santos Neto, E. V.; J. Braz. Chem. Soc. 2008, 19, 978.
  • 27
    Dechaine, G. P.; Gray, M. R.; Energy Fuels 2010, 24, 2795.
  • 28
    Rodgers, R. P.; Hendrickson, C. L.; Emmett, M. R.; Marshall, A. G.; Greaney, M.; Qian, K.; Can. J. Chem. 2001, 79, 546.
  • 29
    Zhao, X.; Liu, Y.; Xu, C.; Yan, Y.; Zhang, Y.; Zhang, Q.; Zhao, S.; Chung, K.; Gray, M. R.; Shi, Q.; Energy Fuels 2013, 27, 2874.
  • 30
    Kim, E.; Cho, E.; Moon, S.; Park, J.-I.; Kim, S.; Energy Fuels 2016, 30, 2932.
  • 31
    Putman, J. C.; Rowland, S. M.; Corilo, Y. E.; McKenna, A. M.; Anal. Chem. 2014, 86, 10708.
  • 32
    Sundararaman, P.; Anal. Chem. 1985, 57, 2204.
  • 33
    Foster, N. S.; Day, J. W.; Filby, R. H.; Alford, A.; Rogers, D.; Org. Geochem. 2002, 33, 907.
  • 34
    Grizzle, P. L.; Sablotny, D. M.; Anal. Chem. 1986, 58, 2389.
  • 35
    Boysen, R. B.; Schabron, J. F.; Energy Fuels 2013, 27, 4654.
  • 36
    Pearce, J. M.; Science 2012, 337, 1303.
  • 37
    Wittbrodt, B. T.; Squires, D. A.; Walbeck, J.; Campbell, E.; Campbell, W. H.; Pearce, J. M.; PLoS One 2015, 10, e0134989.
  • 38
    Kelley, C.; Krolick, A.; Brunner, L.; Burklund, A.; Kahn, D.; Ball, W.; Weber-Shirk, M.; Sensors 2014, 14, 7142.
  • 39
    National Instruments; LabVIEWversion 14; National Instruments, USA, 2015.
  • 40
    Ford, S.; J. Pharm. Biomed. Anal. 2003, 33, 563.
  • 41
    Wandekoken, F. G.; Duyck, C. B.; Fonseca, T. C. O.; Saint’Pierre, T. D.; Spectrochim. Acta, Part B 2016, 119, 1.
  • 42
    de Souza, J. R.; Duyck, C. B.; Fonseca, T. C. O.; Saint’Pierre, T. D.; J. Anal. At. Spectrom. 2012, 27, 1280.
  • 43
    González, P.; Pérez, N.; Knochen, M.; Quim. Nova 2016, 39, 305.
  • 44
    Duyck, C. B.; Saint’Pierre, T. D.; Miekeley, N.; da Fonseca, T. C. O.; Szatmari, P.; Spectrochim. Acta, Part B 2011, 66, 362.

Publication Dates

  • Publication in this collection
    23 May 2019
  • Date of issue
    May 2019

History

  • Received
    12 Oct 2018
  • Accepted
    20 Feb 2019
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br