Quantification of Glyphosate and AMPA by HPLC-ICP-MS/MS and HPLC-DAD: A Comparative Study

Abstract

In this work, the direct determination methodology of glyphosate and aminomethylphosphonic acid (AMPA) by high performance liquid chromatography (HPLC) using inductively coupled plasma with triple quadrupole mass spectrometer (ICP-MS/MS) was compared with the conventional method using diode array detector (HPLC-DAD). Both methods were selective, but the chromatographic analysis using ICP-MS/MS as detector was four times faster than DAD. The methodology using ICP-MS/MS as detector was linear from 27 to 218 µg L-1, while by the conventional method it was 1 to 8 mg L-1. Detection limits of the glyphosate achieved with the HPLC-ICP-MS/MS and HPLC-DAD methods were 8.2 and 300 µg L-1, respectively. The direct determination using ICP-MS/MS as detector showed a lower number of stages in sample preparation because it did not need the derivatization step. Seven real samples were analyzed by HPLC-ICP-MS/MS and the results showed that these analytes were not detected.

Keywords:
aminomethylphosphonic acid; N-(phosphonomethyl)glycine; speciation; derivatization; detection system


Introduction

Glyphosate (N-(phosphonomethyl)glycine) is an herbicide of great agronomic expression due to wide applicability and efficiency in controlling weeds in many crops. In addition, it was considered toxicologically and environmentally safe for many years, which resulted in a considerable increase in agricultural applications, becoming the most widely adopted herbicide in the world.11 Botero-Coy, A. M.; Ibáñez, M.; Sancho, J. V.; Hernández, F.; J. Chromatogr. A 2013, 1292, 132.,22 Marek, L. J.; Koskinen, W. C.; Pest Manage. Sci. 2014, 70, 1158. However, in recent years, this increase has raised doubts and concerns about its increasing use, which in the long run may lead to undesirable impacts on the environment and human health. In 2015, the International Agency for Research on Cancer (IARC) classified glyphosate as a probable carcinogen to humans.33 International Agency for Research on Cancer (IARC); Evaluation of Carcinogenic Risks to Humans: Some Organophosphate Insecticides and Herbicides, vol.112; IARC: Lyon, France, 2015.

Therefore, many studies with glyphosate and its main metabolite, aminomethylphosphonic acid (AMPA), have been conducted to investigate and develop simple, sensitive and efficient methods for extraction and detection of these compounds in several matrices such as water,44 Hanke, I.; Singer, H.; Hollender, J.; Anal. Bioanal. Chem. 2008, 391, 2265.

5 Wang, S.; Liu, B.; Yuan, D.; Ma, J.; Talanta 2016, 161, 700.

6 Pinto, E.; Soares, A. G.; Ferreira, I. M. P. L. V. O.; Anal. Methods 2018, 10, 554.
-77 Demonte, L. D.; Michlig, N.; Gaggiotti, M.; Adam, C.; Beldoménico, G. H. R.; Repetti, M. R.; Sci. Total Environ. 2018, 645, 34. soil,88 Druart, C.; Delhomme, O.; Vaufleury, A.; Ntcho, E.; Millet, M.; Anal. Bioanal. Chem. 2001, 399, 1725.,99 Sun, L.; Kong, D.; Gu, W.; Guo, X.; Tao, W.; Shan, Z.; Wang, Y.; Wang, N.; J. Chromatogr. A 2017, 1502, 8. fruits1010 He, L.; Luo, X.; Jiang, X.; Qu, L.; J. Chromatogr. A 2010, 1217, 5013.

11 Chen, M.; Cao, Z.; Jiang, Y.; Zhu, Z.; J. Chromatogr. A 2013, 1272, 90.
-1212 Khrolenko, M. V.; Wieczorek, P. P.; J. Chromatogr. A 2005, 1093, 111. and vegetables.1313 Koskinen, W. C.; Marek, L. J.; Hall, K. E.; Pest Manage. Sci. 2016, 72, 423.

14 Schrübbers, L. C.; Masís-Mora, M.; Rojas, E. C.; Valverde, B. E.; Christensen, J. H.; Cedergreen, N.; Talanta 2016, 146, 609.
-1515 Oulkar, D. P.; Hingmire, S.; Goon, A.; Jadhav, M.; Ugare, B.; Thekkumpurath, A. S.; Banerjee, K.; J. AOAC Int. 2017, 100, 631. However, chemical properties of glyphosate and AMPA such as high polarity, low solubility in organic solvents, absence of chromophore groups and low volatility make their separation and detection processes laborious.1212 Khrolenko, M. V.; Wieczorek, P. P.; J. Chromatogr. A 2005, 1093, 111.,1313 Koskinen, W. C.; Marek, L. J.; Hall, K. E.; Pest Manage. Sci. 2016, 72, 423.

These characteristics make an additional derivatization step necessary to detect these compounds through spectroscopic techniques (conventional methods), which may promote an increase in their selectivity and sensitivity. Derivatization is still considered the main sample preparation disadvantage, since this procedure increases the number of sample preparation steps, requires expensive reagent use, and may cause analyte loss when there is not reaction efficiency. In addition, all these factors may contribute to an increase in standard deviation and process cost.1616 Arkan, T.; Molnár-Perl, I.; Microchem. J. 2015, 121, 99.

For toxicological and environmental monitoring, several countries and organizations have established significantly different maximum residue limits (MRLs) for glyphosate in drinking water such as 700 µg L-1 by the US EPA,1717 United States Environmental Protection Agency (US EPA); EPA 816-F-09-004, National Primary Drinking Water Regulation; Washington D.C., United States of America, 2009. and less than 0.1 µg L-1 by European legislation1818 International Association of Environmental Analytical Chemistry (IAEAC); Sample Handling of Pesticides in Water; Active: Barcelona, 1994. for this matrix. Therefore, several analytical methods have been developed to determine glyphosate at concentrations below these limits.

Gas chromatography (GC) and high performance liquid chromatography (HPLC) have been the most commonly used techniques for glyphosate and AMPA quantification,1919 Chen, Z.; He, W.; Beer, M.; Megharaj, M.; Naidu, R.; Talanta 2009, 78, 852. and the derivatization step has been necessary in both techniques. In this context, direct determination using inductively coupled plasma with triple quadrupole mass spectrometer (ICP-MS/MS) as a detector for HPLC may be more advantageous, because it does not require the derivatization step. The ICP-MS/MS has been used to determine the heteroatoms instead of compounds, i.e., to determine phosphorus in the glyphosate and AMPA molecules.2020 Nelson, J.; Hopfer, H.; Silva, F.; Wilbur, S.; Chen, J.; Ozawa, K. S.; Wylie, P. L.; J. Agric. Food Chem. 2015, 63, 4478. Moreover, the use of the MS/MS mode eliminates interferences from the matrix, thereby increasing the technique’s sensitivity.2121 Guo, Z. X.; Cai, Q.; Yang, Z.; Rapid Commun. Mass Spectrom. 2007, 21, 1606.

The aim of the present work was to compare two detection systems coupled to HPLC in quantifying glyphosate and AMPA, evaluating analytical performance through selectivity, linearity, limit of detection, analysis and sample preparation time. The evaluated systems were diode array detector HPLC-DAD (conventional method) through the molecular detection of derivatized glyphosate and AMPA, and HPLC-ICP-MS/MS by specific detection of 31P isotope present in these compounds.

Experimental

Reagents and solutions

Glyphosate (99.2% m/m) and aminomethylphosphonic acid (99.0% m/m) were obtained from Sigma-Aldrich (St. Louis, MO, USA). Glyphosate and aminomethylphosphonic acid solutions were prepared at 500 mg L-1 (stock solution) separately in ultrapure water (18 MΩ cm resistivity) obtained by the Milli-Q system (Merck Millipore, Darmstadt, Germany). A solution for each compound was prepared from stock solution at 25 and 5 mg L-1 in the same solvent (working solution). Next, dilutions were made from these solutions to obtain the analytical curves. All solutions were stored at 4 °C in the refrigerator in plastic bottles to avoid the complexation of metal ion standards in glass vials.22 Marek, L. J.; Koskinen, W. C.; Pest Manage. Sci. 2014, 70, 1158. Also, acetonitrile, methanol, 9-fluorenylmethyl chloroformate (FMOC-Cl), phosphoric acid, acetic acid solution (49% m/v), tetrabutylammonium hydroxide, and ammonium acetate (65%) were purchased from Sigma-Aldrich (St. Louis, MO, USA) and sodium borate from QM (São Paulo, Brazil).

Instrumentations

HPLC-DAD

The analysis was performed on an Agilent 1290 system (Agilent Technologies, Germany) consisting of a quaternary pump, an autosampler, a degasser and a diode array detector. ZORBAX SB-C8 (4.6 × 150 mm, 5 µm) was employed for separating glyphosate and AMPA derivatized with FMOC-Cl. The mobile phase composition was 0.2% (v/v) phosphoric acid in water (solvent A) and acetonitrile (solvent B) at a flow rate of 1 mL min−1. The column compartment temperature was set to 25 °C and the injection volume was 20 µL. The program was set for a linear gradient starting from 10% of solvent B to reach 45% of the solvent at 18 min. After that time, there was a decrease of solvent B from 45 to 10% in 2 min, reestablishing the initial conditions for the next run. Detection of FMOC-Cl derivatives of each herbicide was performed at λ = 250, 260 and 264 nm.88 Druart, C.; Delhomme, O.; Vaufleury, A.; Ntcho, E.; Millet, M.; Anal. Bioanal. Chem. 2001, 399, 1725.

Derivatization

Glyphosate and AMPA standards were previously derivatized to provide chromophore groups to the compounds. Aqueous standard solutions (1, 2, 4, 6 and 8 mg L-1) were prepared for calibration curves, where 100 µL of these solutions were transferred to a derivatization vial (0.3 mL capacity) and then derivatized by adding 100 µL of 1.25 mM borate buffer (pH 9) followed by 100 µL of 10 mM FMOC-Cl reagent, and the reaction was allowed to take place for 30 min at 25 °C temperature.2222 Anadón, A.; Martínez-Larrañaga, M. R.; Martínez, M. A.; Castellano, V. J.; Martínez, M.; Martin, M. T.; Nozal, M. J.; Bernal, J. L.; Toxicol. Lett. 2009,190, 91. The derivatized extracts were filtered through a 0.45 µm syringe filter and finally was directly injected into the HPLC-DAD.

HPLC-ICP-MS/MS

The high-performance liquid chromatography module (Model 1260 Infinity Series, Agilent Technologies, Waldbronn, Germany) was equipped with a quaternary pump, autosampler, degasser and thermostatic column compartment.

A ICP-MS/MS model Agilent 8800® (Agilent Technologies, Hachioji, Japan) was used, equipped with a 3-channel low pressure peristaltic pump (sample, internal standard and drain) and sample introduction system consisting of a concentric nebulizer, double pass spray chamber surrounded by a temperature control system, a quartz torch with 2.5 mm diameter injector tube, sampling and skimmer platinum cones with orifices of 1 and 0.4 mm in diameter, respectively. A solid state 27.12 MHz radio frequency generator with helical coils operating between 500 and 1600 Watts of power, fed with argon 99.998% was used for plasma sustaining. The 8800® has a quadrupole for ion selection that will enter in the third generation octopolar reaction system (Octopole Reaction System - ORS3®) pressurized with 99.999% helium or 99.999% O2 to remove the isobaric interferences through discrimination kinetic energy or reaction. After the ORS3®, there is another quadrupole mass spectrometer with hyperbolic bars with a dual stage electron multiplier detection system (pulsed and analog) at the end, which covers the mass spectrum between 2-260 Da.

The first quadrupole was adjusted to detect m/z 31 (31P), as this is the mass of the phosphorus isotope (31P 100%). In addition, ICP-MS/MS enables using a second analysis mass in MS/MS mode due the reaction with O2 in the collision and reaction cell (CRC), which is positioned between two quadrupole analyzers and was used for removing isobaric interferents.

The O2 was introduced as reaction gas in the CRC, and the second quadrupole was adjusted to monitor the reaction product m/z 47 (31P16O+), so that these ions were detected with minimal interference:2020 Nelson, J.; Hopfer, H.; Silva, F.; Wilbur, S.; Chen, J.; Ozawa, K. S.; Wylie, P. L.; J. Agric. Food Chem. 2015, 63, 4478.

(1) P + 31 O 2 16 P 16 31 O + + O 16 H reaction = 3 . 17 e V

The analytical conditions were: ZORBAX SB-C8 (4.6 × 150 mm, 5 µm), injection volume 50 µL, mobile phase composed of 50 mM acetic acid/ammonium acetate solution: 5 mM tetrabutylammonium: 1% methanol with pH = 4.7, isocratic elution with 1.0 mL min-1.2323 Sadi, B. B. M.; Vonderheide, A. P.; Caruso, J. A.; J. Chromatogr. A 2004, 1050, 95. The HPLC-ICP-MS/MS analysis conditions are summarized in Table 1.

Table 1
HPLC-ICP-MS/MS operation conditions

Analytical curves

Aqueous standard solutions (5, 10, 20, 30 and 40 µg L-1 of total phosphorus) were prepared in triplicate for HPLC-ICP-MS/MS analyzes. The analytical curves were obtained by placing the concentration values of glyphosate and AMPA on the abscissa axis and the areas were obtained on the ordinate axes using OriginPro 8 software,2424 Origin Pro 8, version v8.0725 (B725); OriginLab, Northampton, USA, 2007. which provided the determination (R2), angular (a) and linear (b) coefficients of the analytical curves. Extreme values were investigated and excluded by the Jackknife residual test, with a maximum exclusion of 22.2% of the original number of results, and the ordinary least squares method (OLSM) was again applied to the data for each exclusion.

The assumptions related to regression residuals and linear model adjustment were verified to validate the OLSM: (i) normality of residuals by the Ryan-Joiner test; (ii) homoscedasticity of residuals by the Brown-Forsythe test; (iii) independence of residuals by the Durbin-Watson test; and (iv) the F-test to verify the regression significance and linear model adjustment.

Limit of detection

The limit of detection (LOD) should be the lowest concentration detected in the added samples, but not necessarily quantified. The LOD in this study was obtained by signal-to-noise ratio (S/N) and determined as three times this relation.

Real sample preparation

The HPLC-ICP-MS/MS methodology was applied to determine glyphosate and AMPA in water samples from Turmalina city, Minas Gerais, Brazil. This city was chosen for presenting large eucalyptus plantations in its region. We collected water samples from artesian wells, dams, water springs and cisterns. Therefore, these samples had high potential to be contaminated by this herbicide as they were collected near eucalyptus plantation areas. All water samples were collected in 1000 mL polypropylene bottles. Water aliquots were stored at 4 °C immediately after sampling. The water samples were filtered through 0.45 µm pore sized membrane filters (Millipore, Bedford, MA, USA) before analysis. The manipulation and analysis were developed immediately after sampling.

Results and Discussion

We evaluated the analytical performance to compare the two detection systems in the quantification of glyphosate and AMPA through selectivity, analysis time, linearity, limit of detection and sample preparation time.

Selectivity and analysis time

The chromatogram obtained from determining derivatized glyphosate and AMPA by HPLC-DAD showed several chromatographic signals attributed to by-products of the derivatization reaction with FMOC-Cl. However, the method was considered selective and the compounds eluted in less than 20 min, as shown in Figure 1.

Figure 1
Chromatograms obtained from glyphosate (GLY) and AMPA derivatized standards at 6 mg L-1 by HPLC-DAD.

The main wavelengths used in determining glyphosate and AMPA were 260,2525 Chang, S. Y.; Wei, M.; J. Chin. Chem. Soc. 2005, 52, 785. 240,1212 Khrolenko, M. V.; Wieczorek, P. P.; J. Chromatogr. A 2005, 1093, 111. 254, 206, 210 and 360 nm.2626 Corbera, M.; Hidalgo, M.; Salvadó, V.; Microchim. Acta 2006, 153, 203.

27 Peruzzo, P. J.; Porta, A. A.; Ronco, A. E.; Environ. Pollut. 2008, 156, 61.

28 Amelin, V. G.; Bol’Shakov, D. S.; Tretiakov, A. V.; J. Anal. Chem. 2012, 67, 386.
-2929 Qian, K.; Wang, F.; Li, J.; Cao, Y.; Anal. Chim. Acta 2009, 635, 222. The choice of each wavelength was mainly associated to the derivatizing agent used. In this study, the wavelengths 250, 260 and 264 nm were simultaneously evaluated by comparing the slope of the analytical curves (Table S1, Supplementary Information (SI) section Supplementary Information Supplementary data are available free of charge at http://jbcs.sbq.org.br as a PDF file. ). The angular coefficients in each analytical curves were statistically different when evaluated by the t-test (α = 5%), indicating higher sensitivity of the detector to determine these target analytes in 264 nm, which obtained higher angular coefficient. Therefore, this wavelength was adopted for the comparison analyzes with HPLC-ICP-MS/MS.

The chromatogram obtained from the direct determination of glyphosate and AMPA by HPLC-ICP-MS/MS showed only two chromatographic signals assigned to the two target analytes. This result confirmed the selectivity of the method and the compounds eluted in less than 5 min, as shown in Figure 2.

Figure 2
Chromatogram obtained from the direct determination of glyphosate (GLY) and AMPA at the concentration of 50 µg L-1 of total phosphorus (corresponding to 273 µg L-1 of glyphosate and 179 µg L-1 of AMPA) by HPLC-ICP-MS/MS.

In this method, the first quadrupole was programmed to detect P m/z = 31 and the second was adjusted to monitor the reaction product 31P16O+ m/z = 47. The reaction of 31P with O2 in CRC was explored with the aim to avoid isobaric interferents (polyatomic ions)3030 Becker, J. S.; Boulyga, S. F.; Becker, C. P. J.; Buddrus, S.; Przybylski, M.; Anal. Bioanal. Chem. 2003, 375, 561. such as 14N1H16O+, 15N16O+ and 12C1H3 16O+ formed from argon gas plasma, which contribute with the signal intensity of m/z = 31.1919 Chen, Z.; He, W.; Beer, M.; Megharaj, M.; Naidu, R.; Talanta 2009, 78, 852.

Different gases such as helium, hydrogen, oxygen and ammonia could be used in CRC,3131 Bolea-Fernandez, E.; Balcaen, L.; Resano, M.; Vanhaecke, F.; Anal. Bioanal. Chem. 2015, 407, 919. but O2 was chosen in this study because it was verified in some studies that the reaction with oxygen is thermodynamically favorable and there is a significant increase of sensitivity in detecting phosphorous compounds in food matrices and water.2020 Nelson, J.; Hopfer, H.; Silva, F.; Wilbur, S.; Chen, J.; Ozawa, K. S.; Wylie, P. L.; J. Agric. Food Chem. 2015, 63, 4478.,2323 Sadi, B. B. M.; Vonderheide, A. P.; Caruso, J. A.; J. Chromatogr. A 2004, 1050, 95. A recent study also obtained better results using oxygen gas for detecting organophosphorus pollutants in water.3232 Lajin, B.; Goessler, W.; Talanta 2019, 196, 357.

The comparison of these analytical parameters showed that both methods were selective, but the method using ICP-MS/MS as the detector presented a four times shorter analysis time and a cleaner chromatogram.

Linearity

The second analyte parameter evaluated after demonstrating that both methods were selective was linearity. Thus, all outliers were determined and confirmed by the Jackknife residual test. The residual plots and outliers removed for each curve are shown in Figure S1 (SI section Supplementary Information Supplementary data are available free of charge at http://jbcs.sbq.org.br as a PDF file. ). The maximum number of outliers removed was three, corresponding to a limit of 22.2% of the 15 original data. This step allowed for deleting some points that could have an influence on the fitted regression equation. The results obtained for the tests of normality, homoscedasticity and independency of residuals indicated that the use of OLSM were appropriate. Table S2 (SI section Supplementary Information Supplementary data are available free of charge at http://jbcs.sbq.org.br as a PDF file. ) shows the tabulated critical values and the statistics from the normality, homoscedasticity and independency tests.

Also, high significance of the regression was observed for all curves (Table 2), indicating adjustment to the linear model at the concentration levels studied for each analyte in both detection systems.

Table 2
Statistics of variance regression analysis for glyphosate and AMPA curves in the investigated detection systems

Figure 3 presents the x-y plots and the respective OLSM statistics. These plots demonstrate linearity in the ranges from 1 to 8 mg L-1 for derivatized glyphosate and AMPA by the HPLC-DAD determination, and from 5 to 40 µg L-1 of total phosphorus corresponding from 27 to 218 µg L-1 of glyphosate and from 18 to 143 µg L-1 of AMPA by HPLC-ICP-MS/MS.

Figure 3
Calibration curves of glyphosate (GLY) and AMPA by HPLC-ICP-MS/MS and HPLC-DAD in the studied concentration ranges, with respective equations and determination coefficients (R2).

The comparison of these analytical parameters showed that both methods were linear for the two target analytes. However, the methodology using the ICP-MS/MS as detector was linear from 27 to 218 µg L-1, while it was 1 to 8 mg L-1 by the conventional method, showing superiority for the HPLC-ICP-MS/MS in relation to HPLC-DAD.

Limit of detection

In this study, the detection limit achieved for these analytes by HPLC-ICP-MS/MS was 36.5 times lower than HPLC-DAD, as shown in Table 3. These values found in our study were much lower than the MRL established by the US EPA in drinking water (700 µg L-1), but higher than that determined by the European Union (0.1 µg L-1).

Table 3
Comparison of the parameters of the chromatographic systems studied in the determination of glyphosate and AMPA in aqueous samples

A recent study using a similar methodology reached a total phosphorous detection limit of 0.02 to 0.69 µg L-1, corresponding to 0.11 to 3.76 µg L-1 of the glyphosate. These values were lower to those achieved in our study.3232 Lajin, B.; Goessler, W.; Talanta 2019, 196, 357. However, optional gas (1% CO2 in argon) was used in this study for a carbon enhancement effect of up to 3-fold in the limit of detection. This effect has generally been used for determination of arsenic, selenium and phosphorous.3333 Kovačevič, M.; Goessler, W.; Mikac, N.; Veber, M.; Anal. Bioanal. Chem. 2005, 383, 145.

Similarly, the detection limit obtained in determining glyphosate and AMPA by HPLC-DAD was higher than that reached for aqueous samples (9 µg L-1)2929 Qian, K.; Wang, F.; Li, J.; Cao, Y.; Anal. Chim. Acta 2009, 635, 222. and in apples (10 µg kg-1),3434 Qian, K.; Tang, T.; Shi, T.; Li, P.; Li, J.; Cao, Y.; J. Sep. Sci. 2009, 32, 2394. but other derivatizing agents were employed in these studies such as 4-chloro-3,5-dinitrobenzotrifluoride. However, the limits of detection achieved in our study were lower than the MRLs established by the US EPA.

Sample preparation time

The sample preparation includes all the time required for preparing the derivatization reagent solutions such as borate buffer at pH 9 and FMOC-Cl reagent, the time for the derivatization reaction to occur, and the preparation of the analytical curve with five concentrations in triplicate for both detection systems. In addition, the samples were not simultaneously derivatized to avoid different waiting times after derivatization. Thus, the mean sample preparation time for the HPLC-DAD analysis was 200 min due to the derivatization step (Table 3). On the other hand, the mean sample preparation time for determining these analytes by HPLC-ICP-MS/MS was 30 min, because this step was not necessary.

The derivatization step was indispensable in determining glyphosate and AMPA by HPLC-DAD (conventional method), because there is no chromophore group in glyphosate or AMPA molecules. Derivatization is not a simple process, even though it confers higher sensitivity and selectivity to the studied compounds.1616 Arkan, T.; Molnár-Perl, I.; Microchem. J. 2015, 121, 99. Several factors could affect the efficiency of the reaction, hindering the analysis of complex matrices because interferents present in these matrices could decrease the reaction performance and the compound quantification. In addition, the applicability of derivatization agents should be evaluated for each matrix, because the reaction developed for one matrix may not work in another with different characteristics, which contributes to the increased cost, sample preparation and analysis time.1313 Koskinen, W. C.; Marek, L. J.; Hall, K. E.; Pest Manage. Sci. 2016, 72, 423.

The results in this study have shown significant superiority of the HPLC-ICP-MS/MS method for determining glyphosate and AMPA in water, as this method presented a shorter analysis time and absence of the derivatization step. In addition, the specificity of this detector and the shorter number of stages in sample preparation are considerable advantages for using this technique, because it confers reliability to a method. The detection by 31P atom also confers more sensitivity and selectivity at the elemental analysis level, making it possible to obtain smaller LODs than in HPLC-DAD.

Although HPLC-ICP-MS/MS was superior to HPLC-DAD in almost all analytical parameters evaluated in this study, it is important to highlight that the acquisition and maintenance value of HPLC-ICP-MS/MS was much higher than HPLC-DAD, mainly in the consumption of argon gas. Therefore, the choice of the analytical strategy to quantify these compounds should also take this parameter into account.

Real samples

After comparing the two methodologies, HPLC-ICP-MS/MS was used to analyze seven different water samples. All selected samples were collected near eucalyptus plantations. However, the results showed that glyphosate and AMPA were not detected in any of the analyzed samples. Similar results were found in applying HPLC-ICP-MS to water samples from the Ohio River near downtown2323 Sadi, B. B. M.; Vonderheide, A. P.; Caruso, J. A.; J. Chromatogr. A 2004, 1050, 95. in a polluted water sample by ion chromatography (IC) using inductively coupled plasma mass spectrometry (IC-ICP-MS),2121 Guo, Z. X.; Cai, Q.; Yang, Z.; Rapid Commun. Mass Spectrom. 2007, 21, 1606. and in real river water samples using capillary electrophoresis (CE) coupled to ICP-MS.3535 Wuilloud, R. G.; Shah, M.; Kannamkumarath, S. S.; Altamirano, J. C.; Electrophoresis 2005, 26, 1598.

Conclusions

In this paper, the strategy of directly determining glyphosate and AMPA in aqueous samples by HPLC-ICP-MS/MS was compared to the conventional method by HPLC-DAD. The method using HPLC-ICP-MS/MS was faster than the conventional method because it does not require the derivatization step of the target analytes before instrumental analysis. In addition, the limit of instrumental detection was about 36.5 times lower than the limit reached by the diode array detector without previous derivatization with FMOC-Cl. The results for real samples did not detect those analytes in the studied water samples.

Supplementary Information

Supplementary data are available free of charge at http://jbcs.sbq.org.br as a PDF file.

Acknowledgments

The authors thank the Fundação de Amparo à Pesquisa do Estado de Minas Gerais (FAPEMIG) (APQ-01429-16) for the financial support. The authors are also grateful to the Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq) (420637/2016-8 and 307381/2015-3) for the financial support from research fellowships, the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) and the Universidade Federal de Minas Gerais (UFMG) for the infrastructure provided.

References

  • 1
    Botero-Coy, A. M.; Ibáñez, M.; Sancho, J. V.; Hernández, F.; J. Chromatogr. A 2013, 1292, 132.
  • 2
    Marek, L. J.; Koskinen, W. C.; Pest Manage. Sci 2014, 70, 1158.
  • 3
    International Agency for Research on Cancer (IARC); Evaluation of Carcinogenic Risks to Humans: Some Organophosphate Insecticides and Herbicides, vol.112; IARC: Lyon, France, 2015.
  • 4
    Hanke, I.; Singer, H.; Hollender, J.; Anal. Bioanal. Chem 2008, 391, 2265.
  • 5
    Wang, S.; Liu, B.; Yuan, D.; Ma, J.; Talanta 2016, 161, 700.
  • 6
    Pinto, E.; Soares, A. G.; Ferreira, I. M. P. L. V. O.; Anal. Methods 2018, 10, 554.
  • 7
    Demonte, L. D.; Michlig, N.; Gaggiotti, M.; Adam, C.; Beldoménico, G. H. R.; Repetti, M. R.; Sci. Total Environ 2018, 645, 34.
  • 8
    Druart, C.; Delhomme, O.; Vaufleury, A.; Ntcho, E.; Millet, M.; Anal. Bioanal. Chem 2001, 399, 1725.
  • 9
    Sun, L.; Kong, D.; Gu, W.; Guo, X.; Tao, W.; Shan, Z.; Wang, Y.; Wang, N.; J. Chromatogr. A 2017, 1502, 8.
  • 10
    He, L.; Luo, X.; Jiang, X.; Qu, L.; J. Chromatogr. A 2010, 1217, 5013.
  • 11
    Chen, M.; Cao, Z.; Jiang, Y.; Zhu, Z.; J. Chromatogr. A 2013, 1272, 90.
  • 12
    Khrolenko, M. V.; Wieczorek, P. P.; J. Chromatogr. A 2005, 1093, 111.
  • 13
    Koskinen, W. C.; Marek, L. J.; Hall, K. E.; Pest Manage. Sci 2016, 72, 423.
  • 14
    Schrübbers, L. C.; Masís-Mora, M.; Rojas, E. C.; Valverde, B. E.; Christensen, J. H.; Cedergreen, N.; Talanta 2016, 146, 609.
  • 15
    Oulkar, D. P.; Hingmire, S.; Goon, A.; Jadhav, M.; Ugare, B.; Thekkumpurath, A. S.; Banerjee, K.; J. AOAC Int 2017, 100, 631.
  • 16
    Arkan, T.; Molnár-Perl, I.; Microchem. J. 2015, 121, 99.
  • 17
    United States Environmental Protection Agency (US EPA); EPA 816-F-09-004, National Primary Drinking Water Regulation; Washington D.C., United States of America, 2009.
  • 18
    International Association of Environmental Analytical Chemistry (IAEAC); Sample Handling of Pesticides in Water; Active: Barcelona, 1994.
  • 19
    Chen, Z.; He, W.; Beer, M.; Megharaj, M.; Naidu, R.; Talanta 2009, 78, 852.
  • 20
    Nelson, J.; Hopfer, H.; Silva, F.; Wilbur, S.; Chen, J.; Ozawa, K. S.; Wylie, P. L.; J. Agric. Food Chem 2015, 63, 4478.
  • 21
    Guo, Z. X.; Cai, Q.; Yang, Z.; Rapid Commun. Mass Spectrom 2007, 21, 1606.
  • 22
    Anadón, A.; Martínez-Larrañaga, M. R.; Martínez, M. A.; Castellano, V. J.; Martínez, M.; Martin, M. T.; Nozal, M. J.; Bernal, J. L.; Toxicol. Lett 2009,190, 91.
  • 23
    Sadi, B. B. M.; Vonderheide, A. P.; Caruso, J. A.; J. Chromatogr. A 2004, 1050, 95.
  • 24
    Origin Pro 8, version v8.0725 (B725); OriginLab, Northampton, USA, 2007.
  • 25
    Chang, S. Y.; Wei, M.; J. Chin. Chem. Soc 2005, 52, 785.
  • 26
    Corbera, M.; Hidalgo, M.; Salvadó, V.; Microchim. Acta 2006, 153, 203.
  • 27
    Peruzzo, P. J.; Porta, A. A.; Ronco, A. E.; Environ. Pollut 2008, 156, 61.
  • 28
    Amelin, V. G.; Bol’Shakov, D. S.; Tretiakov, A. V.; J. Anal. Chem 2012, 67, 386.
  • 29
    Qian, K.; Wang, F.; Li, J.; Cao, Y.; Anal. Chim. Acta 2009, 635, 222.
  • 30
    Becker, J. S.; Boulyga, S. F.; Becker, C. P. J.; Buddrus, S.; Przybylski, M.; Anal. Bioanal. Chem 2003, 375, 561.
  • 31
    Bolea-Fernandez, E.; Balcaen, L.; Resano, M.; Vanhaecke, F.; Anal. Bioanal. Chem 2015, 407, 919.
  • 32
    Lajin, B.; Goessler, W.; Talanta 2019, 196, 357.
  • 33
    Kovačevič, M.; Goessler, W.; Mikac, N.; Veber, M.; Anal. Bioanal. Chem 2005, 383, 145.
  • 34
    Qian, K.; Tang, T.; Shi, T.; Li, P.; Li, J.; Cao, Y.; J. Sep. Sci 2009, 32, 2394.
  • 35
    Wuilloud, R. G.; Shah, M.; Kannamkumarath, S. S.; Altamirano, J. C.; Electrophoresis 2005, 26, 1598.

Publication Dates

  • Publication in this collection
    20 Jan 2020
  • Date of issue
    Feb 2020

History

  • Received
    15 Mar 2019
  • Accepted
    29 July 2019
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br