On the Use of a TiO2 Sol-Gel for Horseradish Peroxidase Enzyme Immobilization for Development of an Amperometric Biosensor

Abstract

The sol-gel technique has drawn considerable attention in the area of biomolecules immobilization by virtue of its intrinsic properties including chemical inertness, thermal stability, and good biocompatibility. The present work reports the development of a bio-sensing methodology which involves the immobilization of horseradish peroxidase enzyme (HRP) on pencil graphite electrode (PGE) using TiO2 sol-gel and its application for 5-aminosalicylic acid (5-ASA) determination in pharmaceutical formulations. TiO2 sol-gel film was deposited on the PGE surface by dip coating, and the HRP enzyme was immobilized on the film by adsorption and cross-linked binding using glutaraldehyde. The analytical performance of the methodology was investigated through amperometric measurements at -0.08 V in the presence of 5-ASA (electron mediator) and hydrogen peroxide (substrate). The biosensor was successfully applied for 5-ASA determination, where limits of detection and quantification of 3.3 and 10.0 µmol L -1, respectively, were obtained. The application of the methodology for 5-ASA determination in pharmaceutical formulations led to satisfactory recovery rates ranging from 80 to 98%. The biosensor developed in this work is simple, inexpensive and has the potential to be applied for the determination of other pharmaceutical compounds, which also act as electron mediators in the catalytic cycle of HRP.

Keywords:
HRP enzyme; TiO2 sol-gel; pencil graphite electrode; 5-aminosalicylic acid


Introduction

Horseradish peroxidase (HRP) (EC 1.11.1.7) belongs to the superfamily of heme-containing plant peroxidases, which promotes the bioelectrocatalytic reduction of H2O2 and the oxidation of organic and inorganic substrates; although there a great variety of peroxidases, HRP is undoubtedly the most commonly used in practical analytical applications. Besides that, HRP retains its activity over a broad range of pH and temperature, which facilitate its use.11 Freire, R. S.; Pessoa, C. A.; Mello, L. D.; Kubota, L. T.; J. Braz. Chem. Soc. 2003, 14, 230. In fact, this enzyme has been used for a wide range of purposes in environmental (e.g., wastewater remediation) and biological areas (e.g., electrochemical biosensors).22 Liu, S.; Huang, B.; Zheng, G.; Zhang, P.; Li, J.; Yang, B.; Chen, Y.; Liang, L.; Int. J. Biol. Macromol. 2020, 150, 814.

3 Abdulaal, W. H.; Almulaiky, Y. Q.; El-Shishtawy, R. M.; Catalysts 2020, 10, 181.

4 Zhang, C.; Li, H.; Li, C.; Li, Z.; Molecules 2020, 25, 168.
-55 Yao, L.; Zheng, L.; Cai, G.; Wang, S.; Wang, L.; Lin, J.; Sensors 2020, 20, 2738. Electrochemical biosensors offer a great potential for HRP application due to the possibility of combining its high stability, selectivity and sensitivity in low cost chemical analysis methods. Despite their seemingly unstable and delicate structures, enzymes can be resistant like any other chemical catalyst if used properly.66 Lopes, G. R.; Pinto, D. C. G. A.; Silva, A. M. S.; RSC Adv. 2014, 4, 37244. However, it is worth noting that one needs to obtain an efficient interface between the biomolecule and the electronic transducers in order to ensure an effective immobilization of the enzymes on the surfaces of electrodes. When they are bound to an inert support material, bioactive molecules may be rendered insoluble, retaining their catalytic activity, thereby extending their useful life.77 Kandimalla, V.; Tripathi, V. S.; Ju, H.; Crit. Rev. Anal. Chem. 2006, 36, 73.

A wide range of electrode materials have been used as platforms for HRP immobilization; among these materials, those that have been most commonly used include glassy carbon,88 Gao, F.; Yuan, R.; Chai, Y.; Chen, S.; Cao, S.; Tang, M.; J. Biochem. Biophys. Methods 2007, 70, 407.

9 Chen, X.; Chen, Z.; Zhu, J.; Xu, C.; Yan, W.; Yao, C.; Bioelectrochemistry 2011, 82, 87.
-1010 Li, M.; Wu, J.; Su, H.; Tu, Y.; Shang, Y.; He, Y.; Liu, H.; Sensors 2019, 19, 640. glassy carbon modified with carbon nanotube,1111 Ortiz, E.; Gallay, P.; Galicia, L.; Eguílaz, M.; Rivas, G.; Sens. Actuators, B 2019, 292, 254.

12 Zhang, Y.; Yuan, R.; Chai, Y.; Xiang, Y.; Hong, C.; Ran, X.; Biochem. Eng. J. 2010, 51, 102.
-1313 Yang, H.; Gong, C.; Miao, L.; Xu, F.; Int. J. Electrochem. Sci. 2017, 12, 4958. gold,1414 Kafi, A. K. M.; Naqshabandi, M.; Yusoff, M. M.; Crossley, M. J.; Enzyme Microb. Technol. 2018, 113, 67.

15 Mendes, R. K.; Carvalhal, R. F.; Kubota, L. T.; J. Electroanal. Chem. 2008, 612, 164.
-1616 Dai, M.; Huang, T.; Chao, L.; Xie, Q.; Tan, Y.; Chen, C.; Meng, W.; Talanta 2016, 149, 117. carbon paste,1717 Fallah, M.; Rahimnejad, M.; Asghary, M.; Mashkour, M.; Anal. Methods 2019, 12, 33.

18 Raghu, P.; Reddy, T. M.; Gopal, P.; Reddaiah, K.; Sreedhar, N. Y.; Enzyme Microb. Technol. 2014, 57, 8.
-1919 Ding, C.; Zhao, F.; Ren, R.; Lin, J. M.; Talanta 2009, 78, 1148. and platinum.2020 Spychalska, K.; Zając, D.; Cabaj, J.; RSC Adv. 2020, 10, 9079.

21 Baluta, S.; Zając, D.; Szyszka, A.; Malecha, K.; Cabaj, J.; Sensors 2020, 20, 423.
-2222 Al-Ahmed, A.; Ndangili, P. M.; Jahed, N.; Baker, P. G. L.; Iwuoha, E. I.; Sensors 2009, 9, 9965. Over the last decades, pencil graphite-based devices have also been employed as platforms for the construction of biosensors.2323 Wang, J.; Kawde, A. N.; Anal. Chim. Acta 2001, 431, 219. Since the quality of pencil graphite is highly controlled in the manufacturing process, there are rare differences in electrochemical behavior among different pencil graphite electrodes. In addition, pencil graphite electrodes (PGEs) are not as fragile as carbon paste electrodes (CPEs), and they are not as rigid as glassy carbon electrodes (GCEs). Thus, compared to other carbon materials, PGEs are superior in terms of mechanical properties and handling. Considering that pencil graphite is extremely cheap and widely abundant, PGEs can be used as disposable electrodes2424 Aoki, K.; Okamoto, T.; Kaneko, H.; Nozaki, K.; Negishi, A.; J. Electroanal. Chem . 1989, 263, 323.,2525 Nathani, A.; Vishnu, N.; Sharma, C. S.; J. Electrochem. Soc. 2020, 167, 037520. and have been successfully applied toward the immobilization of deoxyribonucleic acid (DNA),2626 Cam, E.; Tanik, N. A.; Cerkez, I.; Demirkan, E.; Aykut, Y.; J. Appl. Polym. Sci. 2018, 135, 45567.

27 Erdem, A.; Congur, G.; Electroanalysis 2018, 30, 67.
-2828 Ensafi, A. A.; Jamei, H. R.; Heydari-Bafrooei, E.; Rezaei, B.; Bioelectrochemistry 2016, 111, 15. antibody,2929 Vural, T.; Yaman, Y. T.; Ozturk, S.; Abaci, S.; Denkbas, E. B.; J. Colloid Interface Sci. 2018, 510, 318.

30 Güner, A.; Çevik, E.; Şenel, M.; Alpsoy, L.; Food Chem. 2017, 229, 358.
-3131 Engel, L.; Baumann, W.; Fresenius’ J. Anal. Chem. 1993, 346, 745. and enzymes3232 Sahyar, B. Y.; Kaplan, M.; Ozsoz, M.; Celik, E.; Otles, S.; Bioelectrochemistry 2019, 130, 107327.

33 Rewatkar, P.; Bandapati, M.; Goel, S.; Int. J. Hydrogen Energy 2019, 44, 31434.
-3434 Sharma, M.; Yadav, P.; Sharma, M.; Carbohydr. Res. 2019, 483, 107749. in the construction of electrochemical biosensors.

Several techniques involving HRP immobilization on electrode materials have been described in the literature;3535 Moyo, M.; Okonkwo, J. O.; Agyei, N. M.; Enzyme Microb. Technol . 2014, 56, 28.

36 Li, Z. L.; Cheng, L.; Zhang, L. W.; Liu, W.; Ma, W. Q.; Liu, L.; Process Saf. Environ. Prot. 2017, 107, 463.

37 Zeng, Y.; Zheng, A.; Wu, J.; Cai, Z.; Huang, A.; Liu, X.; Anal. Chim. Acta 2015, 899, 100.
-3838 Wang, F.; Gong, W.; Wang, L.; Chen, Z.; Microchim. Acta 2015, 182, 1949. these techniques include physical adsorption, covalent attachment, electrostatic interactions and entrapment on polymerized films. Specifically, the sol-gel technique has been widely employed for the immobilization of a variety of biomolecules because of its special properties, which include simplicity of preparation, chemical inertness, physical rigidity, high-thermal stability, biodegradation, and optical transparency.3939 Wang, J.; Anal. Chim. Acta 1999, 399, 21.

40 Gupta, R.; Chaudhury, N. K.; Biosens. Bioelectron. 2007, 22, 2387.
-4141 Jia, J.; Wang, B.; Wu, A.; Cheng, G.; Li, Z.; Dong, S.; Anal. Chem. 2002, 74, 2217. Essentially, the major advantage of a sol-gel system lies in its ability to retain a large content of water. This outstanding property helps ensure a long-term stability for the encapsulated bio-recognition agents or enzyme catalytic centers; apart from that, the process can be performed at room temperature. Other remarkable advantages of the sol-gel system include its biocompatibility and resistance to microbial attack.77 Kandimalla, V.; Tripathi, V. S.; Ju, H.; Crit. Rev. Anal. Chem. 2006, 36, 73.

TiO2 sol-gel based biosensors have been previously employed for the immobilization of different types of enzymes.4242 Hu, H.; Wang, B.; Li, Y.; Wang, P.; Yang, L.; Electroanalysis 2020, 32, 60027.

43 Cui, H. F.; Wu, W. W.; Li, M. M.; Song, X.; Lv, Y.; Zhang, T. T.; Biosens. Bioelectron . 2018, 99, 223.
-4444 Doong, R.; Shih, H.; Biosens. Bioelectron . 2006, 22, 185. In comparison to SiO2 or Al2O3 sol-gel based immobilization matrices, TiO2 presents good biocompatibility, since the titanium coordinates with amine and carboxyl groups present in the enzyme. This intrinsic property of TiO2 is regarded highly beneficial for both the immobilization of enzymes and the maintenance of bioactivity of the system, and thus contributes toward improving the biosensor performance. Titanium colloidal suspensions can be easily produced by hydrolysis of titanium alkoxides; as such, they are regarded viable alternative substrates for HRP immobilization.4545 Chen, X.; Cheng, G.; Dong, S.; Analyst 2001, 126, 1728.,4646 Chen, X.; Dong, S.; Biosens. Bioelectron . 2003, 18, 999.

5-Aminosalicylic acid (5-ASA, mesalazine, CAS 89-57-6) is widely used for the treatment of inflammatory bowel diseases (IBDs), especially for ulcerative colitis and Crohn’s disease.4747 Gotti, R.; Pomponio, R.; Bertucci, C.; Cavrini, V.; J. Chromatogr. A 2001, 916, 175.,4848 Bystrowska, B.; Nowak, J.; Brandys, J.; J. Pharm. Biomed. Anal. 2000, 22, 341. The mesalazine action is uncertain but probably occurs by inhibition of prostaglandin and leukotriene synthesis in the gastrointestinal mucosa.4949 Nobilis, M.; Vybíralová, Z.; Sládková, K.; Lísa, M.; Holcapek, M.; Kvetina, J.; J. Chromatogr. A 2006, 1119, 299. The levels of 5-ASA in pharmaceutical formulations have been traditionally quantified by using chromatographic techniques. Although routinely used, these techniques are relatively expansive and not free of drawbacks since they can often include labor-intensive sample preparation and extended analysis time. In this way, electroanalytical techniques have been also proposed for 5-ASA determination, which contributed for developing a cheaper and more rapid analysis.5050 Uliana, C. V.; Yamanaka, H.; Garbellini, G. S.; Salazar-Banda, G. R.; Quim. Nova 2010, 33, 964.

In a previous work reported in the literature,5151 Uliana, C. V.; Riccardi, C. S.; Yamanaka, H.; Ecletica Quim. 2008, 33, 57. the authors of the present work investigated the electrochemical behavior of HRP in solution in the presence of hydrogen peroxide and 5-aminosalycilic acid (5-ASA) (employed as electron mediator). In another study5050 Uliana, C. V.; Yamanaka, H.; Garbellini, G. S.; Salazar-Banda, G. R.; Quim. Nova 2010, 33, 964. that deserves mentioning, pencil graphite was used for the determination of 5-ASA in commercial medicine. Based on the findings of the aforementioned studies,5050 Uliana, C. V.; Yamanaka, H.; Garbellini, G. S.; Salazar-Banda, G. R.; Quim. Nova 2010, 33, 964.,5151 Uliana, C. V.; Riccardi, C. S.; Yamanaka, H.; Ecletica Quim. 2008, 33, 57. the present work sought to apply HRP as a model for enzyme immobilization on pencil graphite electrodes (PGE) using TiO2 sol-gel. Thus, the central goal of the study was to develop a novel, simple, effective and low-cost HRP immobilization methodology using TiO2 sol-gel modified pencil graphite electrodes (PGEs). The methodology was applied for the determination of 5-ASA in commercial pharmaceutical formulations.

Experimental

Reagents and solutions

Monobasic sodium phosphate, dibasic sodium phosphate, 25% glutaraldehyde (v/v), 30% hydrogen peroxide (m/m), horseradish peroxidase enzyme (EC 1.11.1.7, 1500 U mg -1) and titanium(IV) isopropoxide were acquired from Sigma-Aldrich (Saint Louis, USA). Monohydrate citric acid and potassium chloride were obtained from Mallinckrodt (Phillipsburg, USA). 5-Aminosalicylic acid (5-ASA) was obtained from Acros (Geel, Belgium).

HRP was dissolved in deionized water (1.0 mg mL -1). Solution of 1% glutaraldehyde was prepared in 0.1 mol L -1 phosphate buffer (pH 7.0). All solutions were stored in a freezer at -4 °C. Enzymatic activity was monitored using H2O2 as substrate and 5-ASA as electron mediator. 0.07 mol L -1 H2O2 stock solutions were prepared immediately before use. 5-ASA aqueous stock solutions (1.0 × 10 -3mol L -1) were prepared by dissolving the pure powder in phosphate-citrate buffer at pH 5.0, and heating the solution at a temperature of 60 ºC for 5 min in order to obtain a complete dissolution of the compounds. A 0.1 mol L -1 phosphate-citrate buffer solution at pH 5.0 was used as the supporting electrolyte. The solutions were prepared using high-purity water (resistivity (ρ) = 18.2 MΩ cm).

TiO2 sol-gel was prepared by adding titanium(IV) isopropoxide (20 mL) to a solution of nitric acid in deionized water in order to obtain a resultant solution of Ti:H+:H2O in a molar ratio of 1:0.5:200 (v/v/v). The precipitate obtained was continuously stirred until complete peptization in order to produce a stable suspension, which was dialyzed in deionized water up to pH 3.5, using a Micropore (3500 MW) dialysis membrane.5252 Zanoni, M. V. B.; Sene, J. J.; Anderson, M. A.; J. Photochem. Photobiol., A 2003, 157, 55.,5353 Xu, Q.; Anderson, M. A.; J. Mater. Res. 1991, 6, 1073.

Instrumentation

Type-HB (Faber-Castell) pencil leads with a total length of 60 mm and a diameter of 0.9 mm were obtained from a local bookstore. Initially, the bare pencil leads were cleaned with absorbent paper to remove any particles on the surface and the working area was established (delimiting the boundaries) by wrapping a sealing film (Parafilm M) around the pencil lead at a height of 3.0 mm from the top; the working area covered a geometric area of 0.091 cm2 of the PGE. The other end of the PGE was directly connected to the equipment. Each experiment was performed using a new PGE. Amperometric measurements were performed in 0.1 mol L -1 citrate-phosphate buffer (at pH 5.0) containing 5.0 × 10 -4 mol L -1 5-ASA and 3.5 × 10 -3 mol L -1 H2O2, using an EG&G-PAR model 263 (Maryland, USA) computer-controlled potentiostat. A 5.0 mL one-compartment cell which consisted of a three-electrode system was employed. The three-electrode system was composed of a PGE working electrode, a Ag|AgCl (KClsat) reference electrode, and a platinum wire auxiliary electrode. The morphological characterization of the modified PGE was conducted via high resolution scanning electron microscopy with a field emission gun (SEM-FEG, field emission scanning electron microscope, Jeol (Tokyo, Japan), Model JSM7500). The sol-gel films were deposited on the surface of the PGE by dip coating (Microquímica MQCTL2000-MP, Palhoça, Brazil); this involved 1 dip at a lifting speed of 100 mm min -1.

Electrode modification

Figure 1 summarizes the steps involving PGE modification for the immobilization of HRP and detection of 5-ASA. Initially, sol-gel films were deposited on a clean PGE surface by dip coating (Figures 1a-1b). All the electrodes were allowed to dry at room temperature for 24 h after deposition. Next, the PGE was sequentially immersed in glutaraldehyde (Figure 1c) and HRP solutions (Figure 1d). In between the modification steps, the electrodes were thoroughly washed in 0.1 mol L -1 phosphate buffer solution (at pH 7.0) in order to remove any weakly attached molecules on the electrode surface. Finally, amperometric measurements were performed in the presence of a fixed concentration of H2O2 (substrate) and different concentrations of 5-ASA (electron mediator) (Figure 1e). The appearance of the anodic peak current was associated with the reduction of 5-ASAox generated during the HRP enzymatic process.

Figure 1
Schematic illustration of PGE modification with HRP immobilization and 5-ASA detection: (a) bare PGE, (b) deposition of TiO2 sol-gel film, (c) incubation in the glutaraldehyde solution, (d) HRP immobilization and (e) amperometric measurements in the presence of 5-ASA and H2O2.

To obtain the experimental conditions with the best analytical signal, a two-level-three-factor (233 Abdulaal, W. H.; Almulaiky, Y. Q.; El-Shishtawy, R. M.; Catalysts 2020, 10, 181.) full factorial experiment was developed. A full factorial design can be represented by bk, where k is the number of factors (or variables), and b is the chosen number of levels. Factorial designs with two levels (2k) have been the most commonly used for the development of biosensors; experiments conducted using two-level factorial designs are found to be more efficient than those that investigate one factor at a time.5252 Zanoni, M. V. B.; Sene, J. J.; Anderson, M. A.; J. Photochem. Photobiol., A 2003, 157, 55.,5353 Xu, Q.; Anderson, M. A.; J. Mater. Res. 1991, 6, 1073. In this work, the following factors were evaluated: (i) incubation time of the electrodes in HRP solution (tHRP); (ii) glutaraldehyde concentration (Cglu); and (iii) incubation time of the electrodes in glutaraldehyde solution (tglu). The values chosen for the high and low levels of factors are presented in Table 1. The electrochemical conditions and the concentrations of H2O2 and 5-ASA in the electrochemical cell were kept constant under the different combined sets of conditions (3.5 × 10 -3 mol L -1 of H2O2 and 2.0 × 10 -4 mol L -1 of 5-ASA); this was done with the exclusive goal of determining the influence of the factors investigated in the system. Minitab 15 software5454 Minitab 15; Minitab Inc., State College, PA, USA, 2013. was used to generate the sequence and the combined set of experimental conditions for performing the assays, as well as for the analysis of the signal obtained (current intensity, -I / µA). The magnitude of the influence of each factor on the analytical signal was determined.5555 Brereton, R. G.; Chemometrics: Data Analysis for the Laboratory and Chemical Plant, 1st ed.; Wiley: Chichester, 2003.,5656 Uliana, C. V.; Tognolli, J. O.; Yamanaka, H.; Electroanalysis 2011, 23, 2607. The optimized conditions obtained from the full factorial design analysis were adopted for the subsequent experiments.

Table 1
High and low levels of the factors evaluated in a 23 full factorial design used for the optimization of HRP immobilization on the surface of PGE

5-ASA analysis in commercial pharmaceutical formulations

Calibration curves (n = 3) were obtained by spiking known concentrations of 5-ASA to the electrolytic cell containing 5.0 mL of 0.1 mol L -1 citrate-phosphate buffer solution (at pH 5.0) and 3.5 × 10 -3 mol L -1 H2O2. The limit of detection (LOD) and the limit of quantification (LOQ) were calculated based on the following equations: LOD = 3 × sd/B and LOQ = 10 × sd/B, respectively; where sd is the standard deviation of the average value of the currents obtained for ten blank solutions, and B is the slope of the analytical curve.

Recovery experiments were conducted by adding a known concentration of the target compound to the electrolytic cell containing 5.0 mL of 0.1 mol L -1 citrate-phosphate buffer (at pH 5.0) and 3.5 × 10 -3mol L -1H2O2. The current intensities obtained were directly interpolated in the linear regression equation from the 5-ASA analytical curve, and the concentrations were determined. All measurements were performed in triplicate. The recovery efficiencies (R, in percentage) were calculated based on the following equation: R (%) = (([5-ASA] obtained/[5-ASA] added) × 100), where the value of “[5-ASA] obtained” refers to the concentration obtained by interpolation in the corresponding curve.

The proposed methodology was applied for the determination of 5-ASA in commercial medicinal tablets and enemas. For pharmaceutical formulations in tablets, the Brazilian Pharmacopeia recommends weighing twenty tablets and powdering them.5757 Farmacopeia Brasileira, 6th ed.; Agência Nacional de Vigilância Sanitária (ANVISA): Brasília, 2019. Taking into account the 5-ASA content labeled in the commercial sample formulations, sample solutions of 1.0 × 10 -3 mol L -1 5-ASA were prepared in deionized water by heating at 60 ºC for 5 min. Aliquots of the sample solution corresponding to 5.0 × 10 -5 mol L -1 were transferred to the electrochemical cell containing 5.0 mL of the supporting electrolyte and 3.5 × 10 -3 mol L -1 of H2O2, and the amperograms were recorded.

Results and Discussion

Deposition of TiO 2 sol-gel film on the surface of PGE

The number of sol-gel film depositions on PGE surface was monitored by cyclic voltammetry in the range of +0.5 V to -1.15 V vs. Ag|AgCl (KClsat), at scan rate of 50 mV s -1 using 0.1 mol L -1 phosphate buffer solution of pH 7.0 (Figure 2A). The current intensities at the potential around -0.9 V in both direct and reverse scans increased as the number of sol-gel film depositions increased up to 4 depositions. No significant changes were observed after the fourth deposition onwards. This behavior is associated with the reduction of TiIV to TiIII and the electrons conduction through the oxide.5858 Fabregat-Santiago, F.; Mora-Seró, I.; Garcia-Belmonte, G.; Bisquert, J.; J. Phys. Chem. B 2003, 107, 758. The voltammetric responses observed were found to be typically characteristic of TiO2 thin films; this has been associated with sequential filling of electron states within the oxide.5959 Marken, F.; Bhambra, A. S.; Kim, D. H.; Mortimer, R. J.; Stott, S. J.; Electrochem. Commun. 2004, 6, 1153.

Figure 2
Cyclic voltammograms performed in (A) phosphate buffer at pH 7.0 and (B) 1.0 × 10-2 mol L-1 K4Fe(CN)6 solution using PGE after TiO2 sol-gel deposition of (a) 0; (b) 1; (c) 2; (d) 3 and (e) 4 sol-gel film layers.

To evaluate the influence of the deposition of sol-gel film on the response of K4Fe(CN)6 on PGEs, cyclic voltammograms were recorded in the range of -0.3 V to +0.7 V, at scan rate of 50 mV s -1, using 1.0 × 10 -2 mol L -1 K4Fe(CN)6 solution after a varied number of (sol-gel film) depositions (Figure 2B). As can be noted, a significant decrease was observed in the anodic and cathodic peak current intensities as the number of depositions increased. This behavior suggests that, as the gel thickness increases, the diffusion of the analyte from the solution to the electrode is undermined, and this leads to the decay of the electrochemical signal. Thus, only one sol-gel film deposition (the first deposition) was chosen for conducting all experiments.

The morphological characterization of the sol-gel film deposition on the PGE surface was performed by SEM-FEG. Figures 3a and 3c show the typical ripple and porous morphology of the bare PGE, subjected to 100- and 500-times magnification, respectively. Longitudinal grooves are present in the entire extension of the bare PGE surface as a result of the production process of pencil leads. Figures 3b and 3d are related to the images obtained for the PGE coated with TiO2 sol-gel film (PGE/sol-gel), subjected to 100- and 500-times magnification. Based on these images, one can observe the presence of a brittle film covering the entire PGE surface; this is attributed to the TiO2 sol-gel film. The regions that are completely covered on the PGE provide a stable platform for HRP immobilization; this occurs through the glutaraldehyde over the sol-gel base. By contrast, the cracked regions in the film provide a viable way for electron transfer between the mediator and the electrode.

Figure 3
SEM-FEG images of the bare PGE (a) and (c), subjected to 100- and 500-times magnification, respectively. Images (b) and (d) are related to the modified PGE (PGE/sol-gel), subjected to the same magnification rates.

HRP electrochemical response

For most of the materials employed in electrode modification, direct electron transfer between the materials and HRP enzyme is known to be a kinetically slow process. To solve this sluggish heterogeneous electron transfer problem, one particularly attractive approach reported in the literature involves the use of an electron mediator, a low-molecular weight molecule, which facilitates the electron transfer between an oxidoreductase enzyme active site and a working electrode material.6060 Vadgama, P.; Crump, P. W.; Analyst 1992, 117, 1657.

In the present work, the biosensor analytical signal is derived from the electrochemical behavior of the HRP enzyme in the presence of H2O2 (substrate) and 5-ASA (electron mediator).5151 Uliana, C. V.; Riccardi, C. S.; Yamanaka, H.; Ecletica Quim. 2008, 33, 57. To evaluate the potential for enzymatic reaction involving 5-ASA, cyclic voltammetry measurements were performed using the PGE containing one sol-gel film layer, in the range of +0.30 to -0.30 V, at scan rate of 50 mV s -1, in phosphate-citrate buffer (at pH 5.0) containing 3.5 × 10 -3 mol L -1 H2O2 and 5.0 × 10 -4mol L -1 5-ASA, in the absence and presence of 0.1 mg mL -1 HRP in solution (Figure 4).

Figure 4
Cyclic voltammograms for the application of PGE/sol-gel in phosphate-citrate buffer (pH 5.0) containing 3.5 × 10-3 mol L-1 H2O2 and 5.0 × 10-4 mol L-1 5-ASA in the absence (a) and presence (b) of 0.1 mg mL-1 HRP solution. Inset: current intensities obtained at different HRP concentrations immobilized on the PGE/sol-gel at –0.08 V.

There was a significant increase in current intensity at -0.08 V in the presence of HRP. This result suggests that 5-ASA is oxidized by HRP in the presence of H2O2; and the resulting product from the enzymatic reaction can be reduced on the electrode surface, with maximum cathodic current intensity at -0.08 V. Thus, this potential was chosen for the conduct of subsequent amperometric experiments for monitoring the enzymatic reaction.

Amperometric measurements were performed by varying the HRP concentration immobilized on the sol-gel film-modified PGE (inset Figure 4). Considering the high current intensity obtained for the concentration of 0.6 mg mL -1 HRP solution, this concentration was chosen for conducting subsequent experiments.

Optimization of HRP immobilization by full factorial design

The experimental conditions involved in the HRP immobilization were evaluated through a two-level-three-factor (233 Abdulaal, W. H.; Almulaiky, Y. Q.; El-Shishtawy, R. M.; Catalysts 2020, 10, 181.) full factorial design. The effects of low and high levels on the analytical signal (-I / µA) are presented as plot of main effects (Figure 5a). In this figure (Figure 5a), one can see that the more intense the slope of the line, the greater the magnitude of the main effect. Figure 5a also shows that tHRP and Cglu had a strong positive effect on the system, with the increase of the cathodic current intensity after changing from the low level to the high level, while tglu exhibited a weak positive effect.

Figure 5
Plot of main effects (a) for the factors evaluated in HRP immobilization; and cube plot (b) showing the relationship between the factors and the response (-I / μA), highlighted at the vertices of the cube.

The HRP incubation time is expected to exert a significant influence over the signal, since the current intensity depends primarily on the amount of enzyme immobilized on the PGE surface. The strong positive effect observed for tHRP indicated that the longer incubation time allowed the effective immobilization of a higher amount of HRP on the electrode, leading to a higher signal. The significant influence of Cglu was also directly associated with the amount of HRP immobilized on the PGE, since this reagent was responsible for promoting the crosslink binding between the enzyme and the electrode containing the TiO2 layer. Aldehydes are expected to form Schiff bases upon nucleophilic attack by the ε-amino groups of lysine residues in the protein. The linkage formed by the reaction of glutaraldehyde with an amino group has shown exceptional stability at extreme pHs and temperatures.6161 Migneault, I.; Dartiguenave, C.; Bertrand, M. J.; Waldron, K. C.; BioTechniques 2004, 37, 790. The higher signal observed along with the higher Cglu shows that, presumably, more HRP can be immobilized on the PGE surface when additional glutaraldehyde molecules are available for reaction. By contrast, the weak influence of tglu on the signal indicates that the small glutaraldehyde molecules can rapidly be bound to the hydrophilic sol-gel surface;6262 Wang, X.; Chen, Z.; Li, K.; Wei, X.; Chen, Z.; Ruso, J. M.; Tang, Z.; Liu, Z.; Colloids Surf., A 2019, 560, 298. and the adoption of longer periods of time in this step does not lead to a higher analytical signal. In view of that, the shorter time was chosen for tglu in order to decrease the total amount of time for HRP immobilization.

The cube plot, shown in Figure 5b, is another way of visualizing the impact of the factors on the response. In this plot, one can observe that the use of the optimum experimental conditions (Cglu = 1%; tglu = 1 h and tHRP= 2 h) led to a significant increase in the signal in comparison to other combined sets of conditions, considering the same amount of analyte and substrate in solution. This same figure (Figure 5b) also showed that while the following combined set of conditions (Cglu = 1%; tglu = 2 h and tHRP= 2 h) could also lead to the enhancement of the signal, one would require an additional one hour of treatment. Taking these observations into account, the repeatability of ten independently prepared electrodes was evaluated using the optimum conditions for HRP immobilization. The repeatability test yielded a satisfactory relative standard deviation (RSD) of 4.9%, indicating a good degree of reproducibility of the method developed for HRP immobilization.

Analytical performance of the modified PGE and its application for 5-ASA determination in commercial pharmaceutical formulations

An investigation was carried out regarding the analytical performance of the modified PGE used for the determination of 5-ASA in solution. Amperometric measurements were performed in 0.1 mol L -1 citrate-phosphate buffer solution (pH 5.0) containing 3.5 × 10 -3mol L -1 H2O2 and different concentrations of 5-ASA standard solution. The cathodic current intensity measured at -0.08 V presented a linear relationship with the 5-ASA (mediator) concentration. A 5-ASA analytical curve was constructed in the concentration range of 1.0 × 10 -5 to 10.0 × 10 -5 mol L -1 (Figure 6a). The linear regression equation of the analytical curve was I = - (0.53 ± 0.10) µA - (1.00 ± 0.02) µA L mol -1 [5-ASA], and the correlation coefficient (r) obtained was 0.9978. The LOD and LOQ values obtained were 3.3 and 10.0 µmol L -1, respectively. Analytical experiments were also carried out in order to determine the recovery rates of the method; in this analysis, 5.0 × 10 -5 and 8.0 × 10 -5 mol L -1 concentration solutions of 5-ASA were tested and the corresponding current intensities were directly interpolated in the linear regression equation of the analytical curve. Satisfactory recovery rates of 96.4 and 99.8% were obtained for the tested concentrations, respectively.

Figure 6
(a) Analytical curve for the determination of 5-ASA in the concentration range of 1.0 × 10-5 to 10.0 × 10-5 mol L-1 and (b). Standard addition curve for 5-ASA determination in a commercial pharmaceutical formulation sample (tablet A).

The methodology proposed in this work was applied for 5-ASA determination in commercial medicinal tablets and enemas. Aliquots of the sample solution were transferred to the electrochemical cell and the amperograms were recorded. The final concentration of the analyte was 5.0 × 10 -5 mol L -1 for all tested samples. Initially, the current intensities obtained from the sample analysis were directly interpolated in the analytical curve. However, the 5-ASA concentrations obtained by this procedure were in disagreement with the concentrations found in the drug labels; this may be attributed to a matrix effect on the 5-ASA amperometric responses. In this way, the standard addition method was used for the quantitative determination of 5-ASA. Figure 6b presents the standard addition curve for 5-ASA determination in tablet A, by successive additions of 5-ASA standard solutions to the sample solution. The results obtained for all the samples investigated are presented in Table 2.

Table 2
5-ASA determination in commercial pharmaceutical formulations using the HRP biosensor

A comparison of the analytical performance of the methodology presented in this work with the one proposed by Uliana et al.5050 Uliana, C. V.; Yamanaka, H.; Garbellini, G. S.; Salazar-Banda, G. R.; Quim. Nova 2010, 33, 964. shows that the LOD and LOQ values obtained in the former were higher than those obtained in the latter; this can be explained by the higher sensitivity of the square wave voltammetry technique. However, it is worth pointing out that the degree of sensitivity obtained for the sensor proposed in the present work is completely suitable for 5-ASA determination in commercial pharmaceutical formulations, given the usual content of the compound in these formulations. The recovery rates obtained in both methodologies were found to be similar. According to the Brazilian Pharmacopeia, the acceptable weight variation limit for any solid tablets and enemas containing over 0.250 g of the drug is ± 5.0%.5757 Farmacopeia Brasileira, 6th ed.; Agência Nacional de Vigilância Sanitária (ANVISA): Brasília, 2019. In this case, the samples analyzed using both methodologies are strictly in line with the Brazilian regulatory guidelines for these formulations, with the only exception being tablet B. The further comparison of the analytical performance of the presented methodology with another reported works using HRP enzyme in the development of biosensors to the determination of different oxidizable drugs (resorcinol, epinephrine, pyrogallol, pyrocatecthol, acetaminophen and clozapine) showed similar repeatability and sensitivity (limits of detection in the same order of magnitude found in this work), which confirms the potential of HRP-based biosensors to perform fast, accurate and low cost drugs determination.6363 Erdem, A.; Pabuccuoglu, A.; Merüc, B.; Kerman, K.; Ozsoz, M.; Turk. J. Med. Sci. 2000, 30, 349.,6464 Yu, D.; Blankert, B.; Kauffmann, J.; Biosens. Bioelectron . 2007, 22, 2707.

Conclusions

The present work reported the development of an effective methodology for HRP immobilization which involved the deposition of a TiO2 sol-gel film on the surface of PGE. The immobilization approach adopted was found to be effective in the sense that it helped preserve the HRP activity, while facilitating the electron transfer between the mediator molecule and the electrode. The optimization of the experimental conditions in the HRP immobilization procedure led to a significant enhancement of the sensor signal, thus improving the sensitivity of the method. The modified electrode was successfully applied for 5-ASA determination in commercial pharmaceutical formulations, which presented satisfactory recovery rates comparable to previously reported methodologies. The biosensor developed in this work is simple and cheap, and it is designed to be disposable. Based on the findings of this study, one can conclude that the sensor has potential to be applied for the determination of other pharmaceutical compounds which also act as electron mediators in the catalytic cycle of HRP.

Acknowledgments

The authors are grateful for the financial support granted by the National Council for Scientific and Technological Development (process CNPq 465571/2014-0), Coordination of Superior Level Staff Improvement (process CAPES 88887136426/2017/00) and the São Paulo Research Foundation (process FAPESP 2014/50945-4 and 2008/08990-1) in the development of this research.

References

  • 1
    Freire, R. S.; Pessoa, C. A.; Mello, L. D.; Kubota, L. T.; J. Braz. Chem. Soc 2003, 14, 230.
  • 2
    Liu, S.; Huang, B.; Zheng, G.; Zhang, P.; Li, J.; Yang, B.; Chen, Y.; Liang, L.; Int. J. Biol. Macromol 2020, 150, 814.
  • 3
    Abdulaal, W. H.; Almulaiky, Y. Q.; El-Shishtawy, R. M.; Catalysts 2020, 10, 181.
  • 4
    Zhang, C.; Li, H.; Li, C.; Li, Z.; Molecules 2020, 25, 168.
  • 5
    Yao, L.; Zheng, L.; Cai, G.; Wang, S.; Wang, L.; Lin, J.; Sensors 2020, 20, 2738.
  • 6
    Lopes, G. R.; Pinto, D. C. G. A.; Silva, A. M. S.; RSC Adv 2014, 4, 37244.
  • 7
    Kandimalla, V.; Tripathi, V. S.; Ju, H.; Crit. Rev. Anal. Chem 2006, 36, 73.
  • 8
    Gao, F.; Yuan, R.; Chai, Y.; Chen, S.; Cao, S.; Tang, M.; J. Biochem. Biophys. Methods 2007, 70, 407.
  • 9
    Chen, X.; Chen, Z.; Zhu, J.; Xu, C.; Yan, W.; Yao, C.; Bioelectrochemistry 2011, 82, 87.
  • 10
    Li, M.; Wu, J.; Su, H.; Tu, Y.; Shang, Y.; He, Y.; Liu, H.; Sensors 2019, 19, 640.
  • 11
    Ortiz, E.; Gallay, P.; Galicia, L.; Eguílaz, M.; Rivas, G.; Sens. Actuators, B 2019, 292, 254.
  • 12
    Zhang, Y.; Yuan, R.; Chai, Y.; Xiang, Y.; Hong, C.; Ran, X.; Biochem. Eng. J. 2010, 51, 102.
  • 13
    Yang, H.; Gong, C.; Miao, L.; Xu, F.; Int. J. Electrochem. Sci 2017, 12, 4958.
  • 14
    Kafi, A. K. M.; Naqshabandi, M.; Yusoff, M. M.; Crossley, M. J.; Enzyme Microb. Technol 2018, 113, 67.
  • 15
    Mendes, R. K.; Carvalhal, R. F.; Kubota, L. T.; J. Electroanal. Chem 2008, 612, 164.
  • 16
    Dai, M.; Huang, T.; Chao, L.; Xie, Q.; Tan, Y.; Chen, C.; Meng, W.; Talanta 2016, 149, 117.
  • 17
    Fallah, M.; Rahimnejad, M.; Asghary, M.; Mashkour, M.; Anal. Methods 2019, 12, 33.
  • 18
    Raghu, P.; Reddy, T. M.; Gopal, P.; Reddaiah, K.; Sreedhar, N. Y.; Enzyme Microb. Technol 2014, 57, 8.
  • 19
    Ding, C.; Zhao, F.; Ren, R.; Lin, J. M.; Talanta 2009, 78, 1148.
  • 20
    Spychalska, K.; Zając, D.; Cabaj, J.; RSC Adv 2020, 10, 9079.
  • 21
    Baluta, S.; Zając, D.; Szyszka, A.; Malecha, K.; Cabaj, J.; Sensors 2020, 20, 423.
  • 22
    Al-Ahmed, A.; Ndangili, P. M.; Jahed, N.; Baker, P. G. L.; Iwuoha, E. I.; Sensors 2009, 9, 9965.
  • 23
    Wang, J.; Kawde, A. N.; Anal. Chim. Acta 2001, 431, 219.
  • 24
    Aoki, K.; Okamoto, T.; Kaneko, H.; Nozaki, K.; Negishi, A.; J. Electroanal. Chem . 1989, 263, 323.
  • 25
    Nathani, A.; Vishnu, N.; Sharma, C. S.; J. Electrochem. Soc 2020, 167, 037520.
  • 26
    Cam, E.; Tanik, N. A.; Cerkez, I.; Demirkan, E.; Aykut, Y.; J. Appl. Polym. Sci 2018, 135, 45567.
  • 27
    Erdem, A.; Congur, G.; Electroanalysis 2018, 30, 67.
  • 28
    Ensafi, A. A.; Jamei, H. R.; Heydari-Bafrooei, E.; Rezaei, B.; Bioelectrochemistry 2016, 111, 15.
  • 29
    Vural, T.; Yaman, Y. T.; Ozturk, S.; Abaci, S.; Denkbas, E. B.; J. Colloid Interface Sci 2018, 510, 318.
  • 30
    Güner, A.; Çevik, E.; Şenel, M.; Alpsoy, L.; Food Chem 2017, 229, 358.
  • 31
    Engel, L.; Baumann, W.; Fresenius’ J. Anal. Chem 1993, 346, 745.
  • 32
    Sahyar, B. Y.; Kaplan, M.; Ozsoz, M.; Celik, E.; Otles, S.; Bioelectrochemistry 2019, 130, 107327.
  • 33
    Rewatkar, P.; Bandapati, M.; Goel, S.; Int. J. Hydrogen Energy 2019, 44, 31434.
  • 34
    Sharma, M.; Yadav, P.; Sharma, M.; Carbohydr. Res 2019, 483, 107749.
  • 35
    Moyo, M.; Okonkwo, J. O.; Agyei, N. M.; Enzyme Microb. Technol . 2014, 56, 28.
  • 36
    Li, Z. L.; Cheng, L.; Zhang, L. W.; Liu, W.; Ma, W. Q.; Liu, L.; Process Saf. Environ. Prot 2017, 107, 463.
  • 37
    Zeng, Y.; Zheng, A.; Wu, J.; Cai, Z.; Huang, A.; Liu, X.; Anal. Chim. Acta 2015, 899, 100.
  • 38
    Wang, F.; Gong, W.; Wang, L.; Chen, Z.; Microchim. Acta 2015, 182, 1949.
  • 39
    Wang, J.; Anal. Chim. Acta 1999, 399, 21.
  • 40
    Gupta, R.; Chaudhury, N. K.; Biosens. Bioelectron 2007, 22, 2387.
  • 41
    Jia, J.; Wang, B.; Wu, A.; Cheng, G.; Li, Z.; Dong, S.; Anal. Chem 2002, 74, 2217.
  • 42
    Hu, H.; Wang, B.; Li, Y.; Wang, P.; Yang, L.; Electroanalysis 2020, 32, 60027.
  • 43
    Cui, H. F.; Wu, W. W.; Li, M. M.; Song, X.; Lv, Y.; Zhang, T. T.; Biosens. Bioelectron . 2018, 99, 223.
  • 44
    Doong, R.; Shih, H.; Biosens. Bioelectron . 2006, 22, 185.
  • 45
    Chen, X.; Cheng, G.; Dong, S.; Analyst 2001, 126, 1728.
  • 46
    Chen, X.; Dong, S.; Biosens. Bioelectron . 2003, 18, 999.
  • 47
    Gotti, R.; Pomponio, R.; Bertucci, C.; Cavrini, V.; J. Chromatogr. A 2001, 916, 175.
  • 48
    Bystrowska, B.; Nowak, J.; Brandys, J.; J. Pharm. Biomed. Anal 2000, 22, 341.
  • 49
    Nobilis, M.; Vybíralová, Z.; Sládková, K.; Lísa, M.; Holcapek, M.; Kvetina, J.; J. Chromatogr. A 2006, 1119, 299.
  • 50
    Uliana, C. V.; Yamanaka, H.; Garbellini, G. S.; Salazar-Banda, G. R.; Quim. Nova 2010, 33, 964.
  • 51
    Uliana, C. V.; Riccardi, C. S.; Yamanaka, H.; Ecletica Quim 2008, 33, 57.
  • 52
    Zanoni, M. V. B.; Sene, J. J.; Anderson, M. A.; J. Photochem. Photobiol., A 2003, 157, 55.
  • 53
    Xu, Q.; Anderson, M. A.; J. Mater. Res 1991, 6, 1073.
  • 54
    Minitab 15; Minitab Inc., State College, PA, USA, 2013.
  • 55
    Brereton, R. G.; Chemometrics: Data Analysis for the Laboratory and Chemical Plant, 1st ed.; Wiley: Chichester, 2003.
  • 56
    Uliana, C. V.; Tognolli, J. O.; Yamanaka, H.; Electroanalysis 2011, 23, 2607.
  • 57
    Farmacopeia Brasileira, 6th ed.; Agência Nacional de Vigilância Sanitária (ANVISA): Brasília, 2019.
  • 58
    Fabregat-Santiago, F.; Mora-Seró, I.; Garcia-Belmonte, G.; Bisquert, J.; J. Phys. Chem. B 2003, 107, 758.
  • 59
    Marken, F.; Bhambra, A. S.; Kim, D. H.; Mortimer, R. J.; Stott, S. J.; Electrochem. Commun 2004, 6, 1153.
  • 60
    Vadgama, P.; Crump, P. W.; Analyst 1992, 117, 1657.
  • 61
    Migneault, I.; Dartiguenave, C.; Bertrand, M. J.; Waldron, K. C.; BioTechniques 2004, 37, 790.
  • 62
    Wang, X.; Chen, Z.; Li, K.; Wei, X.; Chen, Z.; Ruso, J. M.; Tang, Z.; Liu, Z.; Colloids Surf., A 2019, 560, 298.
  • 63
    Erdem, A.; Pabuccuoglu, A.; Merüc, B.; Kerman, K.; Ozsoz, M.; Turk. J. Med. Sci 2000, 30, 349.
  • 64
    Yu, D.; Blankert, B.; Kauffmann, J.; Biosens. Bioelectron . 2007, 22, 2707.

Publication Dates

  • Publication in this collection
    02 June 2021
  • Date of issue
    June 2021