Possible stressors in a neonatal intensive care unit at a university hospital

ABSTRACT

Objective:

To investigate possible stressors to which newborns are exposed in the neonatal intensive care unit.

Methods:

The levels of continuous noise were checked by a decibel meter positioned near the ear of the newborn, brightness was observed by a light meter positioned in the incubator in front of the newborn's eyes, and temperature was checked through the incubator display. The evaluations were performed in three periods of the day, with ten measurements taken at one-minute intervals during each shift for the subsequent statistical analysis.

Results:

All shifts showed noise above acceptable levels. Morning (p < 0.001), afternoon (p < 0.05) and night (p < 0.001) showed a significant increase compared to the control. The brightness significantly exceeded the normal range (p < 0.01) in the morning. We observed that only one of the incubators was within the normal temperature limits.

Conclusion:

The noise, brightness and temperature intensities were not in accordance with regulatory standards and thus might be possible stressors to newborns.

Keywords:
Intensive care, neonatal; Critical care; Stress, psychological

RESUMO

Objetivo:

Verificar possíveis fatores estressantes aos quais os recém-nascidos estão expostos na unidade de terapia intensiva neonatal.

Métodos:

Os níveis de ruídos contínuos foram verificados por meio do decibelímetro posicionado próximo ao ouvido do recém-nascido; a luminosidade foi verificada pelo luxímetro posicionado dentro das incubadoras diante dos olhos do recém-nascido; e a verificação da temperatura se deu por meio da visualização do display das incubadoras. As avaliações foram realizadas em três períodos do dia, tendo sido realizadas dez medições com intervalo de 1 minuto em cada turno para posteriores análises estatísticas.

Resultados:

Todos os turnos apresentaram ruídos acima dos níveis aceitáveis. Manhã (p < 0,001), tarde (p < 0,05) e noite (p < 0,001) apresentaram aumento significativo comparado ao controle. A luminosidade excedeu os padrões de normalidade significativamente (p < 0,01) no período da manhã. Quanto à temperatura, observamos que apenas uma das incubadoras encontrava-se dentro dos padrões de normalidade.

Conclusão:

A intensidade dos ruídos, da luminosidade e da temperatura não estavam de acordo com as normas regulatórias, podendo ser assim um possível fator estressante para o recém-nascido.

Descritores:
Unidades de terapia intensiva neonatal; Cuidados críticos; Estresse psicológico

INTRODUCTION

Currently, psychological stress is considered to be one of the main factors responsible for changes in the health status of the global population. These changes manifest through physiological, cognitive and behavioral changes and may lead to the emergence of diseases and even death.(11 Pafaro RC, De Martino MM. Estudo do estresse do enfermeiro com dupla jornada de trabalho em um hospital de oncologia pediátrica de Campinas. Rev Esc Enferm USP. 2004;38(2):152-60.,22 Soares AJ, Alves MG. Cortisol como variável em psicologia da saúde. Psicol Saúde Doenças. 2006;7(2):165-77.) Understanding stress as a psycho-physiological process of the body is important because it allows the diagnosis of responses triggered by the manner in which stimuli are processed.(33 Jorge SR, Santos PB, Stefanello JM. O cortisol salivar como resposta fisiológica ao estresse competitivo: uma revisão sistemática. Rev Educ Fís/UEM. 2010;21(4):677-86.,44 Meyerhof PG. O neonato pré-termo no berçário de cuidados especiais: proposta de intervenção, respeitando sua individualidade, suas fragilidades e suas forças. Rev Bras Crescimento Desenvolv Hum. 1995;5(1/2):17-22.)

The neonatal intensive care unit (NICU) is a stressful environment due to the presence of multiple factors, such as bright light, noise, handlings performed by professionals and little social interaction.(55 Gaspardo CM. Dor em neonatos pré-termo em unidade de terapia intensiva neonatal: avaliação e intervenção com sacarose [Dissertação]. Ribeirão Preto: Faculdade de Medicina de Ribeirão Preto da Universidade de São Paulo; 2006.,66 Nascimento TO, Maranhão DG. Prevenção do estresse neonatal: desafio para a equipe de enfermagem. Rev Enferm UNISA. 2010;11(2):134-7.) Thus, despite the high technology and training of the health teams, these factors contribute to changes in the sleep cycle, stress appearance, discomfort and pain.(77 Weich TM, Ourique AC, Tochetto TM, Franceschi CM. Eficácia de um programa para redução de ruído em unidade de terapia intensiva neonatal. Rev Bras Ter Intensiva. 2011;23(3):327-34.,88 Machado AC, Machado LG, Fonseca TC, Coelho RR. Níveis de ruído em uma unidade de terapia intensiva: avaliação sob a ótica da ergonomia. In: XXI Simpósio de Engenharia de Produção. Bauru, São Paulo, Brasil. 10 a 12 de novembro de 2014.)

A hospital without noise is impossible. However, a hospital with acceptable levels of continuous noise (CN) is a necessary prerogative for the clinical progression of the patient. Therefore, the Associação Brasileira de Normas Técnicas (10152/1987) established in-hospital CN values of 35 and 45 decibels (dB) as desirable and acceptable levels, respectively. Furthermore, the American Academy of Pediatrics set specific values by the Consensus Committee on Recommended Standards for Advanced Neonatal Care, which recommends that CN levels should not exceed 45 dB.(99 Rodarte MD. Exposição e reatividade do prematuro ao ruído intenso durante o cuidado em incubadora [tese]. Ribeirão Preto: Escola de Enfermagem de Ribeirão Preto da Universidade de São Paulo; 2007.

10 Brasil. Ministério da Saúde. Secretaria de Atenção à Saúde. Departamento de Ações Programáticas e Estratégicas. Atenção à saúde do recém-nascido: guia para os profissionais de saúde. Cuidados com o recém-nascido pré-termo. Brasília: Ministério da Saúde; 2011. [Série A. Normas e Manuais Técnicos, 4]. Disponível em: http://www.redeblh.fiocruz.br/media/arn_v4.pdf
http://www.redeblh.fiocruz.br/media/arn_...
-1111 Martins CF, Fialho FA, Dias IV, Amaral JA, Freitas SC. Unidade de terapia intensiva neonatal: o papel da enfermagem na construção de um ambiente terapêutico. R Enferm Cent Oeste Min. 2011;1(2):268-76.)

In addition to noise, bright light should also be considered a stressor because it may affect the states of sleep and wakefulness in premature infants and interfere with the entire hormonal circadian rhythm.(1212 Saraiva CA. Fatores físicos-ambientais e organizacionais em uma unidade de terapia intensiva neonatal: implicações para a saúde do recém-nascido [dissertação]. Porto Alegre: Escola de Engenharia da Universidade Federal do Rio Grande do Sul; 2004.) In the NICU, the use of fluorescent lamps from 40 to 60W/m2 is recommended to allow an accurate assessment of the newborn's skin color and provide high visibility during special procedures while at the same time not interfering with the comfort of the newborns.(77 Weich TM, Ourique AC, Tochetto TM, Franceschi CM. Eficácia de um programa para redução de ruído em unidade de terapia intensiva neonatal. Rev Bras Ter Intensiva. 2011;23(3):327-34.) Technical standards related to illumination (NBR 5413) state that the values of artificial illumination in nurseries should not exceed 100 lux.

Body temperature is the result of the balance between the heat production and elimination mechanisms. In newborns (especially in preterm infants), there may be an imbalance between these mechanisms. For full-term babies, the thermoneutral zone in the first hours of life is between 32 and 34°C, but the thermoneutral range varies depending on birth weight and postnatal gestational age and reaches 35 to 37°C for premature newborns with low birth weights in the first days of life.(77 Weich TM, Ourique AC, Tochetto TM, Franceschi CM. Eficácia de um programa para redução de ruído em unidade de terapia intensiva neonatal. Rev Bras Ter Intensiva. 2011;23(3):327-34.) To increase the survival rate of premature newborns, they are placed in closed chambers with the temperature maintained in a specific range, which reduces oxygen consumption and keeps them warm.(1010 Brasil. Ministério da Saúde. Secretaria de Atenção à Saúde. Departamento de Ações Programáticas e Estratégicas. Atenção à saúde do recém-nascido: guia para os profissionais de saúde. Cuidados com o recém-nascido pré-termo. Brasília: Ministério da Saúde; 2011. [Série A. Normas e Manuais Técnicos, 4]. Disponível em: http://www.redeblh.fiocruz.br/media/arn_v4.pdf
http://www.redeblh.fiocruz.br/media/arn_...
) The same procedure is also suitable for full-term babies who are sick. The incubator heats the atmosphere by convection, providing a warm, stable and uniform environment ranging between 36 and 36.5°C. However, the heated crib emits infrared light that is easily absorbed by the skin, turning it into heat and varying the temperature between 36.5 and 37°C.(1313 Fasolo MI, Moreira RN, Abatti PJ. Avaliação de nível de ruído em incubadora. J Pediatr (Rio J). 1994;70(3):157-62.)

Reduced exposure to these factors is critical for the good prognosis of the newborn, but often these aspects are neglected by health teams.(55 Gaspardo CM. Dor em neonatos pré-termo em unidade de terapia intensiva neonatal: avaliação e intervenção com sacarose [Dissertação]. Ribeirão Preto: Faculdade de Medicina de Ribeirão Preto da Universidade de São Paulo; 2006.) Thus, the present study evaluated possible stress factors (noise, light and temperature) in the neonatal intensive care unit at a university hospital.

METHODS

This study was conducted in Hospital Universitário Sul-Fluminense (HUSF) and approved by the Research Ethics Committee (No. 1161709) at the Universidade Severino Sombra. Parents were requested to sign a free and informed consent form.

The evaluations were performed in three periods [morning (7 am to 9 am), afternoon (13 pm to 15 pm) and night (18 pm to 20 pm)], with 10 measurements taken at one-minute intervals during each 10-minute set. These 10 measurements were only used for the evaluation of lighting and noise. The medical team on shift that day was not informed of the nature of the study in order to avoid any changes in their routine work that could mask the results. The inclusion criterion was to establish that the measurements occurred near the incubators in use at the time of the study.

The measurements were performed with an Instrutherm DEC-460 decibel meter. The equipment was calibrated according to the manufacturer's recommendations with the aid of a standard external acoustic calibrator with a sinusoidal signal of 94 dB at 1 kHz. Then, the device was placed near the ear of the newborn. The decibel meter was set in the compensation circuit "A" and slow response circuit (SLOW) to check the average sound level as indicated for the CN measurements in accordance with regulatory standard NR 15. The reference value recommended by the American Academy of Pediatrics (Consensus Committee on Recommended Standards for Advanced Neonatal Care) and the Associação Brasileira de Normas Técnicas NBR 10152/1987 of 45 dB was used as a control. The CN levels were also evaluated during airway aspiration, which is a common procedure in intensive care units.

The measurements were performed with a factory calibrated Light Meter HS1010A. The device was placed in each incubator in front of the eyes of newborns. A control value of 100 lux was used as recommended by the NBR 5413.

The temperature was verified by the display on the incubator because clinicians commonly use this reading to check the temperature during their routine. Normal reference values were taken from those recommended by the Scopes and Ahmed scale, with suggestions for the ideal temperature based on the term range of neutrality as a function of birth weight and postnatal age.(77 Weich TM, Ourique AC, Tochetto TM, Franceschi CM. Eficácia de um programa para redução de ruído em unidade de terapia intensiva neonatal. Rev Bras Ter Intensiva. 2011;23(3):327-34.)

Statistical analysis

The distributions of data obtained from the sound level meter and light meter were verified by the Kolmogorov-Smirnov test. Then, one-way analysis of variance (ANOVA) with the Bonferroni post hoc test was used to compare the means with GraphPad Prism 5®. The level of significance was set at p < 0.05, and the results are expressed as the mean ± standard error of the mean. For temperature, we only tested whether the identified values fit the Scopes and Ahmed scale.

RESULTS

All shifts had average CN levels above the acceptable limit. The levels during the morning (23.48%; p <0.001), afternoon (15.7%; p <0.05) and night (21%; p <0.001) shifts were higher than the recommendations. Additionally, we observed an increase of 49.7% (p <0.001) in CN while performing a routine procedure (aspiration) compared with the control (Figure 1).

Figure 1
Quantification of continuous noise in the neonatal intensive care unit.

Data presented as the mean ± standard error of the mean. * Significant difference with p < 0.05; *** Difference p < 0.001; both compared to the control. We used 45 dB as the reference value, which is recommended by the American Academy of Pediatrics and Associação Brasileira de Normas Técnicas NRB 10152/1987.


Regarding brightness, we found that the illuminance during the morning shift was 84.8% above the recommended limit (p < 0.01). Significant differences were not observed in the other shifts (Figure 2).

Figure 2
Quantification of illuminance in the neonatal intensive care unit.

Data presented as the mean ± standard error of the mean. ** Significant difference with p < 0.01 compared to the control. For the control, we used 100 lux, the amount recommended by the Associação Brasileira de Normas Técnicas NBR 5413.


Only one of the incubators in use at the time of the study was within the normal temperature limits in all three periods (Table 1).

Table 1
Temperature of the neonatal intensive care unit incubators

DISCUSSION

The NICU is an aggressive environment that is impersonal and difficult to adapt. This environment is full of intense and constant light, noise, temperature changes and disruption of the sleep cycle due to repeated assessments and procedures that occur more or less frequently depending on the severity of the newborn.(1111 Martins CF, Fialho FA, Dias IV, Amaral JA, Freitas SC. Unidade de terapia intensiva neonatal: o papel da enfermagem na construção de um ambiente terapêutico. R Enferm Cent Oeste Min. 2011;1(2):268-76.)

Fasolo et al. conducted a study that found that 33.3% of the assessed incubators were 60 dB above the recommended tolerable levels. Moreover, they observed that the average non-CN generated by manipulation of the incubator access ports ranged from 96.2 dB (opening) to 107 dB (closing). Fasolo et al.(1313 Fasolo MI, Moreira RN, Abatti PJ. Avaliação de nível de ruído em incubadora. J Pediatr (Rio J). 1994;70(3):157-62.) observed that the intensity of CN might vary according to shift.(1515 Nogueira MF, Di Piero KC, Ramos EG, Souza MN, Dutra MV. Mensuração de ruído sonoro em unidades neonatais e incubadoras com recém-nascidos: revisão sistemática de literatura. Rev Latino-Am Enfermagem. 2011;19(1):2012-21.) Heidemann et al. conducted a study in a coronary care unit and observed a positive correlation between CN and stressors.(1616 Heidemann AM, Cândido AP, Kosour C, Costa AR, Dragosavac D. Influência do nível de ruídos na percepção do estresse em pacientes cardíacos. Rev Bras Ter Intensiva. 2011;23(1):62-7.) Sampaio Neto et al. investigated the perception of professionals working in the ICU regarding noise and found that 97.3% reported that the unit had moderate to intense noise.(1717 Sampaio Neto RA, Mesquita FO, Paiva Junior MD, Ramos FF, Andrade FM, Correia Junior MA. Ruídos na unidade de terapia intensiva: quantificação e percepção dos profissionais de saúde. Rev Bras Ter Intensiva. 2010;22(4):369-74.) This study identified CN as above acceptable levels in different shifts.

Consensus 2008 reports that keeping premature infants in continuous shade deprives them of circadian information, which should be received by the end of pregnancy. Periodic low intensity brightness (180 to 200 lux) stimulates the development of the biological clock, which becomes an increasingly important part of neonatal care. Thus, not only can excessive exposure to light negatively affect development but continuous shade is also not indicated.(1818 Consenso aprovado pela Sociedade Portuguesa de Neonatologia nas XXXVI Jornadas Nacionais de Neonatologia. Viseu, 8 de maio de 2008. [citado 2016 Jul 6]. Disponível em: http://www.spp.pt/UserFiles/file/Protocolos/Luz_UCIN.pdf
http://www.spp.pt/UserFiles/file/Protoco...
) Souza et al. conducted a study that found that the artificial lights in the neonatal ICU remained on all the time.(1919 Souza KM, Ferreira SD. Assistência humanizada em UTI neonatal: os sentidos e as limitações identificadas pelos profissionais de saúde. Ciênc Saúde Coletiva. 2010;15(2):471-80.) There was no possibility of reducing brightness at certain times of the day, and the only procedure to reduce brightness involved placing a sheet over the incubator (called a "little roof"). The present study found a significant increase in brightness in the morning (p < 0.01). Additionally, the unit used fabrics (light blockers), which were placed over the incubators continuously. We observed that there was no eye protection for newborns in heated cribs.

According to Martins et al., the temperature of the incubator should be kept in the thermoneutral zone to regularly control the body temperature of the newborn, who will be exposed to relative humidity of 85% in the first week of life. From the third week of life, the relative humidity should be gradually reduced to 60% and maintained until the preterm reaches a weight of 1500 grams.(2020 Martins CP, Tapia CE. A pele do recém-nascido prematuro sob a avaliação do enfermeiro: cuidado norteando a manutenção da integridade cutânea. Rev Bras Enferm. 2009;62(5):778-83.) This study found that most of the incubators were not in accordance with the recommended standards.

Terenan et al. investigated the possible correlation between the quality of life and the severity of diseases through the Medical Outcomes Study 36 (SF-36) and Acute Physiology and Chronic Health Disease Classification System II (APACHE II) questionnaires, respectively, in patients within 72 hours of ICU admission. Despite the study limitations noted by the authors, the study showed a weak correlation between quality of life and severity of disease.(2121 Tereran NP, Zanei SS, Whitaker IY. Qualidade de vida prévia à internação em unidade de terapia intensiva. Rev Bras Ter Intensiva. 2012;24(4):341-6.) Vesz et al. investigated functional and psychological aspects immediately after ICU discharge using the modified Barthel scale and a hospital anxiety and stress questionnaire and observed a high incidence of depressive symptoms, anxiety and sleep disorders.(2222 Vesz PS, Costanzi M, Stolnik D, Dietrich C, Freitas KL, Silva LA, et al. Aspectos funcionais e psicológicos imediatamente após alta da unidade de terapia intensiva: coorte prospectiva. Rev Bras Ter Intensiva. 2013;25(3):218-24.)

CONCLUSION

The continuous noise, light and temperature intensities were not in accordance with regulatory standards and might be stressors for newborns. Special attention is needed regarding the stressors in this environment by the medical team to avoid possible psychophysiological changes that adversely interfere with the newborn's prognosis.

  • Responsible editor: Werther Brunow de Carvalho

REFERÊNCIAS

  • 1
    Pafaro RC, De Martino MM. Estudo do estresse do enfermeiro com dupla jornada de trabalho em um hospital de oncologia pediátrica de Campinas. Rev Esc Enferm USP. 2004;38(2):152-60.
  • 2
    Soares AJ, Alves MG. Cortisol como variável em psicologia da saúde. Psicol Saúde Doenças. 2006;7(2):165-77.
  • 3
    Jorge SR, Santos PB, Stefanello JM. O cortisol salivar como resposta fisiológica ao estresse competitivo: uma revisão sistemática. Rev Educ Fís/UEM. 2010;21(4):677-86.
  • 4
    Meyerhof PG. O neonato pré-termo no berçário de cuidados especiais: proposta de intervenção, respeitando sua individualidade, suas fragilidades e suas forças. Rev Bras Crescimento Desenvolv Hum. 1995;5(1/2):17-22.
  • 5
    Gaspardo CM. Dor em neonatos pré-termo em unidade de terapia intensiva neonatal: avaliação e intervenção com sacarose [Dissertação]. Ribeirão Preto: Faculdade de Medicina de Ribeirão Preto da Universidade de São Paulo; 2006.
  • 6
    Nascimento TO, Maranhão DG. Prevenção do estresse neonatal: desafio para a equipe de enfermagem. Rev Enferm UNISA. 2010;11(2):134-7.
  • 7
    Weich TM, Ourique AC, Tochetto TM, Franceschi CM. Eficácia de um programa para redução de ruído em unidade de terapia intensiva neonatal. Rev Bras Ter Intensiva. 2011;23(3):327-34.
  • 8
    Machado AC, Machado LG, Fonseca TC, Coelho RR. Níveis de ruído em uma unidade de terapia intensiva: avaliação sob a ótica da ergonomia. In: XXI Simpósio de Engenharia de Produção. Bauru, São Paulo, Brasil. 10 a 12 de novembro de 2014.
  • 9
    Rodarte MD. Exposição e reatividade do prematuro ao ruído intenso durante o cuidado em incubadora [tese]. Ribeirão Preto: Escola de Enfermagem de Ribeirão Preto da Universidade de São Paulo; 2007.
  • 10
    Brasil. Ministério da Saúde. Secretaria de Atenção à Saúde. Departamento de Ações Programáticas e Estratégicas. Atenção à saúde do recém-nascido: guia para os profissionais de saúde. Cuidados com o recém-nascido pré-termo. Brasília: Ministério da Saúde; 2011. [Série A. Normas e Manuais Técnicos, 4]. Disponível em: http://www.redeblh.fiocruz.br/media/arn_v4.pdf
    » http://www.redeblh.fiocruz.br/media/arn_v4.pdf
  • 11
    Martins CF, Fialho FA, Dias IV, Amaral JA, Freitas SC. Unidade de terapia intensiva neonatal: o papel da enfermagem na construção de um ambiente terapêutico. R Enferm Cent Oeste Min. 2011;1(2):268-76.
  • 12
    Saraiva CA. Fatores físicos-ambientais e organizacionais em uma unidade de terapia intensiva neonatal: implicações para a saúde do recém-nascido [dissertação]. Porto Alegre: Escola de Engenharia da Universidade Federal do Rio Grande do Sul; 2004.
  • 13
    Fasolo MI, Moreira RN, Abatti PJ. Avaliação de nível de ruído em incubadora. J Pediatr (Rio J). 1994;70(3):157-62.
  • 14
    Scopes JW, Ahmed I. Range of critical temperatures in sick and premature newborn babies. Arch Dis Child. 1966;41(218):417-9.
  • 15
    Nogueira MF, Di Piero KC, Ramos EG, Souza MN, Dutra MV. Mensuração de ruído sonoro em unidades neonatais e incubadoras com recém-nascidos: revisão sistemática de literatura. Rev Latino-Am Enfermagem. 2011;19(1):2012-21.
  • 16
    Heidemann AM, Cândido AP, Kosour C, Costa AR, Dragosavac D. Influência do nível de ruídos na percepção do estresse em pacientes cardíacos. Rev Bras Ter Intensiva. 2011;23(1):62-7.
  • 17
    Sampaio Neto RA, Mesquita FO, Paiva Junior MD, Ramos FF, Andrade FM, Correia Junior MA. Ruídos na unidade de terapia intensiva: quantificação e percepção dos profissionais de saúde. Rev Bras Ter Intensiva. 2010;22(4):369-74.
  • 18
    Consenso aprovado pela Sociedade Portuguesa de Neonatologia nas XXXVI Jornadas Nacionais de Neonatologia. Viseu, 8 de maio de 2008. [citado 2016 Jul 6]. Disponível em: http://www.spp.pt/UserFiles/file/Protocolos/Luz_UCIN.pdf
    » http://www.spp.pt/UserFiles/file/Protocolos/Luz_UCIN.pdf
  • 19
    Souza KM, Ferreira SD. Assistência humanizada em UTI neonatal: os sentidos e as limitações identificadas pelos profissionais de saúde. Ciênc Saúde Coletiva. 2010;15(2):471-80.
  • 20
    Martins CP, Tapia CE. A pele do recém-nascido prematuro sob a avaliação do enfermeiro: cuidado norteando a manutenção da integridade cutânea. Rev Bras Enferm. 2009;62(5):778-83.
  • 21
    Tereran NP, Zanei SS, Whitaker IY. Qualidade de vida prévia à internação em unidade de terapia intensiva. Rev Bras Ter Intensiva. 2012;24(4):341-6.
  • 22
    Vesz PS, Costanzi M, Stolnik D, Dietrich C, Freitas KL, Silva LA, et al. Aspectos funcionais e psicológicos imediatamente após alta da unidade de terapia intensiva: coorte prospectiva. Rev Bras Ter Intensiva. 2013;25(3):218-24.

Publication Dates

  • Publication in this collection
    09 Sept 2016
  • Date of issue
    Jul-Aug 2016

History

  • Received
    14 Jan 2016
  • Accepted
    17 May 2016
Associação de Medicina Intensiva Brasileira - AMIB Rua Arminda, 93 - Vila Olímpia, CEP 04545-100 - São Paulo - SP - Brasil, Tel.: (11) 5089-2642 - São Paulo - SP - Brazil
E-mail: rbti.artigos@amib.com.br