An Indirect Method to Measure Abutment Screw Preload: A Pilot Study Based on Micro-CT Scanning

This study aimed to measure the preload in different implant platform geometries based on micro-CT images. External hexagon (EH) implants and Morse Tapered (MT) implants (n=5) were used for the preload measurement. The abutment screws were scanned in micro-CT to obtain their virtual models, which were used to record their initial length. The abutments were screwed on the implant with a 20 Ncm torque and the set composed by implant, abutment screw and abutment were taken to the micro-CT scanner to obtain virtual slices of the specimens. These slices allowed the measurement of screw lengths after torque application and based on the screw elongation. Preload values were calculated using the Hooke's Law. The preloads of both groups were compared by independent t-test. Removal torque of each specimen was recorded. To evaluate the accuracy of the micro-CT technique, three rods with known lengths were scanned and the length of their virtual model was measured and compared with the original length. One rod was scanned four times to evaluate the measuring method variation. There was no difference between groups for preload (EH = 461.6 N and MT = 477.4 N), but the EH group showed higher removal torque values (13.8±4.7 against 8.2±3.6 Ncm for MT group). The micro-CT technique showed a variability of 0.053% and repeatability showed an error of 0.23 to 0.28%. Within the limitations of this study, there was no difference between external hexagon and Morse taper for preload. The method using micro-CT may be considered for preload calculation.

dental implants; dental prosthesis; torque; screw joint


Resumo

Este estudo teve como objetivo medir a pré-carga em diferentes conexões implante/pilar baseado em imagens de micro-CT. Implantes de hexágono externo (EH) e Cone Morse (MT) (n = 5) foram utilizados para a medição de pré-carga. Os parafusos de pilares foram digitalizados em um micro-CT de alta resolução para obter seus modelos virtuais, que foram utilizados para registrar o comprimento inicial. Os pilares foram parafusados sobre o implante com um torque de 20 Ncm e, o conjunto composto por implante, parafuso do pilar e pilar foi levado para o micro-CT para obter cortes virtuais dos espécimes. Esses cortes permitiram a medida do comprimento dos parafusos após a aplicação do torque. Assim, com base no alongamento dos parafusos, os valores de pré-carga foram calculados usando a Lei de Hooke. A pré-carga de ambos os grupos foram comparados pelo Test-t independente. O torque de remoção de cada espécime foi registrado. Para avaliar a precisão da técnica de micro-CT, três bastões foram escaneados em micro-CT e o comprimento do seu modelo virtual foi comparado com o comprimento original dos bastões. Um bastão foi digitalizado e mensurado quatro vezes para avaliar a variação do método de medição e a sua repetitividade. Não houve diferença entre os grupos para a pré-carga (EH = 461,6 N e MT = 477,4 N), no entanto o grupo EH apresentou maiores valores de torque de afrouxamento do parafuso (13,8 ± 4,7 contra 8,2 ± 3,6 Ncm para o grupo MT). A técnica de micro-CT mostrou uma variabilidade de 0,053% e a repetitividade apresentou um erro de 0,23 a 0,28%. Dentro das limitações deste estudo, não houve diferença entre Hexágono Externo e Cone Morse para pré-carga. O método baseado em imagens de micro-CT pode ser considerado para mensuração da pré-carga.

Introduction

One of the main complications in a single-unit implant supported prosthesis is abutment screw loosening 11. Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JYK.Clinical complications with implants and implant prostheses. J Prosthet Dent 2003;90:121-132. 22. Gracis S, Michalakis K, Vigolo P, Vult von Steyern P, Zwahlen M, Sailer I.Internal vs. external connections for abutments/reconstructions: a systematic review. Clin Oral Implants Res 2012;23:202-216.. This problem occurs due to the reduced torque of the abutment screw 33. Schwarz MS. Mechanical complications of dental implants. Clin Oral Impl Res 2000;11:156-158.. Torque maintenance depends on the preload and this in turn depends on the screw joint stability 44. Binon PP. Implants and components: Entering the new millennium. Int J Oral Maxillofac Implants 2000;15:76-94..

In the process of tightening, the screw elongates, bringing the implant and abutment together. This leads to the development of preload that causes a compressive axial force on the system, known as "clamping force", which then maintains the union between the components 55. Sakaguchi RL, Borgersen SE. Nonlinear contact analysis of preload in dental implant screws. Int J Oral Maxillofac Implants 1995;10:295-302.. The screw has to clamp the joint members together with enough force to prevent separation, slippage and self-loosening when exposed to vibration, shock, and repeated cyclical external loads 44. Binon PP. Implants and components: Entering the new millennium. Int J Oral Maxillofac Implants 2000;15:76-94. 66. Norton MR. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997;8:290-298..

Screw joint stability depends on the geometry of the implant/abutment connection. Internal connections have the advantage of protecting the screw from external non-axial forces, which can lead to loss of preload and screw loosening 77. Sutter F, Weber HP, Sorensen J, Belser U. The new restorative concept of ITI dental implants system: Design and engineering. Int J Periodont Res Dent 1997;13:409-431. 88. Pita MS, Anchieta RB, Barão VAR, Garcia Jr IR, Pedrazzi V, Assunção WG. Prosthetic platforms in implant dentistry. J Craniofac Surg 2011;22:2327-2331..

Considering that the maintenance of preload reduces abutment screw loosening 22. Gracis S, Michalakis K, Vigolo P, Vult von Steyern P, Zwahlen M, Sailer I.Internal vs. external connections for abutments/reconstructions: a systematic review. Clin Oral Implants Res 2012;23:202-216. 99. Khraisat A. Stability of the implant/abutment joint in a single-tooth external-hexagon implant system: Clinical and mechanical review. Clin Implant Dent Relat Res 2004;6:222-229., the measurements of the preload in different implant/abutment joint designs can increase the knowledge of how the abutment screw loosening occurs and might help to prevent this clinical problem with implant prostheses.

Measuring the screw elongation provides the possibility of measuring the preload. This could be achieved by using a specific equation, based on the Hooke´s Law, which allows for the calculation of the force on a member starting from the elastic modulus of the screw material, the cross-sectional area of the screw, and the length of the screw elongation 1010. Haack JE, Sakaguchi RL, Sun T, Coffey J. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536.. Thus, this study presents a new technique used to measure the abutment screw elongation for the preload calculation in different implant platform geometries and also tests the null hypothesis: There is no difference in preload values among different implant-abutment connection designs.

Material and Methods

Two groups were used for the experiment; the composition of these groups is described below (Fig. 1):

Figure 1
Specimen from group EH (left) and group MT (right).

External Hexagon Group (EH)

Five implants with an external hexagon platform, 10 mm high and 3.75 mm diameter (Master Screw; Conexão Sistemas de Prótese Ltda; Arujá, SP, Brazil). One type of pre-fabricated abutment made in cobalt-chromium alloy (Conexão Sistemas de Prótese Ltda) was attached to these implants with hex-headed titanium screws (Conexão Sistemas de Prótese Ltda) with a 20 Ncm torque as recommended by the manufacturer.

Morse Taper Group (MT)

Five Morse indexed tapered implants presenting a platform, 10 mm high and 4 mm diameter (AR Morse; Conexão Sistemas de Prótese Ltda). One type of pre-fabricated abutment made in cobalt-chromium alloy (Conexão Sistemas de Prótese Ltda) was attached to these implants with hex-headed titanium screws (Conexão Sistemas de Prótese Ltda) with a 20 Ncm torque as recommended by the manufacturer.

The abutments were screwed onto the implants, and these specimens were positioned in a vise-grip for the abutment screw tightening with a digital torque wrench (Instrutherm, São Paulo, SP, Brazil). After two minutes, the screw was re-tightened to avoid its surface setting effect and improve the torque maintenance 1111. Breeding LC, Dixon DL, Nelson EW, Tietge JD. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int J Prosthodont 1993;5:435-439. (Fig. 2). Distilled water was used to lubricate the abutment screw and simulate the moisture present in mouth environment.

Figure 2
Torque application; implant was positioned in a vise grip, and the digital torque wrench was used to apply a 20 Ncm torque.

The implant/abutment specimens were placed in a computerized tomography scanner (Skyscan 1172 High Resolution, Brukker, Belgium) for the acquisition of solid models in order to analyze the abutment screw elongation after torque application. The position of the implant was standardized by a silicone matrix (Fig. 3). The scanning process allowed for .tiff format image acquisition. This image set was transposed to .bmp files by the NRecon software (Skyscan, Aartselaar, Belgium) using the Volume of Interest (VOI) tool, which allows the determination of a standardized virtual cube for all specimens. Due to standardization of specimen position, it was possible to ensure the same position of the specimens in the virtual cube, facilitating the image section for the measuring procedure.

Figure 3
Specimen positioned in the Micro-CT chamber; a silicone matrix was used for standardize the position of specimens.

After the solid model construction, the DataViewer software (Skyscan) was used to section the virtual cube and, consequently, the virtual specimens in half, resulting in a 2D image. These images were used to measure the screw length using the Image J software (Figs. 4A and 4B). To quantify the screw elongation, the screw alone was scanned in a micro-CT scanner to register its original length by the same method (Fig. 5). The measurements of each specimen were repeated three times by the same operator.

Figure 4
Images of coronal slices obtained from micro-CT scanning. A: External hexagon implant. B: Morse tapered implant.

Figure 5
Image from micro-CT scanning of the abutment screw.

Method Accuracy and Repeatability

A rod with determined length was inserted into the implant chamber, and this set was taken to the micro-CT scanner to obtain the images by the same method used for the specimens (Fig. 6). This step was repeated four times to evaluate the repeatability of the process in seeking to separate the variation of the scanning process from the variation of the measuring process.

To evaluate the accuracy of the method, three rods with different lengths were used to evaluate if the micro-CT technique would be able to measure such low difference in length. The first rod had 5600 µm in length; the others had the same shape and diameter, but one of them was 25 µm shorter, and the other one 25 µm longer.

Figure 6
Micro-CT image (coronal slice) of a rod in the implant chamber.

Pre-load Calculation

The screw elongation was converted to preload force using the Hooke's Law:

E= σ / ε= (F ⁄ A)/(∆l ⁄ l_0 )

where:

E = modulus of elasticity (N/m2)

σ = stress, is the force F per sectional area A

ε = strain, is the change in length, ∆l, per original length, l_0.

A preload value was obtained for each group from the arithmetic mean of the values of each specimen from the group. The titanium alloy elastic modulus used was 110 GPa and was based on the literature 1212. Colling EW. The physical metallurgy of titanium alloys. Metals Park (OH): American Society for Metals; 1984. 1313. Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent 2001;85:585-598..

Loosening Torque Measurement

The torque required to loose the abutment screw of each specimen was recorded. For this, 72 h after the abutment screw tightening, the specimens were positioned in a vise-grip and the torque to remove the screw was applied with a digital torque wrench (Instrutherm); the values required to loosen the screws were recorded.

Statistical Analysis

From the mean value for each group, it was possible to compare statistically the preload intensity for different implant/abutment joints when the same tightening torque value is applied (comparison between the groups EH and MT). For this, an independent t-test (α=0.05) was used.

Results

Method Repeatability and Accuracy

The results showed a coefficient of variation (CV) of 0.053% among the four scanning procedures made for the same rod (Table 1). The accuracy showed satisfactory results as shown in Table 2, the difference between the shorter rod and the medium rod for the micro-CT measurement was 24 µm (1 µm error), while the difference between the medium and the longer rod was 29 µm (4 µm error). The error between the methods used to measure the rod lengths is also in Table 2, representing a 0.2% to 0.3% error with a consistent positive bias, which means that the error is mostly cancelled when subtracting two values to calculate the screw elongation.

Table 1
Results for length (in µm) obtained from different scanning procedures for the same specimen

Table 2
Results for different rods with different lengths (in µm) and the error between the different methods used for measure the rods (in %)

Loosening Torque

The torque required to loosen the abutment screw is in Table 3. The MT group presented lower torque maintenance.

Table 3
Loosening torque for both groups (Ncm)

Preload Measurement

The mean original screw length obtained from micro-CT measurements was 9,170 µm for MT group and 7,150 µm for EH group. The results for elongation and preload for both groups are presented in Table 4. The EH group presented higher variability. The statistical results showed no statistically significant difference between the groups for preload (p<0.05).

Table 4
Results and standard deviation (S.D.) for elongation (µm and %) and preload (N) for both groups

Discussion

Based on the results, the method using micro-CT scanning may be considered to evaluate the abutment screw elongation, since the variability of 0.053% for a specimen can be considered low. Furthermore, the measurement of specimens with known length, in this case the rods, proved to be possible to measure small changes in length as occurs with screws. However, based on the numerical results of standard deviation, the coefficient of variation for preload was approximately 50% for the EH group and 30% for the MT group, and probably it is the reason for absence of statistical difference between the groups. A larger number of specimens could reduce the high coefficient of variation.

The null hypothesis was confirmed. This study found comparable values for preload among both implant/abutment joints design (External Hexagon and Morse Tapered), as stated by Coppedê et al. 1414. Coppedê AR, Faria AC, Mattos MG, Rodrigues RC, Shibli JA, Ribeiro RF. Mechanical comparison of experimental conical-head abutment screws with conventional flat-head abutment screws for external-hex and internal tri-channel implant connections: An in vitro evaluation of loosening torque. Int J Oral Maxillofac Implants 2013;28:e321-319.. These data disagree the ones from other studies, which found higher preload values and/or higher torque maintenance for internal connections 1515. Weiss EI, Kozak D, Gross MD. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000; 84:194-199. 1616. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to determine implant preload. J Prosthet Dent 2003;90:539-546. 1717. Jorge JRP, Barão VAR, Delben JA, Assunção WG. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling. Int J Oral Maxillofac Implants 2013;28:415-422.. However, the number of specimens used in the present study was small and the comparison with other studies should be done carefully.

Comparing the mean preload values with those obtained by Haack et al. 1010. Haack JE, Sakaguchi RL, Sun T, Coffey J. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536., the present study found slightly higher values. Haack et al. 1010. Haack JE, Sakaguchi RL, Sun T, Coffey J. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536. found a mean preload of 381.5±72.9 N against 461.6 ± 228.2 N obtained in the present study for an external hexagon joint with a titanium screw and a torque of 20 Ncm. This difference could be attributed to the fact that Haack et al. 1010. Haack JE, Sakaguchi RL, Sun T, Coffey J. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536. cut the apical third of the implant, and thus a contact surface between the screw and the implant was lost. Furthermore, the previous study of Haack et al. 1010. Haack JE, Sakaguchi RL, Sun T, Coffey J. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536. did not use lubricant, which may have interfered on the results of preload.

A relationship between preload and abutment screw loosening torque was expected. However, as seen, the preload did not necessarily influence the loosening torque. This may be explained by the friction coefficient between the abutment screw and the implant surfaces. Variations in the fabrication process can lead to different contact surfaces between these parts, causing some difference in the friction coefficient. Furthermore, in this study the preload calculation was based on Hooke's law, as done by Haack et al. 1010. Haack JE, Sakaguchi RL, Sun T, Coffey J. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536.. This calculation did not consider the friction coefficient between the abutment screw and the implant surfaces. Lang et al. 1616. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to determine implant preload. J Prosthet Dent 2003;90:539-546. showed in a finite element analysis that there is an influence of the friction coefficient on the preload intensity.

The torque applied for the abutment installation was not maintained in both implant/abutment joints. The MT group lost 59% while the EH group lost 31% of the applied initial torque. The torque reduction is well documented in the literature 1111. Breeding LC, Dixon DL, Nelson EW, Tietge JD. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int J Prosthodont 1993;5:435-439. 1818. Cantwell A, Hobkirk JA. Preload loss in gold prosthesis-retaining screws as a function of time. Int J Oral Maxillofac Implants 2004;19:124-132. 1919. Kano SC, Paul Binon P, Bonfante G, Curtis DA. Effect of casting procedures on screw loosening in UCLA-type abutments. J Prosthodont 2006;15:77-81. 2020. Delben JA, Gomes EA, Barão VAR, Assunção WG. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int J Oral Maxillofac Implants 2011;26:251-256. and occurs independent of the implant/abutment configuration 1717. Jorge JRP, Barão VAR, Delben JA, Assunção WG. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling. Int J Oral Maxillofac Implants 2013;28:415-422.. The cycling procedure can accelerate the process of torque reduction 99. Khraisat A. Stability of the implant/abutment joint in a single-tooth external-hexagon implant system: Clinical and mechanical review. Clin Implant Dent Relat Res 2004;6:222-229. 2020. Delben JA, Gomes EA, Barão VAR, Assunção WG. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int J Oral Maxillofac Implants 2011;26:251-256., but it was not considered in this pilot study.

The time after torque application can influence the preload value, as demonstrated by Cantwell and Hobkirk 1818. Cantwell A, Hobkirk JA. Preload loss in gold prosthesis-retaining screws as a function of time. Int J Oral Maxillofac Implants 2004;19:124-132., who monitored the preload for 15 h after torque application and found a continuous reduction of preload values. This could be the reason for the lack of correlation between the results obtained for preload from those obtained for loosening torque, as the preload was measured immediately after the torque application while the loosening torque was measured 72 h after torque application.

Since this is a pilot study, there are some limitations that should be considered for future studies. The number of specimens was too low to provide results with reliable statistical power. Furthermore, the utilization of a cycling procedure to simulate the clinical situation must be considered, as this can directly affect the results for preload and torque maintenance. So, a study with a higher number of specimens with cycling procedure is recommended to evaluate more reliably and to compare the preload for different implant/abutment joints using the method based on micro-CT images.

Thus, the method using micro-CT images showed consistent results to measure small changes in length and could be considered to evaluate the abutment screw elongation. Within the limitations of this study, there was no difference between external hexagon and morse taper for preload, despite the higher removal torque values shown by the external hexagon group.

References

  • 1
    Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JYK.Clinical complications with implants and implant prostheses. J Prosthet Dent 2003;90:121-132.
  • 2
    Gracis S, Michalakis K, Vigolo P, Vult von Steyern P, Zwahlen M, Sailer I.Internal vs. external connections for abutments/reconstructions: a systematic review. Clin Oral Implants Res 2012;23:202-216.
  • 3
    Schwarz MS. Mechanical complications of dental implants. Clin Oral Impl Res 2000;11:156-158.
  • 4
    Binon PP. Implants and components: Entering the new millennium. Int J Oral Maxillofac Implants 2000;15:76-94.
  • 5
    Sakaguchi RL, Borgersen SE. Nonlinear contact analysis of preload in dental implant screws. Int J Oral Maxillofac Implants 1995;10:295-302.
  • 6
    Norton MR. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997;8:290-298.
  • 7
    Sutter F, Weber HP, Sorensen J, Belser U. The new restorative concept of ITI dental implants system: Design and engineering. Int J Periodont Res Dent 1997;13:409-431.
  • 8
    Pita MS, Anchieta RB, Barão VAR, Garcia Jr IR, Pedrazzi V, Assunção WG. Prosthetic platforms in implant dentistry. J Craniofac Surg 2011;22:2327-2331.
  • 9
    Khraisat A. Stability of the implant/abutment joint in a single-tooth external-hexagon implant system: Clinical and mechanical review. Clin Implant Dent Relat Res 2004;6:222-229.
  • 10
    Haack JE, Sakaguchi RL, Sun T, Coffey J. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-536.
  • 11
    Breeding LC, Dixon DL, Nelson EW, Tietge JD. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int J Prosthodont 1993;5:435-439.
  • 12
    Colling EW. The physical metallurgy of titanium alloys. Metals Park (OH): American Society for Metals; 1984.
  • 13
    Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent 2001;85:585-598.
  • 14
    Coppedê AR, Faria AC, Mattos MG, Rodrigues RC, Shibli JA, Ribeiro RF. Mechanical comparison of experimental conical-head abutment screws with conventional flat-head abutment screws for external-hex and internal tri-channel implant connections: An in vitro evaluation of loosening torque. Int J Oral Maxillofac Implants 2013;28:e321-319.
  • 15
    Weiss EI, Kozak D, Gross MD. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000; 84:194-199.
  • 16
    Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to determine implant preload. J Prosthet Dent 2003;90:539-546.
  • 17
    Jorge JRP, Barão VAR, Delben JA, Assunção WG. The role of implant/abutment system on torque maintenance of retention screws and vertical misfit of implant-supported crowns before and after mechanical cycling. Int J Oral Maxillofac Implants 2013;28:415-422.
  • 18
    Cantwell A, Hobkirk JA. Preload loss in gold prosthesis-retaining screws as a function of time. Int J Oral Maxillofac Implants 2004;19:124-132.
  • 19
    Kano SC, Paul Binon P, Bonfante G, Curtis DA. Effect of casting procedures on screw loosening in UCLA-type abutments. J Prosthodont 2006;15:77-81.
  • 20
    Delben JA, Gomes EA, Barão VAR, Assunção WG. Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws. Int J Oral Maxillofac Implants 2011;26:251-256.