Design and testing of ECE radiometer for Sinp-Tokamak

Abstract

A multi channel ECE Radiometer, in the frequency range of 75 - 110 GHz, has been designed, constructed and tested for the SINP-Tokamak to diagnose second harmonic X-mode electron cyclotron emission (ECE). This frequency range covers approximately 80% of the plasma cross section in case of a magnetic eld of B = 1.5 T. For the maximum field of the SINP-Tokamak B = 2.0 T, about 50% is covered. The ECE radiation is received by a single horn antenna with a gain of approximately 30 dB. To increase the spatial resolution perpendicular to the line of sight, a focusing lens is employed. The incoming radiation power is split into two branches by a 3dB coupler. Before down conversion of the input spectrum, the upper and the lower side band are selected respectively by means of a high pass and a low pass filter. Behind the filters, identical components are employed. The local oscillator frequency is 92 GHz for both mixers. The IF frequency range after down conversion is 2 to 18 GHz. The IF spectrum is amplified by two amplifiers with a total gain of approximately 60 dB. The outputs of these amplifiers are split into 6 branches. The frequency selection is made by band pass filters and the received power is detected by Schottky diode detectors followed by video amplification and ltering. The data is recorded by an ADC with 1 MHz sampling rate. The radiometer has been tested and calibrated with a high temperature blackbody radiator. The effective input noise temperature for the radiometer channels is found to be in the order of 0.1 eV. The linear regime of operation extends up to an input power which corresponds to an electron temperature of approximately 200 eV. Above this temperature the 1 dB compression point of the Schottky diode detectors is exceeded and the IF gain should be reduced. This can be easily done by incorporating appropriate attenuator behind the IF amplifiers.


Design and Testing of ECE Radiometer for Sinp-Tokamak

R. Ray,

Saha Institute of Nuclear Physics, Calcutta, India

C. A. J. Hugenholtz, and M. J. van de Pol

FOM-Institute for plasma physics 'Rijnhuizen',

P.O. Box 1207.3430 BE, Nieuwegein, The Netherlands

Received on 26 June, 2001

A multi channel ECE Radiometer, in the frequency range of 75 - 110 GHz, has been designed, constructed and tested for the SINP-Tokamak to diagnose second harmonic X-mode electron cyclotron emission (ECE). This frequency range covers approximately 80% of the plasma cross section in case of a magnetic eld of B = 1.5 T. For the maximum field of the SINP-Tokamak B = 2.0 T, about 50% is covered. The ECE radiation is received by a single horn antenna with a gain of approximately 30 dB. To increase the spatial resolution perpendicular to the line of sight, a focusing lens is employed. The incoming radiation power is split into two branches by a 3dB coupler. Before down conversion of the input spectrum, the upper and the lower side band are selected respectively by means of a high pass and a low pass filter. Behind the filters, identical components are employed. The local oscillator frequency is 92 GHz for both mixers. The IF frequency range after down conversion is 2 to 18 GHz. The IF spectrum is amplified by two amplifiers with a total gain of approximately 60 dB. The outputs of these amplifiers are split into 6 branches. The frequency selection is made by band pass filters and the received power is detected by Schottky diode detectors followed by video amplification and ltering. The data is recorded by an ADC with 1 MHz sampling rate. The radiometer has been tested and calibrated with a high temperature blackbody radiator. The effective input noise temperature for the radiometer channels is found to be in the order of 0.1 eV. The linear regime of operation extends up to an input power which corresponds to an electron temperature of approximately 200 eV. Above this temperature the 1 dB compression point of the Schottky diode detectors is exceeded and the IF gain should be reduced. This can be easily done by incorporating appropriate attenuator behind the IF amplifiers.

I Introduction

A radiometer using the technique of heterodyne for measuring microwaves was first reported in 1946 [1]. This technique is used for spectral decomposition of the radiation received by the radiometer. A heterodyne radiometer has several advantages over the other types of receivers. Some of these advantages are high selectivity, high stability, high accuracy and very good signal to noise ratio. Recent developments in microwave technology have been intelligently exploited to measure high frequency up to 170 GHz with much better frequency resolution and signal to noise ratio [2,3]. A spatial resolution of DR/R £ 1 %, a temporal resolution of ³ 100 kHz and a temperature resolution of DT/T £ 1 % are achievable with radiometer [4].

The principles of heterodyne detection has three distinct processes involved in it. First, the radio frequency (RF) signal which is received by the radiometer is mixed with the signal from a local oscillator. In this stage the frequency spectrum of the original signal is translated to a much lower frequency named as intermediate frequency (IF). The resulting intermediate frequency signal is subsequently amplified and finally detected and processed in a video circuit. In the process of frequency down-conversion three essential components are identified. These are local oscillator, mixer and an intermediate frequency output port. If the RF frequency is frf and the local oscillator signal has a frequency of flo respectively then the mixer produces signals of frequencies fif = | frf ± flo | and also of many combinations of the respective harmonics. Only the difference frequency | frf- flo | is of interest. Output signals with high frequency are not desirable and can be filtered out by appropriate filtering circuits.

The super heterodyne radiometer can be utilized to obtain the spatial and temporal profiles of electron temperature in a magnetically confined plasma through a measurement of the electron cyclotron emission (ECE) [5-8]. The gyration of electrons around the magnetic field lines gives rise to emission (in the non relativistic case) at gyro frequency

and its harmonics. In a tokamak, the largest component of the magnetic field, the toroidal magnetic field,

B

T varies inversely with the major radius

R. In the weakly relativistic regime, the emission in a direction to the perpendicular to the magnetic field is broadened due to the magnetic field inhomogeneity. The frequency of emission is therefore directly related to the spatial position. The temperature profile

T (

R) can therefore be obtained by a measurement of the emission spectrum from the tokamak plasma.

For a plasma in thermodynamic equilibrium, the specific intensity of emitted radiation is given by

where

I

B(w) = w

2

k

B

T

e/8p

3

c

2 is the blackbody emission, where

I(w) is intensity of radiation emitted by plasma,

T

e is plasma temperature,

k

B is Boltzmann's constant and t is optical thickness.

Re-writing we get

For an optically thick plasma (t >> 1) there is a reabsorption of emitted radiation and I(w) = IB(w), so that the observed intensity is proportional to the temperature [5]. For t << 1 plasma is optically thin and there is negligible reabsorption of the radiation emitted.

II The sinp radiometer

The SINP tokamak parameters are R0 = 30 cm, a = 7.5 cm, maximum value of toroidal magnetic field = 2 T; peak electron density = 3 × 1019 m-3 and peak electron temperature = 300 eV. [9]. In the equatorial plane, the plasma radius varies from 22.5 cm to 37.5 cm. For a toroidal magnetic field B = 1.3 T at the centre, the second harmonic of cyclotron frequency varies from 58 GHz to 97 GHz.

The optical depth for the second harmonic extraordinary wave has been calculated and has been found to be greater than 1 for the above working conditions. The following radial profiles of density, toroidal magnetic field and temperature have been utilized to calculate the profile dependence of the right hand cut off frequency fR and the upper hybrid frequency fUH [10].

with

f

ce = w

ce/2p.

The accessibility calculations reveal that the second harmonic extraordinary mode can escape the plasma for values of density » 3 × 1019m-3.

A twelve channel ECE Radiometer (Fig. 1), in the frequency range of 75 - 110 GHz has been designed, constructed & tested for SINP-Tokamak to diagnose second harmonic X- mode electron cyclotron emission (ECE). The selected central frequencies for the 12 channels are given by 74.5, 77.5, 80.5, 83.5, 86.5, 89.5, and 94.5, 97.5, 100.5, 103.5, 106.5, 109.5 GHz. The frequency range covers approximately 80 % of the plasma cross-section in case of a magnetic field B = 1.5 T. For the maximum field of the SINP-Tokamak B = 2.0 T, about 50% is covered. The minimum magnetic field is B = 1.1T, where only one channel has a corresponding electron cyclotron resonance layer inside the plasma. It is found that harmonic overlap between 2nd and 3rd harmonic occurs at the inside board of the vessel (22.5 < R < 24.0 cm for B = 1.5T) as is evident from Fig. 2. The ratio of third and second harmonic cyclotron frequency is always less than 0.1%. The selected bandwidth of each channel is 1 GHz, which yields a spatial resolution of DR/R = 1.4% at the low field side and 0.9% at the high field side, since the spatial resolution is given by

where fi is the central frequency of a channel.



The following functional requirements were chosen for the SINP-Tokamak.

(a) Frequency range of 75-110 GHz distributed over 12 channels with centre frequencies of 74.5 etc.

(b) The temperature resolution

and

(c) The spatial resolution

The radiated power from the plasma is given by

where

P is radiated power,

k is the Boltzmann constant,

T

e is plasma temperature and

B

if is bandwidth of IF amplifier.

The calculated value of radiated power for 1 GHz bandwidth is found to be 32 nW or - 45 dBm.

The noise equivalent temperature can be determined with the help of the following equation:

where

F is the overall noise figure of the radiometer and

T

amb is the ambient temperature of approximately 300K (25 meV). The noise figure F is defined as the ratio of the signal- to-noise ratio at the input to the signal-to-noise ratio at the output and can be calculated according to:

where

F

i and

G

i are the noise figure and power gain of the individual components respectively.

The ECE radiation is received by a single horn antenna with a gain of approximately 30 dB. The incoming radiation power is split into two branches by a 3 dB coupler. Before down conversion of the input spectrum, the upper and the lower side band are selected respectively by means of a high pass and a low pass filter. Behind the filters, identical components are employed. The local oscillator frequency is 92 GHz for both mixers. After down conversion the IF frequency range becomes 2 to 18 GHz. The IF signal is amplified by two cascaded amplifiers with a total gain of approximately 60 dB. The output of these amplifiers is split into 6 branches. After frequency selection by means of band pass filters, the received power is detected by Schottky diode detectors. After video amplification and filtering the data is recorded by an ADC with 1 MHz sampling rate. Video low pass filters having cut-off at 300 kHz is used to avoid aliasing. The schematic drawing of the radiometer is shown in Fig. 1.

III Calibration & testing of the radiometer

Several interesting methods are available for the calibration of ECE radiometer [11]. The SINP radiometer has been tested and calibrated with a high temperature blackbody radiator. The radiation was chopped by a 100 Hz chopper. The output signal of the radiometer was averaged by means of a digital oscilloscope, with a maximum integration time of 1000 sweeps. The effective input noise temperature for the radiometer channels was found to be in the order of 0.1 eV. The linear regime of operation extends up to an input power which corresponds to an electron temperature of approximately 200 eV. Above this temperature the 1 dB compression point of the Schottky diode detectors is exceeded and hence the IF gain needs be appropriately reduced. This can be easily done by incorporating an attenuator behind the IF amplifiers.

The calibration of the SAHA radiometer was performed with a high temperature calibration source to provide thermal emission at Thot. The operating temperature of this source can be set to 6000 C maximum. The source is placed in front of the receiving antenna of the radiometer. The distance between the source and the radiometer is approximately 40 cm, which is large enough to be in the Fraunhofer region of the antenna, but small enough to cover the complete cross-section of the antenna pattern by the emitting area of the source. A mechanical chopper with blades large enough to block the emission from the source completely is placed between the antenna and the source. The chopper enables the radiometer to detect the Thot and the room temperature (Tcold) thermal emissions alternately. Both the hot source and the chopper blades are placed with a small angle with respect to the position perpendicular to the line of sight of the radiometer to prevent standing waves to influence the measurements. The output waveform of the radiometer is recorded with a digital oscilloscope.

Since the calibration was done at a relatively low radiation temperature compared to the ECE radiation temperature in a tokamak the linearity of the radiometer was checked with the help of a Backward Wave Oscillator (BWO).

IV Results & Conclusion

The combination of a high spatial and temporal resolution, a high sensitivity and an equivalent input noise temperature of a few eV make the radiometer for the SINP Tokamak a very useful diagnostic. The radiometer is in regular use for the measurement of plasma temperature for toroidal magnetic fields ranging from 1.5 Tesla to nearly 2 Tesla. Efforts have been made to find out the temperature profile. A typical temperature radial profile is shown in Fig. 3.


Acknowledgement

The authors are thankful to J. F. M. van Gelder for his contribution in the primary design of the radiometer. The first author sincerely acknowledges the great help received from the FOM Institute of Plasma Physics in The Netherlands for the fabrication and testing of the SINP Radiometer. He gratefully acknowledges the contributions from Dr. (Mrs.) M. S. Janaki of Plasma Physics Division of SINP in the revision of the manuscript. He expresses his thanks to the members of the ECE group of the SINP Tokamak. The authors are thankful to Tony Donne of FOM Institute for his kind assistance in several stages of completion of the radiometer and also for his valuable comments and suggestions.

[3] W.S. Cheung, F.H. Levien, Artech House inc., Norwood (1985).

  • [1] R.H. Dicke, Rev. Sc. Instrum. 17, 268 (1946).
  • [2] L.N. Ridenour (Editor-in-Chief), Radiation Laboratory Series, Vol. 1-28, McGraw-Hill, New York (1948).
  • [4] J.F.M. van Gelder, Thesis: Electron Cyclotron Wave Absorption Experiments in Tokamak Plasmas, The FOM Institute of Plasma Physics, The Netherlands (1996).
  • [5] G. Bekefi, Radiation Processes in Plasma, Wiley, New York, (1966).
  • [6] A.E. Costley, R.J. Hastie, J.W.M. Paul and J. Chamberlain, Phys. Rev. Lett. 33, 758 (1974).
  • [7] M. Bornatici, R. Cano, O. De Barbieri and F. Engelmann, Nucl. Fusion 23, 1153 (1983).
  • [8] H.J. Hartfuss, T. Geist and M. Hirsch, Plasma Phys. Control. Fusion 39, 1693 (1997).
  • [9] Santwana Raychaudhuri and S. N. Sengupta, Plasma Phys. & Controlled Fusion, 34, 475 (1992).
  • [10] R.P. da Silva, A.M.M. Fonseca, E.R. Calderon and R.M.O. Galvão, Proceeding of the 14th IAEA TCM on Research Using Small Fusion Devices, Sao Paulo, Brazil, June 25-27, (2001).
  • [11] Kazuo Kawahata, Mizuki Sakamoto, Junji Fujita, Hiroshi Matsuo and Kiyomi Sakai, Japanese Journal of Applied Physics, 29, 1824 (1990).

Publication Dates

  • Publication in this collection
    26 Nov 2002
  • Date of issue
    Mar 2002

History

  • Received
    26 June 2001
Sociedade Brasileira de Física Caixa Postal 66328, 05315-970 São Paulo SP - Brazil, Tel.: +55 11 3091-6922, Fax: (55 11) 3816-2063 - São Paulo - SP - Brazil
E-mail: sbfisica@sbfisica.org.br