Efficiency loss in HPGe detectors due to beta and gamma sum coincidence

Abstract

Among the secondary detection effects in gamma-ray spectroscopy with HPGe detectors that reduce the peak area and must be properly taken into account in accurate measurements, the less well known is counting loss by beta and gamma-ray sum coincidence. The fraction of lost counts was estimated assuming that a photon detection event can sum either with the coincident beta-ray or its bremsstrahlung in the detector capsule or surrounding materials. The beta-ray was tracked through all the materials between source and detector. The result was calculated as a sequence of integral expressions evaluated numerically, without resorting to simulation. It is shown that the counting losses due to this effect are negligible for most calibration sources in typical detector arrangements.


APPLIED PHYSICS AND INSTRUMENTATION

Efficiency loss in HPGe detectors due to beta and gamma sum coincidence

Ruy M. CastroI,* * Also at: Laboratório do Acelerador Linear, Instituto de Física, Universidade de São Paulo, Travessa R 187, Cidade Universitária, CEP: 05508-900, São Paulo, SP, Brazil; and Departamento de Matemática e Física, Universidade de Taubaté, Rua Daniel Danelli, s/n, Cep: 12060-440, Taubaté, SP, Brazil ; Vito R. VaninII; Nora L. MaidanaII; Paulo R. PascholatiII; Mauro S. DiasIII; Marina F. KoskinasIII

IDivisão de Sensoriamento Remoto, Instituto de Estudos Avançados, Centro Técnico Aeroespacial, Rodovia dos Tamoios Km 5,5, Torrão de Ouro, Cep 12228-840, São José dos Campos, SP, Brazil

IILaboratório do Acelerador Linear, Instituto de Física, Universidade de São Paulo, Travessa R 187, Cidade Universitária, CEP: 05508-900, São Paulo, SP, Brazil

IIILaboratório de Metrologia Nuclear, Instituto de Pesquisas Energéticas e Nucleares, Travessa R 400,Cidade Universitária, CEP: 05508-900, São Paulo, SP , Brazil

ABSTRACT

Among the secondary detection effects in gamma-ray spectroscopy with HPGe detectors that reduce the peak area and must be properly taken into account in accurate measurements, the less well known is counting loss by beta and gamma-ray sum coincidence. The fraction of lost counts was estimated assuming that a photon detection event can sum either with the coincident beta-ray or its bremsstrahlung in the detector capsule or surrounding materials. The beta-ray was tracked through all the materials between source and detector. The result was calculated as a sequence of integral expressions evaluated numerically, without resorting to simulation. It is shown that the counting losses due to this effect are negligible for most calibration sources in typical detector arrangements.

I. INTRODUCTION

Quantitative gamma-ray spectroscopy of radioactive sources with Hyper-Pure Germanium (HPGe) detectors is affected by many secondary detection effects. This paper deals with the counting losses due to coincidental detection of a gamma-ray and the electron feeding the excited level. Not much attention has been paid to this effect, usually assumed to be insignificant, without a detailed explanation in the few references available.

The counting loss occurs when the electron, after traversing the materials between source and detector, enters the HPGe active region, adding some charge to a simultaneous photon detection event, displacing it from the expected energy value. Another possibility of loss is due to the detection of bremsstrahlung photons generated while the electron follows its path.

A procedure to estimate these sum losses is presented here, in function of the source properties and detector characteristics: beta-energy endpoint; intermediate levels life times; detector size, and Ge dead-layer thickness. The importance for many common calibration sources in a typical source-detector arrangement is shown.

II. COUNTING LOSS

Consider the decay scheme shown in Figure 1, where prompt gamma-rays follow beta transitions. The observed full energy peak area of a gamma-ray transition, already corrected for beta-gamma sum losses, is given by

where:

W = source activity;

eg= gamma-ray detection efficiency;

Pg = gamma-ray emission probability;

Dt = measurement time;

ag = internal conversion coefficient;

fb = beta decay probability and

eb= beta detection efficiency.


A fraction of the events, given by fb·eb in the second term in parentheses in the formula above, corresponds to the correction for beta-gamma sum losses, where the beta detection efficiency, eb, takes into account the geometric efficiency (egeob), combined with the probabilities that a beta (Pb) or an electron bremsstrahlung photon (PBrems) reach the detector:

with the factors entering this formula calculated as follows. The geometric efficiency is estimated from the source to detector distance and the detector area; the probability of the electron reaching the detector active region is determined by the transmission probability through the materials in the electron path, including the detector cap and the Ge dead layer; the probability of a electron Bremsstrahlung reaching the detector active region was calculated as the rate of photon production by a beta hitting the dead layer of the HPGe detector times the transmission probability along the electron trajectory.

A. Beta-gamma sum

This effect will happen whenever a photon is detected together with one electron has enough energy to reach the detector active volume. Since any amount of energy deposited by the electron in the detector simultaneously with the detection of a photon will lead to the loss of the event, Pb will depend on the electron energy threshold to reach the detector, and is given by the integral of the beta spectrum between this threshold and the beta endpoint energy (Ebmax). The electron energy threshold to enter the detector is calculated from the energy loss coefficient in the materials (µ), and the material thickness (x):

For a typical setup, with a radioactive source hold in the normal atmosphere in front of a detector, there are just a few materials involved. The energy loss coefficients for these materials can be seen in Figure 2.


The beta decay spectra was calculated by means of the model of Konopinsk [2] for allowed transitions, see Figure 3. A correction for electron reflection in thick materials [3-5] was applied to the value obtained integrating the beta spectra above the energy threshold.


B. Sum with electron Bremsstrahlung

Since high-energy electrons are more likely to produce bremsstrahlung, some care must be taken to avoid double counting of the losses. Also, the production of bremsstrahlung in the low Z materials forming the source holder and the atmosphere is negligible compared to that in the relatively high Z of the Ge dead layer. Hence, PBrems corresponds to the sum with bremsstrahlung produced by the electrons with energies between that necessary to reach the detector dead layer and the threshold energy Eb. The rate of bremsstrahlung photon production was obtained by interpolation of the table compiled by Seltzer [6]. Figure 4 shows the distribution of bremsstrahlung photons. As most of the electrons have low energy, the production of bremsstrahlung is small and concentrated in the low energy end of the photon spectrum.


III. APPLICATION

This method was applied to estimate the counting losses due to beta and gamma-ray sum coincidence for some usual standard sources assuming a typical HPGe detector-source arrangement, shown schematically in Figure 5.


A. Beta decay of some calibration sources

Since the usual source mounting is made of low Z material and is thin [7-10], then the electron threshold energy to reach the detector does not depend significantly on the particular source and was estimated in about 1.3 MeV.

After checking the requirements for beta and gamma sum (beta endpoint energy greater than the electron threshold energy and half-lives of intermediate levels involved in the decay) we found that, among the following sources used frequently in detector calibration: 46Sc; 60Co; 94Nb; 137Cs; 203Hg; 22Na; 58Co; 95Nb; 141Ce and 152Eu, there are only 2 cases where the sum is possible.

One of the cases happens in the decay of the 60Co, where the 1332 keV level is fed by b- with endpoint energy of 1492 keV with probability of 0.06 %. The other case is the 1474 keV b- decay of the 152Eu to 344 keV level of the 152Gd with 8,2 % probability. Although these beta decays are 1st and 2nd forbidden, the spectrum shape of the allowed transition [2] can be used by for an order of magnitude estimate.

B. Estimation of counting losses

The arrangement shown in Figure 5 has a geometric efficiency of 0.54 %. The probabilities of the detection of betas (Pb) are: 1.0 % and 0.7 % in the cases of 152Eu and 60Co, respectively, while the probability of detection of the bremsstrahlung photon, (PBrems ), is about 0.9 % for both cases.

Therefore, the correction (fb ·eb) can be estimated in about 7·10-6 and 1.2·10-7, of the peak areas of the g 344 keV following 152Eu decay and the g 1332 keV following the 60Co decay, respectively. Thus, the corrections for these cases are negligible.

IV. CONCLUSION

The effect of beta and gamma ray sum will be significant when using HPGe detectors with Al capsule only for radioactive sources where beta transitions have high endpoint energies, well above 1.3 MeV, and feed excited levels of the daughter nucleus. Therefore, it will rarely manifest in usual residual radioactive source measurements, and in none of the common calibration sources that we checked.

Acknowledgments

We wish to acknowledge the CNPq, FAPESP and AIEA for the partial financial support.

Received on 8 June, 2005

  • [1] L. Pages, E. Bertel, H. Joffre, and L. Sklavenitis, Energy Loss, Range, and Bremsstrahlung Yield for 10 keV to 100 MeV Electrons in Various Elements and Chemical Compounds, Atomic Data, 4, 1 (1972)
  • [2] E. Konopinski, The Theory of Beta Radioactivity, Oxford, 1966
  • [3] G. F. Knoll, Radiation Detection and Mesurement John Wiley & Sons, 2Ş Ed., 1989
  • [4] T. Tabata, R. Ito, and S. Okabe, An Empirical Equation for the Backscattering Coefficient of Electrons, Nucl. Inst. and Meth., 94, 509 (1971).
  • [5] Y. Zhou, Z. An, Q. Tang and C. Tang, Reflection of Electrons from Material Surfaces, Radiation Physics and Chemistry, 57, 731 (2000).
  • [6] S. M. Seltzer and M. J. Berger, Bremsstrahlung Spectra From Electrons Interactions with Screened Atomic Nuclei and Orbital electrons, Nucl. Inst. and Meth. in Phys. Res. B 12, 95 (1985).
  • [7] M. F. Koskinas, Características de fontes - Private communication, IPEN, (1997)
  • [8] Amersham, Gamma Reference Source Set, Data Sheet 11188, (1981)
  • [9] Ortec, Quality assurance data Sheet, detector Model No GEM-30185-P, Serial No 34-TP30930A
  • [10] R. M Castro, Transformando o 152Eu em um Padrăo de Calibraçăo de Eficięncia para Deteçăo de Gamas, Phd. Thesis, IFUSP (2003)
  • *
    Also at: Laboratório do Acelerador Linear, Instituto de Física, Universidade de São Paulo, Travessa R 187, Cidade Universitária, CEP: 05508-900, São Paulo, SP, Brazil; and Departamento de Matemática e Física, Universidade de Taubaté, Rua Daniel Danelli, s/n, Cep: 12060-440, Taubaté, SP, Brazil
  • Publication Dates

    • Publication in this collection
      07 Nov 2005
    • Date of issue
      Sept 2005

    History

    • Received
      08 June 2005
    Sociedade Brasileira de Física Caixa Postal 66328, 05315-970 São Paulo SP - Brazil, Tel.: +55 11 3091-6922, Fax: (55 11) 3816-2063 - São Paulo - SP - Brazil
    E-mail: sbfisica@sbfisica.org.br