Randomized, comparative analysis between two tracheal suction systems in newborn

Abstracts

OBJECTIVE: The objective of this study was to quantify and compare variations of oxygen saturation during the whole suction procedure (before, during and after), using two types of suction systems: open (OSS) and closed (CSS). METHODS: A controlled, randomized prospective study was carried out with 39 newborn of gestational ages > to 34 weeks using mechanical ventilation devices with a continuous flow, limited to pressure and cycled in time. Newborn were classified in two groups according to the ventilator parameters. Group I was ventilated using PEEP > to 5 cmH2O and MAP > to 8 cmH2O and Group II using PEEP < of 5 cmH2O and MAP < than 8 cmH2O. RESULTS: No statistically significant differences among the studied variable (oxygen saturation) were found in the population under study, when the two suction systems open and closed were compared , in the two groups. The oxygen saturation was statistically high after the tracheal suction in both groups. CONCLUSION: Both tracheal suction systems can be used without any drawbacks of the OSS in relation to the CSS, since the casuistry is similar to the one presented in this study.

Suction; Oximetry; Physical Therapy; Infant, Newborn; Respiration, artificial


OBJETIVO: Quantificar e comparar as variações da saturação de oxigênio durante todo o processo de aspiração (antes, durante e após o procedimento), utilizando dois tipos de sistema de aspiração: aberto (SAA) e fechado (SAF). MÉTODOS: Foi realizado estudo prospectivo randomizado e controlado com 39 recém-nascidos com idade gestacional > a 34 semanas em uso de ventilação mecânica em aparelhos com fluxo contínuo, limitados à pressão e ciclados a tempo. Os RN foram classificados em dois grupos conforme os parâmetros ventilatórios: O Grupo I foi ventilado utilizando PEEP > a 5 cmH2O e MAP > a 8 cmH2O e o Grupo II PEEP < de 5 cmH2O e MAP < de 8 cmH2O. RESULTADOS: Não se encontraram diferenças estatisticamente significantes quando comparados os dois sistemas de aspiração, aberto e fechado em nenhum dos dois grupos. Houve melhora estatisticamente significante da saturação após o procedimento em ambos grupos. CONCLUSÃO: Ambos os sistemas de aspiração endotraqueal podem ser utilizados sem nenhuma desvantagem do SAA em relação ao SAF, desde que a casuística seja semelhante à do estudo apresentado.

Sucção; Oximetria; Fisioterapia; Recém-nascido; Respiração artificial


ORIGINAL ARTICLE

Randomized comparative analysis between two tracheal suction systems in neonates

Lúcia Cândida Soares de PaulaI,*; Maria Esther Jurfest CecconII

IMestre em Pediatria pela Faculdade de Medicina da Universidade de São Paulo – FMUSP e Fisioterapeuta responsável da Unidade de Cuidados Intensivos Neonatal (UCINE) do Instituto da Criança do HC -FMUSP, São Paulo, SP

IILivre-docente em Pediatria pela Faculdade de Medicina da Universidade de São Paulo - FMUSP e Médica chefe da Unidade de Cuidados Intensivos Neonatal (UCINE) do Instituto da Criança do HC-FMUSP, São Paulo, SP

ABSTRACT

OBJECTIVE: To quantify and compare variations in oxygen saturation throughout the suctioning procedure (before, during, and after) using two endotracheal suction systems: open suction system (OSS) vs. closed suction system (CSS).

METHODS: A prospective randomized controlled study was carried out with 39 newborn infants of gestational age > 34 weeks using pressure-limited, time-cycled, continuous-flow mechanical ventilators. The infants were classified into two groups according to ventilatory parameters: Group I was ventilated using positive end-expiratory pressure (PEEP) > 5 cm H2O and mean airway pressure (MAP) > 8 cm H2O; and Group II using PEEP < 5 cm H2O and MAP < 8 cm H2O.

RESULTS: No statistically significant differences were observed when OSS and CSS were compared in both groups. There was a statistically significant improvement in post-procedure oxygen saturation in both groups.

CONCLUSION: Both endotracheal suction systems can be used with no drawbacks of OSS in relation to CSS, provided the sample is similar to that of the present study.

Key words: Suction. Oximetry. Physical Therapy (Specialty). Infant, Newborn. Respiration, Artificial.

INTRODUCTION

The increased survival of high-risk neonates in recent decades is the result of technological and scientific development in perinatal and neonatal medicine and improvements mainly in respiratory care, with microprocessor-based mechanical ventilators and monitoring equipment appropriate for the neonatal period1.

Ventilatory support may be administered in a noninvasive and invasive manner, and both are important resources in the treatment of infants admitted to the neonatal intensive care unit (NICU). However, especially invasive support, which is the subject of our study, may be associated with some complications2-4.

There is great variation in physical therapy techniques used in neonates, and among them we report endotracheal suction. Endotracheal suctioning can be performed through two systems described below.

The conventional or open suction system requires the opening of the breathing circuit through the disconnection of the endotracheal tube from the mechanical ventilator and insertion of a suction catheter of appropriate size for the diameter of the endotracheal tube.

The closed suction system was developed in the late 1980s and employs a multiple-use suction catheter that is attached to the ventilator circuit, with no need to disconnect it from the patient to perform suction5.

The main objective of endotracheal suctioning is to remove secretions, preventing airway obstruction, atelectasis, and pulmonary infections.6 Although it is recognized that endotracheal suction is essential for patients, its use may produce some adverse effects, such as: decreased oxygenation, changes in heart rate, microatelectasis, bronchoconstriction, increased intracranial pressure, infection, trauma to the tracheobronchial mucosa, pneumothorax, apnea, and even cases of death. Death cases have been reported in studies conducted with adult patients7-12.

Two endotracheal suction systems are commercially available: the open endotracheal suction system (OSS) and the closed endotracheal suction system (CSS). The objective of this study was to compare variations in oxygen saturation throughout the suctioning procedure (before, during, and after) using the two available suction systems.

METHODS

The study was approved by the Research Ethics Committee of Universidade de São Paulo (USP), Brazil, and all parents/legal guardians provided written informed consent. A randomized clinical trial was performed with 39 newborn infants admitted to the NICU at the Children's Institute of USP Hospital das Clínicas. Infants who had cardiac malformations confirmed by echocardiography, tension pneumothorax requiring drainage or not, extensive atelectasis affecting at least one lobe, inhaled nitric oxide (NO), or received high-frequency ventilation were excluded from the study.

The 39 infants studied were classified into two groups according to the need for lower or higher ventilatory parameters, as described below, in order to verify whether the lungs subjected to different inspired oxygen fractions and mean airway pressures would behave differently during and after the suctioning procedure.

Group I – Infants who required the following ventilatory parameters: fraction of inspired oxygen (FIO2) > 50%; positive inspiratory pressure (PIP) > 20 cm; positive end-expiratory pressure (PEEP) > 5; respiratory rate (RR) > 30 bpm; flow > 6 L.

Group II – Infants who required the following ventilatory parameters: FIO2 < 50%; PIP < 20 cm; PEEP < 5; RR < 20 bpm; flow < 6 L.

Randomization was performed before each procedure by using cards identified as OSS and CSS, which were placed inside dark envelopes and randomly distributed to the draw.

OSS was performed according to the following criteria: decrease of saturation below 89%; respiratory distress; pulmonary auscultation; and patient restlessness. In all infants preductal oxygen saturation (SpO2) was measured by pulse oximetry (right radial artery). Multiprogrammable neonatal monitors were used: V24E® (Phillips); Infinity vista XL® (Dräger); and Portal DX2020® (Dixtal). SpO2 was measured at the following time points:

Ta- Before the infant was positioned for suctioning (time required to check the condition of the infant); Tb- IImmediately before suctioning (infant held in position); Tc- During all suctions, Tc1 - after the first suction, Tc2 - after the second suction; Td- Five minutes after the procedure; Te- Ten minutes after the procedure.

ABefore starting the suctioning procedure, the following care procedures were observed and followed:

1) Conditions for fixation of endotracheal tube: it was always kept patent.

2) The length of the tube for fixation after intubation was calculated from the Torchen rule13 (1979), represented by the weight in kilograms6. Immediately after this procedure a radiologic examination was performed to confirm the position of the tip of the tube in the intervertebral space between the second and third thoracic vertebrae (T2 and T3) of the infant.

3) Humidification was maintained by constantly filling the humidifier with water, maintaining the power between 4 and 5 to reach an inspired gas temperature of 37º C, as recommended by the manufacturer Fisher Paykel.

5) The infants were positioned supine 10 minutes before care was started, with the head in the midline and neck in mild extension.

After initial care procedures: preductal SpO2 of the infant was noted down; the infant was preoxygenated, with a 10% increase in previous FIO2; and after a three-minute interval the suctioning procedure was initiated. During suctioning no type of solution, such as saline and/or distilled water, was instilled. A suction catheter #6 or #8, according to the size of the endotracheal tube (2.5, 3.0, 3.5, and 4 cm) was inserted only twice during each procedure, and the time interval between insertions was three minutes. The negative pressure used during suctioning was 15 to 20 cmHg.

Statistical analysis was performed as follows: for qualitative variables absolute and relative frequency were calculated, and for quantitative variables median and minimum and maximum values were calculated. The association of both groups (Groups I and II) with each of the qualitative variables was performed using the chi-square test or Fisher's exact test, when the former was not suitable. The comparison between both groups and quantitative variables was performed using the Mann-Whitney test.

For the variables analyzed, in each of the suction systems, which had more than one measurement over time (SpO2), the average profile was constructed with vertical bars representing the standard error. Repeated-measures analysis of variance (ANOVA) was used for the modeling of these variables, which allowed an adjustment taking into account a possible correlation of data collected.14,15 Based on this model, contrasts were constructed that allowed us to evaluate possible differences between both systems and to evaluate variations in these measurements during monitoring of each system studied. All statistical analyses were performed using SAS 9.1 for Windows. A significance level of 5% was adopted in all statistical analyses, i.e., results with p < 0.05 were considered as statistically significant.

RESULTS

The two groups of infants were homogeneous in terms of sex, newborn classification, weight, and mode of mechanical ventilation (p > 0.05). However, in relation to mode of ventilation, there was a higher percentage of infants receiving intermittent mandatory ventilation (IMV) in Group II compared to the percentage observed in Group I, but the statistical difference was marginally significant (p = 0.055). Group I had a higher percentage of infants in use of vasoactive drugs and sedatives (p < 0.05). There was no significant difference between groups with respect to analgesia (p = 0.182) (Tables 1, 2, and 3 ).

Regarding the parameters studied in Group I, as seen in Table 4 and Figure 1, mean SpO2 at time point Ta was higher in the OSS than in the CSS. A decrease in SpO2 was then observed from Tb until Tc1 in both OSS and CSS. From Tc1 to Tc2, both systems showed an increase in mean SpO2. SpO2 continued to rise from Tc2 to Td in both systems; however, mean SpO2 in the CSS was apparently lower than that observed in the OSS. At time point Te, in both groups, SpO2 was similar to that observed in the previous time point, and the difference in mean SpO2 between systems remained the same. Despite the difference described by repeated-measures ANOVA, the interaction between system and time produced no statistically significant effect, i.e., both systems showed the same SpO2 behavior over time and there was a statistically significant increase in SpO2 in both systems from time point Ta to Te, demonstrating effectiveness of the procedure.


Regarding SpO2 in Group II, as described in Table 5 and Figure 2, both systems showed SpO2 profiles with the same behavior over time, with OSS showing higher mean SpO2 than CSS throughout the monitoring period.


Repeated-measures ANOVA revealed that both systems showed the same SpO2 behavior throughout the evaluation period (p = 0.964). There was no statistically significant difference in mean SpO2 between systems (p > 0.05). Time had a statistically significant effect on both systems evaluated (p < 0.05).

DISCUSSION

The literature reports possible advantages and disadvantages of each suction system. CSS, for example, is believed to induce fewer episodes of hypoxia and its consequences; however, suction catheters used in CSS are more expensive than those used in OSS and are not available in all health care facilities16-18.

In our research, infants in both groups (I and II) showed no statistically significant differences in relation to weight (mean of 3,150 g in Group I and 2,995 g in Group II), sex (10 female infants in Group I and 13 in Group II; 8 male infants in Group I and 8 in Group II), and newborn classification (5 preterm infants in Group I and 2 in Group II; 13 term infants in Group I and 19 in Group II). Therefore, our sample was considered a homogeneous group suitable for comparison. Gestational age greater than 34 weeks was chosen because infants of gestational age less than 34 weeks have immature neurological and cardiorespiratory systems, which could lead to alterations inherent in prematurity.

Diagnoses were diverse, including: respiratory problems, such as pneumonia, bronchiolitis, and pertussis; surgical problems, such as esophageal atresia, omphalocele, gastroschisis, and myelomeningocele; and neurological conditions, such as perinatal asphyxia.

Midazolam was used for sedation, and dipyrone, fentanyl, or a combination of both drugs was used for analgesia. A statistically significant difference was observed between groups regarding sedation and analgesia, being higher in Group I, as shown in Table 3 . This difference is due to the fact that this group was composed of infants requiring increased ventilatory support.

Intubation and mechanical ventilation are major causes of pain and stress in neonates. Mechanical ventilation and related invasive procedures are associated with several biochemical, physiologic, and behavioral alterations, indicating that pain and stress in neonates may lead to clinical instability and an adverse clinical prognosis. In ventilated infants, pain may lead to restlessness and asynchrony between ventilation and spontaneous breathing, resulting in inadequate ventilation. Endotracheal suction is one of these invasive procedures; however, thus far there is no evidence that the use of drugs and non-pharmacological therapy can minimize pain during the procedure.19-20 An important finding in our study is that the sedative doses used allowed spontaneous breathing of the infant, which is an important aspect, since the decrease in spontaneous breathing can lead to larger lung volume loss during suctioning and hinder the recruitment of this volume21.

The care observed during the suctioning procedure derives from the existing data mentioned above and from the experience of the physical therapy team of the NICU, which performs all physical therapy treatments and is present over 24 hours every day. We described in detail the standardization of the procedure aiming to show that the results obtained may be due to the type of method used, which can be reproduced in other health care facilities.

Infants included in this study were classified into two groups according to ventilatory support. The most relevant parameters were PEEP and mean airway pressure (MAP), which maintain alveolar recruitment and oxygenation, respectively. Group I was ventilated using PEEP > 5 cm H2O and MAP > 8 cm H2O, and Group II using PEEP < 5 cm H2O and MAP < 8 cm H2O.

A major concern in relation to endotracheal suction is lung volume loss, which can promote alveolar collapse and, thus, the need to reopen the lung units, which in turn may cause acute lung injury. Arterial oxygenation coincides with changes in lung volume21.

Studies by Maggiore et al. (2003)21, Cereda et al.22 (2001); and Lasocki et al23. (2006) in adult patients with acute lung injury showed decreased lung volume and decreased arterial oxygenation after the use of OSS. Choong et al.24 (2003) corroborate the previous data in a study of children aged between 6 days and 13 years in which OSS and CSS were compared. Those authors showed that total lung volume loss was significantly greater with OSS and children suctioned with this system desaturated to a greater extent; thus, they suggested that CSS should be used in pediatric patients, especially those with significant lung disease, who require high PEEP, to avoid alveolar derecruitment and hypoxia during suctioning.

However, a more recent study by Hoellering et al.25 (2008) with 30 neonates, 20 receiving synchronized intermittent mandatory ventilation (SIMV) and 10 high-frequency oscillatory ventilation (four of these receiving muscle relaxant), in which OSS and CSS were performed in a random order, using same size catheters, same value for negative pressure, and same duration for both procedures, showed that both systems produced a transient reduction in heart rate and SpO2. Those same authors reported that lung volume loss was not affected by the type of system, as well as regaining lung volume was not influenced when patients were receiving SIMV. However, time to regain lung volume was longer with OSS than CSS, when patients were receiving high-frequency ventilation.

Regarding the results of our population, comparisons at six predetermined time points showed no statistically significant differences in SpO2 between the two systems in both groups (Table 1/Figure 1, Table 2/Figure 2). In both systems SpO2 showed a decrease during the suctioning procedure at time points Tc1 and Tc2, although this fall was not clinically significant. Mean values remained within the SpO2 limits recommended for neonates, which is 89-95%26.We believe that the main strategy to prevent and/or minimize hypoxemia induced by endotracheal suction is to limit the procedure time to 15 seconds and preoxygenation to only 10% above the initial FIO2, approaches observed in this study. Other strategies, such as hyperventilation and hyperinsufflation, should be avoided in this population.

A study by Heinze et al27. (2008) analyzed changes in functional residual capacity of 20 postoperative cardiac surgery adult patients after suctioning with OSS and CSS in different ventilatory modes. The study showed a decrease in functional residual capacity after suctioning, regardless of the system used. Similar to our study, SpO2 recovered rapidly after suctioning, regardless of the suction system or ventilatory mode used.

The ability to maintain spontaneous breathing during ventilation so that patients can recruit their own lung volume may be more important than the suctioning method used. The effect of endotracheal suction on lung volume may depend on the ventilatory mode and settings, the suctioning technique and duration, as well as on the ratio between the diameters of the endotracheal tube and the suction catheter21,27.

Therefore, this study provided us with evidence of safety to continue using OSS, which is the most widely used in NICUs, since there were no statistically significant differences in oxygenation. Nevertheless, we point out that patients with more severe lung disease than that observed in our sample, such as patients with diaphragmatic hernia associated with persistent pulmonary hypertension of the newborn, might benefit more from the use of CSS than OSS, as long as the care procedures mentioned in this study are observed.

Conflict of interest: No conflicts of interest declared concerning the publication of this article.

REFERENCES

  • 1. Carlo WA. Permissive hypercapnia and permissive hypoxemia in neonates. J Perinatol. 2007;27:S64-S70.
  • 2. Main E, Castle R, Newham D, Stocks J. Respiratory physiotherapy vs. suction: the effects on respiratory function in ventilated infants and children. Intensive Care Med. 2004;30:1144-51.
  • 3. Woodgate PG, Flenady V. Tracheal suctioning without disconnection in intubated ventilated neonates. Cochrane Database Syst Rev. 2001;(2):CD003065.
  • 4. Morrow BM, Argent AC. A comprehensive review of pediatric endotracheal suctioning: effects, indications, and clinical practice. Pediatr Crit Care Med. 2008;9:465-77.
  • 5. Carlon GC, Fox SJ, Ackerman NJ. Evaluation of a closed-tracheal suction system. Crit Care Med. 1987;15:522-5.
  • 6. Morrow B, Futter M, Argent A. Effect of endotracheal suction on lung dynamics in mechanically-ventilated paediatric patients. Aust J Physiother. 2006;52:121-6.
  • 7. Mosca FA, Colnagghi M, Lattanzio M, Bray M, Pugliese, Fumagalli M. Close versus open endotraceal suctioning in preterm infants: effects on cerebral oxygenation and blood volume. Biol Neonate. 1997;72:9-14.
  • 8. Simbruner G, Coradello H, Fodor M, Havelec L, Lubec G, Pollak A. Effect of tracheal suction on oxygenation, circulation, and lung mechanics in newborn infants. Arch Dis Child. 1981;56:326-30.
  • 9. Young CS. A review of the adverse effects of airway suction. Physioteherapy. 1984;70:104-8.
  • 10. Shah AR, Kurth CD, Gwiazdowski SG, Chance B, Papadopoulos M.Fluctuations in cerebral oxygenation and blood volume during endotracheal suctioning in premature infants. J Pediatr. 1992;120:769-74.
  • 11. Perlman JM, Volpe JJ. Suctioning in the preterm infant: effects on cerebral blood flow velocity, intracranial pressure, and arterial blood pressure. Pediatrics. 1983;72:329-34.
  • 12. Saliba E, Chantepie A, Autret E, Gold F, Pourcelot L, Laugier J. Effects of indomethacin on cerebral hemodynamics at rest and during endotracheal suctioning in preterm neonates. Acta Paediatr Scand. 1991;80:611-5.
  • 13. Tronchin DMR, Tsunechiro MA. Fixação da cânula orotraqueal em recém nascido. Rev Paul Enferm. 2002;21:163-7.
  • 14. Crowder MJ, Hand DJ. Analysis of repeated measures. London: Chapman and Hall; 1990. p. 257.
  • 15. Singer J.M, Andrade D.F. Analysis of longitudinal. In: Sen PK, Rao CR. Bio-environmental and public health statistics. Amsterdam: North Holland; 2000. p.115-60. (Handbook of Statistics nş 18).
  • 16. Thakur A, Buchmiller T, Atkinson J. Bronchial perforation after closed-tube endotracheal suction. J Pediatr Surg. 2000;35:1353-5.
  • 17. Wash CM, Bada HS, Korones SB, Carter MA, Wong SP, Arheart K. Controlled supplemental oxygenation during tracheobronchial hygiene. Nurs Res. 1987;36:211-5.
  • 18. García-Aparicio L, Castañón MC, Tarrado X, Rodríguez L, Iriondo M, Morales L. Bronchial complication of a closed-tube endotracheal suction catheter. J Pediatr Surg. 2002;37:1483-4.
  • 19. Aranda JV, Carlo W, Hummel P, Thomas R, Lehr VT, Anand KJ. Analgesia and sedation during mechanical-ventilation in neonates. Clin Ther. 2005;27:877-99.
  • 20. Cignacco E, Hamers JP, Van Lingen RA, Zimmermann LJ, Müller R, Gessler P, et al. Pain relief in ventilated preterm infants during endotracheal suctioning: a randomized controlled trial. Swiss Med Wkly. 2008;138:635-45.
  • 21. Maggiore SM, Lellouche F, Pigeot J, Taille S, Deye N, Durrmeyer X, Richard JC, Mancebo J, Lemaire F, Brochard L. Prevention of endotracheal suctioning induced alveolar derecruitment in acute lung injury. Am J Respir Crit Care Med. 2003;167:1215-24.
  • 22. Cereda M, Villa F, Colombo E, Greco G, Nacoti M, Pesenti C. Close system endotracheal suctioning maintains lung volume during volume-controlled mechanical ventilation. Intensive Care Med. 2001;27:648-54.
  • 23. Lasocki S, Lu Q, Sartorius A, Fouillat D, Remerand F, Rouby JJ. Open and closed- circuit endotracheal suctioning in acute lung injury: efficiency and effects on gas exchange. Anesthesiology. 2006;104:39-47.
  • 24. Choong K, Chatrkaw P, Frndova H, Cox PN. Comparison of loss in lung volume with open versus in-line catheter endotracheal suctioning. Pediatr Crit Care Med. 2003;4:69-73.
  • 25. Hoellering A, Copnell B, Dargaville PA, Mills JF, Morley CJ, Tingay DG. Lung volume and cardiorespiratory changes during open and close endotracheal suction in ventilated newborn infants. Arch Dis Child Fetal Neonat.. 2008;93:F436-41.
  • 26. Santos AMN. Monitorização do recém nascido com doença pulmonar. In: Alves Filho N, Trindade O. Clínica de perinatologia: aparelho respiratório em neonatologia. Parte I. Rio de Janeiro; Medsi; 2001. p.17-30.
  • 27. Heinze H, Sedemund-Adib B, Heringlake M, Gosch UW. Eichler W. Funcional residual capacity changes after diferent endotracheal suctioning methods. Anesth Analg. 2008;107:941-4.
  • Publication Dates

    • Publication in this collection
      12 Nov 2010
    • Date of issue
      2010

    History

    • Received
      13 Jan 2010
    • Accepted
      03 May 2010
    Associação Médica Brasileira R. São Carlos do Pinhal, 324, 01333-903 São Paulo SP - Brazil, Tel: +55 11 3178-6800, Fax: +55 11 3178-6816 - São Paulo - SP - Brazil
    E-mail: ramb@amb.org.br