Introduction
The recording of otoacoustic emissions (OAE) is a method used for the detection of hearing changes of cochlear origin.1 Among other functions, the cochlea amplifies sounds, which is performed nonlinearly in healthy cochlea, that is, there is a compression mechanism within the cochlea that enables the amplification to occur at different proportions depending on the intensity of the stimulus (weak, medium and strong). For weak sounds the amplification is linear, and for medium intensity sounds the amplification is lower, which is called cochlear compression, Ruggero.2
Direct measurements of cochlear nonlinearity are not possible. As a result, access to the nonlinearity of the human cochlea has to be undertaken with noninvasive and indirect measures, according to Gorga et al..3
With the measurement of distortion-product otoacoustic emissions (DPOAE) it is possible to verify the growth in the response according to the level of intensity of the sound stimulus presented (growth curve), as per Abdala.4
The study of the DPOAE growth function has proven to be a very favorable instrument for access to and understanding of cochlear physiology, especially compression mechanisms and cochlear nonlinearity, as well as the function of outer hair cells. The realization of the DPOAE growth curve, also known as the input/output (IO) growth function makes it possible to obtain the DPOAE threshold.5 The occurrence of DPOAE compression in newborns can be interpreted as an indication of reduction in gain from cochlear amplification with an increase in the stimulus level. The presence of compression in this phase, in addition to suggesting cochlear integrity, is a strong indicator that already at birth the peripheral function enables greater amplification of sounds with a low intensity level, favoring auditory tasks that require temporal resolution ability. Nonlinearity is important for recognizing speech in a noisy environment.6 The sounds are dynamic, with constant changes in the spectrum, amplitude, spatial location and duration parameters. The process of perceiving quick changes in the pattern of the stimulus, integrating information occurring over time, is denominated temporal processing. Knowledge of the occurrence of this pattern at birth confirms the hypothesis of maturity in the peripheral auditory system in the neonatal phase, predisposing the development of temporal resolution ability in this phase.7
Thus, importance of research on the possible clinical applications of this test emerges. Cochlear physiology represents an important subject for study and has become a major motivation in this work: to obtain a more specific understanding of the cochlea and its operation, thereby collaborating when there are losses to these cochlear functions, mainly aimed at the neonatal population, where intervention and early stimulation measures can be developed, as well as complementing newborn hearing screening (NHS). Non-linear amplification of the sound made by the cochlea indicates that it is working normally, especially in relation to the OHC function, and a linear amplification for all levels of stimulus intensity indicates a cochlea with abnormalities.8
In alterations to the middle ear only, the function of the OHCs is not compromised, so the growth curve obtained will have the same configuration as a normal examination, i.e. with nonlinear amplification, presenting only an increased DPOAE threshold. The realization of DPOAE growth curves in babies who “fail” the Newborn Hearing Screening Program would be important and useful to differentiate the presence of amniotic fluid/vernix in the EAM (external auditory meatus) and/or tube dysfunction from a cochlear abnormality,9 optimizing the referral for audiological diagnosis.
Obtaining the thresholds for the emergence of DPOAE helps in the recommendation of hearing aids in the cases of mild to moderate hearing losses, increasing the information provided to the professional who will perform this procedure.10
In addition to its possible applications, this line of research has not been well explored in newborns, meaning that studies in this area and on this specific population are justified.
Objective
To determine the DPOAE emergence thresholds at the frequencies of 2000 Hz and 4000 Hz with a variation in the stimulus of 20 to 65dBNPS.
To establish the slope values obtained in the growth curves at the frequencies of 2000 Hz and 4000 Hz as a function of the intensity.
Methods
This is a prospective, cross-sectional and descriptive study. This was approved by the Research Ethics Committee of São Paulo’s Holy House of Mercy Brotherhood (project 277/10) and was conducted at the institution’s Speech Pathology Sector, with newborns born between March and August 2011. All of the newborns tested were born at São Paulo’s Holy House of Mercy Brotherhood and were tested after being discharged from hospital. At that time, the parents or guardians received a referral for realization of newborn hearing screening at the hospital’s Speech Pathology Sector, at scheduled dates and times, within 4 weeks of birth, enabling investigation within the neonatal period.
Case-based reasoning
Inclusion criteria:
This study included 39 newborns. A total of 14 exams were conducted on each newborn. The total assessment period was an average of 30 minutes per newborn in a single session.
Characterization of the sample
51 newborns were invited to participated in the study, however 12 did not maintain the conditions required to realize the exam either because they woke up during the assessment, did not sleep for testing to begin, or did not respond in the newborn hearing screening after otoacoustic emissions. Thus, 39 newborns participated in the study, 21 (53.8%) male and 18 (46.2%) female. The perinatal characteristics of the newborns can be viewed in Table 1.
Equipment
The equipment used for the collection of the data in the study was the ILO V6 Otodynamics Analyzer, Cochlear Emissions Analyzer ILO292 DP Echoport Version 6, Otodynamics, London. A newborn probe was used for insertion (SGS-8) connected to channel A of the external unit of the ILO equipment.
Procedure
The exams were performed in a soundproof booth. The shape of the stimulus curve was monitored in the ear in ‘real-time’, ensuring the quality and reliability of the data collected. The probe was acoustically calibrated daily in a cubic centimeter cavity, for assessment of newborns. No otoscopy was conducted before the procedure.
In a state of natural sleep, the newborn was positioned to conduct the assessment, starting 50% of the exams in the right ear and 50% in the left ear, thus ensuring no interference from the first ear tested.
The probe was adapted to capture DPOAE growth curves (GC) in the first ear to be tested, results being recorded with the following order of procedures:
DPOAE GC at 2000Hz general paradigm
DPOAE GC at 2000Hz paradigm 1
DPOAE GC at 2000Hz paradigm 2.
DPOAE GC at 4000Hz paradigm 1.
DPOAE GC at 4000Hz paradigm 2.
Where:
Paradigm 1 (P1): L2 from 20 to 40 dB NPS.
Paradigm 2 (P2): L2 from 40 to 65 dB NPS.
General Paradigm (GP): L2 from 35 to 65 dB NPS
Next, the same procedures were repeated in the second ear.
The DPOAE GC were obtained for paradigms P1 and P2 based on the hypothesis that the slope would be higher (greater incline of the curve) in weak sounds (20 to 40dB SPL) and lower (lower incline of the curve) in midrange sounds (40 to 65dB SPL).
The quality of each exam was observed, ensuring equal assessment conditions with regard to the placement of the probes, stimulus intensity, stability and noise level. Thus, although the study of OAE in the neonatal population is delicate, the reliability of the response observed as increased.
The DPOAE growth curve was obtained for both ears. Two pure tones with primary frequencies denominated f1 and f2 were presented. The responses for analysis were obtained at 2f1-f2, with f1/f2 = 1.22. The f2 frequencies chosen were 2000 Hz and 4000 Hz for being the frequencies used most often in studies with this population. 13
The intensities (L2) ranged from 20 to 65 dB SPL divided into three paradigms:
Paradigm1: L2 (intensity of frequency F2) varied from 20 to 40 dB SPL with a 3 dB interval.
Paradigm2: L2 (intensity of frequency F2) varied from 40 to 65 dB SPL with a 3 dB interval.
General paradigm: L2 (intensity of frequency F2) varied from 35 to 65 dB SPL with a 5 dB interval.
The intensity of L1 (intensity of frequency 1) was adjusted using the formula L1= 0.4L2+39 from Kummer et al.14
The exams were finalized when the background noise was equal to or lower than 0 dB SPL after a minimum of three scans.
The following criteria were adopted to analyze the response in the signal to noise ratio (S/N):
Lim0: intensity at which the S/N ratio is greater than or equal to zero – this was called Lim0 (threshold zero).
Lim6: intensity at which the S/N ratio is greater than or equal to six – this was called Lim6 (threshold six).
To obtain the slope, three criteria were established in accordance with the intensities established:
Statistical analysis
The statistical analysis was undertaken using the program SPSS version 13.0.
The sample calculation was performed using Prime software, considering a significance level of 0.05.
For the descriptive analysis, the summary measures for quantitative variables were calculated. For qualitative variables, absolute and relative frequencies were obtained and boxplot graphs were generated, as well as Wilcoxon signed-rank test, adopting a significance level of 5% for all tests.
Results
Study of DPOAE emergence thresholds obtained with the growth curve
The Lim0 and Lim6 thresholds (described in Methods) on the right and left ears at frequencies of 2000 Hz and 4000 Hz were compared (Table 2). It was noted that the differences between the parameters used in the research (Lim0 and Lim6) were significant at the frequencies of 2000 Hz and 4000 Hz in both ears, but when comparing the left ear and right ear the differences were not significant.
TABLE 2 Distribution of emergence thresholds in decibels obtained in the 2 parameters (Lim0 and Lim6) by ear and by frequency
Right ear | Left ear | |||||||
---|---|---|---|---|---|---|---|---|
2000 Hz | 4000 Hz | 2000 Hz | 4000 Hz | |||||
L0 | L6 | L0 | L6 | L0 | L6 | L0 | L6 | |
Minimum | 22 | 25 | 28 | 28 | 19 | 19 | 19 | 19 |
Average | 28.13 | 32.32 | 34.69 | 38.66 | 28.97 | 33.27 | 33.62 | 37.21 |
SD | 6.25 | 7.7 | 6.8 | 8 | 6.4 | 7.2 | 8.4 | 9.4 |
Median | 25 | 28 | 31 | 37 | 28 | 32.5 | 31 | 37 |
Maximum | 53 | 53 | 59 | 44 | 44 | 50 | 62 | 62 |
L0: threshold where the signal to noise ratio is greater than or equal to zero; L6: threshold where the signal to noise ratio is greater than or equal to six; SD: standard deviation; p: significance level.
Descriptive study of the slope obtained with the DPOAE growth curve
A descriptive analysis was undertaken of the slope values obtained with the realization of the DPOAE GC at frequencies of 2000 Hz and 4000 Hz in the three paradigms used in the study (Table 3 and 4).
The values of the slope obtained in paradigms P1 and P2 at 2000 Hz in the right ear and the left ear were compared, and no significant differences were found between them in both ears. When comparing the values of the slope obtained in paradigms P1 and P2 at 4000 Hz in the right ear and the left ear, no significant difference was observed between them. We compared the slope values obtained at frequencies of 2000 Hz and 4000 Hz in paradigm 1 in the right ear, and no significant difference between the frequencies was obtained, with the same occurring in paradigm 2. When comparing the slope values obtained at frequencies of 2000 Hz and 4000 Hz in paradigm 1 and paradigm 2, on the left ear, a significant difference was not obtained between frequencies.
TABLE 3 Distribution of the slope values obtained at 2000 Hz in the 3 paradigms for the right ear and left ear
2000Hz | ||||||
---|---|---|---|---|---|---|
Right ear | Left ear | |||||
P1 | P2 | GP | P1 | P2 | GP | |
Minimum | -0.15 | -0.62 | 2.15 | 0.33 | 0.9 | 1.51 |
Mean | 3.55 | 3.91 | 10.05 | 3.51 | 4.26 | 8.85 |
SD | 2.08 | 2.56 | 6.2 | 3.1 | 2.42 | 4.38 |
Median | 3.16 | 3.21 | 8.98 | 2.33 | 3.74 | 9.16 |
Maximum | 8.08 | 9.19 | 28.06 | 14.59 | 10.96 | 17.03 |
SD: standard deviation; P1: 1 paradigm (stimulus intensity ranging from 20 to 40 dB SPL); P2: paradigm 2 (stimulus intensity ranging from 40 to 65 dB SPL); GP: general paradigm (stimulus intensity ranging from 35 to 65 dB SPL).
TABLE 4 Distribution of the slope values obtained at 4000 Hz in the 2 paradigms for the right ear and left ear
4000Hz | ||||
---|---|---|---|---|
Right ear | Left ear | |||
P1 | P2 | P1 | P2 | |
Minimum | 0.54 | -1.38 | -0.04 | 0.2 |
Mean | 3.97 | 3.7 | 3.01 | 3.06 |
SD | 3.17 | 3.51 | 2.65 | 2.84 |
Median | 2.6 | 2.68 | 2.4 | 2.51 |
Maximum | 11.58 | 15.9 | 12.33 | 14.44 |
SD: standard deviation; P1: 1 paradigm (stimulus intensity ranging from 20 to 40 dB SPL); P2: paradigm 2 (stimulus intensity ranging from 40 to 65 dB SPL).
Discussion
The same sequence presented in the results was maintained for the presentation of the discussion.
Cochlear activity is characterized by the absolute signal of the response, and the criterion for identification of the response depends on signal to noise ratio. Therefore, for analysis of the presence of DPOAE, it is necessary for the cochlear response to be above the noise. There is no consensus in the literature about the recommended signal/noise ratio. Some studies have used an S/N ratio greater than or equal to 3dB, such as Neely et al.15 and Tiradentes et al.16 and there are others who have used an S/N ratio greater than 6dB, such as Gorga et al.8 and Boege and Janssen.17
The equipment used in this study - the ILO V6, analyzes the presence of noise for each pair of frequencies tested considering its mean plus 1 standard deviation. Therefore, the noise level displayed on the monitor screen is higher than the average noise by 6 dB, and therefore the presence of a response can be considered when the S/N ratio is greater than or equal to zero, according to the manufacturer’s protocol.
Analyzing the results when obtaining the DPOAE emergence thresholds with the GC, the results showed a significant difference between the two criteria used (Lim0 and Lim 6) at the two frequencies (2000 Hz and 4000 Hz) in both ears, with the thresholds obtained in Lim0 smaller on average than those obtained in Lim6. According to a study by Dubno,6 significant differences were noted in the emergence thresholds of the DPOAE GC according to the criterion used at 2000 Hz and 4000 Hz. Comparing the results of the emergence thresholds according to the two criteria used (Lim0 and Lim6) significant differences at 2000 Hz were noted in both ears, as well as at 4000 Hz, also in both ears, therefore always observing the criterion used in the study should be emphasized for possible comparisons of results.
The emergence thresholds at 2000 Hz were 30 dB on average, and 35 dB on average at 4000 Hz according to the criteria used (Table 2). These thresholds were lower than those obtained by Dubno6 where the responses emerged at 50 dB at 2000 Hz and 60 dB at 4000 Hz. In this study we considered three thresholds denominated Threshold1, Threshold2 and Threshold3 where:
Threshold1 (Lim1) = lowest intensity studied where the signal/noise ratio is greater than or equal to 3dB;
Threshold2 (Lim2) = lowest intensity studied where the signal/noise ratio is greater than or equal to 3dB and 2F1-F2 with a positive value; and
Threshold3 (Lim3) = lowest intensity studied where the signal/noise ratio is greater than or equal to 3dB and subsequent intensities also present a S/N ratio greater than or equal to 3dB, with the emergence threshold considered when it appeared in 70% of cases.
Therefore, the responses were better at 2000 Hz compared to 4000 Hz such as those obtained in the aforementioned study.
When comparing the left ear with the right ear, the differences were not significant due to the number of newborns composing the sample, which was not sufficient for this purpose. These results compare to those obtained by Leme and Carvallo18 and Costa et al..19 There are studies already established in the literature that show that OAE responses are better in the right ear for the female gender compared to the left ear for the male gender, Leme and Carvallo,18 Aidan et al..20 and Durante et al.21
The results obtained in the DPOAE growth curves in this study were used to calculate the mean responses for each intensity, using 2F1-F2 and the signal to noise ratio at the frequencies of 2000 Hz and 4000 Hz, in order to obtain an average curve profile for newborns and to identify the occurrence of any compression pattern for the responses in this age group. To obtain the mean, the grouped results of the left and right ears at each intensity and frequency were used, given that there was no significant difference between them in this study.
The DPOAE GC were obtained for paradigms P1 and P2 based on the hypothesis that the slope would be higher (greater incline of the curve) in weak sounds (20 to 40dB SPL) and lower (lower incline of the curve) in midrange sounds (40 to 65dB SPL). According to Gorga et al.8 cochlear gain is higher for weaker stimulus levels, decreasing as the level increases. That is, the cochlea is more compressive when stimuli are of medium intensity. In this work, the slope values obtained at 2000 Hz were lower in P1 than P2 in both ears, and at 4000 Hz in the right ear P1 was higher than P2, and in the left ear P1 was lower than P2, as in 2000 Hz. These differences were not significant. However, visually, the slope in P1 is greater than in P2. The slope value was higher on average at 2000 Hz than 4000 Hz (except for the slope obtained in P1 in the right ear at 4000 Hz, which was greater than the corresponding slope at 2000 Hz). If the DPOAE growth curve enables detection of the presence of cochlear compression upon calculating the slope, the numerical results of this study did not make it possible to prove the presence of cochlear compression in the sample, though looking at the growth curve graphs a decrease was seen in the slope of the curves at midrange intensities.
It was not possible to make comparisons with other slope studies obtained in two paradigms, as no work using DPOAE GC in this way was found in the literature. However, a study by Almeida7 noted a change in the incline of the curve of the slope from 65 dB, which would be the compression point for the frequencies studied (the DPOAE GC was outlined with stimuli a variation range between 35 and 75 dB with 5 dB intervals at 2000 Hz and 4000 Hz). Comparing these results with the present study was not possible, because 65dB was the largest stimulus used in the paradigms presented. In studies with adults, the compression point was between 50 and 60 dB.3,15,22
When using the general paradigm the slope values were higher than those obtained in paradigms 1 and 2, but the comparison was not made because of the use of different criteria.
The slope values varied from 3 to 4 on average, reaching 15 in some cases. These maximum values can be attributed to the fact that they were obtained in the newborn population where the OAE responses are better than in other age groups. This fact was discussed with the technical team of the equipment manufacturer (Otodynamic), Peter Bray and Lee Van Middlesworth, who still have no explanation for this finding due to the small number of studies performed with the newborn population. The general paradigm was only applied to part of the sample, and only at 2000 Hz; therefore, it was not compared with the other paradigms (P1 and P2), and the General P slope values were higher than those obtained in P1 and P2. The mean General P slope value was 10.05 in the right ear and 8.85 in the left ear.
Conclusion
The emergence thresholds were on average 30 dB SPL at 2000 Hz and 35 dB SPL at 4000 Hz, therefore the best emergence thresholds were at a frequency of 2000 Hz compared to 4000 Hz.
No difference in the slope values obtained in the DPOAE growth curves were noted, but slopes were steeper at the weakest intensities (P1) and more linear at the moderate intensities (P2) suggesting that the visual analysis of the graph reflects cochlear compression.