Acessibilidade / Reportar erro

The fracture evaluation of NiTi SMA endodontics files

Abstract

The purpose of this study was to evaluate the fracture resistance and the fracture surface of NiTi SMA engine-driven endodontics files submitted to clockwise torsion. The maximum angular deflection and the maximum torque were analyzed without axial loading. The helical plastic deformations and the fracture surface morphology were evaluated by Scanning Electron Microscopic. The results showed that there was a significant statistical difference in the maximum fracture torque and no statistical difference for the angular deflection for the analyzed files. In relation to the maximum torque at the instant of the fracture, one of the brand presented better performance than the other. According to Scanning Electron Microscopic evaluation, all the files showed ductile fracture morphology.

endodontics files; NiTi; SMA; torsion fracture


The fracture evaluation of NiTi SMA endodontics files

Tibério César Uchôa MatheusI, II,* * e-mail: tcmatheus@yahoo.com.br ; Hélio Pereira LopesIV; Diana Santana de AlbuquerqueV; Carlos Nelson EliasII; Antônio Márcio Resende do CarmoVI; Jorge OtuboI; Carlos Sérgio da Costa VianaII, III

IInstituto Tecnológico de Aeronáutica – ITA, Divisão de Engenharia Mecânica Aeronáutica, 12228-900 São José dos Campos - SP, Brazil

IIInstituto Militar de Engenharia, Seção de Engenharia, Mecânica e de Materiais, RJ, Brazil

IIIUniversidade Federal Fluminense, Departamento de Engenharia Metalúrgica, RJ, Brazil

IVUniversidade Estácio de Sá, Departamento de Odontologia Restauradora, RJ, Brazil

VUniversidade de Pernambuco, Departamento de Odontologia Restauradora, PE, Brazil

VIUniversidade Federal de Juiz de Fora, Departamento de Odontologia Restauradora, MG, Brazil

ABSTRACT

The purpose of this study was to evaluate the fracture resistance and the fracture surface of NiTi SMA engine-driven endodontics files submitted to clockwise torsion. The maximum angular deflection and the maximum torque were analyzed without axial loading. The helical plastic deformations and the fracture surface morphology were evaluated by Scanning Electron Microscopic. The results showed that there was a significant statistical difference in the maximum fracture torque and no statistical difference for the angular deflection for the analyzed files. In relation to the maximum torque at the instant of the fracture, one of the brand presented better performance than the other. According to Scanning Electron Microscopic evaluation, all the files showed ductile fracture morphology.

Keywords: endodontics files, NiTi, SMA, torsion fracture

1. Introduction

Nickel-titanium (NiTi) endodontic instruments have been introduced to facilitate the instrumentation of curved canals. NiTi Shape Memory Alloys (SMA) instruments have the combination of good biocompatibility, good mechanical strength, and specific properties, such as the shape memory effect (SME) and superelasticity (SE). NiTi instrument can bend far more than stainless-steel instruments before exceeding their yield point1. This flexibility is an important property that allows the preparation of curved canals while minimizing deviation2. Despite of this increased flexibility, fracture is still a concern to NiTi instruments, and they have been reported to undergo unexpected fail3.

The behavior of NiTi alloys is associated with the stress-induced transformation of the b parent phase, austenite, with the B2 cubic crystal structure, to martensite, with a monoclinic B19’ structure. Appropriate chemical composition and thermomechanical treatments are required in order to stress-induces the martensitic transformation in NiTi alloys4.

The NiTi alloys, with an approximately equiatomic composition, have in addition to high resistance and excellent biocompatibility some special features: the SME and SE. The SME occurs in specific conditions in which the metal is deformed at a certain temperature, the shape recovery temperature is higher than the temperature of deformation. To recover the original shape it is necessary to heat the material above the shape recovery temperature, AF. At SE state as far as the shape recovery temperature is lower than the deformation temperature, the shape is recovered when the load is released without heating.

Because of its shape memory properties, Nitinol has been used for such applicatons as pipe couplings and actuators. Its superelastic properties, on the other hand, are used in eyeglass frames, mobile phone antennas, dental braces, endodontics files and endovascuar stents5.

The fatigue life of a rotary endodontic instrument could be related to the degree to which it is flexed when placed in a curved root canal, with greater flexures leading to a shorter expected life6,7. Torsional fracture occurs when the tip or some part of the endodontic file is hindered in a canal while the instrument shaft continues to rotate. In this case, the elastic limit of the metal is exceeded leading to plastic deformation followed by fracture8,9. Many factors can affect the torsional strength of endodontic files, such as taper, instrument design, alloy composition and stiffness, manufacturing, flexibility, and rotation direction10.

The objective of this study was to compare the performance of the NiTi SMA engine-driven endodontic instruments, when submitted to clockwise torsion test. The evaluated properties were: the angular deflection up to fracture and the maximum torque at the fracture. Also it was evaluated the fracture surface and the shape of spiral of the cutting edge near the immobilizing point of the endodontic instrument after the mechanical testing.

2. Experimental Procedure

The analysed nickel-titanium rotary files were K3 Endo (Sybron Endo – USA) and Profile (Dentsply-Maillefer - Switzerland) (#15 and #35, lenght of 25 mm) with cross section profiles of triple helix and triple U, respectively. Ten files of each size were submitted to the clockwise (CW) rotation. Therefore, 20 files from each manufacturer were tested.

A torsion without axial load was applied to the files with a device attached to Universal Mechanical Testing Machine Emic DL 10000 (Emic Equipments and Systems for Tests Ltd, São Paulo). This apparatus was similar to the one used by others studies11,12 (Figure 1, 2).



The load and the displacement of the thread up to the fracture of the instruments were registered continuously by a microcomputer attached to the testing machine. A chart of the load vs. displacement was obtained.

The determination of the angular deflection and the obtained displacement were converted through the formula:

The maximum torque at the fracture was calculated through the formula:

After fracture occurrence, the SEM analysis (JEOL, model ISM 5800 IV) was carried out to evaluate any possible alterations of the spirals of the cutting edge together with the point of immobilization and the surface characteristics of the fracture.

In order to observe the spiral of the cutting edge near the fracture surface, as well as the morphological characteristics of the fracture surface, micrographies were taken with varying magnification. To analyze the shape of the spiral, photomicrographies with a lower magnification (60 to 200x) was used and to analyse the morphology of the fracture surface, the photomicrographies were taken with 2000x magnification.

3. Results and Discussion

In this study, the evaluation of the angular deflection and the maximum torque were developed in agreement with the ANSI/ADA Specification number 28, 198913. The test for the evaluation of the torque shows to the dentist the maximum load that could be applied to the instrument.

The obtained charts were used to calculate the angular deflection and the maximum torque up to the fracture. The parametric Student’s t-test was applied to the obtained data with 95% of confidence interval. The results are shown in Table 1.

The maximum angular deflection of the K3 Endo and ProFile instruments, revealed that there was no significant difference between the averages of the tested groups. The same results were observed between the instruments K3 Endo #35 (Group I B) and ProFile #35 (Group II B) in relation to the maximum angular deflection. Even though, to some studies11,14, the maximum angular deflection in the fracture, for hand instruments of NiTi, has decreased with the increase of the nominal diameter of the endodontic instrument. Other author observed that the angular deflection is not influenced by the increase of the nominal diameter of the instrument15. This feature may be explained by the fact that all the efforts to standardize the dimensions of the endodontic instruments, could present discrepancy in dimensions of the instruments even with the same number. Besides, they could have deficient surface finishing presenting irregularities and scratches occurred during the manufacturing process and they act as stress concentrator interfering directly in the obtained results of such endodontic instruments when submitted to mechanical tests16.

In the evaluation of the mechanical resistance of the instruments in the torsion test, some different parameters may be used. For some authors11,17, the main parameter is the maximum angular deflection that acts as a safety factor in relation to the fracture of the instrument. Larger the angular deflection of an endodontic instrument, the larger will be its toughness and plastic deformation before starting the fracture. This feature of the material is a safety factor because as far as it goes beyond the limit of resistance to the fracture by torsion of the material18.

As to the maximum fracture torque, it was possible to verify statistical differences for the two brands tested with different diameter, increasing with the increase of the nominal diameter of the endodontic instrument. Similar results were obtained in other studies11,17,18. Concerning to brand, K3 Endo #15 (Group I A) it supported higher torque up to fracture when compared to ProFile #15 brand (Group II A). The same results was presented when it was compared the instruments K3 Endo #35 (Group I B) and ProFile #35 (Group II B).

The source of those differences is related to the variation in transversal cross section, and consequently, with the need of a higher load to fracture the instruments of a larger size. It was verified that the K3 Endo instruments supported a higher maximum torque at the instant of the fracture than the ProFile instruments. This may explain the better performance of triple helix K3 Endo file, which has a larger nucleus when submitted to a torsion effort that promotes the stress distribution progressively between the angle of the cutting edge and the canal file (Figure 3). To a transversal cross section of a triple U instrument (ProFile) (Figure 4), which has a smaller nucleus, the stress is concentrated in the canal file much closer to the center line of the instrument than in triple helix19. This explain the bettter performance of triple helix K3 Endo in relation to triple U Profile.



Qualitative evaluation by SEM showed a plastic deformation of the spirals of the instrument reversing the original direction near the fracture (±3 mm) region, as seen in Figure 5, confirming the earlier works11,21. In the same figure one can see the presence of grooves patterns at the cutting edge which comes from machining stage that can act as stress raisers during loading. The fracture surface presented microvoids and dimples characteristic of tipical ductile fracture as can be seeing in Figure 6. According to Sawaguchi et al.22, fatigue cracks always start at surface irregularities such as scratches as seeing in Figure 5 of this work. It could also start around the inclusions such as TiC and titanium oxide which come from alloy fabrication stage23,24. The fracture surface also showed shear deformation area around the ductile central fibrous region (Figure 5). Similarly, Bahia and Buono25, 2005, observed the presence of a small smooth area near the file edge, associated with nucleation and slow crack propagation.



The frequency of fracture of the NiTi SMA engine-driven endodontic instruments during the preparation of the root canal may be reduced, according to the manufacturing of the instrument with a better control of material quality, more demanding evaluation of the dimensions and a better surface finishing.

4. Conclusions

According to the results, it is conclude that:

  • In relation to the maximum angular deflection, there was a variation when comparing the brands (K

    3 Endo and ProFile) for the same diameters instruments;

  • For maximum torque there was a significant difference between the averages of the tested groups with better perfomance of K

    3 Endo brand compared to ProFile brand;

  • At lower magnification (60 to 200x), all the fractured instrument presented smooth fracture surfaces and perpendicular to their axis. There was a reversion in the direction of the spirals in relation to its original direction when the files were submitted to clockwise torsion. At the cutting edge there were observed grooves patterns from machining that can act as stress raisers during loading; and

  • At higher magnification (2.000x), it was observed that the fracture surface presented typical ductile fracture presenting dimples and microvoids.

Acknowledgments

We would like to acknowledge FAPERJ (grant #E-26/100.130/06-DSC 10) for scholarship to one of the authors and to FAPESP (grant #00/09730-1), FINEP (grant #01.04.0255.00 – CT INFRA 03/2003; PROINFRA 01/05, Protocol 153) for financial support.

Thanks to Dentsply for suppling the samples.

Received: March 19, 2007; Revised: September 25, 2007

  • 1. Walia H, Brantley WA, Gerstein, H. An initial investigation of the bending and torsional properties of Nitinol root canal files. Journal of Endodontics 1988; 14(7):346-51.
  • 2. Glosson CR. et al. A comparison of root canal preparations using NiTi hand, NiTi engine driven, and K Flex endodontic instruments. Journal of Endodontics 1995; 21(3):146-51.
  • 3. Cohen S, Burns RC. Pathways of the pulp 6th ed. St. Louis: Mosby-Year Book, Inc.; 1994. p. 206.
  • 4. Bahia MGA, Martins RC, Gonzalez BM, Buono VTL. Physical and mechanical characterization and the influence of cyclic loading on the behaviour of nickel-titanium wires employed in the manufacture of rotary endodontic instruments. International Endodontics Journal 2005; 38(11):795-801.
  • 5. Mckelvey AL, Ritchie RO. Fatigue-Crack growth behavior in the superelastic and shape-memory alloy nitinol. Metallurgical and Materials Transactions A 2001; 32A:731-43.
  • 6. Pruet J, Clement D, Carnes DJ. Cyclic fatigue testing of nickel-titanium endodontic instruments. Journal of Endodontics 1997; 27(2):77-85.
  • 7. Melo MCC, Bahia MGA, Buono VTL. Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments. Journal of Endodontics 2002; 28(11):765-9.
  • 8. Gambarine, G. Torsional and cyclic fatigue testing of ProFile NiTi rotary instruments. Journal of Evolutionary Dentistry 1999; 2(1):4-14.
  • 9. Blum JY, Machtou P, Micallef JP. Location of contact areas on rotary ProFile instruments in relationship to the forces developed during mechanical preparation on extracted teeth. International Endodontics Journal 1999; 32(2):108-114.
  • 10. Hilt BR, Cunninghan CJ, Shen C, Richards N. Torsional properties of stainless-steel and nickel-titanium files after multiple auto clave sterilizations. Journal of Endodontics 2000; 26(2):76-80.
  • 11. Seto BG, Nicholls JI, Harrington GW. Torsional properties of twisted and machined endodontic files. Journal of Endodontics 1990; 16(8):355-60.
  • 12. Lautenschlager EP. et al. Brittle and ductil torsional failures of endodontic instruments. Journal of Endodontics 1977; 3(5):175-8.
  • 13. American National Standards Institute. American Dental Association. (Council on Dental Materials, Instruments, and Equipment). Specification # 28 for root canal files and reamers, type K, and # 58 for root canal files, type H (Hedströen). Journal American Dental Association 1989; 118(2):239-40.
  • 14. Canalda-Sahli C, Brau-Aguadé E, Berastegui-Jimeno E. A comparison of bending and torsional properties of K-files manufactured with different metallic alloys. International Endodontics Journal 1996; 29(3):185-9.
  • 15. Carmo AMR. et al. Torsional fracture nickel-titanium engine-driven rotatory instruments. Revista Brasileira de Odontologia 2002; 59(3):197-99.
  • 16. Lopes HP, Elias CN. Fracture of engine-driven NiTi endodontic instruments. Theoretic and practical concepts. Revista Brasileira de Odontologia 2001; 58(3):207-10.
  • 17. Rowan MB, Nicholls JI, Steiner J. Torsional properties of stainless steel and nickel-titanium endodontic files. Journal of Endodontics 1996; 22(7):341-45.
  • 18. Lopes HP. et al. Torsional fracture of stainless-steel and nickel-titanium endodontics files. Revista Paulista de Odontologia 2001; 23(2):8-12.
  • 19. Turpin YL. et al. Impact of torsional and bending inertia on root canal instruments. Journal of Endodontics 2001; 27(5):333-36.
  • 20. Lopes HP, Elias CN, Siqueira JF. Instrumentos endodônticos. In: Lopes, HP, Siqueira JF. Endodontia: Biologia e Técnica 2Ş ed., Guanabara Koogan, Rio de Janeiro; 2004.
  • 21. Sattapan B. et al. Defects in rotary nickel-titanium files after clinical use. Journal of Endodontics 2000; 26(3):161-5.
  • 22. Sawaguchi T, Kausträter G, Alejandro Y, Wagner, M.; Eggeler, G. Crack initiation and propagation shape-memory wires in bending-rotation fatigue. Mettalurgical and Materials Transactions A 2003; 34A:2847-60.
  • 23. Otubo J, Rigo OD, Moura Neto C, Kaufman MJ, Mei PR. Scale up of NiTi shape memory alloy production by EBM. Journal of Physique IV 2003; 112:873-76.
  • 24. Otubo J, Rigo OD, Moura Neto C, Mei PR. The effects of Vaccum induction melting and electron beam melting techniques on the purity of NiTi shape memory alloys. Materials Science and Engineering A 2006; 438-40:679-82.
  • 25. Bahia MGA, Buono VTL. Decrease in the fatigue resistance of nickel-titanium rotary instruments after clinical use in curved root canals. Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontics 2005; 100(2):249-255.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      07 Feb 2008
    • Date of issue
      Dec 2007

    History

    • Received
      19 Mar 2007
    • Reviewed
      25 Sept 2007
    ABM, ABC, ABPol UFSCar - Dep. de Engenharia de Materiais, Rod. Washington Luiz, km 235, 13565-905 - São Carlos - SP- Brasil. Tel (55 16) 3351-9487 - São Carlos - SP - Brazil
    E-mail: pessan@ufscar.br