Improvement of Titanium Corrosion Resistance by Coating with Poly-Caprolactone and Poly-Caprolactone/Titanium Dioxide: Potential Application in Heart Valves

Abstract

Heart diseases affect part of world population and generally involves the functioning of valves. Valves replacement is the most common treatment and the choice between synthetic or natural/biological implants depends on several factors. Synthetic implants have greater durability, whereas biological ones are more biocompatible. This work proposes the use of polymeric coating on titanium metal surface to increase implant biocompatibility. Poly-caprolactone (PCL) has demonstrated greater efficacy for biomedical applications due to its biocompatibility. It can easily form films and coat surfaces. Titanium discs were submitted to alkaline and thermal treatments and coated with 1%PCL and 1%PCL+TiO2. For both conditions, PCL crystals were found in titanium coated surface (SEM and EDX) and X-ray diffractogram confirmed PCL coating. Infrared Spectroscopy spectra showed both PCL and TiO2 characteristic peaks. Moreover, corrosion resistance of coated disc has considerably increased, proving the effectiveness of PCL as coating material and its potential application in cardiac valves.

Keywords:
Poly-caprolactone; Titanium dioxide; Titanium; Coating; Heart Valves


1. Introduction

Heart diseases affect a large part of the world population. In Brazil, they were the second cause of death. Certain heart diseases affect the functioning of the valves and different approaches are available for treatment11 Edmunds L. Evolution of prosthetic heart valves. American Heart Journal. 2001;141(5):849-855.. The most common treatment for valve dysfunctions are based on surgical repair or replacement of the damaged valves22 Mansur Filho J. Trombose de prótese valvular e tratamento trombolítico. Revista SOCERJ. 2001;XIV(1):88-94.. The treatment choice depends on the analysis of the particularities of each patient, such as clinical, radiological and hemodynamic conditions, the evolutionary state of the disease and the dimensions of valves dysfunction11 Edmunds L. Evolution of prosthetic heart valves. American Heart Journal. 2001;141(5):849-855.,22 Mansur Filho J. Trombose de prótese valvular e tratamento trombolítico. Revista SOCERJ. 2001;XIV(1):88-94.. For cases where the most indicated treatment is replacement, artificial prostheses (mechanical valves) or biological prostheses (human or animal valves) are employed. The key of mechanical valves are their durability33 Ghanbari H, Viatge H, Kidane A, Burriesci G, Tavakoli M, Seifalian A. Polymeric heart valves: new materials, emerging hopes. Trends in Biotechnology. 2009;27(6):359-367..The main materials used in the manufacture of mechanical valves, such as those produced by BjorkShiley, LilleheiKaster, Omniscience, MedtronicHall or St.Judes, are pyrolytic carbon and titanium. These materials are more resistant, with greater durability without erosion or sharp wear, but still are susceptible to thrombogenic episodes, since these valves provide a surface on which blood clots can form easily11 Edmunds L. Evolution of prosthetic heart valves. American Heart Journal. 2001;141(5):849-855.,44 Gracioli EC, Winter P, Ziulkoski AL, Spilki F, Dullius J, Einloft S, et al. Dispositivos poliméricos cardiovasculares: comportamento termomecânico e viabilidade celular. Matéria (Rio de Janeiro). 2013;18(2):1313-1322.. Thus, a patient who has been implanted with a mechanical valve needs to be take anticoagulant medicine to prevent heart attack and stroke44 Gracioli EC, Winter P, Ziulkoski AL, Spilki F, Dullius J, Einloft S, et al. Dispositivos poliméricos cardiovasculares: comportamento termomecânico e viabilidade celular. Matéria (Rio de Janeiro). 2013;18(2):1313-1322.. To eliminate this lifelong medication need after heart valve replacement, numerous studies have been developed in order to reduce the thrombogenicity of substitute valves 55 Venkatraman S, Boey F, Lao LL. Implanted cardiovascular polymers: Natural, synthetic and bio-inspired. Progress in Polymer Science. 2008;33(9):853-874.,66 Schoen FJ. Cardiovascular pathology: maturation of a discipline and the Society for Cardiovascular Pathology. Cardiovascular Pathology. 2010;19(4):198-200.,77 Liu X, Chu PK, Ding C. Surface modification of titanium, titanium alloys, and related materials for biomedical applications. Materials Science and Engineering: R: Reports. 2004;47(3-4):49-121.,8Catauro M, Bollino F, Papale F. Surface modifications of titanium implants by coating with bioactive and biocompatible poly (ε-caprolactone)/SiO2 hybrids synthesized via sol-gel. Arabian Journal of Chemistry. 2015. In Press..

Polymers (natural and synthetic) have been shown to be potentially applicable to medical devices; however, mechanical properties of these materials might be a disadvantage for application in heart valves. An ideal heart valve substitute should be resistant, durable, hemocompatible, cost-effective and weightless 33 Ghanbari H, Viatge H, Kidane A, Burriesci G, Tavakoli M, Seifalian A. Polymeric heart valves: new materials, emerging hopes. Trends in Biotechnology. 2009;27(6):359-367.

4 Gracioli EC, Winter P, Ziulkoski AL, Spilki F, Dullius J, Einloft S, et al. Dispositivos poliméricos cardiovasculares: comportamento termomecânico e viabilidade celular. Matéria (Rio de Janeiro). 2013;18(2):1313-1322.

5 Venkatraman S, Boey F, Lao LL. Implanted cardiovascular polymers: Natural, synthetic and bio-inspired. Progress in Polymer Science. 2008;33(9):853-874.

6 Schoen FJ. Cardiovascular pathology: maturation of a discipline and the Society for Cardiovascular Pathology. Cardiovascular Pathology. 2010;19(4):198-200.

7 Liu X, Chu PK, Ding C. Surface modification of titanium, titanium alloys, and related materials for biomedical applications. Materials Science and Engineering: R: Reports. 2004;47(3-4):49-121.
-8Catauro M, Bollino F, Papale F. Surface modifications of titanium implants by coating with bioactive and biocompatible poly (ε-caprolactone)/SiO2 hybrids synthesized via sol-gel. Arabian Journal of Chemistry. 2015. In Press.. Therefore, coating of metallic prosthesis with a polymeric material may be an interesting approach to overcome the biocompatibility issue of metals for heart valves applications. In addition, incorporation of inorganic materials in the polymeric coating may also improve mechanical properties of the composites 99 Catauro M, Bollino F, Papale F, Giovanardi R, Veronesi P. Corrosion behavior and mechanical properties of bioactive sol-gel coatings on titanium implants. Materials Science and Engineering: C. 2014;43:375-382.

10 Catauro M, Bollino F, Veronesi P, Lamanna G. Influence of PCL on mechanical properties and bioactivity of ZrO2-based hybrid coatings synthesized by sol-gel dip coating technique. Materials Science and Engineering: C. 2014;39:344-351.

11 Shoja M, Shameli K, Ahmad MB, Zakaria Z. Preparation and characterization of Poly (ε- Caprolactone)/TiO2 micro-composites. Digest Journal of Nanomaterials and Biostructures. 2015;10(2):471-477.
-1212 Muñoz-Bonilla A, Cerrada ML, Fernández-García M, Kubacka A, Ferrer M, Fernández-García M. Biodegradable Polycaprolactone-Titania Nanocomposites: Preparation, Characterization and Antimicrobial Properties. International Journal of Molecular Sciences. 2013;14(5):9249-9266..

Titanium is considered the most biocompatible metal and is currently used on heart valves substitutes. Nevertheless, it is susceptible to corrosion by body fluids, which contain chloride ions and proteins. In the corrosion process, the metal is oxidized to its ionic form and the dissolved oxygen is reduced to hydroxide ions. Thus, byproducts of titanium corrosion may cause inflammatory responses in the body99 Catauro M, Bollino F, Papale F, Giovanardi R, Veronesi P. Corrosion behavior and mechanical properties of bioactive sol-gel coatings on titanium implants. Materials Science and Engineering: C. 2014;43:375-382.. Poly-caprolactone (PCL), an aliphatic polyester, is a synthetic polymer extensively used as biomaterial due to its biocompatibility, bioabsorption, low molecular weight/density and low cost8Catauro M, Bollino F, Papale F. Surface modifications of titanium implants by coating with bioactive and biocompatible poly (ε-caprolactone)/SiO2 hybrids synthesized via sol-gel. Arabian Journal of Chemistry. 2015. In Press.,1313 Smith CM, Roy TD, Bhalkikar A, Li B, Hickman JJ, Church KH. Engineering a titanium and polycaprolactone construct for a biocompatible interface between the body and artificial limb. Tissue Engineering Part A. 2010;16(2):717-724.. Titanium Oxide (TiO2) is an inert and low cost inorganic material used as an additive in cosmetics and medicines because of its biocidal properties1212 Muñoz-Bonilla A, Cerrada ML, Fernández-García M, Kubacka A, Ferrer M, Fernández-García M. Biodegradable Polycaprolactone-Titania Nanocomposites: Preparation, Characterization and Antimicrobial Properties. International Journal of Molecular Sciences. 2013;14(5):9249-9266.. Incorporation of TiO2 in a PCL matrix for further coating of titanium surface would be a great alternative to increase the compatibility/adhesion of PCL/Titanium, since a TiO2 protective layer is naturally formed in the titanium surface. Therefore, the aim of this work is to study the effects of PCL and PCL incorporated with TiO2 coatings on the corrosion resistance of titanium for potential application as cardiac valves. Structural, morphological, chemical and electrostatic characterizations of the coated discs were performed by X-ray Diffractometry (XDR), Scanning Electron Microscopy (SEM), Energy Dispersive X-ray analysis (EDX), Infrared Spectroscopy (FTIR) and Corrosion test. Moreover, PCL and PCL incorporated with TiO2 films were analyzed by Differential Scanning Calorimetry (DSC).

2. Materials and Methods

2.1 Materials

Commercially pure titanium (Ti cp) discs, grade 4, 3mm diameter and 5mm thickness and chemical composition as shown in Table 1 (ASTM F67-061414 ASTM International. ASTM F67-06 - Standard Specification for Unalloyed Titanium, for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700). West Conshohocken: ASTM International; 2006.) was supplied. Poly-caprolactone (Mn 70000-90000, density 1.145 g/mL at 25ºC) was purchased from Sigma- Aldrich (USA). Titanium dioxide particles (TiO2) was synthesized via microwave and ultrasound techniques.

Table 1
Chemical composition of Ti cp (% w/w).

2.2 Methods

2.2.1 Synthesis of TiO2

The synthesis of TiO2 was performed in two steps: Microwave and Ultrasonic Synthesis. The mixture of titanium isopropoxide (98%), citric acid and isopropanol 20% (v/v) was stirred until formation of the metal complex. Then, ethylene glycol was added to the mixture, which was subjected to microwave radiation in order to obtain a polymeric mass. The mass was calcined at 300ºC for 2h. The obtained solid was macerated and calcined again at 700ºC for 2h. The resulted powder was TiO2. Then, 0.4 g of TiO2 was dispersed in 8.6M NaOH solution, sonicated for 2h (Unique, USC 700) and centrifuged. Miliq water was used to wash the decanted solid, followed by the addition of 40mL of methanol and dried on a heating plate.

2.2.2 Alkaline and thermal treatments

Ti cp discs were wet-grounded to 180mesh sandpaper, followed by an ultrasonic bath with detergent, distilled water and isopropyl alcohol. Subsequently, samples were immersed in a 5M NaOH solution and placed in an oven at 60ºC for 24 hours. The discs were submitted to thermal treatment at 600ºC for 1 hour, and cooled to room temperature1515 ASTM International. ASTM E3-95 - Standard Practice for Preparation of Metallographic Specimens. West Conshohocken: ASTM International; 2001..

2.2.3 Characterization of treated/uncoated Ti cp discs

2.2.3.1 Microstructal analysis

The microstructural characterization of alkali and thermally treated Ti cp discs was performed by Optical Microscopy (OM) and Scanning Electron Microscopy (SEM), according to ASTM E3-951515 ASTM International. ASTM E3-95 - Standard Practice for Preparation of Metallographic Specimens. West Conshohocken: ASTM International; 2001.. OM was performed using an optical microscope coupled to the Zeiss brand image analyzer, model AxioVision 4.8.2 SP1. SEM was performed in the FEI microscope, model Inspect 550, coupled to an EDA detector of the brand EDAX, model Apollo X. For SEM analysis, samples were fixed with a carbon tape. A Kroll solution (5% nitric acid, 10% hydrofluoric acid and 85% by volume distilled water) was used to reveal the microstructure.

2.2.3.2 Hardness Testing

The hardness measurements on the Ti cp discs were made in the Durometer, Pantec brand, model RASN RS and RB using a 120º cone penetrator and a force of 150 Kgf.

2.2.4 Preparation of coated Ti cp discs

2.2.4.1 Coating procedure

Treated Ti cp discs were coated by PCL and PCL incorporated with TiO2. Firstly, the polymeric solution was prepared by dissolving PCL in chloroform in the ratio 1: 100 (w/v). This solution was split in two flasks and TiO2 was added to one of them in the ratio 10:100 (w/w; TiO2/PCL) under stirring. Ti cp discs were then immersed in the solutions for few seconds. After removal, the discs were dried at room temperature in a desiccator for 48 hours in order to obtain Ti cp+PCL and Tic p+PCL+TiO2 samples. For characterization purposes, PCL and PCL+TiO2 films were also prepared by solvent evaporation technique. The solutions were cast in a mold and dried at room temperature for 48 hours in order to obtain PCL and PCL+TiO2 samples.

2.2.5 Infrared spectroscopy

Infrared Spectroscopy analysis (FTIR) was performed on the Perkin Elmer Spectrometer Frontier FTIR using an Attenuated Total Reflection (ATR) accessory. No sample preparation was required. Ti cp disc (after alkaline and thermal treatment), PCL film, PCL+ TiO2 film, Ti cp+PCL and Ti cp+PCL+TiO2 were analyzed from 500 to 4000cm-1 in order to identify the characteristic functional groups present in each sample.

2.2.6 X-ray Diffractiometry

X-ray diffraction analysis was performed in a Rigaku Diffractometer; model Ultima IV, with Cu Kα radiation. Ti cp disc, PCL film, PCL+TiO2 film, Ti cp+PCL and Ti cp+PCL+TiO2 were analyzed in a 2θ range from 10º to 90º, with step of 0.02º per 2s/step in order to analyze the crystallinity of the samples.

2.2.7 Thermal analysis

The differential scanning calorimetry (DSC) analysis was performed on the Netzsch equipment, model DSC 200F3, to determine the melt temperature (Tm) and fusion enthalpy (ΔHm) of PCL and PCL+TiO2 films. Films were previously dried in a desiccator for 48 hours and cut into small pieces, totaling approximately 10 mg. Each sample was subjected to heating in an aluminum DSC sealed pan from 30ºC up to 120º C, at heating rate 10ºC min-1, under argon atmosphere. The degree of crystallinity of the PCL-based films was estimated by the following equation:

(1) Xc % = Δ Hm / Δ H 0 m * 100 %

Where:

Xc: crystallinity degree

∆Hm: enthalpy of fusion of the film

∆H0m: enthalpy of fusion of 100% crystalline PCL

2.2.8 Scanning electron microscopy

SEM micrographs of samples were obtained in a FEI microscope, model Inspect 550, coupled to an EDA detector of the brand EDAX, model Apollo X. Films were previously dried in a desiccator for 48 hours and superficial morphology of the samples was analyzed.

2.2.9 Corrosion testing

The open-circuit polarization corrosion (OCP) and potentiodynamic polarization tests were performed using an electrochemical cell, working electrode, platinum counter electrode and the saturated calomel electrode (SCE) and an electrolytic solution (1.5 PBS pH = 7.4 at 36.5ºC). Potentiostatic and Potentialdynamic Anodic Polarization measurement were taken in by varying the corrosion potential (E) from -1.5 V (ECS) to + 3.0 V with a 0.001 V pitch and a 0.001 V / second sweep speed, in order to obtain polarization curves for each sample.

3. Results and Discussion

Titanium discs were submitted to alkaline and thermal treatments in order to clean the surface and improve coatings adhesion1616 Jokanović V, Vilotijević M, Jokanović B, Jenko M, Anžel I, Stamenković D, et al. Investigations of corrosion on the surface of titanium substrate caused by combined alkaline and heat treatment. Corrosion Science. 2014;82:180-190.,1717 Majumdar P, Singh SB, Chakraborty M. The role of heat treatment on microstructure and mechanical properties of Ti-13Zr-13Nb alloy for biomedical load bearing applications. Journal of the Mechanical Behavior of Biomedical Materials. 2011;4(7):1132-1144.. Chemical changes and changes of Ti oxidation states may be resulted from these treatments1717 Majumdar P, Singh SB, Chakraborty M. The role of heat treatment on microstructure and mechanical properties of Ti-13Zr-13Nb alloy for biomedical load bearing applications. Journal of the Mechanical Behavior of Biomedical Materials. 2011;4(7):1132-1144.. Moreover, microstructural characterization should be performed in order to evaluate the consequences of the treatments on the phase analysis.

3.1 Characterization of treated/uncoated Ti cp discs

3.1.1 Microscopic analysis

Commercially pure titanium (Ti cp) micrographs obtained by OM and SEM are shown in Figure 1. It is possible to observe a homogeneous distribution with axial granular structure. The presence of twins, which is a type of crystalline surface defect, where a small displacement of the atoms from their regular positions occurs, is characteristic of the alpha phase of titanium1818 Gammon LM, Briggs RD, Packard JM, Batson KW, Boyer R, Domby CW. Metallography and Microstructures of Titanium and Its Alloys. ASM Handbook. 2004;9:899-917..

Figure 1
Micrograph of Ti cp after alkaline and thermal treatments and reaction with Kroll solution. (a) Micrograph obtained by Optical Microscopy; (b) Micrograph obtained by Scanning Electron Microscopy.

3.1.2 Hardness testing

Ti cp hardness 34.4 ± 1.5 were measured by the Rockwell hardness (HRC). This result is in accordance to the Literature (33.2)1818 Gammon LM, Briggs RD, Packard JM, Batson KW, Boyer R, Domby CW. Metallography and Microstructures of Titanium and Its Alloys. ASM Handbook. 2004;9:899-917.. Hardness is an important property for biomedical application of titanium, since it has to endure tensile forces, compression or shear and does not demonstrate a fragile behavior1616 Jokanović V, Vilotijević M, Jokanović B, Jenko M, Anžel I, Stamenković D, et al. Investigations of corrosion on the surface of titanium substrate caused by combined alkaline and heat treatment. Corrosion Science. 2014;82:180-190.,1717 Majumdar P, Singh SB, Chakraborty M. The role of heat treatment on microstructure and mechanical properties of Ti-13Zr-13Nb alloy for biomedical load bearing applications. Journal of the Mechanical Behavior of Biomedical Materials. 2011;4(7):1132-1144.. Moreover, durability is one of the most important parameter for heart valve application because it involves continuously operation. Tobaruela et al. (2016) investigated the variation of indentation-hardness with the fatigue-life in bioprosthetic heart valves. The authors reported a significant correlation of hardness with fatigue performance, suggesting that hardness may be a reliable indicator of mechanical performance and durability1919 Tobaruela A, Rojo FJ, García Paez JM, Bourges JY, Herrero EJ, Millán I, et al. Indentation hardness: A simple test that correlates with the dissipated-energy predictor for fatigue-life in bovine pericardium membranes for bioprosthetic heart valves. Journal of the Mechanical Behavior of Biomedical Materials. 2016;61:55-61..

3.2 Preparation of coated Ti cp discs

Coating of titanium discs with PCL and PCL incorporated with TiO2 was successful and no bubbles were observed. PCL formed a transparent film in the surface of the disc, while PCL+TiO2 formed a white film. Both films homogeneously coated the titanium surface and remained adhered to it during all characterization procedures.

3.3 Infrared analysis

The presence of PCL and TiO2 in the coated discs was evaluated by FTIR analysis. The spectra of PCL and PCL+TiO2 films, Ti cp, Ti cp+PCL, Ti cp+PCL+TiO2 are shown in Figure 2. The bands located at 762 and 1265 cm-1 are related to the vibration absorption of TiO2. The broad band in the region between 3670-3570 cm-1 refers to the stretching of the OH group. Among the characteristic vibrations of the PCL, the absorption bands 1233, 1107 and 1042 cm-1 correspond to the asymmetric COC vibration. The band at 1726 cm-1 is attributed to the stretching vibration of the carbonyl groups (C=O) present in the PCL. The absorption bands in the range of 2862 and 2942 cm-1 are due to the deformation of CH2 of the PCL interconnected to the inorganic network. The band observed at 1576 cm-1 is attributed to vibrations of hydroxyl group bound to the surficial Ti present in the oxide. The oxygen atom present in the OH group may be coordinated with several neighboring metal atoms. Thus, the superficial hydroxyl may occur in the free form (Ti-OH) linked by hydrogen to each other or linked to chemically adsorbed water molecules in the surface of TiO2. Therefore, interfacial forces between the PCL matrix and the titanium oxide network is mainly established by hydrogen bonds2020 Catauro M, Papale F, Bollino F. Characterization and biological properties of TiO2/PCL hybrid layers prepared via sol-gel dip coating for surface modification of titanium implants. Journal of Non-Crystalline Solids. 2015;415:9-15.. Catauro et al. (2015) also investigated PCL-TiO2 interactions by FTIR analysis2020 Catauro M, Papale F, Bollino F. Characterization and biological properties of TiO2/PCL hybrid layers prepared via sol-gel dip coating for surface modification of titanium implants. Journal of Non-Crystalline Solids. 2015;415:9-15.. The authors were able to confirm these interactions by the changes on the relative intensity of the bands related to the ester groups found in polymer chain2020 Catauro M, Papale F, Bollino F. Characterization and biological properties of TiO2/PCL hybrid layers prepared via sol-gel dip coating for surface modification of titanium implants. Journal of Non-Crystalline Solids. 2015;415:9-15.. According to them, the symmetrical and asymmetrical stretching of the C-O-C bonds and the O-C vibrations can be use as characteristic peaks of PCL2020 Catauro M, Papale F, Bollino F. Characterization and biological properties of TiO2/PCL hybrid layers prepared via sol-gel dip coating for surface modification of titanium implants. Journal of Non-Crystalline Solids. 2015;415:9-15..

Figure 2
Infrared spectra of (black) Ti cp, (blue) Ti cp+PCL, (red) PCL film, (pink) Ticp+PCL+TiO2, and (green) PCL+TiO2 film.

3.4 XRD analysis

XRD analysis also confirmed the coating efficiency of PCL/PCL+TiO2. Figure 3 shows the diffractograms of Ti cp, PCL and PCL+TiO2 films, and the coated discs Ti cp+PCL, Ti cp+PCL+TiO2. The characteristic peaks of the titanium α-phase can be found in the diffractograms of Ti cp, Ti cp+PCL, Ti cp+PCL+TiO2. This α-phase has a compact hexagonal structure. In addition, the diffractograms of PCL, PCL+TiO2, Ti cp+PCL and Ti cp+PCL+TiO2 indicate the presence of characteristic peaks of PCL. PCL presents an X-ray pattern with two intense reflection peaks at 2θ = 21º and 29º that can be easily identified in all diffractograms2121 Wei Z, Wang G, Wang P, Liu L, Qi M. Crystallization behavior of poly(ε-caprolactone)/TiO2 nanocomposites obtained by in situ polymerization. Polymer Engineering & Science. 2012;52(5):1047-1057.. Characteristic peaks of TiO2 depend on the crystallographic phase. XRD patterns of rutile phase exhibited strong diffraction peaks at 27º, 36º and 55º, while anatase phase patterns exhibited strong diffraction peaks at 25ºand 48º2222 Thamaphat K, Limsuwan P, Ngotawornchai B. Phase Characterization of TiO2 Powder by XRD and TEM. Kasetsart Journal (Natural Science). 2008;42:357-361.. Therefore, some of TiO2 diffraction peaks overlap PCL ones. According to Figure 3, TiO2 characteristic peaks were not identified in the Ti cp+PCL+TiO2 diffractogram. This can be attributed to the low concentration of TiO2 dispersed in the PCL matrix as well as to the overlap peaks. Nevertheless, the presence of TiO2 in the Ti cp+PCL+TiO2 diffractogram can be confirmed by the reduced the intensity of the characteristic peak of PCL at 2θ = 21º. This reduction can be due to the interactions between PCL and TiO2.

Figure 3
X-ray Diffractograms of (blue) Ti cp+PCL, (red) PCL film, (pink) Ti cp+PCL+TiO2, and (green) PCL+TiO2 film.

Shoja et al. (2006) reported that PCL characteristic peaks were no longer observed in the XRD of PCL/TiO2 microcomposites with of 15% TiO2 (w/w), indicating that the inorganic networks act by confining PCL crystallization in the films1111 Shoja M, Shameli K, Ahmad MB, Zakaria Z. Preparation and characterization of Poly (ε- Caprolactone)/TiO2 micro-composites. Digest Journal of Nanomaterials and Biostructures. 2015;10(2):471-477..

3.5 Thermal analysis

The interactions between PCL and TiO2 were also evaluated by DSC. Figure 4 shows the DSC curves obtained for PCL and PCL+TiO2 films. According to the curves, the melting temperatures are 62.6ºC and 61.8ºC for PCL and PCL+TiO2 films, respectively. Therefore, the presence of TiO2 did not significantly influence the melting temperature of the composite film. Similar behavior was reported by Wei et al. (2011)2121 Wei Z, Wang G, Wang P, Liu L, Qi M. Crystallization behavior of poly(ε-caprolactone)/TiO2 nanocomposites obtained by in situ polymerization. Polymer Engineering & Science. 2012;52(5):1047-1057.. The authors investigated the crystallization behavior of PCL/TiO2 nanocomposites obtained by in situ polymerization and suggested that the addition of TiO2 into PCL matrix does not affect the melting temperature in the thermodynamic equilibrium2121 Wei Z, Wang G, Wang P, Liu L, Qi M. Crystallization behavior of poly(ε-caprolactone)/TiO2 nanocomposites obtained by in situ polymerization. Polymer Engineering & Science. 2012;52(5):1047-1057..

Figure 4
Differential Scanning Calorimetry curves of (pink) PCL film and (blue) PCL + TiO2 film.

The experimental fusion enthalpy was obtained from the area below the peak in the DSC curve. Contrary to melting temperature, the enthalpy of fusion of PCL+TiO2 film (63.69J/g) was significantly reduced compared to those found for the PCL film (76.27J/g). By equation 1 and using the enthalpy of fusion of 100% crystalline PCL as 135 J/g, it was possible to estimate the degree of crystallinity of the films2323 Borghesi DC, Molina MF, Guerra MA, Campos MGN. Biodegradation Study of a Novel Poly-Caprolactone-Coffee Husk Composite Film. Materials Research. 2016;19(4):752-758.. PCL film showed higher degree of crystallinity (56.5%) than PCL+TiO2 film (47.2%), indicating that TiO2 limited PCL crystallization in the films, as well as observed by XDR analysis. Wei et al. (2011) reported that TiO2 had a great heterogeneous nucleation effect on the crystallization of PCL matrix2121 Wei Z, Wang G, Wang P, Liu L, Qi M. Crystallization behavior of poly(ε-caprolactone)/TiO2 nanocomposites obtained by in situ polymerization. Polymer Engineering & Science. 2012;52(5):1047-1057.. They suggest that TiO2 nanoparticles added to the PCL matrix play two competing roles on PCL crystallization: particles act as heterogeneous nuclei and promote the crystallization process of PCL; and they physically prevent polymer chain mobility and retard PCL crystals growth2121 Wei Z, Wang G, Wang P, Liu L, Qi M. Crystallization behavior of poly(ε-caprolactone)/TiO2 nanocomposites obtained by in situ polymerization. Polymer Engineering & Science. 2012;52(5):1047-1057.. Thus, the reduced crystallinity degree obtained for the PCL+TiO2 film may be due to the higher contribution of TiO2 in reducing PCL mobility and the crystals growth rate.

Chin-San (2004) investigated the in situ polymerization of titanium isopropoxide in poly-caprolactone2424 Wu CS. In situ polymerization of titanium isopropoxide in polycaprolactone: Properties and characterization of the hybrid nanocomposites. Journal of Applied Polymer Science. 2004;92(3):1749-1757.. According to this author, TiO2 influences the strength of interfacial bonding between the polymer chains and affects crystallization of PCL2424 Wu CS. In situ polymerization of titanium isopropoxide in polycaprolactone: Properties and characterization of the hybrid nanocomposites. Journal of Applied Polymer Science. 2004;92(3):1749-1757..

3.6 Scanning electron microscopy

The morphology of PCL crystals was evaluated by SEM. SEM micrographs of PCL and PCL+TiO2 films are shown in Figures 5a and 5b, respectively. Spherulites can be observed in Figure 5a. On the other hand, a large number, but smaller, PCL spherulitus was found in PCL+TiO2 film micrograph (Figure 5b). Thus, the nucleating effect of TiO2 on PCL crystallization can be confirmed by the increase of crystals amount and the reduction on the spherulites diameter. Similar EDS spectra was found for PCL and PCL+TiO2 films, identifying the main chemical elements found in PCL (carbon and oxygen). Titanium was not present in the PCL+TiO2 film spectrum due to the lower concentration of this element in the film composition. The EDS spectra of Ti cp+PCL and Ti cp+ PCL+TiO2 (Figures 5c and 5d) indicate the presence of C and O species in both discs, as characteristics elements of PCL. The presence of Na peaks in these spectra is due to NaOH used in the alkaline treatment of the discs. The darkest background region observed in Figures 4c and 4d corresponds to the Ti matrix. In both micrographs, PCL is uniformly coating all the extension of the discs. Moreover, PCL crystals clusters can be found along the discs.

Figure 5
Micrographs obtained by Scanning Electron Microscopy and EDS spectra obtained at the point A: (a) PCL film, (b) PCL+TiO2 film, (c) Ti cp+PCL, and (d) Ti cp+ PCL+TiO2.

3.7 Corrosion testing

In order to evaluate the efficiency of PCL and PCL+TiO2 coatings on Ti cp discs, corrosion test was performed at physiological conditions (pH 7.4 and 36.5ºC). Corrosion of implants can alter the pH locally, and the release of metallic ions may affect surround cells metabolism.

The electrochemical parameters of the corrosion test (Ecorr - corrosion potential, Icorr - corrosion current and Ipass - passivation current) for PCL and PCL+TiO2 films, Ti cp, Ti cp+PCL, Ti cp+PCL+TiO2 samples are shown in Table 2. In addition, Figure 6 shows the potentiodynamic polarization curves for all tested samples. Ti cp presents a clear region of passivation, with the formation of the passive layer (Ipass = 1.97x10-5A/cm2), which breaks at a potential around 1.5 V2525 Mariano NA, Oliveira RG, Fernandes MA, Rigo ECS. Corrosion behavior of pure titanium in artificial saliva solution. Matéria (Rio de Janeiro). 2009;14(2):878-880.. However, this behavior does not occur to the coated samples Ti cp+PCL and Ti cp+PCL+TiO2 (no rupture of the film was observed during the test).

Table 2
Corrosion Test results at physiological conditions.

Figure 6
Polarization curves obtained at physiological condition: (black) Ti cp, (red) Ti cp+PCL, (blue) Ti cp+PCL+TiO2, (green) PCL film, and (pink) PCL+TiO2 film.

As it can be seen in the polarization curves, both coated samples (Ti cp+PCL and Ti cp+PCL+TiO2) presented similar behavior. Thus, the presence of TiO2 did not alter corrosion resistance. The passivation current density in these coated samples were similar across a wide range of potential, indicating the formation of a compact passive film (possibly titanium oxide) on the surface of Ti cp. Moreover, the higher passivation current density of the coated samples indicates that the passivation happens first in the Ti cp+PCL and Ti cp+PCL+TiO2 than in the Ti cp, protecting the metal from corrosion2525 Mariano NA, Oliveira RG, Fernandes MA, Rigo ECS. Corrosion behavior of pure titanium in artificial saliva solution. Matéria (Rio de Janeiro). 2009;14(2):878-880.,2626 Mariano NA, Oliveira RG, Braga EI, Rigo ECS. Corrosion Characterization of Titanium Alloys by Electrochemical Techniques in Artificial Saliva and SBF Solution. Key Engineering Materials. 2009;396-398:315-318..

Corrosion potential (Ecorr) of Ti cp+PCL and Ti cp+PCL+TiO2 are similar, but higher than Ti cp one. This indicates that coated discs are more stable and that the coatings slow the onset of the corrosive process2525 Mariano NA, Oliveira RG, Fernandes MA, Rigo ECS. Corrosion behavior of pure titanium in artificial saliva solution. Matéria (Rio de Janeiro). 2009;14(2):878-880.. For heart valves application, corrosion stability at physiological condition plays an important role on the implant durability. Therefore, the use of PCL and PCL+TiO2 may protect the implant against body fluids attack and improve its biocompatibility2727 Lam MT, Wu JC. Biomaterial applications in cardiovascular tissue repair and regeneration. Expert Review of Cardiovascular Therapy. 2012;10(8):1039-1049.. Nevertheless, biological tests of coating materials are necessary to subsidize biocompatibility aspects2828 De Viteri VS, Fuentes E. Titanium and Titanium Alloys as Biomaterials. In: Gegner J, ed. Tribology - Fundamentals and Advancements. Rijeka: InTech; 2013. Available from: https://www.intechopen.com/books/tribology-fundamentals-and-advancements/titanium-and-titanium-alloys-as-biomaterials. Access in: 19/12/2017.
https://www.intechopen.com/books/tribolo...
.

4. Conclusions

We successfully coated titanium discs with poly-caprolactone and poly-caprolactone incorporated with titanium dioxide. The coated discs were more resistant to corrosion at physiological conditions. Therefore, poly-caprolactone and poly-caprolactone incorporated with titanium dioxide are potential surface-coating of titanium for heart valve applications.

5. Acknowledgements

The authors thank Brazilian research funding agencies FAPEMIG, CNPq and CAPES for financial.

6. References

  • 1
    Edmunds L. Evolution of prosthetic heart valves. American Heart Journal 2001;141(5):849-855.
  • 2
    Mansur Filho J. Trombose de prótese valvular e tratamento trombolítico. Revista SOCERJ 2001;XIV(1):88-94.
  • 3
    Ghanbari H, Viatge H, Kidane A, Burriesci G, Tavakoli M, Seifalian A. Polymeric heart valves: new materials, emerging hopes. Trends in Biotechnology 2009;27(6):359-367.
  • 4
    Gracioli EC, Winter P, Ziulkoski AL, Spilki F, Dullius J, Einloft S, et al. Dispositivos poliméricos cardiovasculares: comportamento termomecânico e viabilidade celular. Matéria (Rio de Janeiro) 2013;18(2):1313-1322.
  • 5
    Venkatraman S, Boey F, Lao LL. Implanted cardiovascular polymers: Natural, synthetic and bio-inspired. Progress in Polymer Science 2008;33(9):853-874.
  • 6
    Schoen FJ. Cardiovascular pathology: maturation of a discipline and the Society for Cardiovascular Pathology. Cardiovascular Pathology 2010;19(4):198-200.
  • 7
    Liu X, Chu PK, Ding C. Surface modification of titanium, titanium alloys, and related materials for biomedical applications. Materials Science and Engineering: R: Reports 2004;47(3-4):49-121.
  • Catauro M, Bollino F, Papale F. Surface modifications of titanium implants by coating with bioactive and biocompatible poly (ε-caprolactone)/SiO2 hybrids synthesized via sol-gel. Arabian Journal of Chemistry 2015. In Press.
  • 9
    Catauro M, Bollino F, Papale F, Giovanardi R, Veronesi P. Corrosion behavior and mechanical properties of bioactive sol-gel coatings on titanium implants. Materials Science and Engineering: C. 2014;43:375-382.
  • 10
    Catauro M, Bollino F, Veronesi P, Lamanna G. Influence of PCL on mechanical properties and bioactivity of ZrO2-based hybrid coatings synthesized by sol-gel dip coating technique. Materials Science and Engineering: C 2014;39:344-351.
  • 11
    Shoja M, Shameli K, Ahmad MB, Zakaria Z. Preparation and characterization of Poly (ε- Caprolactone)/TiO2 micro-composites. Digest Journal of Nanomaterials and Biostructures 2015;10(2):471-477.
  • 12
    Muñoz-Bonilla A, Cerrada ML, Fernández-García M, Kubacka A, Ferrer M, Fernández-García M. Biodegradable Polycaprolactone-Titania Nanocomposites: Preparation, Characterization and Antimicrobial Properties. International Journal of Molecular Sciences 2013;14(5):9249-9266.
  • 13
    Smith CM, Roy TD, Bhalkikar A, Li B, Hickman JJ, Church KH. Engineering a titanium and polycaprolactone construct for a biocompatible interface between the body and artificial limb. Tissue Engineering Part A 2010;16(2):717-724.
  • 14
    ASTM International. ASTM F67-06 - Standard Specification for Unalloyed Titanium, for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700) West Conshohocken: ASTM International; 2006.
  • 15
    ASTM International. ASTM E3-95 - Standard Practice for Preparation of Metallographic Specimens West Conshohocken: ASTM International; 2001.
  • 16
    Jokanović V, Vilotijević M, Jokanović B, Jenko M, Anžel I, Stamenković D, et al. Investigations of corrosion on the surface of titanium substrate caused by combined alkaline and heat treatment. Corrosion Science 2014;82:180-190.
  • 17
    Majumdar P, Singh SB, Chakraborty M. The role of heat treatment on microstructure and mechanical properties of Ti-13Zr-13Nb alloy for biomedical load bearing applications. Journal of the Mechanical Behavior of Biomedical Materials 2011;4(7):1132-1144.
  • 18
    Gammon LM, Briggs RD, Packard JM, Batson KW, Boyer R, Domby CW. Metallography and Microstructures of Titanium and Its Alloys. ASM Handbook 2004;9:899-917.
  • 19
    Tobaruela A, Rojo FJ, García Paez JM, Bourges JY, Herrero EJ, Millán I, et al. Indentation hardness: A simple test that correlates with the dissipated-energy predictor for fatigue-life in bovine pericardium membranes for bioprosthetic heart valves. Journal of the Mechanical Behavior of Biomedical Materials 2016;61:55-61.
  • 20
    Catauro M, Papale F, Bollino F. Characterization and biological properties of TiO2/PCL hybrid layers prepared via sol-gel dip coating for surface modification of titanium implants. Journal of Non-Crystalline Solids 2015;415:9-15.
  • 21
    Wei Z, Wang G, Wang P, Liu L, Qi M. Crystallization behavior of poly(ε-caprolactone)/TiO2 nanocomposites obtained by in situ polymerization. Polymer Engineering & Science 2012;52(5):1047-1057.
  • 22
    Thamaphat K, Limsuwan P, Ngotawornchai B. Phase Characterization of TiO2 Powder by XRD and TEM. Kasetsart Journal (Natural Science) 2008;42:357-361.
  • 23
    Borghesi DC, Molina MF, Guerra MA, Campos MGN. Biodegradation Study of a Novel Poly-Caprolactone-Coffee Husk Composite Film. Materials Research 2016;19(4):752-758.
  • 24
    Wu CS. In situ polymerization of titanium isopropoxide in polycaprolactone: Properties and characterization of the hybrid nanocomposites. Journal of Applied Polymer Science 2004;92(3):1749-1757.
  • 25
    Mariano NA, Oliveira RG, Fernandes MA, Rigo ECS. Corrosion behavior of pure titanium in artificial saliva solution. Matéria (Rio de Janeiro) 2009;14(2):878-880.
  • 26
    Mariano NA, Oliveira RG, Braga EI, Rigo ECS. Corrosion Characterization of Titanium Alloys by Electrochemical Techniques in Artificial Saliva and SBF Solution. Key Engineering Materials 2009;396-398:315-318.
  • 27
    Lam MT, Wu JC. Biomaterial applications in cardiovascular tissue repair and regeneration. Expert Review of Cardiovascular Therapy 2012;10(8):1039-1049.
  • 28
    De Viteri VS, Fuentes E. Titanium and Titanium Alloys as Biomaterials. In: Gegner J, ed. Tribology - Fundamentals and Advancements Rijeka: InTech; 2013. Available from: https://www.intechopen.com/books/tribology-fundamentals-and-advancements/titanium-and-titanium-alloys-as-biomaterials Access in: 19/12/2017.
    » https://www.intechopen.com/books/tribology-fundamentals-and-advancements/titanium-and-titanium-alloys-as-biomaterials

Publication Dates

  • Publication in this collection
    22 Jan 2018
  • Date of issue
    2017

History

  • Received
    29 Apr 2017
  • Reviewed
    28 Oct 2017
  • Accepted
    07 Dec 2017
ABM, ABC, ABPol UFSCar - Dep. de Engenharia de Materiais, Rod. Washington Luiz, km 235, 13565-905 - São Carlos - SP- Brasil. Tel (55 16) 3351-9487 - São Carlos - SP - Brazil
E-mail: pessan@ufscar.br