Study about the consequences of different sound pressure levels emitted by magnetic resonance imaging equipment

ABSTRACT

Purpose:

to measure the sound pressure level emitted by the magnetic resonance imaging in operation and check the auditory sensations described by individuals submitted to this examination after it has been carried out.

Methods:

the measurement of the deep noise was conducted during image formation in this examination, by means of sound pressure level meter (decibel meter) and the application of a questionnaire for the purpose of checking the hearing sensations presented by those individuals after the exam has been carried out. This procedure was applied to sixty individuals submitted to a magnetic resonance imaging of the skull.

Results:

it was found that the average of sound pressure level presented while the examination was carried out was lower than the level of intensity considered being a risk for auditory damages. However sensations like tinnitus, anxiety, auricular plenitude and relief after the end of the examination were mentioned by the individuals after it had been carried out.

Conclusion:

the use of hearing protectors is relevant as a preventive measure for controlling the level of noise that affects the individual's cochlea, thus minimizing the physical and psycho-social effects resulting from noise exposure.

Keywords:
Noise; Magnetic Resonance Spectroscopy; Protective Devices; Hearing Loss; Hearing

RESUMO

Objetivo:

mensurar o nível de pressão sonora emitido pelo aparelho de ressonância magnética em funcionamento e verificar as sensações auditivas descritas pelos indivíduos submetidos a este exame após a realização do mesmo.

Métodos:

foi realizada a mensuração do ruído produzido, durante a aquisição das imagens, por meio de um medidor de nível de pressão sonora (decibelímetro) e a aplicação de um questionário com a finalidade de verificar as sensações auditivas apresentadas por estes indivíduos após a realização do exame. Tal procedimento foi aplicado a 60 indivíduos submetidos ao exame de ressonância magnética do crânio.

Resultados:

o nível de pressão sonora média apresentada durante a realização do exame foi inferior ao nível de intensidade considerado como risco para lesões auditivas. No entanto, sensações como zumbido, ansiedade, plenitude auricular e alívio após o término do exame foram citados pelos indivíduos após a execução do mesmo.

Conclusão:

é pertinente o uso de protetores auriculares como medida preventiva para o controle do nível de ruído que atinge a cóclea do indivíduo, minimizando desta forma os efeitos físicos e psicossociais decorrentes da exposição ao ruído.

Descritores:
Ruído; Espectroscopia de Ressonância Magnética; Equipamentos de Proteção; Perda Auditiva; Audição

Introduction

The magnetic resonance imaging (MRI) can cause various biological effects when the patient subjected to its execution has been exposed to high sound pressure levels. Several individuals reported feeling headaches, chest pain, fatigue, dizziness, loss of appetite, insomnia and rash. These effects can be justified by the acoustic noise produced during magnetic resonance imaging11. Lufkin RB. Manual de Ressonância Magnética. 2 ed. Rio de Janeiro: Guanabara Koogan; 1999..

The magnetic resonance imaging is a test that uses imaginologic electromagnetic field in imaging, emitting non-ionizing radiation22. Koch HA, Ribeiro ECO, Tonomura ET. Radiologia na formação do médico geral. In: Koch HA, Ribeiro ECO, Tonomura ET. Ressonância magnética. Rio de Janeiro: Revinter; 1997. p. 43-50.),(33. Willian RS. Sistemas de imagens digitais com aplicabilidade na Odontologia [monografia]. Porto Alegre (RS): Universidade Federal do Rio Grande do Sul. Faculdade de Odontologia, Curso de Especialização em Radiologia Odontológica e Imaginologia; 2011.. The received signals are converted into numerical values and reconstructed by computers in axial segmental images, plus coronal, sagittal and oblique views22. Koch HA, Ribeiro ECO, Tonomura ET. Radiologia na formação do médico geral. In: Koch HA, Ribeiro ECO, Tonomura ET. Ressonância magnética. Rio de Janeiro: Revinter; 1997. p. 43-50..

The magnetic resonance imaging equipment consists a magnet system (magnets), coils (transmitter and receiver antennas of radio frequency) and a set of computers. The examination is harmless and painless. For forming the images the patient settles within a large magnet that generates the main magnetic field. Other additional magnets are used to generate secondary magnetic fields that will guide the location and thickness of the segment to be searched. The coil system emits radiofrequency waves (megahertz or MHz) that will excite the tissue hydrogen protons of the segment and, when stopped the stimulation, radiofrequency waves are released, which will be captured by receivers and processed by the computer, thus forming the image22. Koch HA, Ribeiro ECO, Tonomura ET. Radiologia na formação do médico geral. In: Koch HA, Ribeiro ECO, Tonomura ET. Ressonância magnética. Rio de Janeiro: Revinter; 1997. p. 43-50.),(44. Mazzola AA. Ressonância Magnética: princípios de formação da imagem e aplicações em imagem funcional. Rev Bras Fis Med. 2003;3(1):117-29.),(55. McMahon KL, Cowin G, Galloway G. Magnetic resonance imaging: the underlying principles. J Orthop Sports PhysTher. 2011;41(11):806-19..

There are two types of images processed during the examination. Both are called weighted images. The first is T1, called spin-lattice relaxation time, which is more used for the evaluation of regular structures. The second is T2, or spin-spin relaxation used to evaluate pathological conditions like inflammation, cysts or neoplasm66. Foltz WD, Jaffray DA. Principles of Magnetic Resonance Imaging. Radiat Res. 2012;177(4):331-48.. During the formation of the weighted images there is noise generation, and this presents variation in the sound pressure level during the realization of examination22. Koch HA, Ribeiro ECO, Tonomura ET. Radiologia na formação do médico geral. In: Koch HA, Ribeiro ECO, Tonomura ET. Ressonância magnética. Rio de Janeiro: Revinter; 1997. p. 43-50..

Noise can be classified according to the sound pressure level (SPL) presented, and the decibel (dB), the logarithmic unit normally used. SPL is characterized by means of subjective levels. Thus, noise comprises the intensity range from 0 to 19 dB SPL ranks the very quiet level. From 20 to 39 dB SPL intensity ranks the silent level. The quiet intensity level comprises a range between 40 and 59 dB SPL. The noise classified as moderate level is with SPL ranging from 60 to 79 dB. The subjective noise classified as noisy range from 80 to 99 dB SPL. The very loud level of classification is between 100 and 129 dB SPL. The auditory threshold of discomfort is located at this level, approximately, from 120 dB SPL of intensity. The last level is classified as thunderous and covers intensity levels ranging from 130 to 150 dB SPL. It is the threshold of hearing pain. Noise in these intensities are considered intolerable and through a simple exposure it can cause irreversible hearing loss (HL)77. Russo PCI. Noções gerais de acústica e psicoacústica. In: Nudelmann AA, Costa AE, Seligman J, Ibañes NR, editores. PAIR: Perda auditiva induzida pelo ruído. Rio de Janeiro: Revinter; 2001. p. 134-40.),(88. Russo PCI. Bases Físicas da Fonação. In: Russo PCI, editor. Acústica e psicoacústica aplicadas à fonoaudiologia. São Paulo: Lovise; 1999. p. 143-78..

However, the harmful effects of excessive audible noise are not just limited to the auditory organ and it can cause changes in the organic homeostasis, acting through the autonomic nervous system with consequences for organs and systems controlled by the central nervous system, such as cardiovascular and endocrine99. Almeida SIC. Efeitos lesivos do ruído e implicações legais. Acta Awho. 1998; 12(3):121-7.. According to some authors, short periods of noise exposure to sound pressure levels between 90 and 100 dB in normotensive and hypertensive patients, leads to increased blood pressure, on average, 3.3% to 10 7%1010. Kwitko A, Pezzi RG, Silveira MS. Exposição a ruído ocupacional e pressão sangüínea. Rev Bras Otorrinolaringol. 1996;62(2):89-98.. At the hearing, noise can cause a decrease of the thresholds at high frequencies and distortion of acoustic signals of speech and can damage a large extent the auditory pathways from the eardrum to regions of the central nervous system. In the organ of Corti the main changes occur which are responsible for hearing loss induced by noise because their outer hair cells are extremely sensitive to high and prolonged sound pressure, with subsequent enzymatic and energy reduction, as well as decreased oxygen and nutrients. With cell death the space is filled by scar formation, causing the permanent loss of hearing ability1111. Oliveira JAA. Prevenção e proteção contra perda auditiva induzida por ruído. In: Nudelmann AA, Costa EA, Seligman J, Ibanes RN, organizadores. PAIR Perda Auditiva Induzida por Ruído. Rio de Janeiro: Revinter ;2001. p.17-44.),(1212. Araujo SA. Perda auditiva induzida pelo ruído em trabalhadores de metalúrgica. Rev Bras Otorrinolaringol. 2002;68(1):47-52..

Given the proven organic noise changes on the individual's body, the realization of this study aimed to measure the sound pressure level emitted by the MRI machine for examination of the skull and verify the auditory sensations described by patients after completion of the examination, since it is not common the hearing protection in Diagnostic Imaging centers.

Methods

This study was approved by the Ethics and Research Committee of the Pontifical Catholic University of Minas Gerais (CEP: 142). This study was conducted with 60 individuals (30 women and 30 men), aged between 32-65 years old, who underwent the examination of magnetic resonance imaging of the skull in the Diagnostic Imaging Center of Mother Teresa Hospital.

The equipment used to evaluate the sound pressure level was the digital decibel meter of ICEL Manaus brand, DL-4050 model. The MRI machine where the skull examinations were performed was the Siemens brand Magneton Vision Plus model, which operates at 1.5 Tesla. To complement the data we used a questionnaire to obtain information on the hearing of patients before and after the examination.

The measurement of sound pressure level (SPL) of the noise was performed with the decibel meter positioned 60 cm away from the MRI equipment.

The survey was conducted with the decibel meter positioned with the decibel selector switch in the "C" position. This position allows a curve of flat response being recommended for the measurement of generated noise by a determined equipment or machine. The intensity scale used was the "Hi" ranging from 65 to 130 dB with the response time switch in the FAST option "F". This option allows a 200ms response time, simulating the response time of the human ear.

For each skull MRI performed were measured 20 intensities in SPL, divided into four moments of the examination. The first moment is the examination stage where there is no image production. The second moment is the stage where there is production of weighted images T2. The third moment is the stage of the examination in which the instrument calibration occurs. The fourth moment is the examination stage where there is production of weighted images T1. For every moment were recorded 05 SPL. Each examination had an average duration of thirty-two minutes.

For analysis of the results, it was obtained the average of each measured moment, subsequently generating the achievement of the global average of the examination. This procedure was performed for each of the four moments, obtaining as the final result four global values (no image GV, T2 GV, calibration GV, T1 GV). The global values were submitted to an average, achieving the final result in SPL. The results were tabulated and submitted to ANOVA statistic test with 5% significance level. The application of the questionnaire to the patient happened after the examination. The results were obtained through a percentage analysis.

Results

Table 1 shows the global values in SPL obtained in each moment. It is worth mentioning that the values obtained for the no image moment are steadily on the examination. The lowest global value in SPL is represented by the without image moment (63.27 dB SPL) and the highest value in SPL is represented by the calibration moment (77.13 dB SPL). It was found also that the no image moment showed a lower statistically significantly difference (p < 0.05) when compared with other moments. And the remaining moments do not differ from each other (p > 0.05) (Table 1).

Table 1:
Average of global values (Sound Pressure Level), the lowest absolute value and the highest absolute value in the four examination moments by magnetic resonance imaging of the skull, expressed in Decibels - Sound Pressure Level

In relation to the applied questionnaire, it has been verified that all the analyzed individuals (100%) reported no hearing impairment before the examination. Most patients (52 patients - 86.7%) reported feeling of auricular plenitude, followed by symptom relief after the completion of the examination. Forty-eight individuals (80%) reported feeling anxiety during the examination and twenty-eight individuals (46.7%) reported tinnitus during the examination.

Discussion

Considering that noise is a mixture of non-harmonics sounds, which can cause besides an unpleasant sensation, damage a person's health, especially hearing aid, it has become big the concern with regard to exposure to high SPL, especially the occupational77. Russo PCI. Noções gerais de acústica e psicoacústica. In: Nudelmann AA, Costa AE, Seligman J, Ibañes NR, editores. PAIR: Perda auditiva induzida pelo ruído. Rio de Janeiro: Revinter; 2001. p. 134-40.),(88. Russo PCI. Bases Físicas da Fonação. In: Russo PCI, editor. Acústica e psicoacústica aplicadas à fonoaudiologia. São Paulo: Lovise; 1999. p. 143-78.),(1313. Souza MA, Pennacchi SMPL, Silva BCP, Ferreira AJV. Prevalência do zumbido como sintoma da perda auditiva induzida por ruído ocupacional. Rev CEFAC. 2002; 4(3):195-8.. However, despite advances in knowledge about the harmful noise effects, to be the agent commonly found on the workplace and currently be present in the environment of large cities, investments in its control are scarce and localized77. Russo PCI. Noções gerais de acústica e psicoacústica. In: Nudelmann AA, Costa AE, Seligman J, Ibañes NR, editores. PAIR: Perda auditiva induzida pelo ruído. Rio de Janeiro: Revinter; 2001. p. 134-40.),(1414. Basso BV, Campos LA, Thiesen J. Estudo da alteração temporária do limiar auditivo em trabalhadores expostos ao ruído. Rev CEFAC. 2003;5(2):187-91.)-(1616. Pires MGS, Rodrigues PH, Sampaio CCC, Rodrigues CG. Medida do nível de pressão sonora em um centro urbano. J Bras Fonoaudiol. 2002;3(13):263-6..

It is known that the noise effects do not have relation only to their physical characteristics and the exposure time. Other physical changes such as fatigue, interference in communication, tinnitus, nervousness or irritation, attention deficit, headaches and changes in the cardiovascular system may be submitted, taking into account individual factors (individual susceptibility to noise, psychological factors, resonance of external auditory canal and general health) and external factors (vibration, environmental temperature, exposure to chemicals and medicinal drugs) (1313. Souza MA, Pennacchi SMPL, Silva BCP, Ferreira AJV. Prevalência do zumbido como sintoma da perda auditiva induzida por ruído ocupacional. Rev CEFAC. 2002; 4(3):195-8.),(1717. Marchiori MLL, Melo JJ. Comparação das queixas auditivas com relação à exposição ao ruído em componentes de orquestra sinfônica. Pró-fono R Atual Cient. 2001;13(1):9-12.)-(1919. Dreossi FCR, Santos MMT. Investigação sobre o efeito do ruído na inteligibilidade de fala de crianças da quarta série do ensino fundamental. Rev Soc Bras Fonoaudiol. 2004;9(3):136-43.).

Among the above mentioned changes highlight the tinnitus as one of the indications that the individual has been exposed to an overly strong sound stimulus. In some cases it may present as a symptom of a temporary hearing loss1313. Souza MA, Pennacchi SMPL, Silva BCP, Ferreira AJV. Prevalência do zumbido como sintoma da perda auditiva induzida por ruído ocupacional. Rev CEFAC. 2002; 4(3):195-8.),(1717. Marchiori MLL, Melo JJ. Comparação das queixas auditivas com relação à exposição ao ruído em componentes de orquestra sinfônica. Pró-fono R Atual Cient. 2001;13(1):9-12.),(1818. Nudelmann AA, Costa AE, Seligman J, Ibañez NR. PAIR- Perda auditiva induzida pelo ruído. 2 ed. Rio de Janeiro: Revinter; 2001.. In this study it was found 46.7% of this symptomatology, saying this mentioned condition.

Thus, exposure to non-occupational noise has become an object of research, since the increase of its noise levels has been considerable in the last years77. Russo PCI. Noções gerais de acústica e psicoacústica. In: Nudelmann AA, Costa AE, Seligman J, Ibañes NR, editores. PAIR: Perda auditiva induzida pelo ruído. Rio de Janeiro: Revinter; 2001. p. 134-40.),(2020. Vansin R, Ferreira AJ. Avaliação da alteração temporária do limiar auditivo em jovens frequentadores de casas noturnas. Rev CEFAC 2002;(4):219-22.. So this study aims to increase interest about the need for preventive action in relation to non-occupational noise, in this case the noise made by the MRI scanner during the examination.

The NR-15 norm of the 3,214 / 78 decree of Brazilian law defines the limits of daily maximum exposures to certain noise levels. For 8 hours of exposure the allowed sound level is 85 dB, 4 hours of exposure (90 dB), 2 hours of exposure (95 dB), 1 hour exposure (100 dB), 1/2 hour of exposure (105 dB), 1/4 hour of exposure (110 dB). This determination is used to occupational noise controls77. Russo PCI. Noções gerais de acústica e psicoacústica. In: Nudelmann AA, Costa AE, Seligman J, Ibañes NR, editores. PAIR: Perda auditiva induzida pelo ruído. Rio de Janeiro: Revinter; 2001. p. 134-40.),(1717. Marchiori MLL, Melo JJ. Comparação das queixas auditivas com relação à exposição ao ruído em componentes de orquestra sinfônica. Pró-fono R Atual Cient. 2001;13(1):9-12.),(2020. Vansin R, Ferreira AJ. Avaliação da alteração temporária do limiar auditivo em jovens frequentadores de casas noturnas. Rev CEFAC 2002;(4):219-22..

The comparison of the final result of this study (72.91 dB SPL), found in the measurement presented during the magnetic resonance imaging of the skull, with the subjective level of classification of noise by the average intensity values ​​in dB classifies the noise present during the execution of examination at a moderate level. This same classification was found when considering the SPL expressed by the global value of each moment presented in Table 1.

The lowest and highest absolute values ​​found for the without image moment (57.2 dB to 70.9 dB) can be classified as moderate noise level; T2 moment presented peak of noise and the level of which was classified as noisy (91.7 dB SPL), overtaking the danger zone for hearing harmfulness; the calibration moment presented level rotation varying from moderate level (64.4 dB SPL) to the noisy level (83.9 dB SPL). But the T1 moment had presence of the lowest absolute value classifying the subjective level of noise as quiet (59.9 dB SPL), varying with the highest absolute value for the noisy level (82.4 dB SPL).

Considering the final result found (72.91 dB SPL), the measured noise presented minimal risk for hearing loss induced by high levels of sound pressure, as this is presented below, both the amount considered risk (85 dB SPL) and the exposure time to hearing damage. However, despite the low risk of irreversible hearing damage, the noise exposure control requires the use of hearing protection with restriction of choice, since this protection must not contain any induced ferric metal in its composition1515. Manzoni TCRC. Saúde auditiva no trabalho: Um estudo sobre a perda auditiva ocupacional e sua prevenção. Rev CEFAC. 2000;2(2):7-14.),(2121. Avagliano A, Almeida K. Estudo do desempenho de diferentes tipos de protetores auditivos. Rev CEFAC. 2001;3(1):77-87.)-(2323. Ferreira PA, Peloggia SCC, Pirinoto FH, Bernardi APA. Influência da conservação do protetor auricular na prevenção da perda auditiva induzida por níveis de pressão sonora elevados. Rev CEFAC. 2003;5(4)353-9..

As for the changes reported by patients examined by magnetic resonance imaging of the skull, through the provided questionnaire after the examination, it was found that the noise generated by the MRI machine can cause hearing and / or temporary psychosocial changes. However, it should be considered that any invasive procedure or not in which the patient should stand still, lying on a stretcher that slides inside a chamber (magnet) where there will be emission of sounds at different intensities with an average duration of 32 minutes generates discomfort and anxiety to the patient, regardless of the length of the test of time and the noise intensity.

Although the main intervention in the prevention of hearing, physical and / or psychosocial changes caused by noise should be primary, that is, act in suppressing the responsible agent, it can be used as a specific strategy, hearing protection for noise control, since its function covers the reduction of SPL that reach the individual cochlea1414. Basso BV, Campos LA, Thiesen J. Estudo da alteração temporária do limiar auditivo em trabalhadores expostos ao ruído. Rev CEFAC. 2003;5(2):187-91.),(2020. Vansin R, Ferreira AJ. Avaliação da alteração temporária do limiar auditivo em jovens frequentadores de casas noturnas. Rev CEFAC 2002;(4):219-22.)-(2222. Fantazzini ML. Controle de exposição ao ruído através do uso de protetores auriculares: verificação da proteção oferecida por meio de índice de proteção acústica. Rev Bras Saúde Ocup. 1989;(17):3-50..

The noise reduction mechanism for the use of hearing protection works as a barrier which reduces sound waves that come by air to the tympanic membrane2020. Vansin R, Ferreira AJ. Avaliação da alteração temporária do limiar auditivo em jovens frequentadores de casas noturnas. Rev CEFAC 2002;(4):219-22.)-(2222. Fantazzini ML. Controle de exposição ao ruído através do uso de protetores auriculares: verificação da proteção oferecida por meio de índice de proteção acústica. Rev Bras Saúde Ocup. 1989;(17):3-50.. This attenuation has an average of 31.3 dB SPL. It is worth mentioning that in ideal conditions of use, the hearing protector which acts by air does not attenuate more than 40-50 dB at more treble frequencies. However, this performance will depend on the involvement of the physical characteristics of the protector and its relationship to the ear as well as the anatomical and physiological limitations of the user2121. Avagliano A, Almeida K. Estudo do desempenho de diferentes tipos de protetores auditivos. Rev CEFAC. 2001;3(1):77-87.)-(2323. Ferreira PA, Peloggia SCC, Pirinoto FH, Bernardi APA. Influência da conservação do protetor auricular na prevenção da perda auditiva induzida por níveis de pressão sonora elevados. Rev CEFAC. 2003;5(4)353-9..

Thus, it is understood that the most suitable hearing protection for use while performing the MRI scan are the precast plug and molded plug models. The hearing protector of precast plug type is available in standard sizes adapting to the ear canal of several people. It is made of flexible material such as vinyl and silicon. The molded plug model adapts to the individual's conduct by its use, it is modeled by the user directly into the ear canal. This model can also be manufactured in the following materials: cotton, silicone or foam. It is imperative that hearing protectors used during the examination of magnetic resonance imaging of the skull do not contain any metal tool, as this material has attraction to the generated magnetic field during the examination, causing interference in carrying it out1515. Manzoni TCRC. Saúde auditiva no trabalho: Um estudo sobre a perda auditiva ocupacional e sua prevenção. Rev CEFAC. 2000;2(2):7-14.),(2424. Kwitko A, Pezzi RG, Moreira AFS. Ruído industrial: perda auditiva temporária e condutas para conservação da audição. Rev AMRIGS. 1992;36:135-47..

Applying the average value of provided attenuation by hearing protection, it turns out that the intensity of noise in SPL would reduce to 41.61 dB SPL the final result of 72.91 dB SPL, with the use of this equipment during the examination. Thus, the subjective noise level according to the average intensity values would run through the moderate level to the quiet level. Therefore auditory symptoms reported by patients before the application of the questionnaire would be minimized, since the SPL in dB would decay with the intensity of the use of hearing protection.

Conclusion

Considering the obtained results in this study, by measuring the SPL emitted by the device that performs the scan by magnetic resonance image of the skull, and as for the symptoms reported by patients subjected to such examination, it is relevant the use of hearing protectors while performing the MRI scan since it will promote the reduction of SPL that reach the cochlea of the patient. In this way, both the auditory symptoms as the physical and psychosocial symptoms from noise exposure, would be eliminated and / or minimized resulting improvement in the diagnosis quality of the examination and the patient's proprioceptive during its realization as the final result found does not offer risk of irreversible damage to the auditory system.

Referências

  • 1
    Lufkin RB. Manual de Ressonância Magnética. 2 ed. Rio de Janeiro: Guanabara Koogan; 1999.
  • 2
    Koch HA, Ribeiro ECO, Tonomura ET. Radiologia na formação do médico geral. In: Koch HA, Ribeiro ECO, Tonomura ET. Ressonância magnética. Rio de Janeiro: Revinter; 1997. p. 43-50.
  • 3
    Willian RS. Sistemas de imagens digitais com aplicabilidade na Odontologia [monografia]. Porto Alegre (RS): Universidade Federal do Rio Grande do Sul. Faculdade de Odontologia, Curso de Especialização em Radiologia Odontológica e Imaginologia; 2011.
  • 4
    Mazzola AA. Ressonância Magnética: princípios de formação da imagem e aplicações em imagem funcional. Rev Bras Fis Med. 2003;3(1):117-29.
  • 5
    McMahon KL, Cowin G, Galloway G. Magnetic resonance imaging: the underlying principles. J Orthop Sports PhysTher. 2011;41(11):806-19.
  • 6
    Foltz WD, Jaffray DA. Principles of Magnetic Resonance Imaging. Radiat Res. 2012;177(4):331-48.
  • 7
    Russo PCI. Noções gerais de acústica e psicoacústica. In: Nudelmann AA, Costa AE, Seligman J, Ibañes NR, editores. PAIR: Perda auditiva induzida pelo ruído. Rio de Janeiro: Revinter; 2001. p. 134-40.
  • 8
    Russo PCI. Bases Físicas da Fonação. In: Russo PCI, editor. Acústica e psicoacústica aplicadas à fonoaudiologia. São Paulo: Lovise; 1999. p. 143-78.
  • 9
    Almeida SIC. Efeitos lesivos do ruído e implicações legais. Acta Awho. 1998; 12(3):121-7.
  • 10
    Kwitko A, Pezzi RG, Silveira MS. Exposição a ruído ocupacional e pressão sangüínea. Rev Bras Otorrinolaringol. 1996;62(2):89-98.
  • 11
    Oliveira JAA. Prevenção e proteção contra perda auditiva induzida por ruído. In: Nudelmann AA, Costa EA, Seligman J, Ibanes RN, organizadores. PAIR Perda Auditiva Induzida por Ruído. Rio de Janeiro: Revinter ;2001. p.17-44.
  • 12
    Araujo SA. Perda auditiva induzida pelo ruído em trabalhadores de metalúrgica. Rev Bras Otorrinolaringol. 2002;68(1):47-52.
  • 13
    Souza MA, Pennacchi SMPL, Silva BCP, Ferreira AJV. Prevalência do zumbido como sintoma da perda auditiva induzida por ruído ocupacional. Rev CEFAC. 2002; 4(3):195-8.
  • 14
    Basso BV, Campos LA, Thiesen J. Estudo da alteração temporária do limiar auditivo em trabalhadores expostos ao ruído. Rev CEFAC. 2003;5(2):187-91.
  • 15
    Manzoni TCRC. Saúde auditiva no trabalho: Um estudo sobre a perda auditiva ocupacional e sua prevenção. Rev CEFAC. 2000;2(2):7-14.
  • 16
    Pires MGS, Rodrigues PH, Sampaio CCC, Rodrigues CG. Medida do nível de pressão sonora em um centro urbano. J Bras Fonoaudiol. 2002;3(13):263-6.
  • 17
    Marchiori MLL, Melo JJ. Comparação das queixas auditivas com relação à exposição ao ruído em componentes de orquestra sinfônica. Pró-fono R Atual Cient. 2001;13(1):9-12.
  • 18
    Nudelmann AA, Costa AE, Seligman J, Ibañez NR. PAIR- Perda auditiva induzida pelo ruído. 2 ed. Rio de Janeiro: Revinter; 2001.
  • 19
    Dreossi FCR, Santos MMT. Investigação sobre o efeito do ruído na inteligibilidade de fala de crianças da quarta série do ensino fundamental. Rev Soc Bras Fonoaudiol. 2004;9(3):136-43.
  • 20
    Vansin R, Ferreira AJ. Avaliação da alteração temporária do limiar auditivo em jovens frequentadores de casas noturnas. Rev CEFAC 2002;(4):219-22.
  • 21
    Avagliano A, Almeida K. Estudo do desempenho de diferentes tipos de protetores auditivos. Rev CEFAC. 2001;3(1):77-87.
  • 22
    Fantazzini ML. Controle de exposição ao ruído através do uso de protetores auriculares: verificação da proteção oferecida por meio de índice de proteção acústica. Rev Bras Saúde Ocup. 1989;(17):3-50.
  • 23. Ferreira PA, Peloggia SCC, Pirinoto FH, Bernardi APA. Influência da conservação do protetor auricular na prevenção da perda auditiva induzida por níveis de pressão sonora elevados. Rev CEFAC. 2003;5(4)353-9.
  • 24
    Kwitko A, Pezzi RG, Moreira AFS. Ruído industrial: perda auditiva temporária e condutas para conservação da audição. Rev AMRIGS. 1992;36:135-47.
  • Funding: PUC-Minas

Publication Dates

  • Publication in this collection
    Jan-Feb 2016

History

  • Received
    25 May 2015
  • Accepted
    12 Oct 2015
ABRAMO Associação Brasileira de Motricidade Orofacial Rua Uruguaiana, 516, Cep 13026-001 Campinas SP Brasil, Tel.: +55 19 3254-0342 - São Paulo - SP - Brazil
E-mail: revistacefac@cefac.br